The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 912 - 919
1 Aug 2023
Cunningham LJ Walton M Bale S Trail IA

Aims

Reverse total shoulder arthroplasty (rTSA) can be used in complex cases when the glenoid requires reconstruction. In this study, a baseplate with composite bone autograft and a central trabecular titanium peg was implanted, and its migration was assessed for two years postoperatively using radiostereometric analysis (RSA).

Methods

A total of 14 patients who underwent a rTSA with an autograft consented to participate. Of these, 11 had a primary rTSA using humeral head autograft and three had a revision rTSA with autograft harvested from the iliac crest. The mean age of the patients was 66 years (39 to 81). Tantalum beads were implanted in the scapula around the glenoid. RSA imaging (stereographic radiographs) was undertaken immediately postoperatively and at three, six, 12, and 24 months. Analysis was completed using model-based RSA software. Outcomes were collected preoperatively and at two years postoperatively, including the Oxford Shoulder Score, the American Shoulder and Elbow Score, and a visual analogue score for pain. A Constant score was also obtained for the assessment of strength and range of motion.


Bone & Joint Research
Vol. 9, Issue 10 | Pages 645 - 652
5 Oct 2020
Chao C Chen Y Lin J

Aims. To determine whether half-threaded screw holes in a new titanium locking plate design can substantially decrease the notch effects of the threads and increase the plate fatigue life. Methods. Three types (I to III) of titanium locking plates were fabricated to simulate plates used in the femur, tibia, and forearm. Two copies of each were fabricated using full- and half-threaded screw holes (called A and B, respectively). The mechanical strengths of the plates were evaluated according to the American Society for Testing and Materials (ASTM) F382-14, and the screw stability was assessed by measuring the screw removal torque and bending strength. Results. The B plates had fatigue lives 11- to 16-times higher than those of the A plates. Before cyclic loading, the screw removal torques were all higher than the insertion torques. However, after cyclic loading, the removal torques were similar to or slightly lower than the insertion torques (0% to 17.3%), although those of the B plates were higher than those of the A plates for all except the type III plates (101%, 109.8%, and 93.8% for types I, II, and III, respectively). The bending strengths of the screws were not significantly different between the A and B plates for any of the types. Conclusion. Removing half of the threads from the screw holes markedly increased the fatigue life of the locking plates while preserving the tightness of the screw heads and the bending strength of the locking screws. However, future work is necessary to determine the relationship between the notch sensitivity properties and titanium plate design. Cite this article: Bone Joint Res 2020;9(10):645–652


Bone & Joint Research
Vol. 6, Issue 5 | Pages 270 - 276
1 May 2017
Gosiewski JD Holsgrove TP Gill HS

Objectives. Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix. Methods. Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm. 3. , 0.16 g/cm. 3. and 0.24 g/cm. 3. ). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°. Results. The deployable crucifix design (X-Bolt), was more torsionally stable, compared to both the dynamic hip screw (DHS, p = 0.008) and helical blade (DHS Blade, p= 0.008) designs in bone substitute material representative of osteoporotic bone (0.16 g/cm. 3. polyurethane foam). In 0.08 g/cm. 3. density substrate, the crucifix design (X-Bolt) had a higher resistance to torsion than the screw (DHS, p = 0.008). There were no significant differences (p = 0.101) between the implants in 0.24 g/cm. 3. density bone substitute. Conclusions. Our findings indicate that the clinical standard proximal fracture fixator design, the screw (DHS), was the least effective at resisting torsional load, and a novel crucifix design (X-Bolt), was the most effective design in resisting torsional load in bone substitute material with density representative of osteoporotic bone. At other densities the torsional stability was also higher for the X-Bolt, although not consistently significant by statistical analysis. Cite this article: J. D. Gosiewski, T. P. Holsgrove, H. S. Gill. The efficacy of rotational control designs in promoting torsional stability of hip fracture fixation. Bone Joint Res 2017;6:270–276. DOI: 10.1302/2046-3758.65.BJR-2017-0287.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 7 | Pages 864 - 869
1 Jul 2008
Amis AA Oguz C Bull AMJ Senavongse W Dejour D

