Advertisement for orthosearch.org.uk
Results 421 - 440 of 457
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1291 - 1297
1 Oct 2008
De Haan R Pattyn C Gill HS Murray DW Campbell PA De Smet K

We examined the relationships between the serum levels of chromium and cobalt ions and the inclination angle of the acetabular component and the level of activity in 214 patients implanted with a metal-on-metal resurfacing hip replacement. Each patient had a single resurfacing and no other metal in their body. All serum measurements were performed at a minimum of one year after operation. The inclination of the acetabular component was considered to be steep if the abduction angle was greater than 55°.

There were significantly higher levels of metal ions in patients with steeply-inclined components (p = 0.002 for chromium, p = 0.003 for cobalt), but no correlation was found between the level of activity and the concentration of metal ions. A highly significant (p < 0.001) correlation with the arc of cover was found. Arcs of cover of less than 10 mm were correlated with a greater risk of high concentrations of serum metal ions. The arc of coverage was also related to the design of the component and to size as well as to the abduction angle of the acetabular component. Steeply-inclined acetabular components, with abduction angles greater than 55°, combined with a small size of component are likely to give rise to higher serum levels of cobalt and chromium ions. This is probably due to a greater risk of edge-loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1183 - 1190
1 Sep 2009
Kim BS Choi WJ Kim YS Lee JW

Our study describes the clinical outcome of total ankle replacement (TAR) performed in patients with moderate to severe varus deformity. Between September 2004 and September 2007, 23 ankles with a varus deformity ≥ 10° and 22 with neutral alignment received a TAR. Following specific algorithms according to joint congruency, the varus ankles were managed by various additional procedures simultaneously with TAR.

After a mean follow-up of 27 months (12 to 47), the varus ankles improved significantly in all clinical measures (p < 0.0001 for visual analogue scale and American Orthopaedic Foot and Ankle Society score, p = 0.001 for range of movement). No significant differences were found between the varus and neutral groups regarding the clinical (p = 0.766 for visual analogue scale, p = 0.502 for American Orthopaedic Foot and Ankle Society score, p = 0.773 for range of movement) and radiological outcome (p = 0.339 for heterotopic ossification, p = 0.544 for medial cortical reaction, p = 0.128 for posterior focal osteolysis). Failure of the TAR with conversion to an arthrodesis occurred in one case in each group.

The clinical outcome of TAR performed in ankles with pre-operative varus alignment ≥ 10° is comparable with that of neutrally aligned ankles when appropriate additional procedures to correct the deformity are carried out simultaneously with TAR.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1060 - 1063
1 Aug 2007
Singh PJ Perera NS Dega R

We carried out a prospective study over a period of 12 months to measure the exposure to radiation of the hands of a dedicated foot and ankle surgeon. A thermoluminescent dosimeter ring (TLD) was used to measure the cumulative dose of radiation. Fluoroscopy was used in operations on the foot and ankle. The total screening time was 3028 s, with a mean time per procedure of 37.4 s (0.6 to 197). This correlated positively with the number of procedures performed (r = 0.92, p < 0.001), and with the dose of radiation in both the left (r = 0.85, p = 0.0005) and right TLDs (r = 0.59, p = 0.419). There was no significant difference in the dose of radiation between the two hands (t-test, p = 0.62). The total dose to the right TLD over the 12 months was 2.4 millisieverts.

This is a simple and convenient method for evaluating the exposure of a single surgeon to radiation. The radiation detected was well below the annual dose limit set by the International Commission on Radiological Protection.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 430 - 435
1 Apr 2008
Biant LC Bruce WJM Assini JB Walker PM Walsh WR

This prospective study presents the ten-year (5 to 16) clinical and radiological results of 55 primary total hip replacements (THR) using a cementless modular femoral component (S-ROM). All patients had a significant anatomical abnormality which rendered the primary THR difficult.

The mean Harris hip score was 36 (12 to 72) pre-operatively, 83 (44 to 100) at five years, and 85 (45 to 99) at ten years. The Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and short-form (SF)-12 scores were recorded from the year 2000. The mean SF-12 score at five years after surgery was 45.24 (22.74 to 56.58) for the physical component and 54.14 (29.20 to 66.61) for the mental component. By ten years the SF-12 scores were 42.86 (21.59 to 58.95) and 51.03 (33.78 to 61.40), respectively. The mean WOMAC score at five years post-operatively was 25 (0 to 59), and at ten years was 27 (2 to 70).

No femoral components were radiologically loose, although five had osteolysis in Gruen zone 1, three had osteolysis in zone 7, and two showed osteolysis in both zones 1 and 7. No osteolysis was observed around or distal to the prosthetic sleeve. No femoral components were revised, although three hips underwent an acetabular revision and two required a liner exchange. At a mean of ten years’ follow-up the S-ROM femoral component implanted for an anatomically difficult primary THR has excellent clinical and radiological results.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 11 - 15
1 Jan 2008
Jaiswal PK Jagiello J David LA Blunn G Carrington RWJ Skinner JA Cannon SR Briggs TWR

We have managed 27 patients (16 women and 11 men) with a mean age of 68.4 years (50 to 84), with failed total hip replacement and severe proximal femoral bone loss by revision using a distal fix/proximal wrap prosthesis. The mean follow-up was for 55.3 months (25 to 126). The mean number of previous operations was 2.2 (1 to 4). The mean Oxford hip score decreased from 46.2 (38 to 60) to 28.5 (17 to 42) (paired t-test, p < 0.001) and the mean Harris Hip score increased from 30.4 (3 to 57.7) to 71.7 (44 to 99.7) (paired t-test, p < 0.001). There were two dislocations, and in three patients we failed to eradicate previous infection. None required revision of the femoral stem.

This technique allows instant distal fixation while promoting biological integration and restoration of bone stock. In the short term, the functional outcome is encouraging and the complication rates acceptable in this difficult group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 556 - 561
1 May 2008
Glyn-Jones S McLardy-Smith P Gill HS Murray DW

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 4 | Pages 436 - 441
1 Apr 2008
Steffen RT Pandit HP Palan J Beard DJ Gundle R McLardy-Smith P Murray DW Gill HS

Few independent studies have reported the outcome of resurfacing arthroplasty of the hip. The aim of this study was to report the five-year clinical outcome and seven-year survival of an independent series.

A total of 610 Birmingham Hip Resurfacing arthroplasties were performed in 532 patients with a mean age of 51.8 years (16.5 to 81.6). They were followed for between two and eight years; 107 patients (120 hips) had been followed up for more than five years. Two patients were lost to follow-up. At a minimum of five years’ follow-up, 79 of 85 hips (93%) had an excellent or good outcome according to the Harris hip score. The mean Oxford hip score was 16.1 points (sd 7.7) and the mean University of California Los Angeles activity score was 6.6 points (sd 1.9). There were no patients with definite radiological evidence of loosening or of narrowing of the femoral neck exceeding 10% of its width. There were 23 revisions (3.8%), giving an overall survival of 95% (95% confidence interval 85.3 to 99.2) at seven years. Fractured neck of femur in 12 hips was the most common indication for revision, followed by aseptic loosening in four. In three hips (three patients) (0.5%), failure was possibly related to metal debris.

Considering that these patients are young and active these results are good, and support the use of resurfacing. Further study is needed to address the early failures, particularly those related to fracture and metal debris.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 738 - 744
1 Jun 2008
Pang H Lo N Yang K Chong H Yeo S

We have performed a prospective double-blind, randomised controlled trial over two years to evaluate the efficacy and safety of an intra-operative peri-articular injection of triamcinolone acetonide in patients undergoing medial unicondylar knee replacement. We randomised 90 patients into two equal groups. The study group received an injection of triamcinolone acetonide, bupivacaine, and epinephrine into the peri-articular tissues at the end of the operation. The control group received the same injection mixture but without the addition of triamcinolone. The peri-operative analgesic regimen was standardised.

The study group reported a significant reduction in pain (p = 0.014 at 12 hours, p = 0.031 at 18 hours and p = 0.031 at 24 hours) and had a better range of movement (p = 0.023 at three months). There was no significant difference in the rate of infection and no incidence of tendon rupture in either group.

The addition of corticosteroid to the peri-articular injection after unicondylar knee replacement had both immediate and short-term benefits in terms of relief from pain, and rehabilitation with no increased risk of infection.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 3 | Pages 291 - 295
1 Mar 2007
Langdown AJ Pickard RJ Hobbs CM Clarke HJ Dalton DJN Grover ML

We reviewed the initial post-operative radiographs of the Trident acetabulum and identified a problem with seating of the metal-backed ceramic liner. We identified 117 hips in 113 patients who had undergone primary total hip replacement using the Trident shell with a metal-backed alumina liner. Of these, 19 (16.4%) were noted to have incomplete seating of the liner, as judged by plain anteroposterior and lateral radiographs. One case of complete liner dissociation necessitating early revision was not included in the prevalence figures. One mis-seated liner was revised in the early post-operative period and two that were initially incompletely seated were found on follow-up radiographs to have become correctly seated. There may be technical issues with regard to the implanting of this prosthesis of which surgeons should be aware. However, there is the distinct possibility that the Trident shell deforms upon implantation, thereby preventing complete seating of the liner.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1303 - 1308
1 Oct 2007
Nixon M Taylor G Sheldon P Iqbal SJ Harper W

We matched 78 patients with a loose cemented Charnley Elite Plus total hip replacement (THR) by age, gender, race, prosthesis and time from surgery with 49 patients with a well-fixed stable hip replacement, to determine if poor bone quality predisposes to loosening. Clinical, radiological, biomechanical and bone mineral density indicators of bone quality were assessed.

Patients with loose replacements had more pain, were more likely to have presented with atrophic arthritis and to have a history of fragility fracture, narrower femoral cortices and lower peri-prosthetic or lumbar spine bone mineral density (all t-test, p < 0.01). They also tended to be smokers (chi-squared test, p = 0.08). Vitamin-D deficiency was common, but not significantly different between the two groups (t-test, p = 0.31)

In this series of cemented hip replacements performed between 1994 and 1998, aseptic loosening was associated with poor bone quality. Patients with a THR should be screened for osteoporosis and have regular radiological surveillance.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 8 | Pages 1019 - 1024
1 Aug 2007
Hing CB Young DA Dalziel RE Bailey M Back DL Shimmin AJ

Narrowing of the femoral neck after resurfacing arthroplasty of the hip has been described previously in both cemented and uncemented hip resurfacing. The natural history of narrowing of the femoral neck is unknown. We retrospectively measured the diameter of the femoral neck in a series of 163 Birmingham hip resurfacings in 163 patients up to a maximum of six years after operation to determine the extent and progression of narrowing.

There were 105 men and 58 women with a mean age of 52 years (18 to 82). At a mean follow-up of five years, the mean Harris hip score was 94.8 (47 to 100) and the mean flexion of the hip 112.5° (80° to 160°). There was some narrowing of the femoral neck in 77% (125) of the patients reviewed, and in 27.6% (45) the narrowing exceeded 10% of the diameter of the neck. A multiple logistic regression analysis showed a significant association (chi-squared test (derived from logistic regression) p = 0.01) of narrowing with female gender and a valgus femoral neck/shaft angle. There was no significant association between the range of movement, position or size of the component or radiological lucent lines and narrowing of the neck (chi-squared test; p = 0.10 (flexion), p = 0.08 (size of femoral component), p = 0.09 (size of acetabular component), p = 0.71 (femoral component angulation), p = 0.99 (lucent lines)). There was no significant difference between the diameter of the neck at a mean of three years (2.5 to 3.5) and that at five years (4.5 to 5.5), indicating that any change in the diameter of the neck had stabilised by three years (sign rank test, p = 0.60).

We conclude that narrowing of the femoral neck which is found with the Birmingham hip resurfacing arthroplasty is in most cases associated with no adverse clinical or radiological outcome up to a maximum of six years after the initial operation.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 299 - 305
1 Mar 2008
Kim Y

This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading.

A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12).

The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur.

This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1046 - 1049
1 Aug 2005
Shepperd JAN Apthorp H


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1073 - 1076
1 Aug 2005
Cross MJ Parish EN

We prospectively reviewed 1000 consecutive patients who underwent a cementless, hydroxyapatite-coated, stemless, total knee replacement over a period of nine years. Regular post-operative clinical follow-up was performed using the Knee Society score. The mean pre-operative score was 96, improving to 182 and 180 at five and ten years, respectively. To date, there have been seven (0.5%) cases which required revision, primarily for septic loosening (four cases), with low rates of other post-operative complications. The cumulative survival at ten years with revision as the end-point, was 99.14% (95% confidence interval 92.5 to 99.8). These results support the use of hydroxyapatite in a cementless total knee replacement since it can give reliable fixation with an excellent clinical and functional outcome.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 701 - 705
1 Jun 2006
Simpson AHRW Mills L Noble B


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 107 - 113
1 Jan 2008
Scheerlinck T Vandenbussche P Noble PC

Interfacial defects between the cement mantle and a hip implant may arise from constrained shrinkage of the cement or from air introduced during insertion of the stem. Shrinkage-induced interfacial porosity consists of small pores randomly located around the stem, whereas introduced interfacial gaps are large, individual and less uniformly distributed areas of stem-cement separation. Using a validated CT-based technique, we investigated the extent, morphology and distribution of interfacial gaps for two types of stem, the Charnley-Kerboul and the Lubinus SPII, and for two techniques of implantation, line-to-line and undersized.

The interfacial gaps were variable and involved a mean of 6.43% (sd 8.99) of the surface of the stem. Neither the type of implant nor the technique of implantation had a significant effect on the regions of the gaps, which occurred more often over the flat areas of the implant than along the corners of the stems, and were more common proximally than distally for Charnley-Kerboul stems cemented line-to-line. Interfacial defects could have a major effect on the stability and survival of the implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known.

At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%).

The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 9 - 15
1 Jan 2007
Beaulé PE Harvey N Zaragoza E Le Duff MJ Dorey FJ

Because the femoral head/neck junction is preserved in hip resurfacing, patients may be at greater risk of impingement, leading to abnormal wear patterns and pain. We assessed femoral head/neck offset in 63 hips undergoing metal-on-metal hip resurfacing and in 56 hips presenting with non-arthritic pain secondary to femoroacetabular impingement. Most hips undergoing resurfacing (57%; 36) had an offset ratio ≤ 0.15 pre-operatively and required greater correction of offset at operation than the rest of the group. In the non-arthritic hips the mean offset ratio was 0.137 (0.04 to 0.23), with the offset ratio correlating negatively to an increasing α angle. An offset ratio ≤ 0.15 had a 9.5-fold increased relative risk of having an α angle ≥ 50.5°. Most hips undergoing resurfacing have an abnormal femoral head/neck offset, which is best assessed in the sagittal plane.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 10 | Pages 1279 - 1285
1 Oct 2006
Parvizi J Campfield A Clohisy JC Rothman RH Mont MA

Arthritis of the hip in the young adult can be a disabling condition. Recent years have witnessed extensive research related to the management of this condition. This article reviews the current status with regard to aetiology, diagnosis and treatment of arthritis of the hip in the young adult.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 9 | Pages 1143 - 1151
1 Sep 2008
Langton DJ Jameson SS Joyce TJ Webb J Nargol AVF

Increased concentrations of metal ions after metal-on-metal resurfacing arthroplasty of the hip remain a concern. Although there has been no proven link to long-term health problems or early prosthetic failure, variables associated with high metal ion concentrations should be identified and, if possible, corrected. Our study provides data on metal ion levels from a series of 76 consecutive patients (76 hips) after resurfacing arthroplasty with the Articular Surface Replacement. Chromium and cobalt ion concentrations in the whole blood of patients with smaller (≤ 51 mm) femoral components were significantly higher than in those with the larger (≥ 53 mm) components (p < 0.01). Ion concentrations in the former group were significantly related to the inclination (p = 0.01) and anteversion (p = 0.01) of the acetabular component. The same relationships were not significant in the patients with larger femoral components (p = 0.61 and p = 0.49, respectively). Accurate positioning of the acetabular component intra-operatively is essential in order to reduce the concentration of metal ions in the blood after hip resurfacing arthroplasty with the Articular Surface Replacement implant.