Advertisement for orthosearch.org.uk
Results 341 - 359 of 359
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1427 - 1430
1 Nov 2007
Jameson S Reed MR

This paper considers the new financial infrastructure of the National Health Service and provides a resource for orthopaedic surgeons. We describe the importance of accurate documentation and data collection for National Health Service hospital Trust finances and league tables, and support our discussion with examples drawn from our local audit work.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 37 - 43
1 Jan 2009
Hassaballa M Mehendale S Poniatowski S Kalantzis G Smith E Learmonth ID

Loss of bone stock is a major problem in revision surgery of the hip. Impaction bone grafting of the femur is frequently used when dealing with deficient bone stock. In this retrospective study a consecutive series of 68 patients (69 hips) who had revision of a hip replacement with femoral impaction grafting were reviewed. Irradiated bone allograft was used in all hips. Radiological measurement of subsidence of the stem, incorporation of the graft and remodelling was carried out and showed incorporation of the graft in 26 of 69 hips (38%). However, there was no evidence of trabecular remodelling. Moderate subsidence of 5 mm to 10 mm occurred in ten hips (14.5%), and massive subsidence of > 10 mm in five (7.2%).

The results of this study are less favourable than those of others describing studies of revision of the femoral stem using impaction bone grafting. The absence of the characteristic changes of graft remodelling noted in other series raises the question as to whether irradiated bone graft may be a significant factor influencing the post-operative outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 11 | Pages 1422 - 1427
1 Nov 2008
Utting MR Raghuvanshi M Amirfeyz R Blom AW Learmonth ID Bannister GC

We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically.

The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis.

The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years.

Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 776 - 783
1 Jun 2009
Rajesparan K Biant LC Ahmad M Field RE

Tranexamic acid is a fibrinolytic inhibitor which reduces blood loss in total knee replacement. We examined the effect on blood loss of a standardised intravenous bolus dose of 1 g of tranexamic acid, given at the induction of anaesthesia in patients undergoing total hip replacement and tested the potential prothrombotic effect by undertaking routine venography. In all, 36 patients received 1 g of tranexamic acid, and 37 no tranexamic acid. Blood loss was measured directly per-operatively and indirectly post-operatively.

Tranexamic acid reduced the early post-operative blood loss and total blood loss (p = 0.03 and p = 0.008, respectively) but not the intraoperative blood loss. The tranexamic acid group required fewer transfusions (p = 0.03) and had no increased incidence of deep-vein thrombosis. The reduction in early post-operative blood loss was more marked in women (p = 0.05), in whom this effect was dose-related (r = −0.793).

Our study showed that the administration of a standardised pre-operative bolus of 1 g of tranexamic acid was cost-effective in reducing the blood loss and transfusion requirements after total hip replacement, especially in women.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1046 - 1049
1 Aug 2005
Shepperd JAN Apthorp H


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1073 - 1076
1 Aug 2005
Cross MJ Parish EN

We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 296 - 300
1 Mar 2007
van Haaren EH Heyligers IC Alexander FGM Wuisman PIJM

We reviewed the results of 71 revisions of the acetabular component in total hip replacement, using impaction of bone allograft. The mean follow-up was 7.2 years (1.6 to 9.7). All patients were assessed according to the American Academy of Orthopedic Surgeons (AAOS) classification of bone loss, the amount of bone graft required, thickness of the graft layer, signs of graft incorporation and use of augmentation.

A total of 20 acetabular components required re-revision for aseptic loosening, giving an overall survival of 72% (95% CI, 54.4 to 80.5). Of these failures, 14 (70%) had an AAOS type III or IV bone defect. In the failed group, poor radiological and histological graft incorporation was seen.

These results suggest that impaction allografting in acetabular revision with severe bone defects may have poorer results than have previously been reported.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 5 | Pages 609 - 614
1 May 2007
Himanen A Belt EA Lehto MUK Hämäläinen MMJ

We evaluated the survival of moulded monoblock and modular tibial components of the AGC total knee replacement in patients with rheumatoid arthritis. Between 1985 and 1995, 751 knees with this diagnosis were replaced at our institution. A total of 256 tibial components were of the moulded design and 495 of the modular design. The mean follow-up of the moulded subgroup was 9.6 years (0.5 to 14.7), and that of the modular group 7.0 years (0.1 to 14.7).

The groups differed significantly from each other in Larsen grade, cementing of components and patellar resurfacing, but no statistically significant difference between the survival of the components was found (Log rank test, p = 0.91). The cumulative success rate of the moulded group was 96.8% (95% confidence interval 93.6% to 98.4%) at five years and 94.4% (95% confidence interval 90.4% to 96.7%) at ten years, and of the modular group 96.2% (95% confidence interval 94% to 97.6%) and 93.6% (95% confidence interval 89.7% to 96%), respectively. Revision was required in 37 total knee replacements, the main causes were infection, pain, loosening of the tibial component and patellar problems. Survival rates for both components were satisfactory.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1031 - 1035
1 Aug 2007
Dora C Houweling M Koch P Sierra RJ

We have reviewed a group of patients with iliopsoas impingement after total hip replacement with radiological evidence of a well-fixed malpositioned or oversized acetabular component. A consecutive series of 29 patients (30 hips) was assessed. All had undergone a trial of conservative management with no improvement in their symptoms. Eight patients (eight hips) preferred continued conservative management (group 1), and 22 hips had either an iliopsoas tenotomy (group 2) or revision of the acetabular component and debridement of the tendon (group 3), based on clinical and radiological findings. Patients were followed clinically for at least two years, and 19 of the 22 patients (86.4%) who had surgery were contacted by phone at a mean of 7.8 years (5 to 9) post-operatively. Conservative management failed in all eight hips. At the final follow-up, operative treatment resulted in relief of pain in 18 of 22 hips (81.8%), with one hip in group 2 and three in group 3 with continuing symptoms. The Harris Hip Score was significantly better in the combined groups 2 and 3 than in group 1. There was a significant rate of complications in group 3. This group initially had better functional scores, but at final follow-up these were no different from those in group 2.

Tenotomy of the iliopsoas and revision of the acetabular component are both successful surgical options. Iliopsoas tenotomy provided the same functional results as revision of the acetabular component and avoided the risks of the latter procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1005 - 1012
1 Aug 2008
Tsiridis E Pavlou G Charity J Tsiridis E Gie G West R

Comparison of the safety and efficacy of bilateral simultaneous total hip replacement (THR) and that of staged bilateral THR and unilateral THR was conducted using DerSimonian-Laird heterogeneity meta-analysis. A review of the English-language literature identified 23 citations eligible for inclusion. A total of 2063 bilateral simultaneous THR patients were identified. Meta-analysis of homogeneous data revealed no statistically significant differences in the rates of thromboembolic events (p = 0.268 and p = 0.365) and dislocation (p = 0.877) when comparing staged or unilateral with bilateral simultaneous THR procedures. A systematic analysis of heterogeneous data demonstrated that the mean length of hospital stay was shorter after bilateral simultaneous THR. Higher blood transfusion requirements were expected following bilateral simultaneous THR than staged or unilateral THR, and surgical time was not different between groups. This procedure was also found to be economically and functionally efficacious when performed by experienced surgeons in specialist centres.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 12 | Pages 1617 - 1621
1 Dec 2005
O’Shea K Quinlan JF Kutty S Mulcahy D Brady OH

We assessed the outcome of patients with Vancouver type B2 and B3 periprosthetic fractures treated with femoral revision using an uncemented extensively porous-coated implant. A retrospective clinical and radiographic assessment of 22 patients with a mean follow-up of 33.7 months was performed. The mean time from the index procedure to fracture was 10.8 years. There were 17 patients with a satisfactory result. Complications in four patients included subsidence in two, deep sepsis in one, and delayed union in one. Concomitant acetabular revision was required in 19 patients. Uncemented extensively porous-coated femoral stems incorporate distally allowing stable fixation. We found good early survival rates and a low incidence of nonunion using this implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 921 - 927
1 Jul 2005
Glyn-Jones S Gill HS Beard DJ McLardy-Smith P Murray DW

Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability.

Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°, sd 0.42°) and internally rotate (mean 1.28°, sd 0.99°). The CPS-Plus showed no significant valgus rotation (mean 0.2°, sd 0.42°) or internal rotation (mean −0.03°, sd 0.75°). A wider, more rectangular cross-section improves rotational stability and may have a better long-term outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 2 | Pages 189 - 195
1 Feb 2007
Levy JC Virani N Pupello D Frankle M

We report the use of the reverse shoulder prosthesis in the revision of a failed shoulder hemiarthroplasty in 19 shoulders in 18 patients (7 men, 11 women) with severe pain and loss of function. The primary procedure had been undertaken for glenohumeral arthritis associated with severe rotator cuff deficiency.

Statistically significant improvements were seen in pain and functional outcome. After a mean follow-up of 44 months (24 to 89), mean forward flexion improved by 26.4° and mean abduction improved by 35°. There were six prosthesis-related complications in six shoulders (32%), five of which had severe bone loss of the glenoid, proximal humerus or both. Three shoulders (16%) had non-prosthesis related complications.

The use of the reverse shoulder prosthesis provides improvement in pain and function for patients with failure of a hemiarthroplasty for glenohumeral arthritis and rotator cuff deficiency. However, high rates of complications were associated with glenoid and proximal humeral bone loss.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 877 - 882
1 Jul 2006
Hasegawa M Sudo A Uchida A

A modular layered acetabular component (metal-polyethylene-ceramic) was developed in Japan for use in alumina ceramic-on-ceramic total hip replacement. Between May 1999 and July 2000, we performed 35 alumina ceramic-on-ceramic total hip replacements in 30 consecutive patients, using this layered component and evaluated the clinical and radiological results over a mean follow-up of 5.8 years (5 to 6.5). A total of six hips underwent revision, one for infection, two for dislocation with loosening of the acetabular component, two for alumina liner fractures and one for component dissociation with pelvic osteolysis. There were no fractures of the ceramic heads, and no loosening of the femoral or acetabular component in the unrevised hips was seen at final follow-up. Osteolysis was not observed in any of the unrevised hips. The survivorship analysis at six years after surgery was 83%. The layered acetabular component in our experience, has poor durability because of unexpected mechanical failures including alumina liner fracture and component dissociation.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 852 - 857
1 Jul 2006
Board TN Rooney P Kearney JN Kay PR


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1567 - 1573
1 Dec 2006
Kovac S Trebse R Milosev I Pavlovcic V Pisot V

We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean follow-up time for stable stems was 15.1 years (12.1 to 16.6).

Survival of the stem at 15 years was 75.4% (95% confidence interval (CI) 67.3 to 83.5) with aseptic failure (including radiological failure) as the end-point, irrespective of the nature of the head and the quality of the cement mantle. Survival of the stem at 15 years was 79.1% (95% CI 69.8 to 88.4) and 67.1% (95% CI 51.3 to 82.9) with the CoCr alloy and ceramic heads, respectively. The quality of the cement mantle was graded as a function of stem coverage: stems with complete tip coverage (type 1) had an 84.9% (95% CI 77.6 to 92.2) survival at 15 years, compared with those with a poor tip coverage (type 2) which had a survival of only 22.4% (95% CI 2.4 to 42.4). The poor quality of the cement mantle and the implantation of an alumina head substantially lowered the survival of the stem.

In our opinion, further use of the cemented titanium alloy straight-stem femoral components used in our series is undesirable.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1303 - 1308
1 Oct 2006
Johnsen SP Sørensen HT Lucht U Søballe K Overgaard S Pedersen AB

We examined the association between patient-related factors and the risk of initial, short- and long-term implant failure after primary total hip replacement. We used data from the Danish Hip Arthroplasty Registry between 1 January 1995 and 31 December 2002, which gave us a total of 36 984 patients. Separate analyses were carried out for three follow-up periods: 0 to 30 days, 31 days to six months (short term), and six months to 8.6 years after primary total hip replacement (long term). The outcome measure was defined as time to failure, which included re-operation with open surgery for any reason.

Male gender and a high Charlson co-morbidity index score were strongly predictive for failure, irrespective of the period of follow-up. Age and diagnosis at primary total hip replacement were identified as time-dependent predictive factors of failure. During the first 30 days after primary total hip replacement, an age of 80 years or more and hip replacement undertaken as a sequela of trauma, for avascular necrosis or paediatric conditions, were associated with an increased risk of failure. However, during six months to 8.6 years after surgery, being less than 60 years old was associated with an increased risk of failure, whereas none of the diagnoses for primary total hip replacement appeared to be independent predictors.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 562 - 575
1 May 2006
Boileau P Sinnerton RJ Chuinard C Walch G


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 366 - 373
1 Mar 2006
Baumann C Rat AC Osnowycz G Mainard D Delagoutte JP Cuny C Guillemin F

We conducted a multicentre cohort study of 228 patients with osteoarthritis followed up after total hip or knee replacement. Quality of life and patient satisfaction were assessed by self-administered questionnaires. Patient satisfaction was the dependent variable in a multivariate linear regression model. Independent variables included sociodemographic factors, pre- and post-operative clinical characteristics and the pre-operative and post-discharge health-related quality of life.

The mean age of the patients was 69 years (sd 9), and 43.8% were male. Pre- and postoperative clinical characteristics were not associated with satisfaction with health care. Only pre-operative bodily pain (p < 0.01) and pre-operative social functioning (p < 0.05) influenced patient satisfaction with care.

The pre-operative health-related quality of life and patient characteristics have little effect on inpatient satisfaction with care. This suggests that the impact of the care process on satisfaction may be independent of observed and perceived initial patient-related characteristics.