Advertisement for orthosearch.org.uk
Results 1 - 20 of 36
Results per page:
The Bone & Joint Journal
Vol. 107-B, Issue 2 | Pages 181 - 187
1 Feb 2025
Barret H Tiercelin J Godenèche A Charousset C Audebert S Lefebvre Y Gallinet D Barth J Bonnevialle N

Aims. Loosening of the glenoid component in the long term remains an important complication of the anatomical total shoulder arthroplasty (aTSA). The aim of this study was to explore the bony integration of a hybrid glenoid component based on an analysis of CT scans. Methods. In a prospective multicentre study, patients who underwent primary aTSA, whose hybrid design of glenoid component included a fully-polyethylene flanged upper peg and a porous-coated titanium lower peg, and who were reviewed with CT scans between 12 and 24 months postoperatively, were included. Two independent observers reviewed the scans. Bony integration of the upper peg was scored as described by Arnold et al, and integration of the lower peg was scored as described by Gulotta et al. Perforation of the glenoid vault in any plane was also assessed. Results. From an initial group of 120 aTSAs in 116 patients, 104 CT scans were analyzed in 100 patients (four bilateral shoulders, mean age 66 years (SD 11), 62 female and 38 male). Osteolysis around the upper peg was found in 32 patients (32 aTSAs; 31%). Of the remaining patients, 72 had a mean Arnold score of 5.6 points (SD 0.9), and 70 (67%) had perfect integration. The lower peg had a mean Gulotta score of 6.5 points (SD 1.4). There was perfect integration of the lower peg in 70 patients (70 aTSAs; 67%). A total of nine patients (nine aTSAs; 9%) had no bony integration at either peg. There was perforation of the glenoid in an anterior or posterior direction at the level of the upper peg in three and 28 patients, respectively. This occurred at the level of the lower peg in 11 and 18 patients, respectively. The inter- and intraobserver reliability was good (k = 0.782 and 0.86, respectively). No implant breakage occurred at a mean follow-up of 16 months (12 to 24). The clinical outcome was satisfactory at a mean follow-up of 32 months (24 to 35), as assessed by a visual analogue scale score for pain, the Constant-Murley score, Subjective Shoulder Value, and American Shoulder and Elbow Surgeons score. Conclusion. Short-term CT scan analysis of a new hybrid anatomical glenoid component found perfect bony integration around the lower porous coated titanium peg in 90% of patients. The upper polyethylene flanged peg had bony integration in 70 patients (70 aTSAs; 67%). Longer follow-up is needed to analyze the rate of survival of this component. The short-term clinical outcome was satisfactory. Cite this article: Bone Joint J 2025;107-B(2):181–187


The Bone & Joint Journal
Vol. 98-B, Issue 1 | Pages 75 - 80
1 Jan 2016
Merolla G Chin P Sasyniuk TM Paladini P Porcellini G

Aims. We evaluated clinical and radiographic outcomes of total shoulder arthroplasty (TSA) using the second-generation Trabecular Metal (TM) Glenoid component. The first generation component was withdrawn in 2005 after a series of failures were reported. Between 2009 and 2012, 40 consecutive patients with unilateral TSA using the second-generation component were enrolled in this clinical study. The mean age of the patients was 63.8 years (40 to 75) and the mean follow-up was 38 months (24 to 42). Methods. Patients were evaluated using the Constant score (CS), the American Shoulder and Elbow Surgeons (ASES) score and routine radiographs. Results. Significant differences were found between the pre- and post-operative CS (p = 0.003), ASES (p = 0.009) scores and CS subscores of pain (p < 0.001), strength (p < 0.001) and mobility items (p < 0.05). No glenoid or humeral components migrated. Posterior thinning of the keel and slight wear at the polyethylene-TM interface was observed in one patient but was asymptomatic. Radiolucent lines were found around three humeral (< 1.5 mm) and two glenoid components (< 1 mm) and all were asymptomatic. Discussion. TSA with the second-generation TM Glenoid component results in satisfactory to excellent clinical performance, function, and subjective satisfaction at a mean follow-up of about three years. Radiographic changes were few and did not affect the outcome. Take home message: This paper highlights that the second generation Trabecular Metal Glenoid has better outcomes than those reported with the first-generation component.  . Cite this article: Bone Joint J 2016;98-B:75–80


Aims. To report early (two-year) postoperative findings from a randomized controlled trial (RCT) investigating disease-specific quality of life (QOL), clinical, patient-reported, and radiological outcomes in patients undergoing a total shoulder arthroplasty (TSA) with a second-generation uncemented trabecular metal (TM) glenoid versus a cemented polyethylene glenoid (POLY) component. Methods. Five fellowship-trained surgeons from three centres participated. Patients aged between 18 and 79 years with a primary diagnosis of glenohumeral osteoarthritis were screened for eligibility. Patients were randomized intraoperatively to either a TM or POLY glenoid component. Study intervals were: baseline, six weeks, six-, 12-, and 24 months postoperatively. The primary outcome was the Western Ontario Osteoarthritis Shoulder QOL score. Radiological images were reviewed for metal debris. Mixed effects repeated measures analysis of variance for within and between group comparisons were performed. Results. A total of 93 patients were randomized (46 TM; 47 POLY). No significant or clinically important differences were found with patient-reported outcomes at 24-month follow-up. Regarding the glenoid components, there were no complications or revision surgeries in either group. Grade 1 metal debris was observed in three (6.5%) patients with TM glenoids at 24 months but outcomes were not negatively impacted. Conclusion. Early results from this RCT showed no differences in disease-specific QOL, radiographs, complication rates, or shoulder function between uncemented second-generation TM and cemented POLY glenoids at 24 months postoperatively. Revision surgeries and reoperations were reported in both groups, but none attributed to glenoid implant failure. At 24 months postoperatively, Grade 1 metal debris was found in 6.5% of patients with a TM glenoid but did not negatively influence patient-reported outcomes. Longer-term follow-up is needed and is underway. Cite this article: Bone Jt Open 2021;2(9):728–736


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1197 - 1200
1 Sep 2009
Betts HM Abu-Rajab R Nunn T Brooksbank AJ

We describe the longer term clinical and radiological findings in a prospectively followed series of 49 rheumatoid patients (58 shoulders) who had undergone Neer II total shoulder replacement. The early and intermediate results have been published previously. At a mean follow-up of 19.8 years (16.5 to 23.8) 14 shoulders survived. Proximal migration of the humeral component was associated with progressive loosening of the glenoid and humeral components, but was independent of the state of the rotator cuff at the time of operation. Despite these changes the range of movement was preserved. Most patients had little or no pain in the shoulder, could sleep undisturbed and could attend to personal hygiene and grooming


The Bone & Joint Journal
Vol. 106-B, Issue 11 | Pages 1284 - 1292
1 Nov 2024
Moroder P Poltaretskyi S Raiss P Denard PJ Werner BC Erickson BJ Griffin JW Metcalfe N Siegert P

Aims

The objective of this study was to compare simulated range of motion (ROM) for reverse total shoulder arthroplasty (rTSA) with and without adjustment for scapulothoracic orientation in a global reference system. We hypothesized that values for simulated ROM in preoperative planning software with and without adjustment for scapulothoracic orientation would be significantly different.

Methods

A statistical shape model of the entire humerus and scapula was fitted into ten shoulder CT scans randomly selected from 162 patients who underwent rTSA. Six shoulder surgeons independently planned a rTSA in each model using prototype development software with the ability to adjust for scapulothoracic orientation, the starting position of the humerus, as well as kinematic planes in a global reference system simulating previously described posture types A, B, and C. ROM with and without posture adjustment was calculated and compared in all movement planes.


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 1 | Pages 68 - 72
1 Jan 1997
Stewart MPM Kelly IG

We made a prospective study of 58 consecutive Neer II total shoulder replacements in 49 rheumatoid patients. Cemented glenoid and press-fit humeral components had been used. After a mean follow-up of 9.5 years (7 to 13), 11 patients (15 shoulders) had died, one shoulder had been arthrodesed and five patients (five shoulders) had been lost to follow-up. Of the remaining 37 shoulders 29 were painfree or had only slight discomfort, four had pain on unusual activity, and four had moderate or severe pain. There were satisfactory improvements in the mean range of active elevation (53° to 75°) and external rotation (5° to 38°); satisfactory performance of the activities of daily living had been maintained throughout follow-up. Radiographs showed loosening in ten shoulders of nine glenoid and nine humeral components but of these only three had significant symptoms. Three loose glenoid components and two loose humeral components required revision. We consider that the Neer total shoulder replacement provides a reasonable medium to long-term outcome in rheumatoid arthritis, but recommend that the humeral component should be routinely cemented


The Bone & Joint Journal
Vol. 104-B, Issue 1 | Pages 83 - 90
1 Jan 2022
Batten TJ Gallacher S Evans JP Harding RJ Kitson J Smith CD Thomas WJ

Aims. The use and variety of stemless humeral components in anatomical total shoulder arthroplasty (TSA) have proliferated since their advent in 2004. Early outcomes are reassuring but independent mid-term results are scarce. This independent study reports a consecutive series of 143 Eclipse stemless shoulder prostheses with a minimum five-year (5 to 10) follow-up. Methods. Outcomes of 143 procedures undertaken for all indications in 131 patients were reviewed, with subset analysis of those for osteoarthritis (OA) (n = 99). The primary outcome was the Oxford Shoulder Score (OSS) at a minimum of five years. Secondary outcomes were ranges of motion and radiological analysis of humeral radiolucency, rotator cuff failure, and glenoid loosening. Results. Mean OSS at mean follow-up of 6.67 years (5.0 to 10.74) was 40.12 (9 to 48), with no statistically significant difference between those implanted for a non-OA indication and those for OA (p = 0.056) or time-dependent deterioration between two years and five years (p = 0.206). Ranges of motion significantly improved compared with preoperative findings and were maintained between two and five years with a mean external rotation of 38° (SD 18.1, 0 to 100) and forward elevation of 152° (SD 29.9, 90 to 180). Of those components with radiographs suitable for analysis (n = 83), 23 (28%) were found to have a least one humeral radiolucent line, which were predominantly incomplete, less than 2 mm, and in a single anatomical zone. No humeral components were loose. A radiolucent line was present around 22 (15%) of glenoid components, and 15 (10%) of components had failed. Rotator cuff failure was found in 21 (15%) components. The mean time to either glenoid or rotator cuff failure was greater than three years following implantation. Survivorship was 96.4% (95% CI 91.6 to 98.5, number at risk 128) at five years, and 94.3% (95% CI 88.2 to 97.3, number at risk 76) at seven years, both of which compare favourably with best results taken from available registries. Conclusion. Functional and radiological outcomes of the Eclipse stemless TSA are excellent, with no loose humeral components at minimum five-year follow-up. The presence of radiolucent lines is of interest and requires long-term observation but does not impact on the clinical results. Of the eight revisions required, this was predominantly for glenoid and rotator cuff failure. Cite this article: Bone Joint J 2022;104-B(1):83–90


Bone & Joint Open
Vol. 2, Issue 8 | Pages 618 - 630
2 Aug 2021
Ravi V Murphy RJ Moverley R Derias M Phadnis J

Aims

It is important to understand the rate of complications associated with the increasing burden of revision shoulder arthroplasty. Currently, this has not been well quantified. This review aims to address that deficiency with a focus on complication and reoperation rates, shoulder outcome scores, and comparison of anatomical and reverse prostheses when used in revision surgery.

Methods

A Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) systematic review was performed to identify clinical data for patients undergoing revision shoulder arthroplasty. Data were extracted from the literature and pooled for analysis. Complication and reoperation rates were analyzed using a meta-analysis of proportion, and continuous variables underwent comparative subgroup analysis.


Bone & Joint Open
Vol. 2, Issue 1 | Pages 58 - 65
22 Jan 2021
Karssiens TJ Gill JR Sunil Kumar KH Sjolin SU

Aims

The Mathys Affinis Short is the most frequently used stemless total shoulder prosthesis in the UK. The purpose of this prospective cohort study is to report the survivorship, clinical, and radiological outcomes of the first independent series of the Affinis Short prosthesis.

Methods

From January 2011 to January 2019, a total of 141 Affinis Short prostheses were implanted in 127 patients by a single surgeon. Mean age at time of surgery was 68 (44 to 89). Minimum one year and maximum eight year follow-up (mean 3.7 years) was analyzed using the Oxford Shoulder Score (OSS) at latest follow-up. Kaplan-Meier survivorship analysis was performed with implant revision as the endpoint. Most recently performed radiographs were reviewed for component radiolucent lines (RLLs) and proximal humeral migration.


Bone & Joint Research
Vol. 9, Issue 10 | Pages 689 - 700
7 Oct 2020
Zhang A Ma S Yuan L Wu S Liu S Wei X Chen L Ma C Zhao H

Aims

The study aimed to determine whether the microRNA miR21-5p (MiR21) mediates temporomandibular joint osteoarthritis (TMJ-OA) by targeting growth differentiation factor 5 (Gdf5).

Methods

TMJ-OA was induced in MiR21 knockout (KO) mice and wild-type (WT) mice by a unilateral anterior crossbite (UAC) procedure. Mouse tissues exhibited histopathological changes, as assessed by: Safranin O, toluidine blue, and immunohistochemistry staining; western blotting (WB); and quantitative real-time polymerase chain reaction (RT-qPCR). Mouse condylar chondrocytes were transfected with a series of MiR21 mimic, MiR21 inhibitor, Gdf5 siRNA (si-GDF5), and flag-GDF5 constructs. The effects of MiR-21 and Gdf5 on the expression of OA related molecules were evaluated by immunofluorescence, alcian blue staining, WB, and RT-qPCR.


The Bone & Joint Journal
Vol. 102-B, Issue 2 | Pages 232 - 238
1 Feb 2020
Javed S Hadi S Imam MA Gerogiannis D Foden P Monga P

Aims

Accurate measurement of the glenoid version is important in performing total shoulder arthroplasty (TSA). Our aim was to evaluate the Ellipse method, which involves formally defining the vertical mid-point of the glenoid prior to measuring the glenoid version and comparing it with the ‘classic’ Friedman method.

Methods

This was a retrospective study which evaluated 100 CT scans for patients who underwent a primary TSA. The glenoid version was measured using the Friedman and Ellipse methods by two senior observers. Statistical analyses were performed using the paired t-test for significance and the Bland-Altman plot for agreement.


The Bone & Joint Journal
Vol. 101-B, Issue 12 | Pages 1520 - 1525
1 Dec 2019
Clark NJ Samuelsen BT Alentorn-Geli E Assenmacher AT Cofield RH Sperling JW Sánchez-Sotelo J

Aims

Reverse shoulder arthroplasty (RSA) reliably improves shoulder pain and function for a variety of indications. However, the safety and efficacy of RSA in elderly patients is largely unknown. The purpose of this study was to report the mortality, morbidity, complications, reoperations, and outcomes of primary RSA in patients aged > 80 years.

Patients and Methods

Between 2004 and 2013, 242 consecutive primary RSAs were performed in patients aged > 80 years (mean 83.3 years (sd 3.1)). Of these, 53 were lost to follow-up before two years and ten had died within two years of surgery, leaving 179 for analysis of survivorship, pain, motion, and strength at a minimum of two years or until revision surgery. All 242 patients were considered for the analysis of 90-day, one-year, and overall mortality, medical complications (90-day and overall), surgical complications, and reoperations. The indications for surgery included rotator cuff arthropathy, osteoarthritis, fracture, the sequela of trauma, avascular necrosis, and rheumatoid arthritis. A retrospective review of the medical records was performed to collect all variables. Survivorship free of revision surgery was calculated at two and five years.


Bone & Joint 360
Vol. 7, Issue 3 | Pages 21 - 24
1 Jun 2018


The Bone & Joint Journal
Vol. 100-B, Issue 11 | Pages 1493 - 1498
1 Nov 2018
Wagner ER Hevesi M Houdek MT Cofield RH Sperling JW Sanchez-Sotelo J

Aims

Patients with a failed reverse shoulder arthroplasty (RSA) have limited salvage options. The aim of this study was to determine the outcome of revision RSA when used as a salvage procedure for a failed primary RSA.

Patients and Methods

We reviewed all revision RSAs performed for a failed primary RSA between 2006 and 2012, excluding patients with a follow-up of less than two years. A total of 27 revision RSAs were included in the study. The mean age of the patients at the time of revision was 70 years (58 to 82). Of the 27 patients, 14 (52% were female). The mean follow-up was 4.4 years (2 to 10).


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1609 - 1617
1 Dec 2018
Malhas AM Granville-Chapman J Robinson PM Brookes-Fazakerley S Walton M Monga P Bale S Trail I

Aims

We present our experience of using a metal-backed prosthesis and autologous bone graft to treat gross glenoid bone deficiency.

Patients and Methods

A prospective cohort study of the first 45 shoulder arthroplasties using the SMR Axioma Trabecular Titanium (TT) metal-backed glenoid with autologous bone graft. Between May 2013 and December 2014, 45 shoulder arthroplasties were carried out in 44 patients with a mean age of 64 years (35 to 89). The indications were 23 complex primary arthroplasties, 12 to revise a hemiarthroplasty or resurfacing, five for aseptic loosening of the glenoid, and five for infection.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 674 - 679
1 May 2017
Nuttall D Birch A Haines JF Watts AC Trail IA

Aims. Radiostereometric analysis (RSA) allows an extremely accurate measurement of early micromotion of components following arthroplasty. . Patients and Methods. In this study, RSA was used to measure the migration of 11 partially cemented fluted pegged glenoid components in patients with osteoarthritis who underwent total shoulder arthroplasty using an improved surgical technique (seven men, four women, mean age 68). Patients were evaluated clinically using the American Shoulder and Elbow Surgeons (ASES) and Constant-Murley scores and by CT scans two years post-operatively. . Results. There were two patterns of migration, the first showing little, if any, migration and the second showing rotation by > 6° as early as three months post-operatively. At two years, these two groups could be confirmed on CT scans, one with osseointegration around the central peg, and the second with cystic changes. Patients with osteolysis around the central peg were those with early migration and those with osseointegration had minimal early migration. Both groups,however,had similar clinical results. . Conclusion. Rapid early migration associated with focal lucency and absence of osseointegration was observed in three of 11 glenoid components, suggesting that lack of initial stability leads to early movement and failure of osseointegration. Cite this article: Bone Joint J 2017;99-B:674–9


Bone & Joint 360
Vol. 6, Issue 4 | Pages 20 - 22
1 Aug 2017


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 485 - 492
1 Apr 2018
Gauci MO Bonnevialle N Moineau G Baba M Walch G Boileau P

Aims. Controversy about the use of an anatomical total shoulder arthroplasty (aTSA) in young arthritic patients relates to which is the ideal form of fixation for the glenoid component: cemented or cementless. This study aimed to evaluate implant survival of aTSA when used in patients aged < 60 years with primary glenohumeral osteoarthritis (OA), and to compare the survival of cemented all-polyethylene and cementless metal-backed glenoid components. Materials and Methods. A total of 69 consecutive aTSAs were performed in 67 patients aged < 60 years with primary glenohumeral OA. Their mean age at the time of surgery was 54 years (35 to 60). Of these aTSAs, 46 were undertaken using a cemented polyethylene component and 23 were undertaken using a cementless metal-backed component. The age, gender, preoperative function, mobility, premorbid glenoid erosion, and length of follow-up were comparable in the two groups. The patients were reviewed clinically and radiographically at a mean of 10.3 years (5 to 12, . sd. 26) postoperatively. Kaplan–Meier survivorship analysis was performed with revision as the endpoint. Results. A total of 26 shoulders (38%) underwent revision surgery: ten (22%) in the polyethylene group and 16 (70%) in the metal-backed group (p < 0.0001). At 12 years’ follow-up, the rate of implant survival was 74% (. sd.  0.09) for polyethylene components and 24% (. sd.  0.10) for metal-backed components (p < 0.0002). Glenoid loosening or failure was the indication for revision in the polyethylene group, whereas polyethylene wear with metal-on-metal contact, instability, and insufficiency of the rotator cuff were the indications for revision in the metal-backed group. Preoperative posterior subluxation of the humeral head with a biconcave/retroverted glenoid (Walch B2) had an adverse effect on the survival of a metal-backed component. Conclusion. The survival of a cemented polyethylene glenoid component is three times higher than that of a cementless metal-backed glenoid component ten years after aTSA in patients aged < 60 years with primary glenohumeral OA. Patients with a biconcave (B2) glenoid have the highest risk of failure. Cite this article: Bone Joint J 2018;100-B:485–92


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 668 - 672
1 May 2013
Abdel MP Hattrup SJ Sperling JW Cofield RH Kreofsky CR Sanchez-Sotelo J

Instability after arthroplasty of the shoulder is difficult to correct surgically. Soft-tissue procedures and revision surgery using unconstrained anatomical components are associated with a high rate of failure. The purpose of this study was to determine the results of revision of an unstable anatomical shoulder arthroplasty to a reverse design prosthesis. Between 2004 and 2007, 33 unstable anatomical shoulder arthroplasties were revised to a reverse design. The mean age of the patients was 71 years (53 to 86) and their mean follow-up was 42 months (25 to 71). The mean time to revision was 26 months (4 to 164). Pain scores improved significantly (pre-operative visual analogue scale (VAS) of 7.2 (sd 1.6); most recent VAS 2.2 (sd 1.9); p = 0.001). There was a statistically significant increase in mean active forward elevation from 40.2° (sd 27.3) to 97.0° (sd 36.2) (p = 0.001). There was no significant difference in internal (p = 0.93) or external rotation (p = 0.40). Radiological findings included notching in five shoulders (15%) and heterotopic ossification of the inferior capsular region in three (9%). At the last follow-up 31 shoulders (94%) were stable. The remaining two shoulders dislocated at 2.5 weeks and three months post-operatively, respectively. According to the Neer rating system, there were 13 excellent (40%), ten satisfactory (30%) and ten unsatisfactory results (30%). Revision of hemiarthroplasty or anatomical total shoulder replacement for instability using a reverse design prosthesis gives good short-term results.

Cite this article: Bone Joint J 2013;95-B:668–72.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1377 - 1382
1 Oct 2013
Walch G Mesiha M Boileau P Edwards TB Lévigne C Moineau G Young A

Osteoarthritis results in changes in the dimensions of the glenoid. This study aimed to assess the size and radius of curvature of arthritic glenoids. A total of 145 CT scans were analysed, performed as part of routine pre-operative assessment before total shoulder replacement in 91 women and 54 men. Only patients with primary osteoarthritis and a concentric glenoid were included in the study. The CT scans underwent three-dimensional (3D) reconstruction and were analysed using dedicated computer software. The measurements consisted of maximum superoinferior height, anteroposterior width and a best-fit sphere radius of curvature of the glenoid. The mean height was 40.2 mm (. sd. 4.9), the mean width was 29 mm (. sd. 4.3) and the mean radius of curvature was 35.4 mm (. sd. 7.8). The measurements were statistically different in men and women and had a Gaussian distribution with marked variation. All measurements were greater than the known values in normal subjects. With current shoulder replacement systems using a unique backside radius of curvature for the glenoid component, there is a risk of undertaking excessive reaming to adapt the bone to the component resulting in sacrifice of subchondral bone or under-reaming and instability of the component due to a ’rocking horse‘ phenomenon. . Cite this article: Bone Joint J 2013;95-B:1377–82