The purpose of this study was to evaluate whether concerns about the release of metal ions in metal-on-metal total hip replacements (THR) should be extended to patients with metal-bearing total disc replacements (TDR). Cobalt and chromium levels in whole blood and serum were measured in ten patients with a single-level TDR after a mean follow-up of 34.5 months (13 to 61) using inductively-coupled plasma mass spectrometry. These metal ion levels were compared with pre-operative control levels in 81 patients and with metal ion levels 12 months after metal-on-metal THR (n = 21) and resurfacing hip replacement (n = 36). Flexion-extension radiographs were used to verify movement of the TDR. Cobalt levels in whole blood and serum were significantly lower in the TDR group than in either the THR (p = 0.007) or the resurfacing group (p <
0.001). Both chromium levels were also significantly lower after TDR These results suggest that there is minimal cause for concern about high metal ion concentrations after TDR, as the levels appear to be only moderately elevated. However, spinal surgeons using a metal-on-metal TDR should still be aware of concerns expressed in the hip replacement literature about toxicity from elevated metal ion levels, and inform their patients appropriately.
We retrospectively analysed concentrations of chromium and cobalt ions in samples of synovial fluid and whole blood taken from a group of 92 patients with failed current-generation metal-on-metal hip replacements. We applied acid oxidative digestion to our trace metal analysis protocol, which found significantly higher levels of metal ion concentrations in blood and synovial fluid than a non-digestive method. Patients were subcategorised by mode of failure as either ‘unexplained pain’ or ‘defined causes’. Using this classification, chromium and cobalt ion levels were present over a wider range in synovial fluid and not as strongly correlated with blood ion levels as previously reported. There was no significant difference between metal ion concentrations and manufacturer of the implant, nor femoral head size below or above 50 mm. There was a moderately positive correlation between metal ion levels and acetabular component inclination angle as measured on three-dimensional CT imaging. Our results suggest that acid digestion of samples of synovial fluid samples is necessary to determine metal ion concentrations accurately so that meaningful comparisons can be made between studies.
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
Patient warming significantly decreases the risk
of surgical site infection. Recently there have been concerns that forced
air warming may interfere with unidirectional airflow, potentially
posing an increased risk of infection. Our null hypothesis was that
forced air and radiant warming devices do not increase the temperature
and the number of particles over the surgical site when compared
with no warming device. A forced air warming device was compared with
a radiant warming device and no warming device as a control. The
temperature and number of particles were measured over the surgical
site. The theatre was prepared as for a routine lower-limb arthroplasty
operation, and the same volunteer was used throughout the study. Forced air warming resulted in a significant mean increase in
the temperature (1.1°C
The accuracy and precision of two new methods of model-based
radiostereometric analysis (RSA) were hypothesised to be superior
to a plain radiograph method in the assessment of polyethylene (PE)
wear. A phantom device was constructed to simulate three-dimensional
(3D) PE wear. Images were obtained consecutively for each simulated
wear position for each modality. Three commercially available packages
were evaluated: model-based RSA using laser-scanned cup models (MB-RSA),
model-based RSA using computer-generated elementary geometrical
shape models (EGS-RSA), and PolyWare. Precision (95% repeatability
limits) and accuracy (Root Mean Square Errors) for two-dimensional
(2D) and 3D wear measurements were assessed.Objectives
Methods
Pseudotumours (abnormal peri-prosthetic soft-tissue reactions)
following metal-on-metal hip resurfacing arthroplasty (MoMHRA) have
been associated with elevated metal ion levels, suggesting that
excessive wear may occur due to edge-loading of these MoM implants.
This study aimed to quantify The duration and magnitude of edge-loading Objectives
Methods
We determined the midterm survival, incidence
of peri-prosthetic fracture and the enhancement of the width of
the femur when combining struts and impacted bone allografts in
24 patients (25 hips) with severe femoral bone loss who underwent
revision hip surgery. The pre-operative diagnosis was aseptic loosening
in 16 hips, second-stage reconstruction in seven, peri-prosthetic
fracture in one and stem fracture in one
A variety of radiological methods of measuring
version of the acetabular component after total hip replacement (THR)
have been described. The aim of this study was to evaluate the reliability
and validity of six methods (those of Lewinnek; Widmer; Hassan et
al; Ackland, Bourne and Uhthoff; Liaw et al; and Woo and Morrey)
that are currently in use. In 36 consecutive patients who underwent
THR, version of the acetabular component was measured by three independent
examiners on plain radiographs using these six methods and compared
with measurements using CT scans. The intra- and interobserver reliabilities
of each measurement were estimated. All measurements on both radiographs
and CT scans had excellent intra- and interobserver reliability
and the results from each of the six methods correlated well with
the CT measurements. However, measurements made using the methods
of Widmer and of Ackland, Bourne and Uhthoff were significantly
different from the CT measurements (both p <
0.001), whereas
measurements made using the remaining four methods were similar
to the CT measurements. With regard to reliability and convergent
validity, we recommend the use of the methods described by Lewinnek,
Hassan et al, Liaw et al and Woo and Morrey for measurement of version
of the acetabular component.
An ongoing prospective study to investigate failing metal-on-metal
hip prostheses was commenced at our centre in 2008. We report on
the results of the analysis of the first consecutive 126 failed
mated total hip prostheses from a single manufacturer. Analysis was carried out using highly accurate coordinate measuring
to calculate volumetric and linear rates of the articular bearing
surfaces and also the surfaces of the taper junctions. The relationship
between taper wear rates and a number of variables, including bearing
diameter and orientation of the acetabular component, was investigated.Objectives
Methods
We report the use of porous metal acetabular
revision shells in the treatment of contained bone loss. The outcomes of
53 patients with
The purpose of this study was to assess N-acetyl aspartate changes
in the thalamus in patients with osteoarthritis of the hip using
proton magnetic resonance spectroscopy. Nine patients with osteoarthritis of the hip (symptomatic group,
nine women; mean age 61.4 years (48 to 78)) and nine healthy volunteers
(control group, six men, three women; mean age 30.0 years (26 to
38)) underwent proton magnetic resonance spectroscopy to assess
the changes of N-acetyl aspartate in the thalamus. Objectives
Methods
Peripheral nerve injury is an uncommon but serious
complication of hip surgery that can adversely affect the outcome.
Several studies have described the use of electromyography and intra-operative
sensory evoked potentials for early warning of nerve injury. We
assessed the results of multimodal intra-operative monitoring during
complex hip surgery. We retrospectively analysed data collected
between 2001 and 2010 from 69 patients who underwent complex hip
surgery by a single surgeon using multimodal intra-operative monitoring
from a total pool of 7894 patients who underwent hip surgery during
this period. In 24 (35%) procedures the surgeon was alerted to a
possible lesion to the sciatic and/or femoral nerve. Alerts were
observed most frequently during peri-acetabular osteotomy. The surgeon
adapted his approach based on interpretation of the neurophysiological changes.
From 69 monitored surgical procedures, there was only one true positive
case of post-operative nerve injury. There were no false positives
or false negatives, and the remaining 68 cases were all true negative.
The sensitivity for predicting post-operative nerve injury was 100%
and the specificity 100%. We conclude that it is possible and appropriate
to use this method during complex hip surgery and it is effective
for alerting the surgeon to the possibility of nerve injury.
We report the long-term results of revision total
hip replacement using femoral impaction allografting with both uncemented
and cemented Freeman femoral components. A standard design of component
was used in both groups, with additional proximal hydroxyapatite
coating in the uncemented group. A total of 33 hips in 30 patients received
an uncemented component and 31 hips in 30 patients a cemented component.
The mean follow-up was 9.8 years (2 to 17) in the uncemented group
and 6.2 years (1 to 11) in the cemented group. Revision procedures
(for all causes) were required in four patients (four hips) in the
uncemented group and in five patients (five hips) in the cemented
group. Harris hip scores improved significantly in both groups and
were maintained independently of the extent of any migration of
the femoral component within the graft or graft–cement mantle.
We obtained pre-operative and six-month post-operative
Oxford hip (OHS) and knee scores (OKS) for 1523 patients who underwent
total hip replacement and 1784 patients who underwent total knee
replacement. They all also completed a six-month satisfaction question. Scatter plots showed no relationship between pre-operative Oxford
scores and six-month satisfaction scores. Spearman’s rank correlation
coefficients were -0.04 (95% confidence interval (CI) -0.09 to 0.01)
between OHS and satisfaction and 0.04 (95% CI -0.01 to 0.08) between
OKS and satisfaction. A receiver operating characteristic (ROC) curve
analysis was used to identify a cut-off point for the pre-operative
OHS/OKS that identifies whether or not a patient is satisfied with
surgery. We obtained an area under the ROC curve of 0.51 (95% CI
0.45 to 0.56) for hip replacement and 0.56 (95% CI 0.51 to 0.60)
for knee replacement, indicating that pre-operative Oxford scores
have no predictive accuracy in distinguishing satisfied from dissatisfied
patients. In the NHS widespread attempts are being made to use patient-reported
outcome measures (PROMs) data for the purpose of prioritising patients
for surgery. Oxford hip and knee scores have no predictive accuracy
in relation to post-operative patient satisfaction. This evidence
does not support their current use in prioritising access to care.
A 70-year-old man with an uncemented metal-on-polyethylene
total hip prosthesis underwent revision arthroplasty 33 months later
because of pain, swelling and recurrent dislocation. There appeared
to be corrosion and metal release from the prosthetic head, resulting
in pseudotumour formation and severe local soft-tissue destruction.
The corrosion occurred at the junction between the titanium-molybdenum-zirconium-iron
taper and the cobalt-chrome-molybdenum head, but the mechanism was unproven.
We evaluated the incidence of heterotopic ossification
following total ankle replacement to determine whether the degree
of ossification was associated with the clinical outcome. We evaluated
90 ankles in 81 consecutive patients who underwent total ankle replacement,
and heterotopic ossification was assessed according to proportional involvement
of the ankle joint. Correlation analysis was used to investigate
the association between heterotopic ossification and outcome. No significant association was found between the formation of
heterotopic ossification and the clinical outcome. The degree of
heterotopic ossification in the posterior ankle joint was not significantly
correlated with posterior ankle pain (p = 0.929), the American Orthopaedic
Foot and Ankle Society score (p = 0.454) or range of movement (p
= 0.283). This study indicates that caution should be observed in attributing
symptoms and functional limitation to the presence of heterotopic
ossification in the posterior ankle joint when considering excision
of heterotopic bone after total ankle replacement.
The aim of this study was to investigate the
possible benefit of large-head metal-on-metal bearing on a stem
for primary hip replacement compared with a 28 mm diameter conventional
metal-on-polyethylene bearing in a prospective randomised controlled
trial. We investigated cemented stem behaviour between these two
different bearings using Einzel-Bild-Röntgen-Analyse, clinical and
patient reported measures (Harris hip score, Western Ontario and
McMaster Universities osteoarthritis index, Short Form-36 and satisfaction)
and whole blood metal ion levels at two years. A power study indicated
that 50 hips were needed in each group to detect subsidence of >
5 mm at two years with a
p-value of <
0.05. Significant improvement (p <
0.001) was found in the mean
clinical and patient reported outcomes at two years for both groups.
Comparison of outcomes between the groups at two years showed no
statistically significant difference for mean stem migration, clinical
and patient reported outcomes; except overall patient satisfaction which
was higher for metal-on-metal group (p = 0.05). Metal ion levels
were raised above the Medicines and Healthcare products Regulatory
Agency advised safety level (7 µg per litre) in 20% of the metal-on-metal
group and in one patient in metal-on-polyethylene group (who had
a metal-on-metal implant on the contralateral side). Two patients
in the metal-on-metal group were revised, one for pseudotumour and
one for peri-prosthetic fracture. Use of large modular heads is associated with a risk of raised
whole blood metal ion levels despite using a proven bearing from
resurfacing. The head-neck junction or excess stem micromotion are
possibly the weak links warranting further research.
Antibiotic prophylaxis is routinely administered during joint replacement surgery and may predispose patients to
Free vascularised fibular grafting has been reported
to be successful for adult patients with osteonecrosis of the femoral
head (ONFH). However, its benefit in teenage patients with post-traumatic
ONFH has not been determined. We evaluated the effectiveness of
free vascularised fibular grafting in the treatment of this condition
in children and adolescents. We retrospectively analysed 28 hips
in 28 patients in whom an osteonecrotic femoral head had been treated
with free vascularised fibular grafting between 2002 and 2008. Their
mean age was 16.3 years (13 to 19). The stage of the disease at
time of surgery, and results of treatment including pre- and post-operative
Harris hip scores, were studied. We defined clinical failure as
conversion to total hip replacement. All patients were followed
up for a mean of four years (2 to 7). The mean Harris hip score
improved from 60.4 (37 to 84) pre-operatively to 94.2 (87 to 100)
at final follow-up. At the latest follow-up we found improved or
unchanged radiographs in all four initially stage II hips and in
23 of 24 stage III or IV hips. Only one hip (stage V) deteriorated.
No patient underwent total hip replacement. Free vascularised fibular grafting is indicated for the treatment
of post-traumatic ONFH in teenage patients.