Patient-specific cutting guides (PSCGs) are designed
to improve the accuracy of alignment of total knee replacement (TKR).
We compared the accuracy of limb alignment and component positioning
after TKR performed using PSCGs or conventional instrumentation.
A total of 80 patients were randomised to undergo TKR with either of
the different forms of instrumentation, and radiological outcomes
and peri-operative factors such as operating time were assessed.
No significant difference was observed between the groups in terms
of tibiofemoral angle or femoral component alignment. Although the
tibial component in the PSCGs group was measurably closer to neutral
alignment than in the conventional group, the size of the difference
was very small (89.8° ( Cite this article:
The aim of this study was to evaluate the effects of using a
portable, accelerometer-based surgical navigation system (KneeAlign2)
in total knee arthroplasty (TKA) on the alignment of the femoral
component, and blood loss. A total of 241 consecutive patients with primary osteoarthritis
of the knee were enrolled in this prospective, randomised controlled
study. There were 207 women and 34 men. The mean age of the patients
was 74.0 years (57 to 89). The KneeAlign2 system was used for distal
femoral resection in 121 patients (KA2 group) and a conventional intramedullary
femoral guide was used in 120 patients (IM group).Aims
Patients and Methods
Patient specific cutting guides generated by
preoperative Magnetic Resonance Imaging (MRI) of the patient’s extremity
have been proposed as a method of improving the consistency of Total
Knee Arthroplasty (TKA) alignment and adding efficiency to the operative
procedure. The cost of this option was evaluated by quantifying the
savings from decreased operative time and instrument processing
costs compared to the additional cost of the MRI and the guide.
Coronal plane alignment was measured in an unselected consecutive
series of 200 TKAs, 100 with standard instrumentation and 100 with
custom cutting guides. While the cutting guides had significantly lower
total operative time and instrument processing time, the estimated
$322 savings was overwhelmed by the $1,500 additional cost of the
MRI and the cutting guide. All measures of coronal plane alignment
were equivalent between the two groups. The data does not currently
support the proposition that patient specific guides add value to
TKA.
A high radiographic inclination angle (RI) contributes
to accelerated wear and has been associated with dislocation after
total hip arthroplasty (THA). With freehand positioning of the acetabular
component there is a lack of accuracy, with a trend towards a high
radiographic inclination angle. The aim of this study was to investigate
whether the use of a digital protractor to measure the operative
inclination angle (OI) could improve the positioning of the acetabular
component in relation to a ‘safe zone’. We measured the radiographic inclination angles of 200 consecutive
uncemented primary THAs. In the first 100 the component was introduced
freehand and in the second 100 a digital protractor was used to
measure the operative inclination angle. The mean difference between the operative and the radiographic
inclination angles
(∆RI–OI) in the second cohort was 12.3° (3.8° to 19.8°). There was
a strong correlation between the circumference of the hip and ∆RI–OI.
The number of RI outliers was significantly reduced in the protractor
group (p = 0.002). Adjusting the OI, using a digital protractor and taking into
account the circumference of the patient’s hip, improves the RI
significantly (p <
0.001) and does not require additional operating
time. Cite this article:
One goal of total hip arthroplasty is to restore normal hip anatomy.
The aim of this study was to compare displacement of the centre
of rotation (COR) using a standard reaming technique with a technique
in which the acetabulum was reamed immediately peripherally and
referenced off the rim. In the first cohort the acetabulum was reamed to the floor followed
by sequentially larger reamers. In the second cohort the acetabulum
was only reamed peripherally, starting with a reamer the same size
as the native femoral head. Anteroposterior pelvic radiographs were
analysed for acetabular floor depth and vertical and horizontal position
of the COR.Aims
Patients and Methods
This prospective randomised controlled trial was designed to
evaluate the outcome of both the MRI- and CT-based patient-specific
matched guides (PSG) from the same manufacturer. A total of 137 knees in 137 patients (50 men, 87 women) were
included, 67 in the MRI- and 70 in the CT-based PSG group. Their
mean age was 68.4 years (47.0 to 88.9). Outcome was expressed as
the biomechanical limb alignment (centre hip-knee-ankle: HKA-axis)
achieved post-operatively, the position of the individual components
within 3° of the pre-operatively planned alignment, correct planned
implant size and operative data (e.g. operating time and blood loss).Aims
Patients and Methods
This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p <
0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.
A retrospective study was conducted to investigate
the changes in metal ion levels in a consecutive series of Birmingham
Hip Resurfacings (BHRs) at a minimum ten-year follow-up. We reviewed
250 BHRs implanted in 232 patients between 1998 and 2001. Implant
survival, clinical outcome (Harris hip score), radiographs and serum chromium
(Cr) and cobalt (Co) ion levels were assessed. Of 232 patients, 18 were dead (five bilateral BHRs), 15 lost
to follow-up and ten had been revised. The remaining 202 BHRs in
190 patients (136 men and 54 women; mean age at surgery 50.5 years
(17 to 76)) were evaluated at a minimum follow-up of ten years (mean
10.8 years (10 to 13.6)). The overall implant survival at 13.2 years
was 92.4% (95% confidence interval 90.8 to 94.0). The mean Harris
hip score was 97.7 (median 100; 65 to 100). Median and mean ion
levels were low for unilateral resurfacings (Cr: median 1.3 µg/l,
mean
1.95 µg/l (<
0.5 to 16.2); Co: median 1.0 µg/l, mean 1.62 µg/l
(<
0.5 to 17.3)) and bilateral resurfacings (Cr: median 3.2 µg/l,
mean 3.46 µg/l (<
0.5 to 10.0); Co: median 2.3 µg/l, mean 2.66
µg/l (<
0.5 to 9.5)). In 80 unilateral BHRs with sequential ion
measurements, Cr and Co levels were found to decrease significantly
(p <
0.001) from the initial assessment at a median of six years
(4 to 8) to the last assessment at a median of 11 years (9 to 13),
with a mean reduction of 1.24 µg/l for Cr and 0.88 µg/l for Co.
Three female patients had a >
2.5 µg/l increase of Co ions, associated with
head sizes ≤ 50 mm, clinical symptoms and osteolysis. Overall, there
was no significant difference in change of ion levels between genders
(Cr, p = 0.845; Co, p = 0.310) or component sizes (Cr, p = 0.505;
Co, p = 0.370). Higher acetabular component inclination angles correlated
with greater change in ion levels (Cr, p = 0.013; Co, p = 0.002).
Patients with increased ion levels had lower Harris hip scores (p
= 0.038). In conclusion, in well-functioning BHRs the metal ion levels
decreased significantly at ten years. An increase >
2.5 µg/l was
associated with poor function. Cite this article:
The purpose of this study was to measure the
radiological parameters of femoral component alignment of the Oxford
Phase 3 unicompartmental knee replacement (UKR), and evaluate their
effect on clinical outcome. Multiple regression analysis was used
to examine the relative contributions of the radiological assessment
of femoral component alignment in 189 consecutive UKRs performed
by a single surgeon. The American Knee Society scores were compared
between groups, defined as being within or outside recommended tolerances
of the position of the femoral component. For the flexion/extension
position 21 UKRs (11.1%) lay outside the recommended limits, and for
posterior overhang of the femoral component nine (4.8%) lay outside
the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest
canal distance from the distal femoral entry point of the alignment
jig and coronal entry-point position had significant effects on
the flexion/extension position. Pre-operative HKA angle had a significant
influence on posterior overhang of the femoral component. However,
there was no significant difference in American Knee Society scores
relative to the position of the femoral component.
We evaluated the safety and efficacy of total
knee replacement in patients receiving continuous warfarin therapy. We identified 24 consecutive patients receiving long-term warfarin
therapy who underwent total knee replacement between 2006 and 2008
and compared them with a group of age- and gender-matched patients
not on long-term anticoagulation. Primary observations were changes
in haemoglobin, transfusion rates and complications. Secondary observations
were fluctuations in the international normalised ratio (INR) and
post-operative range of movement. There was no significant difference between the two groups in
pre- or post-operative haemoglobin, incidence of transfusion or
incidence of post-operative complications. There were no surgical
delays due to a high INR level. The mean change in INR during the
peri-operative phase was minimal (mean 0.4; Current American College of Chest Physicians guidelines recommend
bridging therapy for high-risk patients receiving oral anticoagulation
and undergoing major orthopaedic procedures. We have shown that
a safe alternative is to continue the steady-state warfarin peri-operatively
in patients on long-term anticoagulation requiring total knee replacement.
We describe the findings at six years in an ongoing prospective clinicoradiological and metal ion study in a cohort of 26 consecutive male patients with unilateral Birmingham Hip Resurfacing arthroplasties with one of two femoral head sizes (50 mm and 54 mm). Their mean age was 52.9 years (29 to 67). We have previously shown an early increase in the 24-hour urinary excretion of metal ions, reaching a peak at six months (cobalt) and one year (chromium) after operation. Subsequently there is a decreasing trend in excretion of both cobalt and chromium. The levels of cobalt and chromium in whole blood also show a significant increase at one year, followed by a decreasing trend until the sixth year.
The role of computer-assisted surgery in maintaining the level of the joint in primary knee joint replacement (TKR) has not been well defined. We undertook a blinded randomised controlled trial comparing joint-line maintenance, functional outcomes, and quality-of-life outcomes between patients undergoing computer-assisted and conventional TKR. A total of 115 patients were randomised (computer-assisted, n = 55; conventional, n = 60). Two years post-operatively no significant correlation was found between computer-assisted and conventional surgery in terms of maintaining the joint line. Those TKRs where the joint line was depressed post-operatively improved the least in terms of functional scores. No difference was detected in terms of quality-of-life outcomes. Change in joint line was found to be related to change in alignment. Change in alignment significantly affects change in joint line and functional scores.
Pseudotumour is a rare but important complication of metal-on-metal hip resurfacing that occurs much more commonly in women than in men. We examined the relationship between head-neck ratio (HNR) and pseudotumour formation in 18 resurfaced hips (18 patients) revised for pseudotumour and 42 asymptomatic control resurfaced hips (42 patients). Patients in whom pseudotumour formation had occurred had higher pre-operative HNR than the control patients (mean 1.37 ( We suggest that reducing the size of the femoral head, made possible by a high pre-operative HNR, increases the risk of impingement and edge loading, and may contribute to high wear and pseudotumour formation. As the incidence of pseudotumour is low in men, it appears safe to perform resurfacing in men. However, this study suggests that it is also reasonable to resurface in women with a pre-operative HNR ≤ 1.3.
Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p <
0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.