The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA
The survivorship method of analysis has been used to compare the failure rate and overall success of 1,430 cemented primary total knee arthroplasties performed at The Hospital for Special Surgery over a 15-year period. There were 224 total condylar prostheses with a polyethylene tibia, 289 of the posterior stabilised type with an all polyethylene tibia, and 917 posterior stabilised with a metal-backed
The optimum cementing technique for the tibial
component in cemented primary total knee replacement (TKR) remains
controversial. The technique of cementing, the volume of cement
and the penetration are largely dependent on the operator, and hence
large variations can occur. Clinical, experimental and computational
studies have been performed, with conflicting results. Early implant
migration is an indication of loosening. Aseptic loosening is the
most common cause of failure in primary TKR and is the product of
several factors. Sufficient penetration of cement has been shown
to increase implant stability. This review discusses the relevant literature regarding all aspects
of the cementing of the
The aim of this study was to determine the risk of tibial eminence avulsion intraoperatively for bi-unicondylar knee arthroplasty (Bi-UKA), with consideration of the effect of implant positioning, overstuffing, and sex, compared to the risk for isolated medial unicondylar knee arthroplasty (UKA-M) and bicruciate-retaining total knee arthroplasty (BCR-TKA). Two experimentally validated finite element models of tibia were implanted with UKA-M, Bi-UKA, and BCR-TKA. Intraoperative loads were applied through the condyles, anterior cruciate ligament (ACL), medial collateral ligament (MCL), and lateral collateral ligament (LCL), and the risk of fracture (ROF) was evaluated in the spine as the ratio of the 95th percentile maximum principal elastic strains over the tensile yield strain of proximal tibial bone.Aims
Methods
The purpose of this study was to evaluate the
change in sagittal tibiotalar alignment after total ankle arthroplasty (TAA)
for osteoarthritis and to investigate factors affecting the restoration
of alignment. . This retrospective study included 119 patients (120 ankles) who
underwent three component TAA using the Hintegra prosthesis. A total
of 63 ankles had anterior displacement of the talus before surgery
(group A), 49 had alignment in the normal range (group B), and eight
had posterior displacement of the talus (group C). Ankles in group
A were further sub-divided into those in whom normal alignment was
restored following TAA (41 ankles) and those with persistent displacement
(22 ankles). Radiographic and clinical results were assessed. Pre-operatively, the alignment in group A was significantly more
varus than that in group B, and the posterior slope of the tibial
plafond was greater (p <
0.01 in both cases). The posterior slope
of the
We performed a CT-based computer simulation study
to determine how the relationship between any inbuilt posterior
slope in the proximal tibial osteotomy and cutting jig rotational
orientation errors affect
In this study we present our experience with
four generations of uncemented total knee arthroplasty (TKA) from Smith
&
Nephew: Tricon M, Tricon LS, Tricon II and Profix, focusing
on the failure rates correlating with each design change. Beginning
in 1984, 380 Tricon M, 435 Tricon LS, 305 Tricon 2 and 588 Profix
were implanted by the senior author. The rate of revision for loosening
was 1.1% for the Tricon M, 1.1% for the Tricon LS, 0.5% for the
Tricon 2 with a HA coated
This systematic review aims to compare the precision of component positioning, patient-reported outcome measures (PROMs), complications, survivorship, cost-effectiveness, and learning curves of MAKO robotic arm-assisted unicompartmental knee arthroplasty (RAUKA) with manual medial unicompartmental knee arthroplasty (mUKA). Searches of PubMed, MEDLINE, and Google Scholar were performed in November 2021 according to the Preferred Reporting Items for Systematic Review and Meta-Analysis statement. Search terms included “robotic”, “unicompartmental”, “knee”, and “arthroplasty”. Published clinical research articles reporting the learning curves and cost-effectiveness of MAKO RAUKA, and those comparing the component precision, functional outcomes, survivorship, or complications with mUKA, were included for analysis.Aims
Methods
A functional anterior cruciate ligament (ACL) or posterior cruciate ligament (PCL) has been assumed to be required for patients undergoing unicompartmental knee arthroplasty (UKA). However, this assumption has not been thoroughly tested. Therefore, this study aimed to assess the biomechanical effects exerted by cruciate ligament-deficient knees with medial UKAs regarding different posterior tibial slopes. ACL- or PCL-deficient models with posterior tibial slopes of 1°, 3°, 5°, 7°, and 9° were developed and compared to intact models. The kinematics and contact stresses on the tibiofemoral joint were evaluated under gait cycle loading conditions.Aims
Methods
The aim of this study was to compare any differences in the primary outcome (biphasic flexion knee moment during gait) of robotic arm-assisted bi-unicompartmental knee arthroplasty (bi-UKA) with conventional mechanically aligned total knee arthroplasty (TKA) at one year post-surgery. A total of 76 patients (34 bi-UKA and 42 TKA patients) were analyzed in a prospective, single-centre, randomized controlled trial. Flat ground shod gait analysis was performed preoperatively and one year postoperatively. Knee flexion moment was calculated from motion capture markers and force plates. The same setup determined proprioception outcomes during a joint position sense test and one-leg standing. Surgery allocation, surgeon, and secondary outcomes were analyzed for prediction of the primary outcome from a binary regression model.Aims
Methods
This study aimed to describe the use of revision knee arthroplasty in Australia and examine changes in lifetime risk over a decade. De-identified individual-level data on all revision knee arthroplasties performed in Australia from 2007 to 2017 were obtained from the Australian Orthopaedic Association National Joint Replacement Registry. Population data and life tables were obtained from the Australian Bureau of Statistics. The lifetime risk of revision surgery was calculated for each year using a standardized formula. Separate calculations were undertaken for males and females.Aims
Methods
Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs. In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.Aims
Methods
Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia. In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee.Aims
Methods
We aimed to obtain anthropometric data on Korean
knees and to compare these with data on commonly available total
knee arthroplasties (TKAs). The dimensions of the femora and tibiae
of 1168 knees were measured intra-operatively. The femoral components
were found to show a tendency toward mediolateral (ML) under-coverage
in small femurs and ML overhang in the large femurs. The ML under-coverage
was most prominent for the small prostheses. The ML/anteroposterior
(ML/AP) ratio of Korean tibiae was greater than that of
We investigated the characteristics of patients
who achieved Japanese-style deep flexion (seiza-sitting) after total knee
replacement (TKR) and measured three-dimensional positioning and
the contact positions of the femoral and
We performed a randomised controlled trial comparing
computer-assisted surgery (CAS) with conventional surgery (CONV)
in total knee replacement (TKR). Between 2009 and 2011 a total of
192 patients with a mean age of 68 years (55 to 85) with osteoarthritis
or arthritic disease of the knee were recruited from four Norwegian
hospitals. At three months follow-up, functional results were marginally
better for the CAS group. Mean differences (MD) in favour of CAS
were found for the Knee Society function score (MD: 5.9, 95% confidence
interval (CI) 0.3 to 11.4, p = 0.039), the Knee Injury and Osteoarthritis
Outcome Score (KOOS) subscales for ‘pain’ (MD: 7.7, 95% CI 1.7 to
13.6, p = 0.012), ‘sports’ (MD: 13.5, 95% CI 5.6 to 21.4, p = 0.001)
and ‘quality of life’ (MD: 7.2, 95% CI 0.1 to 14.3, p = 0.046).
At one-year follow-up, differences favouring CAS were found for
KOOS ‘sports’ (MD: 11.0, 95% CI 3.0 to 19.0, p = 0.007) and KOOS
‘symptoms’ (MD: 6.7, 95% CI 0.5 to 13.0, p = 0.035). The use of
CAS resulted in fewer outliers in frontal alignment (>
3° malalignment),
both for the entire TKR (37.9% vs 17.9%, p = 0.042)
and for the
Simultaneous bilateral total knee arthroplasty (TKA) has been used due to its financial advantages, overall resource usage, and convenience for the patient. The training model where a trainee performs the first TKA, followed by the trainer surgeon performing the second TKA, is a unique model to our institution. This study aims to analyze the functional and clinical outcomes of bilateral simultaneous TKA when performed by a trainee or a supervising surgeon, and also to assess these outcomes based on which side was done by the trainee or by the surgeon. This was a retrospective cohort study of all simultaneous bilateral TKAs performed by a single surgeon in an academic institution between May 2003 and November 2017. Exclusion criteria were the use of partial knee arthroplasty procedures, staged bilateral procedures, and procedures not performed by the senior author on one side and the trainee on another. Primary clinical outcomes of interest included revision and re-revision. Primary functional outcomes included the Oxford Knee Score (OKS) and patient satisfaction scores.Aims
Methods
The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.Aims
Methods
We carried out a prospective investigation into
the radiological outcomes of uncemented Oxford medial compartment
unicondylar replacement in 220 consecutive patients (231 knees)
performed in a single centre with a minimum two-year follow-up.
The functional outcomes using the mean Oxford knee score and the
mean high-activity arthroplasty score were significantly improved
over the pre-operative scores (p <
0.001). There were 196 patients
with a two-year radiological examination performed under fluoroscopic
guidance, aiming to provide images acceptable for analysis of the
bone–implant interface. Of the six tibial zones examined on each
knee on the anteroposterior radiograph, only three had a partial
radiolucent line. All were in the medial aspect of the tibial base plate
(zone 1) and all measured <
1 mm. All of these patients were
asymptomatic. There were no radiolucent lines seen around the femoral
component or on the lateral view. There was one revision for loosening
at one year due to initial inadequate seating of the