Surgeons have commonly used modular femoral heads and stems from
different manufacturers, although this is not recommended by orthopaedic
companies due to the different manufacturing processes. We compared the rate of corrosion and rate of wear at the trunnion/head
taper junction in two groups of retrieved hips; those with mixed
manufacturers (MM) and those from the same manufacturer (SM). We identified 151 retrieved hips with large-diameter cobalt-chromium
heads; 51 of two designs that had been paired with stems from different
manufacturers (MM) and 100 of seven designs paired with stems from
the same manufacturer (SM). We determined the severity of corrosion
with the Goldberg corrosion score and the volume of material loss
at the head/stem junction. We used multivariable statistical analysis
to determine if there was a significant difference between the two
groups. Aims
Materials and Methods
A pelvic discontinuity occurs when the superior
and inferior parts of the hemi-pelvis are no longer connected, which
is difficult to manage when associated with a failed total hip replacement.
Chronic pelvic discontinuity is found in 0.9% to 2.1% of hip revision
cases with risk factors including severe pelvic bone loss, female
gender, prior pelvic radiation and rheumatoid arthritis. Common
treatment options include: pelvic plating with allograft, cage reconstruction,
custom triflange implants, and porous tantalum implants with modular augments.
The optimal technique is dependent upon the degree of the discontinuity,
the amount of available bone stock and the likelihood of achieving
stable healing between the two segments. A method of treating pelvic
discontinuity using porous tantalum components with a distraction
technique that achieves both initial stability and subsequent long-term
biological fixation is described. Cite this article:
The diagnosis of periprosthetic joint infection (PJI) remains
demanding due to limitations of all the available diagnostic tests.
The synovial fluid marker, α-defensin, is a promising adjunct for
the assessment of potential PJI. The purpose of this study was to
investigate the qualitative assessment of α-defensin, using Synovasure
to detect or exclude periprosthetic infection in total joint arthroplasty. We studied 50 patients (28 women, 22 men, mean age 65 years;
20 to 89) with a clinical indication for revision arthroplasty who
met the inclusion criteria of this prospective diagnostic study.
The presence of α-defensin was determined using the qualitative
Synovasure test and compared with standard diagnostic methods for
PJI. Based on modified Musculoskeletal Infection Society (MSIS)
criteria, 13 cases were categorised as septic and 36 as aseptic revisions.
One test was inconclusive.Aims
Patients and Methods
Third-body wear is believed to be one trigger for adverse results
with metal-on-metal (MOM) bearings. Impingement and subluxation
may release metal particles from MOM replacements. We therefore
challenged MOM bearings with relevant debris types of cobalt–chrome
alloy (CoCr), titanium alloy (Ti6Al4V) and polymethylmethacrylate
bone cement (PMMA). Cement flakes (PMMA), CoCr and Ti6Al4V particles (size range
5 µm to 400 µm) were run in a MOM wear simulation. Debris allotments
(5 mg) were inserted at ten intervals during the five million cycle
(5 Mc) test. Objectives
Methods
The conventional method for reconstructing acetabular
bone loss at revision surgery includes using structural bone allograft.
The disadvantages of this technique promoted the advent of metallic
but biocompatible porous implants to fill bone defects enhancing
initial and long-term stability of the acetabular component. This
paper presents the indications, surgical technique and the outcome
of using porous metal acetabular augments for reconstructing acetabular
defects. Cite this article:
Polyethylene wear debris can cause osteolysis
and the failure of total hip arthroplasty. We present the five-year
wear rates of a highly cross-linked polyethylene (X3) bearing surface
when used in conjunction with a 36 mm ceramic femoral head. This was a prospective study of a cohort of 100 THAs in 93 patients.
Pain and activity scores were measured pre- and post-operatively.
Femoral head penetration was measured at two months, one year, two
years and at five years using validated edge-detecting software
(PolyWare Auto). At a mean of 5.08 years (3.93 to 6.01), 85 hips in 78 patients
were available for study. The mean age of these patients was 59.08
years (42 to 73, the mean age of males (n = 34) was 59.15 years,
and females (n = 44) was 59.02 years). All patients had significant
improvement in their functional scores (p <
0.001). The steady
state two-dimensional linear wear rate was 0.109 mm/year. The steady
state volumetric wear rate was 29.61 mm3/year. No significant
correlation was found between rate of wear and age (p = 0.34), acetabular
component size (p = 0.12) or clinical score (p = 0.74). Our study shows low steady state wear rates at five years in
X3 highly cross-linked polyethylene in conjunction with a 36 mm
ceramic femoral head. The linear wear rate was almost identical
to the osteolysis threshold of 0.1 mm/year recommended in the literature. Cite this article:
Joint replacement of the hip and knee remain
very satisfactory operations. They are, however, expensive. The
actual manufacturing of the implant represents only 30% of the final
cost, while sales and marketing represent 40%. Recently, the patents
on many well established and successful implants have expired. Companies
have started producing and distributing implants that purport to
replicate existing implants with good long-term results. The aims of this paper are to assess the legality, the monitoring
and cost saving implications of such generic implants. We also assess
how this might affect the traditional orthopaedic implant companies. Cite this article:
Pelvic discontinuity represents a rare but challenging
problem for orthopaedic surgeons. It is most commonly encountered
during revision total hip replacement, but can also result from
an iatrogentic acetabular fracture during hip replacement. The general
principles in management of pelvic discontinuity include restoration
of the continuity between the ilium and the ischium, typically with
some form of plating. Bone grafting is frequently required to restore
pelvic bone stock. The acetabular component is then impacted, typically
using an uncemented, trabecular metal component. Fixation with multiple
supplemental screws is performed. For larger defects, a so-called
‘cup–cage’ reconstruction, or a custom triflange implant may be
required. Pre-operative CT scanning can greatly assist in planning
and evaluating the remaining bone stock available for bony ingrowth.
Generally, good results have been reported for constructs that restore
stability to the pelvis and allow some form of biologic ingrowth. Cite this article:
Few studies have assessed outcomes following non-metal-on-metal hip arthroplasty (non-MoMHA) revision surgery performed for adverse reactions to metal debris (ARMD). We assessed outcomes following non-MoMHA revision surgery performed for ARMD, and identified predictors of re-revision. We performed a retrospective observational study using data from the National Joint Registry for England and Wales. All non-MoMHAs undergoing revision surgery for ARMD between 2008 and 2014 were included (185 hips in 185 patients). Outcome measures following ARMD revision were intra-operative complications, mortality and re-revision surgery. Predictors of re-revision were identified using Cox regression.Objectives
Methods
Many studies have investigated the kinematics of the lumbar spine and the morphological features of the lumbar discs. However, the segment-dependent immediate changes of the lumbar intervertebral space height during flexion-extension motion are still unclear. This study examined the changes of intervertebral space height during flexion-extension motion of lumbar specimens. First, we validated the accuracy and repeatability of a custom-made mechanical loading equipment set-up. Eight lumbar specimens underwent CT scanning in flexion, neural, and extension positions by using the equipment set-up. The changes in the disc height and distance between adjacent two pedicle screw entry points (DASEP) of the posterior approach at different lumbar levels (L3/4, L4/5 and L5/S1) were examined on three-dimensional lumbar models, which were reconstructed from the CT images.Objectives
Methods
We undertook a retrospective cohort study to
determine clinical outcomes following the revision of metal-on-metal (MoM)
hip replacements for adverse reaction to metal debris (ARMD), and
to identify predictors of time to revision and outcomes following
revision. Between 1998 and 2012 a total of 64 MoM hips (mean age
at revision of 57.8 years; 46 (72%) female; 46 (72%) hip resurfacings
and 18 (28%) total hip replacements) were revised for ARMD at one specialist
centre. At a mean follow-up of 4.5 years (1.0 to 14.6) from revision
for ARMD there were 13 hips (20.3%) with post-operative complications
and eight (12.5%) requiring re-revision. The Kaplan–Meier five-year survival rate for ARMD revision was
87.9% (95% confidence interval 78.9 to 98.0; 19 hips at risk). Excluding
re-revisions, the median absolute Oxford hip score (OHS) following
ARMD revision using the percentage method (0% best outcome and 100%
worst outcome) was 18.8% (interquartile range (IQR) 7.8% to 48.3%),
which is equivalent to 39/48 (IQR 24.8/48 to 44.3/48) when using
the modified OHS. Histopathological response did not affect time
to revision for ARMD (p = 0.334) or the subsequent risk of re-revision
(p = 0.879). Similarly, the presence or absence of a contralateral
MoM hip bearing did not affect time to revision for ARMD (p = 0.066)
or the subsequent risk of re-revision (p = 0.178). Patients revised to MoM bearings had higher rates of re-revision
(five of 16 MoM hips re-revised; p = 0.046), but those not requiring
re-revision had good functional results (median absolute OHS 14.6%
or 41.0/48). Short-term morbidity following revision for ARMD was
comparable with previous reports. Caution should be exercised when choosing
bearing surfaces for ARMD revisions. Cite this article:
Few studies dealing with chondrosarcoma of the pelvis are currently
available. Different data about the overall survival and prognostic
factors have been published but without a detailed analysis of surgery-related complications.
We aimed to analyse the outcome of a series of pelvic chondrosarcomas
treated at a single institution, with particular attention to the
prognostic factors. Based on a competing risk model, our objective
was to identify risk factors for the development of complications. In a retrospective single-centre study, 58 chondrosarcomas (26
patients alive, 32 patients dead) of the pelvis were reviewed. The
mean follow-up was 13 years (one week to 23.1 years).Aims
Patients and Methods
The aim of this systematic review was to report the rate of dislocation
following the use of dual mobility (DM) acetabular components in
primary and revision total hip arthroplasty (THA). A systematic review of the literature according to the Preferred
Reporting Items for Systematic Reviews and Meta-analyses guidelines
was performed. A comprehensive search of Pubmed/Medline, Cochrane
Library and Embase (Scopus) was conducted for English articles between
January 1974 and March 2016 using various combinations of the keywords “dual
mobility”, “dual-mobility”, “tripolar”, “double-mobility”, “double
mobility”, “hip”, “cup”, “socket”. The following data were extracted
by two investigators independently: demographics, whether the operation
was a primary or revision THA, length of follow-up, the design of
the components, diameter of the femoral head, and type of fixation
of the acetabular component.Aims
Materials and Methods
A moderator and panel of five experts led an
interactive session in discussing five challenging and interesting patient
case presentations involving surgery of the hip. The hip pathologies
reviewed included failed open reduction internal fixation of subcapital
femoral neck fracture, bilateral hip disease, evaluation of pain
after metal-on-metal hip arthroplasty, avascular necrosis, aseptic
loosening secondary to osteolysis and polyethylene wear, and management
of ceramic femoral head fracture.
The April 2014 Trauma Roundup360 looks at: is it safe to primarily close dog bite wounds?; conservative transfusion evidence based in hip fracture surgery; tibial nonunion is devastating to quality of life; sexual dysfunction after traumatic pelvic fracture; hemiarthroplasty versus fixation in displaced femoral neck fractures; silver VAC dressings “Gold Standard” in massive wounds; dual plating for talar neck fracture; syndesmosis and fibular length easiest errors in ankle fracture surgery; and dual mobility: stable as a rock in fracture.
The aims of this retrospective study were to report the feasibility
of using 3D-printing technology for patients with a pelvic tumour
who underwent reconstruction. A total of 35 patients underwent resection of a pelvic tumour
and reconstruction using 3D-printed endoprostheses between September
2013 and December 2015. According to Enneking’s classification of
bone defects, there were three Type I lesions, 12 Type II+III lesions,
five Type I+II lesions, two Type I+II+III lesions, ten type I+II+IV
lesions and three type I+II+III+IV lesions. A total of three patients
underwent reconstruction using an iliac prosthesis, 12 using a standard
hemipelvic prosthesis and 20 using a screw-rod connected hemipelvic
prosthesis.Aims
Patients and Methods
The increasing need for total hip replacement
(THR) in an ageing population will inevitably generate a larger number
of revision procedures. The difficulties encountered in dealing
with the bone deficient acetabulum are amongst the greatest challenges
in hip surgery. The failed acetabular component requires reconstruction
to restore the hip centre and improve joint biomechanics. Impaction
bone grafting is successful in achieving acetabular reconstruction
using both cemented and cementless techniques. Bone graft incorporation
restores bone stock whilst providing good component stability. We
provide a summary of the evidence and current literature regarding impaction
bone grafting using both cemented and cementless techniques in revision
THR. Cite this article:
The aim of this randomised controlled study was
to compare functional and radiological outcomes between modern cemented
and uncemented hydroxyapatite coated stems after one year in patients
treated surgically for a fracture of the femoral neck. A total of
141 patients aged >
65 years were included. Patients were randomised
to be treated with a cemented Exeter stem or an uncemented Bimetric
stem. The patients were reviewed at four and 12 months. The cemented group performed better than the uncemented group
for the Harris hip score (78 In conclusion, our data do not support the use of an uncemented
hydroxyapatite coated stem for the treatment of displaced fractures
of the femoral neck in the elderly. Cite this article: