The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 7 | Pages 966 - 969
1 Sep 2004
Shiramizu K Naito M Shitama T Nakamura Y Shitama H

The existing methods of assessing limb lengthening during total hip arthroplasty (THA) are prone to error because the measurements are not parallel to the limb lengthening axis. In order to address this, we designed a caliper to estimate limb lengthening during THA and evaluated its accuracy compared with our previous device, the straight caliper. Limb lengths were measured in 100 patients. The L-shaped caliper was used in 50 cases and the straight caliper in 50. The correlation between intra-operative and post-operative radiographic measurements was significantly improved using the L-shaped device (p < 0.0001, r = 0.934). This method was extremely accurate in predicting changes in limb length due to surgery


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 5 | Pages 835 - 842
1 Sep 1999
Allain J Le Mouel S Goutallier D Voisin MC

Between January 1988 and January 1991 we performed 100 consecutive cemented total hip replacements using a zirconia head, a titanium alloy stem and a polyethylene cup. We reviewed 78 of these hips in 61 patients in detail at a mean of 5.8 years (1 to 9). Aseptic loosening was seen in 11 hips (14%). Eight needed revision. In total, 37 cups (47.5%) showed radiolucent lines, all at the cement-bone interface, with 18 (23%) involving all the interface. Of the 78 femoral implants, 17 (21.7%) showed radiolucent lines, and two, which had a complete line of more than 1 mm thick, definite endocortical osteolyses. There was also an abnormally high incidence of osteolysis of more than 2 mm at the calcar. Survivorship analysis showed that only 63% were in situ at eight years. These worrying results led us to abandon the use of zirconia heads, since at the same hospital, using the same femoral stem, cement and polyethylene cup, but with alumina femoral heads, the survival rate was 93% at nine years. We discuss the possible reasons for the poor performance of zirconia ceramic


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 2 | Pages 191 - 194
1 Mar 1996
McCaskie AW Brown AR Thompson JR Gregg PJ

Three radiological methods are commonly used to assess the outcome of total hip replacement (THR). They aim to record the appearance of lucent areas and migration of the prosthesis in a reproducible manner. Two of them were designed to monitor the implant through time and one to grade the quality of cementing. We have measured the level of inter- and intraobserver agreement in all three systems. We randomised 30 patients to receive either finger packing or retrograde gun cementing during Charnley hip replacements. The postoperative departmental radiographs were evaluated in a blinded study by two orthopaedic trainees, two consultants and two experts in THR. The trainees and consultants repeated the exercise at least two weeks later. We used the unweighted kappa statistic to establish the levels of agreement. In general, intraobserver agreement was moderate but interobserver agreement was poor, with levels similar to or less than those expected by chance. Our results indicate that such systems cannot provide reliable data from centres in different parts of the world, with various levels of surgeon evaluating radiographs at differing time intervals. We discuss the problem and suggest some methods of improvement


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 45 - 49
1 Jan 1992
Fordyce M Ling R

In a prospective, randomised controlled trial, the efficacy of the A-V Impulse System in the prevention of deep-vein thrombosis was investigated in 84 patients who had undergone total hip replacement. The incidence of venographically proven, and clinically significant postoperative deep-vein thrombosis was 40% in the control group and 5% in the treatment group (p less than 0.001). No adverse reactions were recorded


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 5 | Pages 639 - 642
1 Jul 2004
Pitto RP Hamer H Heiss-Dunlop W Kuehle J

Routine prophylaxis for venous thromboembolic disease after total hip replacement (THR) is recommended. Pneumatic compression with foot pumps seems to provide an alternative to chemical agents. However, the overall number of patients investigated in randomised clinical trials has been too small to draw evidence-based conclusions. This randomised clinical trial was carried out to compare the effectiveness and safety of mechanical versus chemical prophylaxis of DVT in patients after THR. Inclusion criteria were osteoarthritis of the hip and age less than 80 years. Exclusion criteria included a history of thromboembolic disease, heart disease, and bleeding diatheses. There were 216 consecutive patients considered for inclusion in the trial who were randomised either for management with the A-V Impulse System foot pump. We excluded 16 patients who did not tolerate continuous use of the foot pump or with low-molecular-weight heparin (LMWH). Patients were monitored for DVT using serial duplex sonography at 3, 10 and 45 days after surgery. DVT was detected in three of 100 patients in the foot-pump group and with six of 100 patients in the LMWH group (p < 0.05). The mean post-operative drainage was 259 ml in the foot-pump group and 328 ml in the LMWH group (p < 0.05). Patients in the foot-pump group had less swelling of the thigh (10 mm compared with 15 mm; p < 0.05). One patient developed heparin-induced thrombocytopenia. This study confirms the effectiveness and safety of mechanical prophylaxis of DVT in THR. Some patients cannot tolerate the foot pump


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 377 - 381
1 May 1998
Ilchmann T Markovic L Joshi A Hardinge K Murphy J Wingstrand H

We analysed in-vivo migration and wear over a long period of all-polyethylene acetabular cups which had not been affected by mechanical loosening. The selection criteria of regular radiological follow-up, good clinical outcome (Charnley score of 5 or 6), continued walking without crutches and no radiological signs of loosening of the acetabular cups were fulfilled by 25 Charnley total hip arthroplasties. Mean migration, measured by the Nunn method, was 0.6 mm in the medial and 0.2 mm in the cranial direction. The mean yearly rate of wear was 0.05 mm and 0.04 mm, with six and two cups having no detectable wear, as measured by the Livermore and Charnley-Cupic methods, respectively. The maximal detected wear was 3.7 mm. There were no changes in the rate of wear with time. Computerised Ein Bild Röntgen Analyse (single-image radiological analysis) measurements of 20 hips indicated plastic deformation of the cups. We conclude that long-term successful cups do not migrate and have a very low rate of wear which was not affected by ageing of the polyethylene. There was no evidence that polyethylene wear alone caused mechanical loosening of the cup but high rates of wear seem to have an adverse prognostic value in terms of the long-term survival of the prosthesis


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 862 - 866
1 Jul 2018
Darrith B Bell JA Culvern C Della Valle CJ

Aims

Accurate placement of the acetabular component is essential in total hip arthroplasty (THA). The purpose of this study was to determine if the ability to achieve inclination of the acetabular component within the ‘safe-zone’ of 30° to 50° could be improved with the use of an inclinometer.

Patients and Methods

We reviewed 167 primary THAs performed by a single surgeon over a period of 14 months. Procedures were performed at two institutions: an inpatient hospital, where an inclinometer was used (inclinometer group); and an ambulatory centre, where an inclinometer was not used as it could not be adequately sterilized (control group). We excluded 47 patients with a body mass index (BMI) of > 40 kg/m2, age of > 68 years, or a surgical indication other than osteoarthritis whose treatment could not be undertaken in the ambulatory centre. There were thus 120 patients in the study, 68 in the inclinometer group and 52 in the control group. The inclination angles of the acetabular component were measured from de-identified plain radiographs by two blinded investigators who were not involved in the surgery. The effect of the use of the inclinometer on the inclination angle was determined using multivariate regression analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 5 | Pages 655 - 660
1 Jul 2003
Tanaka K Tamura J Kawanabe K Shimizu M Nakamura T

We examined the behaviour of alumina ceramic heads in 156 cemented total hip arthroplasties, at a minimum follow-up of eight years. They were divided into three groups according to the size of the femoral head; 22, 26, and 28 mm. We measured polyethylene wear radiologically using a computer-aided technique. The linear wear rate of polyethylene sockets for the 28 mm heads was high (0.156 mm/year), whereas those for the 22 and 26 mm heads were relatively low (0.090 and 0.098 mm/year, respectively). Moreover, the surface roughness data of retrieved femoral heads clearly showed maintenance of an excellent surface finish of the current alumina. We conclude that the alumina ceramic femoral heads currently used are associated with a reduced rate of polyethylene wear


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 700 - 705
1 Jul 2002
English H Timperley AJ Dunlop D Gie G

We report the mid-term results of femoral impaction grafting which was used in 53 patients during the second stage of a two-stage revision for an infected total hip replacement. We reviewed all cases performed between 1989 and 1998. All patients underwent a Girdlestone excision arthroplasty, received local and systemic antibiotics and subsequently underwent reconstruction, using femoral impaction grafting. Four patients had further infection (7.5%), and four died within 24 months of surgery. One patient underwent revision of the stem for a fracture below its tip at ten months. This left 44 patients with a mean follow-up of 53 months (24 to 122). All had improved clinical scores and a satisfactory radiological outcome


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 2 - 8
1 Jun 2019
Aggarwal VK Weintraub S Klock J Stachel A Phillips M Schwarzkopf R Iorio R Bosco J Zuckerman JD Vigdorchik JM Long WJ

Aims

We studied the impact of direct anterior (DA) versus non-anterior (NA) surgical approaches on prosthetic joint infection (PJI), and examined the impact of new perioperative protocols on PJI rates following all surgical approaches at a single institution.

Patients and Methods

A total of 6086 consecutive patients undergoing primary total hip arthroplasty (THA) at a single institution between 2013 and 2016 were retrospectively evaluated. Data obtained from electronic patient medical records included age, sex, body mass index (BMI), medical comorbidities, surgical approach, and presence of deep PJI. There were 3053 male patients (50.1%) and 3033 female patients (49.9%). The mean age and BMI of the entire cohort was 62.7 years (18 to 102, sd 12.3) and 28.8 kg/m2 (13.3 to 57.6, sd 6.1), respectively. Infection rates were calculated yearly for the DA and NA approach groups. Covariates were assessed and used in multivariate analysis to calculate adjusted odds ratios (ORs) for risk of development of PJI with DA compared with NA approaches. In order to determine the effect of adopting a set of infection prevention protocols on PJI, we calculated ORs for PJI comparing patients undergoing THA for two distinct time periods: 2013 to 2014 and 2015 to 2016. These periods corresponded to before and after we implemented a set of perioperative infection protocols.


The Journal of Bone & Joint Surgery British Volume
Vol. 62-B, Issue 4 | Pages 450 - 453
1 Nov 1980
Loudon Charnley J

A method is described of measuring radiological subsidence of a femoral prosthesis in relation to the femur after total hip arthroplasty. The method depends on measuring the distance from the tip of the femoral prosthesis to a fixed point in the bone. Subsidence after the use of a conventional design of femoral stem is compared with that after the use of a stem with a dorsal flange (Cobra). A significant reduction in the incidence and amount of subsidence was found when using the dorsal flange. There was also a notable absence of transverse fractures involving the cement near the tip of the stem, which occurred in 26 per cent of the cases using a conventional prosthesis


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 1 | Pages 45 - 48
1 Jan 1994
Pierchon F Pasquier G Cotten A Fontaine C Clarisse J Duquennoy A

We reviewed radiographs and CT scans of 38 total hip arthroplasties which had dislocated (36 posteriorly; 2 anteriorly) and compared the alignment of the prosthetic components with those of 14 uncomplicated arthroplasties. No difference was found between the alignment of the prosthetic components in the two groups. In the seven patients who had reoperations, the cause of dislocation diagnosed by CT was confirmed in only two cases (one retroversion of the cup and one protruding osteophyte). Muscular imbalance rather than malposition of the components was the major factor determining dislocation. CT allows accurate measurement of cup and neck anteversion but contributes little to preoperative planning


The Journal of Bone & Joint Surgery British Volume
Vol. 73-B, Issue 4 | Pages 559 - 563
1 Jul 1991
Witt J Swann M

Thirteen total hip replacements with titanium alloy femoral components required revision for loosening at an average of two years after implantation. At revision the soft tissues around the implant were darkly stained and a proliferative membrane had invaded the cement-bone interface. The femoral components showed polishing of parts of their shot-blasted surfaces. Histology showed a fibroblastic reaction with abundant titanium lying free and within histiocytes, and a scanty foreign-body giant-cell reaction. Surface analysis of the removed femoral components and chemical analysis of the excised tissues is described. Tissue reaction in response to the metal-wear debris may have contributed to the early failure of these implants


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 5 | Pages 851 - 855
1 Nov 1989
Hope P Kristinsson K Norman P Elson R

We reviewed a series of 91 patients with deep infection of a cemented total hip arthroplasty caused by coagulase-negative staphylococci (C-NS). Of these, 72 were treated by one-stage exchange arthroplasty with a failure rate of 13% due to recurrence of infection. The other 19 patients have started or completed treatment by a two-stage exchange without failure to date. In 27 of the 91 patients multiple strains of C-NS were discovered, many being resistant to previously used antibiotics. The use of gentamicin-containing cement in the primary arthroplasty was significantly associated with the emergence of gentamicin-resistant C-NS in subsequent deep infection. Bacteriological diagnosis of such infections must take into consideration the possibility that multiple strains of the organism are involved


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 1 | Pages 60 - 67
1 Jan 1994
Shanbhag A Jacobs J Glant T Gilbert J Black J Galante J

Interfacial membranes collected at revision from 11 failed uncemented Ti-alloy total hip replacements were examined. Particles in the membranes were characterised by electron microscopy, microchemical spectroscopy and particle size analysis. Most were polyethylene and had a mean size of 0.53 micron +/- 0.3. They were similar to the particles seen in the base resin used in the manufacture of the acetabular implants. Relatively few titanium particles were seen. Fragments of bone, stainless steel and silicate were found in small amounts. Most of the polyethylene particles were too small to be seen by light microscopy. Electron microscopy and spectroscopic techniques are required to provide an accurate description of this debris


The Bone & Joint Journal
Vol. 98-B, Issue 11 | Pages 1425 - 1426
1 Nov 2016
Reed M Haddad FS


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 1 | Pages 47 - 50
1 Jan 1989
Baker A Bitounis V

Electromyographic and clinical studies were performed on patients undergoing total hip replacement by the modified direct lateral (29 hips), the direct lateral (29 hips) and the posterior approaches (21 hips). Assessments were made three months after operation. The Trendelenburg test was positive (Grade II) in eight cases operated upon by the direct lateral route, but in only one of each of the other two groups. Denervation occurred in only five of the 28 hips with abductor weakness without statistical difference between the groups. In the modified direct lateral group, radiological evidence of union of the trochanteric sliver was associated with significantly better abductor function than in those with malunion or non-union


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 2 | Pages 145 - 148
1 Feb 2008
Stockley I Mockford BJ Hoad-Reddick A Norman P

We present a series of 114 patients with microbiologically-proven chronically-infected total hip replacement, treated between 1991 and 2004 by a two-stage exchange procedure with antibiotic-loaded cement, but without the use of a prolonged course of antibiotic therapy. The mean follow-up for all patients was 74 months (2 to 175) with all surviving patients having a minimum follow-up of two years. Infection was successfully eradicated in 100 patients (87.7%), a rate which is similar to that reported by others, but where prolonged adjuvant antibiotic therapy has been used. Using the technique described, a prolonged course of systemic antibiotics does not appear to be essential and the high cost of the administration of antibiotics can be avoided


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 5 | Pages 650 - 654
1 Jul 2003
Clarke MT Lee PTH Villar RN

Metal-on-metal (MOM) is a commonly used bearing notable for its ‘suction fit’ when lubricated. In this study, we examined the capacity for MOM bearings to protect against dislocation after total hip replacement (THR). We undertook a clinical investigation to compare the rate of dislocation of MOM bearings with those of ceramic-on-polyethylene (COP) bearings and found that one MOM bearing dislocated in a series of 109 hips (0.9%) compared with nine of 145 hips (6.2%) in the COP group (p = 0.02). We also performed an in vitro investigation comparing the peak forces generated during forced separation of the two bearings of the same dimensions at velocities from 1 to 50 cm/s. This revealed that the MOM bearing generated significant resistance to separation at all velocities (maximum mean 24 N), whereas the COP did not (maximum mean 1.9 N, p < 0.001). We conclude that MOM bearings are more stable to dislocation than COP bearings as a result of the interfacial forces provided by a thin, lubricating fluid


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 5 | Pages 686 - 690
1 Jul 2001
Jana AK Engh CA Lewandowski PJ Hopper RH Engh CA

We studied the results of total hip arthroplasty (THA) using AML porous-coated femoral components at a mean follow-up of 11 years in a non-selected, consecutive series of patients with rheumatoid arthritis. We reviewed 64 patients with 82 primary THAs using these components. There were seven men (8 hips) and 57 women (74 hips) with a mean age of 55.1 years (24 to 80) at the time of surgery. Nine patients (11 hips) died before the two-year follow-up. Of the remaining 71 hips, only one stem was revised for aseptic loosening. Survivorship for the stems was 98.1% (95% confidence interval (CI) 94.5 to 100.0) at ten years, using a life-table analysis, with revision for any reason as an endpoint. Of the 70 unrevised stems, 66 (94%) had bony ingrowth, while four (6%) were radiologically loose at the most recent follow-up (mean 11.4 years). Our study shows the excellent long-term results which can be achieved with porous-coated femoral components in patients with rheumatoid arthritis