We studied 108 patients (116 hips) who were followed for a minimum of six years (73 to 89 months) after primary total hip arthroplasty using an
We have compared the changes in the pattern of the principal strains in the proximal femur after insertion of eight
We studied the effect of a metal tray with an intramedullary stem on the micromotion of the tibial component in total knee arthroplasty. Of 32
To determine the effect of a change in design of a cementless
ceramic acetabular component in fixation and clinical outcome after
total hip arthroplasty We compared 342 hips (302 patients) operated between 1999 and
2005 with a relatively smooth hydroxyapatite coated acetabular component
(group 1), and 337 hips (310 patients) operated between 2006 and
2011 using a similar acetabular component with a macrotexture on
the entire outer surface of the component (group 2). The mean age of
the patients was 53.5 (14 to 70) in group 1 and 53.0 (15 to 70)
in group 2. The mean follow-up was 12.7 years (10 to 17) for group
1 and 7.2 years (4 to 10) for group 2.Aims
Patients and Methods
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
This study reports on the first 150 consecutive
Oxford cementless unicompartmental knee arthroplasties (UKA) performed
in an independent centre (126 patients). All eligible patients had
functional scores (Oxford knee score and high activity arthroplasty
score) recorded pre-operatively and at two- and five-years of follow-up. Fluoroscopically
aligned radiographs were taken at five years and analysed for any
evidence of radiolucent lines (RLLs), subsidence or loosening. The
mean age of the cohort was 63.6 years (39 to 86) with 81 (53.1%)
males. Excellent functional scores were maintained at five years
and there were no progressive RLLs demonstrated on radiographs.
Two patients underwent revision to a total knee arthroplasty giving
a revision rate of 0.23/100 (95% confidence interval 0.03 to 0.84)
component years with overall component survivorship of 98.7% at
five years. There were a further four patients who underwent further
surgery on the same knee, two underwent bearing exchanges for dislocation
and two underwent lateral UKAs for disease progression. This was
a marked improvement from other UKAs reported in New Zealand Joint
Registry data and supports the designing centre’s early results. Cite this article:
The inherent challenges of total hip replacement
(THR) in children include the choice of implant for the often atypical
anatomical morphology, its fixation to an immature growing skeleton
and the bearing surface employed to achieve a successful long-term
result. We report the medium-term results of 52 consecutive uncemented
THRs undertaken in
35 paediatric patients with juvenile idiopathic arthritis. The mean
age at the time of surgery was 14.4 years (10 to 16). The median
follow-up was 10.5 years (6 to 15). During the study period 13 THRs
underwent revision surgery. With revision as an endpoint, subgroup
analysis revealed 100% survival of the 23 ceramic-on-ceramic THRs
and 55% (16 of 29) of the metal- or ceramic-on-polyethylene. This
resulted in 94% (95% CI 77.8 to 98.4) survivorship of the femoral
component and 62% (95% CI 41.0 to 78.0) of the acetabular component.
Revision of the acetabular component for wear and osteolysis were
the most common reasons for failure accounting for 11 of the 13
revisions. The success seen in patients with a ceramic-on-ceramic articulation
seems to indicate that this implant strategy has the potential to
make a major difference to the long-term outcome in this difficult
group of patients.
We present the 10- to 17-year results of 112 computer-assisted design computer-assisted manufacture femoral components. The total hip replacements were performed between 1992 and 1998 in 111 patients, comprising 53 men and 58 women. Their mean age was 46.2 years (24.6 to 62.2) with a mean follow-up of 13 years (10 to 17). The mean Harris Hip Score improved from 42.4 (7 to 99) to 90.3 (38 to 100), the mean Oxford Hip Score from 43.1 (12 to 59) to 18.2 (12 to 51) and the mean Western Ontario MacMasters University Osteoarthritis Index score from 57.0 (7 to 96) to 11.9 (0 to 85). There was one revision due to failure of the acetabular component but no failures of the femoral component. There were no revisions for aseptic loosening. The worst-case survival in this cohort of custom femoral components at 13.2 years follow-up was 98.2% (95% confidence interval 95 to 99). Overall survival of this series of total hip replacements was 97.3% (95% confidence interval 95 to 99). These results are comparable with the best medium- to long-term results for femoral components used in primary total hip replacement with any means of fixation.
