Advertisement for orthosearch.org.uk
Results 541 - 560 of 1163
Results per page:
The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 185 - 191
1 Feb 2015
Kendrick BJL Kaptein BL Valstar ER Gill HS Jackson WFM Dodd CAF Price AJ Murray DW

The most common reasons for revision of unicompartmental knee arthroplasty (UKA) are loosening and pain. Cementless components may reduce the revision rate. The aim of this study was to compare the fixation and clinical outcome of cementless and cemented Oxford UKAs.

A total of 43 patients were randomised to receive either a cemented or a cementless Oxford UKA and were followed for two years with radiostereometric analysis (RSA), radiographs aligned with the bone–implant interfaces and clinical scores.

The femoral components migrated significantly during the first year (mean 0.2 mm) but not during the second. There was no significant difference in the extent of migration between cemented and cementless femoral components in either the first or the second year. In the first year the cementless tibial components subsided significantly more than the cemented components (mean 0.28 mm (sd 0.17) vs. 0.09 mm (sd 0.19 mm)). In the second year, although there was a small amount of subsidence (mean 0.05 mm) there was no significant difference (p = 0.92) between cemented and cementless tibial components. There were no femoral radiolucencies. Tibial radiolucencies were narrow (< 1 mm) and were significantly (p = 0.02) less common with cementless (6 of 21) than cemented (13 of 21) components at two years. There were no complete radiolucencies with cementless components, whereas five of 21 (24%) cemented components had complete radiolucencies. The clinical scores at two years were not significantly different (p = 0.20).

As second-year migration is predictive of subsequent loosening, and as radiolucency is suggestive of reduced implant–bone contact, these data suggest that fixation of the cementless components is at least as good as, if not better than, that of cemented devices.

Cite this article: Bone Joint J 2015; 97-B:185–91.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 197 - 201
1 Feb 2015
Kallala RF Vanhegan IS Ibrahim MS Sarmah S Haddad FS

Revision total knee arthroplasty (TKA) is a complex procedure which carries both a greater risk for patients and greater cost for the treating hospital than does a primary TKA. As well as the increased cost of peri-operative investigations, blood transfusions, surgical instrumentation, implants and operating time, there is a well-documented increased length of stay which accounts for most of the actual costs associated with surgery.

We compared revision surgery for infection with revision for other causes (pain, instability, aseptic loosening and fracture). Complete clinical, demographic and economic data were obtained for 168 consecutive revision TKAs performed at a tertiary referral centre between 2005 and 2012.

Revision surgery for infection was associated with a mean length of stay more than double that of aseptic cases (21.5 vs 9.5 days, p < 0.0001). The mean cost of a revision for infection was more than three times that of an aseptic revision (£30 011 (sd 4514) vs £9655 (sd 599.7), p < 0.0001).

Current NHS tariffs do not fully reimburse the increased costs of providing a revision knee surgery service. Moreover, especially as greater costs are incurred for infected cases. These losses may adversely affect the provision of revision surgery in the NHS.

Cite this article: Bone Joint J 2015;97-B:197–201.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 192 - 196
1 Feb 2015
Bernhoff K Björck M

We have investigated iatrogenic popliteal artery injuries (PAI) during non arthroplasty knee surgery regarding mechanism of injury, treatment and outcomes, and to identify successful strategies when injury occurs.

In all, 21 iatrogenic popliteal artery injuries in 21 patients during knee surgery other than knee arthroplasty were identified from the Swedish Vascular Registry (Swedvasc) between 1987 and 2011. Prospective registry data were supplemented with case-records, including long-term follow-up. In total, 13 patients suffered PAI during elective surgery and eight during urgent surgery such as fracture fixation or tumour resection. Nine injuries were detected intra-operatively, five within 12 to 48 hours and seven > 48 hours post-operatively (two days to 23 years).

There were 19 open vascular and two endovascular surgical repairs. Two patients died within six months of surgery. One patient required amputation. Only six patients had a complete recovery of whom had the vascular injury detected at time of injury and repaired by a vascular surgeon. Patients sustaining vascular injury during elective procedures are more likely to litigate (p = 0.029).

We conclude that outcomes are poorer when there is a delay of diagnosis and treatment, and that orthopaedic surgeons should develop strategies to detect PAI early and ensure rapid access to vascular surgical support.

Cite this article: Bone Joint J 2015;97-B:192–6.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 147 - 149
1 Feb 2015
Morgan-Jones R Oussedik SIS Graichen H Haddad FS

Revision knee arthroplasty presents a number of challenges, not least of which is obtaining solid primary fixation of implants into host bone. Three anatomical zones exist within both femur and tibia which can be used to support revision implants. These consist of the joint surface or epiphysis, the metaphysis and the diaphysis. The methods by which fixation in each zone can be obtained are discussed. The authors suggest that solid fixation should be obtained in at least two of the three zones and emphasise the importance of pre-operative planning and implant selection.

Cite this article: Bone Joint J 2015;97-B:147–9.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 64 - 70
1 Jan 2015
Hamilton DF Burnett R Patton JT Howie CR Moran M Simpson AHRW Gaston P

Total knee arthroplasty (TKA) is an established and successful procedure. However, the design of prostheses continues to be modified in an attempt to optimise the functional outcome of the patient.

The aim of this study was to determine if patient outcome after TKA was influenced by the design of the prosthesis used.

A total of 212 patients (mean age 69; 43 to 92; 131 female (62%), 81 male (32%)) were enrolled in a single centre double-blind trial and randomised to receive either a Kinemax (group 1) or a Triathlon (group 2) TKA.

Patients were assessed pre-operatively, at six weeks, six months, one year and three years after surgery. The outcome assessments used were the Oxford Knee Score; range of movement; pain numerical rating scales; lower limb power output; timed functional assessment battery and a satisfaction survey. Data were assessed incorporating change over all assessment time points, using repeated measures analysis of variance longitudinal mixed models. Implant group 2 showed a significantly greater range of movement (p = 0.009), greater lower limb power output (p = 0.026) and reduced report of ‘worst daily pain’ (p = 0.003) over the three years of follow-up. Differences in Oxford Knee Score (p = 0.09), report of ‘average daily pain’ (p = 0.57) and timed functional performance tasks (p = 0.23) did not reach statistical significance. Satisfaction with outcome was significantly better in group 2 (p = 0.001).

These results suggest that patient outcome after TKA can be influenced by the prosthesis used.

Cite this article: Bone Joint J 2015;97-B:64–70.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 50 - 55
1 Jan 2015
Zuiderbaan HA Khamaisy S Thein R Nawabi DH Pearle AD

Progressive degenerative changes in the medial compartment of the knee following lateral unicompartmental arthroplasty (UKA) remains a leading indication for revision surgery. The purpose of this study is to evaluate changes in the congruence and joint space width (JSW) of the medial compartment following lateral UKA. The congruence of the medial compartment of 53 knees (24 men, 23 women, mean age 13.1 years; sd 62.1) following lateral UKA was evaluated pre-operatively and six weeks post-operatively, and compared with 41 normal knees (26 men, 15 women, mean age 33.7 years; sd 6.4), using an Interactive closest point algorithm which calculated the congruence index (CI) by performing a rigid transformation that best aligns the digitised tibial and femoral surfaces. Inner, middle and outer JSWs were measured by sub-dividing the medial compartment into four quarters on pre- and post-operative, weight bearing tunnel view radiographs. The mean CI of knees following lateral UKA significantly improved from 0.92 (sd 0.06) pre-operatively to 0.96 (sd 0.02) (p < 0.001) six weeks post-operatively. The mean CI of the healthy control group was 0.99 sd 0.01. Post-operatively, the mean inner JSW increased (p = 0.006) and the outer decreased (p = 0.002). The JSW was restored post-operatively as no significant differences were noted in all three locations compared with the control group (inner JSW p = 0.43; middle JSW p = 0.019, outer JSW p = 0.51).

Our data suggest that a well conducted lateral UKA may improve the congruence and normalise the JSW of the medial compartment, potentially preventing progression of degenerative change.

Cite this article: Bone Joint J 2015;97-B:50–5.


Bone & Joint Research
Vol. 4, Issue 1 | Pages 1 - 5
1 Jan 2015
Vázquez-Portalatín N Breur GJ Panitch A Goergen CJ

Objective

Dunkin Hartley guinea pigs, a commonly used animal model of osteoarthritis, were used to determine if high frequency ultrasound can ensure intra-articular injections are accurately positioned in the knee joint.

Methods

A high-resolution small animal ultrasound system with a 40 MHz transducer was used for image-guided injections. A total of 36 guinea pigs were anaesthetised with isoflurane and placed on a heated stage. Sterile needles were inserted directly into the knee joint medially, while the transducer was placed on the lateral surface, allowing the femur, tibia and fat pad to be visualised in the images. B-mode cine loops were acquired during 100 µl. We assessed our ability to visualise 1) important anatomical landmarks, 2) the needle and 3) anatomical changes due to the injection.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 56 - 63
1 Jan 2015
Abane L Anract P Boisgard S Descamps S Courpied JP Hamadouche M

In this study we randomised 140 patients who were due to undergo primary total knee arthroplasty (TKA) to have the procedure performed using either patient-specific cutting guides (PSCG) or conventional instrumentation (CI).

The primary outcome measure was the mechanical axis, as measured at three months on a standing long-leg radiograph by the hip–knee–ankle (HKA) angle. This was undertaken by an independent observer who was blinded to the instrumentation. Secondary outcome measures were component positioning, operating time, Knee Society and Oxford knee scores, blood loss and length of hospital stay.

A total of 126 patients (67 in the CI group and 59 in the PSCG group) had complete clinical and radiological data. There were 88 females and 52 males with a mean age of 69.3 years (47 to 84) and a mean BMI of 28.6 kg/m2 (20.2 to 40.8). The mean HKA angle was 178.9° (172.5 to 183.4) in the CI group and 178.2° (172.4 to 183.4) in the PSCG group (p = 0.34). Outliers were identified in 22 of 67 knees (32.8%) in the CI group and 19 of 59 knees (32.2%) in the PSCG group (p = 0.99). There was no significant difference in the clinical results (p = 0.95 and 0.59, respectively). Operating time, blood loss and length of hospital stay were not significantly reduced (p = 0.09, 0.58 and 0.50, respectively) when using PSCG.

The use of PSCG in primary TKA did not reduce the proportion of outliers as measured by post-operative coronal alignment.

Cite this article: Bone Joint J 2015;97-B:56–63.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 3 - 9
1 Jan 2015
Hossain FS Konan S Patel S Rodriguez-Merchan EC Haddad FS

The routine use of patient reported outcome measures (PROMs) in evaluating the outcome after arthroplasty by healthcare organisations reflects a growing recognition of the importance of patients’ perspectives in improving treatment. Although widely embraced in the NHS, there are concerns that PROMs are being used beyond their means due to a poor understanding of their limitations.

This paper reviews some of the current challenges in using PROMs to evaluate total knee arthroplasty. It highlights alternative methods that have been used to improve the assessment of outcome.

Cite this article: Bone Joint J 2015;97-B:3–9.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1657 - 1662
1 Dec 2014
Stambough JB Clohisy JC Barrack RL Nunley RM Keeney JA

The aims of this retrospective study were to compare the mid-term outcomes following revision total knee replacement (TKR) in 76 patients (81 knees) < 55 years of age with those of a matched group of primary TKRs based on age, BMI, gender and comorbid conditions. We report the activity levels, functional scores, rates of revision and complications. Compared with patients undergoing primary TKR, those undergoing revision TKR had less improvement in the mean Knee Society function scores (8.14 (–55 to +60) vs 23.3 points (–40 to +80), p < 0.001), a similar improvement in UCLA activity level (p = 0.52), and similar minor complication rates (16% vs 13%, p = 0.83) at a mean follow-up of 4.6 years (2 to 13.4). Further revision surgery was more common among revised TKRs (17% vs 5%, p = 0.02), with deep infection and instability being the most common reasons for failure. As many as one-third of patients aged < 55 years in the revision group had a complication or failure requiring further surgery.

Young patients undergoing revision TKR should be counselled that they can expect somewhat less improvement and a higher risk of complications than occur after primary TKR.

Cite this article: Bone Joint J 2014; 96-B:1657–62


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1623 - 1630
1 Dec 2014
Monk AP Choji K O’Connor JJ Goodfellow† JW Murray DW

We scanned 25 left knees in healthy human subjects using MRI. Multiplanar reconstruction software was used to take measurements of the inferior and posterior facets of the femoral condyles and the trochlea.

A ‘basic circle’ can be defined which, in the sagittal plane, fits the posterior and inferior facets of the lateral condyle, the posterior facet of the medial condyle and the floor of the groove of the trochlea. It also approximately fits both condyles in the coronal plane (inferior facets) and the axial plane (posterior facets). The circle fitting the inferior facet of the medial condyle in the sagittal plane was consistently 35% larger than the other circles and was termed the ‘medial inferior circle’. There were strong correlations between the radii of the circles, the relative positions of the centres of the condyles, the width of the condyles, the total knee width and skeletal measurements including height. There was poor correlation between the radii of the circles and the position of the trochlea relative to the condyles.

In summary, the condyles are approximately spherical except for the inferior facet medially, which has a larger radius in the sagittal plane. The size and position of the condyles are consistent and change with the size of the person. However, the position of the trochlea is variable even though its radius is similar to that of the condyles. This information has implications for understanding anterior knee pain and for the design of knee replacements.

Cite this article: Bone Joint J 2014;96-B:1623–30.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1644 - 1648
1 Dec 2014
Abdel MP Pulido L Severson EP Hanssen AD

Instability in flexion after total knee replacement (TKR) typically occurs as a result of mismatched flexion and extension gaps. The goals of this study were to identify factors leading to instability in flexion, the degree of correction, determined radiologically, required at revision surgery, and the subsequent clinical outcomes. Between 2000 and 2010, 60 TKRs in 60 patients underwent revision for instability in flexion associated with well-fixed components. There were 33 women (55%) and 27 men (45%); their mean age was 65 years (43 to 82). Radiological measurements and the Knee Society score (KSS) were used to assess outcome after revision surgery. The mean follow-up was 3.6 years (2 to 9.8). Decreased condylar offset (p < 0.001), distalisation of the joint line (p < 0.001) and increased posterior tibial slope (p < 0.001) contributed to instability in flexion and required correction at revision to regain stability. The combined mean correction of posterior condylar offset and joint line resection was 9.5 mm, and a mean of 5° of posterior tibial slope was removed. At the most recent follow-up, there was a significant improvement in the mean KSS for the knee and function (both p < 0.001), no patient reported instability and no patient underwent further surgery for instability.

The following step-wise approach is recommended: reduction of tibial slope, correction of malalignment, and improvement of condylar offset. Additional joint line elevation is needed if the above steps do not equalise the flexion and extension gaps.

Cite this article: Bone Joint J 2014;96-B:1644–8.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1649 - 1656
1 Dec 2014
Lindberg-Larsen M Jørgensen CC Bæk Hansen T Solgaard S Odgaard A Kehlet H

We present detailed information about early morbidity after aseptic revision knee replacement from a nationwide study. All aseptic revision knee replacements undertaken between 1st October 2009 and 30th September 2011 were analysed using the Danish National Patient Registry with additional information from the Danish Knee Arthroplasty Registry. The 1218 revisions involving 1165 patients were subdivided into total revisions, large partial revisions, partial revisions and revisions of unicondylar replacements (UKR revisions). The mean age was 65.0 years (27 to 94) and the median length of hospital stay was four days (interquartile range: 3 to 5), with a 90 days re-admission rate of 9.9%, re-operation rate of 3.5% and mortality rate of 0.2%. The age ranges of 51 to 55 years (p = 0.018), 76 to 80 years (p < 0.001) and ≥ 81 years (p < 0.001) were related to an increased risk of re-admission. The age ranges of 76 to 80 years (p = 0.018) and the large partial revision subgroup (p = 0.073) were related to an increased risk of re-operation. The ages from 76 to 80 years (p < 0.001), age ≥ 81 years (p < 0.001) and surgical time > 120 min (p <  0.001) were related to increased length of hospital stay, whereas the use of a tourniquet (p = 0.008) and surgery in a low volume centre (p = 0.013) were related to shorter length of stay.

In conclusion, we found a similar incidence of early post-operative morbidity after aseptic knee revisions as has been reported after primary procedures. This suggests that a length of hospital stay ≤ four days and discharge home at that time is safe following aseptic knee revision surgery in Denmark.

Cite this article: Bone Joint J 2014;96-B:1649–56.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1631 - 1636
1 Dec 2014
Parkkinen M Madanat R Mäkinen TJ Mustonen A Koskinen SK Lindahl J

The role of arthroscopy in the treatment of soft-tissue injuries associated with proximal tibial fractures remains debatable. Our hypothesis was that MRI over-diagnoses clinically relevant associated soft-tissue injuries. This prospective study involved 50 consecutive patients who underwent surgical treatment for a split-depression fracture of the lateral tibial condyle (AO/OTA type B3.1). The mean age of patients was 50 years (23 to 86) and 27 (54%) were female. All patients had MRI and arthroscopy. Arthroscopy identified 12 tears of the lateral meniscus, including eight bucket-handle tears that were sutured and four that were resected, as well as six tears of the medial meniscus, of which five were resected. Lateral meniscal injuries were diagnosed on MRI in four of 12 patients, yielding an overall sensitivity of 33% (95% confidence interval (CI) 11 to 65). Specificity was 76% (95% CI 59 to 88), with nine tears diagnosed among 38 menisci that did not contain a tear. MRI identified medial meniscal injuries in four of six patients, yielding an overall sensitivity of 67% (95% CI 24 to 94). Specificity was 66% (95% CI 50 to 79), with 15 tears diagnosed in 44 menisci that did not contain tears.

MRI appears to offer only a marginal benefit as the specificity and sensitivity for diagnosing meniscal injuries are poor in patients with a fracture. There were fewer arthroscopically-confirmed associated lesions than reported previously in MRI studies.

Cite this article: Bone Joint J 2014;96-B:1631–6.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1637 - 1643
1 Dec 2014
Yang Z Liu H Xie X Tan Z Qin T Kang P

Total knee replacement (TKR) is an effective method of treating end-stage arthritis of the knee. It is not, however, a procedure without risk due to a number of factors, one of which is diabetes mellitus. The purpose of this study was to estimate the general prevalence of diabetes in patients about to undergo primary TKR and to determine whether diabetes mellitus adversely affects the outcome. We conducted a systematic review and meta-analysis according to the Meta-analysis Of Observational Studies in Epidemiology (MOOSE) guidelines. The Odds Ratio (OR) and mean difference (MD) were used to represent the estimate of risk of a specific outcome. Our results showed the prevalence of diabetes mellitus among patients undergoing TKR was 12.2%. Patients with diabetes mellitus had an increased risk of deep infection (OR = 1.61, 95% confidence interval (CI), 1.38 to 1.88), deep vein thrombosis (in Asia, OR = 2.57, 95% CI, 1.58 to 4.20), periprosthetic fracture (OR = 1.89, 95% CI, 1.04 to 3.45), aseptic loosening (OR = 9.36, 95% CI, 4.63 to 18.90), and a poorer Knee Society function subscore (MD = -5.86, 95% CI, -10.27 to -1.46). Surgeons should advise patients specifically about these increased risks when obtaining informed consent and be meticulous about their peri-operative care.

Cite this article: Bone Joint J 2014;96-B:1637–43.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1485 - 1490
1 Nov 2014
Kim CW Seo SS Kim JH Roh SM Lee CR

The aim of this study was to find anatomical landmarks for rotational alignment of the tibial component in total knee replacement (TKR) in a CT-based study. Pre-operative CT scanning was performed on 94 South Korean patients (nine men, 85 women, 188 knees) with osteoarthritis of the knee joint prior to TKR. The tibial anteroposterior (AP) axis was defined as a line perpendicular to the femoral surgical transepicondylar axis and passing through the centre of the posterior cruciate ligament (PCL). The angles between the defined tibial AP axis and anatomical landmarks at various levels of the tibia were measured. The mean values of the angles between the defined tibial AP axis and the line connecting the anterior border of the proximal third of the tibia to the centre of the PCL was -0.2° (-17 to 14.1, sd 4.1). This was very close to the defined tibial axis, and remained so regardless of lower limb alignment and the degree of tibial bowing. Therefore, AP axis defined as described, is a reliable anatomical landmark for rotational alignment of tibial components.

Cite this article: Bone Joint J 2014; 96-B:1485–90.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 122 - 124
1 Nov 2014
Haidukewych GJ Petrie JR Adigweme O

Although the vast majority of patients that undergo total knee replacement have satisfactory outcomes with a generally low complication rate, occasionally a patient will be encountered that has had multiple failed surgeries, and now reaches a crossroad as to whether limb salvage will be acceptable or not.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):122–4.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 96 - 100
1 Nov 2014
Nam D Nunley RM Barrack RL

A national, multi-centre study was designed in which a questionnaire quantifying the degree of patient satisfaction and residual symptoms in patients following total knee replacement (TKR) was administered by an independent, blinded third party survey centre. A total of 90% of patients reported satisfaction with the overall functioning of their knee, but 66% felt their knee to be ‘normal’, with the reported incidence of residual symptoms and functional problems ranging from 33% to 54%. Female patients and patients from low-income households had increased odds of reporting dissatisfaction. Neither the use of contemporary implant designs (gender-specific, high-flex, rotating platform) or custom cutting guides (CCG) with a neutral mechanical axis target improved patient-perceived outcomes. However, use of a CCG to perform a so-called kinematically aligned TKR showed a trend towards more patients reporting their knee to feel ‘normal’ when compared with a so called mechanically aligned TKR

This data shows a degree of dissatisfaction and residual symptoms following TKR, and that several recent modifications in implant design and surgical technique have not improved the current situation.

Cite this article: Bone Joint J 2014;96-B(11 Suppl A):96–100.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1491 - 1497
1 Nov 2014
Howells NR Salmon L Waller A Scanelli J Pinczewski LA

The aim of this study was to examine the functional outcome at ten years following lateral closing wedge high tibial osteotomy for medial compartment osteoarthritis of the knee and to define pre-operative predictors of survival and determinants of functional outcome.

164 consecutive patients underwent high tibial osteotomy between 2000 and 2002. A total of 100 patients (100 knees) met the inclusion criteria and 95 were available for review at ten years. Data were collected prospectively and included patient demographics, surgical details, long leg alignment radiographs, Western Ontario and McMaster Universities osteoarthritis index (WOMAC) and Knee Society scores (KSS) pre-operatively and at five and ten years follow-up.

At ten years, 21 patients had been revised at a mean of five years. Overall Kaplan–Meier survival was 87% (95% confidence interval (CI) 81 to 94) and 79% (95% CI 71 to 87) at five and ten years, respectively. When compared with unrevised patients, those who had been revised had significantly lower mean pre-operative WOMAC Scores (47 (21 to 85) vs 65 (32 to 99), p < 0.001), higher mean age (54 yrs (42 to 61) vs 49 yrs (26 to 66), p = 0.006) and a higher mean BMI (30.2; 25 to 39 vs 27.9; 21 to 36, p = 0.005). Each were found to be risk factors for revision, with hazard ratios of 10.7 (95% CI 4 to 28.6; pre-operative WOMAC < 45), 6.5 (95% CI 2.4 to 17.7; age > 55) and 3.0 (95%CI 1.2 to 7.6; BMI > 30). Survival of patients with pre-operative WOMAC > 45, age < 55 and BMI < 30 was 97% at five and ten years. WOMAC and KSS in surviving patients improved significantly between pre-operative (mean 61; 32 to 99) and five (mean 88; 35 to 100, p = 0.001) and ten years (mean 84; 38 to 100, p = 0.001). Older patients had better functional outcomes overall, despite their higher revision rate.

This study has shown that improved survival is associated with age < 55 years, pre-operative WOMAC scores > 45 and, a BMI < 30. In patients over 55 years of age with adequate pre-operative functional scores, survival can be good and functional outcomes can be significantly better than their younger counterparts. We recommend the routine use of pre-operative functional outcome scores to guide decision-making when considering suitability for high tibial osteotomy.

Cite this article: Bone Joint J 2014;96-B:1491–7.


The Bone & Joint Journal
Vol. 96-B, Issue 11_Supple_A | Pages 78 - 83
1 Nov 2014
Gustke KA

Total knee replacement (TKR) smart tibial trials have load-bearing sensors which will show quantitative compartment pressure values and femoral-tibial tracking patterns. Without smart trials, surgeons rely on feel and visual estimation of imbalance to determine if the knee is optimally balanced. Corrective soft-tissue releases are performed with minimal feedback as to what and how much should be released. The smart tibial trials demonstrate graphically where and how much imbalance is present, so that incremental releases can be performed. The smart tibial trials now also incorporate accelerometers which demonstrate the axial alignment. This now allows the surgeon the option to perform a slight recut of the tibia or femur to provide soft-tissue balance without performing soft-tissue releases. Using a smart tibial trial to assist with soft-tissue releases or bone re-cuts, improved patient outcomes have been demonstrated at one year in a multicentre study of 135 patients (135 knees).