Noise generation has been reported with ceramic-on-ceramic
articulations in total hip replacement (THR). This study evaluated
208 consecutive Delta Motion THRs at a mean follow-up of 21 months
(12 to 35). There were 141 women and 67 men with a mean age of 59
years (22 to 84). Patients were reviewed clinically and radiologically,
and the incidence of noise was determined using a newly described
assessment method. Noise production was examined against range of
movement, ligamentous laxity, patient-reported outcome scores, activity
level and orientation of the acetabular component. There were 143
silent hips (69%), 22 (11%) with noises other than squeaking, 17
(8%) with unreproducible squeaking and 26 (13%) with reproducible
squeaking. Hips with reproducible squeaking had a greater mean range
of movement (p <
0.001) and mean ligament laxity (p = 0.004), smaller
median head size (p = 0.01) and decreased mean acetabular component
inclination (p = 0.02) and anteversion angle (p = 0.02) compared
with the other groups. There was no relationship between squeaking
and age (p = 0.13), height (p = 0.263), weight (p = 0.333), body
mass index (p = 0.643), gender (p = 0.07) or patient outcome score
(p = 0.422). There were no revisions during follow-up. Despite the
surprisingly high incidence of squeaking, all patients remain satisfied
with their hip replacement. Cite this article:
Since the introduction of the National Institute
for Health and Care Excellence (NICE) guidelines on thromboprophylaxis
and the use of extended thromboprophylaxis with new oral agents,
there have been reports of complications arising as a result of
their use. We have looked at the incidence of wound complications
after the introduction of dabigatran for thromboprophylaxis in our
unit. We investigated the rate of venous thromboembolism and wound
leakage in 1728 patients undergoing primary joint replacement, both
before and after the introduction of dabigatran, and following its
subsequent withdrawal from our unit. We found that the use of dabigatran led to a significant increase
in post-operative wound leakage (20% with dabigatran, 5% with a
multimodal regimen; p <
0.001), which also resulted in an increased
duration of hospital stay. The rate of thromboembolism in patients
receiving dabigatran was higher (1.3%) than in those receiving the multimodal
thromboprophylaxis regimen, including low molecular weight heparin
as an inpatient and the extended use of aspirin (0.3%, p = 0.047).
We have ceased the use of dabigatran for thromboprophylaxis in these
patients. Cite this article:
We are currently facing an epidemic of periprosthetic
fractures around the hip. They may occur either during surgery or
post-operatively. Although the acetabulum may be involved, the femur
is most commonly affected. We are being presented with new, difficult
fracture patterns around cemented and cementless implants, and we
face the challenge of an elderly population who may have grossly
deficient bone and may struggle to rehabilitate after such injuries.
The correct surgical management of these fractures is challenging.
This article will review the current choices of implants and techniques
available to deal with periprosthetic fractures of the femur. Cite this article:
We undertook a randomised controlled trial to
compare bipolar hemiarthroplasty (HA) with a novel total hip replacement
(THR) comprising a polycarbonate–urethane (PCU) acetabular component
coupled with a large-diameter metal femoral head for the treatment
of displaced fractures of the femoral neck in elderly patients. Functional
outcome, assessed with the Harris hip score (HHS) at three months
and then annually after surgery, was the primary endpoint. Rates
of revision and complication were secondary endpoints. Based on a power analysis, 96 consecutive patients aged >
70
years were randomised to receive either HA (49) or a PCU-THR (47).
The mean follow-up was 30.1 months (23 to 50) and 28.6 months (22
to 52) for the HA and the PCU group, respectively. The HHS showed no statistically significant difference between
the groups at every follow-up. Higher pain was recorded in the PCU
group at one and two years’ follow-up
(p = 0.006 and p = 0.019, respectively). In the HA group no revision
was performed. In the PCU-THR group six patients underwent revision
and one patient is currently awaiting
re-operation. The three-year survival rate of the PCU-THR group
was 0.841 (95% confidence interval 0.680 to 0.926). Based on our findings we do not recommend the use of the PCU
acetabular component as part of the treatment of patients with fractures
of the femoral neck. Cite this article:
The accuracy and precision of two new methods of model-based
radiostereometric analysis (RSA) were hypothesised to be superior
to a plain radiograph method in the assessment of polyethylene (PE)
wear. A phantom device was constructed to simulate three-dimensional
(3D) PE wear. Images were obtained consecutively for each simulated
wear position for each modality. Three commercially available packages
were evaluated: model-based RSA using laser-scanned cup models (MB-RSA),
model-based RSA using computer-generated elementary geometrical
shape models (EGS-RSA), and PolyWare. Precision (95% repeatability
limits) and accuracy (Root Mean Square Errors) for two-dimensional
(2D) and 3D wear measurements were assessed.Objectives
Methods
A variety of radiological methods of measuring
version of the acetabular component after total hip replacement (THR)
have been described. The aim of this study was to evaluate the reliability
and validity of six methods (those of Lewinnek; Widmer; Hassan et
al; Ackland, Bourne and Uhthoff; Liaw et al; and Woo and Morrey)
that are currently in use. In 36 consecutive patients who underwent
THR, version of the acetabular component was measured by three independent
examiners on plain radiographs using these six methods and compared
with measurements using CT scans. The intra- and interobserver reliabilities
of each measurement were estimated. All measurements on both radiographs
and CT scans had excellent intra- and interobserver reliability
and the results from each of the six methods correlated well with
the CT measurements. However, measurements made using the methods
of Widmer and of Ackland, Bourne and Uhthoff were significantly
different from the CT measurements (both p <
0.001), whereas
measurements made using the remaining four methods were similar
to the CT measurements. With regard to reliability and convergent
validity, we recommend the use of the methods described by Lewinnek,
Hassan et al, Liaw et al and Woo and Morrey for measurement of version
of the acetabular component.
We determined the midterm survival, incidence
of peri-prosthetic fracture and the enhancement of the width of
the femur when combining struts and impacted bone allografts in
24 patients (25 hips) with severe femoral bone loss who underwent
revision hip surgery. The pre-operative diagnosis was aseptic loosening
in 16 hips, second-stage reconstruction in seven, peri-prosthetic
fracture in one and stem fracture in one hip. A total of 14 hips
presented with an Endoklinik grade 4 defect and 11 hips a grade
3 defect. The mean pre-operative Merle D’Aubigné and Postel score
was 5.5 points (1 to 8). The survivorship was 96% (95% confidence interval 72 to 98) at
a mean of 54.5 months (36 to 109). The mean functional score was
17.3 points (16 to 18). One patient in which the strut did not completely
bypass the femoral defect was further revised using a long cemented
stem due to peri-prosthetic fracture at six months post-operatively.
The mean subsidence of the stem was 1.6 mm (1 to 3). There was no
evidence of osteolysis, resorption or radiolucencies during follow-up
in any hip. Femoral width was enhanced by a mean of 41% (19% to
82%). A total of 24 hips had partial or complete bridging of the
strut allografts. This combined biological method was associated with a favourable
survivorship, a low incidence of peri-prosthetic fracture and enhancement
of the width of the femur in revision total hip replacement in patients
with severe proximal femoral bone loss.
Satisfactory primary wound healing following
total joint replacement is essential. Wound healing problems can
have devastating consequences for patients. Assessment of their healing
capacity is useful in predicting complications. Local factors that
influence wound healing include multiple previous incisions, extensive
scarring, lymphoedema, and poor vascular perfusion. Systemic factors
include diabetes mellitus, inflammatory arthropathy, renal or liver
disease, immune compromise, corticosteroid therapy, smoking, and
poor nutrition. Modifications in the surgical technique are necessary
in selected cases to minimise potential wound complications. Prompt
and systematic intervention is necessary to address any wound healing
problems to reduce the risks of infection and other potential complications. Cite this article:
The technique of femoral cement-in-cement revision
is well established, but there are no previous series reporting its
use on the acetabular side at the time of revision total hip replacement.
We describe the technique and report the outcome of 60 consecutive
acetabular cement-in-cement revisions in 59 patients at a mean follow-up
of 8.5 years (5 to 12). All had a radiologically and clinically
well-fixed acetabular cement mantle at the time of revision. During
the follow-up 29 patients died, but no hips were lost to follow-up.
The two most common indications for acetabular revision were recurrent
dislocation (46, 77%) and to complement femoral revision (12, 20%). Of the 60 hips, there were two cases of aseptic loosening of
the acetabular component (3.3%) requiring re-revision. No other
hip was clinically or radiologically loose (96.7%) at the latest
follow-up. One hip was re-revised for infection, four for recurrent
dislocation and one for disarticulation of a constrained component.
At five years the Kaplan-Meier survival rate was 100% for aseptic
loosening and 92.2% (95% CI 84.8 to 99.6), with revision for any cause
as the endpoint. These results support the use of cement-in-cement revision on
the acetabular side in appropriate cases. Theoretical advantages
include preservation of bone stock, reduced operating time, reduced
risk of complications and durable fixation.
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
We report a case of a male patient presenting
with bilateral painful but apparently well-positioned and -fixed
large-diameter metal-on-metal hip replacements four years post-operatively.
Multiple imaging modes revealed a thick-walled, cystic expansile
mass in communication with the hip joint (a pseudotumour). Implant
retrieval analysis and tissue culture eliminated high bearing wear
or infection as causes for the soft-tissue reaction, but noted marked corrosion
of the modular neck taper adaptor and corrosion products in the
tissues. Therefore, we believe corrosion products from the taper
caused by mismatch of the implant components led to pseudotumour
formation requiring revision.
The outcome of total knee replacement (TKR) using
components designed to increase the range of flexion is not fully
understood. The short- to mid-term risk of aseptic revision in high
flexion TKR was evaluated. The endpoint of the study was aseptic
revision and the following variables were investigated: implant
design (high flexion In a cohort of 64 000 TKRs, high flexion components were used
in 8035 (12.5%). The high flexion knees with tibial liners of thickness
>
14 mm had a density of revision of 1.45/100 years of observation,
compared with 0.37/100 in non-high flexion TKR with liners ≤ 14
mm thick. Relative to a standard fixed PS TKR, the NexGen (Zimmer,
Warsaw, Indiana) Gender Specific Female high flexion fixed PS TKR
had an increased risk of revision (hazard ratio (HR) 2.27 (95% confidence
interval (CI) 1.48 to 3.50)), an effect that was magnified when
a thicker tibial insert was used (HR 8.10 (95% CI 4.41 to 14.89)). Surgeons should be cautious when choosing high flexion TKRs,
particularly when thicker tibial liners might be required. Cite this article:
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
Louis Pasteur once said that: “Fortune favours
the prepared mind.” As one of the great scientists who contributed
to the fight against infection, he emphasised the importance of
being prepared at all times to recognise infection and deal with
it. Despite the many scientific discoveries and technological advances,
such as the advent of antibiotics and the use of sterile techniques,
infection continues to be a problem that haunts orthopaedic surgeons
and inflicts suffering on patients. The medical community has implemented many practices with the
intention of preventing infection and treating it effectively when
it occurs. Although high-level evidence may support some of these
practices, many are based on little to no scientific foundation.
Thus, around the world, there is great variation in practices for
the prevention and management of periprosthetic joint infection. This paper summaries the instigation, conduct and findings of
a recent International Consensus Meeting on Surgical Site and Periprosthetic
Joint Infection. Cite this article:
An 81-year-old woman presented with a fracture
in the left femur. She had well-fixed bilateral hip replacements
and had received long-term bisphosphonate treatment. Prolonged bisphosphonate
use has been recently linked with atypical subtrochanteric and diaphyseal
femoral fractures. While the current definition of an atypical fracture
of the femur excludes peri-prosthetic fractures, this case suggests
that they do occur and should be considered in patients with severe
osteopenia. Union of the fracture followed cessation of bisphosphonates
and treatment with teriparatide. Thus, this case calls into question
whether prophylactic intramedullary nailing is sufficient alone
to treat early or completed atypical femoral fractures.
In order to compare the effect of oral apixaban
(a factor Xa inhibitor) with subcutaneous enoxaparin on major venous
thromboembolism and major and non-major clinically relevant bleeding
after total knee and hip replacement, we conducted a pooled analysis
of two previously reported double-blind randomised studies involving 8464
patients. One group received apixaban 2.5 mg twice daily (plus placebo
injection) starting 12 to 24 hours after operation, and the other
received enoxaparin subcutaneously once daily (and placebo tablets)
starting 12 hours (± 3) pre-operatively. Each regimen was continued
for 12 days ( Apixaban 2.5 mg twice daily is more effective than enoxaparin
40 mg once daily without increased bleeding.
Aims. Day-case knee and hip replacement, in which patients are discharged on the day of surgery, has been gaining popularity during the last two decades, and particularly since the COVID-19 pandemic. This systematic review presents the evidence comparing day-case to inpatient-stay surgery. Methods. A systematic literature search was performed of MEDLINE, Embase, and grey literature databases to include all studies which compare day-case with inpatient knee and hip replacement. Meta-analyses were performed where appropriate using a random effects model. The protocol was registered prospectively (PROSPERO CRD42023392811). Results. A total of 38 studies were included, with a total of 83,888 day-case procedures. The studies were predominantly from the USA and Canada, observational, and with a high risk of bias. Day-case patients were a mean of 2.08 years younger (95% CI 1.05 to 3.12), were more likely to be male (odds ratio (OR) 1.3 (95% CI 1.19 to 1.41)), and had a lower mean BMI and American Society of Anesthesiologists grades compared with inpatients. Overall, day-case surgery was associated with significantly lower odds of readmission (OR 0.83 (95% CI 0.73 to 0.96); p = 0.009), subsequent emergency department attendance (OR 0.62 (95% CI 0.48 to 0.79); p < 0.001), and complications (OR 0.7 (95% CI 0.55 to 0.89) p = 0.004), than inpatient surgery. There were no significant differences in the rates of reoperation or mortality. The overall rate of successful same-day discharge for day-case surgery was 85% (95% CI 81 to 88). Patient-reported outcome measures and cost-effectiveness were either equal or favoured day-case. Conclusion. Within the limitations of the literature, in particular the substantial risk of selection bias, the outcomes following day-case knee and hip replacement appear not to be inferior to those following an inpatient stay. The evidence is more robust for unicompartmental knee replacement (UKR) than for total knee replacement (TKR) or total hip replacement (THR). The rate of successful same-day discharge is highest in UKR, followed by TKR and comparatively lower in
Few studies have examined the order in which
a spinal osteotomy and total hip replacement (THR) are to be performed
for patients with ankylosing spondylitis. We have retrospectively
reviewed 28 consecutive patients with ankylosing spondylitis who
underwent both a spinal osteotomy and a
We investigated whether simultaneous bilateral sequential total hip replacement (THR) would increase the rate of mortality and complications compared with unilateral