Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA. A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy.Aims
Methods
The primary objective of this study was to compare migration of the cemented ATTUNE fixed bearing cruciate retaining tibial component with the cemented Press-Fit Condylar (PFC)-sigma fixed bearing cruciate retaining tibial component. The secondary objectives included comparing clinical and radiological outcomes and Patient Reported Outcome Measures (PROMs). A single blinded randomized, non-inferiority study was conducted including 74 patients. Radiostereometry examinations were made after weight bearing, but before hospital discharge, and at three, six, 12, and 24 months postoperatively. PROMS were collected preoperatively and at three, six, 12, and 24 months postoperatively. Radiographs for measuring radiolucencies were collected at two weeks and two years postoperatively.Aims
Methods
Although bone cement is the primary mode of fixation in total knee arthroplasty (TKA), cementless fixation is gaining interest as it has the potential of achieving lasting biological fixation. By 3D printing an implant, highly porous structures can be manufactured, promoting osseointegration into the implant to prevent aseptic loosening. This study compares the migration of cementless, 3D-printed TKA to cemented TKA of a similar design up to two years of follow-up using radiostereometric analysis (RSA) known for its ability to predict aseptic loosening. A total of 72 patients were randomized to either cementless 3D-printed or a cemented cruciate retaining TKA. RSA and clinical scores were evaluated at baseline and postoperatively at three, 12, and 24 months. A mixed model was used to analyze the repeated measurements.Aims
Methods
Cementless unicompartmental knee arthroplasty (UKA) has advantages over cemented UKA, including improved fixation, but has a higher risk of tibial plateau fracture, particularly in Japanese patients. The aim of this multicentre study was to determine when cementless tibial components could safely be used in Japanese patients based on the size and shape of the tibia. The study involved 212 cementless Oxford UKAs which were undertaken in 174 patients in six hospitals. The medial eminence line (MEL), which is a line parallel to the tibial axis passing through the tip of medial intercondylar eminence, was drawn on preoperative radiographs. Knees were classified as having a very overhanging medial tibial condyle if this line passed medial to the medial tibial cortex. They were also classified as very small if a size A/AA tibial component was used.Aims
Methods
Patient-specific instrumentation of total knee arthroplasty (TKA) is a technique permitting the targeting of individual kinematic alignment, but deviation from a neutral mechanical axis may have implications on implant fixation and therefore survivorship. The primary objective of this randomized controlled study was to compare the fixation of tibial components implanted with patient-specific instrumentation targeting kinematic alignment (KA+PSI) A total of 47 patients due to undergo TKA were randomized to KA+PSI (n = 24) or MA+CAS (n = 23). In the KA+PSI group, there were 16 female and eight male patients with a mean age of 64 years (Aims
Patients and Methods
It is unknown whether kinematic alignment (KA) objectively improves knee balance in total knee arthroplasty (TKA), despite this being the biomechanical rationale for its use. This study aimed to determine whether restoring the constitutional alignment using a restrictive KA protocol resulted in better quantitative knee balance than mechanical alignment (MA). We conducted a randomized superiority trial comparing patients undergoing TKA assigned to KA within a restrictive safe zone or MA. Optimal knee balance was defined as an intercompartmental pressure difference (ICPD) of 15 psi or less using a pressure sensor. The primary endpoint was the mean intraoperative ICPD at 10° of flexion prior to knee balancing. Secondary outcomes included balance at 45° and 90°, requirements for balancing procedures, and presence of tibiofemoral lift-off.Aims
Methods
The aims of this study were to determine the proportion of patients
with outlier varus or valgus alignment in kinematically aligned
total knee arthroplasty (TKA), whether those with outlier varus
or valgus alignment have higher forces in the medial or lateral
compartments of the knee than those with in-range alignment and
whether measurements of the alignment of the limb, knee and components
predict compartment forces. The intra-operative forces in the medial and lateral compartments
were measured with an instrumented tibial insert in 67 patients
who underwent a kinematically aligned TKA during passive movement.
