A prospective cohort of 222 patients who underwent revision hip replacement between April 2001 and March 2004 was evaluated to determine predictors of function, pain and activity level between one and two years post-operatively, and to define quality of life outcomes using validated patient reported outcome tools. Predictive models were developed and proportional odds regression analyses were performed to identify factors that predict quality of life outcomes at one and two years post-operatively. The dependent outcome variables were the Western Ontario and McMaster Osteoarthritis Index (WOMAC) function and pain scores, and University of California Los Angeles activity scores. The independent variables included patient demographics, operative factors, and objective quality of life parameters, including pre-operative WOMAC, and the Short Form-12 mental component score. There was a significant improvement ( Predictors of quality of life outcomes after revision hip replacement were established. Although some patient-specific and surgery-specific variables were important, age, gender, Charnley class and pre-operative WOMAC function score had the most robust associations with outcome.
Custom-made intercalary endoprostheses may be used for the reconstruction of diaphyseal defects following the resection of bone tumours. The aim of this study was to determine the survival of intercalary endoprostheses with a lap joint design, and to evaluate the clinical results, complications and functional outcome. We retrospectively reviewed six consecutive patients, three of whom underwent limb salvage with intercalary endoprostheses of the tibia, two of the femur, and one of the humerus. Their mean age was 42 years (28 to 64). The mean follow-up was 21.6 months (9 to 58). The humeral prosthesis required revision at 14 months owing to aseptic loosening. There were no implant-related failures. Musculoskeletal Tumour Society functional outcome scores indicated that patients achieved 90% of premorbid function. Custom intercalary endoprostheses result in reconstructions comparable with, if not better than, those of allografts. Using this design of implant reduces the incidence of early complications and difficulties experienced with previous versions.
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.
Avascular necrosis of the femoral head creates considerable morbidity in successful renal transplant recipients who are generally young and expect active lifestyles. Total hip replacement is considered the treatment of choice in these patients, but surgeons may be wary because of a supposed increase in the risk of infection and other complications. A review of the literature reveals that cemented hip arthroplasty provides good to excellent functional outcomes for renal transplant patients. Most authors have found that the risk of infection is not increased despite chronic immunosuppression, but the rates of general complications are and should be anticipated and treated. There is a high rate of early failure in these patients because of their young age and diffuse osteopenia as a result of secondary hyperparathyroidism related to the underlying renal disease and chronic steroid use. Recent studies have found that despite decreased bone stock in these patients, porous-coated prostheses are not contraindicated.
The aim of this prospective randomised study
was to compare the clinical and radiological results of a cemented
all-polyethylene Ultima acetabular component with those of a cementless
porous-coated acetabular component (PFC) following total hip replacement
(THR). A total of 287 patients received either a polyethylene acetabular
component (group A) or a cobalt–chromium porous-coated component
(group B) with an identical cemented femoral component and 28 mm
cobalt-chromium head, thus making it the largest study of its type.
Patients were evaluated radiologically and clinically using the
Harris hip score (HHS). Group A comprised 183 patients (73 male,
110 female) with a mean age of
71.3 years (55 to 89). Group B comprised 104 patients (48 male,
56 female) with a mean age of 69.8 years (56 to 89). A total of
16 patients (13 in Group A, three in Group B) did not have post-operative
data for analysis. The mean follow-up in group A was 7.52 years
(0.4 to 15.0) and in Group B 7.87 years (0.5 to 14.0). At final follow-up the mean HHS was similar between groups A
and B (74.5 (25 to 100) and 78.0 (37 to 100), respectively; p =
0.068). The total number of revisions for any cause was 28, 17 of
which were in group A and 11 in group B. The ten-year survivorship
was 86.8% (95% confidence interval (CI) 78.4 to 92.1) and 89.2%
(95% CI 78.3 to 94.8) for groups A and B, respectively (log-rank
p-value = 0.938). A total of 20 cemented and two cementless acetabular
components had evidence of acetabular radiolucencies or acetabular
component migration at last follow-up (p = 0.001). These results indicate that patients with a cemented all-polyethylene
and cementless porous-coated polyethylene lined acetabular component
have similar long-term clinical outcomes.
We retrospectively reviewed 21 patients (22 shoulders) who presented with deep infection after surgery to the shoulder, 17 having previously undergone hemiarthroplasty and five open repair of the rotator cuff. Nine shoulders had undergone previous surgical attempts to eradicate their infection. The diagnosis of infection was based on a combination of clinical suspicion (16 shoulders), positive frozen sections (>
5 polymorphonuclear leukocytes per high-power field) at the time of revision (15 shoulders), positive intra-operative cultures (18 shoulders) or the pre-operative radiological appearances. The patients were treated by an extensive debridement, intravenous antibiotics, and conversion to a reverse shoulder prosthesis in either a single- (10 shoulders) or a two-stage (12 shoulders) procedure. At a mean follow-up of 43 months (25 to 66) there was no evidence of recurrent infection. All outcome measures showed statistically significant improvements. Mean abduction improved from 36.1° (
The best treatment for the active and lucid elderly patient with a displaced intracapsular fracture of the femoral neck is still controversial. Randomised controlled trials have shown that a primary total hip replacement is superior to internal fixation as regards the need for secondary surgery, hip function and health-related quality of life. Despite good results achieved with total hip replacement in this group, most orthopaedic surgeons still advocate hemiarthroplasty for this injury. We studied 120 patients with a mean age of 81 years (70 to 90) with an acute displaced intracapsular fracture of the femoral neck. They were randomly allocated to be treated with either a bipolar hemiarthroplasty or total hip replacement. Outcome measurements included peri-operative data, general and hip-specific complications, hip function and health-related quality of life. The patients were reviewed at four and 12 months. The duration of surgery was longer in the total hip replacement group (102 minutes (70 to 151)) These results indicate that a total hip replacement provides better function than a bipolar hemiarthroplasty as soon as one year post-operatively, without increasing the complication rate. We recommend total hip replacement as the primary treatment for this group of patients.
We reviewed the long-term radiological outcome,
complications and revision operations in 19 children with quadriplegic
cerebral palsy and hip dysplasia who underwent combined peri-iliac
osteotomy and femoral varus derotation osteotomy. They had a mean
age of 7.5 years (1.6 to 10.9) and comprised 22 hip dislocations
and subluxations. We also studied the outcome for the contralateral
hip. At a mean follow-up of 11.7 years (10 to 15.1) the Melbourne
cerebral palsy (CP) hip classification was grade 2 in 16 hips, grade
3 in five, and grade 5 in one. There were five complications seen
in four hips (21%, four patients), including one dislocation, one
subluxation, one coxa vara with adduction deformity, one subtrochanteric
fracture and one infection. A recurrent soft-tissue contracture occurred
in five hips and ten required revision surgery. In pre-adolescent children with quadriplegic cerebral palsy good
long-term outcomes can be achieved after reconstruction of the hip;
regular follow-up is required.
Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
Primary uncemented femoral stems reported to the Norwegian arthroplasty register between 1987 and 2005 were included in this prospective observational study. There were 11 516 hips (9679 patients) and 14 different designs of stem. Kaplan-Meier survival probabilities and Cox regression were used to analyse the data. With aseptic loosening as the end-point, all currently used designs performed excellently with survival of 96% to 100% at ten years. With the end-point as stem revision for any cause, the long-term results of the different designs varied from poor to excellent, with survival at 15 years ranging between 29% and 97%. Follow-up for longer than seven years was needed to identify some of the poorly-performing designs. There were differences between the stems; the Corail, used in 5456 hips, was the most frequently used stem with a survival of 97% at 15 years. Male gender was associated with an increased risk of revision of × 1.3 (95% confidence interval 1.05 to 1.52), but age and diagnosis had no influence on the results. Overall, modern uncemented femoral stems performed well. Moderate differences in survival between well-performing stems should be interpreted with caution since the differences may be caused by factors other than the stem itself.
An international faculty of orthopaedic surgeons
presented their work on the current challenges in hip surgery at
the London Hip Meeting which was attended by over
400 delegates. The topics covered included femoroacetabular impingement, thromboembolic
phenomena associated with hip surgery, bearing surfaces (including metal-on-metal
articulations), outcomes of hip replacement surgery and revision
hip replacement. We present a concise report of the current opinions
on hip surgery from this meeting with appropriate references to
the current literature.
Tranexamic acid (TEA), an inhibitor of fibrinolysis,
reduces blood loss after routine total knee replacement (TKR). However,
controversy persists regarding the dosage and timing of administration
of this drug during surgery. We performed a prospective randomised
controlled study to examine the optimum blood-saving effect of TEA
in minimally invasive TKR. We randomly assigned 151 patients who underwent unilateral minimally
invasive TKR to three groups: 1) a placebo group (50 patients);
2) a one-dose TEA group (52 patients), who received one injection
of TEA (10 mg/kg) intra-operatively on deflation of the tourniquet;
and 3) a two-dose TEA group (49 patients), who received two injections
of TEA (10 mg/kg) given pre-operatively and intra-operatively. Total
blood loss was calculated from the maximum loss of haemoglobin.
All patients were followed clinically for the presence of venous
thromboembolism (VTE). The mean total blood loss was significantly higher in the placebo
group than in the other two groups (1222 ml (845 to 2043) Our prospective randomised controlled study showed that one intra-operative
injection of TEA is effective for blood conservation after minimally
invasive TKR.
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other polished tapered stem designs.
The lateral compartment is predominantly affected
in approximately 10% of patients with osteoarthritis of the knee. The
anatomy, kinematics and loading during movement differ considerably
between medial and lateral compartments of the knee. This in the
main explains the relative protection of the lateral compartment
compared with the medial compartment in the development of osteoarthritis.
The aetiology of lateral compartment osteoarthritis can be idiopathic,
usually affecting the femur, or secondary to trauma commonly affecting
the tibia. Surgical management of lateral compartment osteoarthritis
can include osteotomy, unicompartmental knee replacement and total
knee replacement. This review discusses the biomechanics, pathogenesis
and development of lateral compartment osteoarthritis and its management. Cite this article:
We describe the results of 81 consecutive revision
total hip replacements with impaction grafting in 79 patients using
a collared polished chrome–cobalt stem, customised in length according
to the extent of distal bone loss. Our hypothesis was that the features
of this stem would reduce the rate of femoral fracture and subsidence
of the stem. The mean follow-up was 12 years (8 to 15). No intra-operative
fracture or significant subsidence occurred. Only one patient suffered
a post-operative diaphyseal fracture, which was associated with
a fall. All but one femur showed incorporation of the graft. No
revision for aseptic loosening was recorded. The rate of survival of the femoral component at 12 years, using
further femoral revision as the endpoint, was 100% (95% confidence
interval (CI) 95.9 to 100), and at nine years using re-operation
for any reason as the endpoint, was 94.6% (95% CI 92.0 to 97.2). These results suggest that a customised cemented polished stem
individually adapted to the extent of bone loss and with a collar
may reduce subsidence and the rate of fracture while maintaining
the durability of the fixation.
Plasma levels of cobalt and chromium ions and
Metal Artefact Reduction Sequence (MARS)-MRI scans were performed
on patients with 209 consecutive, unilateral, symptomatic metal-on-metal
(MoM) hip arthroplasties. There was wide variation in plasma cobalt
and chromium levels, and MARS-MRI scans were positive for adverse reaction
to metal debris (ARMD) in 84 hips (40%). There was a significant
difference in the median plasma cobalt and chromium levels between
those with positive and negative MARS-MRI scans (p <
0.001).
Compared with MARS-MRI as the potential reference standard for the
diagnosis of ARMD, the sensitivity of metal ion analysis for cobalt
or chromium with a cut-off of >
7 µg/l was 57%. The specificity was
65%, positive predictive value was 52% and the negative predictive
value was 69% in symptomatic patients. A lowered threshold of >
3.5 µg/l for cobalt and chromium ion levels improved the sensitivity
and negative predictive value to 86% and 74% but at the expense
of specificity (27%) and positive predictive value (44%). Metal ion analysis is not recommended as a sole indirect screening
test in the surveillance of symptomatic patients with a MoM arthroplasty.
The investigating clinicians should have a low threshold for obtaining
cross-sectional imaging in these patients, even in the presence
of low plasma metal ion levels.
In a prospective multicentre study we investigated
variations in pain management used by knee arthroplasty surgeons
in order to compare the differences in pain levels among patients
undergoing total knee replacements (TKR), and to compare the effectiveness
of pain management protocols. The protocols, peri-operative levels
of pain and patient satisfaction were investigated in 424 patients
who underwent TKR in 14 hospitals. The protocols were highly variable
and peri-operative pain levels varied substantially, particularly
during the first two post-operative days. Differences in levels
of pain were greatest during the night after TKR, when visual analogue
scores ranged from 16.9 to 94.3 points. Of the methods of managing pain, the combined use of peri-articular
infiltration and nerve blocks provided better pain relief than other
methods during the first two post-operative days. Patients managed
with peri-articular injection plus nerve block, and epidural analgesia
were more likely to have higher satisfaction at two weeks after TKR.
This study highlights the need to establish a consistent pain management
strategy after TKR.
There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
Peri-prosthetic patellar fracture following resurfacing
as part of total knee replacement (TKR) is an infrequent yet challenging
complication. This case-control study was performed to identify
clinical, radiological and surgical factors that increase the risk
of developing a spontaneous patellar fracture after TKR. Patellar
fractures were identified in 74 patients (88 knees) from a series
of 7866 consecutive TKRs conducted between 1998 and 2009. After excluding
those with a previous history of extensor mechanism realignment
or a clear traumatic event, a metal-backed patella, any uncemented
component or subsequent infection, the remaining 64 fractures were
compared with a matched group of TKRs with an excellent outcome
defined by the Knee Society score. The mean age of patients with
a fracture was 70 years (51 to 81) at the time of TKR. Patellar
fractures were detected at a mean of 13.4 months (2 to 84) after
surgery. The incidence of patellar fracture was found to be strongly
associated with the number of previous knee operations, greater
pre-operative mechanical malalignment, smaller post-operative patellar
tendon length, thinner post-resection patellar thickness, and a
lower post-operative Insall-Salvati ratio. An understanding of the risk factors associated with spontaneous
patellar fracture following TKR provides a valuable insight into
prevention of this challenging complication.
We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6)