The Journal of Bone & Joint Surgery British Volume
Vol. 66-B, Issue 1 | Pages 55 - 62
1 Jan 1984
Grimer R Karpinski M Edwards A

One hundred and three sequential Stanmore knee replacements were reviewed retrospectively on two occasions with a maximal follow-up period of nine years three months. This knee prosthesis, which is hinged, was successful in alleviating pain, stabilising an unstable knee and modestly increasing the arc of flexion. Walking capacity was increased and flexion contractures were reduced. There were seven cases of infection and four of fracture around the prosthesis. All these proved difficult to treat and two knees with both fracture and infection needed amputation. Eight knees were revised for aseptic loosening and a further 14 were found to have radiological signs of loosening. The results have been analysed by the methods advocated by Tew and Waugh and give a cumulative success rate of 80 per cent at seven years, provided success is judged solely by whether the prosthesis is still in situ. The role of the Stanmore knee as a primary arthroplasty is discussed


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 6 | Pages 809 - 817
1 Aug 2003
Halliday BR English HW Timperley AJ Gie GA Ling RSM

We report the results of cancellous femoral impaction grafting with cement in revision hip arthroplasty in all patients from one centre who had undergone surgery more than five years previously. A total of 32 surgeons undertook femoral impaction grafting in 207 patients (226 hips). There were no deaths attributable to the revision surgery; 33 patients with 35 functioning hips died with less than five years’ follow-up. One patient was lost to follow-up. Two hips (1%) developed early postoperative infection. Of the 12 stems which underwent a further surgical procedure for aseptic failure, ten were for femoral fracture and two for loosening. Survivorship with any further femoral operation as the endpoint was 90.5% (confidence intervals, 82 to 98) and using femoral reoperation for symptomatic aseptic loosening as the endpoint, the survivorship was 99.1% (confidence intervals, 96 to 100) at 10 to 11 years. As a consequence of the experience in this series, we have modified our technique with an increased use of longer stems with impacted allograft. Long stems are indicated when the host bone around the tip of a short stem is compromised, in patients with major loss of bone stock, or when a femoral fracture occurs


The Journal of Bone & Joint Surgery British Volume
Vol. 57-B, Issue 1 | Pages 59 - 62
1 Feb 1975
Phillips H Taylor JG

Eighty-three Waildius arthroplasties, performed by one surgeon as a salvage operation on the knee joint between 1966 and 1972, were independently reviewed. The fifty-seven living patients with sixty-seven arthroplasties were interviewed and examined and the clinical records of the deceased patients were inspected. Sixty-seven arthroplasties (81 per cent) were successful and sixteen failed (19 per cent). Acrylic cement was used to secure the prosthesis on eight occasions only. There were two primary infections (24 per cent) and two delayed (24 per cent). Major loosening occurred in three arthroplasties (36 per cent). Minor loosening was compatible with a good result. Arthrodesis was successful on the two occasions on which it became necessary to remove the implant. There were no disasters. On the basis of these results it is considered that the Walidius arthroplasty can justifiably be offered as an alternative to primary arthrodesis of the knee


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 920 - 924
1 Jul 2006
Mori T Kudo H Iwano K Juji T

We studied 11 patients (14 elbows) with gross rheumatoid deformity of the elbow, treated by total arthroplasty using the Kudo type-5 unlinked prosthesis, and who were evaluated between five and 11 years after operation. Massive bone defects were augmented by autogenous bone grafts. There were no major complications such as infection, subluxation or loosening. In most elbows relief from pain and stability were achieved. The results, according to the Mayo Elbow Performance Score, were excellent in eight, good in five and fair in one. In most elbows there was minimal or no resorption of the grafted bone. There were no radiolucent lines around the stems of the cementless components. This study shows that even highly unstable rheumatoid elbows can be replaced successfully using an unlinked prosthesis, with augmentation by grafting for major defects of bone


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 362 - 365
1 May 1995
Raut V Siney P Wroblewski B

We reviewed prospectively cemented stem revision in 106 patients with severe femoral endosteal bone lysis without infection. Bone grafts were not used in any of the patients. The minimum follow-up was three years (mean 6 years 4 months). At the last review 76.4% of the patients were free of pain and 17.9% had only mild or occasional discomfort; radiographs showed well-fixed stable stems in 101 (95.3%). An intramedullary cement plug was used at revision to improve stability in 97.7%. There was new endosteal osteolysis after revision in 17 patients; only two had severe changes. Seven hips (6.6%) required a second revision; only four of these (3.8%) were for stem loosening. Survivorship of the revised stem, using radiological evidence of stem loosening as the end point, was 95.8% at seven years. The results of stem revision arthroplasty using cement in the presence of massive endosteal cavitation are satisfactory


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 3 | Pages 387 - 391
1 Apr 2002
Sandén B Olerud C Petrén-Mallmin M Larsson S

We investigated the effects of hydroxyapatite (HA) coating on the purchase of pedicle screws. A total of 23 consecutive patients undergoing lumbar fusion was randomly assigned to one of three treatment groups. The first received uncoated stainless-steel screws, the second screws which were partly coated with HA, and the third screws which were fully coated. The insertion torque was recorded. After 11 to 16 months, 21 screws had been extracted. The extraction torque was recorded. Radiographs were taken to assess fusion and to detect loosening of the screws. At removal, the extraction torques exceeded the upper limit of the torque wrench (600 Ncm) for many HA-coated screws. The calculated mean extraction torque was 29 ± 36 Ncm for the uncoated group, 447 ± 114 Ncm for the partly-coated group and 574 ± 52 Ncm for the fully-coated group. There were significant differences between all three groups (p < 0.001). There were more radiolucent zones surrounding the uncoated screws than the HA-coated screws (p < 0.001). HA coating of pedicle screws resulted in improved fixation with reduced risk of loosening of the screws


The Journal of Bone & Joint Surgery British Volume
Vol. 76-B, Issue 2 | Pages 252 - 257
1 Mar 1994
Ferguson G Cabanela M Ilstrup D

From 1969 through 1982, 305 hips in 290 patients had total hip arthroplasty for failed femoral intertrochanteric osteotomy. Of these, we reviewed 215 hips (70.5%) with a minimum follow-up of five years. The results were good or excellent in 79%, but there were technical problems at operation in 23% and a total perioperative complication rate of 11.8%. Late complications occurred in 13.1% including seven late infections (3.2%). At a mean follow-up of ten years, 39 hips had been revised (18.1%), there was probable loosening in 19.5% of stems and 12.6% of cups and possible loosening in 11.4% of stems and 7.2% of cups. The cumulative probability of failure at ten years was 20.6%. We recommend that intertrochanteric osteotomy be advised, planned and executed bearing in mind the possible need for a subsequent total hip arthroplasty, and that screws and plates should be routinely removed soon after union of the osteotomy


Bone & Joint Research
Vol. 1, Issue 4 | Pages 64 - 70
1 Apr 2012
Ritter MA Davis KE Meding JB Farris A

Objectives. The purpose of this study was to examine the effect of posterior cruciate ligament (PCL) retention, PCL recession, and PCL excision during cruciate-retaining total knee replacement. Methods. A total of 3018 anatomic graduated component total knee replacements were examined; 1846 of these retained the PCL, 455 PCLs were partially recessed, and in 717 the PCL was completely excised from the back of the tibia. Results. Clinical scores between PCL groups favored excision for flexion (p < 0.0001), and recession and retention for stairs (p < 0.0001). There was a mild difference in long-term all-cause aseptic survivorship between PCL-retained (96.4% at 15 years) combined with PCL-recessed groups (96.6% at 15 years) when compared with the PCL-excised group (95.0% at 15 years) (p = 0.0411, Wilcoxon; p = 0.0042, log-rank), as well as tibial or femoral loosening, which reported prosthesis survival of 97.8% at 15 years for PCL-retained knees, 98.2% for recessed knees, and 96.4% for excised knees (p = 0.0934, Wilcoxon; p = 0.0202, log-rank). Conclusions. Despite some trade-off in clinical performance, if the PCL is detached at the time of operation, conversion to a posterior-stabilised prosthesis may not be necessarily required as long as stability in the anteroposterior and coronal planes is achieved


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 1 | Pages 77 - 82
1 Jan 2002
Ikävalko M Lehto MUK Repo A Kautiainen H Hämäläinen M

We present the results of 525 primary Souter elbow arthroplasties undertaken in 406 patients between 1982 and 1997. There were 372 women and 34 men with a mean age of 57 years; 119 patients had a bilateral procedure. The elbows were affected by chronic inflammatory disease, usually rheumatoid arthritis, which had been present for a mean of 24.7 years (2 to 70). In about 30% the joints were grossly destroyed with significant loss of bone. In 179 elbows the ulnar components were metal-backed and retentive; in the remaining 346, with better bone stock, non-retentive, all-polyethylene prostheses were used. Because of complications, 108 further operations were required in 82 patients. During the early years the incidence of complications was higher. Dislocation was the indication for 30 further procedures in 26 patients. Thirty patients underwent 33 revision procedures for aseptic loosening, 12 had 29 operations because of deep infection, two for superficial infection, and 14 further operations were done for other reasons. The cumulative rate of success, without aseptic loosening, five and ten years after surgery, was 96% and 85%, respectively


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 3 | Pages 531 - 539
1 May 1998
Goodman SB Huie P Song Y Schurman D Maloney W Woolson S Sibley R

The tissues surrounding 65 cemented and 36 cementless total joint replacements undergoing revision were characterised for cell types by immunohistochemistry and for cytokine expression by in situ hybridisation. We identified three distinct groups of revised implants: loose implants with ballooning radiological osteolysis, loose implants without osteolysis, and well-fixed implants. In the cemented series, osteolysis was associated with increased numbers of macrophages (p = 0.0006), T-lymphocyte subgroups (p = 0.03) and IL-1 (p = 0.02) and IL-6 (p = 0.0001) expression, and in the cementless series with increased numbers of T-lymphocyte subgroups (p = 0.005) and increased TNFα expression (p = 0.04). For cemented implants, the histological, histochemical and cytokine profiles of the interface correlated with the clinical and radiological grade of loosening and osteolysis. Our findings suggest that there are different biological mechanisms of loosening and osteolysis for cemented and cementless implants. T-lymphocyte modulation of macrophage function may be an important interaction at prosthetic interfaces


The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 5 | Pages 652 - 658
1 Sep 1992
Quinn J Joyner C Triffitt J Athanasou N

Macrophages and their fused products are commonly found at the polymethylmethacrylate cement-bone interface, but it is not known if they contribute directly to the osteolysis associated with loosening of the cemented prosthesis. We isolated mononuclear phagocytes from granulomas formed by subcutaneous implantation of polymethylmethacrylate into mice and incubated them on bone slices in which they formed resorption lacunae after co-culture for seven to 14 days with both marrow stromal cells and osteoblast-like cells (in the presence of 1 alpha,25-dihydroxyvitamin D3 and dexamethasone). Increased numbers of tartrate-resistant acid phosphatase-positive mononuclear and multinucleated cells formed in these cultures. Both in the presence and absence of stromal cells, macrophages produced extensive superficial roughening of the bone surface. Polymethylmethacrylate-induced macrophages are thus capable of low-grade surface and high-grade lacunar osteolysis, the latter requiring the presence of specific hormonal and stromal cell elements. These two forms of bone resorption could account for the pathogenesis and clinical patterns associated with loosening of the cemented prosthesis


The Bone & Joint Journal
Vol. 101-B, Issue 7 | Pages 880 - 888
1 Jul 2019
Wei R Guo W Yang R Tang X Yang Y Ji T Liang H

Aims

The aim of this study was to describe the use of 3D-printed sacral endoprostheses to reconstruct the pelvic ring and re-establish spinopelvic stability after total en bloc sacrectomy (TES) and to review its outcome.

Patients and Methods

We retrospectively reviewed 32 patients who underwent TES in our hospital between January 2015 and December 2017. We divided the patients into three groups on the basis of the method of reconstruction: an endoprosthesis group (n = 10); a combined reconstruction group (n = 14), who underwent non-endoprosthetic combined reconstruction, including anterior spinal column fixation; and a spinopelvic fixation (SPF) group (n = 8), who underwent only SPF. Spinopelvic stability, implant survival (IS), intraoperative haemorrhage rate, and perioperative complication rate in the endoprosthesis group were documented and compared with those of other two groups.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 2 | Pages 183 - 189
1 Mar 1997
Pihlajamäki H Myllynen P Böstman O

We analysed the complications encountered in 102 consecutive patients who had posterolateral lumbosacral fusion performed with transpedicular screw and rod fixation for non-traumatic disorders after a minimum of two years. Of these, 40 had spondylolysis and spondylolisthesis, 42 a degenerative disorder, 14 instability after previous laminectomy and decompression, and six pain after nonunion of previous attempts at spinal fusion without internal fixation. There were 75 multilevel and 27 single-level fusions. There were 76 individual complications in 48 patients, and none in the other 54. The complications seen were screw misplacement, coupling failure of the device, wound infection, nonunion, permanent neural injury, and loosening, bending and breakage of screws. Screw breakage or loosening was more common in patients with multilevel fusions (p < 0.001). Screws of 5 mm diameter should not be used for sacral fixation. Forty-six patients had at least one further operation for one or several complications, including 20 fusion procedures for nonunion. The high incidence of complications is a disadvantage of this technically-demanding method


The Journal of Bone & Joint Surgery British Volume
Vol. 56-B, Issue 4 | Pages 626 - 642
1 Nov 1974
Evans EM Freeman MAR Miller AJ Vernon-Roberts B

1. Evidence is presented which suggests that after total joint replacement bone necrosis and consequent loosening of the prosthesis may be due to the development of sensitivity to the metals used. 2. Nine patients, from a total of fourteen with loose prostheses, were found to be metal sensitive by skin-patch testing. In twenty-four patients with intact prostheses no sensitivity was demonstrated. 3. In material from the joints of sensitive patients the metal content was raised. 4. Examination of this material showed necrosis of bone and soft tissue following obliterative changes in the vascular supply. 5. Similar reactions were found following the injection of cobalt into a sensitive patient. 6. The release of metal around a prosthesis is greatest where metal rubs against itself. 7. We conclude that prostheses in which metal articulates with polyethylene should be preferred; that any patient in whom loosening or fragmentation occurs should be patch tested; and that if sensitivity is found the implant should be removed


The Journal of Bone & Joint Surgery British Volume
Vol. 54-B, Issue 4 | Pages 590 - 599
1 Nov 1972
Andersson GBJ Freeman MAR Swanson SAV

1. Prosthetic acetabular cups of the Charnley and McKee-Farrar designs were cemented into cadaveric pelves using different procedures for preparing the acetabulum. 2. The torsional moments needed to loosen these cups were measured. 3. The torsional moments so measured were found to be from about four to more than twenty times higher than the frictional moments measured in independent tests on the two designs of prosthesis. 4. It is argued that late looseness of the acetabular component after total hip replacement, in the absence of infection, seems most likely to be due to thermal damage to the bone occurring at the time of polymerisation of the cement, and to subsequent bone resorption. 5. Surgical preparation of the acetabulum should include removal of all the articular cartilage and cleaning of the acetabular fossa, but the drilling of additional holes in the floor of the acetabulum seems unimportant. 6. The possibility of fatigue fracture in bone as a factor contributing to late loosening is an argument in favour of metal-on-polyethylene prostheses with their lower frictional moments, although the importance of this factor cannot be estimated


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 1 | Pages 66 - 73
1 Jan 1996
Hilding MB Lanshammar H Ryd L

We report a prospective study of gait and tibial component migration in 45 patients with osteoarthritis treated by total knee arthroplasty (TKA). Migration was measured over two years using roentgen stereophotogrammetry. We used the previously established threshold of 200 μm migration in the second postoperative year to distinguish two groups: a risk group of 15 patients and a stable group of 28 patients. We performed gait analysis before operation and at six months and at two years after TKA. On all three occasions we found significant differences between the two groups in the mean sagittal plane moments of the knee joint. The risk group walked with higher peak flexion moments than the stable group. The two groups were not discriminated by any clinical or radiological criteria or other gait characteristics. The relationship which we have found between gait with increased flexion moments and risk of tibial component loosening warrants further study as regards the aetiology of prosthetic loosening and possible methods of influencing its incidence


The Journal of Bone & Joint Surgery British Volume
Vol. 86-B, Issue 2 | Pages 190 - 194
1 Mar 2004
Bizot P Hannouche D Nizard R Witvoet J Sedel L

Between 1990 and 1992, we implanted 71 hybrid alumina-on-alumina hip arthroplasties in 62 consecutive patients under the age of 55 years, with a mean age of 46 years at surgery. There were 56 primary and 15 secondary procedures. The prostheses involved a cemented titanium alloy stem, a 32 mm alumina head, and a press-fit metal-backed socket with an alumina insert. Three patients (four hips) died from unrelated causes. Four hips had revision surgery for either deep infection, unexplained persistent pain, fracture of the alumina head, or aseptic loosening of the socket. The nine-year survival rate was 93.7% with revision for any cause as the end-point and 98.4% with revision for aseptic loosening as the end-point. The outcome in the surviving patients (50 patients, 57 hips) with a minimum five-year follow-up (mean eight years) was excellent in 47 hips (82.5%), very good in eight (14%), good in one and fair in one. A thin, partial, lucent line, mainly in zone III was present in 38% of the sockets and one socket had a complete lucency less than 1 mm thick. One stem had isolated femoral osteolysis. There was no detectable component migration nor acetabular osteolysis. This hybrid arthroplasty gave satisfactory medium-term results in active patients. The press-fit metal-backed socket appeared to have reliable fixation in alumina-on-alumina hip arthroplasty. The excellent results using cemented fixation of the stem may be related to the low production of wear debris


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 3 | Pages 320 - 323
1 Mar 2005
Little CP Ruiz AL Harding IJ McLardy-Smith P Gundle R Murray DW Athanasou NA

We present the histological findings of bone retrieved from beneath the femoral components of failed metal-on-metal hip resurfacing arthroplasties. Of a total of 377 patients who underwent resurfacing arthroplasty, 13 required revision; for fracture of the femoral neck in eight, loosening of a component in three and for other reasons in two. None of these cases had shown histological evidence of osteonecrosis in the femoral bone at the time of the initial implantation. Bone from the remnant of the femoral head showed changes of osteonecrosis in all but one case at revision. In two cases of fracture which occurred within a week of implantation, the changes were compatible with early necrosis of the edge of the fracture. In the remaining six fractures, there were changes of established osteonecrosis. In all but one of the non-fracture cases, patchy osteonecrosis was seen. We conclude that histological evidence of osteonecrosis is a common finding in failed resurfaced hips. Given that osteonecrosis is extensive in resurfaced femoral heads which fail by fracture, it is likely to play a role in the causation of these fractures


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 3 | Pages 389 - 392
1 Apr 2003
Misra AN Hussain MRA Fiddian NJ Newton G

We randomised 129 knees which were to be replaced using a standard posterior-cruciate-ligament (PCL)-retaining cemented total knee replacement into two groups. In one the PCL was retained in the normal way and in the other it was resected. They were well matched, with a predominance of women, and a mean age of 67 years. There was no statistically significant difference in the Hospital for Special Surgery scores at a mean of 57 months (56 to 60) between the two groups although 21 patients (24 knees) were lost to follow-up. Relief from pain, correction of deformity, range of movement, stability and strength were comparable in both. Radiological assessment showed femoral rollback in approximately 20% of knees with a slightly higher incidence in the PCL-resected group. There was no significant loosening detected in either group at review at two years. At five years, one knee in the PCL-retained group had been revised because of infection and one patient in each group was awaiting revision for loosening. Our findings have shown no significant difference in the five-year results for a PCL-retaining total knee replacement if the PCL is excised or preserved. This suggests two important points. First, the PCL is not functional in most patients with a total knee replacement even when retained. Secondly, patients with an excised PCL show a good result with a PCL-retaining implant, thereby questioning the need for a posterior stabilised design in such a situation


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1230 - 1237
1 Oct 2019
Kayani B Konan S Horriat S Ibrahim MS Haddad FS

Aims

The aim of this study was to assess the effect of posterior cruciate ligament (PCL) resection on flexion-extension gaps, mediolateral soft-tissue laxity, fixed flexion deformity (FFD), and limb alignment during posterior-stabilized (PS) total knee arthroplasty (TKA).

Patients and Methods

This prospective study included 110 patients with symptomatic osteoarthritis of the knee undergoing primary robot-assisted PS TKA. All operations were performed by a single surgeon using a standard medial parapatellar approach. Optical motion capture technology with fixed femoral and tibial registration pins was used to assess gaps before and after PCL resection in extension and 90° knee flexion. Measurements were made after excision of the anterior cruciate ligament and prior to bone resection. There were 54 men (49.1%) and 56 women (50.9%) with a mean age of 68 years (sd 6.2) at the time of surgery. The mean preoperative hip-knee-ankle deformity was 4.1° varus (sd 3.4).