Periprosthetic femoral fracture (PFF) is a potentially
devastating complication after total hip arthroplasty, with historically
high rates of complication and failure because of the technical
challenges of surgery, as well as the prevalence of advanced age
and comorbidity in the patients at risk. This study describes the short-term outcome after revision arthroplasty
using a modular, titanium, tapered, conical stem for PFF in a series
of 38 fractures in 37 patients. The mean age of the cohort was 77 years (47 to 96). A total of
27 patients had an American Society of Anesthesiologists grade of
at least 3. At a mean follow-up of 35 months (4 to 66) the mean
Oxford Hip Score (OHS) was 35 (15 to 48) and comorbidity was significantly
associated with a poorer OHS. All fractures united and no stem needed
to be revised. Three hips in three patients required further surgery
for infection, recurrent PFF and recurrent dislocation and three
other patients required closed manipulation for a single dislocation.
One stem subsided more than 5 mm but then stabilised and required
no further intervention. In this series, a modular, tapered, conical stem provided a versatile
reconstruction solution with a low rate of complications. Cite this article:
We investigated the rate of polyethylene wear of a cementless acetabular component at different periods of follow-up in order to test the hypothesis than an irrecoverable deformation process (creep) was followed by an initially low, but gradually increasing wear rate. We studied prospectively 93 uncemented total hip arthroplasties in 83 patients (mean age 50 years (22 to 63)) with a mean follow-up of 8.2 years (3 to 12). We measured the penetration of the femoral head from radiographs taken immediately after surgery at three, six and nine years, or at the latest follow-up. The median wear rate was 0.17 mm per year in the first three years, a finding which we considered to be caused by creep. Thereafter, the rate of wear declined to 0.07 mm per year (four- to six-year period) and then increased to 0.17 mm per year (seven to nine years) and 0.27 mm per year (more than nine years), which we considered to be a reflection of genuine polyethylene wear. After the nine-year follow-up the wear rates were higher in patients with marked osteolysis. We found no relationship between the inclination angle of the acetabular component or femoral head orientation and the rate of wear. No acetabular component required revision.
In developmental dysplasia of the hip (DDH),
a bone defect is often observed superior to the acetabulum after
the reconstruction at the level of the true acetabulum during total
hip replacement (THR). However, the essential amount of uncemented
acetabular component coverage required for a satisfactory outcome
remains controversial. The purpose of this study was to assess the
stability and function of acetabular components with a lack of coverage >
30% (31% to 50%). A total of 760 DDH patients underwent THR with
acetabular reconstruction at the level of the true floor. Lack of
coverage above the acetabular component of >
30% occurred in 56
patients. Intra-operatively, autogenous morcellised bone grafts
were used to fill the uncovered portion. Other than two screws inserted through
the acetabular shell, no additional structural supports were used
in these hips. In all, four patients were lost to follow-up. Therefore,
52 patients (52 hips, 41 women and 11 men) with a mean age of 60.1
years (42 to 78) were available for this study at a mean of 4.8
years (3 to 7). There were no instances of prosthesis revision or
marked loosening during the follow-up. The Harris hip score improved
from a mean of 40.7 points ( Cite this article:
We treated 50 consecutive patients with infected total hip arthroplasties according to a standard protocol. Previous surgery to eradicate the infection had been attempted in 13 patients and discharging sinuses were present in 20. Aspiration arthrography was routinely carried out before our interventions. The first stage was a meticulous removal of all foreign and potentially infected material. Samples were taken for culture and a thorough lavage carried out. Antibiotic-loaded beads were placed in the femoral shaft and an antibiotic-loaded cement ball in the acetabulum. At the second stage an uncemented arthroplasty was introduced. Bone allograft was used in 18 patients. The interval between procedures was usually three weeks, but this was extended if the wound was slow to heal or there was extensive bony destruction. Appropriate antibiotics were given for three months. At a mean follow-up of 5.8 years the rate of reinfection was 8% (4 patients). Two of these patients have had another, successful, two-stage revision. At this medium-term review, a satisfactory clinical and radiological outcome was obtained in all except two patients.
This study compared the primary stability of two commercially
available acetabular components from the same manufacturer, which
differ only in geometry; a hemispherical and a peripherally enhanced
design (peripheral self-locking (PSL)). The objective was to determine
whether altered geometry resulted in better primary stability. Acetabular components were seated with 0.8 mm to 2 mm interference
fits in reamed polyethylene bone substrate of two different densities
(0.22 g/cm3 and 0.45 g/cm3). The primary stability
of each component design was investigated by measuring the peak
failure load during uniaxial pull-out and tangential lever-out tests.Objective
Methods
Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.
Polyethylene wear of acetabular components is a key factor in the development of periprosthetic osteolysis and wear at the articular surface has been well documented and quantified, but fewer data are available about changes which occur at the backside of the liner. At revision surgery for loosening of the femoral component we retrieved 35 conventional modular acetabular liners of the same design. Linear and volumetric articular wear, backside volumetric change and the volume of the screw-head indentations were quantified. These volumes, clinical data and the results from radiological The rate of backside volumetric change was found to be 2.8% of the rate of volumetric articular wear and decreased with increasing liner size. Migrated acetabular components showed significantly higher rates of backside volumetric change plus screw-head indentations than those without migration. The backside volumetric change was at least ten times larger than finite-element simulation had suggested. In a stable acetabular component with well-anchored screws, the amount of backside wear should not cause clinical problems. Impingement of the screw-heads could produce more wear particles than those generated at the liner-shell interface. Because the rate of backside volumetric change is only 2.8% of the rate of volumetric articular wear and since creep is likely to contribute a significant portion to this, the debris generated by wear at the backside of the liner may not be sufficient to create a strong osteolytic response.
We report a high rate of failure of the Ring polyethylene cementless cup caused largely by granulomatous osteolysis. We have reviewed 126 prostheses inserted from 1986 to 1992 at from 11 to 90 months after surgery. There was radiological evidence of osteolytic granulomas adjacent to the external surface of the cup in 32%, appearing on average at three years from operation. In a subgroup of 59 prostheses followed for at least four years the incidence of such changes was 54%. A total of 27 cups (22%) have required revision, 21 for granulomatous loosening at an average follow-up of five years. In the retrieved prostheses there was obvious polyethylene abrasion and histological examination confirmed the presence of polyethylene wear debris. We found no significant correlation of osteolysis with cup size, although smaller cups were predominant among those having revision.
We report the clinical and tribological performance of 67 ceramic acetabular prostheses implanted between 1976 and 1979 without bone cement. They articulated with ceramic femoral heads mounted on mental femoral stems. After a mean elapsed period of 144 months, 59 sockets were radiographically stable but two showed early signs and six showed late signs of loosening. Four of the loose sockets have been revised. Histological analysis of the retrieved tissue showed a fibrous membrane around all the implants, with fibrocartilage in some. There was no bone ingrowth, and the fibrous membrane was up to 6 mm thick and infiltrated with lymphocytes, plasma cells, and macrophages. Intra- and extracellular birefringent wear particles were seen. Tribological analysis showed total wear rates in two retrieved alumina-on-alumina joints of 2.6 microns per year in a stable implant and 68 microns in a loose implant. Survival analysis showed a revision rate of 12.4% at 136 months.
The Ring plastic-on-metal total hip replacement was introduced with the intention of combining the advantages of the cemented low-friction arthroplasty with those of a cementless system. Fourteen hundred and eighty-eight of these procedures have been reviewed. The results of a two to seven-year follow-up, and of a group which has completed at least five years are presented. There was an excellent clinical result in 93% of the whole series, and in 87% of the five-year group. When graded on the Charnley scale there was an average score of 5.86 for pain, 5.96 for function and 5.9 for range of movement, which compares favourably with cemented replacements. The deep infection rate was zero, and the revision rate 1.5% for the whole series. Erosive problems occurred with the smaller acetabular components and the reasons are discussed. The ease and success of exchange have been maintained.
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
We describe the clinical and radiological results
of cementless primary total hip replacement (THR) in 25 patients
(18 women and seven men; 30 THRs) with severe developmental dysplasia
of the hip (DDH). Their mean age at surgery was 47 years (23 to
89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe
type IV. Cementless acetabular components with standard polyethylene
liners were introduced as close to the level of the true acetabulum
as possible. The modular cementless S-ROM femoral component was
used with a low resection of the femoral neck. A total of 21 patients (25 THRs) were available for review at
a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified
Harris hip score improved from 46 points pre-operatively to 90 at
final follow up (p <
0.001). A total of 15 patients (17 THRs; 57%) underwent revision of the
acetabular component at a mean of 14.6 years (7 to 20.8), all for
osteolysis. Two patients (two THRs) had symptomatic loosening. No
patient underwent femoral revision. Survival with revision of either
component for any indication was 81% at 15 years (95% CI 60.1 to
92.3), with 21 patients at risk. This technique may reduce the need for femoral osteotomy in severe
DDH, while providing a good long-term functional result. Cite this article: