Aims. To analyse the effectiveness of debridement and implant retention
(DAIR) in patients with hip
The aim of this study was to assess the role
of synovial C-reactive protein (CRP) in the diagnosis of chronic periprosthetic
hip infection. We prospectively collected synovial fluid from 89
patients undergoing revision hip arthroplasty and measured synovial
CRP, serum CRP, erythrocyte sedimentation rate (ESR), synovial white
blood cell (WBC) count and synovial percentages of polymorphonuclear
neutrophils (PMN). Patients were classified as septic or aseptic
by means of clinical, microbiological, serum and synovial fluid
findings. The high viscosity of the synovial fluid precluded the
analyses in nine patients permitting the results in 80 patients
to be studied. There was a significant difference in synovial CRP
levels between the septic (n = 21) and the aseptic (n = 59) cohort.
According to the receiver operating characteristic curve, a synovial
CRP threshold of 2.5 mg/l had a sensitivity of 95.5% and specificity
of 93.3%. The area under the curve was 0.96. Compared with serum
CRP and ESR, synovial CRP showed a high diagnostic value. According
to these preliminary results, synovial CRP may be a useful parameter
in diagnosing chronic
Total knee arthroplasty (TKA) is known to lead
to a reduction in
Aims. Infection following total hip or knee arthroplasty is a serious
complication. We noted an increase in post-operative infection in
cases carried out in temporary operating theatres. We therefore
compared those cases performed in standard and temporary operating
theatres and examined the deep
Aims. Advocates of debridement, antibiotics and implant retention (DAIR)
in hip
The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless polished taper slip stem (PTS) for management of an intracapsular hip fracture. This retrospective study included hip fracture patients aged 50 years and older treated with Exeter (PTS) bipolar hemiarthroplasty between 2019 and 2022. Patient demographics, place of domicile, fracture type, delirium status, American Society of Anesthesiologists (ASA) grade, length of stay, and mortality were collected. Reoperation and mortality were recorded up to a median follow-up of 29.5 months (interquartile range 12 to 51.4). Cox regression was performed to evaluate independent risk factors associated with reoperation and mortality.Aims
Methods
It is unclear whether there is a limit to the amount of distal bone required to support fixation of supracondylar periprosthetic femoral fractures. This retrospective multicentre study evaluated lateral locked plating of
The aim of this study was to determine whether fixation, as opposed to revision arthroplasty, can be safely used to treat reducible Vancouver B type fractures in association with a cemented collarless polished tapered femoral stem (the Exeter). This retrospective cohort study assessed 152 operatively managed consecutive unilateral Vancouver B fractures involving Exeter stems; 130 were managed with open reduction and internal fixation (ORIF) and 22 with revision arthroplasty. Mean follow-up was 6.5 years (SD 2.6; 3.2 to 12.1). The primary outcome measure was revision of at least one component. Kaplan–Meier survival analysis was performed. Regression analysis was used to identify risk factors for revision following ORIF. Secondary outcomes included any reoperation, complications, blood transfusion, length of hospital stay, and mortality.Aims
Methods
Aims.
Infection is a leading indication for revision
arthroplasty. Established criteria used to diagnose prosthetic joint infection
(PJI) include a range of laboratory tests. Leucocyte esterase (LE)
is widely used on a colorimetric reagent strip for the diagnosis
of urinary tract infections. This inexpensive test may be used for
the diagnosis or exclusion of PJI. Aspirates from 30 total hip arthroplasties
(THAs) and 79 knee arthroplasties (KA) were analysed for LE activity. Semi-quantitative
reagent strip readings of 15, 70, 125 and 500 white blood cells
(WBC) were validated against a manual synovial white cell count
(WCC). A receiver operating characteristic (ROC) curve was constructed
to determine the optimal cut-off point for the semi-quantitative
results. Based on established criteria, six THAs and 15 KAs were
classified as infected. The optimal cut-off point for the diagnosis
of PJI was 97 WBC. The closest semi-quantitative reading for a positive
result was 125 WBC, achieving a sensitivity of 81% and a specificity
of 93%. The positive and negative predictive values of the LE test
strip were 74% and 95% respectively. The LE reagent strip had a high specificity and negative predictive
value. A negative result may exclude PJI and negate the need for
further diagnostic tests. Cite this article:
Surgeons remain concerned that ceramic hip prostheses may fail catastrophically if either the head or the liner is fractured. We report two patients, each with a ceramic-on-ceramic total hip replacement who sustained high-energy trauma sufficient to cause a displaced
The use of plate-and-cable constructs to treat
The Unified Classification System (UCS) was introduced
because of a growing need to have a standardised universal classification
system of
In a prospective study of 14 patients undergoing total hip replacement we have used dual-energy X-ray absorptiometry (DEXA) to investigate remodelling of the bone around two different designs of cementless femoral prosthesis. The bone mineral density (BMD) was measured at 12-weekly intervals for a year. Eight patients (group A) had a stiff, collarless implant and six (group B) a flexible isoelastic implant. Patients in group A showed a decrease in BMD from 14 weeks after operation. By 12 months, the mean loss in BMD was 27%, both medially and laterally to the proximal part of the implant. Those in group B showed an overall increase in BMD which reached a mean of 12.6% on the lateral side of the distal portion of the implant. Our results support the current concepts of the effects of stem stiffness and flexibility on
We report our experience of staged revision surgery
for the treatment of infected total elbow arthroplasty (TEA). Between
1998 and 2010 a consecutive series of 33 patients (34 TEAs) underwent
a first-stage procedure with the intention to proceed to second-stage
procedure when the infection had been controlled. A single first-stage procedure
with removal of the components and cement was undertaken for 29
TEAs (85%), followed by the insertion of antibiotic-impregnated
cement beads, and five (15%) required two or more first-stage procedures.
The most common organism isolated was coagulase-negative A second-stage procedure was performed for 26 TEAs (76%); seven
patients (seven TEAs, 21%) had a functional resection arthroplasty
with antibiotic beads There were three recurrent infections (11.5%) in those patients
who underwent a second-stage procedure. The infection presented
at a mean of eight months (5 to 10) post-operatively. The mean Mayo
Elbow Performance Score (MEPS) in those who underwent a second stage
revision without recurrent infection was 81.1 (65 to 95). Staged revision surgery is successful in the treatment of patients
with an infected TEA and is associated with a low rate of recurrent
infection. However, when infection does occur, this study would
suggest that it becomes apparent within ten months of the second
stage procedure. Cite this article:
The World Health Organization (WHO) and the Centre
for Disease Control and Prevention (CDC) recently published guidelines
for the prevention of surgical site infection. The WHO guidelines,
if implemented worldwide, could have an immense impact on our practices
and those of the CDC have implications for healthcare policy in
the United States. Our aim was to review the strategies for prevention of periprosthetic
joint infection in light of these and other recent guidelines. Cite this article: