Advertisement for orthosearch.org.uk
Results 321 - 340 of 493
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 94 - 99
1 Jan 2007
Bottner F Wegner A Winkelmann W Becker K Erren M Götze C

This prospective study evaluates the role of new laboratory markers in the diagnosis of deep implant infection in 78 patients (41 men and 37 women) with a revision total knee or hip replacement.

The mean age at the time of operation was 64.0 years (19 to 90). Intra-operative cultures showed that 21 patients had a septic and 57 an aseptic total joint replacement. The white blood cell count, the erythrocyte sedimentation rate and levels of C-reactive protein, interleukin-6, procalcitonin and tumour necrosis factor (TNF)-α were measured in blood samples before operation. The diagnostic cut-off values were determined by Received Operating Characteristic curve analysis.

C-reactive protein (> 3.2 md/dl) and interleukin-6 (> 12 pg/ml) have the highest sensitivity (0.95). Interleukin-6 is less specific than C-reactive protein (0.87 vs 0.96). Combining C-reactive protein and interleukin-6 identifies all patients with deep infection of the implant. Procalcitonin (> 0.3 ng/ml) and TNF-α (> 40 ng/ml) are very specific (0.98 vs 0.94) but have a low sensitivity (0.33 vs 0.43).

The combination of C-reactive protein and interleukin-6 measurement provide excellent screening tests for infection of a deep implant. A highly specific marker such as procalcitonin and pre-operative aspiration of the joint might be useful in identifying patients with true positive C-reactive protein and/or interleukin-6 levels.


The Bone & Joint Journal
Vol. 97-B, Issue 10 | Pages 1338 - 1344
1 Oct 2015
te Stroet MAJ Keurentjes JC Rijnen WHC Gardeniers JWM Verdonschot N Slooff TJJH Schreurs BW

We present the results of 62 consecutive acetabular revisions using impaction bone grafting and a cemented polyethylene acetabular component in 58 patients (13 men and 45 women) after a mean follow-up of 27 years (25 to 30). All patients were prospectively followed. The mean age at revision was 59.2 years (23 to 82).

We performed Kaplan–Meier (KM) analysis and also a Competing Risk (CR) analysis because with long-term follow-up, the presence of a competing event (i.e. death) prevents the occurrence of the endpoint of re-revision.

A total of 48 patients (52 hips) had died or had been re-revised at final review in March 2011. None of the deaths were related to the surgery. The mean Harris hip score of the ten surviving hips in ten patients was 76 points (45 to 99).

The KM survivorship at 25 years for the endpoint ‘re-revision for any reason’ was 58.0% (95% confidence interval (CI) 38 to 73) and for ‘re-revision for aseptic loosening’ 72.1% (95% CI 51 to 85). With the CR analysis we calculated the KM analysis overestimates the failure rate with respectively 74% and 93% for these endpoints. The current study shows that acetabular impaction bone grafting revisions provide good clinical results at over 25 years.

Cite this article: Bone Joint J 2015;97-B:1338–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 11 - 13
1 Nov 2012
Cuckler JM

Hip implant retrieval analysis is the most important source of insight into the performance of new materials and designs of hip arthroplasties. Even the most rigorous in vitro testing will not accurately simulate the behavior of implant materials and new designs of prosthetic arthroplasties. Retrieval analysis has revealed such factors as the effects of gamma-in-air sterilisation of polyethylene, fatigue failure mechanisms of polymethylmethacrylate bone cement, fretting corrosion of Morse taper junctions, third body wear effects of both hard-on-hard and hard-on-soft bearing couples, and the effects of impingement of components on the full spectrum of bearing surfaces, none of which was predicted by pre-implantation in vitro testing of these materials and combinations. The temporal sequence of the retrieval process is approximately six years from first implantation through retrieval analysis, laboratory investigation, and publication of results, and thus, in addition to rigorous clinical evaluation, represents the true development and insight cycle for new designs and materials.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 535 - 537
1 Apr 2007
Evangelista GT Fulkerson E Kummer F Di Cesare PE

During open reduction of an irreducible anterior dislocation of a total hip replacement with an Oxinium femoral head, it was observed that the head had been significantly damaged. Gross and scanning electron microscopic examination revealed cracking, gouging, and delamination of the surface. Because of the risk which this poses for damaging the polyethylene acetabular liner, it is strongly recommended that patients with this type of prosthetic head be carefully monitored after a dislocation.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1493 - 1499
1 Nov 2015
Pandit H Hamilton TW Jenkins C Mellon SJ Dodd CAF Murray DW

This prospective study reports the 15-year survival and ten-year functional outcome of a consecutive series of 1000 minimally invasive Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women, 52%, mean age 66 years; 32 to 88). These were implanted by two surgeons involved with the design of the prosthesis to treat anteromedial osteoarthritis and spontaneous osteonecrosis of the knee, which are recommended indications. Patients were prospectively identified and followed up independently for a mean of 10.3 years (5.3 to 16.6).

At ten years, the mean Oxford Knee Score was 40 (standard deviation (sd) 9; 2 to 48): 79% of knees (349) had an excellent or good outcome. There were 52 implant-related re-operations at a mean of 5.5 years (0.2 to 14.7). The most common reasons for re-operation were arthritis in the lateral compartment (2.5%, 25 knees), bearing dislocation (0.7%, seven knees) and unexplained pain (0.7%, seven knees). When all implant-related re-operations were considered as failures, the ten-year rate of survival was 94% (95% confidence interval (CI) 92 to 96) and the 15-year survival rate 91% (CI 83 to 98). When failure of the implant was the endpoint the 15-year survival was 99% (CI 96 to 100).

This is the only large series of minimally invasive UKAs with 15-year survival data. The results support the continued use of minimally invasive UKA for the recommended indications.

Cite this article: Bone Joint J 2015;97-B:1493–99.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 2 | Pages 196 - 201
1 Feb 2010
Hallan G Dybvik E Furnes O Havelin LI

The Norwegian Arthroplasty Register has shown that several designs of uncemented femoral stems give good or excellent survivorship. The overall findings for uncemented total hip replacement however, have been disappointing because of poor results with the use of metal-backed acetabular components. In this study, we exclusively investigated the medium-to long-term performance of primary uncemented metal-backed acetabular components.

A total of 9113 primary uncemented acetabular components were implanted in 7937 patients between 1987 and 2007. These were included in a prospective, population-based observational study. All the implants were modular and metal-backed with ultra-high-molecular-weight polyethylene liners. The femoral heads were made of stainless steel, cobalt-chrome (CoCr) alloy or alumina ceramic. In all, seven different designs of acetabular component were evaluated by the Kaplan-Meier survivorship method and Cox regression analysis.

Most acetabular components performed well up to seven years. When the endpoint was revision of the acetabular component because of aseptic loosening, the survival ranged between 87% and 100% at ten years. However, when the endpoint was revision for any reason, the survival estimates were 81% to 92% for the same implants at ten years. Aseptic loosening, wear, osteolysis and dislocation were the main reasons for the relatively poor overall performance of the acetabular components. Prostheses with alumina heads performed slightly better than those with stainless steel or CoCr alloy in subgroups.

Whereas most acetabular components performed well at seven years, the survivorship declined with longer follow-up. Fixation was generally good. None of the metal-backed uncemented acetabular components with ultra-high-molecular-weight polyethylene liners in our study had satisfactory long-term results because of high rates of wear, osteolysis, aseptic loosening and dislocation.


Bone & Joint 360
Vol. 4, Issue 3 | Pages 12 - 14
1 Jun 2015

The June 2015 Knee Roundup360 looks at: Cruciate substituting versus retaining knee replacement; What’s behind the psychology of anterior cruciate ligament (ACL) reconstruction?; Is there a difference in total knee arthroplasty risk of revision in highly crosslinked versus conventional polyethylene?; Unicompartmental knee arthroplasty: is age the missing variable?; Satisfaction rates following total knee arthroplasty; Is knee alignment dynamic?; Unicompartmental knee arthroplasty: cemented or cementless?; Can revision knee services pay?


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1641 - 1648
1 Dec 2012
Baker PN Jameson SS Deehan DJ Gregg PJ Porter M Tucker K

Current analysis of unicondylar knee replacements (UKRs) by national registries is based on the pooled results of medial and lateral implants. Consequently, little is known about the differential performance of medial and lateral replacements and the influence of each implant type within these pooled analyses. Using data from the National Joint Registry for England and Wales (NJR) we aimed to determine the proportion of UKRs implanted on the lateral side of the knee, and their survival and reason for failure compared with medial UKRs. By combining information on the side of operation with component details held on the NJR, we were able to determine implant laterality (medial versus lateral) for 32 847 of the 35 624 unicondylar replacements (92%) registered before December 2010. Of these, 2052 (6%) were inserted on the lateral side of the knee. The rates of survival at five years were 93.1% (95% confidence interval (CI) 92.7 to 93.5) for medial and 93.0% (95% CI 91.1 to 94.9) for lateral UKRs (p = 0.49). The rates of failure remained equivalent after adjusting for patient age, gender, American Society of Anesthesiologists (ASA) grade, indication for surgery and implant design using Cox’s proportional hazards method (hazard ratio for lateral relative to medial replacement = 0.88 (95% CI 0.69 to 1.13); p = 0.32). Aseptic loosening/lysis and unexplained pain were the main reasons for revision in both groups, although the reasons did vary depending on whether a mobile- or a fixed-bearing design was used. At a maximum of eight years the mid-term survival rates of medial and lateral UKRs are similar.


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 611 - 616
1 May 2015
Shin WC Lee SM Lee KW Cho HJ Lee JS Suh KT

There is no single standardised method of measuring the orientation of the acetabular component on plain radiographs after total hip arthroplasty. We assessed the reliability and accuracy of three methods of assessing anteversion of the acetabular component for 551 THAs using the PolyWare software and the methods of Liaw et al, and of Woo and Morrey. All measurements of the three methods had excellent intra- and inter-observer reliability. The values of the PolyWare software, which determines version of the acetabular component by edge detection were regarded as the reference standard. Although the PolyWare software and the method of Liaw et al were similarly precise, the method of Woo and Morrey was significantly less accurate (p < 0.001). The method of Liaw et al seemed to be more accurate than that of Woo and Morrey when compared with the measurements using the PolyWare software. If the qualified lateral radiograph was selected, anteversion measured using the method of Woo and Morrey was considered to be relatively reliable.

Cite this article: Bone Joint J 2015; 97-B:611–16.


The Bone & Joint Journal
Vol. 97-B, Issue 9 | Pages 1183 - 1191
1 Sep 2015
Briggs TWR Hanna SA Kayani B Tai S Pollock RC Cannon SR Blunn GW Carrington RWJ

The long term biological effects of wear products following total hip arthroplasty (THA) are unclear. However, the indications for THA are expanding, with increasingly younger patients undergoing the procedure.

This prospective, randomised study compared two groups of patients undergoing THA after being randomised to receive one of two different bearing surfaces: metal-on-polyethylene (MoP) n = 22 and metal-on-metal (MoM) n = 23. We investigated the relationship between three variables: bearing surface (MoP vs MoM), whole blood levels of chromium (Cr) and cobalt (Co) and chromosomal aberrations in peripheral lymphocyte pre-operatively and at one, two and five years post-surgery.

Our results demonstrated significantly higher mean cobalt and chromium (Co and Cr) blood levels in the MoM group at all follow-up points following surgery (p < 0.01), but there were no significant differences in the chromosomal aberration indices between MoM and MoP at two or five years (two years: p = 0.56, p = 0.08, p = 0.91, p = 0.51 and five years: p = 0.086, p = 0.73, p = 0.06, p = 0.34) for translocations, breaks, loss and gain of chromosomes respectively. Regression analysis showed a strong linear relationship between Cr levels and the total chromosomal aberration indices in the MoM group (R2 = 0.90016), but this was not as strong for Co (R2 = 0.68991). In the MoP group, the analysis revealed a poor relationship between Cr levels and the total chromosomal aberration indices (R2 = 0.23908) but a slightly stronger relationship for Co (R2 = 0.64292). Across both groups, Spearman’s correlation detected no overall association between Co and Cr levels and each of the studied chromosomal aberrations. There remains no clear indication which THA bearing couple is the most biocompatible, especially in young active patients. While THA continues to be very successful at alleviating pain and restoring function, the long-term biological implications of the procedure still require further scrutiny.

Cite this article: Bone Joint J 2015;97-B:1183–91.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 911 - 916
1 Jul 2015
Del Balso C Teeter MG Tan SC Lanting BA Howard JL

Tribocorrosion at the head–neck taper interface – so-called ‘taperosis’ – may be a source of metal ions and particulate debris in metal-on-polyethylene total hip arthroplasty (THA).

We examined the effect of femoral head length on fretting and corrosion in retrieved head–neck tapers in vivo for a minimum of two years (mean 8.7 years; 2.6 to 15.9). A total of 56 femoral heads ranging from 28 mm to 3 mm to 28 mm + 8 mm, and 17 femoral stems featuring a single taper design were included in the study. Fretting and corrosion were scored in three horizontally oriented concentric zones of each taper by stereomicroscopy.

Head length was observed to affect fretting (p = 0.03), with 28 mm + 8 mm femoral heads showing greater total fretting scores than all other head lengths. The central zone of the femoral head bore taper was subject to increased fretting damage (p = 0.01), regardless of head length or stem offset. High-offset femoral stems were associated with greater total fretting of the bore taper (p = 0.04).

Increased fretting damage is seen with longer head lengths and high-offset femoral stems, and occurs within a central concentric zone of the femoral head bore taper. Further investigation is required to determine the effect of increased head size, and variations in head–neck taper design.

Cite this article: Bone Joint J 2015; 97-B:911–16.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 84 - 86
1 Nov 2014
Russell RD Huo MH Jones RE

Patellofemoral complications are common after total knee replacement (TKR). Leaving the patellar unsurfaced after TKR may lead to complications such as anterior knee pain, and re-operation to surface it. Complications after patellar resurfacing include patellar fracture, aseptic loosening, patellar instability, polyethylene wear, patellar clunk and osteonecrosis. Historically, patellar complications account for one of the larger proportions of causes of failure in TKR, however, with contemporary implant designs, complication rates have decreased. Most remaining failures relate to patellofemoral tracking. Understanding the causes of patellofemoral maltracking is essential to prevent these complications as well as manage them when they occur.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):84–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 859 - 864
1 Jul 2009
Gwynne-Jones DP Garneti N Wainwright C Matheson JA King R

We reviewed the results at nine to 13 years of 125 total hip replacements in 113 patients using the monoblock uncemented Morscher press-fit acetabular component. The mean age at the time of operation was 56.9 years (36 to 74). The mean clinical follow-up was 11 years (9.7 to 13.5) and the mean radiological follow-up was 9.4 years (7.7 to 13.1). Three hips were revised, one immediately for instability, one for excessive wear and one for deep infection.

No revisions were required for aseptic loosening. A total of eight hips (7.0%) had osteolytic lesions greater than 1 cm, in four around the acetabular component (3.5%). One required bone grafting behind a well-fixed implant. The mean wear rate was 0.11 mm/year (0.06 to 0.78) and was significantly higher in components with a steeper abduction angle.

Kaplan-Meier survival curves at 13 years showed survival of 96.8% (95% confidence interval 90.2 to 99.0) for revision for any cause and of 95.7% (95% confidence interval 88.6 to 98.4) for any acetabular re-operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 11 | Pages 1522 - 1526
1 Nov 2010
McGrory BJ McGrory CP Barbour L Barbour B

Dislocation is a common and well-studied complication after total hip replacement. However, subluxation, which we define as a clinically recognised episode of incomplete movement of the femoral head outside the acetabulum with spontaneous reduction, has not been studied previously. Out of a total of 2521 hip replacements performed over 12 years by one surgeon, 30 patients experienced subluxations which occurred in 31 arthroplasties. Data were collected prospectively with a minimum follow-up of two years. Subluxation occurred significantly more frequently after revision than after primary hip replacement, and resolved in 19 of 31 cases (61.3%). In six of the 31 hips (19.4%) the patient subsequently dislocated the affected hip, and in six hips (19.4%) intermittent subluxation continued. Four patients had a revision operation for instability, three for recurrent dislocation and one for recurrent subluxation. Clinical and radiological comparisons with a matched group of stable total hips showed no correlation with demographic or radiological parameters.

Patients with subluxing hips reported significantly more concern that their hip would dislocate, more often changed their behaviour to prevent instability and had lower postoperative Harris hip scores than patients with stable replacements.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 45 - 49
1 Jan 2015
Tokarski AT Novack TA Parvizi J

We hypothesised that the use of tantalum (Ta) acetabular components in revision total hip arthroplasty (THA) was protective against subsequent failure due to infection. We identified 966 patients (421 men, 545 women and 990 hips) who had undergone revision THA between 2000 and 2013. The mean follow up was 40.2 months (3 months to 13.1 years). The mean age of the men and women was 62.3 years (31 to 90) and 65.1 years (25 to 92), respectively.

Titanium (Ti) acetabular components were used in 536 hips while Ta components were used in 454 hips. In total, 73 (7.3%) hips experienced subsequent acetabular failure. The incidence of failure was lower in the Ta group at 4.4% (20/454) compared with 9.9% (53/536) in the Ti group (p < 0.001, odds ratio 2.38; 95% CI 1.37 to 4.27). Among the 144 hips (64 Ta, 80 Ti) for which revision had been performed because of infection, failure due to a subsequent infection was lower in the Ta group at 3.1% (2/64) compared with 17.5% (14/80) for the Ti group (p = 0.006).

Thus, the use of Ta acetabular components during revision THA was associated with a lower incidence of failure from all causes and Ta components were associated with a lower incidence of subsequent infection when used in patients with periprosthetic joint infection.

Cite this article: Bone Joint J 2015;97-B:45–9.


Bone & Joint Research
Vol. 4, Issue 1 | Pages 6 - 10
1 Jan 2015
Goudie ST Deakin AH Deep K

Objectives

Acetabular component orientation in total hip arthroplasty (THA) influences results. Intra-operatively, the natural arthritic acetabulum is often used as a reference to position the acetabular component. Detailed information regarding its orientation is therefore essential. The aim of this study was to identify the acetabular inclination and anteversion in arthritic hips.

Methods

Acetabular inclination and anteversion in 65 symptomatic arthritic hips requiring THA were measured using a computer navigation system. All patients were Caucasian with primary osteoarthritis (29 men, 36 women). The mean age was 68 years (SD 8). Mean inclination was 50.5° (SD 7.8) in men and 52.1° (SD 6.7) in women. Mean anteversion was 8.3° (SD 8.7) in men and 14.4° (SD 11.6) in women.


The Bone & Joint Journal
Vol. 96-B, Issue 7 | Pages 855 - 856
1 Jul 2014
Haddad FS Oussedik S


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11 | Pages 1540 - 1545
1 Nov 2012
Reigstad O Lütken T Grimsgaard C Bolstad B Thorkildsen R Røkkum M

The Motec cementless modular metal-on-metal ball-and-socket wrist arthroplasty was implanted in 16 wrists with scaphoid nonunion advanced collapse (SNAC; grades 3 or 4) and 14 wrists with scapholunate advanced collapse (SLAC) in 30 patients (20 men) with severe (grades 3 or 4) post-traumatic osteoarthritis of the wrist. The mean age of the patients was 52 years (31 to 71). All prostheses integrated well radiologically. At a mean follow-up of 3.2 years (1.1 to 6.1) no luxation or implant breakage occurred. Two wrists were converted to an arthrodesis for persistent pain. Loosening occurred in one further wrist at five years post-operatively. The remainder demonstrated close bone–implant contact. The clinical results were good, with markedly decreased Disabilities of the Arm Shoulder and Hand (DASH) and pain scores, and increased movement and grip strength. No patient used analgesics and most had returned to work.

Good short-term function was achieved using this wrist arthroplasty in a high-demand group of patients with post-traumatic osteoarthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 4 | Pages 451 - 458
1 Apr 2009
Hooper GJ Rothwell AG Stringer M Frampton C

We have reviewed the rate of revision of fully cemented, hybrid and uncemented primary total hip replacements (THRs) registered in the New Zealand Joint Registry between 1999 and December 2006 to determine whether there was any statistically significant difference in the early survival and reason for revision in these different types of fixation. The percentage rate of revision was calculated per 100 component years and compared with the reason for revision, the type of fixation and the age of the patients.

Of the 42 665 primary THRs registered, 920 (2.16%) underwent revision requiring change of at least one component. Fully-cemented THRs had a lower rate of revision when considering all causes for failure (p < 0.001), but below the age of 65 years uncemented THRs had a lower rate (p < 0.01). The rate of revision of the acetabular component for aseptic loosening was less in the uncemented and hybrid groups compared with that in the fully cemented group (p < 0.001), and the rate of revision of cemented and uncemented femoral components was similar, except in patients over 75 years of age in whom revision of cemented femoral components was significantly less frequent (p < 0.02). Revision for infection was more common in patients aged below 65 years and in cemented and hybrid THRs compared with cementless THRs (p < 0.001). Dislocation was the most common cause of revision for all types of fixation and was more frequent in both uncemented acetabular groups (p < 0.001). The experience of the surgeon did not affect the findings.

Although cemented THR had the lowest rate of revision for all causes in the short term (90 days), uncemented THR had the lowest rate of aseptic loosening in patients under 65 years of age and had rates comparable with international rates of aseptic loosening in those over 65 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1215 - 1221
1 Sep 2010
Sakai T Ohzono K Nishii T Miki H Takao M Sugano N

The long-term results of grafting with hydroxyapatite granules for acetabular deficiency in revision total hip replacement are not well known. We have evaluated the results of revision using a modular cup with hydroxyapatite grafting for Paprosky type 2 and 3 acetabular defects at a minimum of ten years’ follow-up. We retrospectively reviewed 49 acetabular revisions at a mean of 135 months (120 to 178). There was one type 2B, ten 2C, 28 3A and ten 3B hips. With loosening as the endpoint, the survival rate was 74.2% (95% confidence interval 58.3 to 90.1). Radiologically, four of the type 3A hips (14%) and six of the type 3B hips (60%) showed aseptic loosening with collapse of the hydroxyapatite layer, whereas no loosening occurred in type 2 hips. There was consolidation of the hydroxyapatite layer in 33 hips (66%). Loosening was detected in nine of 29 hips (31%) without cement and in one of 20 hips (5%) with cement (p = 0.03, Fisher’s exact probability test). The linear wear and annual wear rate did not correlate with loosening.

These results suggest that the long-term results of hydroxyapatite grafting with cement for type 2 and 3A hips are encouraging.