Wear debris released from bearing surfaces has been shown to
provoke negative immune responses in the recipient. Excessive wear
has been linked to early failure of prostheses. Analysis using coordinate
measuring machines (CMMs) can provide estimates of total volumetric
material loss of explanted prostheses and can help to understand
device failure. The accuracy of volumetric testing has been debated,
with some investigators stating that only protocols involving hundreds
of thousands of measurement points are sufficient. We looked to
examine this assumption and to apply the findings to the clinical
arena. We examined the effects on the calculated material loss from
a ceramic femoral head when different CMM scanning parameters were
used. Calculated wear volumes were compared with gold standard gravimetric
tests in a blinded study. Objectives
Methods
We investigated factors that were thought to be associated with an increased incidence of squeaking of ceramic-on-ceramic total hip replacements. Between June 1997 and December 2008 the three senior authors implanted 2406 primary total hip replacements with a ceramic-on-ceramic bearing surface. The mean follow-up was 10.6 years. The diagnosis was primary osteoarthritis in each case, and no patient had undergone previous surgery to the hip. We identified 74 squeaking hips (73 patients) giving an incidence of 3.1% at a mean follow-up of 9.5 years (4.1 to 13.3). Taller, heavier and younger patients were significantly more likely to have hips that squeaked. Squeaking hips had a significantly higher range of post-operative internal (p = 0.001) and external rotation (p = 0.003) compared with silent hips. Patients with squeaking hips had significantly higher activity levels (p = 0.009). A squeaking hip was not associated with a significant difference in patient satisfaction (p = 0.24) or Harris hip score (p = 0.34). Four implant position factors enabled good prediction of squeaking. These were high acetabular component inclination, high femoral offset, lateralisation of the hip centre and either high or low acetabular component anteversion. This is the largest study to date to examine patient factors and implant position factors that predispose to squeaking of a ceramic-on-ceramic hip. The results suggest that factors which increase the mechanical forces across the hip joint and factors which increase the risk of neck-to-rim impingement, and therefore edge-loading, are those that predispose to squeaking.
Most published randomised controlled trials which
compare the rates of wear of conventional and cross-linked (XL) polyethylene
(PE) in total hip arthroplasty (THA) have described their use with
a cementless acetabular component. We conducted a prospective randomised study to assess the rates
of penetration of two distinct types of PE in otherwise identical
cemented all-PE acetabular components. A total of 100 consecutive patients for THA were randomised to
receive an acetabular component which had been either highly XL
then remelted or moderately XL then annealed. After a minimum of eight years follow-up, 38 hips in the XL group
and 30 hips in the annealed group had complete data (mean follow-up
of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively).
In the XL group, the steady state rate of penetration from one year
onwards was -0.0002 mm/year ( These results show that the yearly linear rate of femoral head
penetration can be significantly reduced by using a highly XLPE
cemented acetabular component. Cite this article:
We performed 114 consecutive primary total hip arthroplasties with a cementless expansion acetabular component in 101 patients for advanced osteonecrosis of the femoral head. The mean age of the patients at surgery was 51 years (36 to 62) and the mean length of follow-up was 110 months (84 to 129). The mean pre-operative Harris hip score of 47 points improved to 93 points at final follow-up. The polyethylene liner was exchanged in two hips during this period and one broken acetabular component was revised. The mean linear wear rate of polyethylene was 0.07 mm/year and peri-acetabular osteolysis was seen in two hips (1.9%). Kaplan-Meier analysis indicated that the survival of the acetabular component without revision was 97.8% (95% confidence interval 0.956 to 1.000) at ten years. Our study has shown that the results of THA with a cementless expansion acetabular component and an alumina-polyethylene bearing surface are good.
This study reports the results of 38 total hip
arthroplasties (THAs) in 33 patients aged <
50 years, using the
JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.
This represents an update of previous reports of the same cohort
at ten and 16 years, which were reported in 2004 and 2009, respectively.
We describe the survival, radiological and functional outcomes at
a mean follow-up of 21 years (17 to 25). Of the surviving 34 THAs,
one underwent femoral revision for peri-prosthetic fracture after
21 years, and one patient (one hip) was lost to follow-up. Using
aseptic loosening as the end-point, 12 hips (31.5%) needed acetabular
revision but none needed femoral revision, demonstrating 100% survival
(95% confidence interval 89 to 100). In young patients with high demands, the Furlong HAC–coated femoral
component gives excellent long-term results. Cite this article:
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article:
Bioengineering reasons for increased wear and failure of metal-on-metal (MoM) bearings in hip prostheses have been described. Low wear occurs in MoM hips when the centre of the femoral head is concentric with the centre of the acetabular component and the implants are correctly positioned. Translational or rotational malpositioning of the components can lead to the contact-patch of the femoral component being displaced to the rim of the acetabular component, resulting in a ten- to 100-fold increase in wear and metal ion levels. This may cause adverse tissue reactions, loosening of components and failure of the prosthesis.
Large femoral heads have become popular in total
hip replacement (THR) as a method of reducing the risk of dislocation.
However, if large heads are used in ceramic-on-ceramic THR, the
liner must be thinner, which may increase the risk of fracture.
To compare the rates of ceramic fracture and dislocation between
28 mm and 32 mm ceramic heads, 120 hips in 109 patients (51 men
and 58 women, mean age 49.2 years) were randomised to THR with either
a 28 mm or a 32 mm ceramic articulation. A total of 57/60 hips assigned
to the 28 mm group and 55/60 hips assigned to the 32 mm group were
followed for at least five years. No ceramic component fractures
occured in any patient in either group. There was one dislocation
in the 32 mm group and none in the 28 mm group (p = 0.464). No hip
had detectable wear, focal osteolysis or prosthetic loosening. In
our small study the 32 mm ceramic articulation appeared to be safe
in terms of ceramic liner fracture. Cite this article:
This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading. A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12). The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur. This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.
Metallosis is a rare cause of failure after total knee replacement and has only previously been reported when there has been abnormal metal-on-metal contact. We describe 14 patients (15 knees) whose total knee replacement required revision for a new type of early failure caused by extensive metallosis. A modification of a cementless rotating platform implant, which had previously had excellent long-term survival, had been used in each case. The change was in the form of a new porous-beaded surface on the femoral component to induce cementless fixation, which had been used successfully in the fixation of acetabular and tibial components. This modification appeared to have resulted in metallosis due to abrasive two-body wear. The component has subsequently been recalled and is no longer in use. The presentation, investigation, and findings at revision are described and a possible aetiology and its implications are discussed.
We describe the clinical and radiological results
of cementless primary total hip replacement (THR) in 25 patients
(18 women and seven men; 30 THRs) with severe developmental dysplasia
of the hip (DDH). Their mean age at surgery was 47 years (23 to
89). In all, 21 hips had Crowe type III dysplasia and nine had Crowe
type IV. Cementless acetabular components with standard polyethylene
liners were introduced as close to the level of the true acetabulum
as possible. The modular cementless S-ROM femoral component was
used with a low resection of the femoral neck. A total of 21 patients (25 THRs) were available for review at
a mean follow-up of 18.7 years (15.8 to 21.8). The mean modified
Harris hip score improved from 46 points pre-operatively to 90 at
final follow up (p <
0.001). A total of 15 patients (17 THRs; 57%) underwent revision of the
acetabular component at a mean of 14.6 years (7 to 20.8), all for
osteolysis. Two patients (two THRs) had symptomatic loosening. No
patient underwent femoral revision. Survival with revision of either
component for any indication was 81% at 15 years (95% CI 60.1 to
92.3), with 21 patients at risk. This technique may reduce the need for femoral osteotomy in severe
DDH, while providing a good long-term functional result. Cite this article:
We assessed the orientation of the acetabular
component in 1070 primary total hip arthroplasties with hard-on-soft, small
diameter bearings, aiming to determine the size and site of the
target zone that optimises outcome. Outcome measures included complications,
dislocations, revisions and ΔOHS (the difference between the Oxford
Hip Scores pre-operatively and five years post-operatively). A wide
scatter of orientation was observed (2 This study demonstrated that with traditional technology surgeons
can only reliably achieve a target zone of ±15°. As the optimal
zone to diminish the risk of dislocation is also ±15°, surgeons
should be able to achieve this. This is the first study to demonstrate
that optimal orientation of the acetabular component improves the
functional outcome. However, the target zone is small (± 5°) and
cannot, with current technology, be consistently achieved. Cite this article:
We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for zirconium oxide ceramic/Enduron and 0.40 mm for zirconium oxide ceramic/Hylamer. The difference between the groups was significant (p <
0.008) except for stainless steel/Hylamer At present, 12 patients have undergone a revision procedure, four at five years and eight thereafter. No patient who received a stainless steel/Enduron articulation at their primary replacement required revision. Conflicting results have been reported about the performance of the zirconium oxide ceramic femoral head, but our findings suggest that it should not be used with a polymethylmethacrylate acetabular component. Hylamer has already been withdrawn from the market.
We report the clinical and radiological outcome of 86 revisions of cemented hip arthroplasties using JRI-Furlong hydroxyapatite-ceramic-coated acetabular and femoral components. The acetabular component was revised in 62 hips and the femoral component in all hips. The mean follow-up was 12.6 years and no patient was lost to follow-up. The mean age of the patients was 71.2 years. The mean Harris hip and Oxford scores were 82 (59 to 96) and 23.4 (14 to 40), respectively. The mean Charnley modification of the Merle d’Aubigné and Postel score was 5 (3 to 6) for pain, 4.9 (3 to 6) for movement and 4.4 (3 to 6) for mobility. Migration of the acetabular component was seen in two hips and the mean acetabular inclination was 42.6°. The mean linear polyethylene wear was 0.05 mm/year. The mean subsidence of the femoral component was 1.9 mm and stress shielding was seen in 23 (28%) with bony ingrowth in 76 (94%). Heterotopic ossification was seen in 12 hips (15%). There were three re-revisions, two for deep sepsis and one for recurrent dislocation and there were no re-revisions for aseptic loosening. The mean EuroQol EQ-5D description scores and health thermometer scores were 0.69 (0.51 to 0.89) and 79 (54 to 95), respectively. With an end-point of definite or probable loosening, the probability of survival at 12 years was 93.9% and 95.6% for the acetabular and femoral components, respectively. Overall survival at 12 years, with removal or further revision of either component for any reason as the end-point, was 92.3%. Our study supports the continued use of this arthroplasty and documents the durability of hydroxyapatite-ceramic-coated components.
The long-term results of grafting with hydroxyapatite granules for acetabular deficiency in revision total hip replacement are not well known. We have evaluated the results of revision using a modular cup with hydroxyapatite grafting for Paprosky type 2 and 3 acetabular defects at a minimum of ten years’ follow-up. We retrospectively reviewed 49 acetabular revisions at a mean of 135 months (120 to 178). There was one type 2B, ten 2C, 28 3A and ten 3B hips. With loosening as the endpoint, the survival rate was 74.2% (95% confidence interval 58.3 to 90.1). Radiologically, four of the type 3A hips (14%) and six of the type 3B hips (60%) showed aseptic loosening with collapse of the hydroxyapatite layer, whereas no loosening occurred in type 2 hips. There was consolidation of the hydroxyapatite layer in 33 hips (66%). Loosening was detected in nine of 29 hips (31%) without cement and in one of 20 hips (5%) with cement (p = 0.03, Fisher’s exact probability test). The linear wear and annual wear rate did not correlate with loosening. These results suggest that the long-term results of hydroxyapatite grafting with cement for type 2 and 3A hips are encouraging.
Symptomatic hip osteonecrosis is a disabling
condition with a poorly understood aetiology and pathogenesis. Numerous
treatment options for hip osteonecrosis are described, which include
non-operative management and joint preserving procedures, as well
as total hip replacement (THR). Non-operative or joint preserving
treatment may improve outcomes when an early diagnosis is made before
the lesion has become too large or there is radiographic evidence
of femoral head collapse. The presence of a crescent sign, femoral
head flattening, and acetabular involvement indicate a more advanced-stage
disease in which joint preserving options are less effective than
THR. Since many patients present after disease progression, primary
THR is often the only reliable treatment option available. Prior
to the 1990s, outcomes of THR for osteonecrosis were poor. However,
according to recent reports and systemic reviews, it is encouraging
that with the introduction of newer ceramic and/or highly cross-linked
polyethylene bearings as well as highly-porous fixation interfaces,
THR appears to be a reliable option in the management of end-stage
arthritis following hip osteonecrosis in this historically difficult
to treat patient population. Cite this article:
We report on 397 consecutive revision total hip
replacements in 371 patients with a mean clinical and radiological follow-up
of 12.9 years (10 to 17.7). The mean age at surgery was 69 years
(37 to 93). A total of 28 patients (8%) underwent further revision,
including 16 (4%) femoral components. In all 223 patients (56%,
233 hips) died without further revision and 20 patients (5%, 20
hips) were lost to follow-up. Of the remaining patients, 209 (221
hips) were available for clinical assessment and 194 (205 hips)
for radiological review at mean follow-up of 12.9 years (10 to 17.7). The mean Harris Hip Score improved from 58.7 (11 to 92) points
to 80.7 (21 to 100) (p <
0.001) and the mean Merle d’Aubigné and
Postel hip scores at final follow-up were 4.9 (2 to 6), 4.5 (2 to
6) and 4.3 (2 to 6), respectively for pain, mobility and function.
Radiographs showed no lucencies around 186 (90.7%) femoral stems
with stable bony ingrowth seen in 199 stems (97%). The survival
of the S-ROM femoral stem at 15 years with revision for any reason as
the endpoint was 90.5% (95% confidence interval (CI) 85.7 to 93.8)
and with revision for aseptic loosening as the endpoint 99.3% (95%
CI 97.2 to 99.8). We have shown excellent long-term survivorship and good clinical
outcome of a cementless hydroxyapatite proximally-coated modular
femoral stem in revision hip surgery. Cite this article:
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
A total of 397 hips were randomised to receive
Metasul metal-on-metal (MoM), metal-on-conventional polyethylene (MoP)
or ceramic-on-polyethylene (CoP) bearings using a cemented triple-tapered
polished femoral component (MS-30). There were 129 MoM hips in 123
patients (39 male and 84 female, mean age 63.3 years (40.7 to 72.9)),
137 MoP hips in 127 patients (39 male and 88 female, mean age 62.8
years (24.5 to 72.7)) and 131 CoP hips in 124 patients (51 male
and 73 female, mean age 63.9 years (30.6 to 73.8)). All acetabular
components were cemented Weber polyethylene components with the
appropriate inlay for the MoM articulation. Clinical evaluation
was undertaken using the Harris hip score (HHS) and radiological
assessments were made at two, five and seven years. The HHS and radiological
analysis were available for 341 hips after seven years. The MoM
group had the lowest mean HHS (p = 0.124), a higher rate of revision
(p <
0.001) and a higher incidence of radiolucent lines in unrevised
hips (p <
0.001). In all, 12 revisions had been performed in
12 patients: eight in the MoM group (four for infection, four for
aseptic loosening, three in the MoP group (one each of infection,
dislocation and pain) and one in the CoP group (infection). Our findings reveal no advantage to the MoM bearing and identified
a higher revision rate and a greater incidence of radiolucent lines
than with the other articulations. We recommend that patients with
a 28 mm Metasul MoM bearing be followed carefully. Cite this article:
Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article: