Since 1996 more than one million metal-on-metal
articulations have been implanted worldwide. Adverse reactions to
metal debris are escalating. Here we present an algorithmic approach
to patient management. The general approach to all arthroplasty
patients returning for follow-up begins with a detailed history,
querying for pain, discomfort or compromise of function. Symptomatic
patients should be evaluated for intra-articular and extra-articular
causes of pain. In large head MoM arthroplasty, aseptic loosening
may be the source of pain and is frequently difficult to diagnose.
Sepsis should be ruled out as a source of pain. Plain radiographs
are evaluated to rule out loosening and osteolysis, and assess component
position. Laboratory evaluation commences with erythrocyte sedimentation
rate and C-reactive protein, which may be elevated. Serum metal
ions should be assessed by an approved facility. Aspiration, with
manual cell count and culture/sensitivity should be performed, with
cloudy to creamy fluid with predominance of monocytes often indicative
of failure. Imaging should include ultrasound or metal artifact
reduction sequence MRI, specifically evaluating for fluid collections
and/or masses about the hip. If adverse reaction to metal debris
is suspected then revision to metal or ceramic-on-polyethylene is indicated
and can be successful. Delay may be associated with extensive soft-tissue
damage and hence poor clinical outcome.
Medium-term results of the Discovery elbow replacement
are presented. We reviewed 51 consecutive primary Discovery total
elbow replacements (TERs) implanted in 48 patients. The mean age
of the patients was 69.2 years (49 to 92), there were 19 males and
32 females (37%:63%) The mean follow-up was 40.6 months (24 to 69).
A total of six patients were lost to follow-up. Statistically significant
improvements in range movement and Oxford Elbow Score were found
(p <
0.001). Radiolucent lines were much more common in, and
aseptic loosening was exclusive to, the humeral component. Kaplan–Meier
survivorship at five years was 92.2% (95% CI 74.5% to 96.4%) for
aseptic loosening. In four TERs, periprosthetic infection occurred
resulting in failure. A statistically significant association between
infection and increased BMI was found (p = 0.0268). Triceps failure
was more frequent after the Mayo surgical approach and TER performed
after previous trauma surgery. No failures of the implant were noted. Our comparison shows that the Discovery has early clinical results
that are similar to other semi-constrained TERs. We found continued
radiological surveillance with particular focus on humeral lucency
is warranted and has not previously been reported. Despite advances
in the design of total elbow replacement prostheses, rates of complication
remain high. Cite this article:
We have designed a prospective study to evaluate
the usefulness of prolonged incubation of cultures from sonicated
orthopaedic implants. During the study period 124 implants from
113 patients were processed (22 osteosynthetic implants, 46 hip
prostheses, 54 knee prostheses, and two shoulder prostheses). Of
these, 70 patients had clinical infection; 32 had received antibiotics
at least seven days before removal of the implant. A total of 54 patients
had sonicated samples that produced positive cultures (including
four patients without infection). All of them were positive in the
first seven days of incubation. No differences were found regarding
previous antibiotic treatment when analysing colony counts or days
of incubation in the case of a positive result. In our experience, extending
incubation of the samples to 14 days does not add more positive
results for sonicated orthopaedic implants (hip and knee prosthesis
and osteosynthesis implants) compared with a conventional seven-day incubation
period. Cite this article:
The June 2013 Foot &
Ankle Roundup360 looks at: soft-tissue pain following arthroplasty; pigmented villonodular synovitis of the foot and ankle; ankles, allograft and arthritis; open calcaneal fracture; osteochondral lesions in the longer term; severe infections in diabetic feet; absorbable first ray fixation; and showering after foot surgery.
Wear of polyethylene inserts plays an important role in failure
of total knee replacement and can be monitored Before revision, the minimum joint space width values and their
locations on the insert were measured in 15 fully weight-bearing
radiographs. These measurements were compared with the actual minimum
thickness values and locations of the retrieved tibial inserts after
revision. Introduction
Method
Malpositioning of the trochanteric entry point
during the introduction of an intramedullary nail may cause iatrogenic
fracture or malreduction. Although the optimal point of insertion
in the coronal plane has been well described, positioning in the
sagittal plane is poorly defined. The paired femora from 374 cadavers were placed both in the anatomical
position and in internal rotation to neutralise femoral anteversion.
A marker was placed at the apparent apex of the greater trochanter,
and the lateral and anterior offsets from the axis of the femoral
shaft were measured on anteroposterior and lateral photographs. Greater
trochanteric morphology and trochanteric overhang were graded. The mean anterior offset of the apex of the trochanter relative
to the axis of the femoral shaft was 5.1 mm ( Placement of the entry position at the apex of the greater trochanter
in the anteroposterior view does not reliably centre an intramedullary
nail in the sagittal plane. Based on our findings, the site of insertion
should be about 5 mm posterior to the apex of the trochanter to
allow for its anterior offset. Cite this article:
We retrospectively reviewed the records of 1150
computer-assisted total knee replacements and analysed the clinical
and radiological outcomes of 45 knees that had arthritis with a
pre-operative recurvatum deformity. The mean pre-operative hyperextension
deformity of 11° (6° to 15°), as measured by navigation at the start
of the operation, improved to a mean flexion deformity of 3.1° (0°
to 7°) post-operatively. A total of 41 knees (91%) were managed
using inserts ≤ 12.5 mm thick, and none had mediolateral laxity
>
2 mm from a mechanical axis of 0° at the end of the surgery. At
a mean follow-up of 26.4 months (13 to 48) there was significant
improvement in the mean Knee Society, Oxford knee and Western Ontario
and McMaster Universities Osteoarthritis Index scores compared with
the pre-operative values. The mean knee flexion improved from 105°
(80° to 125°) pre-operatively to 131° (120° to 145°), and none of
the limbs had recurrent recurvatum. These early results show that total knee replacement using computer
navigation and an algorithmic approach for arthritic knees with
a recurvatum deformity can give excellent radiological and functional
outcomes without recurrent deformity.
Platelet-derived growth factor (PDGF) is known
to stimulate osteoblast or osteoprogenitor cell activity. We investigated
the effect of locally applied PDGF from poly- These results indicate that local application of PDGF from biodegradable
PDLLA-coated implants significantly accelerates fracture healing
in experimental animals. Further development may help fracture healing
in the clinical situation.
The purpose of this study was to identify factors
that predict implant cut-out after cephalomedullary nailing of intertrochanteric
and subtrochanteric hip fractures, and to test the significance
of calcar referenced tip-apex distance (CalTAD) as a predictor for
cut-out. We retrospectively reviewed 170 consecutive fractures that had
undergone cephalomedullary nailing. Of these, 77 met the inclusion
criteria of a non-pathological fracture with a minimum of 80 days
radiological follow-up (mean 408 days; 81 days to 4.9 years). The
overall cut-out rate was 13% (10/77). The significant parameters in the univariate analysis were tip-apex
distance (TAD) (p <
0.001), CalTAD (p = 0.001), cervical angle
difference (p = 0.004), and lag screw placement in the anteroposterior
(AP) view (Parker’s ratio index) (p = 0.003). Non-significant parameters
were age (p = 0.325), gender (p = 1.000), fracture side (p = 0.507),
fracture type (AO classification) (p = 0.381), Singh Osteoporosis
Index (p = 0.575), lag screw placement in the lateral view (p =
0.123), and reduction quality (modified Baumgaertner’s method) (p = 0.575).
In the multivariate analysis, CalTAD was the only significant measurement
(p = 0.001). CalTAD had almost perfect inter-observer reliability
(interclass correlation coefficient (ICC) 0.901). Our data provide the first reported clinical evidence that CalTAD
is a predictor of cut-out. The finding of CalTAD as the only significant
parameter in the multivariate analysis, along with the univariate
significance of Parker’s ratio index in the AP view, suggest that
inferior placement of the lag screw is preferable to reduce the
rate of cut-out. Cite this article:
The June 2014 Oncology Roundup360 looks at: Infection still a problem in endoprosthetic reconstruction; massive allografts not as successful as we perhaps think; curopsy for aneurysmal bone cysts?; lengthening prosthesis: days are numbered; new WHO classification in brief; proximal tumours and fluid levels: bad news; infection is predictable in orthopaedic oncology; psychosocial support key in oncological outcomes.
In this randomised controlled trial, we evaluated
the role of elastic compression using ankle injury stockings (AIS)
in the management of fractures of the ankle. A total of 90 patients
with a mean age of 47 years (16 to 79) were treated within 72 hours
of presentation with a fracture of the ankle, 31 of whom were treated
operatively and 59 conservatively, were randomised to be treated
either with compression by AIS plus an Aircast boot or Tubigrip
plus an Aircast boot. Male to female ratio was 36:54. The primary
outcome measure was the functional Olerud–Molander ankle score (OMAS).
The secondary outcome measures were; the American Orthopaedic Foot
and Ankle Society score (AOFAS); the Short Form (SF)-12v2 Quality
of Life score; and the frequency of deep vein thrombosis (DVT). Compression using AIS reduced swelling of the ankle at all time
points and improved the mean OMAS score at six months to 98 (95%
confidence interval (CI) 96 to 99) compared with a mean of 67 (95%
CI 62 to 73) for the Tubigrip group (p <
0.001). The mean AOFAS
and SF-12v2 scores at six months were also significantly improved
by compression. Of 86 patients with duplex imaging at four weeks,
five (12%) of 43 in the AIS group and ten (23%) of 43 in the Tubigrip
group developed a DVT (p = 0.26). Compression improved functional outcome and quality of life following
fracture of the ankle. DVTs were frequent, but a larger study would
be needed to confirm that compression with AISs reduces the incidence
of DVT. Cite this article:
The June 2014 Wrist &
Hand Roundup360 looks at: aart throwing not quite as we thought; two-gear, four-bar linkage in the wrist?; assessing outcomes in distal radial fractures; gold standard Swanson’s?; multistrand repairs of unclear benefit in flexor tendon release; for goodness’ sake, leave the thumb alone in scaphoid fractures; horizons in carpal tunnel surgery; treading the Essex-Lopresti tightrope; wrist replacement in trauma? and radial shortening reliable in the long term for Kienbock’s disease
The August 2012 Trauma Roundup360 looks at: pelvic fractures, thromboembolism and the Japanese; venous thromboembolism risk after pelvic and acetabular fractures; the displaced clavicular fracture; whether to use a nail or plate for the displaced fracture of the distal tibia; the dangers of snowboarding; how to predict the outcome of lower leg blast injuries; compressive external fixation for the displaced patellar fracture; broken hips in Morocco; and spinal trauma in mainland China.
The April 2014 Shoulder &
Elbow Roundup360 looks at: arthroscopic capsular release successful after six months; MCIC in cuff surgery; analgesia following arthroscopic cuff repair; platelet-rich fibrin; and cuff tear and suprascapular nerve neuropathy?
It is important to be able to identify patients
with an increased risk of venous thromboembolism (VTE) in order
to minimise the risk of an event. We investigated the incidence
and risk factors for post-operative VTE in 168 consecutive patients
with a malignancy of the lower limb. The period of study included
ten months before and 12 months after the introduction of chemical
thromboprophylaxis. All data about the potential risk factors were identified
and classified into three groups (patient-, surgery- and tumour-related).
The outcome measure was a thromboembolic event within 90 days of
surgery. Of the 168 patients, eight (4.8%) had a confirmed symptomatic
deep-vein thrombosis and one (0.6%) a fatal pulmonary embolism.
Of the 28 variables tested, age >
60 years, higher American Society
of Anesthesiologists grade and metastatic tumour were independent
risk factors for VTE. The overall rate of symptomatic VTE was not significantly
different between patients who received chemical thromboprophylaxis
and those who did not. Knowledge of these risk factors may be of
value in improving the surgical outcome of patients with a malignancy
of the lower limb. Cite this article:
In a global environment of rising costs and limited funds, robotic and computer-assisted orthopaedic technologies could provide the means to drive a necessary revolution in arthroplasty productivity. Robots have been used to operate on humans for 20 years, but the adoption of the technology has lagged behind that of the manufacturing industry. The use of robots in surgery should enable cost savings by reducing instrumentation and inventories, and improving accuracy. Despite these benefits, the orthopaedic community has been resistant to change. If the ergonomics and economics are right, robotic technology just might transform the provision of joint replacement.
We systematically reviewed all the evidence published
in the English language on proximal interphalangeal joint (PIPJ)
replacement, to determine its effectiveness on the function of the
hand and the associated post-operative complications. Original studies were selected if they reported clinical outcome
with a minimum of one year’s follow-up. Quality was assessed using
the Cowley systematic review criteria modified for finger-joint
replacements. Of 319 articles identified, only five were adequately
reported according to our quality criteria; there were no randomised
controlled trials. PIPJ replacements had a substantial effect size
on hand pain of -23.2 (95% confidence interval (CI) -27.3 to -19.1)
and grip strength 1.2 (95% CI -10.7 to 13.1), and a small effect
on range of movement 0.2 (95% CI -0.4 to 0.8). A dorsal approach
was most successful. Post-operative loosening occurred in 10% (95%
CI 3 to 30) of ceramic and 12.5% (95% CI 7 to 21) of pyrocarbon
replacements. Post-operative complications occurred in 27.8% (95%
CI 20 to 37). We conclude that the effectiveness of PIPJ replacement has not
been established. Small observational case studies and short-term
follow-up, together with insufficient reporting of patient data,
functional outcomes and complications, limit the value of current
evidence. We recommend that a defined core set of patients, surgical and
outcome data for this intervention be routinely and systematically
collected within the framework of a joint registry.
The April 2014 Research Roundup360 looks at: scientific writing needed in orthopaedic papers; antiseptics and osteoblasts; thromboembolic management in orthopaedic patients; nicotine and obesity in post-operative complications; defining the “Patient Acceptable Symptom State”; and cheap and nasty implants of poor quality.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.