Metal-on-metal (MoM) hip resurfacing was introduced into clinical
practice because it was perceived to be a better alternative to
conventional total hip replacement for young and active patients.
However, an increasing number of reports of complications have arisen
focusing on design and orientation of the components, the generation
of metallic wear particles and serum levels of metallic ions. The
procedure introduced a combination of two elements: large-dimension
components and hard abrasive particles of metal wear. The objective
of our study was to investigate the theory that microseparation
of the articular surfaces draws in a high volume of bursal fluid
and its contents into the articulation, and at relocation under
load would generate high pressures of fluid ejection, resulting
in an abrasive water jet. This theoretical concept using MoM resurfacing components (head
diameter 55 mm) was modelled mathematically and confirmed experimentally
using a material-testing machine that pushed the head into the cup
at a rate of 1000 mm/min until fully engaged.Objectives
Methods
We investigated the relationship of the inferior gluteal nerve to gluteus maximus by dissecting the muscle in 12 fresh-frozen and formalin-treated cadavers. The anatomy was recorded using still digital photography. The course of the inferior gluteal nerve was carefully traced and was noted to enter the deep surface of gluteus maximus approximately 5 cm from the tip of the greater trochanter of the femur. The susceptibility of the nerve to injury during a posterior approach to the hip may be explained by its close relationship to the deep surface of gluteus maximus. It is easily damaged before it has been seen if the muscle is split and parted more than 5 cm from the tip of the greater trochanter of the femur. We suggest that a modified posterior approach be used to expose the hip to avoid damage to this nerve.
Alumina–alumina bearings are among the most resistant
to wear in total hip replacement. Examination of their surfaces
is one way of comparing damage caused by wear of hip joints simulated We found that long-term alumina wear in association with a loose
acetabular component could be categorised into three groups. Of
20 specimens, four had ‘low wear’, eight ‘crescent wear’ and eight
‘severe wear’, which was characterised by a change in the physical
shape of the bearing and a loss of volume. This suggests that the
wear in alumina–alumina bearings in association with a loose acetabular
component may be variable in pattern, and may explain, in part,
why the wear of a ceramic head
The aim of this study was to compare a third-generation
cementing procedure for glenoid components with a new technique
for cement pressurisation. In 20 pairs of scapulae, 20 keeled and
20 pegged glenoid components were implanted using either a third-generation
cementing technique (group 1) or a new pressuriser (group 2). Cement penetration
was measured by three-dimensional (3D) analysis of micro-CT scans.
The mean 3D depth of penetration of the cement was significantly
greater in group 2 (p <
0.001). The mean thickness of the cement
mantle for keeled glenoids was 2.50 mm (2.0 to 3.3) in group 1 and
5.18 mm (4.4 to 6.1) in group 2, and for pegged glenoids it was 1.72 mm
(0.9 to 2.3) in group 1 and 5.63 mm (3.6 to 6.4) in group 2. A cement
mantle <
2 mm was detected less frequently in group 2 (p <
0.001). Using the cement pressuriser the proportion of cement mantles
<
2 mm was significantly reduced compared with the third-generation
cementing technique.
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with
We have compared four computer-assisted methods to measure penetration of the femoral head into the acetabular component in total hip replacement. These were the Martell Hip Analysis suite 7.14, Rogan HyperOrtho, Rogan View Pro-X and Roman v1.70. The images used for the investigation comprised 24 anteroposterior digital radiographs and 24 conventional acetate radiographs which were scanned to provide digital images. These radiographs were acquired from 24 patients with an uncemented total hip replacement with a follow-up of approximately eight years (mean 8.1; 6.3 to 9.1). Each image was measured twice by two blinded observers. The mean annual rates of penetration of the femoral head measured in the eight-year single image analysis were: Martell, 0.24 (SD 0.19); HyperOrtho, 0.12 (SD 0.08); View Pro-X, 0.12 (SD 0.06); Roman, 0.12 (SD 0.07). In paired analysis of the six-month and eight-year radiographs: Martell, 0.35 (SD 0.22); HyperOrtho, 0.15 (SD 0.13); View Pro-X, 0.11 (SD 0.06); Roman, 0.11 (SD 0.07). The intra- and inter-observer variability for the paired analysis was best for View Pro-X and Roman software, with intraclass correlations of 0.97, 0.87 and 0.96, 0.87, respectively, and worst for HyperOrtho and Martell, with intraclass correlations of 0.46, 0.13 and 0.33, 0.39, respectively. The Roman method proved the most precise and the most easy to use in clinical practice and the software is available free of charge. The Martell method showed the lowest precision, indicating a problem with its edge detection algorithm on digital images.
Magnification of anteroposterior radiographs of the pelvis is variable. To improve the accuracy of templating, reliable and radiographer-friendly methods of scaling are necessary. We assessed two methods of scaling digital radiographs of the pelvis: placing a coin of known diameter in the plane of interest between the patient’s thighs, and using a caliper to measure the bony width of the pelvis. A total of 39 patients who had recently undergone hemiarthroplasty of the hip or total hip replacement were enrolled in the study. The accuracy of the methods was assessed by comparing the actual diameter of the head of the prosthesis with the measured on-screen value. The coin method was within a mean of 1.12% (0% to 2.38%) of the actual measurement, the caliper group within 6.99% (0% to 16.67%). The coin method was significantly more accurate (p <
0.001). It was also reliable and radiographer friendly. We recommend it as the method of choice for scaling radiographs of the pelvis before hip surgery.
Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications. We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre. At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip. The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.
The use of allograft struts and cerclage wire, possibly augmented by plate fixation, for the treatment of Vancouver type-B1 peri-prosthetic fractures around a total hip replacement has been strongly advocated. We examined our results using plate fixation without allograft struts and compared them with the results of the use of struts alone or when combined with plate fixation. Of 20 consecutive patients with type-B1 fractures treated by open reduction and plate fixation, 19 were available for follow-up. The fractures healed in 18 patients with a mean time to weight-bearing of ten weeks (4 to 19). There were no cases of infection or malunion. Nonunion occurred in one patient and required a second plate fixation to achieve union. Safe, cost-effective treatment of Vancouver type-B1 fractures can be performed by plate fixation without the addition of cortical struts. This procedure may allow earlier weight-bearing than allograft strut fixation alone.
Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse - femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be ± 1.6 mm for the x-direction and ± 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.
The complications of impaction bone grafting in revision hip replacement includes fracture of the femur and subsidence of the prosthesis. In this We found that the peak loads and hoop strains transmitted to the femoral cortex during graft compaction and subsidence of the stem in subsequent mechanical testing were reduced. This innovative technique has the potential to reduce the risk of intra-operative fracture and to improve graft compaction and therefore prosthetic stability.
The Morscher-Spotorno (MS-30) femoral stem is a stainless-steel, straight, three-dimensionally tapered, collarless implant for cemented fixation in total hip replacement. We report the results at ten years of a consecutive series of 124 total hip replacements in 121 patients with the matt-surfaced MS-30 stem and an alumina ceramic head of 28-mm diameter. All the stems were fixed with Palacos bone cement with gentamicin using a modern cementing technique. They were combined with an uncemented, press-fit cup. The mean period of observation was 10.2 years (8.3 to 12.1) and no patient was lost to follow-up. Twenty-seven patients (22%) died with the implant None of the stems or cups had been revised. The Harris hip score was excellent or good in 97% (85 hips) and moderate in 3% (three hips). Radiologically, six hips (6.8%) had osteolysis adjacent to the stem, mostly in Gruen zone 7. Twenty (22.7%) showed one or more radiolucent lines. Twenty-two stems (25%) had subsided by 2 mm to 5 mm. In these cases two showed osteolysis (9.1%) with subsidence and four without (6.1%). Radiolucent lines were seen in seven with migration (31.8%) and in 13 without (19.7%). No infections and no acetabular osteolysis were observed. The clinical results were excellent with survivorship after ten years of 100% and only a slightly statistically non-significant higher rate of osteolysis and radiolucency in cases of subsidence.
The effects of the method of fixation and interface conditions on the biomechanics of the femoral component of the Birmingham hip resurfacing arthroplasty were examined using a highly detailed three-dimensional computer model of the hip. Stresses and strains in the proximal femur were compared for the natural femur and for the femur resurfaced with the Birmingham hip resurfacing. A comparison of cemented
We evaluated the use of a stemmed acetabular component in the treatment of severe acetabular deficiency during revision and complex primary total hip arthroplasty. There were 31 hips of which 24 were revisions (20 for aseptic loosening, four for infection) and the remainder were complex primary arthroplasties. At a mean follow-up of 10.7 years (6 to 12.8), no component had been revised for aseptic loosening; one patient had undergone a revision of the polyethylene liner for wear. There was one failure because of infection. At the latest follow-up, the cumulative survival rate for aseptic loosening, with revision being the end-point, was 100%; for radiographic loosening it was 92% and for infection and radiographic loosening it was 88%. These results justify the continued use of this stemmed component for the reconstruction of severe acetabular deficiency.
Concerns have been reported to the United Kingdom
National Patient Safety Agency, warning that cementing the femoral component
during hip replacement surgery for fracture of the proximal femur
may increase peri-operative mortality. The National Hip Fracture Database collects demographic and outcome
data about patients with a fracture of the proximal femur from over
100 participating hospitals in the United Kingdom. We conducted
a mixed effects logistic regression analysis of this dataset to
determine whether peri-operative mortality was increased in patients
who had undergone either hemiarthroplasty or total hip replacement
using a cemented femoral component. A total of 16 496 patients from
129 hospitals were included in the analysis, which showed a small
but significant adjusted survival benefit associated with cementing
(odds ratio 0.83, 95% confidence interval 0.72 to 0.96). Other statistically
significant variables in predicting death at discharge, listed in
order of magnitude of effect, were gender, American Society of Anesthesiologists
grade, age, walking accompanied outdoors and arthroplasty. Interaction
terms between cementing and these other variables were sequentially
added to, but did not improve, the model. This study has not shown an increase in peri-operative mortality
as a result of cementing the femoral component in patients requiring
hip replacement following fracture of the proximal femur.
We developed a method of applying vibration to the impaction bone grafting process and assessed its effect on the mechanical properties of the impacted graft. Washed morsellised bovine femoral heads were impacted into shear test rings. A range of frequencies of vibration was tested, as measured using an accelerometer housed in a vibration chamber. Each shear test was repeated at four different normal loads to generate stress-strain curves. The Mohr-Coulomb failure envelope from which shear strength and interlocking values are derived was plotted for each test. The experiments were repeated with the addition of blood in order to replicate a saturated environment. Graft impacted with the addition of vibration at all frequencies showed improved shear strength when compared with impaction without vibration, with 60 Hz giving the largest effect. Under saturated conditions the addition of vibration was detrimental to the shear strength of the aggregate. The civil-engineering principles of particulate settlement and interlocking also apply to impaction bone grafting. Although previous studies have shown that vibration may be beneficial in impaction bone grafting on the femoral side, our study suggests that the same is not true in acetabular revision.
We carried out a prospective study of 118 hydroxyapatite-coated, cementless total knee replacements in patients who were ≤ 55 years of age and who had primary (92; 78%) or post-traumatic (26; 22%) osteoarthritis. The mean period of follow-up was 7.9 years (5 to 12.5). The Knee Society clinical scores improved from a pre-operative mean of 98 (0 to 137) to a mean of 185 (135 to 200) at five years, and 173 (137 to 200) at ten years. There were two revisions of the tibial component because of aseptic loosening, and one case of polyethylene wear requiring further surgery. There was no osteolysis or progressive radiological loosening of any other component. At 12 years, the overall rate of implant survival was 97.5% (excluding exchange of spacer) and 92.1% (including exchange of spacer). Cementless total knee replacement can achieve excellent long-term results in young, active patients with osteoarthritis. In contrast to total hip replacement, polyethylene wear, osteolysis and loosening of the prosthesis were not major problems for these patients, although it is possible that this observation could change with longer periods of follow-up.
The efficacy of circumpatellar electrocautery in reducing the incidence of post-operative anterior knee pain is unknown. We conducted a single-centre, outcome-assessor and patient-blinded, parallel-group, randomised, controlled trial to compare circumpatellar electrocautery with no electrocautery in total knee replacement in the absence of patellar resurfacing. Patients requiring knee replacement for primary osteoarthritis were randomly assigned circumpatellar electrocautery (intervention group) or no electrocautery (control group). The primary outcome measure was the incidence of anterior knee pain. A secondary measure was the standardised clinical and patient-reported outcomes determined by the American Knee Society scores and the Western Ontario and McMaster Universities (WOMAC) osteoarthritis index. A total of 131 knees received circumpatellar electrocautery and 131 had no electrocautery. The overall incidence of anterior knee pain at follow-up at one year was 26% (20% to 31%), with 19% (12% to 26%) in the intervention group and 32% (24% to 40%) in the control group (p = 0.02). The relative risk reduction from electrocautery was 40% (9% to 61%) and the number needed to treat was 7.7 (4.3 to 41.4). The intervention group had a better mean total WOMAC score at follow-up at one year compared with the control group (16.3 (0 to 77.7) Our study suggests that in the absence of patellar resurfacing electrocautery around the margin of the patella improves the outcome of total knee replacement.
We conducted a systematic review and meta-analysis
of randomised controlled trials evaluating the effect of tranexamic
acid (TXA) upon blood loss and transfusion in primary total knee
replacement. The review used the generic evaluation tool designed
by the Cochrane Bone, Joint and Muscle Trauma Group. A total of
19 trials were eligible: 18 used intravenous administration, one
also evaluated oral dosing and one trial evaluated topical use.
TXA led to a significant reduction in the proportion of patients
requiring blood transfusion (risk ratio (RR) 2.56, 95% confidence
interval (CI) 2.1 to 3.1, p <
0.001; heterogeneity I2 =
75%; 14 trials, 824 patients). Using TXA also reduced total blood
loss by a mean of 591 ml (95% CI 536 to 647, p <
0.001; I2 =
78%; nine trials, 763 patients). The clinical interpretation of
these findings is limited by substantial heterogeneity. However,
subgroup analysis of high-dose (>
4 g) TXA showed a plausible consistent
reduction in blood transfusion requirements (RR 5.33; 95% CI 2.44
to 11.65, p <
0.001; I2 = 0%), a finding that should
be confirmed by a further well-designed trial. The current evidence
from trials does not support an increased risk of deep-vein thrombosis
(13 trials, 801 patients) or pulmonary embolism (18 trials, 971
patients) due to TXA administration.