Advertisement for orthosearch.org.uk
Results 421 - 440 of 490
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 12 | Pages 1592 - 1596
1 Dec 2011
Babis GC Sakellariou VI Chatziantoniou AN Soucacos PN Megas P

We report the results of 62 hips in 62 patients (17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent revision of the acetabular component of a total hip replacement due to aseptic loosening between May 2003 and November 2007. All hips had a Paprosky type IIIa acetabular defect. Acetabular revision was undertaken using a Procotyl E cementless oblong implant with modular side plates and a hook combined with impaction allografting.

At a mean follow-up of 60.5 months (36 to 94) with no patients lost to follow-up and one died due to unrelated illness, the complication rate was 38.7%. Complications included aseptic loosening (19 hips), deep infection (3 hips), broken hook and side plate (one hip) and a femoral nerve palsy (one hip). Further revision of the acetabular component was required in 18 hips (29.0%) and a further four hips (6.4%) are currently loose and awaiting revision.

We observed unacceptably high rates of complication and failure in our group of patients and cannot recommend this implant or technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 281 - 286
1 Feb 2012
Chana R Esposito C Campbell PA Walter WK Walter WL

We report a case of a male patient presenting with bilateral painful but apparently well-positioned and -fixed large-diameter metal-on-metal hip replacements four years post-operatively. Multiple imaging modes revealed a thick-walled, cystic expansile mass in communication with the hip joint (a pseudotumour). Implant retrieval analysis and tissue culture eliminated high bearing wear or infection as causes for the soft-tissue reaction, but noted marked corrosion of the modular neck taper adaptor and corrosion products in the tissues. Therefore, we believe corrosion products from the taper caused by mismatch of the implant components led to pseudotumour formation requiring revision.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1618 - 1624
1 Dec 2012
Daurka JS Malik AK Robin DA Witt JD

The inherent challenges of total hip replacement (THR) in children include the choice of implant for the often atypical anatomical morphology, its fixation to an immature growing skeleton and the bearing surface employed to achieve a successful long-term result. We report the medium-term results of 52 consecutive uncemented THRs undertaken in 35 paediatric patients with juvenile idiopathic arthritis. The mean age at the time of surgery was 14.4 years (10 to 16). The median follow-up was 10.5 years (6 to 15). During the study period 13 THRs underwent revision surgery. With revision as an endpoint, subgroup analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral component and 62% (95% CI 41.0 to 78.0) of the acetabular component. Revision of the acetabular component for wear and osteolysis were the most common reasons for failure accounting for 11 of the 13 revisions.

The success seen in patients with a ceramic-on-ceramic articulation seems to indicate that this implant strategy has the potential to make a major difference to the long-term outcome in this difficult group of patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 671 - 677
1 May 2012
Raiss P Sowa B Bruckner T Eck S Woerz S Rohr K Rickert M Kasten P

The aim of this study was to compare a third-generation cementing procedure for glenoid components with a new technique for cement pressurisation. In 20 pairs of scapulae, 20 keeled and 20 pegged glenoid components were implanted using either a third-generation cementing technique (group 1) or a new pressuriser (group 2). Cement penetration was measured by three-dimensional (3D) analysis of micro-CT scans. The mean 3D depth of penetration of the cement was significantly greater in group 2 (p < 0.001). The mean thickness of the cement mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and 5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm (0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement mantle < 2 mm was detected less frequently in group 2 (p < 0.001). Using the cement pressuriser the proportion of cement mantles < 2 mm was significantly reduced compared with the third-generation cementing technique.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 51 - 55
1 Jan 2012
Masterson S Lidder S Scott G

We report the long-term results of revision total hip replacement using femoral impaction allografting with both uncemented and cemented Freeman femoral components. A standard design of component was used in both groups, with additional proximal hydroxyapatite coating in the uncemented group. A total of 33 hips in 30 patients received an uncemented component and 31 hips in 30 patients a cemented component. The mean follow-up was 9.8 years (2 to 17) in the uncemented group and 6.2 years (1 to 11) in the cemented group. Revision procedures (for all causes) were required in four patients (four hips) in the uncemented group and in five patients (five hips) in the cemented group. Harris hip scores improved significantly in both groups and were maintained independently of the extent of any migration of the femoral component within the graft or graft–cement mantle.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 158 - 162
1 Feb 2012
Sternheim A Backstein D Kuzyk PRT Goshua G Berkovich Y Safir O Gross AE

We report the use of porous metal acetabular revision shells in the treatment of contained bone loss. The outcomes of 53 patients with 50% acetabular bleeding host bone contact were compared with a control group of 49 patients with > 50% to 85% bleeding host bone contact. All patients were treated with the same type of trabecular metal acetabular revision shell. The mean age at revision was 62.4 years (42 to 80) and the mean follow-up of both groups was 72.4 months (60 to 102). Clinical, radiological and functional outcomes were assessed. There were four (7.5%) mechanical failures in the 50% host bone contact group and no failures in the > 50% host bone contact group (p = 0.068). Out of both groups combined there were four infections (3.9%) and five recurrent dislocations (4.9%) with a stable acetabular component construct that were revised to a constrained liner. Given the complexity of the reconstructive challenge, porous metal revision acetabular shells show acceptable failure rates at five to ten years’ follow-up in the setting of significant contained bone defects. This favourable outcome might be due to the improved initial stability achieved by a high coefficient of friction between the acetabular implant and the host bone, and the high porosity, which affords good bone ingrowth.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 459 - 465
1 Apr 2012
Nikolaou VS Edwards MR Bogoch E Schemitsch EH Waddell JP

The ideal bearing surface for young patients undergoing total hip replacement (THR) remains controversial. We report the five-year results of a randomised controlled trial comparing the clinical and radiological outcomes of 102 THRs in 91 patients who were <  65 years of age. These patients were randomised to receive a cobalt–chrome on ultra-high-molecular-weight polyethylene, cobalt–chrome on highly cross-linked polyethylene, or a ceramic-on-ceramic bearing. In all, 97 hip replacements in 87 patients were available for review at five years. Two hips had been revised, one for infection and one for peri-prosthetic fracture.

At the final follow-up there were no significant differences between the groups for the mean Western Ontario and McMaster Universities osteoarthritis index (pain, p = 0.543; function, p = 0.10; stiffness, p = 0.99), Short Form-12 (physical component, p = 0.878; mental component, p = 0.818) or Harris hip scores (p = 0.22). Radiological outcomes revealed no significant wear in the ceramic group.

Comparison of standard and highly cross-linked polyethylene, however, revealed an almost threefold difference in the mean annual linear wear rates (0.151 mm/year versus 0.059 mm/year, respectively) (p < 0.001).


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 322 - 327
1 Mar 2012
Morley JR Blake SM Hubble MJW Timperley AJ Gie GA Howell JR

The removal of all prosthetic material and a two-stage revision procedure is the established standard management of an infected total hip replacement (THR). However, the removal of well-fixed femoral cement is time-consuming and can result in significant loss of bone stock and femoral shaft perforation or fracture. We report our results of two-stage revision THR for treating infection, with retention of the original well-fixed femoral cement mantle in 15 patients, who were treated between 1989 and 2002. Following partial excision arthroplasty, patients received local and systemic antibiotics and underwent reconstruction and re-implantation at a second-stage procedure, when the infection had resolved.

The mean follow-up of these 15 patients was 82 months (60 to 192). Two patients had positive microbiology at the second stage and were treated with six weeks of appropriate antibiotics; one of these developed recurrent infection requiring further revision. Successful eradication of infection was achieved in the remaining 14 patients.

We conclude that when two-stage revision is used for the treatment of peri-prosthetic infection involving a THR, a well-fixed femoral cement mantle can be safely left in situ, without compromising the treatment of infection. Advantages of this technique include a shorter operating time, reduced loss of bone stock and a technically more straightforward second-stage procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 562 - 575
1 May 2006
Boileau P Sinnerton RJ Chuinard C Walch G


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 4 | Pages 510 - 516
1 Apr 2012
Hickey BA Kempshall PJ Metcalfe AJ Forster MC

As part of the national initiative to reduce waiting times for joint replacement surgery in Wales, the Cardiff and Vale NHS Trust referred 224 patients to the NHS Treatment Centre in Weston-Super-Mare for total knee replacement (TKR). A total of 258 Kinemax TKRs were performed between November 2004 and August 2006. Of these, a total of 199 patients (232 TKRs, 90%) have been followed up for five years. This cohort was compared with 258 consecutive TKRs in 250 patients, performed at Cardiff and Vale Orthopaedic Centre (CAVOC) over a similar time period. The five year cumulative survival rate was 80.6% (95% confidence interval (CI) 74.0 to 86.0) in the Weston-Super-Mare cohort and 95.0% (95% CI 90.2 to 98.2) in the CAVOC cohort with revision for any reason as the endpoint. The relative risk for revision at Weston-Super-Mare compared with CAVOC was 3.88 (p < 0.001). For implants surviving five years, the mean Oxford knee scores (OKS) and mean EuroQol (EQ-5D) scores were similar (OKS: Weston-Super-Mare 29 (2 to 47) vs CAVOC 29.8 (3 to 48), p = 0.61; EQ-5D: Weston-Super-Mare 0.53 (-0.38 to 1.00) vs CAVOC 0.55 (-0.32 to 1.00), p = 0.79). Patients with revised TKRs had significantly lower Oxford knee and EQ-5D scores (p < 0.001).

The results show a higher revision rate for patients operated at Weston-Super-Mare Treatment Centre, with a reduction in functional outcome and quality of life after revision. This further confirms that patients moved from one area to another for joint replacement surgery fare poorly.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 1 | Pages 80 - 85
1 Jan 2012
Malone AA Sanchez JS Adams R Morrey B

We report the effectiveness of revision of total elbow replacement by re-cementing. Between 1982 and 2004, 53 elbows in 52 patients were treated with re-cementing of a total elbow replacement into part or all of the existing cement mantle or into the debrided host-bone interface, without the use of structural bone augmentation or a custom prosthesis. The original implant revision was still in situ and functional in 42 of 53 elbows (79%) at a mean of 94.5 months (26 to 266) after surgery. In 31 of these 42 elbows (74%) the Mayo Elbow Performance Score was good or excellent. Overall, of the 53 elbows, 18 (34%) required re-operation, ten (19%) for loosening. A classification system was developed to identify those not suitable for revision by this technique, and using this we have showed that successful re-implantation is statistically correlated to properly addressing the bone deficiency for both the humeral (p = 0.005) and the ulnar (p = 0.039) components.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 315 - 321
1 Mar 2012
Coulter G Young DA Dalziel RE Shimmin AJ

We report the findings of an independent review of 230 consecutive Birmingham hip resurfacings (BHRs) in 213 patients (230 hips) at a mean follow-up of 10.4 years (9.6 to 11.7). A total of 11 hips underwent revision; six patients (six hips) died from unrelated causes; and 13 patients (16 hips) were lost to follow-up. The survival rate for the whole cohort was 94.5% (95% confidence interval (CI) 90.1 to 96.9). The survival rate in women was 89.1% (95% CI 79.2 to 94.4) and in men was 97.5% (95% CI 92.4 to 99.2). Women were 1.4 times more likely to suffer failure than men. For each millimetre increase in component size there was a 19% lower chance of a failure. The mean Oxford hip score was 45.0 (median 47.0, 28 to 48); mean University of California, Los Angeles activity score was 7.4 (median 8.0, 3 to 9); mean patient satisfaction score was 1.4 (median 1.0, 0 to 9). A total of eight hips had lysis in the femoral neck and two hips had acetabular lysis. One hip had progressive radiological changes around the peg of the femoral component. There was no evidence of progressive neck narrowing between five and ten years.

Our results confirm that BHR provides good functional outcome and durability for men, at a mean follow-up of ten years. We are now reluctant to undertake hip resurfacing in women with this implant.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 9 | Pages 1217 - 1222
1 Sep 2011
Bonner TJ Eardley WGP Patterson P Gregg PJ

Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR.

A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by > 3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47).

We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.


Bone & Joint Research
Vol. 2, Issue 6 | Pages 102 - 111
1 Jun 2013
Patel RA Wilson RF Patel PA Palmer RM

Objectives

To review the systemic impact of smoking on bone healing as evidenced within the orthopaedic literature.

Methods

A protocol was established and studies were sourced from five electronic databases. Screening, data abstraction and quality assessment was conducted by two review authors. Prospective and retrospective clinical studies were included. The primary outcome measures were based on clinical and/or radiological indicators of bone healing. This review specifically focused on non-spinal orthopaedic studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 11 | Pages 1479 - 1486
1 Nov 2011
Park J Kim Y

The purpose of this prospective, randomised study was to evaluate the clinical and radiological results comparing the identical cemented or cementless NexGen total knee prostheses implanted bilaterally in the same patient. Sequential simultaneous bilateral total knee replacements were performed in 50 patients (100 knees). There were 39 women and 11 men with a mean age of 58.4 years (51 to 67) who received a cemented prosthesis in one knee and a cementless prosthesis in the other. The mean follow-up was 13.6 years (13 to 14). At final review, the mean Knee Society scores (96.2 (82 to 100) versus 97.7 (90 to 100)), the mean Western Ontario and McMaster Universities osteoarthritis index (34.5 (4 to 59) versus 35.6 (5 to 51)), the mean ranges of knee movement (124° (100° to 140°) versus 128° (110° to 140°)), mean patient satisfaction (8.1 (sd 1.9) versus 8.3 (sd 1.7)), and radiological results were similar in both groups. The rate of survival of the femoral components was 100% in both groups at 14 years. The rate of survival of the cemented tibial component was 100% and 98% in the cementless tibial component. No osteolysis was identified in either group. Our data have shown no advantage of cementless over cemented components in total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1355 - 1361
1 Oct 2011
Bollars P Luyckx J Innocenti B Labey L Victor J Bellemans J

High-flexion total knee replacement (TKR) designs have been introduced to improve flexion after TKR. Although the early results of such designs were promising, recent literature has raised concerns about the incidence of early loosening of the femoral component. We compared the minimum force required to cause femoral component loosening for six high-flexion and six conventional TKR designs in a laboratory experiment.

Each TKR design was implanted in a femoral bone model and placed in a loading frame in 135° of flexion. Loosening of the femoral component was induced by moving the tibial component at a constant rate of displacement while maintaining the same angle of flexion. A stereophotogrammetric system registered the relative movement between the femoral component and the underlying bone until loosening occurred.

Compared with high-flexion designs, conventional TKR designs required a significantly higher force before loosening occurred (p < 0.001). High-flexion designs with closed box geometry required significantly higher loosening forces than high-flexion designs with open box geometry (p = 0.0478). The presence of pegs further contributed to the fixation strength of components.

We conclude that high-flexion designs have a greater risk for femoral component loosening than conventional TKR designs. We believe this is attributable to the absence of femoral load sharing between the prosthetic component and the condylar bone during flexion.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 648 - 655
1 May 2012
Jolles BM Grzesiak A Eudier A Dejnabadi H Voracek C Pichonnaz C Aminian K Martin E

This study compared the outcome of total knee replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses during the first post-operative year and at five years’ follow-up, using gait parameters as a new objective measure. This double-blind randomised controlled clinical trial included 55 patients with mobile-bearing (n = 26) and fixed-bearing (n = 29) prostheses of the same design, evaluated pre-operatively and post-operatively at six weeks, three months, six months, one year and five years. Each participant undertook two walking trials of 30 m and completed the EuroQol questionnaire, Western Ontario and McMaster Universities osteoarthritis index, Knee Society score, and visual analogue scales for pain and stiffness. Gait analysis was performed using five miniature angular rate sensors mounted on the trunk (sacrum), each thigh and calf. The study population was divided into two groups according to age (≤ 70 years versus > 70 years).

Improvements in most gait parameters at five years’ follow-up were greater for fixed-bearing TKRs in older patients (> 70 years), and greater for mobile-bearing TKRs in younger patients (≤ 70 years). These findings should be confirmed by an extended age controlled study, as the ideal choice of prosthesis might depend on the age of the patient at the time of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 10 | Pages 1382 - 1388
1 Oct 2011
Amirfeyz R Stanley D

We studied, ten patients (11 elbows) who had undergone 14 allograft-prosthesis composite reconstructions following failure of a previous total elbow replacement with massive structural bone loss. There were nine women and one man with a mean age of 64 years (40 to 84), who were reviewed at a mean of 75 months (24 to 213). One patient developed a deep infection after 26 months and had the allograft-prosthesis composite removed, and two patients had mild pain. The median flexion-extension arc was 100° (95% confidence interval (CI) 76° to 124°). With the exception of the patient who had the infected failure, all the patients could use their elbows comfortably without splints or braces for activities of daily living. The mean Mayo Elbow Performance Index improved from 9.5 (95% CI 4.4 to 14.7) pre-operatively to 74 (95% CI 62.4 to 84.9) at final review.

Radiologically, the rate of partial resorption was similar in the humeral and ulnar allografts (three of six and four of eight, respectively; p > 0.999). The patterns of resorption, however, were different. Union at the host-bone-allograft junction was also different between the humeral and ulnar allografts (one of six and seven of eight showing union, respectively; p = 0.03).

At medium-term follow-up, allograft-prosthesis composite reconstruction appears to be a useful salvage technique for failed elbow replacements with massive bone loss. The effects of allograft resorption and host-bone-allograft junctional union on the longevity of allograft-prosthesis composite reconstruction, however, remain unknown, and it is our view that these patients should remain under long-term regular review.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 1 | Pages 34 - 38
1 Jan 2011
Charity JAF Tsiridis E Sheeraz A Howell JR Hubble MJW Timperley AJ Gie GA

We evaluated all cases involving the combined use of a subtrochanteric derotational femoral shortening osteotomy with a cemented Exeter stem performed at our institution. With severe developmental dysplasia of the hip an osteotomy is often necessary to achieve shortening and derotation of the proximal femur. Reduction can be maintained with a 3.5 mm compression plate while the implant is cemented into place. Such a plate was used to stabilise the osteotomy in all cases. Intramedullary autograft helps to prevent cement interposition at the osteotomy site and promotes healing. There were 15 female patients (18 hips) with a mean age of 51 years (33 to 75) who had a Crowe IV dysplasia of the hip and were followed up for a mean of 114 months (52 to 168). None was lost to follow-up. All clinical scores were collected prospectively. The Charnley modification of the Merle D’Aubigné-Postel scores for pain, function and range of movement showed a statistically significant improvement from a mean of 2.4 (1 to 4), 2.3 (1 to 4), 3.4 (1 to 6) to 5.2 (3 to 6), 4.4 (3 to 6), 5.2 (4 to 6), respectively. Three acetabular revisions were required for aseptic loosening; one required femoral revision for access. One osteotomy failed to unite at 14 months and was revised successfully. No other case required a femoral revision. No postoperative sciatic nerve palsy was observed.

Cemented Exeter femoral components perform well in the treatment of Crowe IV dysplasia with this procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 4 | Pages 535 - 539
1 Apr 2010
Cazeneuve JF Cristofari D

We have previously described the short-term outcome of the use of reverse shoulder arthroplasty in the treatment of acute complex proximal humeral fractures in the elderly. We now report the clinical and radiological outcome of 36 fractures at a mean of 6.6 years (1 to 16). Previously, at a mean follow-up of 6 years (1 to 12) the mean Constant score was 58.5; this was reduced to 53 points with the further follow-up. A total of 23 patients (63%) had radiological evidence of loosening of the glenoid component. Nevertheless, only one patient had aseptic loosening of the baseplate at 12 years’ follow-up.

The reduction in the mean Constant score with longer follow-up and the further development of scapular notching is worrying.

New developments in design, bearing surfaces and surgical technique, and further follow-up, will determine whether reverse shoulder arthroplasty has a place in the management of complex proximal humeral fractures in the elderly.