We have reviewed 70 Harris-Galante uncemented acetabular components implanted as hybrid hip replacements with cemented stems between 1991 and 1995 in 53 patients whose mean age was 40 years (19 to 49). The mean follow-up was for 13.6 years (12 to 16) with no loss to follow-up. We assessed the patients both clinically and radiologically. The mean Oxford hip score was 20 (12 to 46) and the mean Harris hip score 81 (37 to 100) at the final review. Radiologically, 27 hips (39%) had femoral osteolysis, 13 (19%) acetabular osteolysis, and 31 (44%) radiolucent lines around the acetabular component. Kaplan-Meier survival curves were constructed for the outcomes of revision of the acetabular component, revision of the component and polyethylene liner, and impending revision for progressive osteolysis. The cumulative survival for revision of the acetabular component was 94% (95% confidence interval 88.4 to 99.7), for the component and liner 84% (95% confidence interval 74.5 to 93.5) and for impending revision 55.3% (95% confidence interval 40.6 to 70) at 16 years. Uncemented acetabular components with polyethylene liners undergo silent lysis and merit regular long-term radiological review.
Aims. The aim of this study was to compare the survivorship and radiographic
outcomes at ten-year follow-up of three prospective consecutive
series of patients each of which received a different design of
cementless femoral components for total hip arthroplasty (THA). Patients and Methods. In Cohort 1, 91 consecutive patients (100 hips) underwent THA
with a cementless
Aims. Femoral revision component subsidence has been identified as predicting early failure in revision hip surgery. This comparative cohort study assessed the potential risk factors of subsidence in two commonly used femoral implant designs. Methods. A comparative cohort study was undertaken, analyzing a consecutive series of patients following revision total hip arthroplasties using either a tapered-modular (TM) fluted titanium or a
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium
The purpose of this study was to assess the long-term results (more than ten years) of two types of cemented ulnar component with type-5 Kudo total elbow arthroplasty in a consecutive series of 56 patients (60 elbows) with rheumatoid arthritis, and to compare the results in elbows above and below a Larsen grade IV. There was no radiolucency around the humeral component. Patients in whom a metal-backed ulnar component and a
One of the most controversial issues in total knee replacement is whether or not to resurface the patella. In order to determine the effects of different designs of femoral component on the conformity of the patellofemoral joint, five different knee prostheses were investigated. These were Low Contact Stress, the Miller-Galante II, the NexGen, the
We implanted cylinders of cobalt-chrome or titanium, with smooth or porous surfaces, into rabbit bones which had been inoculated with suspensions of Staphylococcus aureus in various doses. The bacterial concentration required to produce infection of
There are few reports describing the technique
of managing acetabular protrusio in primary total hip replacement. Most
are small series with different methods of addressing the challenges
of significant medial and proximal migration of the joint centre,
deficient medial bone and reduced peripheral bony support to the
acetabular component. We describe our technique and the clinical
and radiological outcome of using impacted morsellised autograft
with a
There are limited long-term studies reporting on outcomes of the Zimmer Modular Revision (ZMR) stem, and concerns remain regarding failure. Our primary aim was to determine long-term survival free from all-cause revision and stem-related failure for this modular revision stem in revision total hip arthroplasty (THA). Secondary aims included evaluating radiological and functional outcomes. We retrospectively identified all patients in our institutional database who underwent revision THA using the ZMR system from January 2000 to December 2007. We included 106 patients (108 hips) with a mean follow-up of 14.5 years (2.3 to 22.3). Mean patient age was 69.2 years (37.0 to 89.4), and 51.9% were female (n = 55). Indications for index revision included aseptic loosening (73.1%), infection (16.7%), fracture (9.3%), and stem fracture (0.9%). Kaplan-Meier analysis was used to determine the all-cause and stem-related failure revision-free survival. At most recent follow-up, Oxford Hip Scores (OHS) were collected, and radiological stem stability was determined using the Engh classification.Aims
Methods
We reviewed 123 second-generation uncemented total hip replacements performed on 115 patients by a single surgeon between 1993 and 1994. The acetabular component used in all cases was a fully
Impaction allograft is an established method of securing initial stability of an implant in arthroplasty. Subsequent bone integration can be prolonged, and the volume of allograft may not be maintained. Intermittent administration of parathyroid hormone has an anabolic effect on bone and may therefore improve integration of an implant. Using a canine implant model we tested the hypothesis that administration of parathyroid hormone may improve osseointegration of implants surrounded by bone graft. In 20 dogs a cylindrical
Uncemented implants are now commonly used at reimplantation of a two-stage revision total hip arthoplasty (THA) following periprosthetic joint infection (PJI). However, there is a paucity of data on the performance of the most commonly used uncemented femoral implants – modular fluted tapered (MFT) femoral components – in this setting. This study evaluated implant survival, radiological results, and clinical outcomes in a large cohort of reimplantation THAs using MFT components. We identified 236 reimplantation THAs from a single tertiary care academic institution from September 2000 to September 2020. Two designs of MFT femoral components were used as part of an established two-stage exchange protocol for the treatment of PJI. Mean age at reimplantation was 65 years (SD 11), mean BMI was 32 kg/m2 (SD 7), and 46% (n = 109) were female. Mean follow-up was seven years (SD 4). A competing risk model accounting for death was used.Aims
Methods
Six
Total hip replacement for high dislocation of the hip joint remains technically difficult in terms of preparation of the true acetabulum and restoration of leg length. We describe our experience of cementless total hip replacement combined with a subtrochanteric femoral shortening osteotomy in 20 hips with Crowe grade IV dislocation with a mean follow-up of 8.1 years (4 to 11.5). There was one man and 17 women with a mean age of 55 years (44 to 69) at the time of the operation. After placment of the acetabular component at the site of the natural acetabulum, a cementless
The incidence of deep vein thrombosis in 244 patients who had total knee replacement has been studied. In 120 the prosthesis was cemented and in 124 it was cementless. In all cases the replacement was primary and a
The aim of this study was to compare the pattern of initial fixation and changes in periprosthetic bone mineral density (BMD) between patients who underwent total hip arthroplasty (THA) using a traditional fully hydroxyapatite (HA)-coated stem (T-HA group) and those with a newly introduced fully HA-coated stem (N-HA group). The study included 36 patients with T-HA stems and 30 with N-HA stems. Dual-energy X-ray absorptiometry was used to measure the change in periprosthetic BMD, one and two years postoperatively. The 3D contact between the stem and femoral cortical bone was evaluated using a density-mapping system, and clinical assessment, including patient-reported outcome measurements, was recorded.Aims
Methods
The primary outcome was investigating differences in wear, as measured by femoral head penetration, between cross-linked vitamin E-diffused polyethylene (vE-PE) and cross-linked polyethylene (XLPE) acetabular component liners and between 32 and 36 mm head sizes at the ten-year follow-up. Secondary outcomes included acetabular component migration and patient-reported outcome measures (PROMs) such as the EuroQol five-dimension questionnaire, 36-Item Short-Form Health Survey, Harris Hip Score, and University of California, Los Angeles Activity Scale (UCLA). A single-blinded, multi-arm, 2 × 2 factorial randomized controlled trial was undertaken. Patients were recruited between May 2009 and April 2011. Radiostereometric analyses (RSAs) were performed from baseline to ten years. Of the 220 eligible patients, 116 underwent randomization, and 82 remained at the ten-year follow-up. Eligible patients were randomized into one of four interventions: vE-PE acetabular liner with either 32 or 36 mm femoral head, and XLPE acetabular liner with either 32 or 36 mm femoral head. Parameters were otherwise identical except for acetabular liner material and femoral head size.Aims
Methods
Between 1986 and 1991, 106 patients (127 knees) underwent uncemented knee arthroplasty for osteoarthritis. There were 106 total knee arthroplasties and 21 medial unicompartmental knee arthroplasties. The arthroplasties were evaluated for aseptic loosening during the year 2000. For total arthroplasty we used 77
We identified five (2.3%) fractures of the stem in a series of 219 revision procedures using a cementless, cylindrical, extensively
The aim of this study was to evaluate the survival of a collarless, straight, hydroxyapatite-coated femoral stem in total hip arthroplasty (THA) at a minimum follow-up of 20 years. We reviewed the results of 165 THAs using the Omnifit HA system in 138 patients, performed between August 1993 and December 1999. The mean age of the patients at the time of surgery was 46 years (20 to 77). Avascular necrosis was the most common indication for THA, followed by ankylosing spondylitis and primary osteoarthritis. The mean follow-up was 22 years (20 to 31). At 20 and 25 years, 113 THAs in 91 patients and 63 THAs in 55 patients were available for review, respectively, while others died or were lost to follow-up. Kaplan-Meier analysis was performed to evaluate the survival of the stem. Radiographs were reviewed regularly, and the stability of the stem was evaluated using the Engh classification.Aims
Methods