Advertisement for orthosearch.org.uk
Results 81 - 100 of 309
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 6 | Pages 908 - 913
1 Nov 1997
Masterson EL Masri BA Duncan CP Rosenberg A Cabanela M Gross M

An analysis of the cement mantle obtained with the Exeter impaction allografting system at one centre showed that it was either deficient or absent in almost 47% of Gruen zones. We therefore examined the mantle obtained using this system at another hospital and compared the results with those from the CPT and Harris Precoat Systems at other centres. The surgical indications for the procedure and the patient details were broadly similar in all four hospitals. There was some variation in the frequency of use of cortical strut allografts, cerclage wires and wire mesh to supplement the impaction allograft. Analysis of the cement mantles showed that when uncertain Gruen zones were excluded, the incidence of zones with areas of absence or deficiency of the cement was 47% and 50%, respectively, for the two centres using the Exeter system, 21% for the CPT system and 18% for the Harris Precoat system. We measured the difference in size between the proximal allograft impactors and the definitive prosthesis for each system. The Exeter system impactors are shorter than the definitive prosthesis and taper sharply so that the cavity created is inadequate, especially distally. The CPT proximal impactors are considerably longer than the definitive prosthesis and are designed to give a mantle of approximately 2 mm medially and laterally and 1.5 mm anteriorly and posteriorly. The Harris Precoat proximal impactors allow for a mantle with a circumference of 0.75 mm in the smaller sizes and 1 mm in the larger. Many reports link the longevity of a cemented implant to the adequacy of the cement mantle. For this reason, femoral impaction systems require careful design to achieve a cement mantle which is uninterrupted in its length and adequate in its thickness. Our results suggest that some current systems require modification


Bone & Joint 360
Vol. 11, Issue 6 | Pages 6 - 11
1 Dec 2022
Roberton A Stocker M Phillips J


Bone & Joint Open
Vol. 5, Issue 6 | Pages 452 - 456
1 Jun 2024
Kennedy IW Rooney EJ Ryan PJ Siva S Kennedy MJ Wheelwright B Young D Meek RMD

Aims

Femoral periprosthetic fractures are rising in incidence. Their management is complex and carries a high associated mortality. Unlike native hip fractures, there are no guidelines advising on time to theatre in this group. We aim to determine whether delaying surgical intervention influences morbidity or mortality in femoral periprosthetic fractures.

Methods

We identified all periprosthetic fractures around a hip or knee arthroplasty from our prospectively collated database between 2012 and 2021. Patients were categorized into early or delayed intervention based on time from admission to surgery (early = ≤ 36 hours, delayed > 36 hours). Patient demographics, existing implants, Unified Classification System fracture subtype, acute medical issues on admission, preoperative haemoglobin, blood transfusion requirement, and length of hospital stay were identified for all patients. Complication and mortality rates were compared between groups.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 946 - 952
1 Sep 2023
Dhawan R Young DA Van Eemeren A Shimmin A

Aims

The Birmingham Hip Resurfacing (BHR) arthroplasty has been used as a surgical treatment of coxarthrosis since 1997. We present 20-year results of 234 consecutive BHRs performed in our unit.

Methods

Between 1999 and 2001, there were 217 patients: 142 males (65.4%), mean age 52 years (18 to 68) who had 234 implants (17 bilateral). They had patient-reported outcome measures collected, imaging (radiograph and ultrasound), and serum metal ion assessment. Survivorship analysis was performed using Kaplan-Meier estimates. Revision for any cause was considered as an endpoint for the analysis.


Bone & Joint 360
Vol. 12, Issue 4 | Pages 20 - 23
1 Aug 2023

The August 2023 Foot & Ankle Roundup360 looks at: Achilles tendon rupture: surgery or conservative treatment for the high-demand patient?; First ray amputation in diabetic patients; Survival of ankle arthroplasty in the UK; First metatarsophalangeal joint fusion and flat foot correction; Intra-articular corticosteroid injections with or without hyaluronic acid in the management of subtalar osteoarthritis; Factors associated with nonunion of post-traumatic subtalar arthrodesis; The Mayo Prosthetic Joint Infection Risk Score for total ankle arthroplasty.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 958 - 963
1 Jul 2017
Mamarelis G Key S Snook J Aldam C

Aims. Hip hemiarthroplasty is a standard treatment for intracapsular proximal femoral fractures in the frail elderly. In this study we have explored the implications of early return to theatre, within 30 days, on patient outcome following hip hemiarthroplasty. Patients and Methods. We retrospectively reviewed the hospital records of all hip hemiarthroplasties performed in our unit between January 2010 and January 2015. Demographic details, medical backround, details of the primary procedure, complications, subsequent procedures requiring return to theatre, re-admissions, discharge destination and death were collected. Results. A total of 705 procedures were included; 428 Austin Moore and 277 Exeter Trauma Stems were used. A total of 34 fractures (in 33 patients) required early return to theatre within 30 days. Age, gender, laterality, time from admission to primary procedure, American Society of Anesthesiologists grade, and implant type were similar for those requiring early return to theatre and those who did not. Early return to theatre was associated with a significantly higher length of stay (mean 33.6 days (7 to 107) versus 18.6 days (0 to 152), p < 0.001), re-admission rate (38.2% versus 8.6%, p < 0.001), and subsequent revision rate (17.6% versus 1.3%, p < 0.001). We found no difference in level of care required on discharge or mortality. Conclusion. Proximal femoral fractures are common in the elderly population, with far-reaching medical and economic implications. Factors such as infection or dislocation may require early return to theatre, and this is associated with outcomes which may be both medically and economically detrimental. This illustrates the importance of avoiding early complications to improve longer term outcome. Return to theatre within 30 days is associated with longer length of stay, higher re-admission rate, and higher subsequent revision rate. It may be a useful short-term quality indicator for longer term outcome measures following hip hemiarthroplasty for intracapsular fractures of the proximal femur. Cite this article: Bone Joint J 2017;99-B:958–63


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 802 - 807
1 Aug 2024
Kennedy IW Sinnerton R Jeyakumar G Kane N Young D Meek RMD

Aims

The number of revision arthroplasties being performed in the elderly is expected to rise, including revision for infection. The primary aim of this study was to measure the treatment success rate for octogenarians undergoing revision total hip arthroplasty (THA) for periprosthetic joint infection (PJI) compared to a younger cohort. Secondary outcomes were complications and mortality.

Methods

Patients undergoing one- or two-stage revision of a primary THA for PJI between January 2008 and January 2021 were identified. Age, sex, BMI, American Society of Anesthesiologists grade, Charlson Comorbidity Index (CCI), McPherson systemic host grade, and causative organism were collated for all patients. PJI was classified as ‘confirmed’, ‘likely’, or ‘unlikely’ according to the 2021 European Bone and Joint Infection Society criteria. Primary outcomes were complications, reoperation, re-revision, and successful treatment of PJI. A total of 37 patients aged 80 years or older and 120 patients aged under 80 years were identified. The octogenarian group had a significantly lower BMI and significantly higher CCI and McPherson systemic host grades compared to the younger cohort.


The Bone & Joint Journal
Vol. 105-B, Issue 5 | Pages 511 - 517
1 May 2023
Petrie MJ Panchani S Al-Einzy M Partridge D Harrison TP Stockley I

Aims

The duration of systemic antibiotic treatment following first-stage revision surgery for periprosthetic joint infection (PJI) after total hip arthroplasty (THA) is contentious. Our philosophy is to perform an aggressive debridement, and to use a high local concentration of targeted antibiotics in cement beads and systemic prophylactic antibiotics alone. The aim of this study was to assess the success of this philosophy in the management of PJI of the hip using our two-stage protocol.

Methods

The study involved a retrospective review of our prospectively collected database from which we identified all patients who underwent an intended two-stage revision for PJI of the hip. All patients had a diagnosis of PJI according to the major criteria of the Musculoskeletal Infection Society (MSIS) 2013, a minimum five-year follow-up, and were assessed using the MSIS working group outcome-reporting tool. The outcomes were grouped into ‘successful’ or ‘unsuccessful’.


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 435 - 441
1 May 2024
Angelomenos V Mohaddes M Kärrholm J Malchau H Shareghi B Itayem R

Aims

Refobacin Bone Cement R and Palacos R + G bone cement were introduced to replace the original cement Refobacin Palacos R in 2005. Both cements were assumed to behave in a biomechanically similar fashion to the original cement. The primary aim of this study was to compare the migration of a polished triple-tapered femoral stem fixed with either Refobacin Bone Cement R or Palacos R + G bone cement. Repeated radiostereometric analysis was used to measure migration of the femoral head centre. The secondary aims were evaluation of cement mantle, stem positioning, and patient-reported outcome measures.

Methods

Overall, 75 patients were included in the study and 71 were available at two years postoperatively. Prior to surgery, they were randomized to one of the three combinations studied: Palacos cement with use of the Optivac mixing system, Refobacin with use of the Optivac system, and Refobacin with use of the Optipac system. Cemented MS30 stems and cemented Exceed acetabular components were used in all hips. Postoperative radiographs were used to assess the quality of the cement mantle according to Barrack et al, and the position and migration of the femoral stem. Harris Hip Score, Oxford Hip Score, Forgotten Joint Score, and University of California, Los Angeles Activity Scale were collected.


The Bone & Joint Journal
Vol. 107-B, Issue 1 | Pages 50 - 57
1 Jan 2025
Hussein Y Iljazi A Sørensen MS Overgaard S Petersen MM

Aims

Dislocation is a major concern following total hip arthroplasty (THA) for osteoarthritis (OA). Both dual-mobility components and standard acetabular components with large femoral heads are used to reduce the risk of dislocation. We investigated whether dual-mobility components are superior to standard components in reducing the two-year dislocation and revision risk in a propensity-matched sample from the Danish Hip Arthroplasty Register (DHR).

Methods

This population-based cohort study analyzed data from the DHR and the Danish National Patient Register. We included all patients undergoing primary THA for OA from January 2010 to December 2019 with either dual-mobility or standard acetabular components with metal-on-polyethylene or ceramic-on-polyethylene articulations with a 36 mm femoral head. The samples were propensity score-matched on patient and implant characteristics. The primary outcome was the difference in the absolute risk of dislocation within two years, with a secondary outcome of the difference in the absolute risk of revision surgery of any cause within the same timeframe. The cumulative incidence of dislocation was calculated using the Aalen-Johansen estimator, while the difference in absolute risk was estimated using absolute risk regression (ARR).


Bone & Joint 360
Vol. 12, Issue 3 | Pages 18 - 22
1 Jun 2023

The June 2023 Foot & Ankle Roundup360 looks at: Nail versus plate fixation for ankle fractures; Outcomes of first ray amputation in diabetic patients; Vascular calcification on plain radiographs of the ankle to diagnose diabetes mellitus; Elderly patients with ankle fracture: the case for early weight-bearing; Active treatment for Frieberg’s disease: does it work?; Survival of ankle arthroplasty; Complications following ankle arthroscopy.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1617 - 1625
1 Dec 2013
Schmitz MWJL Timmer C Rijnen WHC Gardeniers JWM Schreurs BW

Despite the worldwide usage of the cemented Contemporary acetabular component (Stryker), no published data are available regarding its use in patients aged < 50 years. We undertook a mid- to long-term follow-up study, including all consecutive patients aged < 50 years who underwent a primary total hip replacement using the Contemporary acetabular component with the Exeter cemented stem between January 1999 and January 2006. There were 152 hips in 126 patients, 61 men and 65 women, mean age at surgery 37.6 years (16 to 49 yrs). One patient was lost to follow-up. Mean clinical follow-up of all implants was 7.6 years (0.9 to 12.0). All clinical questionnaire scores, including Harris hip score, Oxford hip score and several visual analogue scales, were found to have improved. The eight year survivorship of all acetabular components for the endpoints revision for any reason or revision for aseptic loosening was 94.4% (95% confidence interval (CI) 89.2 to 97.2) and 96.4% (95% CI 91.6 to 98.5), respectively. Radiological follow-up was complete for 146 implants. The eight year survival for the endpoint radiological loosening was 93.1% (95% CI 86.2 to 96.6). Three surviving implants were considered radiologically loose but were asymptomatic. The presence of acetabular osteolysis (n = 17, 11.8%) and radiolucent lines (n = 20, 13.9%) in the 144 surviving cups indicates a need for continued observation in the second decade of follow-up in order to observe their influence on long-term survival. The clinical and radiological data resulting in a ten-year survival rate > 90% in young patients support the use of the Contemporary acetabular component in this specific patient group. Cite this article: Bone Joint J 2013;95-B:1617–25


Bone & Joint Open
Vol. 3, Issue 9 | Pages 710 - 715
5 Sep 2022
Khan SK Tyas B Shenfine A Jameson SS Inman DS Muller SD Reed MR

Aims

Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes.

Methods

Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.


Bone & Joint Open
Vol. 4, Issue 9 | Pages 659 - 667
1 Sep 2023
Nasser AAHH Osman K Chauhan GS Prakash R Handford C Nandra RS Mahmood A

Aims

Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade.

Methods

Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.


Bone & Joint Open
Vol. 4, Issue 1 | Pages 38 - 46
17 Jan 2023
Takami H Takegami Y Tokutake K Kurokawa H Iwata M Terasawa S Oguchi T Imagama S

Aims

The objectives of this study were to investigate the patient characteristics and mortality of Vancouver type B periprosthetic femoral fractures (PFF) subgroups divided into two groups according to femoral component stability and to compare postoperative clinical outcomes according to treatment in Vancouver type B2 and B3 fractures.

Methods

A total of 126 Vancouver type B fractures were analyzed from 2010 to 2019 in 11 associated centres' database (named TRON). We divided the patients into two Vancouver type B subtypes according to implant stability. Patient demographics and functional scores were assessed in the Vancouver type B subtypes. We estimated the mortality according to various patient characteristics and clinical outcomes between the open reduction internal fixation (ORIF) and revision arthroplasty (revision) groups in patients with unstable subtype.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1475 - 1480
1 Nov 2015
Inngul C Blomfeldt R Ponzer S Enocson A

The aim of this randomised controlled study was to compare functional and radiological outcomes between modern cemented and uncemented hydroxyapatite coated stems after one year in patients treated surgically for a fracture of the femoral neck. A total of 141 patients aged > 65 years were included. Patients were randomised to be treated with a cemented Exeter stem or an uncemented Bimetric stem. The patients were reviewed at four and 12 months. The cemented group performed better than the uncemented group for the Harris hip score (78 vs 70.7, p = 0.004) at four months and for the Short Musculoskeletal Function Assesment Questionnaire dysfunction score at four (29.8 vs 39.2, p = 0.007) and 12 months (22.3 vs 34.9, p = 0.001). The mean EQ-5D index score was better in the cemented group at four (0.68 vs 0.53, p = 0.001) and 12 months (0.75 vs 0.58, p = < 0.001) follow-up. There were nine intra-operative fractures in the uncemented group and none in the cemented group. . In conclusion, our data do not support the use of an uncemented hydroxyapatite coated stem for the treatment of displaced fractures of the femoral neck in the elderly. Cite this article: Bone Joint J 2015;97-B:1475–80


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 82 - 87
1 Jan 2023
Barrie A Kent B

Aims

Management of displaced paediatric supracondylar elbow fractures remains widely debated and actual practice is unclear. This national trainee collaboration aimed to evaluate surgical and postoperative management of these injuries across the UK.

Methods

This study was led by the South West Orthopaedic Research Division (SWORD) and performed by the Supra Man Collaborative. Displaced paediatric supracondylar elbow fractures undergoing surgery between 1 January 2019 and 31 December 2019 were retrospectively identified and their anonymized data were collected via Research Electronic Data Capture (REDCap).


Bone & Joint Research
Vol. 11, Issue 5 | Pages 270 - 277
6 May 2022
Takegami Y Seki T Osawa Y Imagama S

Aims

Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems.

Methods

We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes.


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 94 - 99
1 Jan 2015
Grammatopoulos G Wilson HA Kendrick BJL Pulford EC Lippett J Deakin M Andrade AJ Kambouroglou G

National Institute of Clinical Excellence guidelines state that cemented stems with an Orthopaedic Data Evaluation Panel (ODEP) rating of > 3B should be used for hemiarthroplasty when treating an intracapsular fracture of the femoral neck. These recommendations are based on studies in which most, if not all stems, did not hold such a rating. . This case-control study compared the outcome of hemiarthroplasty using a cemented (Exeter) or uncemented (Corail) femoral stem. These are the two prostheses most commonly used in hip arthroplasty in the UK. Data were obtained from two centres; most patients had undergone hemiarthroplasty using a cemented Exeter stem (n = 292/412). Patients were matched for all factors that have been shown to influence mortality after an intracapsular fracture of the neck of the femur. Outcome measures included: complications, re-operations and mortality rates at two, seven, 30 and 365 days post-operatively. Comparable outcomes for the two stems were seen. . There were more intra-operative complications in the uncemented group (13% vs 0%), but the cemented group had a greater mortality in the early post-operative period (n = 6). There was no overall difference in the rate of re-operation (5%) or death (365 days: 26%) between the two groups at any time post-operatively. This study therefore supports the use of both cemented and uncemented stems of proven design, with an ODEP rating of 10A, in patients with an intracapsular fracture of the neck of the femur. Cite this article: Bone Joint J 2015;97-B:94–9


The Journal of Bone & Joint Surgery British Volume
Vol. 70-B, Issue 3 | Pages 358 - 363
1 May 1988
Shelley P Wroblewski B

An experimental evaluation was made of the efficacy of an unflanged cup, an ogee-flanged cup and the Exeter pressuriser in the pressurisation of acetabular cement. Only a very modest injection pressure could be generated through an unflanged socket and even this was readily lost as the socket "bottomed out". The ogee-flanged socket gave a consistently high injection pressure which could be maintained throughout the process of polymerisation. The Exeter pressuriser gave marginally better results. Of the two cements tested, Palacos with gentamicin showed better intrusion than CMW Type 1 cement, but it also allowed greater extrusion once pressure was released. The importance of maintaining a continuous pressure on the cement throughout polymerisation is emphasised