Between 1990 and 2000, 123 hips in 110 patients were reconstructed for aseptic loosening using impaction bone grafting with frozen, irradiated, morsellised femoral heads and cemented acetabular components. This series was reported previously at a mean follow-up of five years. We have extended this follow-up and now describe the outcome of 86 hips in 74 patients at a mean of ten years. There have been 19 revisions, comprising nine for infection, seven for aseptic loosening and three for dislocation. In surviving acetabular reconstructions, union of the graft had occurred in 64 of 67 hips (95.5%). Survival analysis for all indications at ten years was 83.3% (95% confidence interval (CI) 68 to 89) and 71.3% (95% CI 58 to 84) at 15 years. Acetabular reconstruction using irradiated allograft and a cemented acetabular component is an effective method of reconstruction, providing results in the medium- to long-term comparable with those of reported series where non-irradiated freshly-frozen bone was used.
When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article:
The hip joint is commonly involved in multiple epiphyseal dysplasia and patients may require total hip replacement before the age of 30 years. We retrospectively reviewed nine patients (16 hips) from four families. The diagnosis of multiple epiphyseal dysplasia was based on a family history, genetic counselling, clinical features and radiological findings. The mean age at surgery was 32 years (17 to 63), with a mean follow-up of 15.9 years (5.5 to 24). Of the 16 hips, ten required revision at a mean of 12.5 years (5 to 15) consisting of complete revision of the acetabular component in three hips and isolated exchange of the liner in seven. No femoral component has loosened or required revision during the period of follow-up. With revision for any reason, the 15-year survival was only 11.4% (95% confidence interval 1.4 to 21.4). However, when considering revision of the acetabular shell in isolation the survival at ten years was 93.7% (95% confidence interval 87.7 to 99.7), reducing to 76.7% (95% confidence interval 87.7 to 98.7) at 15 and 20 years, respectively.
This was a retrospective analysis of the medium-
to long-term results of 46 TC3 Sigma revision total knee replacements
using long uncemented stems in press-fit mode. Clinical and radiological analysis took place pre-operatively,
at two years post-operatively, and at a mean follow-up of 8.5 years
(4 to 12). The mean pre-operative International Knee Society (IKS)
clinical score was 42 points (0 to 74), improving to 83.7 (52 to
100) by the final follow-up. The mean IKS score for function improved
from 34.3 points (0 to 80) to 64.2 (15 to 100) at the final follow-up.
At the final follow-up 30 knees (65.2%) had an excellent result, seven
(15.2%) a good result, one (2.2%) a medium and eight (17.4%) a poor
result. There were two failures, one with anteroposterior instability
and one with aseptic loosening. The TC3 revision knee system, when used with press-fit for long
intramedullary stems and cemented femoral and tibial components,
in both septic and aseptic revisions, results in a satisfactory
clinical and radiological outcome, and has a good medium- to long-term
survival rate.
This pilot study reports the clinical outcomes
of a combination of partial subscapularis tendon transfer and small-head
hemiarthroplasty in patients with rotatator cuff arthropathy. A
total of 30 patients (30 shoulders; eight men and 22 women) with
a mean age of 74 years (55 to 84) were assessed at a mean follow-up
of 31 months (24 to 60). The inclusion criteria were painful cuff
tear arthropathy with normal deltoid function and a non-degenerative subscapularis
muscle and tendon and a preserved teres minor. Outcome was assessed using the University of California Los Angeles
score, the Japanese Orthopaedic Association score, and the Oxford
Shoulder Score. Radiographic measurements included the centre of
rotation distance and the length of the deltoid. All clinical scores were significantly improved post-operatively.
The active flexion and external rotation improved significantly
at the most recent follow-up (p <
0.035). Although the mean centre
of rotation distance changed significantly (p <
0.001), the mean
length of the deltoid did not change significantly from the pre-operative
value (p = 0.29). The change in the length of the deltoid with <
100° flexion was significantly less than that with >
100° (p <
0.001). Progressive erosion of the glenoid was seen in four patients.
No patient required revision or further surgery. A combination of partial subscapularis tendon transfer and small-head
hemiarthroplasty effectively restored function and relieved pain
in patients with rotator cuff arthropathy. Cite this article: 2015;97-B:1090–5.
We report our experience of performing an elbow
hemiarthroplasty in the treatment of comminuted distal humeral fractures
in the elderly patients. A cohort of 42 patients (three men and 39 women, mean age 72;
56 to 84) were reviewed at a mean of 34.3 months (24 to 61) after
surgery. Functional outcome was measured with the Mayo Elbow Performance
Score (MEPS) and range of movement. The disabilities of the arm,
shoulder and hand questionnaire (DASH) was used as a patient rated
evaluation. Complications and ulnar nerve function were recorded.
Plain radiographs were obtained to assess prosthetic loosening,
olecranon wear and heterotopic bone formation. The mean extension deficit was 23.5° (0° to 60°) and mean flexion
was 126.8° (90° to 145°) giving a mean arc of 105.5° (60° to 145°).
The mean MEPS was 90 (50 to 100) and a mean DASH score of 20 (0
to 63). Four patients had additional surgery for limited range of
movement and one for partial instability. One elbow was revised
due to loosening, two patients had sensory ulnar nerve symptoms,
and radiographic signs of mild olecranon wear was noted in five
patients. Elbow hemiarthroplasty for comminuted intra-articular distal
humeral fractures produces reliable medium-term results with functional
outcome and complication rates, comparable with open reduction and
internal fixation and total elbow arthroplasty. Cite this article:
Partial knee arthroplasty (PKA), either medial
or lateral unicompartmental knee artroplasty (UKA) or patellofemoral arthroplasty
(PFA) are a good option in suitable patients and have the advantages
of reduced operative trauma, preservation of both cruciate ligaments
and bone stock, and restoration of normal kinematics within the
knee joint. However, questions remain concerning long-term survival.
The goal of this review article was to present the long-term results
of medial and lateral UKA, PFA and combined compartmental arthroplasty
for multicompartmental disease. Medium- and long-term studies suggest
reasonable outcomes at ten years with survival greater than 95% in
UKA performed for medial osteoarthritis or osteonecrosis, and similarly
for lateral Cite this article:
Metal-on-metal total hip replacement has been targeted at younger patients with anticipated long-term survival, but the effect of the production of metal ions is a concern because of their possible toxicity to cells. We have reviewed the results of the use of the Ultima hybrid metal-on-metal total hip replacement, with a cemented polished tapered femoral component with a 28 mm diameter and a cobalt-chrome (CoCr) modular head, articulating with a 28 mm CoCr acetabular bearing surface secured in a titanium alloy uncemented shell. Between 1997 and 2004, 545 patients with 652 affected hips underwent replacement using this system. Up to 31 January 2008, 90 (13.8%) hips in 82 patients had been revised. Pain was the sole reason for revision in 44 hips (48.9%) of which 35 had normal plain radiographs. Peri-prosthetic fractures occurred in 17 hips (18.9%) with early dislocation in three (3.3%) and late dislocation in 16 (17.8%). Infection was found in nine hips (10.0%). At operation, a range of changes was noted including cavities containing cloudy fluid under pressure, necrotic soft tissues with avulsed tendons and denuded osteonecrotic upper femora. Corrosion was frequently observed on the retrieved cemented part of the femoral component. Typically, the peri-operative findings confirmed those found on pre-operative metal artefact reduction sequence MRI and histological examination showed severe necrosis. Metal artefact reduction sequence MRI proved to be useful when investigating these patients with pain in the absence of adverse plain radiological features.
Between 2002 and 2008, 130 consecutive ankles were replaced with an hydroxyapatite (HA) and titanium-HA-coated Ankle Evolutive System total ankle prosthesis. Plain radiographs were analysed by two independent observers. Osteolytic lesions were classified by their size and location, with cavities >
10 mm in diameter considered to be ‘marked’. CT scanning was undertaken in all patients with marked osteolysis seen on the plain radiographs. Osteolytic lesions were seen on the plain films in 48 (37%) and marked lesions in 27 (21%) ankles. The risk for osteolysis was found to be 3.1 (95% confidence interval 1.6 to 5.9) times higher with implants with Ti-HA porous coating. Care should be taken with ankle arthroplasty until more is known about the reasons for these severe osteolyses.
We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component. At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
Despite many claims of good wear properties following
total knee replacement (TKR) with an oxidised zirconium (OxZr) femoral
component, there are conflicting clinical results. We hypothesised
that there would be no difference in either the mid-term clinical
and radiological outcomes or the characteristics of the polyethylene
wear particles (weight, size and shape) in patients using an OxZr
or cobalt-chrome (CoCr) femoral component. In all 331 patients underwent
bilateral TKR, receiving an OxZr femoral component in one knee and
a CoCr femoral component in the other. The mean follow-up was 7.5
years (6 to 8). Following aspiration, polyethylene wear particles
were analysed using thermogravimetric methods and scanning electron
microscopy. At the most recent follow-up, the mean Knee Society
score, Western Ontario and McMaster Universities Osteoarthritis
Index score, range of movement and satisfaction score were not significantly
different in the two groups. The mean weight, size, aspect ratio
and roundness of the aspirated wear particles were similar for each
femoral component. Survivorship of the femoral, tibial and patellar
components was 100% in both groups. In the absence of evidence of an advantage in the medium term
we cannot justify the additional expense of an OxZr femoral component.
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.
Primary total knee arthroplasty (TKA) is a reliable
procedure with reproducible long-term results. Nevertheless, there
are conditions related to the type of patient or local conditions
of the knee that can make it a difficult procedure. The most common
scenarios that make it difficult are discussed in this review. These
include patients with many previous operations and incisions, and
those with severe coronal deformities, genu recurvatum, a stiff knee,
extra-articular deformities and those who have previously undergone
osteotomy around the knee and those with chronic dislocation of
the patella. Each condition is analysed according to the characteristics of
the patient, the pre-operative planning and the reported outcomes. When approaching the difficult primary TKA surgeons should use
a systematic approach, which begins with the review of the existing
literature for each specific clinical situation. Cite this article:
Our aim in this prospective study was to evaluate the outcome of total shoulder replacement in the treatment of young and middle-aged active patients with primary glenohumeral osteoarthritis. We reviewed 21 patients (21 shoulders) with a mean age of 55 years (37 to 60). The mean follow-up was seven years (5 to 9). The same anatomical, third-generation, cemented implant had been used in all patients. All the patients were evaluated radiologically and clinically using the Constant and Murley score. No patients required revision. In one a tear of the supraspinatus tendon occurred. Overall, 20 patients (95%) were either very satisfied (n = 18) or satisfied (n = 2) with the outcome. Significant differences (p <
0.0001) were found for all categories of the Constant and Murley score pre- and post-operatively. The mean Constant and Murley score increased from 24.1 points (10 to 45) to 64.5 points (39 to 93), and the relative score from 30.4% (11% to 50%) to 83% (54% to 116%). No clinical or radiological signs of loosening of the implant were seen. For young and middle-aged patients with osteoarthritis, third-generation total shoulder replacement is a viable method of treatment with a low rate of complications and excellent results in the mid-term.
The outcome of 219 revision total hip arthroplasties
(THAs) in 98 male and 121 female patients, using 137 long length
and 82 standard length cemented collarless double-taper femoral
stems in 211 patients, with a mean age of 72 years (30 to 90) and
mean follow-up of six years (two to 18) have been described previously.
We have extended the follow-up to a mean of 13 years (8 to 20) in
this cohort of patients in which the pre-operative bone deficiency Paprosky
grading was IIIA or worse in 79% and 73% of femurs with long and
standard stems, respectively. For the long stem revision group, survival to re-revision for
aseptic loosening at 14 years was 97% (95% confidence interval (CI)
91 to 100) and in patients aged >
70 years, survival was 100%. Two
patients (two revisions) were lost to follow-up and 86 patients
with 88 revisions had died. Worst-case analysis for survival to
re-revision for aseptic loosening at 14 years was 95% (95% CI 89
to 100) and 99% (95% CI 96 to 100) for patients aged >
70 years. One
additional long stem was classified as loose radiographically but
not revised. For the standard stem revision group, survival to re-revision
for aseptic loosening at 14 years was 91% (95% CI 83 to 99). No
patients were lost to follow-up and 49 patients with 51 hips had
died. No additional stems were classified as loose radiographically. Femoral revision using a cemented collarless double-taper stem,
particularly with a long length stem, and in patients aged >
70
years, continues to yield excellent results up to 20 years post-operatively,
including in hips with considerable femoral metaphyseal bone loss. Cite this article:
This was a randomised controlled trial studying
the safety of a new short metaphyseal fixation (SMF) stem. We hypothesised
that it would have similar early clinical results and micromovement
to those of a standard-length tapered Synergy metaphyseal fixation
stem. Using radiostereometric analysis (RSA) we compared the two
stems in 43 patients. A short metaphyseal fixation stem was used
in 22 patients and a Synergy stem in 21 patients. No difference
was found in the clinical outcomes pre- or post-operatively between
groups. RSA showed no significant differences two years post-operatively
in mean micromovement between the two stems (except for varus/valgus tilt
at p = 0.05) (subsidence 0.94 mm ( Cite this article:
Persistent groin pain after seemingly successful
total hip replacement (THR) appears to have become more common.
Recent studies have indicated a high incidence after metal-on-polyethylene
and metal-on-metal conventional THR and it has been documented in
up to 18% of patients after metal-on-metal resurfacing. There are many
causes, including acetabular loosening, stress fracture, and iliopsoas
tendonitis and impingement. The evaluation of this problem requires
a careful history and examination, plain radiographs and an algorithmic approach
to special diagnostic imaging and tests. Non-operative treatment
is not usually successful. Specific operative treatment depending
on the cause of the pain usually involves revision of the acetabular
component, iliopsoas tenotomy or other procedures, and is usually
successful. Here, an appropriate algorithm is described.
Deficiencies of acetabular bone stock at revision hip replacement were reconstructed with two different types of allograft using impaction bone grafting and a Burch-Schneider reinforcement ring. We compared a standard frozen non-irradiated bone bank allograft (group A) with a freeze-dried irradiated bone allograft, vitalised with autologous marrow (group B). We studied 78 patients (79 hips), of whom 87% (69 hips) had type III acetabular defects according to the American Academy of Orthopaedic Surgeons classification at a mean of 31.4 months (14 to 51) after surgery. At the latest follow-up, the mean Harris hip score was 69.9 points (13.5 to 97.1) in group A and 71.0 points (11.5 to 96.5) in group B. Each hip showed evidence of trabeculation and incorporation of the allograft with no acetabular loosening. These results suggest that the use of an acetabular reinforcement ring and a living composite of sterile allograft and autologous marrow appears to be a method of reconstructing acetabular deficiencies which gives comparable results to current forms of treatment.