Screw plugs have been reported to increase the fatigue strength of stainless steel locking plates. The objective of this study was to examine and compare this effect between stainless steel and titanium locking plates. Custom-designed locking plates with identical structures were fabricated from stainless steel and a titanium alloy. Three types of plates were compared: type I unplugged plates; type II plugged plates with a 4 Nm torque; and type III plugged plates with a 12 Nm torque. The stiffness, yield strength, and fatigue strength of the plates were investigated through a four-point bending test. Failure analyses were performed subsequently.Objectives
Methods
The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a large femoral head (38 mm to 52 mm) and the other with
a conventional femoral head (28 mm or 32 mm). We compared clinical
outcome, blood metal ion levels, and the incidence of pseudotumour in
the two groups. Between December 2009 and December 2011, 62 patients underwent
MoM THA with a large femoral head (Magnum group) and 57 patients
an MoM THA with a conventional femoral head (conventional group).
Clinical outcome was assessed using the Harris Hip score, University
of California, Los Angeles (UCLA) activity score and EuroQol-5D
(EQ-5D). Blood metal ion levels were measured and MRI scans were
analyzed at a minimum of five years postoperatively.Aims
Patients and Methods
Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (Aims
Patients and Methods
Aims
Patients and Methods
The use of a noninvasive growing endoprosthesis in the management
of primary bone tumours in children is well established. However,
the efficacy of such a prosthesis in those requiring a revision
procedure has yet to be established. The aim of this series was
to present our results using extendable prostheses for the revision
of previous endoprostheses. All patients who had a noninvasive growing endoprosthesis inserted
at the time of a revision procedure were identified from our database.
A total of 21 patients (seven female patients, 14 male) with a mean
age of 20.4 years (10 to 41) at the time of revision were included.
The indications for revision were mechanical failure, trauma or infection
with a residual leg-length discrepancy. The mean follow-up was 70
months (17 to 128). The mean shortening prior to revision was 44 mm
(10 to 100). Lengthening was performed in all but one patient with
a mean lengthening of 51 mm (5 to 140).Aims
Patients and Methods
Opening wedge high tibial osteotomy (HTO) is an established surgical procedure for the treatment of early-stage knee arthritis. Other than infection, the majority of complications are related to mechanical factors – in particular, stimulation of healing at the osteotomy site. This study used finite element (FE) analysis to investigate the effect of plate design and bridging span on interfragmentary movement (IFM) and the influence of fracture healing on plate stress and potential failure. A 10° opening wedge HTO was created in a composite tibia. Imaging and strain gauge data were used to create and validate FE models. Models of an intact tibia and a tibia implanted with a custom HTO plate using two different bridging spans were validated against experimental data. Physiological muscle forces and different stages of osteotomy gap healing simulating up to six weeks postoperatively were then incorporated. Predictions of plate stress and IFM for the custom plate were compared against predictions for an industry standard plate (TomoFix).Objectives
Materials and Methods
The aim of this study was to report the long-term outcome and
implant survival of the lateral resurfacing elbow (LRE) arthroplasty
in the treatment of elbow arthritis. We reviewed a consecutive series of 27 patients (30 elbows) who
underwent LRE arthroplasty between December 2005 and January 2008.
There were 15 women and 12 men, with a mean age of 61 years (25
to 82). The diagnosis was primary hypotrophic osteoarthritis (OA)
in 12 patients (14 elbows), post-traumatic osteoarthritis (PTOA)
in five (five elbows) and rheumatoid arthritis (RA) in ten patients
(11 elbows). The mean clinical outcome scores including the Mayo
Elbow Performance Score (MEPS), the American Shoulder and Elbow
Surgeons elbow score (ASES-e), the mean range of movement and the
radiological outcome were recorded at three, six and 12 months and
at a mean final follow-up of 8.3 years (7.3 to 9.4). A one sample Aim
Patients and Methods
Previous studies have suggested that metal-on-metal (MoM) Pinnacle (DePuy Synthes, Warsaw, Indiana) hip arthroplasties implanted after 2006 exhibit higher failure rates. This was attributed to the production of implants with reduced diametrical clearances between their bearing surfaces, which, it was speculated, were outside manufacturing tolerances. This study aimed to better understand the performance of Pinnacle Systems manufactured before and after this event. A total of 92 retrieved MoM Pinnacle hips were analyzed, of which 45 were implanted before 2007, and 47 from 2007 onwards. The ‘pre-2007’ group contained 45 implants retrieved from 21 male and 24 female patients, with a median age of 61.3 years (interquartile range (IQR) 57.1 to 65.5); the ‘2007 onwards’ group contained 47 implants retrieved from 19 male and 28 female patients, with a median age of 61.8 years (IQR 58.5 to 67.8). The volume of material lost from their bearing and taper surfaces was measured using coordinate and roundness measuring machines. These outcomes were then compared statistically using linear regression models, adjusting for potentially confounding factors.Objectives
Methods
To report our experience with trunnion corrosion following metal-on-polyethylene
total hip arthroplasty, in particular to report the spectrum of
presentation and determine the mean time to presentation. We report the presenting symptoms and signs, intraoperative findings,
and early results and complications of operative treatment in nine
patients with a mean age of 74 years (60 to 86). The onset of symptoms
was at a mean of seven years (3 to 18) after index surgery.Aims
Patients and Methods
It has been suggested that mobile-bearing total knee arthroplasty
(TKA) might lead to better outcomes by accommodating some femorotibial
rotational mismatch, thereby reducing contact stresses and polyethylene
wear. The aim of this study was to determine whether there is a
difference between fixed- and mobile-bearing versions of a contemporary
TKA with respect to durability, range of movement (ROM) and function,
ten years postoperatively. A total of 240 patients who were enrolled in this randomized
controlled trial (RCT) underwent a primary cemented TKA with one
of three tibial components (all-polyethylene fixed-bearing, modular
metal-backed fixed-bearing and mobile-bearing). Patients were reviewed
at a median follow-up of ten years (IQR 9.2 to 10.4).Aims
Patients and Methods
It is important to consider sagittal pelvic rotation when introducing
the acetabular component at total hip arthroplasty (THA). The purpose
of this study was to identify patients who are at risk of unfavourable
pelvic mobility, which could result in poor outcomes after THA. A consecutive series of 4042 patients undergoing THA had lateral
functional radiographs and a low-dose CT scan to measure supine
pelvic tilt, pelvic incidence, standing pelvic tilt, flexed-seated
pelvic tilt, standing lumbar lordotic angle, flexed-seated lumbar
lordotic angle, and lumbar flexion. Changes in pelvic tilt from
supine-to-standing positions and supine-to-flexed-seated positions
were determined. A change in pelvic tilt of 13° between positions was
deemed unfavourable as it alters functional anteversion by 10° and
effectively places the acetabular component outside the safe zone
of orientation.Aims
Patients and Methods
The crucial role of the radial head in the stability of the elbow in terrible triad injury is acknowledged. This retrospective study aims to compare the results of resection of a severely comminuted radial head with or without prosthetic arthroplasty as part of the reconstruction for this injury. The outcome of radial head resection was compared with prosthetic arthroplasty in 29 and 15 patients with terrible triad injuries, respectively. There were ten female patients (34.5%) in the resection group and six female patients (40%) in the prosthesis group. The mean age was 40.7 years (Aims
Patients and Methods
Reconstruction of the acetabulum after resection of a periacetabular
malignancy is technically challenging and many different techniques
have been used with varying success. Our aim was to prepare a systematic
review of the literature dealing with these techniques in order
to clarify the management, the rate of complications and the outcomes. A search of PubMed and MEDLINE was conducted for English language
articles published between January 1990 and February 2017 with combinations
of key search terms to identify studies dealing with periacetabular
resection with reconstruction in patients with a malignancy. Studies
in English that reported radiographic or clinical outcomes were
included. Data collected from each study included: the number and
type of reconstructions, the pathological diagnosis of the lesions,
the mean age and follow-up, gender distribution, implant survivorship, complications,
functional outcome, and mortality. The results from individual studies
were combined for the general analysis, and then grouped according
to the type of reconstruction. Aims
Patients and Methods
Magnetically controlled growing rod (MCGR) systems use non-invasive
spinal lengthening for the surgical treatment of early-onset scoliosis
(EOS). The primary aim of this study was to evaluate the performance
of these devices in the prevention of progression of the deformity.
A secondary aim was to record the rate of complications. An observational study of 31 consecutive children with EOS, of
whom 15 were male, who were treated between December 2011 and October
2017 was undertaken. Their mean age was 7.7 years (2 to 14). The
mean follow-up was 47 months (24 to 69). Distractions were completed
using the tailgating technique. The primary outcome measure was
correction of the radiographic deformity. Secondary outcomes were
growth, functional outcomes and complication rates.Aims
Patients and Methods
The primary aim of this study was to define the standard minimum
follow-up required to produce a reliable estimate of the rate of
re-operation after radial head arthroplasty (RHA). The secondary
objective was to define the leading reasons for re-operation. Four electronic databases, between January 2000 and March 2017
were searched. Articles reporting reasons for re-operation (Group
I) and results (Group II) after RHA were included. In Group I, a
meta-analysis was performed to obtain the standard minimum follow-up,
the mean time to re-operation and the reason for failure. In Group
II, the minimum follow-up for each study was compared with the standard
minimum follow-up.Aims
Materials and Methods
This study reports the outcomes of a technique of soft-tissue coverage and Chopart amputation for severe crush injuries of the forefoot. Between January 2012 to December 2016, 12 patients (nine male; three female, mean age 38.58 years; 26 to 55) with severe foot crush injury underwent treatment in our institute. All patients were followed-up for at least one year. Their medical records, imaging, visual analogue scale score, walking ability, complications, and functional outcomes one year postoperatively based on the American Orthopedic Foot and Ankle Society (AOFAS) and 36-Item Short-Form Health Survey (SF-36) scores were reviewed.Aims
Patients and Methods