Advertisement for orthosearch.org.uk
Results 141 - 160 of 180
Results per page:
The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1480 - 1483
1 Nov 2013
Hooper GJ Gilchrist N Maxwell R March R Heard A Frampton C

We studied the bone mineral density (BMD) and the bone mineral content (BMC) of the proximal tibia in patients with a well-functioning uncemented Oxford medial compartment arthroplasty using the Lunar iDXA bone densitometer. Our hypothesis was that there would be decreased BMD and BMC adjacent to the tibial base plate and increased BMD and BMC at the tip of the keel.

There were 79 consecutive patients (33 men, 46 women) with a mean age of 65 years (44 to 84) with a minimum two-year follow-up (mean 2.6 years (2.0 to 5.0)) after unilateral arthroplasty, who were scanned using a validated standard protocol where seven regions of interest (ROI) were examined and compared with the contralateral normal knee. All had well-functioning knees with a mean Oxford knee score of 43 (14 to 48) and mean Knee Society function score of 90 (20 to 100), showing a correlation with the increasing scores and higher BMC and BMD values in ROI 2 in the non-implanted knee relative to the implanted knee (p = 0.013 and p = 0.015, respectively).

The absolute and percentage changes in BMD and BMC were decreased in all ROIs in the implanted knee compared with the non-implanted knee, but this did not reach statistical significance. Bone loss was markedly less than reported losses with total knee replacement.

There was no significant association with side, although there was a tendency for the BMC to decrease with age in men. The BMC was less in the implanted side relative to the non-implanted side in men compared with women in ROI 2 (p = 0.027), ROI 3 (p = 0.049) and ROI 4 (p = 0.029).

The uncemented Oxford medial compartment arthroplasty appears to allow relative preservation of the BMC and BMD of the proximal tibia, suggesting that the implant acts more physiologically than total knee replacement. Peri-prosthetic bone loss is an important factor in assessing long-term implant stability and survival, and the results of this study are encouraging for the long-term outcome of this arthroplasty.

Cite this article: Bone Joint J 2013;95-B:1480–3.


The aims of this study were to assess the efficacy of a newly designed radiological technique (the radial groove view) for the detection of protrusion of screws in the groove for the extensor pollicis longus tendon (EPL) during plating of distal radial fractures. We also aimed to determine the optimum position of the forearm to obtain this view. We initially analysed the anatomy of the EPL groove by performing three-dimensional CT on 51 normal forearms. The mean horizontal angle of the groove was 17.8° (14° to 23°). We found that the ideal position of the fluoroscopic beam to obtain this view was 20° in the horizontal plane and 5° in the sagittal plane.

We then intra-operatively assessed the use of the radial groove view for detecting protrusion of screws in the EPL groove in 93 fractures that were treated by volar plating. A total of 13 protruding screws were detected. They were changed to shorter screws and these patients underwent CT scans of the wrist immediately post-operatively. There remained one screw that was protruding. These findings suggest that the use of the radial groove view intra-operatively is a good method of assessing the possible protrusion of screws into the groove of EPL when plating a fracture of the distal radius.

Cite this article: Bone Joint J 2013;95-B:1372–6.


Bone & Joint 360
Vol. 2, Issue 5 | Pages 27 - 29
1 Oct 2013

The October 2013 Shoulder & Elbow Roundup360 looks at: Deltoid impairment not necessarily a contra-indication for shoulder arthroplasty; The tricky radiograph; Not so asymptomatic cuff tears; Total shoulder arthroplasty: kinder on the glenoid; Barbotage for calcific tendonitis; What happens to the arthritic glenoid?; Two screws a screw too few?; Sloppy hinge best for elbow arthroplasty.


Bone & Joint 360
Vol. 2, Issue 3 | Pages 25 - 27
1 Jun 2013

The June 2013 Wrist & Hand Roundup360 looks at: whether size is a limitation; cancellous bone grafting in scaphoid nonunion; the Kienböck’s dichotomy; late displacement of the distal radius; flexor slide for finger contracture; aesthetic syndactyly; flexor tendon repair; and fixation of trapeziometacarpal cups.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 2 | Pages 185 - 189
1 Feb 2009
Pandit H Jenkins C Beard DJ Gallagher J Price AJ Dodd CAF Goodfellow JW Murray DW

We randomised 62 knees to receive either cemented or cementless versions of the Oxford unicompartmental knee replacement. The implants used in both arms of the study were similar, except that the cementless components were coated with porous titanium and hydroxyapatite. The tibial interfaces were studied with fluoroscopically-aligned radiographs.

At one year there was no difference in clinical outcome between the two groups. Narrow radiolucent lines were seen at the bone-implant interfaces in 75% of cemented tibial components. These were partial in 43%, and complete in 32%. In the cementless implants, partial radiolucencies were seen in 7% and complete radiolucencies in none. These differences are statistically significant (p < 0.0001) and imply satisfactory bone ingrowth into the cementless implants.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 10 | Pages 1344 - 1351
1 Oct 2005
Field RE Rushton N

The Cambridge Cup has been designed to replace the horseshoe-shaped articular cartilage of the acetabulum and the underlying subchondral bone. It is intended to provide physiological loading with minimal resection of healthy bone.

The cup has been used in 50 women with displaced, subcapital fractures of the neck of the femur. In 24 cases, the cup was coated with hydroxyapatite. In 26, the coating was removed before implantation in order to simulate the effect of long-term resorption.

The mean Barthel index and the Charnley-modified Merle d’Aubigné scores recovered to their levels before fracture. We reviewed 30 women at two years, 21 were asymptomatic and nine reported minimal pain. The mean scores deteriorated slightly after five years reflecting the comorbidity of advancing age. Patients with the hydroxyapatite-coated components remained asymptomatic, with no wear or loosening. The uncoated components migrated after four years and three required revision. This trial shows good early results using a novel, hydroxyapatite-coated, physiological acetabular component.


The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 147 - 156
1 Feb 2014
Rajpura A Kendoff D Board TN

We reviewed the literature on the currently available choices of bearing surface in total hip replacement (THR). We present a detailed description of the properties of articulating surfaces review the understanding of the advantages and disadvantages of existing bearing couples. Recent technological developments in the field of polyethylene and ceramics have altered the risk of fracture and the rate of wear, although the use of metal-on-metal bearings has largely fallen out of favour, owing to concerns about reactions to metal debris. As expected, all bearing surface combinations have advantages and disadvantages. A patient-based approach is recommended, balancing the risks of different options against an individual’s functional demands.

Cite this article: Bone Joint J 2014;96-B:147–56.


The Bone & Joint Journal
Vol. 95-B, Issue 7 | Pages 911 - 916
1 Jul 2013
Small SR Ritter MA Merchun JG Davis KE Rogge RD

Stress shielding resulting in diminished bone density following total knee replacement (TKR) may increase the risk of migration and loosening of the prosthesis. This retrospective study was designed to quantify the effects of the method of fixation on peri-prosthetic tibial bone density beneath cemented and uncemented tibial components of similar design and with similar long-term survival rates. Standard radiographs taken between two months and 15 years post-operatively were digitised from a matched group of TKRs using cemented (n = 67) and uncemented (n = 67) AGC tibial prostheses. Digital radiograph densitometry was used to quantify changes in bone density over time. Age, length of follow-up, gender, body mass index and alignment each significantly influenced the long-term pattern of peri-prosthetic bone density. Similar long-term changes in density irrespective of the method of fixation correlated well with the high rate of survival of this TKR at 20 years, and suggest that cemented and uncemented fixation are both equally viable.

Cite this article: Bone Joint J 2013;95-B:911–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 351 - 356
1 Mar 2009
Parratte S Argenson JA Pearce O Pauly V Auquier P Aubaniac J

We retrospectively reviewed 35 cemented unicompartmental knee replacements performed for medial unicompartmental osteoarthritis of the knee in 31 patients ≤50 years old (mean 46, 31 to 49). Patients were assessed clinically and radiologically using the Knee Society scores at a mean follow-up of 9.7 years (5 to 16) and survival at 12 years was calculated. The mean Knee Society Function Score improved from 54 points (25 to 64) pre-operatively to 89 (80 to 100) post-operatively (p < 0.0001). Six knees required revision, four for polyethylene wear treated with an isolated exchange of the tibial insert, one for aseptic loosening and one for progression of osteoarthritis.

The 12-year survival according to Kaplan-Meier was 80.6% with revision for any reason as the endpoint. Despite encouraging clinical results, polyethylene wear remains a major concern affecting the survival of unicompartmental knee replacement in patients younger than 50.


Bone & Joint 360
Vol. 2, Issue 2 | Pages 18 - 20
1 Apr 2013

The April 2013 Wrist & Hand Roundup360 looks at: whether botox is just for Hollywood; supercharging nerve repairs; YouTube research; options for Keinbock’s disease; volar plates; driving and plasters; symptomatic radial malunion; and MRI and acute scaphoid fractures.


The Bone & Joint Journal
Vol. 95-B, Issue 8 | Pages 1075 - 1082
1 Aug 2013
Choi GW Kim HJ Yeo ED Song SY

In a retrospective study we compared 32 HINTEGRA total ankle replacements (TARs) and 35 Mobility TARs performed between July 2005 and May 2010, with a minimum follow-up of two years. The mean follow-up for the HINTEGRA group was 53 months (24 to 76) and for the Mobility group was 34 months (24 to 45). All procedures were performed by a single surgeon.

There was no significant difference between the two groups with regard to the mean AOFAS score, visual analogue score for pain or range of movement of the ankle at the latest follow-up. Most radiological measurements did not differ significantly between the two groups. However, the most common grade of heterotopic ossification (HO) was grade 3 in the HINTEGRA group (10 of 13 TARs, 76.9%) and grade 2 in the Mobility group (four of seven TARs, 57.1%) (p = 0.025). Although HO was more frequent in the HINTEGRA group (40.6%) than in the Mobility group (20.0%), this was not statistically significant (p = 0.065).The difference in peri-operative complications between the two groups was not significant, but intra-operative medial malleolar fractures occurred in four (11.4%) in the Mobility group; four (12.5%) in the HINTEGRA group and one TAR (2.9%) in the Mobility group failed (p = 0.185).

Cite this article: Bone Joint J 2013;95-B:1075–82.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 750 - 756
1 Jun 2009
Mannan K Scott G

We describe the survivorship of the Medial Rotation total knee replacement (TKR) at ten years in 228 cemented primary replacements implanted between October 1994 and October 2006, with their clinical and radiological outcome. This implant has a highly congruent medial compartment, with the femoral component represented by a portion of a sphere which articulates with a matched concave surface on the medial side of the tibial insert.

There were 78 men (17 bilateral TKRs) and 111 women (22 bilateral TKRs) with a mean age of 67.9 years (28 to 90). All the patients were assessed clinically and radiologically using the American Knee Society scoring systems. The mean follow-up was for six years (1 to 13) with only two patients lost to follow-up and 34 dying during the period of study, one of whom had required revision for infection.

There were 11 revisions performed in total, three for aseptic loosening, six for infection, one for a periprosthetic fracture and one for a painful but well-fixed replacement performed at another centre.

With revision for any cause as the endpoint, the survival at ten years was 94.5% (95% CI 85.1 to 100), and with aseptic loosening as the endpoint 98.4% (95% CI 93 to 100). The mean American Knee Society score improved from 47.6 (0 to 88) to 72.2 (26 to 100) and for function from 45.1 (0 to 100) to 93.1 (45 to 100). Radiological review failed to detect migration in any of the surviving knees.

The clinical and radiological results of the Medial Rotation TKR are satisfactory at ten years. The increased congruence of the medial compartment has not led to an increased rate of loosening and continued use can be supported.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 327 - 332
1 Mar 2009
García-Rey E García-Cimbrelo E Cordero-Ampuero J

We reviewed 111 hemispherical Duraloc series-500 acetabular components with a minimum follow-up of 12 years. The mean clinical and radiological follow-up was 13.4 years (12 to 15). A Profile hydroxyapatite-coated anatomical femoral component was used in each case. Six patients had a late dislocation, for whom the polyethylene liner was exchanged. Each acetabular component was well fixed and all femoral components showed signs of bone ingrowth. The mean rate of femoral head penetration was 0.10 mm/year (0.021 to 0.481). The probability of not developing femoral cortical hypertrophy and proximal osteopenia by 12 years was 80.2% (95% confidence interval, 72.7 to 87.6) and 77.5% (95% confidence interval, 69.7 to 85.2), respectively. Despite these good clinical results, further follow-up is needed to determine whether these prostheses will loosen with time.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1339 - 1347
1 Oct 2013
Scott CEH Eaton MJ Nutton RW Wade FA Pankaj P Evans SL

As many as 25% to 40% of unicompartmental knee replacement (UKR) revisions are performed for pain, a possible cause of which is proximal tibial strain. The aim of this study was to examine the effect of UKR implant design and material on cortical and cancellous proximal tibial strain in a synthetic bone model. Composite Sawbone tibiae were implanted with cemented UKR components of different designs, either all-polyethylene or metal-backed. The tibiae were subsequently loaded in 500 N increments to 2500 N, unloading between increments. Cortical surface strain was measured using a digital image correlation technique. Cancellous damage was measured using acoustic emission, an engineering technique that detects sonic waves (‘hits’) produced when damage occurs in material.

Anteromedial cortical surface strain showed significant differences between implants at 1500 N and 2500 N in the proximal 10 mm only (p < 0.001), with relative strain shielding in metal-backed implants. Acoustic emission showed significant differences in cancellous bone damage between implants at all loads (p = 0.001). All-polyethylene implants displayed 16.6 times the total number of cumulative acoustic emission hits as controls. All-polyethylene implants also displayed more hits than controls at all loads (p < 0.001), more than metal-backed implants at loads ≥ 1500 N (p < 0.001), and greater acoustic emission activity on unloading than controls (p = 0.01), reflecting a lack of implant stiffness. All-polyethylene implants were associated with a significant increase in damage at the microscopic level compared with metal-backed implants, even at low loads. All-polyethylene implants should be used with caution in patients who are likely to impose large loads across their knee joint.

Cite this article: Bone Joint J 2013;95-B:1339–47.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1284 - 1290
1 Oct 2008
Ihle M Mai S Pfluger D Siebert W

In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection.

Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 85 - 89
1 Nov 2012
Drexler M Dwyer T Marmor M Abolghasemian M Sternheim A Cameron HU

In this study we present our experience with four generations of uncemented total knee arthroplasty (TKA) from Smith & Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing on the failure rates correlating with each design change. Beginning in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix were implanted by the senior author. The rate of revision for loosening was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the Tricon 2 with a HA coated tibial component, and 1.3% for the Profix TKA. No loosening of the femoral component was seen with the Tricon M, Tricon LS or Tricon 2, with no loosening seen of the tibial component with the Profix TKA. Regarding revision for wear, the incidence was 13.1% for the Tricon M, 6.6% for the Tricon LS, 2.3% for the Tricon 2, and 0% for the Profix. These results demonstrate that improvements in the design of uncemented components, including increased polyethylene thickness, improved polyethylene quality, and the introduction of hydroxyapatite coating, has improved the outcomes of uncemented TKA over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 786 - 789
1 Jun 2005
Grobler GP Learmonth ID Bernstein BP Dower BJ

We retrospectively reviewed, ten years after surgery, 100 consecutive total hip replacements in which the Duraloc 300 cup had been used. Post-operative radiographs were analysed for placement of the cup and interface gaps and follow-up radiographs for lucent lines, osteolysis, wear and migration.

All the components were found to be stable with no evidence of loosening. The mean rate of wear was 0.12 mm/year. Three hips developed acetabular osteolysis at the level of the apex hole. Two have successfully undergone bone grafting without removal of the implants and one patient is awaiting surgery. The Duraloc 300 cup has a survival of 100% at ten years with no aseptic loosening and a low incidence of pelvic osteolysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 34 - 38
1 Jan 2007
Epinette J Manley MT

This study describes 146 primary total knee replacements, either fully or partially coated with hydroxyapatite of which 74 knees in 68 patients were available for clinical and radiological assessment at a mean of 11.2 years (10 to 15). The global failure rate was 1.37% and survival rate with mechanical failure as the end-point was 98.14%. Radiological assessment indicated intimate contact between bone and the hydroxyapatite coating. Over time the hydroxyapatite coating appears to encourage filling of interface gaps remaining after surgery. Our results compare favourably with those of series describing cemented or porous-coated knee replacements, and suggest that fixation with hydroxyapatite is a reliable option in primary total knee replacement.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 7 | Pages 937 - 941
1 Jul 2011
Bae J Oh J Chon C Oh C Hwang J Yoon Y

We evaluated the biomechanical properties of two different methods of fixation for unstable fractures of the proximal humerus. Biomechanical testing of the two groups, locking plate alone (LP), and locking plate with a fibular strut graft (LPSG), was performed using seven pairs of human cadaveric humeri. Cyclical loads between 10 N and 80 N at 5 Hz were applied for 1 000 000 cycles. Immediately after cycling, an increasing axial load was applied at a rate of displacement of 5 mm/min. The displacement of the construct, maximum failure load, stiffness and mode of failure were compared.

The displacement was significantly less in the LPSG group than in the LP group (p = 0.031). All maximum failure loads and measures of stiffness in the LPSG group were significantly higher than those in the LP group (p = 0.024 and p = 0.035, respectively). In the LP group, varus collapse and plate bending were seen. In the LPSG group, the humeral head cut out and the fibular strut grafts fractured. No broken plates or screws were seen in either group.

We conclude that strut graft augmentation significantly increases both the maximum failure load and the initial stiffness of this construct compared with a locking plate alone.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 484 - 489
1 Apr 2011
Charalambous CP Abiddin Z Mills SP Rogers S Sutton P Parkinson R

The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint.

The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use.