The mean of the forces at full extension, 45° and 90° of flexion
determined the force in the medial and lateral compartments. Measurements
of the alignment of the limb and the components included the hip-knee-ankle
(HKA) angle, proximal medial tibial angle (PMTA), and distal lateral
femoral angle (DLFA). Measurements of the alignment of the knee
and the components included the tibiofemoral angle (TFA), tibial
component angle (TCA) and femoral component angle (FCA). Alignment was
measured on post-operative, non-weight-bearing anteroposterior (AP)
scanograms and categorised as varus or valgus outlier or in-range
in relation to mechanically aligned criteria.Aims
Patients and Methods
The purpose of the present study was to compare patient-specific instrumentation (PSI) and conventional surgical instrumentation (CSI) for total knee arthroplasty (TKA) in terms of early implant migration, alignment, surgical resources, patient outcomes, and costs. The study was a prospective, randomized controlled trial of 50 patients undergoing TKA. There were 25 patients in each of the PSI and CSI groups. There were 12 male patients in the PSI group and seven male patients in the CSI group. The patients had a mean age of 69.0 years (Aims
Patients and Methods
The optimal method of tibial component fixation remains uncertain
in total knee arthroplasty (TKA). Hydroxyapatite coatings have been
applied to improve bone ingrowth in uncemented designs, but may
only coat the directly accessible surface. As peri-apatite (PA)
is solution deposited, this may increase the coverage of the implant
surface and thereby fixation. We assessed the tibial component fixation
of uncemented PA-coated TKAs Patients were randomised to PA-coated or cemented TKAs. In 60
patients (30 in each group), radiostereometric analysis of tibial
component migration was evaluated as the primary outcome at baseline,
three months post-operatively and at one, two and five years. A
linear mixed-effects model was used to analyse the repeated measurements.Aims
Patients and Methods
The purpose of this study was to measure the
radiological parameters of femoral component alignment of the Oxford
Phase 3 unicompartmental knee replacement (UKR), and evaluate their
effect on clinical outcome. Multiple regression analysis was used
to examine the relative contributions of the radiological assessment
of femoral component alignment in 189 consecutive UKRs performed
by a single surgeon. The American Knee Society scores were compared
between groups, defined as being within or outside recommended tolerances
of the position of the femoral component. For the flexion/extension
position 21 UKRs (11.1%) lay outside the recommended limits, and for
posterior overhang of the femoral component nine (4.8%) lay outside
the range. The pre-operative hip/knee/ankle (HKA) angle, narrowest
canal distance from the distal femoral entry point of the alignment
jig and coronal entry-point position had significant effects on
the flexion/extension position. Pre-operative HKA angle had a significant
influence on posterior overhang of the femoral component. However,
there was no significant difference in American Knee Society scores
relative to the position of the femoral component.
We conducted a randomised controlled trial to assess the accuracy
of positioning and alignment of the components in total knee arthroplasty
(TKA), comparing those undertaken using standard intramedullary
cutting jigs and those with patient-specific instruments (PSI). There were 64 TKAs in the standard group and 69 in the PSI group. The post-operative hip-knee-ankle (HKA) angle and positioning
was investigated using CT scans. Deviation of >
3° from the planned
position was regarded as an outlier. The operating time, Oxford
Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.Aims
Patients and Methods
Our aim was to compare the degree of patellar descent and alteration in angle of the inclination of the tibial plateau in lateral closing-wedge and medial opening-wedge high tibial osteotomy (HTO) in 51 consecutive patients with osteoarthritis of the medial compartment and varus malalignment. Patellar height was measured by the Insall-Salvati (IS) and the Blackburne-Peel (BP) ratios. The tibial inclination was determined by the Moore-Harvey (MH) method. Multivariate linear regression analysis was used to determine the influence of the type of HTO (closing After an opening-wedge HTO the patellar height was significantly more decreased (mean post-operative difference: IS = 0.15; 95% confidence interval (CI) 0.06 to 0.23; BP = 0.11; 95% CI 0.05 to 0.18) compared with a closing-wedge HTO. The angle of tibial inclination differed significantly (mean post-operative difference MH = −6.40°; 95% CI −8.74 to −4.02) between the two HTO techniques, increasing after opening-wedge HTO and decreasing after closing-wedge HTO. There was no clinically-relevant difference in the intra- and interobserver variability of measurements of patellar height either before or after HTO.
The aim of this study was to evaluate the risk
factors for dislocation of the bearing after a mobile-bearing Oxford medial
unicompartmental knee replacement (UKR) and to test the hypothesis
that surgical factors, as measured from post-operative radiographs,
are associated with its dislocation From a total of 480 UKRs performed between 2001 and 2012, in
391 patients with a mean age of 66.5 years (45 to 82) (316 female,
75 male), we identified 17 UKRs where bearing dislocation occurred.
