Modern metal-on-metal (MoM) hip resurfacing arthroplasty (HRA), while achieving good results with well-orientated, well-designed components in ideal patients, is contraindicated in women, men with head size under 50 mm, or metal hypersensitivity. These patients currently have no access to the benefits of HRA. Highly crosslinked polyethylene (XLPE) has demonstrated clinical success in total hip arthroplasty (THA) and, when used in HRA, potentially reduces metal ion-related sequelae. We report the early performance of HRA using a direct-to-bone cementless mono-bloc XLPE component coupled with a cobalt-chrome femoral head, in the patient group for whom HRA is currently contraindicated. This is a cross-sectional, observational assessment of 88 consecutive metal-on-XLPE HRAs performed in 84 patients between 2015 and 2018 in three centres (three surgeons, including the designer surgeon). Mean follow-up is 1.6 years (0.7 to 3.9). Mean age at operation was 56 years (Objectives
Methods
This single-centre observational study aimed to describe the results of extensive bone impaction grafting of the whole acetabular cavity in combination with an uncemented component in acetabular revisions performed in a standardized manner since 1993. Between 1993 and 2013, 370 patients with a median age of 72 years (interquartile range (IQR) 63 to 79 years) underwent acetabular revision surgery. Of these, 229 were more than ten years following surgery and 137 were more than 15 years. All revisions were performed with extensive use of morcellized allograft firmly impacted into the entire acetabular cavity, followed by insertion of an uncemented component with supplementary screw fixation. All types of reoperation were captured using review of radiographs and medical charts, combined with data from the local surgical register and the Swedish Hip Arthroplasty Register.Aims
Methods
Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients.Aims
Patients and Methods
The aim of this study was to evaluate blood metal ion levels, leucocyte profiles, and serum cytokines in patients with a total hip arthroplasty (THA) involving modular dual-mobility components. A total of 39 patients were recruited, with clinical follow-up of up to two years. Outcome was assessed using the Harris Hip Score (HHS, the 12-Item Short-Form Health Survey (SF-12), the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and a visual analogue scale (VAS) for pain. Blood concentrations of cobalt (Co), chromium (Cr), and serum cytokines were measured. Subpopulations of leucocytes were analyzed by flow cytometry.Aims
Patients and Methods
Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for adverse local tissue reactions (ALTR) to metal. We determined the early- to mid-term serum Co and Cr levels after modular dual-mobility components were used in revision and complex primary THAs, and specifically included patients revised for ALTR. Serum Co and Cr levels were measured prospectively in 24 patients with a modular dual-mobility construct and a ceramic femoral head. Patients with CoCr heads or contralateral THAs with CoCr heads were excluded. The mean age was 63 years (35 to 83), with 13 patients (54%) being female. The mean follow-up was four years (2 to 7). Indications for modular dual-mobility were prosthetic joint infection treated with two-stage exchange and subsequent reimplantation (n = 8), ALTR revision (n = 7), complex primary THA (n = 7), recurrent instability (n = 1), and periprosthetic femoral fracture (n = 1). The mean preoperative Co and Cr in patients revised for an ALTR were 29.7 μg/l (2 to 146) and 21.5 μg/l (1 to 113), respectively.Aims
Patients and Methods
To determine ten-year failure rates following 36 mm metal-on-metal
(MoM) Pinnacle total hip arthroplasty (THA), and identify predictors
of failure. We retrospectively assessed a single-centre cohort of 569 primary
36 mm MoM Pinnacle THAs (all Corail stems) followed up since 2012
according to Medicines and Healthcare Products Regulation Agency
recommendations. All-cause failure rates (all-cause revision, and
non-revised cross-sectional imaging failures) were calculated, with predictors
for failure identified using multivariable Cox regression.Aims
Patients and Methods
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
Ceramic-on-metal (CoM) is a relatively new bearing
combination for total hip arthroplasty (THA) with few reported outcomes.
A total of 287 CoM THAs were carried out in 271 patients (mean age
55.6 years (20 to 77), 150 THAs in female patients, 137 in male)
under the care of a single surgeon between October 2007 and October
2009. With the issues surrounding metal-on-metal bearings the decision
was taken to review these patients between March and November 2011,
at a mean follow-up of 34 months (23 to 45) and to record pain,
outcome scores, radiological analysis and blood ion levels. The
mean Oxford Hip Score was 19.2 (12 to 53), 254 patients with 268
hips (95%) had mild/very mild/no pain, the mean angle of inclination
of the acetabular component was 44.8o (28o to
63o), 82 stems (29%) had evidence of radiolucent lines
of >
1 mm in at least one Gruen zone and the median levels of cobalt
and chromium ions in the blood were 0.83 μg/L (0.24 μg/L to 27.56 μg/L)
and 0.78 μg/L (0.21 μg/L to 8.84 μg/L), respectively. The five-year
survival rate is 96.9% (95% confidence interval 94.7% to 99%). Due to the presence of radiolucent lines and the higher than
expected levels of metal ions in the blood, we would not recommend
the use of CoM THA without further long-term follow-up. We plan
to monitor all these patients regularly. Cite this article:
We investigated the incidence of soft-tissue
lesions after small head metal-on-metal total hip replacement (MoM THR).
Between December 1993 and May 1999, 149 patients (195 hips) underwent
primary cementless MoM THR. During the follow-up period, three patients (five THRs) died
and eight patients (14 THRs) were lost to follow-up. We requested
that all patients undergo CT evaluation. After exclusion of five
patients (six THRs) who had undergone a revision procedure, and
22 (28 THRs) who were unwilling to take part in this study, 111
patients (142 THRs) were evaluated. There were 63 men (88 THRs)
and 48 women (54 THRs) with a mean age of 45.7 years (37 to 56)
at the time of surgery. The mean follow-up was 15.4 years (13 to
19). A soft-tissue lesion was defined as an abnormal peri-prosthetic
collection of fluid, solid lesion or asymmetrical soft-tissue mass. At final follow-up, soft-tissue lesions were found in relation
to 28 THRs (19.7%), including 25 solid and three cystic lesions.
They were found in 20 men and eight women; 26 lesions were asymptomatic
and two were symptomatic. The mean maximal diameter of the soft-tissue
lesion was 42.3 mm (17 to 135). The relatively high rate of soft-tissue lesions
observed with small head MoM THR remains a concern. Cite this article:
Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene
THAs, and four primary osteoarthritis synovial membranes. The latter
were analyzed to assess the amount of tissue fibrosis that might
have been present at the time of implantation to enable evaluation,
in relation to implantation time, of any subsequent response in
the tissues.Aims
Patients and Methods
For this retrospective cohort study, patients aged ≤ 30 years
(very young) who underwent total hip arthroplasty (THA) were compared
with patients aged ≥ 60 years (elderly) to evaluate the rate of
revision arthroplasty, implant survival, the indications for revision,
the complications, and the patient-reported outcomes. We retrospectively reviewed all patients who underwent primary
THA between January 2000 and May 2015 from our institutional database.
A total of 145 very young and 1359 elderly patients were reviewed.
The mean follow-up was 5.3 years (1 to 18). Logistic generalized
estimating equations were used to compare characteristics and the revision
rate. Survival was evaluated using Kaplan–Meier curves and hazard
rates were created using Cox regression.Aims
Patients and Methods
The aim of this study was to identify predictors of return to
work (RTW) after revision lower limb arthroplasty in patients of
working age in the United Kingdom. We assessed 55 patients aged ≤ 65 years after revision total
hip arthroplasty (THA). There were 43 women and 12 men with a mean
age of 54 years (23 to 65). We also reviewed 30 patients after revision
total knee arthroplasty (TKA). There were 14 women and 16 men with
a mean age of 58 years (48 to 64). Preoperatively, age, gender,
body mass index, social deprivation, mode of failure, length of
primary implant survival, work status and nature, activity level
(University of California, Los Angeles (UCLA) score), and Oxford
Hip and Knee Scores were recorded. Postoperatively, RTW status,
Oxford Hip and Knee Scores, EuroQol-5D (EQ-5D), UCLA score, and
Work, Osteoarthritis and Joint-Replacement Questionnaire (WORQ)
scores were obtained. Univariate and multivariate analysis was performed.Aims
Patients and Methods
We present a case series of five patients who had revision surgery
following magnetic controlled growing rods (MGCR) for early onset
scoliosis. Metallosis was found during revision in four out of five
patients and we postulated a mechanism for rod failure based on
retrieval analysis. Retrieval analysis was performed on the seven explanted rods.
The mean duration of MCGR from implantation to revision was 35 months
(17 to 46). The mean age at revision was 12 years (7 to 15; four
boys, one girl).Aims
Patients and Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods