Advertisement for orthosearch.org.uk
Results 41 - 44 of 44
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1197 - 1202
1 Sep 2005
Fawzy E Mandellos G De Steiger R McLardy-Smith P Benson MKD Murray D

We followed up 76 consecutive hips with symptomatic acetabular dysplasia treated by acetabular shelf augmentation for a mean period of 11 years. Survival analysis using conversion to hip replacement as an end-point was 86% at five years and 46% at ten years. Forty-four hips with slight or no narrowing of the joint space pre-operatively had a survival of 97% at five and 75% at ten years. This was significantly higher (p = 0.0007) than that of the 32 hips with moderate or severe narrowing of the joint-space, which was 76% at five and 22% at ten years. There was no significant relationship between survival and age (p = 0.37) or the pre- and post-operative centre-edge (p = 0.39) and acetabular angles (p = 0.85).

Shelf acetabuloplasty is a reliable, safe procedure offering medium-term symptomatic relief for adults with acetabular dysplasia. The best results were achieved in patients with mild and moderate dysplasia of the hip with little arthritis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 4 | Pages 441 - 448
1 Apr 2007
Vendittoli P Mottard S Roy AG Dupont C Lavigne M

We evaluated the concentrations of chromium and cobalt ions in blood after metal-on-metal surface replacement arthroplasty using a wrought-forged, high carbon content chromium-cobalt alloy implant in 64 patients. At one year, mean whole blood ion levels were 1.61 μg/L (0.4 to 5.5) for chromium and 0.67 μg/L (0.23 to 2.09) for cobalt. The pre-operative ion levels, component size, female gender and the inclination of the acetabular component were inversely proportional to the values of chromium and/or cobalt ions at one year postoperatively. Other factors, such as age and level of activity, did not correlate with the levels of metal ions. We found that the levels of the ions in the serum were 1.39 and 1.37 times higher for chromium and cobalt respectively than those in the whole blood.

The levels of metal ions obtained may be specific to the hip resurfacing implant and reflect its manufacturing process.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 6 | Pages 721 - 726
1 Jun 2006
Girard J Lavigne M Vendittoli P Roy AG

We have compared the biomechanical nature of the reconstruction of the hip in conventional total hip arthroplasty (THA) and surface replacement arthroplasty (SRA) in a randomised study involving 120 patients undergoing unilateral primary hip replacement. The contralateral hip was used as a control.

Post-operatively, the femoral offset was significantly increased with THA (mean 5.1 mm; −2.8 to 11.6) and decreased with SRA (mean −3.3 mm; −8.9 to 8.2). Femoral offset was restored within sd 4 mm in 14 (25%) of those with THA and in 28 (57%) of the patients receiving SRA (p < 0.001). In the THA group, the leg was lengthened by a mean of 2.6 mm (−6.04 to +12.9), whereas it was shortened by a mean of 1.9 mm (−7.1 to +2.05) in the SRA group, compared with the contralateral side. Leg-length inequality was restored within sd 4 mm in 42 (86%) of the SRA and 33 (60%) of the THA patients. The radiological parameters of acetabular reconstruction were similar in both groups.

Restoration of the normal proximal femoral anatomy was more precise with SRA. The enhanced stability afforded by the use of a large-diameter femoral head avoided over-lengthening of the limb or increased offset to improve soft-tissue tension as occurs sometimes in THA. In a subgroup of patients with significant pre-operative deformity, restoration of the normal hip anatomy with lower pre-operative femoral offset or significant shortening of the leg was still possible with SRA.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 9 | Pages 1149 - 1154
1 Sep 2007
Lian Y Yoo M Pei F Cho Y Cheng J Chun S

We performed 52 total hip replacements in 52 patients using a cementless acetabular component combined with a circumferential osteotomy of the medial acetabular wall for the late sequelae of childhood septic arthritis of the hip. The mean age of the patients at operation was 44.5 years (22 to 66) and the mean follow-up was 7.8 years (5 to 11.8). The mean improvement in the Harris Hip Score was 29.6 points (19 to 51) at final follow-up. The mean cover of the acetabular component was 98.5% (87.8% to 100%). The medial acetabular wall was preserved with a mean thickness of 8.3 mm (1.7 to 17.4) and the mean length of abductor lever arm increased from 43.4 mm (19.1 to 62) to 54.2 mm (36.5 to 68.6). One acetabular component was revised for loosening and osteolysis 4.5 years postoperatively, and one had radiolucent lines in all acetabular zones at final review. Kaplan-Meier survival was 94.2% (95% confidence interval 85.8% to 100%) at 7.3 years, with revision or radiological loosening as an end-point when two hips were at risk.

A cementless acetabular component combined with circumferential medial acetabular wall osteotomy provides favourable results for acetabular reconstruction in patients who present with late sequelae of childhood septic hip arthritis.