We report the incidence of radiolucent lines (RLLs) using two
flanged acetabular components at total hip arthroplasty (THA) and
the effect of the Rim Cutter. We performed a retrospective review of 300 hips in 292 patients
who underwent primary cemented THA. A contemporary flanged acetabular
component was used with (group 1) and without (group 2) the use
of the Rim Cutter and the Rimfit acetabular component was used with
the Rim Cutter (group 3). RLLs and clinical outcomes were evaluated
immediately post-operatively and at five years post-operatively.Aims
Patients and Methods
This study aimed to characterise and qualitatively grade the severity of the corrosion particles released into the hip joint following taper corrosion. The 26 cases examined were CoC/ABG Modular (n = 13) and ASR/SROM (n = 13). Blood serum metal ion levels were collected before and after revision surgery. The haematoxylin and eosin tissue sections were graded on the presence of fibrin exudates, necrosis, inflammatory cells and corrosion products. The corrosion products were identified based on visible observation and graded on abundance. Two independent observers blinded to the clinical patient findings scored all cases. Elemental analysis was performed on corrosion products within tissue sections. X-Ray diffraction was used to identify crystalline structures present in taper debris.Objectives
Methods
Our aim was to report survivorship data and lessons learned with
the Corail/Pinnacle cementless total hip arthroplasty (THA) system. Between August 2005 and March 2015, a total of 4802 primary cementless
Corail/Pinnacle THAs were performed in 4309 patients. In March 2016,
we reviewed these hips from a prospectively maintained database.Aims
Patients and Methods
We sought to determine whether cobalt-chromium alloy (CoCr) femoral
stem tapers (trunnions) wear more than titanium (Ti) alloy stem
tapers (trunnions) when used in a large diameter (LD) metal-on-metal
(MoM) hip arthroplasty system. We performed explant analysis using validated methodology to
determine the volumetric material loss at the taper surfaces of
explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy
(n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs
with a rough male taper surface and a nominal included angle close
to 5.666° were included. Multiple regression modelling was undertaken
using taper angle, taper roughness, bearing diameter (horizontal
lever arm) as independent variables. Material loss was mapped using
a coordinate measuring machine, profilometry and scanning electron
microscopy.Aims
Patients and Methods
We performed a retrospective study to determine
the effect of osteoporosis on the functional outcome of osteoporotic
distal radial fractures treated with a volar locking plate. Between
2009 and 2012 a total of 90 postmenopausal women with an unstable
fracture of the distal radius treated with a volar locking plate
were studied. Changes in the radiological parameters of 51 patients
with osteoporosis (group 1, mean age 66.9, mean T-score –3.16 ( We found that osteoporosis does not have a negative effect on
the functional outcome and additional analysis did not show a correlation
between T-score and outcome. Cite this article:
In the time since Letournel popularised the surgical
treatment of acetabular fractures, more than 25 years ago, there
have been many changes within the field, related to patients, surgical
technique, implants and post-operative care. However, the long-term
outcomes appear largely unchanged. Does this represent stasis or
have the advances been mitigated by other negative factors? In this
article we have attempted to document the recent changes within
the surgery of patients with a fracture involving the acetabulum,
outline contemporary management, and identify the major problem
areas where further research is most needed. Cite this article:
Our aim was to evaluate the radiographic characteristics of patients
undergoing total hip arthroplasty (THA) for the potential of posterior
bony impingement using CT simulations. Virtual CT data from 112 patients who underwent THA were analysed.
There were 40 men and 72 women. Their mean age was 59.1 years (41
to 76). Associations between radiographic characteristics and posterior
bony impingement and the range of external rotation of the hip were
evaluated. In addition, we investigated the effects of pelvic tilt
and the neck/shaft angle and femoral offset on posterior bony impingement.Aims
Patients and Methods
Inflammation of the retrocalcaneal bursa (RB) is a common clinical problem, particularly in professional athletes. RB inflammation is often treated with corticosteroid injections however a number of reports suggest an increased risk of Achilles tendon (AT) rupture. The aim of this cadaveric study was to describe the anatomical connections of the RB and to investigate whether it is possible for fluid to move from the RB into AT tissue. A total of 20 fresh-frozen AT specimens were used. In ten specimens, ink was injected into the RB. The remaining ten specimens were split into two groups to be injected with radiological contrast medium into the RB either with or without ultrasonography guidance (USG).Objectives
Methods
A number of studies have reported satisfactory
results from the isolated revision of an acetabular component. However,
many of these studies reported only the short- to intermediate-term
results of heterogeneous bearing surfaces in a mixed age group. We present our experience of using a ceramic-on-ceramic (CoC)
bearing for isolated revision of an uncemented acetabular component
in 166 patients (187 hips) who were under the age of 50 years at
the time of revision. There were 78 men and 88 women with a mean
age of 47.4 years (28 to 49). The most common reason for revision
was polyethylene wear and acetabular osteolysis in 123 hips (66%),
followed by aseptic loosening in 49 hips (26%). We report the clinical and radiological outcome, complication
rate, and survivorship of this group. The mean duration of follow-up
was 15.6 years (11 to 19). The mean pre-operative Harris hip score was 33 points (1 to 58),
and improved to a mean of 88 points (51 to 100) at follow-up. The
mean pre-operative total Western Ontario and McMaster Universities
Osteoarthritis Index score was 63.2 (43 to 91) and improved to 19.8
points (9 to 61) post-operatively. Overall, 153 of 166 patients
(92%) were satisfied with their outcome. Kaplan–Meier survivorship
analysis, with revision or radiological evidence of implant failure
(13 patients, 8%) as end-points, was 92% at 15 years (95% confidence
interval 0.89 to 0.97). Isolated revision of a cementless acetabular component using
a CoC bearing gives good results in patients under 50 years of age. Cite this article:
The mainstay of surgical correction of hallux valgus is first
metatarsal osteotomy, either proximally or distally. We present
a technique of combining a distal chevron osteotomy with a proximal
opening wedge osteotomy, for the correction of moderate to severe
hallux valgus. We reviewed 45 patients (49 feet) who had undergone double osteotomy.
Outcome was assessed using the American Orthopaedic Foot and Ankle
Society (AOFAS) and the Short Form (SF) -36 Health Survey scores.
Radiological measurements were undertaken to assess the correction. The mean age of the patients was 60.8 years (44.2 to 75.3). The
mean follow-up was 35.4 months (24 to 51).Aims
Patients and Methods
We sought to establish whether an oxidised zirconium (OxZr) femoral
component causes less loss of polyethylene volume than a cobalt
alloy (CoCr) femoral component in total knee arthroplasty. A total of 20 retrieved tibial inserts that had articulated with
OxZr components were matched with 20 inserts from CoCr articulations
for patient age, body mass index, length of implantation, and revision
diagnosis. Changes in dimensions of the articular surfaces were compared
with those of pristine inserts using laser scanning. The differences
in volume between the retrieved and pristine surfaces of the two
groups were calculated and compared.Aims
Materials and Methods
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8o (28o to
63o), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of radiolucent lines and the higher than
expected levels of metal ions in the blood, we would not recommend
the use of CoM THA without further long-term follow-up. We plan
to monitor all these patients regularly. Cite this article:
There are many guidelines that help direct the management of
patients with metal-on-metal (MOM) hip arthroplasties. We have undertaken
a study to compare the management of patients with MOM hip arthroplasties in
different countries. Six international tertiary referral orthopaedic centres were
invited to participate by organising a multi-disciplinary team (MDT)
meeting, consisting of two or more revision hip arthroplasty surgeons
and a musculoskeletal radiologist. A full clinical dataset including
history, blood tests and imaging for ten patients was sent to each
unit, for discussion and treatment planning. Differences in the
interpretation of findings, management decisions and rationale for
decisions were compared using quantitative and qualitative methods.Aims
Methods
We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised adverse local tissue reaction due to corrosion at the
trunnion and pseudotumour formation. The diagnosis was not suspected in nine of the ten patients (six
female/four male; mean age 66 years), despite treatment in a specialist
unit (mean time from index surgery to revision was 58 months, 36
to 84). It was identified at revision surgery and subsequently confirmed
by histological examination of resected tissue. Pre-operative assessment
and culture results ruled out infection. A variety of treatment
strategies were used, including resection of the pseudotumour and
efforts to avoid recurrent dislocation. Aims
Methods
The aim of this study was to compare the post-operative radiographic
and clinical outcomes between kinematically and mechanically aligned
total knee arthroplasties (TKAs). A total of 60 TKAs (30 kinematically and 30 mechanically aligned)
were performed in 60 patients with varus osteoarthritis of the knee
using a navigation system. The angles of orientation of the joint
line in relation to the floor, the conventional and true mechanical
axis (tMA) (the line from the centre of the hip to the lowest point
of the calcaneus) were compared, one year post-operatively, on single-leg
and double-leg standing long leg radiographs between the groups.
The range of movement and 2011 Knee Society Scores were also compared
between the groups at that time.Aims
Patients and Methods
The orientation of the acetabular component is
influenced not only by the orientation at which the surgeon implants
the component, but also the orientation of the pelvis at the time
of implantation. Hence, the orientation of the pelvis at set-up
and its movement during the operation, are important. During 67
hip replacements, using a validated photogrammetric technique, we
measured how three surgeons orientated the patient’s pelvis, how
much the pelvis moved during surgery, and what effect these had
on the final orientation of the acetabular component. Pelvic orientation
at set-up, varied widely (mean (± 2, standard deviation ( Cite this article:
Despite a lack of long-term follow-up, there
is an increasing trend towards using femoral heads of large diameter
in total hip replacement (THR), partly because of the perceived
advantage of lower rates of dislocation. However, increasing the
size of the femoral head is not the only way to reduce the rate
of dislocation; optimal alignment of the components and repair of
the posterior capsule could achieve a similar effect. In this prospective study of 512 cemented unilateral THRs (Male:Female
230:282) performed between 2004 and 2011, we aimed to determine
the rate of dislocation in patients who received a 22 mm head on
a 9/10 Morse taper through a posterior approach with capsular repair
and using the transverse acetabular ligament (TAL) as a guide for the
alignment of the acetabular component. The mean age of the patients
at operation was 67 years (35 to 89). The mean follow-up was 2.8
years (0.5 to 6.6). Pre- and post-operative assessment included
Oxford hip, Short Form-12 and modified University of California
Los Angeles and Merle D’Aubigne scores. The angles of inclination
and anteversion of the acetabular components were measured using
radiological software. There were four dislocations (0.78%), all
of which were anterior. In conclusion, THR with a 22 mm diameter head performed through
a posterior approach with capsular repair and using the TAL as a
guide for the alignment of the acetabular component was associated
with a low rate of dislocation. Cite this article: