Advertisement for orthosearch.org.uk
Results 341 - 360 of 800
Results per page:
The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 897 - 901
1 Aug 2019
Konan S Alazzawi S Yoon B Cha Y Koo K

Ceramic bearings have several desirable properties, such as resistance to wear, hardness, and biocompatibility, that favour it as an articulating surface in hip arthroplasty. However, ceramic fracture remains a concern. We have reviewed the contemporary literature, addressing the factors that can influence the incidence of ceramic bearing surface fracture.

Cite this article: Bone Joint J 2019;101-B:897–901.


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 6 | Pages 908 - 910
1 Nov 1991
Beaver R Schemitsch E Gross A

We report a case of late dissociation of a one-piece metal-backed acetabular component, associated with a prolific soft-tissue reaction to the polyethylene debris. The polyethylene liner was not visible on the radiographs. The metal-backed shell could not be removed because of bone ingrowth. We recommend that modular systems be utilised in cementless arthroplasty of the hip and that radiographic markers be incorporated in the polyethylene of the acetabular cup


The Bone & Joint Journal
Vol. 101-B, Issue 6 | Pages 724 - 731
1 Jun 2019
Bernthal NM Upfill-Brown A Burke ZDC Ishmael CR Hsiue P Hori K Hornicek F Eckardt JJ

Aims

Aseptic loosening is a major cause of failure in cemented endoprosthetic reconstructions. This paper presents the long-term outcomes of a custom-designed cross-pin fixation construct designed to minimize rotational stress and subsequent aseptic loosening in selected patients. The paper will also examine the long-term survivorship and modes of failure when using this technique.

Patients and Methods

A review of 658 consecutive, prospectively collected cemented endoprosthetic reconstructions for oncological diagnoses at a single centre between 1980 and 2017 was performed. A total of 51 patients were identified with 56 endoprosthetic implants with cross-pin fixation, 21 of which were implanted following primary resection of tumour. Locations included distal femoral (n = 36), proximal femoral (n = 7), intercalary (n = 6), proximal humeral (n = 3), proximal tibial (n = 3), and distal humeral (n = 1).


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 631 - 635
1 Jul 2002
Norton MR Yarlagadda R Anderson GH

We report catastrophic early failure of a cemented total hip replacement comprising a modular femoral component with a Zirconia ceramic head and an acetabular component of cross-linked ultra-high molecular-weight polyethylene (Hylamer). Between 1995 and 1999 we implanted 29 hips in 26 patients with a mean age of 49.2 years. Survivorship analysis in this group revealed a failure rate of 67.6% at five years. All hips which failed did so because of aseptic loosening with progressive osteolysis or radiolucencies. We therefore recommend early and regular review of all patients with this combination of implants and early revision surgery in order to avoid massive bone loss


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 3 | Pages 324 - 334
1 Apr 2002
Williams HDW Browne G Gie GA Ling RSM Timperley AJ Wendover NA

We describe our experience with the implantation of 325 Exeter Universal hips. The fate of every implant was known. The procedures were undertaken by surgeons of widely differing experience. At follow-up at 12 years, survivorship with revision of the femoral component for aseptic loosening as the endpoint was 100% (95% CI 98 to 100). Survivorship with revision of the acetabular component for aseptic loosening as the endpoint was 96.86% (95% CI 93.1 to 98.9) and that with any reoperation as the endpoint 91.74% (95% CI 87.7 to 95.8). No adverse features have emerged as a consequence of the modular connection between the head and neck of the implant


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1356 - 1361
1 Nov 2019
Chalmers BP Mehrotra KG Sierra RJ Pagnano MW Taunton MJ Abdel MP

Aims

Knee osteonecrosis in advanced stages may lead to joint degeneration. Total knee arthroplasty (TKA) for osteonecrosis has traditionally been associated with suboptimal results. We analyzed outcomes of contemporary TKAs for osteonecrosis, with particular emphasis on: survivorship free from aseptic loosening, any revision, and any reoperation plus the clinical outcomes, complications, and radiological results.

Patients and Methods

In total, 156 patients undergoing 167 primary TKAs performed for osteonecrosis between 2004 and 2014 at a single institution were reviewed. The mean age at index TKA was 61 years (14 to 93) and the mean body mass index (BMI) was 30 kg/m2 (18 to 51) The mean follow-up was six years (2 to 12). A total of 110 TKAs (66%) were performed for primary osteonecrosis and 57 TKAs (34%) for secondary osteonecrosis. Overall, 15 TKAs (9%) had tibial stems, while 12 TKAs (7%) had femoral stems. Posterior-stabilized designs were used in 147 TKAs (88%) of TKAs. Bivariate Cox regression analysis was conducted to identify risk factors for revision and reoperation.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 345 - 350
1 Mar 1998
Ă–nsten I Carlsson Ă…S Besjakov J

We used radiostereometric analysis to compare wear rates between uncemented porous sockets and cemented all-polyethylene sockets in a series of 102 hips randomised for either a Harris-Galante or a Charnley cup. Wear was evaluated in 95 hips at a mean of five years (2 to 7). All hips had a cemented, 22 mm head mono-bloc Charnley stem. The mean annual wear rate was 0.09 mm in the Charnley sockets and 0.10 mm in the Harris-Galante sockets, with no statistically significant differences in wear, migration or rotation. We conclude that, up to five years, the wear characteristics of the modular and porous Harris-Galante socket resemble that of the Charnley socket


Bone & Joint Open
Vol. 1, Issue 3 | Pages 29 - 34
13 Mar 2020
Stirling P Middleton SD Brenkel IJ Walmsley PJ

Introduction

The primary aim of this study was to describe a baseline comparison of early knee-specific functional outcomes following revision total knee arthroplasty (TKA) using metaphyseal sleeves with a matched cohort of patients undergoing primary TKA. The secondary aim was to compare incidence of complications and length of stay (LOS) between the two groups.

Methods

Patients undergoing revision TKA for all diagnoses between 2009 and 2016 had patient-reported outcome measures (PROMs) collected prospectively. PROMs consisted of the American Knee Society Score (AKSS) and Short-Form 12 (SF-12). The study cohort was identified retrospectively and demographics were collected. The cohort was matched to a control group of patients undergoing primary TKA.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 3 | Pages 344 - 349
1 Apr 2004
McCombe P Williams SA

We selected randomly a consecutive series of 162 patients requiring hip replacement to receive either a cementless, hemispherical, modular, titanium acetabular cup or a cemented, all-polyethylene cup. These replacements were performed by two surgeons in four general hospitals. The same surgical technique was used and a 26 mm metal-head femoral component was used in every case. After exclusions, 115 hips were studied for differences in rates of wear and osteolysis. The mean clinical follow-up was eight years and the mean radiological follow-up, 6.5 years. The cementless cups wore at a mean rate of 0.15 mm per year and the cemented cups at 0.07 mm per year. This difference was significant (p < 0.0001). Our findings in this mid-term study suggest that cementless cups wear more than cemented cups


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 85 - 91
1 Jan 1996
Önsten I Carlsson ÅS Sanzén L Besjakov J

A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis


The Bone & Joint Journal
Vol. 101-B, Issue 3 | Pages 311 - 316
1 Mar 2019
Löchel J Janz V Hipfl C Perka C Wassilew GI

Aims

The use of trabecular metal (TM) shells supported by augments has provided good mid-term results after revision total hip arthroplasty (THA) in patients with a bony defect of the acetabulum. The aim of this study was to assess the long-term implant survivorship and radiological and clinical outcomes after acetabular revision using this technique.

Patients and Methods

Between 2006 and 2010, 60 patients (62 hips) underwent acetabular revision using a combination of a TM shell and augment. A total of 51 patients (53 hips) had complete follow-up at a minimum of seven years and were included in the study. Of these patients, 15 were men (29.4%) and 36 were women (70.6%). Their mean age at the time of revision THA was 64.6 years (28 to 85). Three patients (5.2%) had a Paprosky IIA defect, 13 (24.5%) had a type IIB defect, six (11.3%) had a type IIC defect, 22 (41.5%) had a type IIIA defect, and nine (17%) had a type IIIB defect. Five patients (9.4%) also had pelvic discontinuity.


Bone & Joint Research
Vol. 8, Issue 10 | Pages 489 - 494
1 Oct 2019
Klasan A Bäumlein M Dworschak P Bliemel C Neri T Schofer MD Heyse TJ

Objectives

Periprosthetic femoral fractures (PFFs) have a higher incidence with cementless stems. The highest incidence among various cementless stem types was observed with double-wedged stems. Short stems have been introduced as a bone-preserving alternative with a higher incidence of PFF in some studies. The purpose of this study was a direct load-to-failure comparison of a double-wedged cementless stem and a short cementless stem in a cadaveric fracture model.

Methods

Eight hips from four human cadaveric specimens (age mean 76 years (60 to 89)) and eight fourth-generation composite femurs were used. None of the cadaveric specimens had compromised quality (mean T value 0.4 (-1.0 to 5.7)). Each specimen from a pair randomly received either a double-wedged stem or a short stem. A materials testing machine was used for lateral load-to-failure test of up to a maximal load of 5000 N.


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 1 | Pages 73 - 77
1 Jan 1994
Jasty M Bragdon C Lee K Hanson A Harris W

Fifty-four cobalt-chrome alloy femoral heads were retrieved at revision surgery or at post-mortem; 23 came from uncemented, eight from hybrid and 23 from cemented hip arthroplasties. The uncemented and hybrid implants had porous coating, metal backing and modular femoral heads; with one exception none of the cemented implants had any of these. Twenty-five of the 31 heads from uncemented and hybrid arthroplasties, and 11 of the 23 heads from cemented arthroplasties showed surface damage involving more than 25% of the surface. Scanning electron microscopy revealed multidirectional fine scratches 1 micron to 10 microns in depth and width which appeared to have been made by fine, hard particles. There was a higher rate of such damage in the uncemented and hybrid arthroplasties than in the cemented implants, suggesting that the abrasive particles were mainly released from the metal, rather than from the cement or polyethylene components of the implants


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 3 | Pages 401 - 406
1 Apr 2002
Mittermayer F Windhager R Dominkus M Krepler P Schwameis E Sluga M Kotz R Strasser G

In 251 patients over a period of 15 years an uncemented Kotz modular femoral and tibial reconstruction mega prosthesis was implanted after resection of a malignant tumour of the lower limb. Twenty-one patients (8.4%) underwent revision for aseptic loosening, again using an uncemented prosthesis, and five of these required a further revision procedure. The median follow-up time from the first revision was 60 months (11 to 168) and after a second revision, 33 months (2 to 50). The probability of a patient avoiding aseptic loosening for ten years was 96% for a proximal femoral, 76% for a distal femoral and 85% for a proximal tibial implant. At the time of follow-up all radiographs were assessed according to the International Symposium of Limb Salvage criteria. The first radiological signs of aseptic loosening were always seen at the most proximal or distal part of the anchorage stem at a mean of 12 months (4 to 23) after the first implantation. Using the Musculoskeletal Tumor Society score for evaluation, the clinical results showed a mean of 88% of normal function


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 285 - 292
1 Mar 2020
Tanaka A Katagiri H Murata H Wasa J Miyagi M Honda Y Takahashi M

Aims

The aim of this study is to evaluate the clinical results of operative intervention for femoral metastases which were selected based on expected survival and to discuss appropriate surgical strategies.

Methods

From 2002 to 2017, 148 consecutive patients undergoing surgery for femoral metastasis were included in this study. Prognostic risk assessments were performed according to the Katagiri and revised Katagiri scoring system. In general, the low-risk group underwent resection and reconstruction with endoprosthetic replacement (EPR), while the high-risk group underwent internal fixation (IF) and radiation therapy. For the intermediate-risk group, the operative choice depended on the patient’s condition, degree of bone destruction, and radio-sensitivity. Overall survival, local failure, walking ability, and systemic complications were evaluated.


The Bone & Joint Journal
Vol. 102-B, Issue 1 | Pages 64 - 71
1 Jan 2020
Tsuda Y Fujiwara T Stevenson JD Parry MC Tillman R Abudu A

Aims

The purpose of this study was to report the long-term results of extendable endoprostheses of the humerus in children after the resection of a bone sarcoma.

Methods

A total of 35 consecutive patients treated with extendable endoprosthetic replacement of the humerus in children were included. There were 17 boys and 18 girls in the series with a median age at the time of initial surgery of nine years (interquartile range (IQR) 7 to 11).


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 787 - 792
1 Jul 2019
Goto K Kuroda Y Kawai T Kawanabe K Matsuda S

Aims

In the 1990s, a bioactive bone cement (BABC) containing apatite-wollastonite glass-ceramic (AW-GC) powder and bisphenol-a-glycidyl methacrylate resin was developed at our hospital. In 1996, we used BABC to fix the acetabular component in primary total hip arthroplasty (THA) in 20 patients as part of a clinical trial. The purpose of this study was to investigate the long-term results of primary THA using BABC.

Patients and Methods

A total of 20 patients (three men and 17 women) with a mean age of 57.4 years (40 to 71), a mean body weight of 52.3 kg (39 to 64), and a mean body mass index (BMI) of 23.0 kg/m2 (19.8 to 28.6) were evaluated clinically and radiologically. Survival analyses were undertaken, and wear analyses were carried out using a computer-aided method.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 123 - 126
1 Jun 2019
El-Husseiny M Masri B Duncan C Garbuz DS

Aims

We investigated the long-term performance of the Tripolar Trident acetabular component used for recurrent dislocation in revision total hip arthroplasty. We assessed: 1) rate of re-dislocation; 2) incidence of complications requiring re-operation; and 3) Western Ontario and McMaster Universities osteoarthritis index (WOMAC) pain and functional scores.

Patients and Methods

We retrospectively identified 111 patients who had 113 revision tripolar constrained liners between 1994 and 2008. All patients had undergone revision hip arthroplasty before the constrained liner was used: 13 after the first revision, 17 after the second, 38 after the third, and 45 after more than three revisions. A total of 75 hips (73 patients) were treated with Tripolar liners due to recurrent instability with abductor deficiency, In addition, six patients had associated cerebral palsy, four had poliomyelitis, two had multiple sclerosis, two had spina bifida, two had spondyloepiphyseal dysplasia, one had previous reversal of an arthrodesis, and 21 had proximal femoral replacements. The mean age of patients at time of Tripolar insertions was 72 years (53 to 89); there were 69 female patients (two bilateral) and 42 male patients. All patients were followed up for a mean of 15 years (10 to 24). Overall, 55 patients (57 hips) died between April 2011 and February 2018, at a mean of 167 months (122 to 217) following their tripolar liner implantation. We extracted demographics, implant data, rate of dislocations, and incidence of other complications.


Bone & Joint 360
Vol. 8, Issue 4 | Pages 5 - 13
1 Aug 2019
Middleton R Khan T Alvand A


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 4 | Pages 608 - 614
1 Jul 1995
Schiller C Windhager R Fellinger E Salzer-Kuntschik M Kaider A Kotz R

From 1986 to 1991 we fitted 20 children with endoprostheses after resection of malignant bone tumours of the leg; six have reached skeletal maturity and are the subject of this study. Reconstruction of defects in growing limbs in which the eventual shortening can be predicted requires the use of extendable prostheses. The mean age at operation was 11 years (9.2 to 13.7) and the average follow-up period was 6.3 years (4.3 to 7.6). The diagnosis was osteosarcoma in five patients and Ewing's sarcoma in one. All tumours were Enneking stage-IIB. When seen for follow-up all patients were free from disease. The extendable implants used included the Pafford-Lewis prosthesis and the Kotz Modular Femur Tibia Reconstruction system with a compatible, newly-designed growth module. Telescope-like elongation of the prostheses was performed by insertion of a screwdriver through a small skin incision. Active epiphyseal growth in the adjacent growth plate was preserved by using prosthetic stems with a smooth surface. The mean length gained was 13.15 cm (4.5 to 19.5) requiring 53 planned procedures. Seven revision operations were necessary for complications. Functional evaluation showed excellent and good results in all cases. Stress-shielding at the site of anchorage of the prosthesis was more pronounced than in adults. Implantation of extendable endoprostheses in children provides a reasonable alternative to rotationplasty, but limb salvage requires more operations