We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain
It is increasingly appreciated that coordinated regulation of angiogenesis and osteogenesis is needed for bone formation. How this regulation is achieved during peri-implant bone healing, such as osseointegration, is largely unclear. This study examined the relationship between angiogenesis and osteogenesis in a unique model of osseointegration of a mouse tibial implant by pharmacologically blocking the vascular endothelial growth factor (VEGF) pathway. An implant was inserted into the right tibia of 16-week-old female C57BL/6 mice (n = 38). Mice received anti-VEGF receptor-1 (VEGFR-1) antibody (25 mg/kg) and VEGF receptor-2 (VEGFR-2) antibody (25 mg/kg; n = 19) or an isotype control antibody (n = 19). Flow cytometric (n = 4/group) and immunofluorescent (n = 3/group) analyses were performed at two weeks post-implantation to detect the distribution and density of CD31hiEMCNhi endothelium. RNA sequencing analysis was performed using sorted CD31hiEMCNhi endothelial cells (n = 2/group). Osteoblast lineage cells expressing osterix (OSX) and osteopontin (OPN) were also detected with immunofluorescence. Mechanical pull-out testing (n = 12/group) was used at four weeks post-implantation to determine the strength of the bone-implant interface. After pull-out testing, the tissue attached to the implant surface was harvested. Whole mount immunofluorescent staining of OSX and OPN was performed to determine the amount of osteoblast lineage cells.Aims
Materials and Methods
Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (Aims
Patients and Methods
Short-stemmed femoral implants have been used for total hip arthroplasty (THA) in young and active patients to conserve bone, provide physiological loading, and reduce the incidence of thigh pain. Only short- to mid-term results have been presented and there have been concerns regarding component malalignment, incorrect sizing, and subsidence. This systematic review reports clinical and radiological outcomes, complications, revision rates, and implant survival in THA using short-stemmed femoral components. A literature review was performed using the EMBASE, Medline, and Cochrane databases. Strict inclusion and exclusion criteria were used to identify studies reporting clinical and radiological follow-up for short-stemmed hip arthroplasties.Aims
Materials and Methods
Loosening is a well-known complication in the fixation of fractures using devices such as locking plates or unilateral fixators. It is believed that high strains in the bone at the bone-screw interface can initiate loosening, which can result in infection, and further loosening. Here, we present a new theory of loosening of implants. The time-dependent response of bone subjected to loads results in interfacial deformations in the bone which accumulate with cyclical loading and thus accentuates loosening. We used an ‘ideal’ bone-screw system, in which the screw is subjected to cyclical lateral loads and trabecular bone is modelled as non-linear viscoelastic and non-linear viscoelastic-viscoplastic material, based on recent experiments, which we conducted.Aims
Methods
The advent of trabecular metal (TM) augments has revolutionized
the management of severe bone defects during acetabular reconstruction.
The purpose of this study was to evaluate patients undergoing revision
total hip arthroplasty (THA) with the use of TM augments for reconstruction
of Paprosky IIIA and IIIB defects. A retrospective study was conducted at four centres between August
2008 and January 2015. Patients treated with TM augments and TM
shell for a Paprosky grade IIIA or IIIB defect, in the absence of
pelvic discontinuity, and who underwent revision hip arthroplasty
with the use of TM augments were included in the study. A total
of 41 patients with minimum follow-up of two years were included
and evaluated using intention-to-treat analysis.Aims
Patients and Methods
Aims
Patients and Methods
The traditional transosseus flexor hallucis longus (FHL) tendon
transfer for patients with Achilles tendinopathy requires two incisions
to harvest a long tendon graft. The use of a bio-tenodesis screw
enables a short graft to be used and is less invasive, but lacks
supporting evidence about its biomechanical behaviour. We aimed,
in this study, to compare the strength of the traditional transosseus
tendon-to-tendon fixation with tendon-to-bone fixation using a tenodesis
screw, in cyclical loading and ultimate load testing. Tendon grafts were undertaken in 24 paired lower-leg specimens
and randomly assigned in two groups using fixation with a transosseus
suture (suture group) or a tenodesis screw (screw group). The biomechanical
behaviour was evaluated using cyclical and ultimate loading tests.
The Student’s Aims
Materials and Methods
The aim of this study was to examine the results of the acetabular
distraction technique in achieving implantation of a stable construct,
obtaining biological fixation, and producing healing of chronic
pelvic discontinuity at revision total hip arthroplasty. We identified 32 patients treated between 2006 and 2013 who underwent
acetabular revision for a chronic pelvic discontinuity using acetabular
distraction, and who were radiographically evaluated at a mean of
62 months (25 to 160). Of these patients, 28 (87.5%) were female.
The mean age at the time of revision was 67 years (44 to 86). The patients
represented a continuous series drawn from two institutions that
adhered to an identical operative technique.Aims
Patients and Methods
Severe acetabular bone loss and pelvic discontinuity (PD) present particular challenges in revision total hip arthroplasty. To deal with such complex situations, cup-cage reconstruction has emerged as an option for treating this situation. We aimed to examine our success in using this technique for these anatomical problems. We undertook a retrospective, single-centre series of 35 hips in 34 patients (seven male, 27 female) treated with a cup-cage construct using a trabecular metal shell in conjunction with a titanium cage, for severe acetabular bone loss between 2011 and 2015. The mean age at the time of surgery was 70 years (42 to 85) and all patients had an acetabular defect graded as Paprosky Type 2C through to 3B, with 24 hips (69%) having PD. The mean follow-up was 47 months (25 to 84).Aims
Patients and Methods
The purpose of this retrospective study was to evaluate the minimum
five-year outcome of revision total hip arthroplasty (THA) using
the Kerboull acetabular reinforcement device (KARD) in patients
with Paprosky type III acetabular defects and destruction of the
inferior margin of the acetabulum. We identified 36 patients (37 hips) who underwent revision THA
under these circumstances using the KARD, fresh frozen allograft
femoral heads, and reconstruction of the inferior margin of the
acetabulum. The Merle d’Aubigné system was used for clinical assessment.
Serial anteroposterior pelvic radiographs were used to assess migration
of the acetabular component.Aims
Patients and Methods
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
We compared the clinical and radiological outcomes of using a
polyetheretherketone cage with (TiPEEK) and without a titanium coating
(PEEK) for instrumented transforaminal lumbar interbody fusion (TLIF). We conducted a randomised clinical pilot trial of 40 patients
who were scheduled to undergo a TLIF procedure at one or two levels
between L2 and L5. The Oswestry disability index (ODI), EuroQoL-5D,
and back and leg pain were determined pre-operatively, and at three,
six, and 12 months post-operatively. Fusion rates were assessed
by thin slice CT at three months and by functional radiography at
12 months.Aims
Materials and Methods
After intercalary resection of a bone tumour from the femur,
reconstruction with a vascularized fibular graft (VFG) and massive
allograft is considered a reliable method of treatment. However,
little is known about the long-term outcome of this procedure. The
aims of this study were to determine whether the morbidity of this
procedure was comparable to that of other reconstructive techniques,
if it was possible to achieve a satisfactory functional result, and
whether biological reconstruction with a VFG and massive allograft
could achieve a durable, long-lasting reconstruction. A total of 23 patients with a mean age of 16 years (five to 40)
who had undergone resection of an intercalary bone tumour of the
femur and reconstruction with a VFG and allograft were reviewed
clinically and radiologically. The mean follow-up was 141 months
(24 to 313). The mean length of the fibular graft was 18 cm (12 to
29). Full weight-bearing without a brace was allowed after a mean
of 13 months (seven to 26).Aims
Patients and Methods
To assess the extent of osteointegration in two designs of shoulder
resurfacing implants. Bony integration to the Copeland cylindrical
central stem design and the Epoca RH conical-crown design were compared. Implants retrieved from six patients in each group were pair-matched.
Mean time to revision surgery of Copeland implants was 37 months
(standard deviation (Aims
Patients and Methods
The Corail stem has good long-term results. After four years
of using this stem, we have detected a small group of patients who
have presented with symptomatic metaphyseal debonding. The aim of
this study was to quantify the incidence of this complication, to
delineate the characteristics of patients presenting with this complication
and to compare these patients with asymptomatic controls to determine
any important predisposing factors. Of 855 Corail collarless cementless stems implanted for osteoarthritis,
18 presented with symptomatic metaphyseal debonding. A control group
of 74 randomly selected patients was assembled. Clinical and radiological
parameters were measured and a logistic regression model was created
to evaluate factors associated with metaphyseal debonding.Aims
Patients and Methods
The aim of this study was to compare early functional and health
related quality of life outcomes (HRQoL) in patients who have undergone
total hip arthroplasty (THA) using a bone conserving short stem
femoral component and those in whom a conventional length uncemented
component was used. Outcome was assessed using a validated performance
based outcome instrument as well as patient reported outcome measures
(PROMs). We prospectively analysed 33 patients whose THA involved a contemporary
proximally porous coated tapered short stem femoral component and
53 patients with a standard conventional femoral component, at a
minimum follow-up of two years. The mean follow-up was 31.4 months
(24 to 39). Patients with poor proximal femoral bone quality were
excluded. The mean age of the patients was 66.6 years (59 to 77)
and the mean body mass index was 30.2 kg/m2 (24.1 to
41.0). Outcome was assessed using the Oxford Hip Score (OHS) and
the University College Hospital (UCH) hip score which is a validated
performance based instrument. HRQoL was assessed using the EuroQol
5D (EQ-5D).Aims
Patients and Methods
Acetabular bone loss is a challenging problem
facing the revision total hip replacement surgeon. Reconstruction
of the acetabulum depends on the presence of anterosuperior and
posteroinferior pelvic column support for component fixation and
stability. The Paprosky classification is most commonly used when
determining the location and degree of acetabular bone loss. Augments
serve the function of either providing primary construct stability
or supplementary fixation. When a pelvic discontinuity is encountered we advocate the use
of an acetabular distraction technique with a jumbo cup and modular
porous metal acetabular augments for the treatment of severe acetabular
bone loss and associated chronic pelvic discontinuity. Cite this article:
Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone. Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.Objectives
Methods