Advertisement for orthosearch.org.uk
Results 61 - 80 of 437
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 5 | Pages 689 - 691
1 Sep 1996
Søballe K Overgaard S


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 2 | Pages 334 - 335
1 Mar 1993
Nolan P Wilson D Mollan R


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 6 | Pages 760 - 768
1 Jun 2011
ten Broeke RHM Alves A Baumann A Arts JJC Geesink RGT

Four uncemented Symax hip stems were extracted at three weeks and nine, 13 and 32 months, respectively, for reasons other than loosening. The reasons for implant removal were infection in two cases, recurrent dislocation in one and acetabular fracture in one. They were analysed to assess the effect and behaviour of an electrochemically deposited, completely resorbable biomimetic BONIT-hydroxyapatite (HA) coating (proximal part) and a DOTIZE surface treatment (distal part) using qualitative histology, quantitative histomorphometry and scanning electron microscopy (SEM). Early and direct bone-implant bonding with signs of active remodelling of bone and the HA coating were demonstrated by histology and SEM. No loose BONIT-HA particles or delamination of the coating were observed, and there was no inflammation or fibrous interposition at the interface.

Histomorphometry showed bone-implant contact varying between 26.5% at three weeks and 83.5% at 13 months at the HA-coated implant surface. The bone density in the area of investigation was between 24.6% at three weeks and 41.1% at 32 months. The DOTIZE surface treatment of the distal part of the stem completely prevented tissue and bone apposition in all cases, thereby optimising proximal stress transfer.

The overall features of this implant, in terms of geometry and surface texture, suggest a mechanically stable design with a highly active biomimetic coating, resulting in rapid and extensive osseo-integration, exclusively in the metaphyseal part of the stem. Early remodelling of the HA coating does not seem to have a detrimental effect on short-term bone-implant coupling. There were no adverse effects identified from either the BONIT-HA coating or the DOTIZE surface treatment.


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 8 | Pages 1118 - 1123
1 Nov 2004
Singh S Trikha SP Edge AJ

We describe the clinical and radiological results of 38 total hip replacements (THR) using the JRI Furlong hydroxyapatite-ceramic (HAC)-coated femoral component in patients younger than 50 years. The mean age at the time of operation was 42 years (22 to 49) and the mean length of follow-up was ten years (63 to 170 months). All patients receiving a Furlong HAC THR were entered into the study regardless of the primary pathology including patients who had undergone previous hip surgery.

The mean Harris hip score improved from 44 before operation to 92 at the last postoperative review. After 12 years the cumulative surivival for the stem was 100% (95% confidence interval 89 to 100). No femoral component was revised. Our results show that the Furlong HAC implant gives excellent long-term results in young patients with high demands.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 1044 - 1044
1 Nov 1997
Fulford P


The Bone & Joint Journal
Vol. 98-B, Issue 9 | Pages 1289 - 1296
1 Sep 2016
McNally MA Ferguson JY Lau ACK Diefenbeck M Scarborough M Ramsden AJ Atkins BL

Aims

Chronic osteomyelitis may recur if dead space management, after excision of infected bone, is inadequate. This study describes the results of a strategy for the management of deep bone infection and evaluates a new antibiotic-loaded biocomposite in the eradication of infection from bone defects.

Patients and Methods

We report a prospective study of 100 patients with chronic osteomyelitis, in 105 bones. Osteomyelitis followed injury or surgery in 81 patients. Nine had concomitant septic arthritis. 80 patients had comorbidities (Cierny-Mader (C-M) Class B hosts). Ten had infected nonunions.

All patients were treated by a multidisciplinary team with a single-stage protocol including debridement, multiple sampling, culture-specific systemic antibiotics, stabilisation, dead space filling with the biocomposite and primary skin closure.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 569 - 573
1 May 2014
Sullivan MP McHale KJ Parvizi J Mehta S

Nanotechnology is the study, production and controlled manipulation of materials with a grain size < 100 nm. At this level, the laws of classical mechanics fall away and those of quantum mechanics take over, resulting in unique behaviour of matter in terms of melting point, conductivity and reactivity. Additionally, and likely more significant, as grain size decreases, the ratio of surface area to volume drastically increases, allowing for greater interaction between implants and the surrounding cellular environment. This favourable increase in surface area plays an important role in mesenchymal cell differentiation and ultimately bone–implant interactions.

Basic science and translational research have revealed important potential applications for nanotechnology in orthopaedic surgery, particularly with regard to improving the interaction between implants and host bone. Nanophase materials more closely match the architecture of native trabecular bone, thereby greatly improving the osseo-integration of orthopaedic implants. Nanophase-coated prostheses can also reduce bacterial adhesion more than conventionally surfaced prostheses. Nanophase selenium has shown great promise when used for tumour reconstructions, as has nanophase silver in the management of traumatic wounds. Nanophase silver may significantly improve healing of peripheral nerve injuries, and nanophase gold has powerful anti-inflammatory effects on tendon inflammation.

Considerable advances must be made in our understanding of the potential health risks of production, implantation and wear patterns of nanophase devices before they are approved for clinical use. Their potential, however, is considerable, and is likely to benefit us all in the future.

Cite this article: Bone Joint J 2014; 96-B: 569–73.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1610 - 1616
1 Dec 2013
Epinette J Asencio G Essig J Llagonne B Nourissat C

We report a multicentre prospective consecutive study assessing the long-term outcome of the proximally hydroxyapatite (HA)-coated ABG II monobloc femoral component in a series of 1148 hips in 1053 patients with a mean age at surgery of 64.77 years (22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At latest follow-up, the mean total Harris hip score was 94.7 points (sd; 6.87; 49 to 100), and the mean Merle d’Aubigné–Postel score was 17.6 points (sd 1.12; 7 to 18). The mean total Engh radiological score score was 21.54 (sd 5.77; 3.5 to 27), with 95.81% of ‘confirmed ingrowth’, according to Engh’s classification. With aseptic loosening or pain as endpoints, three AGB II stems (0.26%) failed, giving a 99.7% survival rate (se 0.002; 95% confidence interval (CI) 0.994 to 1) at 14 years’ follow-up. The survival of patients ≤ 50 years of age (99.0% (se 0.011; 95% CI 0.969 to 1)) did not differ significantly from those of patients aged > 50 years (99.8% (se 0.002; 95% CI 0.994 to 1)). This study confirmed the excellent long-term results currently achieved with the ABG II proximally HA-coated monobloc stem.

Cite this article: Bone Joint J 2013;95-B:1610–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 2 | Pages 368 - 369
1 Mar 1999
NELISSEN RGHH VALSTAR ER ROZING PM


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 467 - 471
1 Apr 2013
Sandiford N Doctor C Rajaratnam SS Ahmed S East DJ Miles K Butler-Manuel A Shepperd JAN

We present the extended follow-up (≥ 20 years) of a series of fully hydroxyapatite-coated femoral components used in 72 primary total hip replacements (THRs). Earlier results of this cohort have been previously published. All procedures were performed between 1986 and 1991. The series involved 45 women and 15 men with 12 bilateral procedures. Their mean age at the time of surgery was 60 years (46 to 80) and the mean duration of follow-up was 22.5 years (20 to 25). At final follow-up, the mean Merle d’Aubigné and Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0 to 5.0) for pain, mobility and function, respectively. Of the patients 92% were very satisfied at the time of final follow-up.

There were seven revisions: six of the acetabular component for aseptic loosening and one of both the stem and the acetabular component for loosening due to deep infection. The survival of this prosthesis at 22.5 years with revision for any reason as the endpoint was 91.7% (95% confidence interval (CI) 84 to 99). Survival with aseptic loosening of the stem as the endpoint was 100% (95% CI 90 to 100).

This prosthesis provides pain relief in the long term. Survival of this component is comparable to the best results for primary THR with any means of fixation.

Cite this article: Bone Joint J 2013;95-B:467–71.


The Bone & Joint Journal
Vol. 95-B, Issue 2 | Pages 181 - 187
1 Feb 2013
Liddle AD Pandit H O’Brien S Doran E Penny ID Hooper GJ Burn PJ Dodd CAF Beverland DE Maxwell AR Murray DW

The Cementless Oxford Unicompartmental Knee Replacement (OUKR) was developed to address problems related to cementation, and has been demonstrated in a randomised study to have similar clinical outcomes with fewer radiolucencies than observed with the cemented device. However, before its widespread use it is necessary to clarify contraindications and assess the complications. This requires a larger study than any previously published.

We present a prospective multicentre series of 1000 cementless OUKRs in 881 patients at a minimum follow-up of one year. All patients had radiological assessment aligned to the bone–implant interfaces and clinical scores. Analysis was performed at a mean of 38.2 months (19 to 88) following surgery. A total of 17 patients died (comprising 19 knees (1.9%)), none as a result of surgery; there were no tibial or femoral loosenings. A total of 19 knees (1.9%) had significant implant-related complications or required revision. Implant survival at six years was 97.2%, and there was a partial radiolucency at the bone–implant interface in 72 knees (8.9%), with no complete radiolucencies. There was no significant increase in complication rate compared with cemented fixation (p = 0.87), and no specific contraindications to cementless fixation were identified.

Cementless OUKR appears to be safe and reproducible in patients with end-stage anteromedial osteoarthritis of the knee, with radiological evidence of improved fixation compared with previous reports using cemented fixation.

Cite this article: Bone Joint J 2013;95-B:181–7.


The Bone & Joint Journal
Vol. 96-B, Issue 3 | Pages 345 - 349
1 Mar 2014
Liddle AD Pandit HG Jenkins C Lobenhoffer P Jackson WFM Dodd CAF Murray DW

The cementless Oxford unicompartmental knee replacement has been demonstrated to have superior fixation on radiographs and a similar early complication rate compared with the cemented version. However, a small number of cases have come to our attention where, after an apparently successful procedure, the tibial component subsides into a valgus position with an increased posterior slope, before becoming well-fixed. We present the clinical and radiological findings of these six patients and describe their natural history and the likely causes. Two underwent revision in the early post-operative period, and in four the implant stabilised and became well-fixed radiologically with a good functional outcome.

This situation appears to be avoidable by minor modifications to the operative technique, and it appears that it can be treated conservatively in most patients.

Cite this article: Bone Joint J 2014;96-B:345–9.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 8 | Pages 1067 - 1070
1 Aug 2012
Melton JTK Mayahi R Baxter SE Facek M Glezos C

A consecutive series of patients with a hydroxyapatite-coated uncemented total knee replacement (TKR) performed by a single surgeon between 1992 and 1995 was analysed. All patients were invited for clinical review and radiological assessment. Revision for aseptic loosening was the primary outcome. Assessment was based on the Knee Society clinical score (KSS) and an independent radiological analysis. Of 471 TKRs performed in 356 patients, 432 TKRs in 325 patients were followed for a mean of 16.4 years (15 to 18). The 39 TKRs in 31 patients lost to follow-up had a mean KSS of 176 (148 to 198) at a mean of ten years. There were revisions in 26 TKRs (5.5%), of which 11 (2.3%) were for aseptic loosening. Other further surgery was carried out on 49 TKRs (10.4%) including patellar resurfacing in 20, arthrolysis in 19, manipulation under anaesthetic in nine and extensor mechanism reconstruction in one.

Survivorship at up to 18 years without aseptic loosening was 96% (95% confidence interval 91.9 to 98.1), at which point the mean KSS was 176 (134 to 200). Of 110 knees that underwent radiological evaluation, osteolysis was observed in five (4.5%), one of which was revised.

These data indicate that uncemented hydroxyapatite-coated TKR can achieve favourable long-term survivorship, at least as good as that of cemented designs.


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 745 - 745
1 Jul 1998
BAUER TW


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 4 | Pages 745 - 746
1 Jul 1998
FURLONG R


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 2 | Pages 267 - 272
1 Mar 1998
Morscher EW Hefti A Aebi U

One concern about the fixation of HA-coated implants is the possible disintegration of the surface, with the migration of HA granules into the joint space, producing third-body wear.

We report a study of six revisions of HA-coated polyethylene RM cups at 9 to 14 years after successful primary arthroplasty. In all six hips, we found HA granules embedded in the articulating surface of the polyethylene, with abrasive wear of the cup and the metal femoral head. The cup had loosened in four hips and three showed severe osteolysis of the proximal femur.

Third-body wear due to HA particles from implant coating may produce severe clinical problems with few early warning signs. Further clinical, radiological and histological observations are needed to determine the possible incidence of this late complication in the various types of coating of a variety of substrates.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years.

The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial.

Cite this article: Bone Joint J 2014;96-B:263–9.


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 136 - 143
1 Feb 2024
van der Lelij TJN Marang-van de Mheen PJ Kaptein BL Koster LA Ljung P Nelissen RGHH Toksvig-Larsen S

Aims. The objective of this study was to compare the two-year migration and clinical outcomes of a new cementless hydroxyapatite (HA)-coated titanium acetabular shell with its previous version, which shared the same geometrical design but a different manufacturing process for applying the titanium surface. Methods. Overall, 87 patients undergoing total hip arthroplasty (THA) were randomized to either a Trident II HA or Trident HA shell, each cementless with clusterholes and HA-coating. All components were used in combination with a cemented Exeter V40 femoral stem. Implant migration was measured using radiostereometric analysis (RSA), with radiographs taken within two days of surgery (baseline), and at three, 12, and 24 months postoperatively. Proximal acetabular component migration was the primary outcome measure. Clinical scores and patient-reported outcome measures (PROMs) were collected at each follow-up. Results. Mean proximal migrations at three, 12, and 24 months were 0.08 mm (95% confidence interval (CI) 0.03 to 0.14), 0.11 mm (95% CI 0.06 to 0.16), and 0.14 mm (95% CI 0.09 to 0.20), respectively, in the Trident II HA group, versus 0.11 mm (95% CI 0.06 to 0.16), 0.12 mm (95% CI 0.07 to 0.17), and 0.14 mm (95% CI 0.09 to 0.19) in the Trident HA group (p = 0.875). No significant differences in translations or rotations between the two designs were found in any other direction. Clinical scores and PROMs were comparable between groups, except for an initially greater postoperative improvement in Hip disability and Osteoarthritis Outcome Symptoms score in the Trident HA group (p = 0.033). Conclusion. The Trident II clusterhole HA shell has comparable migration with its predecessor, the Trident hemispherical HA cluster shell, suggesting a similar risk of long-term aseptic loosening. Cite this article: Bone Joint J 2024;106-B(2):136–143


Bone & Joint Research
Vol. 12, Issue 7 | Pages 412 - 422
4 Jul 2023
Ferguson J Bourget-Murray J Hotchen AJ Stubbs D McNally M

Aims. Dead-space management, following dead bone resection, is an important element of successful chronic osteomyelitis treatment. This study compared two different biodegradable antibiotic carriers used for dead-space management, and reviewed clinical and radiological outcomes. All cases underwent single-stage surgery and had a minimum one-year follow-up. Methods. A total of 179 patients received preformed calcium sulphate pellets containing 4% tobramycin (Group OT), and 180 patients had an injectable calcium sulphate/nanocrystalline hydroxyapatite ceramic containing gentamicin (Group CG). Outcome measures were infection recurrence, wound leakage, and subsequent fracture involving the treated segment. Bone-void filling was assessed radiologically at a minimum of six months post-surgery. Results. The median follow-up was 4.6 years (interquartile range (IQR) 3.2 to 5.4; range 1.3 to 10.5) in Group OT compared to 4.9 years (IQR 2.1 to 6.0; range 1.0 to 8.3) in Group CG. The groups had similar defect sizes following excision (both mean 10.9 cm. 3. (1 to 30)). Infection recurrence was higher in Group OT (20/179 (11.2%) vs 8/180 (4.4%), p = 0.019) than Group CG, as was early wound leakage (33/179 (18.4%) vs 18/180 (10.0%), p = 0.024) and subsequent fracture (11/179 (6.1%) vs 1.7% (3/180), p = 0.032). Group OT cases had an odds ratio 2.9-times higher of developing any one of these complications, compared to Group CG (95% confidence interval 1.74 to 4.81, p < 0.001). The mean bone-void healing in Group CG was better than in Group OT, in those with ≥ six-month radiological follow-up (73.9% vs 40.0%, p < 0.001). Conclusion. Local antibiotic carrier choice affects outcome in chronic osteomyelitis surgery. A biphasic injectable carrier with a slower dissolution time was associated with better radiological and clinical outcomes compared to a preformed calcium sulphate pellet carrier. Cite this article: Bone Joint Res 2023;12(7):412–422


Bone & Joint Open
Vol. 3, Issue 12 | Pages 991 - 997
23 Dec 2022
McPherson EJ Stavrakis AI Chowdhry M Curtin NL Dipane MV Crawford BM

Aims. Large acetabular bone defects encountered in revision total hip arthroplasty (THA) are challenging to restore. Metal constructs for structural support are combined with bone graft materials for restoration. Autograft is restricted due to limited volume, and allogenic grafts have downsides including cost, availability, and operative processing. Bone graft substitutes (BGS) are an attractive alternative if they can demonstrate positive remodelling. One potential product is a biphasic injectable mixture (Cerament) that combines a fast-resorbing material (calcium sulphate) with the highly osteoconductive material hydroxyapatite. This study reviews the application of this biomaterial in large acetabular defects. Methods. We performed a retrospective review at a single institution of patients undergoing revision THA by a single surgeon. We identified 49 consecutive patients with large acetabular defects where the biphasic BGS was applied, with no other products added to the BGS. After placement of metallic acetabular implants, the BGS was injected into the remaining bone defects surrounding the new implants. Patients were followed and monitored for functional outcome scores, implant fixation, radiological graft site remodelling, and revision failures. Results. Mean follow-up was 39.5 months (36 to 71), with a significant improvement in post-revision function compared to preoperative function. Graft site remodelling was rated radiologically as moderate in 31 hips (63%) and strong in 12 hips (24%). There were no cases of complete graft site dissolution. No acetabular loosening was identified. None of the patients developed clinically significant heterotopic ossification. There were twelve reoperations: six patients developed post-revision infections, three experienced dislocations, two sustained periprosthetic femur fractures, and one subject had femoral component aseptic loosening. Conclusion. Our series reports bone defect restoration with the sole use of a biphasic injectable BGS in the periacetabular region. We did not observe significant graft dissolution. We emphasize that successful graft site remodelling requires meticulous recipient site preparation. Cite this article: Bone Jt Open 2022;3(12):991–997