Results 1 - 35 of 35
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 74-B, Issue 1 | Pages 159 - 160
1 Jan 1992
Lacroix H du Bois N van Biezen F


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 623 - 627
1 May 2005
Kim Y

Between January 1984 and January 1986, 131 porous-coated anatomic total hip replacements were performed in 119 consecutive patients. Of these, 100 patients (110 hips) who were alive at a minimum of 18 years after replacement were included in the study. The mean age of the patients at surgery was 48.4 years and that of surviving patients at the latest follow-up was 67.8 years. The mean duration of the clinical and radiological follow-up was 19.4 years (18 to 20).

The mean Harris hip score initially improved from 55 points before to 95 points at two years after operation, but subsequently decreased to 91 points after six years, 87 points after 11.2 years, and 85 points after 19.4 years. At the final follow-up, 23 hips (18%) of the entire cohort and 21% of survivors had undergone revision because of loosening or osteolysis of the acetabular component and eight hips (6%) of the entire cohort and 7% of survivors for loosening or osteolysis of the femoral component. Only four femoral components (4%) were revised for isolated aseptic loosening without osteolysis and two (2%) for recurrent dislocation. On the basis of these long-term results, the porous-coated anatomic femoral component survived for a minimum of 18 years after operation while the acetabular component was less durable. The findings identify the principles of uncemented joint replacement which can be applied to current practice.


The Journal of Bone & Joint Surgery British Volume
Vol. 77-B, Issue 3 | Pages 369 - 373
1 May 1995
Savilahti S Myllyneva I Lindholm T Pajamaki K Nevalainen J Laippala P

We report the clinical outcome and survival of 100 total hip arthroplasties with Link RS cementless hip prostheses after a mean of 63 months (51 to 93). The average preoperative Merle d'Aubigne hip score was 9.4 points and the average score at the latest follow-up was 15.3 points. Thirteen prostheses have been revised and seven await revision. Survivorship analysis based on intention to revise showed only 68% survival at seven years (95% CI 53% to 81%). The survivals of femoral and acetabular components verified to be loose at surgery were 78% and 89% respectively at seven years. A number of other patients have pain and it seems likely that there will be substantial reduction in these survival figures in the future.


The Journal of Bone & Joint Surgery British Volume
Vol. 61-B, Issue 4 | Pages 422 - 423
1 Nov 1979
Hunter G

Sixty-five total hip arthroplastics were reinserted after sepsis around the hip, positive cultures being obtained from fifty-six. Although 65 per cent of patients still have their implant in position, only sixteen of sixty-five (25 per cent) show an excellent or good result on a Harris rating. Twenty-three of sixty-five (35 per cent) subsequently required an excision arthroplasty. The indications and contraindications for this procedure are given.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 6 | Pages 884 - 891
1 Nov 1996
Malchau H Herberts P Wang YX Kärrholm J Romanus B

We enrolled 98 patients (107 hips) with a mean age of 47 years (SD 8.6) into a prospective study of the Madreporic Lord THR; 34 hips had primary and 73 secondary osteoarthritis.

After ten years, the survival rate using revision as the endpoint for failure was 70% (±9) for the cup and 98% (±0.3) for the stem. The combined clinical and radiological survival rates were 46% (±11) and 81% (±10), respectively. Osteoporosis due to stress-shielding was observed in the proximal femur. Hips with radiologically dense bone postoperatively showed the most pronounced bone loss.

We recommend continued radiological follow-up of patients with this type of implant to allow revision to be performed before there is severe bony destruction of the pelvis.


The Journal of Bone & Joint Surgery British Volume
Vol. 59-B, Issue 3 | Pages 298 - 301
1 Aug 1977
Clegg J

Twenty-nine patients with thirty pseudarthroses after removal of infected total hip prostheses have been reviewed one to six years later. Six hips still had a discharging sinus. Complete removal of all cement was found to be essential for healing, and various points in the technique of its clearance have been made. In this respect radio-opaque cement was a great advantage, and lateral guttering of the femur was most effective. Compared with the situation before replacement considerable relief of pain was obtained in most patients but there was much less improvement in function.


The Bone & Joint Journal
Vol. 105-B, Issue 9 | Pages 1020 - 1029
1 Sep 2023
Trouwborst NM ten Duis K Banierink H Doornberg JN van Helden SH Hermans E van Lieshout EMM Nijveldt R Tromp T Stirler VMA Verhofstad MHJ de Vries JPPM Wijffels MME Reininga IHF IJpma FFA

Aims

The aim of this study was to investigate the association between fracture displacement and survivorship of the native hip joint without conversion to a total hip arthroplasty (THA), and to determine predictors for conversion to THA in patients treated nonoperatively for acetabular fractures.

Methods

A multicentre cross-sectional study was performed in 170 patients who were treated nonoperatively for an acetabular fracture in three level 1 trauma centres. Using the post-injury diagnostic CT scan, the maximum gap and step-off values in the weightbearing dome were digitally measured by two trauma surgeons. Native hip survival was reported using Kaplan-Meier curves. Predictors for conversion to THA were determined using Cox regression analysis.


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1326 - 1331
1 Oct 2013
Eilander W Harris SJ Henkus HE Cobb JP Hogervorst T

Orientation of the acetabular component influences wear, range of movement and the incidence of dislocation after total hip replacement (THR). During surgery, such orientation is often referenced to the anterior pelvic plane (APP), but APP inclination relative to the coronal plane (pelvic tilt) varies substantially between individuals. In contrast, the change in pelvic tilt from supine to standing (dPT) is small for nearly all individuals. Therefore, in THR performed with the patient supine and the patient’s coronal plane parallel to the operating table, we propose that freehand placement of the acetabular component placement is reliable and reflects standing (functional) cup position. We examined this hypothesis in 56 hips in 56 patients (19 men) with a mean age of 61 years (29 to 80) using three-dimensional CT pelvic reconstructions and standing lateral pelvic radiographs. We found a low variability of acetabular component placement, with 46 implants (82%) placed within a combined range of 30° to 50° inclination and 5° to 25° anteversion. Changing from the supine to the standing position (analysed in 47 patients) was associated with an anteversion change < 10° in 45 patients (96%). dPT was < 10° in 41 patients (87%). In conclusion, supine THR appears to provide reliable freehand acetabular component placement. In most patients a small reclination of the pelvis going from supine to standing causes a small increase in anteversion of the acetabular component.

Cite this article: Bone Joint J 2013;95-B:1326–31.


The Bone & Joint Journal
Vol. 95-B, Issue 1 | Pages 23 - 30
1 Jan 2013
Kiernan S Hermann KL Wagner P Ryd L Flivik G

Progressive retroversion of a cemented stem is predictive of early loosening and failure. We assessed the relationship between direct post-operative stem anteversion, measured with CT, and the resulting rotational stability, measured with repeated radiostereometric analysis over ten years. The study comprised 60 cemented total hip replacements using one of two types of matt collared stem with a rounded cross-section. The patients were divided into three groups depending on their measured post-operative anteversion (< 10°, 10° to 25°, >  25°). There was a strong correlation between direct post-operative anteversion and later posterior rotation. At one year the < 10° group showed significantly more progressive retroversion together with distal migration, and this persisted to the ten-year follow-up. In the < 10° group four of ten stems (40%) had been revised at ten years, and an additional two stems (20%) were radiologically loose. In the ‘normal’ (10° to 25°) anteversion group there was one revised (3%) and one loose stem (3%) of a total of 30 stems, and in the > 25° group one stem (5%) was revised and another loose (5%) out of 20 stems. This poor outcome is partly dependent on the design of this prosthesis, but the results strongly suggest that the initial rotational position of cemented stems during surgery affects the subsequent progressive retroversion, subsidence and eventual loosening. The degree of retroversion may be sensitive to prosthetic design and stem size, but < 10° of anteversion appears deleterious to the long-term outcome for cemented hip prosthetic stems.

Cite this article: Bone Joint J 2013;95-B:23–30.


The Bone & Joint Journal
Vol. 105-B, Issue 7 | Pages 775 - 782
1 Jul 2023
Koper MC Spek RWA Reijman M van Es EM Baart SJ Verhaar JAN Bos PK

Aims

The aims of this study were to determine if an increasing serum cobalt (Co) and/or chromium (Cr) concentration is correlated with a decreasing Harris Hip Score (HHS) and Hip disability and Osteoarthritis Outcome Score (HOOS) in patients who received the Articular Surface Replacement (ASR) hip resurfacing arthroplasty (HRA), and to evaluate the ten-year revision rate and show if sex, inclination angle, and Co level influenced the revision rate.

Methods

A total of 62 patients with an ASR-HRA were included and monitored yearly postoperatively. At follow-up, serum Co and Cr levels were measured and the HHS and the HOOS were scored. In addition, preoperative patient and implant variables and the need for revision surgery were recorded. We used a linear mixed model to relate the serum Co and Cr levels to different patient-reported outcome measures (PROMs). For the survival analyses we used the Kaplan-Meier and Cox regression model.


Bone & Joint Research
Vol. 5, Issue 10 | Pages 461 - 469
1 Oct 2016
Liu YK Deng XX Yang H

Objectives. The cytotoxicity induced by cobalt ions (Co. 2+. ) and cobalt nanoparticles (Co-NPs) which released following the insertion of a total hip prosthesis, has been reported. However, little is known about the underlying mechanisms. In this study, we investigate the toxic effect of Co. 2+. and Co-NPs on liver cells, and explain further the potential mechanisms. Methods. Co-NPs were characterised for size, shape, elemental analysis, and hydrodynamic diameter, and were assessed by Transmission Electron Microscope, Scanning Electron Microscope, Energy Dispersive X-ray Spectroscopy and Dynamic Light Scattering. BRL-3A cells were used in this study. Cytotoxicity was evaluated by MTT and lactate dehydrogenase release assay. In order to clarify the potential mechanisms, reactive oxygen species, Bax/Bcl-2 mRNA expression, IL-8 mRNA expression and DNA damage were assessed on BRL-3A cells after Co. 2+. or Co-NPs treatment. Results. Results showed cytotoxic effects of Co. 2+. and Co-NPs were dependent upon time and dosage, and the cytotoxicity of Co-NPs was greater than that of Co. 2+. In addition, Co-NPs elicited a significant (p < 0.05) reduction in cell viability with a concomitant increase in lactic dehydrogenase release, reactive oxygen species generation, IL-8 mRNA expression, Bax/Bcl-2 mRNA expression and DNA damage after 24 hours of exposure. Conclusion. Co-NPs induced greater cytotoxicity and genotoxicity in BRL-3A cells than Co. 2+. Cell membrane damage, oxidative stress, immune inflammation and DNA damage may play an important role in the effects of Co-NPs on liver cells. Cite this article: Y. K. Liu, X. X. Deng, H.L. Yang. Cytotoxicity and genotoxicity in liver cells induced by cobalt nanoparticles and ions. Bone Joint Res 2016;5:461–469. DOI: 10.1302/2046-3758.510.BJR-2016-0016.R1


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1292 - 1303
1 Dec 2022
Polisetty TS Jain S Pang M Karnuta JM Vigdorchik JM Nawabi DH Wyles CC Ramkumar PN

Literature surrounding artificial intelligence (AI)-related applications for hip and knee arthroplasty has proliferated. However, meaningful advances that fundamentally transform the practice and delivery of joint arthroplasty are yet to be realized, despite the broad range of applications as we continue to search for meaningful and appropriate use of AI. AI literature in hip and knee arthroplasty between 2018 and 2021 regarding image-based analyses, value-based care, remote patient monitoring, and augmented reality was reviewed. Concerns surrounding meaningful use and appropriate methodological approaches of AI in joint arthroplasty research are summarized. Of the 233 AI-related orthopaedics articles published, 178 (76%) constituted original research, while the rest consisted of editorials or reviews. A total of 52% of original AI-related research concerns hip and knee arthroplasty (n = 92), and a narrative review is described. Three studies were externally validated. Pitfalls surrounding present-day research include conflating vernacular (“AI/machine learning”), repackaging limited registry data, prematurely releasing internally validated prediction models, appraising model architecture instead of inputted data, withholding code, and evaluating studies using antiquated regression-based guidelines. While AI has been applied to a variety of hip and knee arthroplasty applications with limited clinical impact, the future remains promising if the question is meaningful, the methodology is rigorous and transparent, the data are rich, and the model is externally validated. Simple checkpoints for meaningful AI adoption include ensuring applications focus on: administrative support over clinical evaluation and management; necessity of the advanced model; and the novelty of the question being answered.

Cite this article: Bone Joint J 2022;104-B(12):1292–1303.


The Bone & Joint Journal
Vol. 104-B, Issue 12 | Pages 1369 - 1378
1 Dec 2022
van Rijckevorsel VAJIM de Jong L Verhofstad MHJ Roukema GR

Aims

Factors associated with high mortality rates in geriatric hip fracture patients are frequently unmodifiable. Time to surgery, however, might be a modifiable factor of interest to optimize clinical outcomes after hip fracture surgery. This study aims to determine the influence of postponement of surgery due to non-medical reasons on clinical outcomes in acute hip fracture surgery.

Methods

This observational cohort study enrolled consecutively admitted patients with a proximal femoral fracture, for which surgery was performed between 1 January 2018 and 11 January 2021 in two level II trauma teaching hospitals. Patients with medical indications to postpone surgery were excluded. A total of 1,803 patients were included, of whom 1,428 had surgery < 24 hours and 375 had surgery ≥ 24 hours after admission.


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims

Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors.

Methods

We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test.


The Journal of Bone & Joint Surgery British Volume
Vol. 67-B, Issue 3 | Pages 413 - 415
1 May 1985
Nielsen H Veth R Oldhoff J Koops H Scales J

We report a patient with a peri-acetabular chondrosarcoma which was treated by resection of part of the pelvis and the femoral head. This was followed by replacement with a custom-made titanium alloy pelvic portion and a conventional total hip prosthesis. The operation was carried out in two stages, and yielded a good functional and cosmetic result


The Journal of Bone & Joint Surgery British Volume
Vol. 71-B, Issue 1 | Pages 58 - 62
1 Jan 1989
Langlais F Vielpeau C

Four en-bloc resections for malignant tumours of the hip, the peri-acetabular region and the iliac wing were reconstructed using an irradiated hemipelvic allograft together with a total hip prosthesis. Technical aspects include the use of an anterior Enneking approach which excises the previous biopsy site, division and re-attachment of the iliac crest and fixation of the prosthesis using a modified acetabular cup and three polypropylene artificial ligaments to increase the stability of the joint. Weight-bearing was allowed at three months. The oncological and clinical results were satisfactory after a mean follow-up of 19 months, with walking distances of 500 metres to two kilometres without pain


The Journal of Bone & Joint Surgery British Volume
Vol. 59-B, Issue 3 | Pages 293 - 297
1 Aug 1977
Hunter G Dandy D

Deep infection, the most serious local complication of total hip replacement, prompted a study of the records of 135 patients (137 hips) thus afflicted in a nationwide survey of Canada. Particular attention has been paid to the natural history of the infection, and the problems of diagnosis are described. Twenty-one patients died after the insertion, or removal, of the prosthesis, and of the survivors of the original 135 patients only eighteen have been able to retain the prosthesis without further problems with the wound. The remaining patients had the prosthesis removed, and most dry wounds. Certain suggestions are made on management. The advice that a second total hip prosthesis should be inserted after a deep infection of the first implant is not supported


The Journal of Bone & Joint Surgery British Volume
Vol. 75-B, Issue 3 | Pages 468 - 474
1 May 1993
Schuller H Dalstra M Huiskes R Marti R

In acetabular dysplasia, fixation of the acetabular component of a cemented total hip prosthesis may be insecure and superolateral bone grafts are often used to augment the acetabular roof. We used finite element analysis to study the mechanical importance of the lateral acetabular roof and found that the lateral acetabular rim plays an important role in the load transfer of the pelvic bone. When the superlateral rim was lacking, the load shifted to the posterosuperior rim and to the area of pubic support, and the stresses in all materials, especially in the cement and in the trabecular bone, increased greatly. At the cement-bone interface the tilting component of the shear stress increased threefold. In a model in which the dysplastic acetabulum was augmented by a rigidly fixed, load-transmitting bone graft, the stresses were considerably diminished


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 4 | Pages 705 - 709
1 Jul 1999
Hara T Hayashi K Nakashima Y Kanemaru T Iwamoto Y

We have studied the effect of hydroxyapatite (HA) coating in 15 ovariectomised and 15 normal rats which had had a sham procedure. Twenty-four weeks after operation, HA-coated implants were inserted into the intramedullary canal of the right femur and uncoated implants into the left femur. The prostheses were removed four weeks after implantation. Twelve specimens in each group had mechanical push-out tests. Sagittal sections of the other three were evaluated by SEM. The bone mineral density (BMD) of the dissected left tibia was measured by dual-energy x-ray absorptiometry. The difference in BMD between the control and ovariectomised tibiae was 35.01 mg/cm. 2. (95% CI, 26.60 to 43.42). The push-out strength of the HA-coated implants was higher than that of the uncoated implants in both groups (p < 0.0001), but the HA-coated implants of the ovariectomised group had a reduction in push-out strength of 40.3% compared with the control group (p < 0.0001). Our findings suggest that HA-coated implants may improve the fixation of a cementless total hip prosthesis but that the presence of osteoporosis may limit the magnitude of this benefit


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 912 - 917
1 Sep 1998
Granchi D Verri E Ciapetti G Stea S Savarino L Sudanese A Mieti M Rotini R Dallari D Zinghi G Montanaro L

Our aim was to determine if the serum levels of bone-resorbing cytokines (IL-1β, TNF-α, IL-6, GM-CSF) are altered in patients with aseptic loosening of a total hip prosthesis, and if such levels are influenced by the type of implant. We determined cytokine levels in sera from 35 patients before revision for failed total hip arthroplasty and compared them with those in 25 healthy donors. We also assessed the soluble receptor of interleukin-2 (sIL-2r) in serum as an indication of a specific immune reaction against the implant. Our findings showed that the sIL-2r and TNF-α serum level did not change. The IL-6 level was not significantly altered, but was higher in patients with TiAlV prostheses than in those with a CrCoMo implant and in patients with cemented prostheses. The IL-1β level was found to be higher in those with a TiAlV cemented prosthesis than in the control group (p = 0.0001) and other groups of patients (p = 0.003 v uncemented TiAlV, p = 0.01 v cemented CrCoMo, p = 0.001 v uncemented CrCoMo). The GM-CSF level significantly increased in patients compared with healthy subjects (p = 0.008), and it was higher in those with cemented than with uncemented implants (p = 0.01). Only patients with cementless CrCoMo prostheses had levels of GM-CSF similar to those of the control group. The highest GM-CSF concentrations were observed in patients treated with non-steroidal anti-inflammatory drugs (NSAIDs) in the last months before revision (p = 0.04). In addition, when massive osteolysis was observed, the level of GM-CSF tended to decrease to that of the control group


Bone & Joint Research
Vol. 8, Issue 8 | Pages 387 - 396
1 Aug 2019
Alt V Rupp M Lemberger K Bechert T Konradt T Steinrücke P Schnettler R Söder S Ascherl R

Objectives

Preclinical data showed poly(methyl methacrylate) (PMMA) loaded with microsilver to be effective against a variety of bacteria. The purpose of this study was to assess patient safety of PMMA spacers with microsilver in prosthetic hip infections in a prospective cohort study.

Methods

A total of 12 patients with prosthetic hip infections were included for a three-stage revision procedure. All patients received either a gentamicin-PMMA spacer (80 g to 160 g PMMA depending on hip joint dimension) with additional loading of 1% (w/w) of microsilver (0.8 g to 1.6 g per spacer) at surgery 1 followed by a gentamicin-PMMA spacer without microsilver at surgery 2 or vice versa. Implantation of the revision prosthesis was carried out at surgery 3.


Bone & Joint Research
Vol. 7, Issue 11 | Pages 609 - 619
1 Nov 2018
Pijls BG Sanders IMJG Kuijper EJ Nelissen RGHH

Objectives

Prosthetic joint infection (PJI) is a devastating complication following total joint arthroplasty. Non-contact induction heating of metal implants is a new and emerging treatment for PJI. However, there may be concerns for potential tissue necrosis. It is thought that segmental induction heating can be used to control the thermal dose and to limit collateral thermal injury to the bone and surrounding tissues. The purpose of this study was to determine the thermal dose, for commonly used metal implants in orthopaedic surgery, at various distances from the heating centre (HC).

Methods

Commonly used metal orthopaedic implants (hip stem, intramedullary nail, and locking compression plate (LCP)) were heated segmentally using an induction heater. The thermal dose was expressed in cumulative equivalent minutes at 43°C (CEM43) and measured with a thermal camera at several different distances from the HC. A value of 16 CEM43 was used as the threshold for thermal damage in bone.


The Bone & Joint Journal
Vol. 100-B, Issue 10 | Pages 1392 - 1398
1 Oct 2018
Willeumier JJ van de Sande MAJ van der Wal RJP Dijkstra PDS

Aims

The aim of this study was to assess the current trends in the estimation of survival and the preferred forms of treatment of pathological fractures among national and international general and oncological orthopaedic surgeons, and to explore whether improvements in the management of these patients could be identified in this way.

Materials and Methods

All members of the Dutch Orthopaedic Society (DOS) and the European Musculoskeletal Oncology Society (EMSOS) were invited to complete a web-based questionnaire containing 12 cases.


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 455 - 460
1 Apr 2018
Mumith A Thomas M Shah Z Coathup M Blunn G

Increasing innovation in rapid prototyping (RP) and additive manufacturing (AM), also known as 3D printing, is bringing about major changes in translational surgical research.

This review describes the current position in the use of additive manufacturing in orthopaedic surgery.

Cite this article: Bone Joint J 2018;100-B:455-60.



Bone & Joint Research
Vol. 6, Issue 2 | Pages 113 - 122
1 Feb 2017
Scholes SC Hunt BJ Richardson VM Langton DJ Smith E Joyce TJ

Objectives

The high revision rates of the DePuy Articular Surface Replacement (ASR) and the DePuy ASR XL (the total hip arthroplasty (THA) version) have led to questions over the viability of metal-on-metal (MoM) hip joints. Some designs of MoM hip joint do, however, have reasonable mid-term performance when implanted in appropriate patients. Investigations into the reasons for implant failure are important to offer help with the choice of implants and direction for future implant designs. One way to assess the performance of explanted hip prostheses is to measure the wear (in terms of material loss) on the joint surfaces.

Methods

In this study, a coordinate measuring machine (CMM) was used to measure the wear on five failed cementless Biomet Magnum/ReCap/ Taperloc large head MoM THAs, along with one Biomet ReCap resurfacing joint. Surface roughness measurements were also taken. The reason for revision of these implants was pain and/or adverse reaction to metal debris (ARMD) and/or elevated blood metal ion levels.


Bone & Joint Research
Vol. 5, Issue 6 | Pages 206 - 214
1 Jun 2016
Malak TT Broomfield JAJ Palmer AJR Hopewell S Carr A Brown C Prieto-Alhambra D Glyn-Jones S

Objectives

High failure rates of metal-on-metal hip arthroplasty implants have highlighted the need for more careful introduction and monitoring of new implants and for the evaluation of the safety of medical devices. The National Joint Registry and other regulatory services are unable to detect failing implants at an early enough stage. We aimed to identify validated surrogate markers of long-term outcome in patients undergoing primary total hip arthroplasty (THA).

Methods

We conducted a systematic review of studies evaluating surrogate markers for predicting long-term outcome in primary THA. Long-term outcome was defined as revision rate of an implant at ten years according to National Institute of Health and Care Excellence guidelines. We conducted a search of Medline and Embase (OVID) databases. Separate search strategies were devised for the Cochrane database and Google Scholar. Each search was performed to include articles from the date of their inception to June 8, 2015.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 160 - 163
1 Feb 2015
Biemond JE Venkatesan S van Hellemondt GG

The long-term survival of the cementless Spotorno CLS femoral component in patients aged > 50 years at the time of arthroplasty was investigated. Survivorship analysis of a consecutive series of 85 patients (100 hips; under 50 years of age at a mean follow-up of 18.4 years (16.3 to 20.8)) was performed. The clinical and radiographic outcomes were satisfactory. The overall rate of survival of the femoral component was 93.5% (95% confidence interval (CI), 90.9 to 96.1) after 19 years. Survival with revision for aseptic loosening as the end point was 95.7% (95% CI 93.6 to 97.8%) at 19 years.

This study demonstrates an excellent long-term survival of the Spotorno CLS femoral component after 16 to 20 years in young patients undergoing total hip arthroplasty.

Cite this article: Bone Joint J 2015;97-B:160–3.


The Bone & Joint Journal
Vol. 96-B, Issue 11 | Pages 1478 - 1484
1 Nov 2014
Garcia-Rey E Cruz-Pardos A Madero R

A total of 31 patients, (20 women, 11 men; mean age 62.5 years old; 23 to 81), who underwent conversion of a Girdlestone resection-arthroplasty (RA) to a total hip replacement (THR) were compared with 93 patients, (60 women, 33 men; mean age 63.4 years old; 20 to 89), who had revision THR surgery for aseptic loosening in a retrospective matched case-control study. Age, gender and the extent of the pre-operative bone defect were similar in all patients. Mean follow-up was 9.3 years (5 to 18).

Pre-operative function and range of movement were better in the control group (p = 0.01 and 0.003, respectively) and pre-operative leg length discrepancy (LLD) was greater in the RA group (p < 0.001). The post-operative clinical outcome was similar in both groups except for mean post-operative LLD, which was greater in the study group (p = 0.003). There was a significant interaction effect for LLD in the study group (p < 0.001). A two-way analysis of variance showed that clinical outcome depended on patient age (patients older than 70 years old had worse pre-operative pain, p = 0.017) or bone defect (patients with a large acetabular bone defect had higher LLD, p = 0.006, worse post-operative function p = 0.009 and range of movement, p = 0.005), irrespective of the group.

Despite major acetabular and femoral bone defects requiring complex surgical reconstruction techniques, THR after RA shows a clinical outcome similar to those obtained in aseptic revision surgery for hips with similar sized bone defects.

Cite this article: Bone Joint J 2014;96-B:1478–84.


Bone & Joint 360
Vol. 3, Issue 2 | Pages 20 - 22
1 Apr 2014

The April 2014 Trauma Roundup360 looks at: is it safe to primarily close dog bite wounds?; conservative transfusion evidence based in hip fracture surgery; tibial nonunion is devastating to quality of life; sexual dysfunction after traumatic pelvic fracture; hemiarthroplasty versus fixation in displaced femoral neck fractures; silver VAC dressings “Gold Standard” in massive wounds; dual plating for talar neck fracture; syndesmosis and fibular length easiest errors in ankle fracture surgery; and dual mobility: stable as a rock in fracture.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 10 | Pages 1363 - 1369
1 Oct 2010
Garcia-Cimbrelo E Garcia-Rey E Cruz-Pardos A Madero R

Revision surgery of the hip was performed on 114 hips using an extensively porous-coated femoral component. Of these, 95 hips (94 patients) had a mean follow-up of 10.2 years (5 to 17). No cortical struts were used and the cortical index and the femoral cortical width were measured at different levels.

There were two revisions for aseptic loosening. Survivorship at 12 years for all causes of failure was 96.9% (95% confidence interval 93.5 to 100) in the best-case scenario. Fibrous or unstable fixation was associated with major bone defects. The cortical index (p = 0.045) and the lateral cortical thickness (p = 0.008) decreased at the proximal level over time while the medial cortex increased (p = 0.001) at the proximal and distal levels. An increase in the proximal medial cortex was found in patients with an extended transtrochanteric osteotomy (p = 0.026) and in those with components shorter than 25 cm (p = 0.008).

The use of the extensively porous-coated femoral component can provide a solution for difficult cases in revision surgery. Radiological bony ingrowth is common. Although without clinical relevance at the end of follow-up, the thickness of the medial femoral cortex often increased while that of the lateral cortex decreased. In cases in which a shorter component was used and in those undertaken using an extended trochanteric osteotomy, there was a greater increase in thickness of the femoral cortex over time.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 5 | Pages 701 - 706
1 May 2010
Fennema P Lubsen J

Survival analysis is an important tool for assessing the outcome of total joint replacement. The Kaplan-Meier method is used to estimate the incidence of revision of a prosthesis over time, but does not account appropriately for competing events which preclude revision. In the presence of competing death, this method will lead to statistical bias and the curve will lose its interpretability. A valid comparison of survival results between studies using the method is impossible without accounting for different rates of competing events. An alternative and easily applicable approach, the cumulative incidence of competing risk, is proposed. Using three simulated data sets and realistic data from a cohort of 406 consecutive cementless total hip prostheses, followed up for a minimum of ten years, both approaches were compared and the magnitude of potential bias was highlighted. The Kaplan-Meier method overestimated the incidence of revision by almost 4% (60% relative difference) in the simulations and more than 1% (31.3% relative difference) in the realistic data set. The cumulative incidence of competing risk approach allows for appropriate accounting of competing risk and, as such, offers an improved ability to compare survival results across studies.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 6 | Pages 785 - 789
1 Jun 2007
Ross R Mirza AH Norris HE Khatri M

Between January 1990 and December 2000 we carried out 226 SB Charité III disc replacements for lumbar disc degeneration in 160 patients. They were reviewed at a mean follow-up of 79 months (31 to 161) to determine the clinical and radiological outcome. The clinical results were collected by an independent observer, who was not involved in patient selection, treatment or follow-up, using a combination of outcome measures, including the Oswestry Disability Index. Pain was recorded using a visual analogue score, and the most recent radiographs were reviewed.

Survival of the device was analysed by the Kaplan-Meier method and showed a cumulative survival of 35% at 156 months when radiological failure was taken as the endpoint. The mean improvement in the Oswestry disability index scores after disc replacement was 14% (6% to 21%) and the mean improvement in the pain score was 1.6 (0.46 to 2.73), both falling below the clinically significant threshold. Removal of the implant was required in 12 patients, four because of implant failure.

These poor results indicate that further use of this implant is not justified.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 3 | Pages 299 - 305
1 Mar 2008
Kim Y

This study reviewed the results of a cementless anatomical femoral component to give immediate post-operative stability, and with a narrow distal section in order not to contact the femoral cortex in the diaphysis, ensuring exclusively metaphyseal loading.

A total of 471 patients (601 hips) who had a total hip replacement between March 1995 and February 2002 were included in the study. There were 297 men and 174 women. The mean age at the time of operation was 52.7 years (28 to 63). Clinical and radiological evaluation were performed at each follow-up. Bone densitometry was carried out on all patients two weeks after operation and at the final follow-up examination. The mean follow-up was 8.8 years (5 to 12).

The mean pre-operative Harris hip score was 41 points (16 to 54), which improved to a mean of 96 (68 to 100) at the final follow-up. No patient complained of thigh pain at any stage. No acetabular or femoral osteolysis was observed and no hip required revision for aseptic loosening of either component. Deep infection occurred in two hips (0.3%) which required revision. One hip (0.2%) required revision of the acetabular component for recurrent dislocation. Bone mineral densitometry revealed a minimal bone loss in the proximal femur.

This cementless anatomical femoral component with metaphyseal loading but without distal fixation produced satisfactory fixation and encourages proximal femoral loading.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 8 | Pages 1068 - 1072
1 Aug 2005
Morag G Zalzal P Liberman B Safir O Flint M Gross AE

Our aim was to determine if the height of the cup, lateralisation or the abduction angle correlated with functional outcome or survivorship in revision total hip replacement in patients with a previous diagnosis of developmental dysplasia of the hip. A retrospective investigation of 51 patients (63 hips) who had undergone revision total hip replacement was performed. The mean duration of follow-up was 119 months. Forty-one patients (52 hips) were available for both determination of functional outcome and survivorship analysis. Ten patients (11 hips) were only available for survivorship analysis.

The height of the cup was found to have a statistically significant correlation with functional outcome and a high hip centre correlated with a worse outcome score. Patients with a hip centre of less than 3.5 cm above the anatomical level had a statistically better survivorship of the cup than those with centres higher than this. Restoration of the height of the centre of the hip to as near the anatomical position as possible improved functional outcome and survivorship of the cup.