Ankle replacements have improved significantly since the first reported attempt at resurfacing of the talar dome in 1962. We are now at a stage where ankle replacement offers a viable option in the treatment of end-stage ankle arthritis. As the procedure becomes more successful, it is important to reflect and review the current surgical outcomes. This allows us to guide our patients in the treatment of end-stage ankle arthritis. What is the better surgical treatment – arthrodesis or replacement?
The Advance Medial-Pivot total knee arthroplasty (TKA) was designed
to reflect contemporary data regarding the kinematics of the knee.
We wished to examine the long-term results obtained with this prosthesis
by extending a previous evaluation. We retrospectively evaluated prospectively collected data from
225 consecutive patients (41 men and 184 women; mean age at surgery
71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up
of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both
subjective and objective) and radiological outcome were assessed.
Pre- and post-operative clinical and radiographic data were available
at regular intervals for all patients. A total of ten patients (4.4%;
ten TKAs) were lost to follow-up.Aims
Patients and Methods
The term mid-flexion instability has entered
the orthopaedic literature as a concept, but has not been confirmed
as a distinct clinical entity. The term is used freely, sometimes
as a synonym for flexion instability. However, the terms need to
be clearly separated. A cadaver study published in 1990 associated
joint line elevation with decreased stability at many angles of
flexion, but that model was not typical of clinical scenarios. The
literature is considered and it is proposed that the more common
entity of an uncorrected flexion contracture after a measured resection arthroplasty
technique is more likely to produce clinical findings that suggest
instability mid-flexion. It is proposed that the clinical scenario encountered is generalised
instability, with the appearance of stability in full extension
from tight posterior structures. This paper seeks to clarify whether mid-flexion instability exists
as an entity distinct from other commonly recognised forms of instability. Cite this article:
We conducted a randomised controlled trial to assess the accuracy
of positioning and alignment of the components in total knee arthroplasty
(TKA), comparing those undertaken using standard intramedullary
cutting jigs and those with patient-specific instruments (PSI). There were 64 TKAs in the standard group and 69 in the PSI group. The post-operative hip-knee-ankle (HKA) angle and positioning
was investigated using CT scans. Deviation of >
3° from the planned
position was regarded as an outlier. The operating time, Oxford
Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.Aims
Patients and Methods
Our aim was to examine the clinical and radiographic outcomes
in 257 consecutive Oxford unicompartmental knee arthroplasties (OUKAs)
(238 patients), five years post-operatively. A retrospective evaluation was undertaken of patients treated
between April 2008 and October 2010 in a regional centre by two
non-designing surgeons with no previous experience of UKAs. The
Oxford Knee Scores (OKSs) were recorded and fluoroscopically aligned
radiographs were assessed post-operatively at one and five years.Aims
Patients and Methods
A retrospective review was performed of patients
undergoing primary cementless total knee replacement (TKR) using
porous tantalum performed by a group of surgical trainees. Clinical
and radiological follow-up involved 79 females and 26 males encompassing
115 knees. The mean age was 66.9 years (36 to 85). Mean follow-up
was 7 years (2 to 11). Tibial and patellar components were porous
tantalum monoblock implants, and femoral components were posterior
stabilised (PS) in design with cobalt–chromium fibre mesh. Radiological
assessments were made for implant positioning, alignment, radiolucencies,
lysis, and loosening. There was 95.7% survival of implants. There
was no radiological evidence of loosening and no osteolysis found.
No revisions were performed for aseptic loosening. Average tibial
component alignment was 1.4° of varus (4°of valgus to 9° varus),
and 6.2° (3° anterior to 15° posterior) of posterior slope. Mean
femoral component alignment was 6.6° (1° to 11°) of valgus. Mean tibiofemoral
alignment was 5.6° of valgus (7° varus to 16° valgus). Patellar
tilt was a mean of 2.4° lateral (5° medial to 28° lateral). Patient
satisfaction with improvement in pain was 91%. Cementless TKR incorporating
porous tantalum yielded good clinical and radiological outcomes
at a mean of follow-up of seven-years. Cite this article:
Post-operative migration of cemented acetabular components as
measured by radiostereometric analysis (RSA) has a strong predictive
power for late, aseptic loosening. Also, radiolucent lines predict
late loosening. Migration has been reduced by systemic bisphosphonate
treatment in randomised trials of hip and knee arthroplasty. Used
as a local treatment, a higher local dose of bisphosphonate can
be achieved without systemic exposure. We wished to see if this
principle could be applied usefully in total hip arthroplasty (THA). In this randomised placebo-controlled, double-blinded trial with
60 participants, we compressed gauze soaked in bisphosphonate solution
(ibandronate) or saline against the acetabular bone bed immediately
before cementing the acetabular component. RSA, classification of
radiolucent lines, the Harris Hip Score (HHS) and the Western Ontario McMasters
Universities Osteoarthritis Index (WOMAC) were carried out at three-,
six-, 12-, and 24-month follow-up.Aims
Patients and Methods
The aim of this study was to evaluate the risk
factors for dislocation of the bearing after a mobile-bearing Oxford medial
unicompartmental knee replacement (UKR) and to test the hypothesis
that surgical factors, as measured from post-operative radiographs,
are associated with its dislocation From a total of 480 UKRs performed between 2001 and 2012, in
391 patients with a mean age of 66.5 years (45 to 82) (316 female,
75 male), we identified 17 UKRs where bearing dislocation occurred.
The post-operative radiological measurements of the 17 UKRs and
51 matched controls were analysed using conditional logistic regression analysis.
The post-operative radiological measurements included post-operative
change in limb alignment, the position of the femoral and tibial
components, the resection depth of the proximal tibia, and the femoral component-posterior
condyle classification. We concluded that a post-operative decrease in the posterior
tibial slope relative to the pre-operative value was the only significant
determinant of dislocation of the bearing after medial Oxford UKR
(odds ratio 1.881; 95% confidence interval 1.272 to 2.779). A post-operative
posterior tibial slope <
8.45° and a difference between the pre-operative
and post-operative posterior tibial slope of >
2.19° may increase
the risk of dislocation. Cite this article:
The October 2014 Knee Roundup360 looks at: microfracture equivalent to OATS; examination better than MRI in predicting hamstrings re-injury; a second view on return to play with hamstrings injuries; dislocation risks in the Oxford Unicompartmental Knee; what about the tibia?; getting on top of lateral facet pain post TKR; readmission in TKR; patient-specific instrumentation; treating infrapatellar saphenous neuralgia; and arthroscopy in the middle-aged.
Unicompartmental knee arthroplasty (UKA) is a potential treatment
for isolated bone on bone osteoarthritis when limited to a single
compartment. The risk for revision of UKA is three times higher
than for total knee arthroplasty (TKA). The aim of this review was
to discuss the different revision options after UKA failure. A search was performed for English language articles published
between 2006 and 2016. After reviewing titles and abstracts, 105
papers were selected for further analysis. Of these, 39 papers were
deemed to contain clinically relevant data to be included in this review.Objectives
Materials and Methods
Valgus knee deformity can present a number of
unique surgical challenges for the total knee arthroplasty (TKA)
surgeon. Understanding the typical patterns of bone and soft-tissue pathology
in the valgus arthritic knee is critical for appropriate surgical
planning. This review aims to provide the knee arthroplasty surgeon
with an understanding of surgical management strategies for the
treatment of valgus knee arthritis. Lateral femoral and tibial deficiencies, contracted lateral soft
tissues, attenuated medial soft tissues, and multiplanar deformities
may all be present in the valgus arthritic knee. A number of classifications
have been reported in order to guide surgical management, and a variety
of surgical strategies have been described with satisfactory clinical
results. Depending on the severity of the deformity, a variety of
TKA implant designs may be appropriate for use. Regardless of an operating surgeon’s preferred surgical strategy,
adherence to a step-wise approach to deformity correction is advised. Cite this article:
Collateral ligament release is advocated in total
knee arthroplasty (TKA) to deal with significant coronal plane deformities,
but is also associated with significant disadvantages. We describe steps to avoid release of the collateral (superficial
medial and lateral collateral) ligaments during TKA in severely
deformed knees, while correcting deformity and balancing the knee. Cite this article:
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
Patient specific instrumentation (PSI) uses advanced
imaging of the knee (CT or MRI) to generate individualised cutting
blocks aimed to make the procedure of total knee arthroplasty (TKA)
more accurate and efficient. However, in this era of healthcare
cost consciousness, the value of new technologies needs to be critically
evaluated. There have been several comparative studies looking at
PSI Cite this article:
We have previously reported the short-term radiological
results of a randomised controlled trial comparing kinematically
aligned total knee replacement (TKR) and mechanically aligned TKR,
along with early pain and function scores. In this study we report
the two-year clinical results from this trial. A total of 88 patients
(88 knees) were randomly allocated to undergo either kinematically
aligned TKR using patient-specific guides, or mechanically aligned
TKR using conventional instruments. They were analysed on an intention-to-treat
basis. The patients and the clinical evaluator were blinded to the
method of alignment. At a minimum of two years, all outcomes were better for the kinematically
aligned group, as determined by the mean Oxford knee score (40 (15
to 48) In this study, the use of a kinematic alignment technique performed
with patient-specific guides provided better pain relief and restored
better function and range of movement than the mechanical alignment
technique performed with conventional instruments. Cite this article:
Because there have been no standard methods to determine pre-operatively
the thickness of resection of the proximal tibia in unicompartmental
knee arthroplasty (UKA), information about the relationship between
the change of limb alignment and the joint line elevation would
be useful for pre-operative planning. The purpose of this study
was to clarify the correlation between the change of limb alignment
and the change of joint line height at the medial compartment after
UKA. A consecutive series of 42 medial UKAs was reviewed retrospectively.
These patients were assessed radiographically both pre- and post-operatively
with standing anteroposterior radiographs. The thickness of bone
resection at the proximal tibia and the distal femur was measured
radiographically. The relationship between the change of femorotibial
angle (δFTA) and the change of joint line height, was analysed.Objectives
Methods
Instability in flexion after total knee replacement
(TKR) typically occurs as a result of mismatched flexion and extension
gaps. The goals of this study were to identify factors leading to
instability in flexion, the degree of correction, determined radiologically,
required at revision surgery, and the subsequent clinical outcomes.
Between 2000 and 2010, 60 TKRs in 60 patients underwent revision
for instability in flexion associated with well-fixed components.
There were 33 women (55%) and 27 men (45%); their mean age was 65
years (43 to 82). Radiological measurements and the Knee Society
score (KSS) were used to assess outcome after revision surgery.
The mean follow-up was 3.6 years (2 to 9.8). Decreased condylar
offset (p <
0.001), distalisation of the joint line (p <
0.001)
and increased posterior tibial slope (p <
0.001) contributed
to instability in flexion and required correction at revision to regain
stability. The combined mean correction of posterior condylar offset
and joint line resection was 9.5 mm, and a mean of 5° of posterior
tibial slope was removed. At the most recent follow-up, there was
a significant improvement in the mean KSS for the knee and function
(both p <
0.001), no patient reported instability and no patient
underwent further surgery for instability. The following step-wise approach is recommended: reduction of
tibial slope, correction of malalignment, and improvement of condylar
offset. Additional joint line elevation is needed if the above steps
do not equalise the flexion and extension gaps. Cite this article:
The aim of the current study was to analyse the effects of posterior cruciate ligament (PCL) deficiency on forces of the posterolateral corner structure and on tibiofemoral (TF) and patellofemoral (PF) contact force under dynamic-loading conditions. A subject-specific knee model was validated using a passive flexion experiment, electromyography data, muscle activation, and previous experimental studies. The simulation was performed on the musculoskeletal models with and without PCL deficiency using a novel force-dependent kinematics method under gait- and squat-loading conditions, followed by probabilistic analysis for material uncertain to be considered.Objectives
Methods
In this study we randomised 140 patients who
were due to undergo primary total knee arthroplasty (TKA) to have the
procedure performed using either patient-specific cutting guides
(PSCG) or conventional instrumentation (CI). The primary outcome measure was the mechanical axis, as measured
at three months on a standing long-leg radiograph by the hip–knee–ankle
(HKA) angle. This was undertaken by an independent observer who
was blinded to the instrumentation. Secondary outcome measures were
component positioning, operating time, Knee Society and Oxford knee
scores, blood loss and length of hospital stay. A total of 126 patients (67 in the CI group and 59 in the PSCG
group) had complete clinical and radiological data. There were 88
females and 52 males with a mean age of 69.3 years (47 to 84) and
a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA
angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4
to 183.4) in the PSCG group (p = 0.34). Outliers were identified
in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%)
in the PSCG group (p = 0.99). There was no significant difference
in the clinical results (p = 0.95 and 0.59, respectively). Operating time,
blood loss and length of hospital stay were not significantly reduced
(p = 0.09, 0.58 and 0.50, respectively) when using PSCG. The use of PSCG in primary TKA did not reduce the proportion
of outliers as measured by post-operative coronal alignment. Cite this article: