Numerous complications following total knee replacement (TKR)
relate to the patellofemoral (PF) joint, including pain and patellar
maltracking, yet the options for A total of three knees with end-stage osteoarthritis and three
knees that had undergone TKR at more than one year’s follow-up were
investigated. In each knee, sequential biplane radiological images
were acquired from the sagittal direction (i.e. horizontal X-ray
source and 10° below horizontal) for a sequence of eight flexion
angles. Three-dimensional implant or bone models were matched to
the biplane images to compute the six degrees of freedom of PF tracking
and TF kinematics, and other clinical measures.Objectives
Methods
The patellar clunk syndrome describes painful catching, grinding or jumping of the patella when the knee moves from a flexed to an extended position after total knee replacement (TKR). The posterior stabilised TKR had been noted to have a higher incidence of this problem. Mobile-bearing posteriorly stabilised TKRs have been introduced to improve
Objective patellar instability has been correlated with dysplasia of the femoral trochlea. This in vitro study tested the hypothesis that trochleoplasty would increase patellar stability and normalise the kinematics of a knee with a dysplastic trochlea. Six fresh-frozen knees were loaded via the heads of the quadriceps. The patella was displaced 10 mm laterally and the displacing force was measured from 0° to 90° of flexion.
Aims. Between 15% and 20% of patients remain dissatisfied following total knee arthroplasty (TKA). The SAIPH knee system (MatOrtho, Surrey, United Kingdom) is a medial ball and socket TKA that has been designed to replicate native knee kinematics in order to maximize the range of movement, stability, and function. This system is being progressively introduced in a stepwise fashion, with this study reporting the mid-term clinical and radiological outcomes. Patients and Methods. A retrospective review was undertaken of the first 100 consecutive patients with five-year follow-up following SAIPH TKA performed by the senior authors. The data that were collected included the demographics of the patients, clinical findings, the rate of intraoperative ligamentous release, patient-reported outcome measures (PROMS), radiological assessment, complications, and all-cause revision. Revision data were cross-checked with a national registry. Results. A total of 100 TKAs in 92 patients were included. Three patients died (three TKAs) and a further two TKAs were revised. Of the remaining 95 TKAs, five-year follow-up data were available for 81 TKAs (85%) in 87 patients. There were significant improvements in all PROMs and high satisfaction. The mean ROM at final follow-up was from 0° (full extension) to 124° flexion. There were seven major complications (7%): one infection, two deep vein thromboses, one cerebrovascular event, and two patients with stiffness requiring a manipulation under anaesthesia. Two patients required a lateral retinacular release to optimize
There are many reasons why a total knee replacement
(TKR) may fail and qualify for revision. Successful revision surgery
depends as much on accurate assessment of the problem TKR as it
does on revision implant design and surgical technique. Specific
modes of failure require specific surgical solutions. Causes of
failure are often presented as a list or catalogue, without a system
or process for making a decision. In addition, strict definitions
and consensus on modes of failure are lacking in published series
and registry data. How we approach the problem TKR is an essential
but neglected aspect of understanding knee replacement surgery.
It must be carried out systematically, comprehensively and efficiently.
Eight modes of failure are described: 1) sepsis; 2) extensor discontinuity;
3) stiffness; 4) tibial- femoral instability; 5)
We report a prospective analysis of clinical
outcome in patients treated with medial patellofemoral ligament
(MPFL) reconstruction using an autologous semitendinosus graft.
The technique includes superolateral portal arthroscopic assessment
before and after graft placement to ensure correct graft tension
and
In 11 paediatric patients (seven girls and four
boys, from 12 to 15 years old) with unilateral obligatory patellar dislocation
and ligamentous laxity vastus medialis advancement, lateral release,
partial patellar ligament transposition and Galeazzi semitendinosus
tenodesis was undertaken to stabilise the patella. The diagnostic criterion
for ligamentous laxity was based on the Beighton scale. Outcomes
were evaluated radiologically and functionally by measurement of
the range of knee movement and isokinetic testing. The evaluation
also included the Lysholm knee scale. Follow-up studies took place
at a mean of 8.1 years (5 to 15) post-operatively. Normal
The lateral subvastus approach combined with an osteotomy of the tibial tubercle is a recognised, but rarely used approach for total knee replacement (TKR). A total of 32 patients undergoing primary TKR was randomised into two groups, in one of which the lateral subvastus approach combined with a tibial tubercle osteotomy and in the other the medial parapatellar approach were used. The patients were assessed radiologically and clinically using measurement of the range of movement, a visual analogue patient satisfaction score, the Western Ontario McMasters University Osteoarthritis Index and the American Knee Society score. Four patients were lost to the complete follow-up at two years. At two years there were no significant differences between the groups in any of the parameters for clinical outcome. In the lateral approach group there was one complication due to displacement of the tibial tubercle osteotomy and two osteotomies took more than six months to unite. In the medial approach group, one patient had a partial tear of the quadriceps. There was a significantly greater incidence of lateral patellar subluxation in the medial approach group (3 of 12) compared with the lateral approach group (0 of 16) (p = 0.034), but without any apparent clinical detriment. We conclude that the lateral approach with tibial tubercle osteotomy is a safe technique with an outcome comparable with that of the medial parapatellar approach for TKR, but the increased surgical time and its specific complications do not support its routine use. It would seem to be more appropriate to reserve this technique for patients in whom problems with
Advanced 3D imaging and CT-based navigation have emerged as valuable tools to use in total knee arthroplasty (TKA), for both preoperative planning and the intraoperative execution of different philosophies of alignment. Preoperative planning using CT-based 3D imaging enables more accurate prediction of the size of components, enhancing surgical workflow and optimizing the precision of the positioning of components. Surgeons can assess alignment, osteophytes, and arthritic changes better. These scans provide improved insights into the patellofemoral joint and facilitate tibial sizing and the evaluation of implant-bone contact area in cementless TKA. Preoperative CT imaging is also required for the development of patient-specific instrumentation cutting guides, aiming to reduce intraoperative blood loss and improve the surgical technique in complex cases. Intraoperative CT-based navigation and haptic guidance facilitates precise execution of the preoperative plan, aiming for optimal positioning of the components and accurate alignment, as determined by the surgeon’s philosophy. It also helps reduce iatrogenic injury to the periarticular soft-tissue structures with subsequent reduction in the local and systemic inflammatory response, enhancing early outcomes. Despite the increased costs and radiation exposure associated with CT-based navigation, these many benefits have facilitated the adoption of imaged based robotic surgery into routine practice. Further research on ultra-low-dose CT scans and exploration of the possible translation of the use of 3D imaging into improved clinical outcomes are required to justify its broader implementation. Cite this article:
This systematic review and meta-analysis aimed to compare the influence of patellar resurfacing following cruciate-retaining (CR) and posterior-stabilized (PS) total knee arthroplasty (TKA) on the incidence of anterior knee pain, knee-specific patient-reported outcome measures, complication rates, and reoperation rates. A systematic review of MEDLINE, PubMed, and Google Scholar was performed to identify randomized controlled trials (RCTs) according to search criteria. Search terms used included: arthroplasty, replacement, knee (Mesh), TKA, prosthesis, patella, patellar resurfacing, and patellar retaining. RCTs that compared patellar resurfacing versus unresurfaced in primary TKA were included for further analysis. Studies were evaluated using the Scottish Intercollegiate Guidelines Network assessment tool for quality and minimization of bias. Data were synthesized and meta-analysis performed.Aims
Methods
Total knee arthroplasty (TKR) using a medial capsular approach gives worse results in arthritic knees with valgus deformity than in those in varus, usually because of swelling, poor wound healing and stiffness, instability, recurrent valgus deformity and poor
The Q angle is an important determinant of
Magnetic resonance imaging was used to analyse the patellofemoral relationships during the first 30 degrees of knee flexion in women with recurrent patellar dislocation. The patellofemoral joints were imaged both sagittally and axially with the knee flexed 0 degrees, 10 degrees, 20 degrees, and 30 degrees. At the beginning of knee flexion the sulcus angle was greater than in unaffected women, the lateral patellofemoral angle was smaller, the patella displaced further laterally, tilted more laterally and the congruence angle was directed more laterally. At 30 degrees of knee flexion these differences were less marked than at 0 degree to 10 degrees. Logistic regression analysis showed that the sulcus angle at 10 degrees of knee flexion was the most diagnostic feature, indicating that there is an anatomical predisposition to recurrent dislocation and that pathological
The transepicondylar axis is a well-established reference for the determination of femoral component rotation in total knee arthroplasty (TKA). However, when severe bone loss is present in the femoral condyles, rotational alignment can be more complicated. There is a lack of validated landmarks in the supracondylar region of the distal femur. Therefore, the aim of this study was to analyze the correlation between the surgical transepicondylar axis (sTEA) and the suggested dorsal cortex line (DCL) in the coronal plane and the inter- and intraobserver reliability of its CT scan measurement. A total of 75 randomly selected CT scans were measured by three experienced surgeons independently. The DCL was defined in the coronal plane as a tangent to the dorsal femoral cortex located 75 mm above the joint line in the frontal plane. The difference between sTEA and DCL was calculated. Descriptive statistics and angulation correlations were generated for the sTEA and DCL, as well as for the distribution of measurement error for intra- and inter-rater reliability.Aims
Methods
The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design. This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.Aims
Methods
The February 2023 Children’s orthopaedics Roundup360 looks at: Trends in management of paediatric distal radius buckle fractures; Pelvic osteotomy in patients with previous sacral-alar-iliac fixation; Sacral-alar-iliac fixation in patients with previous pelvic osteotomy; Idiopathic toe walking: an update on natural history, diagnosis, and treatment; A prediction model for treatment decisions in distal radial physeal injuries: a multicentre retrospective study; Angular deformities after percutaneous epiphysiodesis for leg length discrepancy; MRI assessment of anterior coverage is predictive of future radiological coverage; Predictive scoring for recurrent patellar instability after a first-time patellar dislocation.
This study aimed to evaluate if total knee arthroplasty (TKA) femoral components aligned in either mechanical alignment (MA) or kinematic alignment (KA) are more biomimetic concerning trochlear sulcus orientation and restoration of trochlear height. Bone surfaces from 1,012 CT scans of non-arthritic femora were segmented using a modelling and analytics system. TKA femoral components (Triathlon; Stryker) were virtually implanted in both MA and KA. Trochlear sulcus orientation was assessed by measuring the distal trochlear sulcus angle (DTSA) in native femora and in KA and MA prosthetic femoral components. Trochlear anatomy restoration was evaluated by measuring the differences in medial, lateral, and sulcus trochlear height between native femora and KA and MA prosthetic femoral components.Aims
Methods
The surgical target for optimal implant positioning in robotic-assisted total knee arthroplasty remains the subject of ongoing discussion. One of the proposed targets is to recreate the knee’s functional behaviour as per its pre-diseased state. The aim of this study was to optimize implant positioning, starting from mechanical alignment (MA), toward restoring the pre-diseased status, including ligament strain and kinematic patterns, in a patient population. We used an active appearance model-based approach to segment the preoperative CT of 21 osteoarthritic patients, which identified the osteophyte-free surfaces and estimated cartilage from the segmented bones; these geometries were used to construct patient-specific musculoskeletal models of the pre-diseased knee. Subsequently, implantations were simulated using the MA method, and a previously developed optimization technique was employed to find the optimal implant position that minimized the root mean square deviation between pre-diseased and postoperative ligament strains and kinematics.Aims
Methods
Patient dissatisfaction is not uncommon following primary total knee arthroplasty. One proposed method to alleviate this is by improving knee kinematics. Therefore, we aimed to answer the following research question: are there significant differences in knee kinematics based on the design of the tibial insert (cruciate-retaining (CR), ultra-congruent (UC), or medial congruent (MC))? Overall, 15 cadaveric knee joints were examined with a CR implant with three different tibial inserts (CR, UC, and MC) using an established knee joint simulator. The effects on coronal alignment, medial and lateral femoral roll back, femorotibial rotation, bony rotations (femur, tibia, and patella), and patellofemoral length ratios were determined.Aims
Methods
The mid-term results of kinematic alignment (KA) for total knee arthroplasty (TKA) using image derived instrumentation (IDI) have not been reported in detail, and questions remain regarding ligamentous stability and revisions. This paper aims to address the following: 1) what is the distribution of alignment of KA TKAs using IDI; 2) is a TKA alignment category associated with increased risk of failure or poor patient outcomes; 3) does extending limb alignment lead to changes in soft-tissue laxity; and 4) what is the five-year survivorship and outcomes of KA TKA using IDI? A prospective, multicentre, trial enrolled 100 patients undergoing KA TKA using IDI, with follow-up to five years. Alignment measures were conducted pre- and postoperatively to assess constitutional alignment and final implant position. Patient-reported outcome measures (PROMs) of pain and function were also included. The Australian Orthopaedic Association National Joint Arthroplasty Registry was used to assess survivorship.Aims
Methods
Evaluating musculoskeletal conditions of the lower limb and understanding the pathophysiology of complex bone kinematics is challenging. Static images do not take into account the dynamic component of relative bone motion and muscle activation. Fluoroscopy and dynamic MRI have important limitations. Dynamic CT (4D-CT) is an emerging alternative that combines high spatial and temporal resolution, with an increased availability in clinical practice. 4D-CT allows simultaneous visualization of bone morphology and joint kinematics. This unique combination makes it an ideal tool to evaluate functional disorders of the musculoskeletal system. In the lower limb, 4D-CT has been used to diagnose femoroacetabular impingement, patellofemoral, ankle and subtalar joint instability, or reduced range of motion. 4D-CT has also been used to demonstrate the effect of surgery, mainly on patellar instability. 4D-CT will need further research and validation before it can be widely used in clinical practice. We believe, however, it is here to stay, and will become a reference in the diagnosis of lower limb conditions and the evaluation of treatment options. Cite this article:
This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX mobile bearing, MUTARS; Implantcast, Germany) versus the BPKS (Brehm, Germany) hinged TKA (n = 27), in which the weight placed on the MoM hinge is diffused through a polyethylene (PE) inlay, reducing the direct load on the MoM hinge. Serum cobalt and chromium ion concentrations were assessed after minimum follow-up of 12 months, as well as functional outcome and quality of life.Aims
Methods
Rotating-hinge knee prostheses are commonly used to reconstruct the distal femur after resection of a tumour, despite the projected long-term burden of reoperation due to complications. Few studies have examined the factors that influence their failure and none, to our knowledge, have used competing risk models to do so. The purpose of this study was to determine the risk factors for failure of a rotating-hinge knee distal femoral arthroplasty using the Fine-Gray competing risk model. We retrospectively reviewed 209 consecutive patients who, between 1991 and 2016, had undergone resection of the distal femur for tumour and reconstruction using a rotating-hinge knee prosthesis. The study endpoint was failure of the prosthesis, defined as removal of the femoral component, the tibial component, or the bone-implant fixation; major revision (exchange of the femoral component, tibial component, or the bone-implant fixation); or amputation.Aims
Methods
In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis.Aims
Methods
The primary aim of this study was to compare the postoperative systemic inflammatory response in conventional jig-based total knee arthroplasty (conventional TKA) versus robotic-arm assisted total knee arthroplasty (robotic TKA). Secondary aims were to compare the macroscopic soft tissue injury, femoral and tibial bone trauma, localized thermal response, and the accuracy of component positioning between the two treatment groups. This prospective randomized controlled trial included 30 patients with osteoarthritis of the knee undergoing conventional TKA versus robotic TKA. Predefined serum markers of inflammation and localized knee temperature were collected preoperatively and postoperatively at six hours, day 1, day 2, day 7, and day 28 following TKA. Blinded observers used the Macroscopic Soft Tissue Injury (MASTI) classification system to grade intraoperative periarticular soft tissue injury and bone trauma. Plain radiographs were used to assess the accuracy of achieving the planned postioning of the components in both groups.Aims
Methods
Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction. A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity).Aims
Methods
Metaphyseal fixation during revision total knee arthroplasty (TKA) is important, but potentially difficult when using historical designs of cone. Material and manufacturing innovations have improved the size and shape of the cones which are available, and simplified the required bone preparation. In a large series, we assessed the implant survivorship, radiological results, and clinical outcomes of new porous 3D-printed titanium metaphyseal cones featuring a reamer-based system. We reviewed 142 revision TKAs in 139 patients using 202 cones (134 tibial, 68 femoral) which were undertaken between 2015 and 2016. A total of 60 involved tibial and femoral cones. Most cones (149 of 202; 74%) were used for Type 2B or 3 bone loss. The mean age of the patients was 66 years (44 to 88), and 76 (55 %) were female. The mean body mass index (BMI) was 34 kg/m2 (18 to 60). The patients had a mean of 2.4 (1 to 8) previous operations on the knee, and 68 (48%) had a history of prosthetic infection. The mean follow-up was 2.4 years (2 to 3.6).Aims
Methods
The primary aim of this study was to describe a baseline comparison of early knee-specific functional outcomes following revision total knee arthroplasty (TKA) using metaphyseal sleeves with a matched cohort of patients undergoing primary TKA. The secondary aim was to compare incidence of complications and length of stay (LOS) between the two groups. Patients undergoing revision TKA for all diagnoses between 2009 and 2016 had patient-reported outcome measures (PROMs) collected prospectively. PROMs consisted of the American Knee Society Score (AKSS) and Short-Form 12 (SF-12). The study cohort was identified retrospectively and demographics were collected. The cohort was matched to a control group of patients undergoing primary TKA.Introduction
Methods
It remains controversial whether patellofemoral joint pathology is a contraindication to lateral unicompartmental knee arthroplasty (UKA). This study aimed to evaluate the effect of preoperative radiological degenerative changes and alignment on patient-reported outcome scores (PROMs) after lateral UKA. Secondarily, the influence of lateral UKA on the alignment of the patellofemoral joint was studied. A consecutive series of patients who underwent robotic arm-assisted fixed-bearing lateral UKA with at least two-year follow-up were retrospectively reviewed. Radiological evaluation was conducted to obtain a Kellgren Lawrence (KL) grade, an Altman score, and alignment measurements for each knee. Postoperative PROMs were assessed using the Kujala (Anterior Knee Pain Scale) score, Knee Injury and Osteoarthritis Outcome Score Joint Replacement (KOOS JR), and satisfaction levels.Aims
Methods
The aims of this study were to determine the effect of osteophyte excision on deformity correction and soft tissue gap balance in varus knees undergoing computer-assisted total knee arthroplasty (TKA). A total of 492 consecutive, cemented, cruciate-substituting TKAs performed for varus osteoarthritis were studied. After exposure and excision of both cruciates and menisci, it was noted from operative records the corrective interventions performed in each case. Knees in which no releases after the initial exposure, those which had only osteophyte excision, and those in which further interventions were performed were identified. From recorded navigation data, coronal and sagittal limb alignment, knee flexion range, and medial and lateral gap distances in maximum knee extension and 90° knee flexion with maximal varus and valgus stresses, were established, initially after exposure and excision of both cruciate ligaments, and then also at trialling. Knees were defined as ‘aligned’ if the hip-knee-ankle axis was between 177° and 180°, (0° to 3° varus) and ‘balanced’ if medial and lateral gaps in extension and at 90° flexion were within 2 mm of each other.Aims
Methods
We aimed to examine the long-term mechanical survivorship, describe the modes of all-cause failure, and identify risk factors for mechanical failure of all-polyethylene tibial components in endoprosthetic reconstruction. This is a retrospective database review of consecutive endoprosthetic reconstructions performed for oncological indications between 1980 and 2019. Patients with all-polyethylene tibial components were isolated and analyzed for revision for mechanical failure. Outcomes included survival of the all-polyethylene tibial component, revision surgery categorized according to the Henderson Failure Mode Classification, and complications and functional outcome, as assessed by the Musculoskeletal Tumor Society (MSTS) score at the final follow-up.Aims
Methods
Aims
Patients and Methods
There is little literature about total knee arthroplasty (TKA) after distal femoral osteotomy (DFO). Consequently, the purpose of this study was to analyze the outcomes of TKA after DFO, with particular emphasis on: survivorship free from aseptic loosening, revision, or any re-operation; complications; radiological results; and clinical outcome. We retrospectively reviewed 29 patients (17 women, 12 men) from our total joint registry who had undergone 31 cemented TKAs after a DFO between 2000 and 2012. Their mean age at TKA was 51 years (22 to 76) and their mean body mass index 32 kg/m2 (20 to 45). The mean time between DFO and TKA was ten years (2 to 20). The mean follow-up from TKA was ten years (2 to 16). The prostheses were posterior-stabilized in 77%, varus-valgus constraint (VVC) in 13%, and cruciate-retaining in 10%. While no patient had metaphyseal fixation (e.g. cones or sleeves), 16% needed a femoral stem.Aims
Patients and Methods
The aim of this study was to investigate the changes in femoral
trochlear morphology following surgical correction of recurrent
patellar dislocation associated with trochlear dysplasia in children. A total of 23 patients with a mean age of 9.6 years (7 to 11)
were included All had bilateral recurrent patellar dislocation associated
with femoral trochlear dysplasia. The knee with traumatic dislocation
at the time of presentation or that had dislocated most frequently
was treated with medial patellar retinacular plasty (Group S). The
contralateral knee served as a control and was treated conservatively
(Group C). All patients were treated between October 2008 and August
2013. The mean follow-up was 48.7 months (43 to 56). Axial CT scans
were undertaken in all patients to assess the trochlear morphological
characteristics on a particular axial image which was established
at the point with the greatest epicondylar width based on measurements
preoperatively and at the final follow-up.Aims
Patients and Methods
The aim of this study was to determine and compare the congruency of the articular surface contact area of the patellofemoral joint (PFJ) during both active and passive movement of the knee with the use of an MRI mapping technique in both the stable and unstable PFJ. A prospective case-control MRI imaging study of patients with a history of PFJ instability and a control group of volunteers without knee symptoms was performed. The PFJs were imaged with the use of an MRI scan during both passive and active movement from 0° through to 40° of flexion. The congruency through measurement of the contact surface area was mapped in 5-mm intervals on axial slices. In all, 40 patients were studied. The case group included 31 patients with symptomatic patellofemoral instability and the control group of nine asymptomatic volunteers. The ages were well matched between the case and control groups. The mean age was 25 years (16 to 42; Aims
Patients and Methods
The aim of this study was to evaluate cartilaginous patellotrochlear congruence and patellofemoral alignment parameters after deepening trochleoplasty in severe trochlear dysplasia. The study group comprised 20 patients (two male, 18 female; mean age 24 years (16 to 39)) who underwent deepening trochleoplasty and medial patellofemoral ligament (MPFL) reconstruction for the treatment of recurrent lateral patellar dislocation due to severe trochlear dysplasia (Dejour type B to D). Pre- and postoperative MRI investigations of the study group were compared with MRI data of 20 age- and gender-matched control patients (two male, 18 female; mean age 27 years (18 to 44)) regarding the patellotrochlear contact ratio, patellotrochlear contact area, posterior patellar edge-trochlear groove ratio, and patellar tilt.Aims
Patients and Methods
The primary aim of this study was to compare the knee-specific functional outcome of patellofemoral arthroplasty with total knee arthroplasty (TKA) in the management of patients with patellofemoral osteoarthritis. A total of 54 consecutive Avon patellofemoral arthroplasties were identified and propensity-score-matched to a group of 54 patients undergoing a TKA with patellar resurfacing for patellofemoral osteoarthritis. The Oxford Knee Score (OKS), the 12-Item Short-Form Health Survey (SF-12), and patient satisfaction were collected at a mean follow up of 9.2 years (8 to 15). Survival was defined by revision or intention to revise.Aims
Patients and Methods
It has been suggested that mobile-bearing total knee arthroplasty
(TKA) might lead to better outcomes by accommodating some femorotibial
rotational mismatch, thereby reducing contact stresses and polyethylene
wear. The aim of this study was to determine whether there is a
difference between fixed- and mobile-bearing versions of a contemporary
TKA with respect to durability, range of movement (ROM) and function,
ten years postoperatively. A total of 240 patients who were enrolled in this randomized
controlled trial (RCT) underwent a primary cemented TKA with one
of three tibial components (all-polyethylene fixed-bearing, modular
metal-backed fixed-bearing and mobile-bearing). Patients were reviewed
at a median follow-up of ten years (IQR 9.2 to 10.4).Aims
Patients and Methods
The Advance Medial-Pivot total knee arthroplasty (TKA) was designed
to reflect contemporary data regarding the kinematics of the knee.
We wished to examine the long-term results obtained with this prosthesis
by extending a previous evaluation. We retrospectively evaluated prospectively collected data from
225 consecutive patients (41 men and 184 women; mean age at surgery
71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up
of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both
subjective and objective) and radiological outcome were assessed.
Pre- and post-operative clinical and radiographic data were available
at regular intervals for all patients. A total of ten patients (4.4%;
ten TKAs) were lost to follow-up.Aims
Patients and Methods
The use of hinged implants in primary total knee
replacement (TKR) should be restricted to selected indications and mainly
for elderly patients. Potential indications for a rotating hinge
or pure hinge implant in primary TKR include: collateral ligament
insufficiency, severe varus or valgus deformity (>
20°) with necessary
relevant soft-tissue release, relevant bone loss including insertions
of collateral ligaments, gross flexion-extension gap imbalance,
ankylosis, or hyperlaxity. Although data reported in the literature
are inconsistent, clinical results depend on implant design, proper
technical use, and adequate indications. We present our experience
with a specific implant type that we have used for over 30 years
and which has given our elderly patients good mid-term results.
Because revision of implants with long cemented stems can be very
challenging, an effort should be made in the future to use shorter stems
in modular versions of hinged implants. Cite this article:
Valgus knee deformity can present a number of
unique surgical challenges for the total knee arthroplasty (TKA)
surgeon. Understanding the typical patterns of bone and soft-tissue pathology
in the valgus arthritic knee is critical for appropriate surgical
planning. This review aims to provide the knee arthroplasty surgeon
with an understanding of surgical management strategies for the
treatment of valgus knee arthritis. Lateral femoral and tibial deficiencies, contracted lateral soft
tissues, attenuated medial soft tissues, and multiplanar deformities
may all be present in the valgus arthritic knee. A number of classifications
have been reported in order to guide surgical management, and a variety
of surgical strategies have been described with satisfactory clinical
results. Depending on the severity of the deformity, a variety of
TKA implant designs may be appropriate for use. Regardless of an operating surgeon’s preferred surgical strategy,
adherence to a step-wise approach to deformity correction is advised. Cite this article:
It is not clear whether anterior knee pain and osteoarthritis
(OA) of the patellofemoral joint (PFJ) are contraindications to
medial unicompartmental knee arthroplasty (UKA). Our aim was to
investigate the long-term outcome of a consecutive series of patients,
some of whom had anterior knee pain and PFJ OA managed with UKA. We assessed the ten-year functional outcomes and 15-year implant
survival of 805 knees (677 patients) following medial mobile-bearing
UKA. The intra-operative status of the PFJ was documented and, with
the exception of bone loss with grooving to the lateral side, neither
the clinical or radiological state of the PFJ nor the presence of
anterior knee pain were considered a contraindication. The impact
of radiographic findings and anterior knee pain was studied in a
subgroup of 100 knees (91 patients).Aims
Patients and Methods
We present the ten- to 15-year follow-up of 31
patients (34 knees), who underwent an Elmslie-Trillat tibial tubercle osteotomy
for chronic, severe patellar instability, unresponsive to non-operative
treatment. The mean age of the patients at the time of surgery was
31 years (18 to 46) and they were reviewed post-operatively, at
four years (2 to 8) and then at 12 years (10 to 15). All patients
had pre-operative knee radiographs and Cox and Insall knee scores. Superolateral
portal arthroscopy was performed per-operatively to document chondral
damage and after the osteotomy to assess the stability of the patellofemoral
joint. A total of 28 knees (82%) had a varying degree of damage
to the articular surface. At final follow-up 25 patients (28 knees)
were available for review and underwent clinical examination, radiographs
of the knee, and Cox and Insall scoring. Six patients who had no
arthroscopic chondral abnormality showed no or only early signs
of osteoarthritis on final radiographs; while 12 patients with lower
grade chondral damage (grade 1 to 2) showed early to moderate signs
of osteoarthritis and six out of ten knees with higher grade chondral
damage (grade 3 to 4) showed marked evidence of osteoarthritis;
four of these had undergone a knee replacement. In the 22 patients
(24 knees) with complete follow-up, 19 knees (79.2%) were reported
to have a good or excellent outcome at four years, while 15 knees
(62.5%) were reported to have the same at long-term follow-up. The
functional and radiological results show that the extent of pre-operatively
sustained chondral damage is directly related to the subsequent
development of patellofemoral osteoarthritis. Cite this article:
Malrotation of the femoral component can result in post-operative complications in total knee arthroplasty (TKA), including patellar maltracking. Therefore, we used computational simulation to investigate the influence of femoral malrotation on contact stresses on the polyethylene (PE) insert and on the patellar button as well as on the forces on the collateral ligaments. Validated finite element (FE) models, for internal and external malrotations from 0° to 10° with regard to the neutral position, were developed to evaluate the effect of malrotation on the femoral component in TKA. Femoral malrotation in TKA on the knee joint was simulated in walking stance-phase gait and squat loading conditions.Objectives
Materials and Methods
Seven stiff total knee arthroplasties are presented
to illustrate the roles of: 1) manipulation under general anesthesia;
2) multiple concurrent diagnoses in addition to stiffness; 3) extra-articular
pathology; 4) pain as part of the stiffness triad (pain and limits
to flexion or extension); 5) component internal rotation; 6) multifactorial
etiology; and 7) surgical exposure in this challenging clinical
problem.
We have previously reported the short-term radiological
results of a randomised controlled trial comparing kinematically
aligned total knee replacement (TKR) and mechanically aligned TKR,
along with early pain and function scores. In this study we report
the two-year clinical results from this trial. A total of 88 patients
(88 knees) were randomly allocated to undergo either kinematically
aligned TKR using patient-specific guides, or mechanically aligned
TKR using conventional instruments. They were analysed on an intention-to-treat
basis. The patients and the clinical evaluator were blinded to the
method of alignment. At a minimum of two years, all outcomes were better for the kinematically
aligned group, as determined by the mean Oxford knee score (40 (15
to 48) In this study, the use of a kinematic alignment technique performed
with patient-specific guides provided better pain relief and restored
better function and range of movement than the mechanical alignment
technique performed with conventional instruments. Cite this article:
We have investigated the errors in the identification of the transepicondylar axis and the anteroposterior axis between a minimally-invasive and a conventional approach in four fresh-frozen cadaver knees. The errors in aligning the femoral prosthesis were compared with the reference transepicondylar axis as established by CT. The error in the identification of the transepicondylar axis was significantly higher in the minimal approach (4.5° of internal rotation,
A key to the success of revision total knee arthroplasty
(TKA) is a safe surgical approach using an exposure that minimises
complications. In most patients, a medial parapatellar arthrotomy
with complete synovectomy is sufficient. If additional exposure
is needed, a quadriceps snip performed through the quadriceps tendon
often provides the additional exposure required. It is simple to
perform and does not alter the post-operative rehabilitative protocol.
In rare cases, in which additional exposure is needed, or when removal
of a cemented long-stemmed tibial component is required, a tibial
tubercle osteotomy (TTO) may be used. Given the risk of post-operative
extensor lag, a V-Y quadricepsplasty is rarely indicated and usually
considered only if TTO is not possible. Cite this article:
The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint. The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use.
Instability is a common indication for early
revision after both primary and revision total knee arthroplasty
(TKA), accounting for up to 20% in the literature. The number of
TKAs performed annually continues to climb exponentially, thus having
an effective algorithm for treatment is essential. This relies on
a thorough pre- and intra-operative assessment of the patient. The
underlying cause of the instability must be identified initially
and subsequently, the surgeon must be able to balance the flexion
and extension gaps and be comfortable using a variety of constrained
implants. This review describes the assessment of the unstable TKA, and
the authors’ preferred form of treatment for these difficult cases
where the source of instability is often multifactorial. Cite this article:
Malrotation of the femoral component is a cause of patellofemoral maltracking after total knee arthroplasty. Its precise effect on the patellofemoral mechanics has not been well quantified. We have developed an in vitro method to measure the influence of patellar maltracking on contact. Maltracking was induced by progressively rotating the femoral component either internally or externally. The contact mechanics were analysed using Tekscan. The results showed that excessive malrotation of the femoral component, both internally and externally, had a significant influence on the mechanics of contact. The contact area decreased with progressive maltracking, with a concomitant increase in contact pressure. The amount of contact area that carries more than the yield stress of ultra-high molecular weight polyethylene significantly increases with progressive maltracking. It is likely that the elevated pressures noted in malrotation could cause accelerated and excessive wear of the patellar button.
We describe the mid-term results of a prospective study of total knee replacement in severe valgus knees using an osteotomy of the lateral femoral condyle and computer navigation. There were 15 knees with a mean valgus deformity of 21° (17° to 27°) and a mean follow-up of 28 months (24 to 60). A cemented, non-constrained fixed bearing, posterior-cruciate-retaining knee prosthesis of the same design was used in all cases (Columbus-B. Braun; Aesculap, Tuttlingen, Germany). All the knees were corrected to a mean of 0.5° of valgus (0° to 2°). Flexion of the knee had been limited to a mean of 85° (75° to 110°) pre-operatively and improved to a mean of 105° (90° to 130°) after operation. The mean Knee Society score improved from 37 (30 to 44) to 90 points (86 to 94). Osteotomy of the lateral femoral condyle combined with computer-assisted surgery gave an excellent mid-term outcome in patients undergoing total knee replacement in the presence of severe valgus deformity.
Component malalignment can be associated with
pain following total knee replacement (TKR). Using MRI, we reviewed
50 patients with painful TKRs and compared them with a group of
16 asymptomatic controls to determine the feasibility of using MRI
in evaluating the rotational alignment of the components. Using
the additional soft-tissue detail provided by this modality, we
also evaluated the extent of synovitis within these two groups.
Angular measurements were based on the femoral transepicondylar
axis and tibial tubercle. Between two observers, there was very
high interobserver agreement in the measurements of all values.
Patients with painful TKRs demonstrated statistically significant
relative internal rotation of the femoral component (p = 0.030).
There was relative internal rotation of the tibial to femoral component
and combined excessive internal rotation of the components in symptomatic
knees, although these results were significant only with one of
the observers (p = 0.031). There was a statistically significant
association between the presence and severity of synovitis and painful
TKR (p <
0.001). MRI is an effective modality in evaluating component rotational
alignment.
Patellofemoral complications are common after
total knee replacement (TKR). Leaving the patellar unsurfaced after
TKR may lead to complications such as anterior knee pain, and re-operation
to surface it. Complications after patellar resurfacing include
patellar fracture, aseptic loosening, patellar instability, polyethylene
wear, patellar clunk and osteonecrosis. Historically, patellar complications
account for one of the larger proportions of causes of failure in
TKR, however, with contemporary implant designs, complication rates
have decreased. Most remaining failures relate to patellofemoral
tracking. Understanding the causes of patellofemoral maltracking
is essential to prevent these complications as well as manage them
when they occur. Cite this article:
A series of 100 consecutive osteoarthritic patients was randomised to undergo total knee replacement using a Miller-Galante II prosthesis, with or without a cemented polyethylene patellar component. Knee function was evaluated using the American Knee Society score, Western Ontario and McMaster University Osteoarthritis index, specific patellofemoral-related questions and radiographic evaluation until the fourth post-operative year, then via questionnaire until ten years post-operatively. A ten-point difference in the American Knee Society score between the two groups was considered a significant change in knee performance, with α and β levels of 0.05. The mean age of the patients in the resurfaced group was 71 years (53 to 88) and in the non-resurfaced group was 73 years (54 to 86). After ten years 22 patients had died, seven were suffering from dementia, three declined further participation and ten were lost to follow-up. Two patients in the non-resurfaced group subsequently had their patellae resurfaced. In the resurfaced group one patient had an arthroscopic lateral release. There was no significant difference between the two treatment groups: both had a similar deterioration of scores with time, and no further patellofemoral complications were observed in either group. We are unable to recommend routine patellar resurfacing in osteoarthritic patients undergoing total knee replacement on the basis of our findings.
We used three-dimensional movement analysis by computer modelling of knee flexion from 0° to 50° in 14 knees in 12 patients with recurrent patellar dislocation and in 15 knees in ten normal control subjects to compare the The patients had greater values of spin from 20° to 50°, while there were no statistically significant differences in flexion and tilt. The patients also had greater percentage patellar shift from 0° to 50°, percentage tubercle shift at 0° and 10° and patellar inclination from 0° to 50° with a smaller oval-shaped contact area from 20° to 50° moving downwards on the lateral facet. Patellar movement analysis using a three-dimensional computer model is useful to clearly demonstrate differences between patients with recurrent dislocation of the patella and normal control subjects.
The management of osteoarthritis of the knee associated with patellar instability secondary to external tibial torsion >
45° is challenging. Patellofemoral biomechanics in these patients cannot be achieved by intra-articular correction using standard techniques of total knee replacement. We reviewed seven patients (eight knees) with recurrent patellar dislocation and one with bilateral irreducible lateral dislocation who had undergone simultaneous total knee replacement and internal tibial derotational osteotomy. All had osteoarthritis and severe external tibial torsion. The mean follow-up was for 47.2 months (24 to 120). The mean objective and functional Knee Society scores improved significantly (p = 0.0001) from 29.7 and 41.5 pre-operatively to 71.4 and 73.5 post-operatively, respectively. In all patients the osteotomies healed and patellar stability was restored. Excessive external tibial torsion should be identified and corrected in patients with osteoarthritis and patellar instability. Simultaneous internal rotation osteotomy of the tibia and total knee replacement is a technically demanding but effective treatment for such patients.
We have reviewed the literature to establish the role of lateral retinacular release in the management of disorders of the extensor apparatus of the knee. The scientific evidence for intervention is explored and reports on outcome are discussed.
This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1). The PFC Sigma knee system appears to provide excellent results in the medium term.
The requirement for release of collateral ligaments to achieve a stable, balanced total knee replacement has been reported to arise in about 50% to 100% of procedures. This wide range reflects a lack of standardised quantitative indicators to determine the necessity for a release. Using recent advances in computerised navigation, we describe two navigational predictors which provide quantitative measures that can be used to identify the need for release. The first was the ability to restore the mechanical axis before any bone resection was performed and the second was the discrepancy in the measured medial and lateral joint spaces after the tibial osteotomy, but before any femoral resection. These predictors showed a significant association with the need for collateral ligament release (p <
0.001). The first predictor using the knee stress test in extension showed a sensitivity of 100% and a specificity of 98% and the second, the difference between medial and lateral gaps in millimetres, a sensitivity of 83% and a specificity of 95%. The use of the two navigational predictors meant that only ten of the 93 patients required collateral ligament release to achieve a stable, neutral knee.
We undertook a study in which 138 female patients with a mean age of 71.2 years (51 to 82) received a standard NexGen CR-flex prosthesis in one knee and a gender-specific NexGen CR-flex prosthesis in the other. The mean follow-up period was 3.25 years (3.1 to 3.5). The aspect ratios of the standard and gender-specific prostheses were compared with that of the distal femur. The mean post-operative Knee Society knee scores were 94 (70 to 100) and 93 (70 to 100) points and the function scores were 83 (60 to 100) and 84 (60 to 100) points for the standard implants and the gender-specific designs, respectively. The mean post-operative Western Ontario and McMaster Universities score was 26.4 points (0 to 76). Patient satisfaction, the radiological results and the complication rates were similar in the two groups. In those with a standard prosthesis, the femoral component was closely matched in 80 knees (58.0%), overhung in 14 (10.1%) and undercovered the bone in 44 (31.9%). In those with a gender-specific prosthesis, it was closely matched in 15 knees (10.9%) and undercovered the bone in 123 (89.1%). Since we found no significant differences between the two groups with regard to the clinical and radiological results, patient satisfaction or complication rate, the goal of the design of the gender-specific CR-flex prosthesis to improve the outcome was not achieved in our patients.
Substantial healthcare resources have been devoted
to computer navigation and patient-specific instrumentation systems
that improve the reproducibility with which neutral mechanical alignment
can be achieved following total knee replacement (TKR). This choice of
alignment is based on the long-held tenet that the alignment of
the limb post-operatively should be within 3° of a neutral mechanical
axis. Several recent studies have demonstrated no significant difference
in survivorship when comparing well aligned Review of the literature suggests that a neutral mechanical axis
remains the optimal guide to alignment. Cite this article:
Pain is the main indication for performing total knee replacement (TKR). In most patients after TKR there is an improvement, but a few continue to have pain. Generally, the cause of the pain can be addressed when it is identified. However, unexplained pain can be more difficult to manage because revision surgery is likely to be unrewarding in this group. In our study of 622 cemented TKRs in 512 patients with a mean age of 69 years (23 to 90) treated between January 1995 and August 1998, we identified 24 patients (knees) with unexplained pain at six months. This group was followed for five years (data was unavailable for 18 knees) and ten patients (55.5%) went on to show an improvement without intervention. In the case of unexplained pain, management decisions must be carefully considered, but reassurance can be offered to patients that the pain will improve in more than half with time.
We describe our experience with the ‘four-in-one’ procedure for habitual dislocation of the patella in five children (six knees). All the patients presented with severe generalised ligamentous laxity and aplasia of the trochlear groove. All had a lateral release, proximal ‘tube’ realignment of the patella, semitendinosus tenodesis and transfer of the patellar tendon. The mean age at the time of the operation was 6.1 years (4.9 to 6.9), and the patients were followed up for a mean of 54.5 months (31 to 66). The clinical results were evaluated using the Kujala score. There has been no recurrence of dislocation. All the patients have returned to full activities and the parents and children were satisfied with the clinical results. The mean Kujala score was 95.3 (88 to 98). Two patients had marginal skin necrosis which healed after debridement and secondary closure. These early results in this small group have shown that the ‘four-in-one’ procedure is effective in the treatment of obligatory dislocation of the patella in children with severe ligamentous laxity and trochlear aplasia.
At least four ways have been described to determine
femoral component rotation, and three ways to determine tibial component
rotation in total knee replacement (TKR). Each method has its advocates
and each has an influence on knee kinematics and the ultimate short
and long term success of TKR. Of the four femoral component methods,
the author prefers rotating the femoral component in flexion to
that amount that establishes a stable symmetrical flexion gap. This
judgement is made after the soft tissues of the knee have been balanced
in extension. Of the three tibial component methods, the author prefers rotating
the tibial component into congruency with the established femoral
component rotation with the knee is in extension. This yields a
rotationally congruent articulation during weight-bearing and should
minimise the torsional forces being transferred through a conforming tibial
insert, which could lead to wear to the underside of the tibial
polyethylene. Rotating platform components will compensate for any
mal-rotation, but can still lead to pain if excessive tibial insert
rotation causes soft-tissue impingement. Cite this article:
Isolated patellofemoral arthritis is a common
condition and there are varying opinions on the most effective treatments.
Non-operative and operative treatments have failed to demonstrate
effective long-term treatment for those in an advanced stage of
the condition. Newer designs and increased technology in patellofemoral replacement
(PFR) have produced more consistent outcomes. This has led to a
renewed enthusiasm for this procedure. Newer PFR prostheses have
addressed the patellar maltracking issues plaguing some of the older designs.
Short-term results with contemporary prostheses and new technology
are described here. Cite this article:
The Oxford medial unicompartmental knee replacement was designed to reproduce normal mobility and forces in the knee, but its detailed effect on the patellofemoral joint has not been studied previously. We have examined the effect on patellofemoral mechanics of the knee by simultaneously measuring patellofemoral kinematics and forces in 11 cadaver knee specimens in a supine leg-extension rig. Comparison was made between the intact normal knee and sequential unicompartmental and total knee replacement. Following medial mobile-bearing unicompartmental replacement in 11 knees, patellofemoral kinematics and forces did not change significantly from those in the intact knee across any measured parameter. In contrast, following posterior cruciate ligament retaining total knee replacement in eight knees, there were significant changes in patellofemoral movement and forces. The Oxford device appears to produce near-normal patellofemoral mechanics, which may partly explain the low incidence of complications with the extensor mechanism associated with clinical use.
We report the mid-term results of a new patellofemoral arthroplasty for established isolated patellofemoral arthritis. We have reviewed the experience of 109 consecutive patellofemoral resurfacing arthroplasties in 85 patients who were followed up for at least five years. The five-year survival rate, with revision as the endpoint, was 95.8% (95% confidence interval 91.8% to 99.8%). There were no cases of loosening of the prosthesis. At five years the median Bristol pain score improved from 15 of 40 points (interquartile range 5 to 20) pre-operatively, to 35 (interquartile range 20 to 40), the median Melbourne score from 10 of 30 points (interquartile range 6 to 15) to 25 (interquartile range 20 to 29), and the median Oxford score from 18 of 48 points (interquartile range 13 to 24) to 39 (interquartile range 24 to 45). Successful results, judged on a Bristol pain score of at least 20 at five years, occurred in 80% (66) of knees. The main complication was radiological progression of arthritis, which occurred in 25 patients (28%) and emphasises the importance of the careful selection of patients. These results give increased confidence in the use of patellofemoral arthroplasty.
An abnormal lateral position of the tibial tuberosity causes distal malalignment of the extensor mechanism of the knee and can lead to lateral tracking of the patella causing anterior knee pain or objective patellar instability, characterised by recurrent dislocation. Computer tomography is used for a precise pre-operative assessment of the tibial tubercle-trochlear groove distance. A distance of more than 15 mm is considered to be pathological and an indication for surgery in symptomatic patients. In a prospective study we performed a subtle transfer of the tibial tuberosity according to the information gained from the pre-operative CT scan. This method was applied to two groups of patients, those with painful lateral tracking of the patella, and those with objective patellar instability. We evaluated the clinical results in 30 patients in each group. The outcome was documented at 3, 12 and 24 months using the Lysholm scale, the Kujala score, and a visual analogue pain score. Post-operatively, all but one patient in the instability group who had a patellar dislocation requiring further surgery reported good improvement with no further subluxation or dislocation. All patients in both groups had a marked improvement in pain and functional score. Two patients sustained a tibial fracture six and seven weeks after surgery. One patient suffered a per-operative fracture of the tibial tubercle which later required further fixation. If carefully performed, this type of transfer of the tibial tubercle appears to be a satisfactory technique for the treatment of patients with an increased tibial tubercle-trochlear groove distance and who present with symptoms related to lateral maltracking of the patella.
The October 2013 Knee Roundup360 looks at: Make it easy, release the MCL; Do patients remember clinical information in day surgery?; Osteoarthritis and arthroscopy?; How best to double your bundles; When to operate for infection; Cementless unicompartment knee replacement?; Tibial tubercle-trochlear groove confusion; Tarts, cherries and osteoarthritis
We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.
The appearance of the ‘grand-piano sign’ on the anterior resected surface of the femur has been considered to be a marker for correct femoral rotational alignment during total knee replacement. Our study was undertaken to assess quantitatively the morphological patterns on the resected surface after anterior femoral resection with various angles of external rotation, using a computer-simulation technique. A total of 50 right distal femora with varus osteoarthritis in 50 Korean patients were scanned using computerised tomography. Computer image software was used to simulate the anterior femoral cut, which was applied at an external rotation of 0°, 3° and 6° relative to the posterior condylar axis, and parallel to the surgical and clinical epicondylar axes in each case. The morphological patterns on the resected surface were quantified and classified as the ‘grand-piano sign’, ‘the boot sign’ and the ‘butterfly sign’. The surgeon can use the analogy of these quantified sign patterns to ensure that a correct rotational alignment has been obtained intra-operatively.
Disruption of the extensor mechanism in total
knee arthroplasty may occur by tubercle avulsion, patellar or quadriceps
tendon rupture, or patella fracture, and whether occurring intra-operatively
or post-operatively can be difficult to manage and is associated
with a significant rate of failure and associated complications.
This surgery is frequently performed in compromised tissues, and
repairs must frequently be protected with cerclage wiring and/or
augmentation with local tendon (semi-tendinosis, gracilis) which
may also be used to treat soft-tissue loss in the face of chronic
disruption. Quadriceps rupture may be treated with conservative
therapy if the patient retains active extension. Component loosening
or loss of active extension of 20° or greater are clear indications
for surgical treatment of patellar fracture. Acute patellar tendon
disruption may be treated by primary repair. Chronic extensor failure
is often complicated by tissue loss and retraction can be treated
with medial gastrocnemius flaps, achilles tendon allografts, and
complete extensor mechanism allografts. Attention to fixing the
graft in full extension is mandatory to prevent severe extensor
lag as the graft stretches out over time.
Hypermobility is an acknowledged risk factor
for patellar instability. In this case control study the influence
of hypermobility on clinical outcome following medial patellofemoral
ligament (MPFL) reconstruction for patellar instability was studied. A total of 25 patients with hypermobility as determined by the
Beighton criteria were assessed and compared with a control group
of 50 patients who were matched for age, gender, indication for
surgery and degree of trochlear dysplasia. The patients with hypermobility
had a Beighton Score of ≥ 6; the control patients had a score of <
4. All patients underwent MPFL reconstruction performed using semitendinosus
autograft and a standardised arthroscopically controlled technique.
The mean age of the patients was 25 years (17 to 49) and the mean
follow-up was 15 months (6 to 30). Patients with hypermobility had a significant improvement in
function following surgery, with reasonable rates of satisfaction,
perceived improvement, willingness to repeat and likelihood of recommendation.
Functional improvements were significantly less than in control
patients (p <
0.01). Joint hypermobility is not a contraindication to MPFL reconstruction
although caution is recommended in managing the expectations of
patients with hypermobility before consideration of surgery.
We evaluated the rates of survival and cause
of revision of seven different brands of cemented primary total
knee replacement (TKR) in the Norwegian Arthroplasty Register during
the years 1994 to 2009. Revision for any cause, including resurfacing
of the patella, was the primary endpoint. Specific causes of revision
were secondary outcomes. Three posterior cruciate-retaining (PCR) fixed modular-bearing
TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing
posterior cruciate-sacrificing TKRs were investigated in a total
of 17 782 primary TKRs. The median follow-up for the implants ranged
from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from
89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative
to the fixed modular-bearing Profix knee (the most frequently used
TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision
for aseptic tibial loosening was higher in the mobile-bearing LCS
Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the
LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing
Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular
bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared
with the Profix. These implants (except AGC Universal) also had
an increased risk of revision for femoral loosening (RR 2.3
(95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95%
CI 1.1 to 11.0), respectively). These results suggest that aseptic
loosening is related to design in TKR. Cite this article:
The incidence of anterior knee pain following
total knee replacement (TKR) is reported to be as high as 49%. The source
of the pain is poorly understood but the soft tissues around the
patella have been implicated. In theory circumferential electrocautery denervates the patella
thereby reducing efferent pain signals. However, there is mixed
evidence that this practice translates into improved outcomes. We aimed to investigate the clinical effect of intra-operative
circumpatellar electrocautery in patients undergoing TKR using the
LCS mobile bearing or Kinemax fixed bearing TKR. A total of 200
patients were randomised to receive either circumpatellar electrocautery
(diathermy) or not (control). Patients were assessed by visual analogue
scale (VAS) for anterior knee pain and Oxford knee score (OKS) pre-operatively
and three months, six months and one year post-operatively. Patients
and assessors were blinded. There were 91 patients in the diathermy group and 94 in the control.
The mean VAS improvement at one year was 3.9 in both groups (control;
-10 to 6, diathermy; We found no relevant effect of patellar electrocautery on either
VAS anterior knee pain or OKS for patients undergoing LCS and Kinemax
TKR.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
In posterior stabilised total knee replacement
(TKR) a larger femoral component is sometimes selected to manage the
increased flexion gap caused by resection of the posterior cruciate
ligament. However, concerns remain regarding the adverse effect
of the increased anteroposterior dimensions of the femoral component
on the patellofemoral (PF) joint. Meanwhile, the gender-specific
femoral component has a narrower and thinner anterior flange and
is expected to reduce the PF contact force. PF contact forces were
measured at 90°, 120°, 130° and 140° of flexion using the NexGen
Legacy Posterior Stabilized (LPS)-Flex Fixed Bearing Knee system
using Standard, Upsized and Gender femoral components during TKR.
Increasing the size of the femoral component significantly increased
mean PF forces at 120°, 130° and 140° of flexion (p = 0.005, p <
0.001 and p <
0.001, respectively). No difference was found in
contact force between the Gender and the Standard components. Among
the patients who had overhang of the Standard component, mean contact
forces with the Gender component were slightly lower than those
of the Standard component, but no statistical difference was found
at 90°, 120°, 130° or 140° of flexion (p = 0.689, 0.615, 0.253 and
0.248, respectively). Upsized femoral components would increase PF forces in deep knee
flexion. Gender-specific implants would not reduce PF forces.
The lateral compartment is predominantly affected
in approximately 10% of patients with osteoarthritis of the knee. The
anatomy, kinematics and loading during movement differ considerably
between medial and lateral compartments of the knee. This in the
main explains the relative protection of the lateral compartment
compared with the medial compartment in the development of osteoarthritis.
The aetiology of lateral compartment osteoarthritis can be idiopathic,
usually affecting the femur, or secondary to trauma commonly affecting
the tibia. Surgical management of lateral compartment osteoarthritis
can include osteotomy, unicompartmental knee replacement and total
knee replacement. This review discusses the biomechanics, pathogenesis
and development of lateral compartment osteoarthritis and its management. Cite this article:
Complications involving the patellofemoral joint,
caused by malrotation of the femoral component during total knee replacement,
are an important cause of persistent pain and failure leading to
revision surgery. The aim of this study was to determine and quantify
the influence of femoral component malrotation on patellofemoral
wear, and to determine whether or not there is a difference in the
rate of wear of the patellar component when articulated against
oxidised zirconium (OxZr) and cobalt-chrome (CoCr) components. An The results suggest that patellar maltracking due to an internally
rotated femoral component leads to an increased mean patellar wear.
Although not statistically significant, the mean wear production
may be lower for OxZr than for CoCr components.
Correct positioning and alignment of components during primary total knee replacement (TKR) is widely accepted to be an important predictor of patient satisfaction and implant durability. This retrospective study reports the effect of the post-operative mechanical axis of the lower limb in the coronal plane on implant survival following primary TKR. A total of 501 TKRs in 396 patients were divided into an aligned group with a neutral mechanical axis (± 3°) and a malaligned group where the mechanical axis deviated from neutral by >
3°. At 15 years’ follow-up, 33 of 458 (7.2%) TKRs were revised for aseptic loosening. Kaplan-Meier survival analysis showed a weak tendency towards improved survival with restoration of a neutral mechanical axis, but this did not reach statistical significance (p = 0.47). We found that the relationship between survival of a primary TKR and mechanical axis alignment is weaker than that described in a number of previous reports.
The efficacy of circumpatellar electrocautery in reducing the incidence of post-operative anterior knee pain is unknown. We conducted a single-centre, outcome-assessor and patient-blinded, parallel-group, randomised, controlled trial to compare circumpatellar electrocautery with no electrocautery in total knee replacement in the absence of patellar resurfacing. Patients requiring knee replacement for primary osteoarthritis were randomly assigned circumpatellar electrocautery (intervention group) or no electrocautery (control group). The primary outcome measure was the incidence of anterior knee pain. A secondary measure was the standardised clinical and patient-reported outcomes determined by the American Knee Society scores and the Western Ontario and McMaster Universities (WOMAC) osteoarthritis index. A total of 131 knees received circumpatellar electrocautery and 131 had no electrocautery. The overall incidence of anterior knee pain at follow-up at one year was 26% (20% to 31%), with 19% (12% to 26%) in the intervention group and 32% (24% to 40%) in the control group (p = 0.02). The relative risk reduction from electrocautery was 40% (9% to 61%) and the number needed to treat was 7.7 (4.3 to 41.4). The intervention group had a better mean total WOMAC score at follow-up at one year compared with the control group (16.3 (0 to 77.7) Our study suggests that in the absence of patellar resurfacing electrocautery around the margin of the patella improves the outcome of total knee replacement.
Among the variety of differential diagnoses for chronic patellar tendinopathy, isolated tuberculosis is extremely rare. We report such a case, without any evident primary contiguous or distant focus, in a 31-year-old immunocompetent male.
Isolated patellofemoral osteoarthritis can be a disabling disease. When conservative treatment fails, surgical options can be unpredictable and may be considered too aggressive for middle-aged and active people. We analysed the clinical and radiological results of a new coronal osteotomy involving thinning of the patella in a selected group of patients with isolated patellofemoral osteoarthritis. Since 1991, 31 patients (35 knees) have been treated, of whom 34 were available for follow-up at a mean of 9.1 years. The Knee Society Score, the Patellar score and the Short-form-36 questionnaire were used for clinical evaluation. We also examined the radiological features to confirm bone consolidation and assess the progression of osteoarthritis. A significant improvement in the functional scores and radiological parameters was noted. All patients except one were satisfied with the operation. Radiological progression of the patellofemoral osteoarthritis was slowed but radiological femorotibial osteoarthritis progressed in 23 (65%) cases, with a total knee replacement becoming necessary in four cases without technical problems in resurfacing the patella. We compared the results with other forms of surgical treatment reported in the literature. This treatment offers good clinical and radiological results, presenting an alternative method of managing patellofemoral osteoarthritis.
Segmental resection of malignant bone disease in the femoral diaphysis with subsequent limb reconstruction is a major undertaking. This is a retrospective review of 23 patients who had undergone limb salvage by endoprosthetic replacement of the femoral diaphysis for a primary bone tumour between 1989 and 2005. There were 16 males and seven females, with a mean age of 41.3 years (10 to 68). The mean overall follow-up was for 97 months (3 to 240), and 120 months (42 to 240) for the living patients. The cumulative patient survival was 77% (95% confidence interval 63% to 95%) at ten years. Survival of the implant, with failure of the endoprosthesis as an endpoint, was 85% at five years and 68% (95% confidence interval 42% to 92%) at ten years. The revision rate was 22% and the overall rate of re-operation was 26%. Complications included deep infection (4%), breakage of the prosthesis (8%), periprosthetic fracture (4%), aseptic loosening (4%), local recurrence (4%) and metastases (17%). The 16 patients who retained their diaphyseal endoprosthesis had a mean Musculoskeletal Tumour Society score of 87% (67% to 93%). They were all able to comfortably perform most activities of daily living. Femoral diaphyseal endoprosthetic replacement is a viable option for reconstruction following segmental resection of malignant bone disease. It allows immediate weight-bearing, is associated with a good long-term functional outcome, has an acceptable complication and revision rate and, most importantly, does not appear to compromise patient survival.
We carried out three total knee replacements with proximal realignment in two patients with severe osteoarthritis of the knee and congenital dislocation of the patella. During the operation, the femur and the tibia were cut according to the recommendations of the manufacturer of the implant. The femoral component was placed in external rotation and the centre of the tibial component aligned in relation to the tibial tuberosity with regard to rotation and translation. After making the bone cuts, the iliotibial tract was detached from Gerdy’s tubercle, the popliteus tendon divided, and the biceps femoris tendon elongated by Z-plasty. After the trial implants were positioned, a proximal re-alignment procedure was performed. One knee had deep infection. There was one dislocated patella which was repositioned, and walking ability was improved in all knees. We believe that the rotational alignment of a prosthesis is as important as the soft-tissue surgery.
This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3° of internal rotation (25.4° internal to 13.9° external rotation) in the painful group and 2.2° of external rotation (8.5° internal to 18.2° external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9° compared with none in the painfree group (p <
0.001). Additionally, six femoral components in the painful group were internally rotated more than 6° compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.
We present a 25-year-old patient with juvenile rheumatoid arthritis and ankylosis of both hips and both knees treated by staged bilateral hip and knee arthroplasty. She was followed up for 18 months. We discuss the pre-operative planning, surgical details and post-operative rehabilitation.
We describe a cohort of patients with a high rate of mid-term failure following Kinemax Plus total knee replacement inserted between 1998 and 2001. This implant has been recorded as having a survival rate of 96% at ten years. However, in our series the survival rate was 75% at nine years. This was also significantly lower than that of subsequent consecutive series of PFC Sigma knee replacements performed by the same surgeon. No differences were found in the clinical and radiological parameters between the two groups. At revision the most striking finding was polyethylene wear. An independent analysis of the polyethylene components was therefore undertaken. Scanning electron microscopy revealed type 2 fusion defects in the ultra-high molecular weight polyethylene (UHMWPE), which indicated incomplete boundary fusion. Other abnormalities consistent with weak UHMWPE particle interface strength were present in both the explanted inserts and in unused inserts from the same period. We consider that these type 2 fusion defects are the cause of the early failure of the Kinemax implants. This may represent a manufacturing defect resulting in a form of programmed polyethylene failure.
The management of patients with a painful total knee replacement requires careful assessment and a stepwise approach in order to diagnose the underlying pathology accurately. The management should include a multidisciplinary approach to the patient’s pain as well as addressing the underlying aetiology. Pain should be treated with appropriate analgesia, according to the analgesic ladder of the World Health Organisation. Special measures should be taken to identify and to treat any neuropathic pain. There are a number of intrinsic and extrinsic causes of a painful knee replacement which should be identified and treated early. Patients with unexplained pain and without any recognised pathology should be treated conservatively since they may improve over a period of time and rarely do so after a revision operation.
We investigated the clinical and radiological outcome of trochleaplasty for recurrent patellar dislocation in association with trochlear dysplasia in 38 consecutive patients (45 knees) with a mean follow-up of 8.3 years (4 to 14). None had recurrence of dislocation after trochleaplasty. Post-operatively, patellofemoral pain, present pre-operatively in only 35 knees, became worse in 15 (33.4%), remained unchanged in four (8.8%) and improved in 22 (49%). Four knees which had no pain pre-operatively (8.8%) continued to have no pain. A total of 33 knees were available for radiological assessment. Post-operatively, all but two knees (93.9%) had correction of trochlear dysplasia radiologically but degenerative changes of the patellofemoral joint developed in 30% (10) of the knees. We conclude that recurrent patellar dislocation associated with trochlear dysplasia can be treated successfully by trochleaplasty, but the impact on patellofemoral pain and the development of patellofemoral osteoarthritis is less predictable. Overall, subjective patient satisfaction with restored patellofemoral stability after trochleaplasty appeared to outweigh its possible sequelae.
The rotational alignment of the tibia is an unresolved issue in knee replacement. A poor functional outcome may be due to malrotation of the tibial component. Our aim was to find a reliable method for positioning the tibial component in knee replacement. CT scans of 19 knees were reconstructed in three dimensions and orientated vertically. An axial plane was identified 20 mm below the tibial spines. The centre of each tibial condyle was calculated from ten points taken round the condylar cortex. The tibial tubercle centre was also generated as the centre of the circle which best fitted eight points on the outside of the tubercle in an axial plane at the level of its most prominent point. The derived points were identified by three observers with errors of 0.6 mm to 1 mm. The medial and lateral tibial centres were constant features (radius 24 mm ( Alignment of the knee when based on this anatomical axis was more reliable than either the posterior surfaces or any axis involving the tubercle which was the least reliable landmark in the region.
Achieving deep flexion after total knee replacement remains a challenge. In this study we compared the soft-tissue tension and tibiofemoral force in a mobile-bearing posterior cruciate ligament-sacrificing total knee replacement, using equal flexion and extension gaps, and with the gaps increased by 2 mm each. The tests were conducted during passive movement in five cadaver knees, and measurements of strain were made simultaneously in the collateral ligaments. The tibiofemoral force was measured using a customised mini-force plate in the tibial tray. Measurements of collateral ligament strain were not very sensitive to changes in the gap ratio, but tibiofemoral force measurements were. Tibiofemoral force was decreased by a mean of 40% (
Stiffness is an uncommon but potentially debilitating complication following total knee replacement (TKR). The treatment of this condition remains difficult and controversial. We present the results of 13 patients who underwent open arthrolysis for stiffness. The mean time between TKR and arthrolysis was 14 months. The mean follow-up was 7.2 years (2 to 10). The mean range of movement prior to arthrolysis was 55°. This increased to 91°, six months after arthrolysis (p <
0.005). The improved range of movement was maintained during the follow-up period. No patient has required revision of their components. We have found arthrolysis to be a useful and successful approach to post-TKR stiffness.
We prospectively assessed the benefits of using either a range-of-movement technique or an anatomical landmark method to determine the rotational alignment of the tibial component during total knee replacement. We analysed the cut proximal tibia intraoperatively, determining anteroposterior axes by the range-of-movement technique and comparing them with the anatomical anteroposterior axis. We found that the range-of-movement technique tended to leave the tibial component more internally rotated than when anatomical landmarks were used. In addition, it gave widely variable results (mean 7.5°; 2° to 17°), determined to some extent by which posterior reference point was used. Because of the wide variability and the possibilities for error, we consider that it is inappropriate to use the range-of-movement technique as the sole method of determining alignment of the tibial component during total knee replacement.
Evaluation of patients with painful total knee replacement requires a thorough clinical examination and relevant investigations in order to reach a diagnosis. Awareness of the common and uncommon problems leading to painful total knee replacement is useful in the diagnostic approach. This review article aims to act as a guide to the evaluation of patients with painful total knee replacement.
The clinical results of bilateral total knee replacement staged at a one-week interval during a single hospital admission were compared with bilateral total knee replacements performed under the same anaesthetic and with bilateral total knee replacements performed during two separate admissions. The data were retrospectively reviewed. All operations had been performed by the same surgeon using the same design of prosthesis at a single institution. The operative time and length of stay for the one-week staged group were comparable with those of the separate admission group but longer than for the patients treated under one anaesthetic. There was a low rate of complications and good clinical outcome in all groups at a mean follow-up of four years (1 to 7.2). The group staged at a one-week interval had the least blood loss (p = 0.004). With appropriate patient selection, bilateral total knee replacement performed under a single anaesthetic, or staged at a one-week interval, is a safe and effective method to treat bilateral arthritis of the knee.
We have examined the differences in clinical outcome of total knee replacement (TKR) with and without patellar resurfacing in a prospective, randomised study of 181 osteoarthritic knees in 142 patients using the Profix total knee system which has a femoral component with features considered to be anatomical and a domed patellar implant. The procedures were carried out between February 1998 and November 2002. A total of 159 TKRs in 142 patients were available for review at a mean of four years (3 to 7). The patients and the clinical evaluator were blinded in this prospective study. Evaluation was undertaken annually by an independent observer using the knee pain scale and the Knee Society clinical rating system. Specific evaluation of anterior knee pain, stair-climbing and rising from a seated to a standing position was also undertaken. No benefit was shown of TKR with patellar resurfacing over that without resurfacing with respect to any of the measured outcomes. In 22 of 73 knees (30.1%) with and 18 of 86 knees (20.9%) without patellar resurfacing there was some degree of anterior knee pain (p = 0.183). No revisions related to the patellofemoral joint were performed in either group. Only one TKR in each group underwent a re-operation related to the patellofemoral joint. A significant association between knee flexion contracture and anterior knee pain was observed in those knees with patellar resurfacing (p = 0.006).