There is great variability in acetabular component
orientation following hip replacement. The aims of this study were
to compare the component orientation at impaction with the orientation
measured on post-operative radiographs and identify factors that
influence the difference between the two. A total of 67 hip replacements
(52 total hip replacements and 15 hip resurfacings) were prospectively
studied. Intra-operatively, the orientation of the acetabular component
after impaction relative to the operating table was measured using
a validated stereo-photogrammetry protocol. Post-operatively, the
radiographic orientation was measured; the mean inclination/anteversion
was 43° ( This study demonstrated that in order to achieve a specific radiographic
orientation target, surgeons should implant the acetabular component
5° less inclined and 8° more anteverted than their target. Great
variability (2 Cite this article:
The April 2014 Knee Roundup360 looks at: mobile compression as good as chemical thromboprophylaxis; patellar injury with MIS knee surgery; tibial plateau fracture results not as good as we thought; back and knee pain; metaphyseal sleeves may be the answer in revision knee replacement; oral tranexamic acid; gentamycin alone in antibiotic spacers; and whether the jury is still out on unloader braces.
We reviewed the clinical outcome of arthroscopic femoral osteochondroplasty for cam femoroacetabular impingement performed between August 2005 and March 2009 in a series of 40 patients over 60 years of age. The group comprised 26 men and 14 women with a mean age of 65 years (60 to 82). The mean follow-up was 30 months (12 to 54). The mean modified Harris hip score improved by 19.2 points (95% confidence interval 13.6 to 24.9; p <
0.001) while the mean non-arthritic hip score improved by 15.0 points (95% confidence interval 10.9 to 19.1, p <
0.001). Seven patients underwent total hip replacement after a mean interval of 12 months (6 to 24 months) at a mean age of 63 years (60 to 70). The overall level of satisfaction was high with most patients indicating that they would undergo similar surgery in the future to the contralateral hip, if indicated. No serious complications occurred. Arthroscopic femoral osteochondroplasty performed in selected patients over 60 years of age, who have hip pain and mechanical symptoms resulting from cam femoroacetabular impingement, is beneficial with a minimal risk of complications at a mean follow-up of 30 months.
Cubitus varus is the most frequent complication
following the treatment of supracondylar humeral fractures in children.
We investigated data from publications reporting on the surgical
management of cubitus varus found in electronic searches of Ovid/MEDLINE
and Cochrane Library databases. In 894 children from 40 included
studies, the mean age at initial injury was 5.7 years (3 to 8.6)
and 9.8 years (4 to 15.7) at the time of secondary correction. The four
osteotomy techniques were classified as lateral closing wedge, dome,
complex (multiplanar) and distraction osteogenesis. A mean angular
correction of 27.6º (18.5° to 37.0°) was achieved across all classes
of osteotomy. The meta-analytical summary estimate for overall rate
of good to excellent results was 87.8% (95% CI 84.4 to 91.2). No technique
was shown to significantly affect the surgical outcome, and the
risk of complications across all osteotomy classes was 14.5% (95%
CI 10.6 to 18.5). Nerve palsies occurred in 2.53% of cases (95%
CI 1.4 to 3.6), although 78.4% were transient. No one technique
was found to be statistically safer or more effective than any other. Cite this article:
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
The April 2014 Trauma Roundup360 looks at: is it safe to primarily close dog bite wounds?; conservative transfusion evidence based in hip fracture surgery; tibial nonunion is devastating to quality of life; sexual dysfunction after traumatic pelvic fracture; hemiarthroplasty versus fixation in displaced femoral neck fractures; silver VAC dressings “Gold Standard” in massive wounds; dual plating for talar neck fracture; syndesmosis and fibular length easiest errors in ankle fracture surgery; and dual mobility: stable as a rock in fracture.
An international faculty of orthopaedic surgeons
presented their work on the current challenges in hip surgery at
the London Hip Meeting which was attended by over
400 delegates. The topics covered included femoroacetabular impingement, thromboembolic
phenomena associated with hip surgery, bearing surfaces (including metal-on-metal
articulations), outcomes of hip replacement surgery and revision
hip replacement. We present a concise report of the current opinions
on hip surgery from this meeting with appropriate references to
the current literature.
We have investigated the accuracy of the templating of digital radiographs in planning total hip replacement using two common object-based calibration methods with the ball placed laterally (method 1) or medially (method 2) and compared them with two non-object-based methods. The latter comprised the application of a fixed magnification of 121% (method 3) and calculation of magnification based on the object-film-distance (method 4). We studied the post-operative radiographs of 57 patients (19 men, 38 women, mean age 73 years (53 to 89)) using the measured diameter of the prosthetic femoral head and comparing it with the true value. Both object-based methods (1 and 2) produced large errors (mean/maximum: 2.55%/17.4% and 2.04%/6.46%, respectively). Method 3 applying a fixed magnification and method 4 (object-film-distance) produced smaller errors (mean/maximum 1.42%/5.22% and 1.57%/4.24%, respectively; p <
0.01). The latter results were clinically relevant and acceptable when planning was allowed to within one implant size. Object-based calibration (methods 1 and 2) has fundamental problems with the correct placement of the calibration ball. The accuracy of the fixed magnification (method 3) matched that of object-film-distance (method 4) and was the most reliable and efficient calibration method in digital templating.
Antegrade nailing of proximal humeral fractures
using a straight nail can damage the bony insertion of the supraspinatus
tendon and may lead to varus failure of the construct. In order
to establish the ideal anatomical landmarks for insertion of the
nail and their clinical relevance we analysed CT scans of bilateral
proximal humeri in 200 patients (mean age 45.1 years ( We therefore emphasise the need for ‘fastidious’ pre-operative
planning to minimise this risk. Cite this article:
We have reviewed 42 patients who had revision of metal-on-metal resurfacing procedures, mostly because of problems with the acetabular component. The revisions were carried out a mean of 26.2 months (1 to 76) after the initial operation and most of the patients (30) were female. Malpositioning of the acetabular component resulted in 27 revisions, mostly because of excessive abduction (mean 69.9°; 56° to 98°) or insufficient or excessive anteversion. Seven patients had more than one reason for revision. The mean increase in the diameter of the component was 1.8 mm (0 to 4) when exchange was needed. Malpositioning of the components was associated with metallosis and a high level of serum ions. The results of revision of the femoral component to a component with a modular head were excellent, but four patients had dislocation after revision and four required a further revision.
We report a prospective analysis of clinical
outcome in patients treated with medial patellofemoral ligament
(MPFL) reconstruction using an autologous semitendinosus graft.
The technique includes superolateral portal arthroscopic assessment
before and after graft placement to ensure correct graft tension
and patellar tracking before fixation. Between October 2005 and
October 2010, a total of 201 consecutive patients underwent 219 procedures.
Follow-up is presented for 211 procedures in 193 patients with a
mean age of 26 years (16 to 49), and mean follow-up of 16 months
(6 to 42). Indications were atraumatic recurrent patellar dislocation
in 141 patients, traumatic recurrent dislocation in 50, pain with
subluxation in 14 and a single dislocation with persistent instability
in six. There have been no recurrent dislocations/subluxations.
There was a statistically significant improvement between available pre-
and post-operative outcome scores for 193 patients (all p <
0.001).
Female patients with a history of atraumatic recurrent dislocation
and all patients with history of previous surgery had a significantly
worse outcome (all p <
0.05). The indication for surgery, degree
of dysplasia, associated patella alta, time from primary dislocation
to surgery and evidence of associated cartilage damage at operation
did not result in any significant difference in outcome. This series adds considerably to existing evidence that MPFL
reconstruction is an effective surgical procedure for selected patients
with patellofemoral instability.
In this study, we evaluated patient-reported
outcomes, the rate of revision and the indications for revision
following resurfacing hemiarthroplasty of the shoulder in patients
with osteoarthritis. All patients with osteoarthritis who underwent primary resurfacing
hemiarthroplasty and reported to the Danish Shoulder Arthroplasty
Registry (DSR), between January 2006 and December 2010 were included.
There were 772 patients (837 arthroplasties) in the study. The Western
Ontario Osteoarthritis of the Shoulder (WOOS) index was used to
evaluate patient-reported outcome 12 months (10 to 14) post-operatively.
The rates of revision were calculated from the revisions reported
to the DSR up to December 2011 and by checking deaths with the Danish National
Register of Persons. A complete questionnaire was returned by 688 patients (82.2%).
The mean WOOS was 67 (0 to 100). A total of 63 hemiarthroplasties
(7.5%) required revision; the cumulative five-year rate of revision
was 9.9%. Patients aged <
55 years had a statistically significant
inferior WOOS score, which exceeded the minimal clinically important
difference, compared with older patients (mean difference 14.2 (8.8;
95% CI 19.6; p <
0.001), but with no increased risk of revision.
There was no significant difference in the mean WOOS or the risk
of revision between designs of resurfacing hemiarthroplasty. Cite this article:
We compared the clinical and radiological outcomes
of two cementless femoral stems in the treatment of patients with
a Garden III or IV fracture of the femoral neck. A total of 70 patients At final follow-up there were no statistically significant differences
between the short anatomical and the conventional stems with regard
to the mean Harris hip score (85.7 (66 to 100) Our study demonstrated that despite the poor bone quality in
these elderly patients with a fracture of the femoral neck, osseo-integration
was obtained in all hips in both groups. However, the incidence
of thigh pain, pulmonary microemboli and peri-prosthetic fracture
was significantly higher in the conventional stem group than in
the short stem group.
In 2012 we reviewed a consecutive series of 92
uncemented THRs performed between 1986 and 1991 at our institution
using the CLS Spotorno stem, in order to assess clinical outcome
and radiographic data at a minimum of 21 years. The series comprised
92 patients with a mean age at surgery of 59.6 years (39 to 77)
(M:F 43;49). At the time of this review, seven (7.6%) patients had died and
two (2.2%) were lost to follow-up. The 23-year Kaplan–Meier survival
rates were 91.5% (95% confidence intervals (CI) 85.4% to 97.6%;
55 hips at risk) and 80.3% (95% CI, 71.8% to 88.7%; 48 hips at risk)
respectively, with revision of the femoral stem or of any component
as endpoints. At the time of this review, 76 patients without stem
revision were assessed clinically and radiologically (mean follow-up
24.0 years (21.5 to 26.5)). For the 76 unrevised hips the mean Harris
hip score was 87.1 (65 to 97). Femoral osteolysis was detected in
five hips (6.6%) only in Gruen zone 7. Undersized stems were at
higher risk of revision owing to aseptic loosening (p = 0.0003).
Patients implanted with the stem in a varus position were at higher risk
of femoral cortical hypertrophy and thigh pain (p = 0.0006 and p
= 0.0007, respectively). In our study, survival, clinical outcome and radiographic data
remained excellent in the third decade after implantation. Nonetheless,
undersized stems were at higher risk of revision owing to aseptic
loosening. Cite this article:
We describe the results at five years of a prospective study of a new tri-tapered polished, cannulated, cemented femoral stem implanted in 51 patients (54 hips) with osteoarthritis. The mean age and body mass index of the patients was 74 years and 27.9, respectively. Using the anterolateral approach, half of the stems were implanted by a consultant orthopaedic surgeon and half by six different registrars. There were three withdrawals from the study because of psychiatric illness, a deep infection and a recurrent dislocation. Five deaths occurred prior to five-year follow-up and one patient withdrew from clinical review. In the remaining 51 hips the mean pre-operative Oxford hip score was 47 points which decreased to 19 points at five years (45 hips). Of the stems 49 (98%) were implanted within 1° of neutral in the femoral canal. The mean migration of the stem at five years was 1.9 mm and the survivorship for aseptic loosening was 100%. There was no significant difference in outcome between the consultant and registrar groups. At five years, the results were comparable with those of other polished, tapered, cemented stems. Long-term surveillance continues.
Two-stage exchange remains the gold standard
for treatment of peri-prosthetic joint infection after total hip replacement
(THR). In the first stage, all components and associated cement
if present are removed, an aggressive debridement is undertaken
including a complete synovectomy, and an antibiotic-loaded cement
spacer is put in place. Patients are then treated with six weeks
of parenteral antibiotics, followed by an ‘antibiotic free period’
to help ensure the infection has been eradicated. If the clinical
evaluation and serum inflammatory markers suggest the infection
has resolved, then the second stage can be completed, which involves
removal of the cement spacer, repeat debridement, and placement
of a new THR. Cite this article:
The removal of well-fixed bone cement from the femoral canal during revision of a total hip replacement (THR) can be difficult and risks the loss of excessive bone stock and perforation or fracture of the femoral shaft. Retaining the cement mantle is attractive, yet the technique of cement-in-cement revision is not widely practised. We have used this procedure at our hospital since 1989. The stems were removed to gain a better exposure for acetabular revision, to alter version or leg length, or for component incompatibility. We studied 136 hips in 134 patients and followed them up for a mean of eight years (5 to 15). A further revision was required in 35 hips (25.7%), for acetabular loosening in 26 (19.1%), sepsis in four, instability in three, femoral fracture in one and stem fracture in one. No femoral stem needed to be re-revised for aseptic loosening. A cement-in-cement revision of the femoral stem is a reliable technique in the medium term. It also reduces the risk of perforation or fracture of the femoral shaft.
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
We report the results of 62 hips in 62 patients
(17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent
revision of the acetabular component of a total hip replacement
due to aseptic loosening between May 2003 and November 2007. All
hips had a Paprosky type IIIa acetabular defect. Acetabular revision
was undertaken using a Procotyl E cementless oblong implant with
modular side plates and a hook combined with impaction allografting. At a mean follow-up of 60.5 months (36 to 94) with no patients
lost to follow-up and one died due to unrelated illness, the complication
rate was 38.7%. Complications included aseptic loosening (19 hips),
deep infection (3 hips), broken hook and side plate (one hip) and
a femoral nerve palsy (one hip). Further revision of the acetabular component
was required in 18 hips (29.0%) and a further four hips (6.4%) are
currently loose and awaiting revision. We observed unacceptably high rates of complication and failure
in our group of patients and cannot recommend this implant or technique.