The post-operative radiological measurements of the 17 UKRs and
51 matched controls were analysed using conditional logistic regression analysis.
The post-operative radiological measurements included post-operative
change in limb alignment, the position of the femoral and tibial
components, the resection depth of the proximal tibia, and the femoral component-posterior
condyle classification. We concluded that a post-operative decrease in the posterior
tibial slope relative to the pre-operative value was the only significant
determinant of dislocation of the bearing after medial Oxford UKR
(odds ratio 1.881; 95% confidence interval 1.272 to 2.779). A post-operative
posterior tibial slope <
8.45° and a difference between the pre-operative
and post-operative posterior tibial slope of >
2.19° may increase
the risk of dislocation. Cite this article:
This study evaluated the results of a physeal-sparing technique of intra-articular anterior cruciate ligament (ACL) reconstruction in skeletally immature patients, with particular reference to growth disturbance. Between 1992 and 2007, 57 children with a mean age of 12.2 years (6.8 to 14.5) underwent ACL reconstruction using the same technique. At a mean of 5.5 years (2 to 14) after surgery, 56 patients underwent clinical and radiological evaluation. At that time, 49 patients (87.5%) had reached bony maturity and 53 (95%) achieved A or B according to the IKDC 2000 classification. Four patients had stopped participation in sports because of knee symptoms, and three patients (5.4%) had a subsequent recurrent ACL injury. There was no clinical or radiological evidence of growth disturbance after a mean growth in stature of 20.0 cm (3 to 38). This study demonstrates that ACL reconstruction sparing the physes in children is a safe technique protecting against meniscal tears and giving better results than reconstruction in adults, without causing significant growth disturbance.
We retrospectively reviewed the records of 1150
computer-assisted total knee replacements and analysed the clinical
and radiological outcomes of 45 knees that had arthritis with a
pre-operative recurvatum deformity. The mean pre-operative hyperextension
deformity of 11° (6° to 15°), as measured by navigation at the start
of the operation, improved to a mean flexion deformity of 3.1° (0°
to 7°) post-operatively. A total of 41 knees (91%) were managed
using inserts ≤ 12.5 mm thick, and none had mediolateral laxity
>
2 mm from a mechanical axis of 0° at the end of the surgery. At
a mean follow-up of 26.4 months (13 to 48) there was significant
improvement in the mean Knee Society, Oxford knee and Western Ontario
and McMaster Universities Osteoarthritis Index scores compared with
the pre-operative values. The mean knee flexion improved from 105°
(80° to 125°) pre-operatively to 131° (120° to 145°), and none of
the limbs had recurrent recurvatum. These early results show that total knee replacement using computer
navigation and an algorithmic approach for arthritic knees with
a recurvatum deformity can give excellent radiological and functional
outcomes without recurrent deformity.
Percutaneous physiodesis is an established technique for treating mild leg-length discrepancy and problems of expected extreme height. Angular deformities resulting from incomplete physeal arrest have been reported, and little is known about the time interval from percutaneous physiodesis to actual physeal arrest. This procedure was carried out in ten children, six with leg-length discrepancy and four with expected extreme height. Radiostereometric analysis was used to determine the three-dimensional dynamics of growth retardation. Errors of measurement of translation were less than 0.05 mm and of rotation less than 0.06°. Physeal arrest was obtained in all but one child within 12 weeks after physiodesis and no clinically-relevant angular deformities occurred. This is a suitable method for following up patients after percutaneous physiodesis. Incomplete physeal arrest can be detected at an early stage and the procedure repeated before corrective osteotomy is required.
We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation.
We have investigated the benefits of patient
specific instrument guides, applied to osteotomies around the knee. Single,
dual and triple planar osteotomies were performed on tibias or femurs
in 14 subjects. In all patients, a detailed pre-operative plan was
prepared based upon full leg standing radiographic and CT scan information.
The planned level of the osteotomy and open wedge resection was
relayed to the surgery by virtue of a patient specific guide developed
from the images. The mean deviation between the planned wedge angle
and the executed wedge angle was 0° (-1 to 1, Cite this article: