This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83). Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem. The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4). This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.
Dislocation is one of the most common causes
of patient and surgeon dissatisfaction following hip replacement
and to treat it, the causes must first be understood. Patient factors
include age greater than 70 years, medical comorbidities, female
gender, ligamentous laxity, revision surgery, issues with the abductors,
and patient education. Surgeon factors include the annual quantity
of procedures and experience, the surgical approach, adequate restoration
of femoral offset and leg length, component position, and soft-tissue
or bony impingement. Implant factors include the design of the head
and neck region, and so-called skirts on longer neck lengths. There
should be offset choices available in order to restore soft-tissue
tension. Lipped liners aid in gaining stability, yet if improperly placed
may result in impingement and dislocation. Late dislocation may
result from polyethylene wear, soft-tissue destruction, trochanteric
or abductor disruption and weakness, or infection. Understanding
the causes of hip dislocation facilitates prevention in a majority
of instances. Proper pre-operative planning includes the identification
of patients with a high offset in whom inadequate restoration of
offset will reduce soft-tissue tension and abductor efficiency.
Component position must be accurate to achieve stability without impingement.
Finally, patient education cannot be over-emphasised, as most dislocations occur
early, and are preventable with proper instructions. Cite this article:
This systematic review of the literature summarises
the clinical experience with ceramic-on-ceramic hip bearings over
the past 40 years and discusses the concerns that exist in relation
to the bearing combination. Loosening, fracture, liner chipping
on insertion, liner canting and dissociation, edge-loading and squeaking
have all been reported, and the relationship between these issues
and implant design and surgical technique is investigated. New design
concepts are introduced and analysed with respect to previous clinical
experience.
The Unified Classification System (UCS) emphasises
the key principles in the assessment and management of peri-prosthetic
fractures complicating partial or total joint replacement. We tested the inter- and intra-observer agreement for the UCS
as applied to the pelvis and femur using 20 examples of peri-prosthetic
fracture in 17 patients. Each subtype of the UCS was represented
by at least one case. Specialist orthopaedic surgeons (experts)
and orthopaedic residents (pre-experts) assessed reliability on
two separate occasions. For the pelvis, the UCS showed inter-observer agreement of 0.837
(95% confidence intervals (CI) 0.798 to 0.876) for the experts and
0.728 (95% CI 0.689 to 0.767) for the pre-experts. The intra-observer
agreement for the experts was 0.861 (95% CI 0.760 to 0.963) and
0.803 (95% 0.688 to 0.918) for the pre-experts. For the femur, the
UCS showed an inter-observer kappa value of 0.805 (95% CI 0.765
to 0.845) for the experts and a value of 0.732 (95% CI 0.690 to 0.773)
for the pre-experts. The intra-observer agreement was 0.920 (95%
CI 0.867 to 0.973) for the experts, and 0.772 (95% CI 0.652 to 0.892)
for the pre-experts. This corresponds to a substantial and ‘almost
perfect’ inter- and intra-observer agreement for the UCS for peri-prosthetic
fractures of the pelvis and femur. We hope that unifying the terminology of these injuries will
assist in their assessment, treatment and outcome. Cite this article:
The August 2014 Hip &
Pelvis Roundup360 looks at: Serial MRIs best for pseudotumour surveillance; Is ultrasound good enough for MOM follow-up?; Does weight loss in obese patients help?; Measuring acetabular anteversion on plain films; Two-stage one-stage too many in fungal hip revisions? and 35 is the magic number in arthroplasty.