Objective patellar instability has been correlated with dysplasia of the femoral trochlea. This in vitro study tested the hypothesis that trochleoplasty would increase patellar stability and normalise the kinematics of a knee with a dysplastic trochlea. Six fresh-frozen knees were loaded via the heads of the quadriceps. The patella was displaced 10 mm laterally and the displacing force was measured from 0° to 90° of flexion. Patellar tracking was measured from 0° to 130° of knee flexion using magnetic sensors. These tests were repeated after raising the central anterior trochlea to simulate dysplasia, and repeated again after performing a trochleoplasty on each specimen. The simulated dysplasia significantly reduced stability from that of the normal knee (p < 0.001). Trochleoplasty significantly increased the stability (p < 0.001), so that it did not then differ significantly from the normal knee (p = 0.244). There were small but statistically significant changes in patellar tracking (p< 0.001). This study has provided objective biomechanical data to support the use of trochleoplasty in the treatment of patellar instability associated with femoral trochlear dysplasia


Bone & Joint 360
Vol. 10, Issue 5 | Pages 7 - 10
1 Oct 2021
Morris DLJ Cresswell T Espag M Tambe AA Clark DI Ollivere BJ


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 4 | Pages 577 - 582
1 Apr 2005
Senavongse W Amis AA

Normal function of the patellofemoral joint is maintained by a complex interaction between soft tissues and articular surfaces. No quantitative data have been found on the relative contributions of these structures to patellar stability. Eight knees were studied using a materials testing machine to displace the patella 10 mm laterally and medially and measure the force required. Patellar stability was tested from 0° to 90° knee flexion with the quadriceps tensed to 175 N. Four conditions were examined: intact, vastus medialis obliquus relaxed, flat lateral condyle, and ruptured medial retinaculae. Abnormal trochlear geometry reduced the lateral stability by 70% at 30° flexion, while relaxation of vastus medialis obliquus caused a 30% reduction. Ruptured medial retinaculae had the largest effect at 0° flexion with 49% reduction. There was no effect on medial stability. There is a complex interaction between these structures, with their contributions to loss of lateral patellar stability varying with knee flexion


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 11 | Pages 1459 - 1465
1 Nov 2009
Luites JWH Brinkman J Wymenga AB van Heerwaarden RJ

Valgus high tibial osteotomy for osteoarthritis of the medial compartment of the knee can be performed using medial opening- and lateral closing-wedge techniques. The latter have been thought to offer greater initial stability. We measured and compared the stability of opening- and closing-wedge osteotomies fixed by TomoFix plates using radiostereometry in a series of 42 patients in a prospective, randomised clinical trial. There were no differences between the opening- and closing-wedge groups in the time to regain knee function and full weight-bearing. Pain and knee function were significantly improved in both groups without any differences between them. All the osteotomies united within one year. Radiostereometry showed no clinically relevant movement of bone or differences between either group. Medial opening-wedge high tibial osteotomy secured by a TomoFix plate offers equal stability to a lateral closing-wedge technique. Both give excellent initial stability and provide significantly improved knee function and reduction in pain, although the opening-wedge technique was more likely to produce the intended correction


Bone & Joint Research
Vol. 5, Issue 10 | Pages 453 - 460
1 Oct 2016
Ernstbrunner L Werthel J Hatta T Thoreson AR Resch H An K Moroder P

Objectives. The bony shoulder stability ratio (BSSR) allows for quantification of the bony stabilisers in vivo. We aimed to biomechanically validate the BSSR, determine whether joint incongruence affects the stability ratio (SR) of a shoulder model, and determine the correct parameters (glenoid concavity versus humeral head radius) for calculation of the BSSR in vivo. Methods. Four polyethylene balls (radii: 19.1 mm to 38.1 mm) were used to mould four fitting sockets in four different depths (3.2 mm to 19.1mm). The SR was measured in biomechanical congruent and incongruent experimental series. The experimental SR of a congruent system was compared with the calculated SR based on the BSSR approach. Differences in SR between congruent and incongruent experimental conditions were quantified. Finally, the experimental SR was compared with either calculated SR based on the socket concavity or plastic ball radius. Results. The experimental SR is comparable with the calculated SR (mean difference 10%, . sd. 8%; relative values). The experimental incongruence study observed almost no differences (2%, . sd. 2%). The calculated SR on the basis of the socket concavity radius is superior in predicting the experimental SR (mean difference 10%, . sd. 9%) compared with the calculated SR based on the plastic ball radius (mean difference 42%, . sd. 55%). Conclusion. The present biomechanical investigation confirmed the validity of the BSSR. Incongruence has no significant effect on the SR of a shoulder model. In the event of an incongruent system, the calculation of the BSSR on the basis of the glenoid concavity radius is recommended. Cite this article: L. Ernstbrunner, J-D. Werthel, T. Hatta, A. R. Thoreson, H. Resch, K-N. An, P. Moroder. Biomechanical analysis of the effect of congruence, depth and radius on the stability ratio of a simplistic ‘ball-and-socket’ joint model. Bone Joint Res 2016;5:453–460. DOI: 10.1302/2046-3758.510.BJR-2016-0078.R1


Bone & Joint Research
Vol. 2, Issue 12 | Pages 264 - 269
1 Dec 2013
Antoniades G Smith EJ Deakin AH Wearing SC Sarungi M

Objective. This study compared the primary stability of two commercially available acetabular components from the same manufacturer, which differ only in geometry; a hemispherical and a peripherally enhanced design (peripheral self-locking (PSL)). The objective was to determine whether altered geometry resulted in better primary stability. Methods. Acetabular components were seated with 0.8 mm to 2 mm interference fits in reamed polyethylene bone substrate of two different densities (0.22 g/cm. 3. and 0.45 g/cm. 3. ). The primary stability of each component design was investigated by measuring the peak failure load during uniaxial pull-out and tangential lever-out tests. Results. There was no statistically significant difference in seating force (p = 0.104) or primary stability (pull-out p = 0.171, lever-out p = 0.087) of the two components in the low-density substrate. Similarly, in the high-density substrate, there was no statistically significant difference in the peak pull-out force (p = 0.154) or lever-out moment (p = 0.574) between the designs. However, the PSL component required a significantly higher seating force than the hemispherical cup in the high-density bone analogue (p = 0.006). Conclusions. Higher seating forces associated with the PSL design may result in inadequate seating and increased risk of component malpositioning or acetabular fracture in the intra-operative setting in high-density bone stock. Our results, if translated clinically, suggest that a purely hemispherical geometry may have an advantage over a peripherally enhanced geometry in high density bone stock. Cite this article: Bone Joint Res 2013;2:264–9


Bone & Joint Research
Vol. 6, Issue 5 | Pages 296 - 306
1 May 2017
Samara E Moriarty TF Decosterd LA Richards RG Gautier E Wahl P

Objectives. Thermal stability is a key property in determining the suitability of an antibiotic agent for local application in the treatment of orthopaedic infections. Despite the fact that long-term therapy is a stated goal of novel local delivery carriers, data describing thermal stability over a long period are scarce, and studies that avoid interference from specific carrier materials are absent from the orthopaedic literature. Methods. In this study, a total of 38 frequently used antibiotic agents were maintained at 37°C in saline solution, and degradation and antibacterial activity assessed over six weeks. The impact of an initial supplementary heat exposure mimicking exothermically curing bone cement was also tested as this material is commonly used as a local delivery vehicle. Antibiotic degradation was assessed by liquid chromatography coupled to mass spectrometry, or by immunoassays, as appropriate. Antibacterial activity over time was determined by the Kirby-Bauer disk diffusion assay. Results. The heat exposure mimicking curing bone cement had minimal effect on stability for most antibiotics, except for gentamicin which experienced approximately 25% degradation as measured by immunoassay. Beta-lactam antibiotics were found to degrade quite rapidly at 37°C regardless of whether there was an initial heat exposure. Excellent long-term stability was observed for aminoglycosides, glycopeptides, tetracyclines and quinolones under both conditions. Conclusions. This study provides a valuable dataset for orthopaedic surgeons considering local application of antibiotics, and for material scientists looking to develop next-generation controlled or extended-release antibiotic carriers. Cite this article: E. Samara, T. F. Moriarty, L. A. Decosterd, R. G. Richards, E. Gautier, P. Wahl. Antibiotic stability over six weeks in aqueous solution at body temperature with and without heat treatment that mimics the curing of bone cement. Bone Joint J 2017;6:296–306. DOI: 10.1302/2046-3758.65.BJR-2017-0276.R1


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 2 | Pages 173 - 177
1 Mar 1989
Rehnberg L Olerud C

We measured the stability of fixation in femoral neck fractures treated with von Bahr screws, investigated the influence of impaction and correlated peroperative stability with the clinical results. Stability was measured at operation using a metal probe fitted with strain gauges. Its tip was anchored in the subchondral bone of the femoral head and its lateral end was fixed in the lateral femoral cortex. The shearing force produced by longitudinal compression applied to the foot of the operated leg was recorded. The results in 41 consecutive patients all followed for 30 months, showed that fractures with early loosening or nonunion had all had significantly poorer stability than the fractures that had healed. Impaction improved stability in only 23 out of the 41 fractures; in the others stability had deteriorated or was unchanged


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 713 - 718
1 May 2012
Kaiser MM Zachert G Wendlandt R Eggert R Stratmann C Gros N Schulze-Hessing M Rapp M

Elastic stable intramedullary nailing (ESIN) is generally acknowledged to be the treatment of choice for displaced diaphyseal femoral fractures in children over the age of three years, although complication rates of up to 50% are described. Pre-bending the nails is recommended, but there are no published data to support this. Using synthetic bones and a standardised simulated fracture, we performed biomechanical testing to determine the influence on the stability of the fracture of pre-bending the nails before implantation. Standard ESIN was performed on 24 synthetic femoral models with a spiral fracture. In eight cases the nails were inserted without any pre-bending, in a further eight cases they were pre-bent to 30° and in the last group of eight cases they were pre-bent to 60°. Mechanical testing revealed that pre-bending to 60° produced a significant increase in the stiffness or stability of the fracture. Pre-bending to 60° showed a significant positive influence on the stiffness compared with unbent nails. Pre-bending to 30° improved stiffness only slightly. These findings validate the recommendations for pre-bending, but the degree of pre-bend should exceed 30°. Adopting higher degrees of pre-bending should improve stability in spiral fractures and reduce the complications of varus deformity and shortening


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 760 - 767
1 Jul 2019
Galea VP Rojanasopondist P Laursen M Muratoglu OK Malchau H Bragdon C

Aims. Vitamin E-diffused, highly crosslinked polyethylene (VEPE) and porous titanium-coated (PTC) shells were introduced in total hip arthroplasty (THA) to reduce the risk of aseptic loosening. The purpose of this study was: 1) to compare the wear properties of VEPE to moderately crosslinked polyethylene; 2) to assess the stability of PTC shells; and 3) to report their clinical outcomes at seven years. Patients and Methods. A total of 89 patients were enrolled into a prospective study. All patients received a PTC shell and were randomized to receive a VEPE liner (n = 44) or a moderately crosslinked polyethylene (ModXLPE) liner (n = 45). Radiostereometric analysis (RSA) was used to measure polyethylene wear and component migration. Differences in wear were assessed while adjusting for body mass index, activity level, acetabular inclination, anteversion, and head size. Plain radiographs were assessed for radiolucency and patient-reported outcome measures (PROMs) were administered at each follow-up. Results. In total, 73 patients (82%) completed the seven-year visit. Mean seven-year linear proximal penetration was -0.07 mm (. sd. 0.16) and 0.00 mm (. sd. 0.22) for the VEPE and ModXLPE cohorts, respectively (p = 0.116). PROMs (p = 0.310 to 0.807) and radiolucency incidence (p = 0.330) were not different between the polyethylene cohorts. The mean proximal shell migration rate was 0.04 mm per year (. sd. 0.09). At seven years, patients with radiolucency (34%) demonstrated greater migration (mean difference: 0.6 mm (. sd. 0.2); p < 0.001). PROMs were lower for patients with radiolucency and greater proximal migration (p = 0.009 to p = 0.045). No implants were revised for aseptic loosening. Conclusion. This is the first randomized controlled trial to report seven-year RSA results for VEPE. All wear rates were below the previously reported osteolysis threshold (0.1 mm per year). PTC shells demonstrated acceptable primary stability through seven years, as indicated by low migration and lack of aseptic loosening. However, patients with acetabular radiolucency were associated with higher shell migration and lower PROM scores. Cite this article: Bone Joint J 2019;101-B:760–767


The Journal of Bone & Joint Surgery British Volume
Vol. 46-B, Issue 4 | Pages 630 - 647
1 Nov 1964
Garden RS

1. Practical experience has shown that subcapital fractures of the femur unite freely if reduction is stable and fixation is secure. 2. Stable reduction is obtained when the muscular and gravitational forces tending to redisplace the fracture are opposed by equal and opposite counterforces, and inherent stability is believed to depend upon the integrity of the flared cortical buttress at the postero-inferior junction of the femoral neck and head. 3. In the stable subcapital fracture a state of equilibrium is reached when the forward and upward thrust of the fixation appliance in the femoral head is opposed by the counterthrust of the closely applied and cleanly broken fragments at the postero-inferior aspect of the fracture. When the postero-inferior cortical buttress is comminuted, inherent stability is lost, lateral rotation deformity recurs and the fixation device is avulsed from the cancellous bone of the head. 4. Stability may be restored by reduction in the "valgus" position, by various forms of osteotomy, by refashioning the fracture fragments or by a postero-inferiorly positioned bone graft. Theoretically, stability may also be obtained by a double lever system of fixation in which an obliquely placed fixation device or bone graft is combined with a horizontally disposed wire, pin, nail or screw crossing it anteriorly. Multilever fixation by three or more threaded wires or pins inserted at different angles and lying in contact at their point of crossing may likewise provide stability. 5. Fixation by two crossed screws has been chosen for clinical trial in 100 displaced subcapital fractures. Imperfect positioning of the screws in seven patients has been followed by early breakdown of reduction and non-union, but satisfactory positioning has been associated with radiological union in fifty patients who have been observed for twelve months or more. 6. Ultimate breakdown in some of these fractures is certain to follow avascular necrosis, and this complication has already been seen in a few patients treated by cross screw fixation more than two years ago. It is also expected that non-union will occur in some of those patients still under observation for less than a year. Even so, these preliminary findings indicate a percentage of union far greater than that obtained by previous methods of treatment, and, although statistically inadequate, they are presented in support of the belief that it should no longer be considered impossible to achieve the same percentage of union in subcapital fractures of the femur as we are accustomed to expect in the treatment of fractures elsewhere. It is not implied, however, that this ideal will be reached merely by the adoption of some form of double or multilever fixation, and much will continue to depend upon the quality of the radiographic services, the precision of reduction and the perfection of operative technique. 7. Every advance in our understanding of what is meant by "perfection of operative technique" lends increasing support to the ultimate truth of Watson-Jones's (1941) dictum: "A perfect result may be expected from a technically perfect operation; an imperfect result is due to imperfect technique." But the simple and foolproof method of fixation which will end the search for technical perfection in the treatment of the displaced subcapital fracture has yet to be evolved, and many questions remain to be answered about this injury. Nevertheless, it is clear that the surgeon should now be prepared to attribute early mechanical failure in the treatment of femoral neck fractures to his own shortcomings, and the temptation to blame capital ischaemia for every disaster should be resisted


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 7 | Pages 921 - 927
1 Jul 2005
Glyn-Jones S Gill HS Beard DJ McLardy-Smith P Murray DW

Polished, tapered stems are now widely used for cemented total hip replacement and many such designs have been introduced. However, a change in stem geometry may have a profound influence on stability. Stems with a wide, rectangular proximal section may be more stable than those which are narrower proximally. We examined the influence of proximal geometry on stability by comparing the two-year migration of the Exeter stem with a more recent design, the CPS-Plus, which has a wider shoulder and a more rectangular cross-section. The hypothesis was that these design features would increase rotational stability. Both stems subsided approximately 1 mm relative to the femur during the first two years after implantation. The Exeter stem was found to rotate into valgus (mean 0.2°, . sd. 0.42°) and internally rotate (mean 1.28°, . sd. 0.99°). The CPS-Plus showed no significant valgus rotation (mean 0.2°, . sd. 0.42°) or internal rotation (mean −0.03°, . sd. 0.75°). A wider, more rectangular cross-section improves rotational stability and may have a better long-term outcome


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1662 - 1668
1 Nov 2021
Bhanushali A Chimutengwende-Gordon M Beck M Callary SA Costi K Howie DW Solomon LB

Aims

The aims of this study were to compare clinically relevant measurements of hip dysplasia on radiographs taken in the supine and standing position, and to compare Hip2Norm software and Picture Archiving and Communication System (PACS)-derived digital radiological measurements.

Methods

Preoperative supine and standing radiographs of 36 consecutive patients (43 hips) who underwent periacetabular osteotomy surgery were retrospectively analyzed from a single-centre, two-surgeon cohort. Anterior coverage (AC), posterior coverage (PC), lateral centre-edge angle (LCEA), acetabular inclination (AI), sharp angle (SA), pelvic tilt (PT), retroversion index (RI), femoroepiphyseal acetabular roof (FEAR) index, femoroepiphyseal horizontal angle (FEHA), leg length discrepancy (LLD), and pelvic obliquity (PO) were analyzed using both Hip2Norm software and PACS-derived measurements where applicable.


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 3 | Pages 372 - 376
1 May 1992
Curtis M Jinnah R Wilson V Hungerford D

The osseo-integration of an uncemented acetabular component depends on its initial stability. This is usually provided by under-reaming of the acetabulum. We have assessed the fixation of 52 mm porous-coated hemispherical prostheses inserted into cadaveric acetabula under-reamed by 1, 2, 3 and 4 mm. We tested the torsional stability of fixation, after preloading with 686 N in compression, by measuring the torque required to produce 1 degree and 2 degrees of rotation. Under-reaming by 2 mm and 3 mm gave significantly better fixation than 1 mm (p less than 0.01, p less than 0.02). Insertion after under-reaming of 4 mm caused some fractures. To obtain maximum interference fit and optimal implant stability, we recommend the use of an implant 2 mm or 3 mm larger than the last reamer


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 1 | Pages 24 - 26
1 Jan 1999
Vyskocil P Gerber C Bamert P

The radiolucent lines and the stability of the components of 66 knee arthroplasties were assessed by six orthopaedic surgeons on conventional anteroposterior and lateral radiographs and on fluoroscopic views which had been taken on the same day. The examiners were blinded as to the patients and clinical results. The interpretation of the radiographs was repeated after five months. On fluoroscopically-assisted radiographs four of the six examiners identified significantly more radiolucent lines for the femoral component (p < 0.05) and one significantly more for the tibial implant. Five examiners rated more femoral components as radiologically loose on fluoroscopically-assisted radiographs (p = 0.0008 to 0.0154), but none did so for the tibial components. The mean intra- and interobserver kappa values were higher for fluoroscopically-assisted radiographs for both components. We have shown that fluoroscopically-assisted radiographs allow more reproducible, and therefore reliable, detection of radiolucent lines in total knee arthroplasty. Assessment of the stability of the components is significantly influenced by the radiological technique used. Conventional radiographs are not adequate for evaluation of the stability of total knee arthroplasty and should be replaced by fluoroscopically-assisted films


The Bone & Joint Journal
Vol. 97-B, Issue 5 | Pages 642 - 648
1 May 2015
Hunt NC Ghosh KM Blain AP Rushton SP Longstaff LM Deehan DJ

The aim of this study was to compare the maximum laxity conferred by the cruciate-retaining (CR) and posterior-stabilised (PS) Triathlon single-radius total knee arthroplasty (TKA) for anterior drawer, varus–valgus opening and rotation in eight cadaver knees through a defined arc of flexion (0º to 110º). The null hypothesis was that the limits of laxity of CR- and PS-TKAs are not significantly different.

The investigation was undertaken in eight loaded cadaver knees undergoing subjective stress testing using a measurement rig. Firstly the native knee was tested prior to preparation for CR-TKA and subsequently for PS-TKA implantation. Surgical navigation was used to track maximal displacements/rotations at 0º, 30º, 60º, 90º and 110° of flexion. Mixed-effects modelling was used to define the behaviour of the TKAs.

The laxity measured for the CR- and PS-TKAs revealed no statistically significant differences over the studied flexion arc for the two versions of TKA. Compared with the native knee both TKAs exhibited slightly increased anterior drawer and decreased varus-valgus and internal-external roational laxities. We believe further study is required to define the clinical states for which the additional constraint offered by a PS-TKA implant may be beneficial.

Cite this article: Bone Joint J 2015; 97-B:642–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 86 - 92
1 Jan 2012
Jeon IH Sanchez-Sotelo J Zhao K An KN Morrey BM

We undertook this study to determine the minimum amount of coronoid necessary to stabilise an otherwise intact elbow joint. Regan–Morrey types II and III, plus medial and lateral oblique coronoid fractures, collectively termed type IV fractures, were simulated in nine fresh cadavers. An electromagnetic tracking system defined the three-dimensional stability of the ulna relative to the humerus. The coronoid surface area accounts for 59% of the anterior articulation. Alteration in valgus, internal and external rotation occurred only with a type III coronoid fracture, accounting for 68% of the coronoid and 40% of the entire articular surface. A type II fracture removed 42% of the coronoid articulation and 25% of the entire articular surface but was associated with valgus and external rotational changes only when the radial head was removed, thereby removing 67% of the articular surface. We conclude that all type III fractures, as defined here, are unstable, even with intact ligaments and a radial head. However, a type II deficiency is stable unless the radial head is removed. Our study suggests that isolated medial-oblique or lateral-oblique fractures, and even a type II fracture with intact ligaments and a functional radial head, can be clinically stable, which is consistent with clinical observation.