We reviewed seven children with torticollis due to refractory atlanto-axial rotatory fixation who were treated in a halo vest. Pre-operative three-dimensional CT and sagittal CT imaging showed deformity of the superior articular process of C2 in all patients. The mean duration of halo vest treatment was 67 days (46 to 91). The mean follow-up was 34 months (8 to 73); at the latest review six patients demonstrated remodelling of the deformed articular process. The other child, who had a more severe deformity, required C1-2 fusion. We suggest that patients with atlanto-axial rotatory fixation who do not respond to conservative treatment and who have deformity of the superior articular process of C2 should undergo manipulative reduction and halo-vest fixation for two to three months to induce remodelling of the deformed superior articular process before C1-2 fusion is considered.
Given the increasing number of total hip arthroplasty
procedures being performed annually, it is imperative that orthopaedic
surgeons understand factors responsible for instability. In order
to treat this potentially complex problem, we recommend correctly
classifying the type of instability present based on component position, abductor
function, impingement, and polyethylene wear. Correct classification
allows the treating surgeon to choose the appropriate revision option
that ultimately will allow for the best potential outcome. Cite this article:
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article:
We prospectively evaluated the long-term outcome of 158 consecutive patients who underwent revision total hip replacement using uncemented computer-assisted design-computer-assisted manufacture femoral components. There were 97 men and 61 women. Their mean age was 63.1 years (34.6 to 85.9). The mean follow-up was 10.8 years (10 to 12). The mean Oxford, Harris and Western Ontario and McMaster hip scores improved from 41.1, 44.2 and 52.4 pre-operatively to 18.2, 89.3 and 12.3, respectively (p <
0.0001, for each). Six patients required further surgery. The overall survival of the femoral component was 97% (95% confidence interval 94.5 to 99.7). These results are comparable to those of previously published reports for revision total hip replacement using either cemented or uncemented components.
The optimal management of intracapsular fractures of the femoral
neck in independently mobile patients remains open to debate. Successful
fixation obviates the limitations of arthroplasty for this group
of patients. However, with fixation failure rates as high as 30%,
the outcome of revision surgery to salvage total hip arthroplasty
(THA) must be considered. We carried out a systematic review to
compare the outcomes of salvage THA and primary THA for intracapsular
fractures of the femoral neck. We performed a Preferred Reporting Items for Systematic Reviews
and Meta-Analysis (PRISMA) compliant systematic review, using the
PubMed, EMBASE and Cochrane libraries databases. A meta-analysis
was performed where possible, and a narrative synthesis when a meta-analysis
was not possible.Aims
Patients and Methods
Revision of a cemented hemiarthroplasty of the
hip may be a hazardous procedure with high rates of intra-operative complications.
Removing well-fixed cement is time consuming and risks damaging
already weak bone or perforating the femoral shaft. The cement-in-cement
method avoids removal of intact cement and has shown good results
when used for revision total hip arthroplasty (THA). The use of
this technique for the revision of a hemiarthroplasty to THA has
not been previously reported. A total of 28 consecutive hemiarthroplasties (in 28 patients)
were revised to a THA using an Exeter stem and the cement-in-cement
technique. There were four men and 24 women; their mean age was
80 years (35 to 93). Clinical and radiographic data, as well as
operative notes, were collected prospectively and no patient was
lost to follow-up. Four patients died within two years of surgery. The mean follow
up of the remainder was 70 months (25 to 124). Intra-operatively
there was one proximal perforation, one crack of the
femoral calcar and one acetabular fracture. No femoral components
have required subsequent revision for aseptic loosening or are radiologically loose. Four patients with late complications (14%) have since undergone
surgery (two for a peri-prosthetic fracture, and one each for deep
infection and recurrent dislocation) resulting in an overall major
rate of complication of 35.7%. The cement-in-cement technique provides reliable femoral fixation
in this elderly population and may reduce operating time and rates
of complication. Cite this article:
This study identifies early risk factors for symptomatic nonunion
of displaced midshaft fractures of the clavicle that aid identification
of an at risk group who may benefit from surgery. We performed a retrospective study of 88 patients aged between
16 and 60 years that were managed non-operatively. Aims
Methods
We undertook a review of the literature relating to the two basic stem designs in use in cemented hip replacement, namely loaded tapers or force-closed femoral stems, and the composite beam or shape-closed designs. The associated stem fixation theory as understood from It is clear that both design principles are capable of producing successful long-term results, providing that their specific requirements of stem metallurgy, shape and surface finish, preparation of the bone and handling of the cement are observed.
Most published randomised controlled trials which
compare the rates of wear of conventional and cross-linked (XL) polyethylene
(PE) in total hip arthroplasty (THA) have described their use with
a cementless acetabular component. We conducted a prospective randomised study to assess the rates
of penetration of two distinct types of PE in otherwise identical
cemented all-PE acetabular components. A total of 100 consecutive patients for THA were randomised to
receive an acetabular component which had been either highly XL
then remelted or moderately XL then annealed. After a minimum of eight years follow-up, 38 hips in the XL group
and 30 hips in the annealed group had complete data (mean follow-up
of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively).
In the XL group, the steady state rate of penetration from one year
onwards was -0.0002 mm/year ( These results show that the yearly linear rate of femoral head
penetration can be significantly reduced by using a highly XLPE
cemented acetabular component. Cite this article:
We reviewed the outcome of 69 uncemented, custom-made,
distal femoral endoprosthetic replacements performed in 69 patients
between 1994 and 2006. There were 31 women and 38 men with a mean
age at implantation of 16.5 years (5 to 37). All procedures were
performed for primary malignant bone tumours of the distal femur.
At a mean follow-up of 124.2 months (4 to 212), 53 patients were
alive, with one patient lost to follow-up. All nine implants (13.0%)
were revised due to aseptic loosening at a mean of 52 months (8
to 91); three implants (4.3%) were revised due to fracture of the
shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic
fracture. Bone remodelling associated with periosteal cortical thinning
adjacent to the uncemented intramedullary stem was seen in 24 patients
but this did not predispose to failure. All aseptically loose implants
in this series were diagnosed to be loose within the first five
years. The results from this study suggest that custom-made uncemented
distal femur replacements have a higher rate of aseptic loosening
compared to published results for this design when used with cemented
fixation. Loosening of uncemented replacements occurs early indicating
that initial fixation of the implant is crucial. Cite this article:
We have analysed from the Finnish Arthroplasty Register the long-term survivorship of the 12 most commonly-used cemented implants between 1980 and 2005 in patients aged 55 years or older with osteoarthritis. Only two designs of femoral component, the Exeter Universal and the Müller Straight femoral component had a survivorship of over 95% at ten years with revision for aseptic loosening as the endpoint. At 15 years of the femoral and acetabular component combinations, only the Exeter Universal/Exeter All-poly implant had a survival rate of over 90% with revision for aseptic loosening as the endpoint. In the subgroup of patients aged between 55 and 64 years, survivorship overall was less than 90% at ten years. The variation in the long-term rates of survival of different cemented hip implants was considerable in patients aged 55 years or older. In those aged between 55 and 64 years, none of the cemented prostheses studied yielded excellent long-term survival rates (≥ 90% at 15 years).
We present the extended follow-up (≥ 20 years)
of a series of fully hydroxyapatite-coated femoral components used in
72 primary total hip replacements (THRs). Earlier results of this
cohort have been previously published. All procedures were performed
between 1986 and 1991. The series involved 45 women and 15 men with
12 bilateral procedures. Their mean age at the time of surgery was
60 years (46 to 80) and the mean duration of follow-up was 22.5
years (20 to 25). At final follow-up, the mean Merle d’Aubigné and
Postel hip scores were 5.5 (4.5 to 6), 3.8 (3.5 to 5) and 3.3 (3.0
to 5.0) for pain, mobility and function, respectively. Of the patients
92% were very satisfied at the time of final follow-up. There were seven revisions: six of the acetabular component for
aseptic loosening and one of both the stem and the acetabular component
for loosening due to deep infection. The survival of this prosthesis
at 22.5 years with revision for any reason as the endpoint was 91.7%
(95% confidence interval (CI) 84 to 99). Survival with aseptic loosening
of the stem as the endpoint was 100% (95% CI 90 to 100). This prosthesis provides pain relief in the long term. Survival
of this component is comparable to the best results for primary
THR with any means of fixation. Cite this article:
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article:
The aim of this study was to investigate the
relationship between the geometry of the proximal femur and the incidence
of intra-operative fracture during uncemented total hip arthroplasty
(THA). We studied the pre-operative CT scans of 100 patients undergoing
THA with an uncemented femoral component. We measured the anteroposterior
and mediolateral dimensions at the level of division of the femoral
neck to calculate the aspect ratio of the femur. Wide variations
in the shape of the femur were observed, from round, to very narrow
elliptic. The femurs of women were narrower than those of men (p
<
0.0001) and small femurs were also narrower than large ones.
Patients with an intra-operative fracture of the calcar had smaller
and narrower femurs than those without a fracture (p <
0.05)
and the implanted Corail stems were smaller in those with a fracture
(mean size 9 The variability of the shape of the femoral neck at the level
of division contributes to the understanding of the causation of
intra-operative fractures in uncemented THA. Cite this article:
In 2004 we described the ten-year prospective results of 38 total hip replacements using the Furlong hydroxyapatite-ceramic-coated femoral component in 35 patients <
50 years old. We have now reviewed the surviving 35 arthroplasties in 33 patients at a mean of 16 years (10.3 to 19.9). The mean age of the surviving patients at the time of operation was 41.3 years (26.0 to 49.0). Of these, eight have undergone revision of their acetabular component for aseptic loosening. None of the femoral components has had revision for aseptic loosening giving a survival rate of 100% at 16 years (95% confidence interval 89% to 100%). The Furlong hydroxyapatite-ceramic-coated femoral component gives excellent long-term survival in young and active patients.
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.
Fractures of the occipital condyle are rare. Their prompt diagnosis is crucial since there may be associated cranial nerve palsies and cervical spinal instability. The fracture is often not visible on a plain radiograph. We report the case of a 21-year-old man who sustained an occipital condylar fracture without any associated cranial nerve palsy or further injuries. We have also reviewed the literature on this type of injury, in order to assess the incidence, the mechanism and the association with head and cervical spinal injuries as well as classification systems, options for treatment and outcome.
Unlinked, linked and convertible total elbow
arthroplasties (TEAs) are currently available. This study is the
first to report the clinical results of the convertible Latitude
TEA. This was a retrospective study of a consecutive cohort of 63
patients (69 primary TEAs) with a mean age of 60 years (23 to 87).
Between 2006 and 2008 a total of 19 men and 50 women underwent surgery.
The mean follow-up was 43 months (8 to 84). The range of movement,
function and pain all improved six months post-operatively and either
continued to improve slightly or reached a plateau thereafter. The
complication rate is similar to that reported for other TEA systems.
No loosening was seen. Remarkable is the disengagement of the radial
head component in 13 TEAs (31%) with a radial head component implanted. Implantation of both the linked and the unlinked versions of
the Latitude TEA results in improvement of function and decreased
pain, and shows high patient satisfaction at mid-term follow-up. Cite this article:
Total knee replacement (TKR) smart tibial trials
have load-bearing sensors which will show quantitative compartment
pressure values and femoral-tibial tracking patterns. Without smart
trials, surgeons rely on feel and visual estimation of imbalance
to determine if the knee is optimally balanced. Corrective soft-tissue
releases are performed with minimal feedback as to what and how
much should be released. The smart tibial trials demonstrate graphically
where and how much imbalance is present, so that incremental releases
can be performed. The smart tibial trials now also incorporate accelerometers
which demonstrate the axial alignment. This now allows the surgeon
the option to perform a slight recut of the tibia or femur to provide
soft-tissue balance without performing soft-tissue releases. Using
a smart tibial trial to assist with soft-tissue releases or bone
re-cuts, improved patient outcomes have been demonstrated at one
year in a multicentre study of 135 patients (135 knees). Cite this article: