We report a prospective study of the use of intramedullary bone blocks to improve the fixation of a matt-finish femoral stem in Charnley low-friction arthroplasties. There were 379 patients (441 hips), but at a minimum follow-up of ten years there were 258 arthroplasties in 221 patients including some which had been revised. The mean age at surgery was 41 years (17 to 51) and the mean follow-up was 13.4 years (1 to 20 including the early revisions). Nine stems (3.5%) had been revised for aseptic loosening, but there were no stem fractures.
The aim of this study was to report the long-term results of rotational acetabular osteotomy (RAO) for symptomatic hip dysplasia in patients aged younger than 21 years at the time of surgery. We evaluated 31 patients (37 hips) aged younger than 21 years at the time of surgery retrospectively. There were 29 female and two male patients. Their mean age at the time of surgery was 17.4 years (12 to 21). The mean follow-up was 17.9 years (7 to 30). The RAO was combined with a varus or valgus femoral osteotomy or a greater trochanteric displacement in eight hips, as instability or congruence of the hip could not be corrected adequately using RAO alone.Aims
Patients and Methods
We reviewed the radiographs of 864 Charnley and STH (Zimmer) cemented total hip arthroplasties with a mean follow-up of seven years (maximum 16 years).
Patients with a failed reverse shoulder arthroplasty (RSA) have limited salvage options. The aim of this study was to determine the outcome of revision RSA when used as a salvage procedure for a failed primary RSA. We reviewed all revision RSAs performed for a failed primary RSA between 2006 and 2012, excluding patients with a follow-up of less than two years. A total of 27 revision RSAs were included in the study. The mean age of the patients at the time of revision was 70 years (58 to 82). Of the 27 patients, 14 (52% were female). The mean follow-up was 4.4 years (2 to 10).Aims
Patients and Methods
The Norwegian Arthroplasty Register recorded 24,408 primary total hip replacements from 1987 to 1993; 2907 of them (13%) were performed with uncemented femoral components. We have compared the results of eight different designs, each used in more than 100 patients.
Between January 1988 and January 1991 we performed 100 consecutive cemented total hip replacements using a zirconia head, a titanium alloy stem and a polyethylene cup. We reviewed 78 of these hips in 61 patients in detail at a mean of 5.8 years (1 to 9). Aseptic loosening was seen in 11 hips (14%). Eight needed revision. In total, 37 cups (47.5%) showed radiolucent lines, all at the cement-bone interface, with 18 (23%) involving all the interface. Of the 78 femoral implants, 17 (21.7%) showed radiolucent lines, and two, which had a complete line of more than 1 mm thick, definite endocortical osteolyses. There was also an abnormally high incidence of osteolysis of more than 2 mm at the calcar.
The primary aim of the study was to perform an analysis to identify the cost per quality-adjusted life-year (QALY) of robot-assisted unicompartmental knee arthroplasty (rUKA) relative to manual total knee arthroplasty (TKA) and unicompartmental knee arthroplasty (UKA) for patients with isolated medial compartment osteoarthritis (OA) of the knee. Secondary aims were to assess how case volume and length of hospital stay influenced the relative cost per QALY. A Markov decision analysis was performed, using known parameters for costs, outcomes, implant survival, and mortality, to assess the cost-effectiveness of rUKA relative to manual TKA and UKA for patients with isolated medial compartment OA of the knee with a mean age of 65 years. The influence of case volume and shorter hospital stay were assessed.Aims
Patients and Methods
We have reviewed 105 Liverpool Mark II knee replacements in 71 patients to assess survival and long-term results. Assessment was both clinical and radiological, using a modification of the British Orthopaedic Association knee function assessment chart, and analysis was by the
Modular dual mobility (DM) prostheses in which a cobalt-chromium liner is inserted into a titanium acetabular shell ( This was a prospective study of patients between 18 and 65 years of age, with a body mass index (BMI) < 35 kg/m2 and University of California at Los Angeles (UCLA) activity score > 6, who received a modular cobalt-chromium acetabular liner, highly crosslinked polyethylene mobile bearing, and cementless titanium femoral stem for their primary THA. Patients with a history of renal disease and metal hardware elsewhere in the body were excluded. A total of 43 patients (30 male, 13 female; mean age 52.6 years (Aims
Patients and Methods
It has been suggested that mobile-bearing total knee arthroplasty
(TKA) might lead to better outcomes by accommodating some femorotibial
rotational mismatch, thereby reducing contact stresses and polyethylene
wear. The aim of this study was to determine whether there is a
difference between fixed- and mobile-bearing versions of a contemporary
TKA with respect to durability, range of movement (ROM) and function,
ten years postoperatively. A total of 240 patients who were enrolled in this randomized
controlled trial (RCT) underwent a primary cemented TKA with one
of three tibial components (all-polyethylene fixed-bearing, modular
metal-backed fixed-bearing and mobile-bearing). Patients were reviewed
at a median follow-up of ten years (IQR 9.2 to 10.4).Aims
Patients and Methods
The calcar femorale is a vertical plate of bone lying deep to the lesser trochanter and is formed as a result of traction of the iliopsoas which separates the femoral cortex into two distinct layers, the calcar femorale and the medial femoral cortex. They fuse together proximally to form the medial femoral neck. A stem placed centrally will abut against the calcar femorale with little or no space for cement. Clearing of the calcar will offer space for a cement layer, which will support the stem proximally on the posterior aspect. We compared two consecutive groups of Charnley low-friction arthroplasties, with and without clearing of the calcar. In 330 patients who had an arthroplasty without clearing the calcar, there were ten revisions for aseptic loosening of the stem and six other stems were considered ‘definitely loose’, giving a rate of failure of 4.8%. In 111 patients in whom the calcar was cleared there was only one revision for aseptic loosening and no stems were classed as ‘definitely loose’, giving a rate of failure of 0.9%.
Unicompartmental knee arthroplasty (UKA) is one surgical option for treating symptomatic medial osteoarthritis. Clinical studies have shown the functional benefits of UKA; however, the optimal alignment of the tibial component is still debated. The purpose of this study was to evaluate the effects of tibial coronal and sagittal plane alignment in UKA on knee kinematics and cruciate ligament tension, using a musculoskeletal computer simulation. The tibial component was first aligned perpendicular to the mechanical axis of the tibia, with a 7° posterior slope (basic model). Subsequently, coronal and sagittal plane alignments were changed in a simulation programme. Kinematics and cruciate ligament tensions were simulated during weight-bearing deep knee bend and gait motions. Translation was defined as the distance between the most medial and the most lateral femoral positions throughout the cycle.Objectives
Methods
We treated 31 consecutive patients of mean age 43 years (22 to 59) with severe osteoarthritis of the hip secondary to dysplasia by valgus-extension osteotomy. The clinical results were analysed for factors which may have affected the outcome. The procedure was carried out only on one side and was combined with a Chiari pelvic osteotomy in four patients. The angle of valgus was between 30 and 40° with 10 to 20° of extension. The mean follow-up was 15 years (12 to 18). Clinical evaluation using Charnley’s modification of the system of Merle d’Aubigné and Postel showed that the mean score for pain had improved from 2.3 before operation to 4.2 at 15 years, and function from 2.3 to 4.2. Range of movement had decreased from 4.2 to 3.5. Fifteen patients (48.4%) had a pain score of 5 or 6 and were evaluated as satisfactory. Kaplan-Meier
We reviewed retrospectively 50 consecutive total hip arthroplasties with proximal femoral replacement prostheses in 49 patients. All the patients had massive bone loss. There were 31 women and 18 men who had undergone a mean of 3.1 hip operations. At the last review, 15 patients had died and two had been lost to follow-up, so that 32 patients (33 hips) were available for assessment. The mean age of the patients at the time of the operation was 60.6 years. The mean length of clinical follow-up was 11.1 +/- 4 years, and mean radiographic follow-up was 7.6 +/- 3.2 years. The average preoperative Harris hip score was 46 +/- 13 points, which improved to 80 +/- 10 points after one year and to 76 +/- 16 points at the latest follow-up. The most frequent complication was dislocation, which occurred in 11 hips. Four femoral components and seven acetabular prostheses were revised because of aseptic loosening. With revision at the end point,
This study compares the PFC total knee arthroplasty (TKA) system in a prospective randomized control trial (RCT) of the mobile-bearing rotating-platform (RP) TKA against the fixed-bearing (FB) TKA. This is the largest RCT with the longest follow-up where cruciate-retaining PFC total knee arthroplasties are compared in a non-bilateral TKA study. A total of 167 patients (190 knees with 23 bilateral cases), were recruited prospectively and randomly assigned, with 91 knees receiving the RP and 99 knees receiving FB. The mean age was 65.5 years (48 to 82), the mean body mass index (BMI) was 29.7 kg/m2 (20 to 52) and 73 patients were female. The Knee Society Score (KSS), Knee Society Functional Score (KSFS), Oxford Knee Score (OKS), Western Ontario and McMaster Universities Arthritis Index (WOMAC), and 12-Item Short-Form Health Survey Physical and Mental Component Scores (SF-12 PCS, SF-12 MCS) were gathered and recorded preoperatively, at five-years’ follow-up, and at ten years’ follow-up. Additionally, Knee Injury and Osteoarthritis Outcome Scores (KOOS) were collected at five- and ten-year follow-ups. The prevalence of radiolucent lines (RL) on radiographs and implant survival were recorded at five- and ten-year follow-ups.Aims
Patients and Methods
Early implant migration measured with radiostereometric analysis (RSA) has been proposed as a useful predictor of long-term fixation of tibial components in total knee arthroplasty. Evaluation of actual long-term fixation is of interest for cemented components, as well as for cementless fixation, which may offer long-term advantages once osseointegration has occurred. The objective of this study was to compare the long-term migration with one- and two-year migration to evaluate the predictive ability of short-term migration data and to compare migration and inducible displacement between cemented and cementless (porous metal monoblock) components at least ten years postoperatively. Patients who had participated in RSA migration studies with two-year follow-up were recruited to return for a long-term follow-up, at least ten years from surgery. Two cemented tibial designs from two manufacturers and one porous metal monoblock cementless tibial design were studied. At the long-term follow-up, patients had supine RSA examinations to determine migration and loaded examinations (single leg stance) to determine inducible displacement. In total, 79 patients (54 female) returned, with mean time since surgery of 12 years (10 to 14). There were 58 cemented and 21 cementless tibial components.Aims
Patients and Methods