This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1). The PFC Sigma knee system appears to provide excellent results in the medium term.
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene
wear A total of 120 patients, 85 women and 35 men with a mean age
of 70 years (59 to 80) who were scheduled for primary cemented total
hip arthroplasty were randomly allocated to four study groups. Patients
were blinded to their group assignment and received either a conventional
polyethylene (CPE) or a highly cross-linked (HXL) acetabular component
of identical design. On the femoral side patients received a 28
mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated
RSA examinations over five years. Clinical outcome was measured
using the Harris hip score. There was no difference in polyethylene wear between the two
head materials when used with either of the two types of acetabular
component (p = 0.3 to 0.6). When comparing the two types of polyethylene
there was a significant difference in favour of HXLPE, regardless
of the head material used (p <
0.001). In conclusion, we found no advantage of OxZr over CoCr in terms
of polyethylene wear after five years of follow-up. Our findings
do not support laboratory results which have shown a reduced rate
of wear with OxZr. They do however add to the evidence on the better
resistance to wear of HXLPE over CPE. Cite this article:
A method is described by which an artificial limb can be fitted as soon as an above-knee or below-knee amputation wound is healed, by using preformed sockets which enable the prosthetist to fit the limb within the hour. The method has been found safe, and better than other methods previously described to facilitate the early mobilisation and rehabilitation of the amputee.
We report the results of 62 hips in 62 patients
(17 males, 45 females) with mean age of 62.4 years (37 to 81), who underwent
revision of the acetabular component of a total hip replacement
due to aseptic loosening between May 2003 and November 2007. All
hips had a Paprosky type IIIa acetabular defect. Acetabular revision
was undertaken using a Procotyl E cementless oblong implant with
modular side plates and a hook combined with impaction allografting. At a mean follow-up of 60.5 months (36 to 94) with no patients
lost to follow-up and one died due to unrelated illness, the complication
rate was 38.7%. Complications included aseptic loosening (19 hips),
deep infection (3 hips), broken hook and side plate (one hip) and
a femoral nerve palsy (one hip). Further revision of the acetabular component
was required in 18 hips (29.0%) and a further four hips (6.4%) are
currently loose and awaiting revision. We observed unacceptably high rates of complication and failure
in our group of patients and cannot recommend this implant or technique.
Aims.
Aims. The aim of this study was to investigate the rate of revision for distal femoral arthroplasty (DFA) performed as a primary procedure for native knee fractures using data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR). Methods. Data from the AOANJRR were obtained for DFA performed as primary procedures for native knee fractures from 1 September 1999 to 31 December 2020. Pathological fractures and revision for failed internal fixation were excluded. The five prostheses identified were the Global
Aims. Taper corrosion has been widely reported to be problematic for
The December 2023 Foot & Ankle Roundup. 360. looks at: Subchondral bone cysts remodel after correction of varus deformity in ankle arthritis; 3D-printed
Aims. United Classification System (UCS) B2 and B3 periprosthetic fractures in total hip arthroplasties (THAs) have been commonly managed with
Aims. Several different designs of hemiarthroplasty are used to treat intracapsular fractures of the proximal femur, with large variations in costs. No clinical benefit of
Aims. The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore
The August 2024 Oncology Roundup. 360. looks at: What factors are associated with osteoarthritis after cementation for benign aggressive bone tumour of the knee joint: a systematic review and meta-analysis; Recycled bone grafts treated with extracorporeal irradiation or liquid nitrogen freezing after malignant tumour resection; Intercalary resection of the tibia for primary bone tumours: are vascularized fibula autografts with or without allografts a durable reconstruction?; 3D-printed
Aims. Dislocation remains a leading cause of failure following revision total hip arthroplasty (THA). While dual-mobility (DM) bearings have been shown to mitigate this risk, options are limited when retaining or implanting an uncemented shell without
Aims. In 2015, we published the results of our ceramic-on-metal (CoM) total hip arthroplasties (THAs) performed between October 2007 and July 2009 with a mean follow-up of 34 months (23 to 45) and a revision rate of 3.1%. The aim of this paper is to present the longer-term outcomes. Methods. A total of 264 patients were reviewed at a mean of 5.8 years (4.6 to 7.2) and 10.1 years (9.2 to 10.6) to determine revision rate, pain, outcome scores, radiological analysis, and blood ion levels. Those who were unwilling or unable to travel were contacted by telephone. Results. The all-cause revision rate at six years was 3.1% (eight THAs), increasing to 8.8% (18 THAs) at ten years. Of these, there were four and then seven bearing-related revisions at six and ten years, respectively. There was a statistically significant deterioration in the visual analogue scale pain score and Oxford Hip Score (OHS) between six and ten years. There were 18 CoM THAs in 17 patients who had a cobalt or chromium level over 4 ppb and ten CoM THAs in nine patients who had a cobalt or chromium level higher than 7 ppb with a statistically significant increase in chromium levels only between the two timepoints. Overall, 84 stems (39.1%) had significant radiolucent lines at ten years compared to 65 (25.5%) at six years. Conclusion. When compared to the original review, there has been a significant deterioration in pain score, OHS, radiograph appearance, and, most critically, survival has fallen to 91.2%, which does not meet the Orthopaedic Data Evaluation Panel (ODEP) 10 A* 95% threshold. Although this bearing is no longer on the market, 2.5% were bearing-related revisions, which have relevance to the discussion around
Aims. In metal-on-metal (MoM) hip arthroplasties and resurfacings, mechanically induced corrosion can lead to elevated serum metal ions, a local inflammatory response, and formation of pseudotumours, ultimately requiring revision. The size and diametral clearance of anatomical (ADM) and
Aims. The proximal tibia (PT) is the anatomical site most frequently affected by primary bone tumours after the distal femur. Reconstruction of the PT remains challenging because of the poor soft-tissue cover and the need to reconstruct the extensor mechanism. Reconstructive techniques include implantation of massive endoprosthesis (megaprosthesis), osteoarticular allografts (OAs), or allograft-prosthesis composites (APCs). Methods. This was a retrospective analysis of clinical data relating to patients who underwent proximal tibial arthroplasty in our regional bone tumour centre from 2010 to 2018. Results. A total of 76 patients fulfilled the inclusion criteria and were included in the study. Mean age at surgery was 43.2 years (12 to 86 (SD 21)). The mean follow-up period was 60.1 months (5.4 to 353). In total 21 failures were identified, giving an overall failure rate of 27.6%. Prosthesis survival at five years was 75.5%, and at ten years was 59%. At last follow-up, mean knee flexion was 89.8° (SD 36°) with a mean extensor lag of 18.1° (SD 24°). In univariate analysis, factors associated with better survival of the prosthesis were a malignant or metastatic cancer diagnosis (versus benign), with a five- and ten-year survival of 78.9% and 65.7% versus 37.5% (p = 0.045), while in-hospital length of stay longer than nine days was also associated with better prognosis with five- and ten-year survival rates at 84% and 84% versus 60% and 16% (p < 0.001). In multivariate analysis, only in-hospital length of stay was associated with longer survival (hazard ratio (HR) 0.23, 95% confidence interval (CI) 0.08 to 0.66). Conclusion. We have shown that proximal tibial arthroplasty with endoprosthesis is a safe and reliable method for reconstruction in patients treated for orthopaedic oncological conditions. Either
Aims. Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods. An institutional review board-approved retrieval archive was surveyed for
Aims. Cutibacterium acnes (C. acnes; previously known as Propionibacterium acnes or P. acnes) periprosthetic hip and knee infections are under-reported. While culture contamination with C. acnes occurs, true infections are important to recognize and treat. We sought to describe the demographics and treatment outcomes of patients with C. acnes periprosthetic joint infections (PJIs) of the hip and knee. Methods. Patients with C. acnes PJI between January 2005 and December 2018 were retrospectively reviewed utilizing the institutional total joint registry. Patients with monomicrobial PJI and two or more positive cultures were considered to have true C. acnes PJI. Patients with polymicrobial infection or with only one positive culture were excluded. This resulted in 35 PJIs (21 hips and 14 knees); the patients’ mean age was 63 years (35 to 84) and 15 (43%) were female. Mean follow-up was five years (1 to 14). Results. The median time to positive culture was five days (IQR 5 to 6) and median synovial fluid cell count was 22,583 cells (IQR 15,200 to 53,231). The median ESR was 25 mm/hr (IQR 7 to 37), and CRP was 15 mg/l (IQR 3 to 29). Of the 35 PJIs, 18 (51%) were treated with chronic antibiotic suppression without surgical intervention, and the remainder were treated with two-stage exchange arthroplasty. The two-year survival free of any revision was 94%. Four patients failed treatment due to symptomatic infection, with three treated with two-stage exchange and one treated with irrigation and debridement with
Aims. The number of revision total ankle arthroplasties (TAAs) which are undertaken is increasing. Few studies have reported the survival after this procedure. The primary aim of this study was to analyze the survival of revision ankle arthroplasties using large datasets. Secondary aims were to summarize the demographics of the patients, the indications for revision TAA, further operations, and predictors of survival. Methods. The study combined data from the National Joint Registry and NHS Digital to report the survival of revision TAA. We have previously reported the failure rates and risk factors for failure after TAA, and the outcome of fusion after a failed TAA, using the same methodology. Survival was assessed using life tables and Kaplan Meier graphs. Cox proportional hazards regression models were fitted to compare failure rates. Results. A total of 228 patients underwent revision TAA. The mean follow-up was 2.6 years (SD 2.0). The mean time between the initial procedure and revision was 2.3 years (SD 1.8). The most commonly used implant was the Inbone which was used in 81 patients. A total of 29 (12.7%) failed; nine (3.9%) patients underwent a further revision, 19 (8.3%) underwent a fusion, and one (0.4%) had an amputation. The rate of survival was 95.4% (95% confidence interval (CI) 91.6 to 97.5) at one year, 87.7% (95% CI 81.9 to 91.7; n = 124) at three years and 77.5% (95% CI 66.9 to 85.0; n = 57) at five years. Revision-specific implants had a better survival than when primary implants were used at revision. A total of 50 patients (21.9%) had further surgery; 19 (8.3%) underwent reoperation in the first 12 months. Cox regression models were prepared. In crude analysis the only significant risk factors for failure were the use of cement (hazard ratio (HR) 3.02 (95% CI 1.13 to 8.09)) and the time since the primary procedure (HR 0.67 (95% CI 0.47 to 0.97)). No risk factors for failure were identified in multivariable Cox regression modelling. Conclusion. Revision TAAs have good medium term survival and low rates of further surgery. New
Aims. With increasing burden of revision hip arthroplasty (THA), one of the major challenges is the management of proximal femoral bone loss associated with previous multiple surgeries. Proximal femoral arthroplasty (PFA) has already been popularized for tumour surgeries. Our aim was to describe the outcome of using PFA in these demanding non-neoplastic cases. Methods. A retrospective review of 25 patients who underwent PFA for non-neoplastic indications between January 2009 and December 2015 was undertaken. Their clinical and radiological outcome, complication rates, and survival were recorded. All patients had the Stanmore Implant –