Despite a lack of long-term follow-up, there
is an increasing trend towards using femoral heads of large diameter
in total hip replacement (THR), partly because of the perceived
advantage of lower rates of dislocation. However, increasing the
size of the femoral head is not the only way to reduce the rate
of dislocation; optimal alignment of the components and repair of
the posterior capsule could achieve a similar effect. In this prospective study of 512 cemented unilateral THRs (Male:Female
230:282) performed between 2004 and 2011, we aimed to determine
the rate of dislocation in patients who received a 22 mm head on
a 9/10 Morse taper through a posterior approach with capsular repair
and using the transverse acetabular ligament (TAL) as a guide for the
alignment of the acetabular component. The mean age of the patients
at operation was 67 years (35 to 89). The mean follow-up was 2.8
years (0.5 to 6.6). Pre- and post-operative assessment included
Oxford hip, Short Form-12 and modified University of California
Los Angeles and Merle D’Aubigne scores. The angles of inclination
and anteversion of the acetabular components were measured using
radiological software. There were four dislocations (0.78%), all
of which were anterior. In conclusion, THR with a 22 mm diameter head performed through
a posterior approach with capsular repair and using the TAL as a
guide for the alignment of the acetabular component was associated
with a low rate of dislocation. Cite this article:
Metal-on-metal hip arthroplasty gained significant
favor in the first decade of the millennium. However, the past several
years have seen increasing reports of failure, pseudotumor and other
adverse reactions. This study presents the results of a single center’s
15-year experience with metal-on-metal total hip arthroplasty as
strong evidence that metal-on-metal is going, going, gone.
We conducted a retrospective study to assess
the prevalence of adverse reactions to metal debris (ARMD) in patients
operated on at our institution with metal-on-metal (MoM) total hip
replacements with 36 mm heads using a Pinnacle acetabular shell.
A total of 326 patients (150 males, 175 hips; 176 females, 203 hips)
with a mean age of 62.7 years (28 to 85) and mean follow-up of 7.5
years (0.1 to 10.8) participating in our in-depth modern MoM follow-up
programme were included in the study, which involved recording whole
blood cobalt and chromium ion measurements, Oxford hip scores (OHS)
and plain radiographs of the hip and targeted cross-sectional imaging. Elevated
blood metal ion levels (>
5 parts per billion) were seen in 32 (16.1%)
of the 199 patients who underwent unilateral replacement. At 23
months after the start of our modern MoM follow-up programme, 29
new cases of ARMD had been revealed. Hence, the nine-year survival
of this cohort declined from 96% (95% CI 95 to 98) with the old
surveillance routine to 86% (95% CI 82 to 90) following the new
protocol. Although ARMD may not be as common in 36 mm MoM THRs as
in those with larger heads, these results support the Medicines
and Healthcare Products Regulatory Agency guidelines on regular
reviews and further investigations, and emphasise the need for specific
a follow-up programme for patients with MoM THRs. Cite this article:
The August 2012 Hip &
Pelvis Roundup360 looks at: whether cemented hip replacement might be bad for your health; highly cross-linked polyethylene; iHOT-33 - a new hip outcome measure; hamstring injuries; total hip replacement; stemmed metal-on-metal THR; bipolar hemiarthroplasty, neuromuscular disease and dislocation; the high risk of secondary hemiarthroplasty; and whether we have to repair the labrum after all?
Revision knee arthroplasty presents a number
of challenges, not least of which is obtaining solid primary fixation
of implants into host bone. Three anatomical zones exist within
both femur and tibia which can be used to support revision implants.
These consist of the joint surface or epiphysis, the metaphysis
and the diaphysis. The methods by which fixation in each zone can
be obtained are discussed. The authors suggest that solid fixation
should be obtained in at least two of the three zones and emphasise
the importance of pre-operative planning and implant selection. Cite this article:
The February 2014 Hip &
Pelvis Roundup360 looks at: length of stay; cementless metaphyseal fixation; mortality trends in over 400,000 total hip replacements; antibiotics in hip fracture surgery; blood supply to the femoral head after dislocation; resurfacing and THR in metal-on-metal replacement; diabetes and hip replacement; bone remodelling over two decades following hip replacement; and whether bisphosphonates affect acetabular fixation.
The issues surrounding raised levels of metal
ions in the blood following large head metal-on-metal total hip replacement
(THR), such as cobalt and chromium, have been well documented. Despite
the national popularity of uncemented metal-on-polyethylene (MoP)
THR using a large-diameter femoral head, few papers have reported
the levels of metal ions in the blood following this combination.
Following an isolated failure of a 44 mm Trident–Accolade uncemented
THR associated with severe wear between the femoral head and the
trunnion in the presence of markedly elevated levels of cobalt ions
in the blood, we investigated the relationship between modular femoral head
diameter and the levels of cobalt and chromium ions in the blood
following this THR. A total of 69 patients received an uncemented Trident–Accolade
MoP THR in 2009. Of these, 43 patients (23 men and 20 women, mean
age 67.0 years) were recruited and had levels of cobalt and chromium
ions in the blood measured between May and June 2012. The patients
were then divided into three groups according to the diameter of
the femoral head used: 12 patients in the 28 mm group (controls),
18 patients in the 36 mm group and 13 patients in the 40 mm group.
A total of four patients had identical bilateral prostheses in situ
at phlebotomy: one each in the 28 mm and 36 mm groups and two in
the 40 mm group. There was a significant increase in the mean levels of cobalt
ions in the blood in those with a 36 mm diameter femoral head compared
with those with a 28 mm diameter head (p = 0.013). The levels of
cobalt ions in the blood were raised in those with a 40 mm diameter
head but there was no statistically significant difference between
this group and the control group (p = 0.152). The levels of chromium
ions in the blood were normal in all patients. The clinical significance of this finding is unclear, but we
have stopped using femoral heads with a diameter of ≤ 36 mm, and
await further larger studies to clarify whether, for instance, this
issue particularly affects this combination of components. Cite this article:
We present the clinical results and survivorship
of consecutive 100 Birmingham Hip Resurfacings in 90 patients at
a minimum follow-up of ten years. All procedures were carried out
by an independent surgeon who commenced a prospective study in 1998.
Patients were assessed clinically using the Western Ontario and
McMaster Universities osteoarthritis index, Short-Form 36, Harris
hip score and University of California, Los Angeles activity score. Radiological
analysis was performed by an independent observer and blood metal
ion levels concentrations were measured at ten years post-operatively
in 62 patients. The median acetabular component inclination was
46.2° (34° to 59°) and anteversion 11.0° (0° to 30°). The median
chromium concentration in the unilateral group was 1.74 (0.41 to
15.23) and for the bilateral group was 2.98 (1.57 to 18.01). The
equivalent values for cobalt were 1.67 (0.54 to 20.4) and 1.88 (1.38
to 19.32). In total there were eight failures giving an overall survival
at ten years of 92% (95% confidence interval (CI) 86.7 to 97.3).
BHR in male patients had an improved survivorship of 94.6% (95%
CI 89.4 to 100) compared with females at 84.6% (95% CI 70.7 to 98.5),
but this did not reach statistical significance (p = 0.119). Four
of the nine BHRs with a bearing diameter of 42 mm failed. The overall results were consistent with data produced from other
centres in that the clinical outcome of large male patients was
extremely encouraging, whereas the survival of the smaller joints
was less satisfactory.
The femoral head receives blood supply mainly
from the deep branch of the medial femoral circumflex artery (MFCA).
In previous studies we have performed anatomical dissections of
16 specimens and subsequently visualised the arteries supplying
the femoral head in
55 healthy individuals. In this further radiological study we compared
the arterial supply of the femoral head in 35 patients (34 men and
one woman, mean age 37.1 years (16 to 64)) with a fracture/dislocation
of the hip with a historical control group of 55 hips. Using CT
angiography, we identified the three main arteries supplying the femoral
head: the deep branch and the postero-inferior nutrient artery both
arising from the MFCA, and the piriformis branch of the inferior
gluteal artery. It was possible to visualise changes in blood flow
after fracture/dislocation. Our results suggest that blood flow is present after reduction
of the dislocated hip. The deep branch of the MFCA was patent and
contrast-enhanced in 32 patients, and the diameter of this branch
was significantly larger in the fracture/dislocation group than
in the control group (p = 0.022). In a subgroup of ten patients
with avascular necrosis (AVN) of the femoral head, we found a contrast-enhanced
deep branch of the MFCA in eight hips. Two patients with no blood
flow in any of the three main arteries supplying the femoral head
developed AVN. Cite this article:
The February 2014 Research Roundup360 looks at: blood supply to the femoral head after dislocation; diabetes and hip replacement; bone remodelling over two decades following hip replacement; sham surgery as good as arthroscopic meniscectomy; distraction in knee osteoarthritis; whether joint replacement prevent cardiac events; tranexamic acid and knee replacement haemostasis; cartilage colonisation in bipolar ankle grafts; CTs and proof of fusion; atorvastatin for muscle re-innervation after sciatic nerve transection; microfracture and short-term pain in cuff repair; promising early results from L-PRF augmented cuff repairs; and fatty degeneration in a rodent model.
The spiral blade modification of the Dynamic
Hip Screw (DHS) was designed for superior biomechanical fixation
in the osteoporotic femoral head. Our objective was to compare clinical
outcomes and in particular the incidence of loss of fixation. In a series of 197 consecutive patients over the age of 50 years
treated with DHS-blades (blades) and 242 patients treated with conventional
DHS (screw) for AO/OTA 31.A1 or A2 intertrochanteric fractures were
identified from a prospectively compiled database in a level 1 trauma
centre. Using propensity score matching, two groups comprising 177
matched patients were compiled and radiological and clinical outcomes
compared. In each group there were 66 males and 111 females. Mean
age was 83.6 (54 to 100) for the conventional DHS group and 83.8
(52 to 101) for the blade group. Loss of fixation occurred in two blades and 13 DHSs. None of
the blades had observable migration while nine DHSs had gross migration
within the femoral head before the fracture healed. There were two
versus four implant cut-outs respectively and one side plate pull-out
in the DHS group. There was no significant difference in mortality
and eventual walking ability between the groups. Multiple logistic
regression suggested that poor reduction (odds ratio (OR) 11.49,
95% confidence intervals (CI) 1.45 to 90.9, p = 0.021) and fixation
by DHS (OR 15.85, 95%CI 2.50 to 100.3, p = 0.003) were independent
predictors of loss of fixation. The spiral blade design may decrease the risk of implant migration
in the femoral head but does not reduce the incidence of cut-out
and reoperation. Reduction of the fracture is of paramount importance
since poor reduction was an independent predictor for loss of fixation
regardless of the implant being used. Cite this article:
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
Ceramic-on-ceramic bearings in hip replacement have low rates of wear and are increasingly being used in young adults. Our aim was to determine the incidence of audible phenomena or other bearing-related complications. We retrospectively analysed 250 ceramic-on-ceramic hip replacements in 224 patients which had been implanted between April 2000 and December 2007. The mean age of the patients at operation was 44 years (14 to 83) and all the operations were performed using the same surgical technique at a single centre. At a mean follow-up of 59 months (24 to 94), the mean Oxford hip score was 40.89 (11 to 48). There were six revisions, three of which were for impingement-related complications. No patient reported squeaking, but six described grinding or clicking, which was usually associated with deep flexion. No radiological evidence of osteolysis or migration of the components was observed in any hip. The early to mid-term results of contemporary ceramic-on-ceramic hip replacement show promising results with few concerns in terms of noise and squeaking. Positioning of the acetabular component remains critical in regard to the reduction of other impingement-related complications.
Metal-on-metal resurfacing of the hip (MoMHR)
has enjoyed a resurgence in the last decade, but is now again in question
as a routine option for osteoarthritis of the hip. Proponents of
hip resurfacing suggest that its survival is superior to that of
conventional hip replacement (THR), and that hip resurfacing is
less invasive, is easier to revise than THR, and provides superior
functional outcomes. Our argument serves to illustrate that none
of these proposed advantages have been realised and new and unanticipated
serious complications, such as pseudotumors, have been associated
with the procedure. As such, we feel that the routine use of MoMHR
is not justified. Cite this article:
The June 2013 Hip &
Pelvis Roundup360 looks at: failure in metal-on-metal arthroplasty; minimal hip approaches; whether bisphosphonates improve femoral bone stock following arthroplasty; whether more fat means more operative time; surgical infection; vascularised fibular graft for osteonecrosis; subclinical SUFE; and dentists, hips and antibiotics.