Advertisement for orthosearch.org.uk
Results 1 - 59 of 59
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 976 - 982
1 Nov 1998
Psychoyios V Crawford RW Murray DW O’Connor JJ

Many designs of unicompartmental knee replacement show early and mid-term failure due to polyethylene wear. We studied the wear rate of congruent polyethylene meniscal bearings retrieved from failed Oxford unicompartmental knee replacements. We examined 16 bearings, 0.8 to 12.8 years after implantation, measuring their thickness and comparing it with that of 14 unused bearings. The mean rate of penetration, which included the effects of wear at both upper and lower surfaces, was 0.036 mm per year (maximum 0.08). Bearings as thin as 3.5 mm wore no faster than thicker models, but ten with evidence of impingement had greater wear. The six bearings with no impingement showed a mean rate of penetration of 0.01 mm per year. In unicompartmental knee replacement, careful implantation of fully congruous meniscal bearings can avoid failure due to polyethylene wear


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 486 - 492
1 Apr 2013
Breeman S Campbell MK Dakin H Fiddian N Fitzpatrick R Grant A Gray A Johnston L MacLennan GS Morris RW Murray DW

There is conflicting evidence about the merits of mobile bearings in total knee replacement, partly because most randomised controlled trials (RCTs) have not been adequately powered. We report the results of a multicentre RCT of mobile versus fixed bearings. This was part of the knee arthroplasty trial (KAT), where 539 patients were randomly allocated to mobile or fixed bearings and analysed on an intention-to-treat basis. The primary outcome measure was the Oxford Knee Score (OKS) plus secondary measures including Short Form-12, EuroQol EQ-5D, costs, cost-effectiveness and need for further surgery. There was no significant difference between the groups pre-operatively: mean OKS was 17.18 (. sd. 7.60) in the mobile-bearing group and 16.49 (. sd. 7.40) in the fixed-bearing group. At five years mean OKS was 33.19 (. sd. 16.68) and 33.65 (. sd. 9.68), respectively. There was no significant difference between trial groups in OKS at five years (-1.12 (95% confidence interval -2.77 to 0.52) or any of the other outcome measures. Furthermore, there was no significant difference in the proportion of patients with knee-related re-operations or in total costs. In this appropriately powered RCT, over the first five years after total knee replacement functional outcomes, re-operation rates and healthcare costs appear to be the same irrespective of whether a mobile or fixed bearing is used. Cite this article: Bone Joint J 2013;95-B:486–92


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1261 - 1269
1 Jul 2021
Burger JA Zuiderbaan HA Sierevelt IN van Steenbergen L Nolte PA Pearle AD Kerkhoffs GMMJ

Aims. Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. Methods. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed. Results. In hospitals performing less than 100 cases per year, cemented mobile bearings reported comparable adjusted risks of revision as uncemented mobile bearings. However, in hospitals performing more than 100 cases per year, the adjusted risk of revision was higher for cemented mobile bearings compared to uncemented mobile bearings (hazard ratio 1.78 (95% confidence interval 1.34 to 2.35)). The adjusted risk of revision between cemented fixed bearing and uncemented mobile bearing was comparable, independent of annual hospital volume. In addition, 12.3% of uncemented mobile bearing, 20.3% of cemented mobile bearing, and 41.5% of uncemented fixed bearing revisions were for tibial component loosening. The figures for instability were 23.6%, 14.5% and 11.7%, respectively, and for periprosthetic fractures were 10.0%, 2.8%, and 3.5%. Bearing exchange was the type of revision in 40% of uncemented mobile bearing, 24.3% of cemented mobile bearing, and 5.3% cemented fixed bearing revisions. Conclusion. The findings of this study demonstrated improved survival with use of uncemented compared to cemented mobile bearings in medial UKA, only in those hospitals performing more than 100 cases per year. Cemented fixed bearings reported comparable survival results as uncemented mobile bearings, regardless of the annual hospital volume. The high rates of instability, periprosthetic fractures, and bearing exchange in uncemented mobile bearings emphasize the need for further research. Cite this article: Bone Joint J 2021;103-B(7):1261–1269


Bone & Joint Research
Vol. 8, Issue 11 | Pages 535 - 543
1 Nov 2019
Mohammad HR Campi S Kennedy JA Judge A Murray DW Mellon SJ

Objectives. The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. Methods. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation. Results. The wear rate for 1900 resin machined (n = 19), 1050 machined (n = 21), and 1050 moulded bearings (n = 23) were 60 µm/year (. sd. 42), 76 µm/year (. sd. 32), and 57 µm/year (. sd. 30), respectively. There was no significant difference between 1900 machined and 1050 machined (p = 0.20), but 1050 moulded had significantly less wear than the 1050 machined (p = 0.05). Increasing femoral (p < 0.001) and tibial (p < 0.001) component size were associated with increasing wear. Conclusion. Wear rate is similar with 1050 and 1900 resin, but lower with moulded bearings than machined bearings. The currently used Phase 3 bearings wear rate is low (1050 moulded, 57 µm/year), but higher than the previously reported Phase 2 bearings (1900 moulded, 20 µm/year). This is unlikely to be due to the change in polyethylene but may relate to the minimally invasive approach used with the Phase 3. This approach, as well as improving function and thus increasing activity levels, may increase the risk of surgical errors, such as impingement or bearing overhang, which can increase wear. Surgeons should aim to use 4 mm thick bearings rather than 3 mm thick bearings in young patients, unless they are small and need conservative bone resections. Cite this article: Bone Joint Res 2019;8:535–543


The Bone & Joint Journal
Vol. 106-B, Issue 8 | Pages 808 - 816
1 Aug 2024
Hall AJ Cullinan R Alozie G Chopra S Greig L Clarke J Riches PE Walmsley P Ohly NE Holloway N

Aims. Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice. Methods. This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m. 2. (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis. Results. Kaplan-Meier analysis demonstrated no difference in implant survival according to sex, age, or obesity status. Ten-year PROMs were available for 196/387 (50.6%) surviving patients and were excellent: mean Oxford Knee Score 34.4 (95% CI 32.7 to 36.1); mean Forgotten Joint Score (FJS) 51.2 (95% CI 16.1 to 86.3); mean EuroQol five-dimension five-level questionnaire score 69.9 (95% CI 46.8 to 93.0); 141/196 (71.9%) achieved the 22-point FJS patient-acceptable symptom state (PASS); and 156/196 (79.6%) were “very satisfied or satisfied”. Conclusion. This is the only large study reporting ten-year implant survival and functional outcomes of TKA using a cemented single-radius design and with a CS tibial bearing construct. The findings of excellent implant survival, safety, and functional outcomes indicate that this combination is a safe and effective option in routine TKA. Further investigation of this single-radius design TKA with CS tibial bearings with well-matched patient study groups will allow further insight into the performance of these implants. Cite this article: Bone Joint J 2024;106-B(8):808–816


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 470 - 475
1 Apr 2011
Kendrick BJL Simpson DJ Kaptein BL Valstar ER Gill HS Murray DW Price AJ

The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent, mobile, polyethylene bearing. Wear of polyethylene is a significant cause of revision surgery in UKR in the first decade, and the incidence increases in the second decade. Our study used model-based radiostereometric analysis to measure the combined wear of the upper and lower bearing surfaces in 13 medial-compartment Oxford UKRs at a mean of 20.9 years (17.2 to 25.9) post-operatively. The mean linear penetration of the polyethylene bearing was 1.04 mm (0.307 to 2.15), with a mean annual wear rate of 0.045 mm/year (0.016 to 0.099). The annual wear rate of the phase-2 bearings (mean 0.022 mm/year) was significantly less (p = 0.01) than that of phase-1 bearings (mean 0.07 mm/year). The linear wear rate of the Oxford UKR remains very low into the third decade. We believe that phase-2 bearings had lower wear rates than phase-1 implants because of the improved bearing design and surgical technique which decreased the incidence of impingement. We conclude that the design of the Oxford UKR gives low rates of wear in the long term


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 33 - 39
1 Jul 2019
Lachiewicz PF O’Dell JA

Aims. There is insufficient evidence to recommend the use of alternative polyethylene bearings in modular, fixed-bearing total knee arthroplasty (TKA). The purpose of this study was to compare standard polyethylene (SP) and highly crosslinked polyethylene (XLP) tibial liners in posterior-stabilized TKA, with osteolysis as the primary outcome and clinical results and the rate of re-operation as the secondary outcomes. Patients and Methods. This is a single-surgeon, prospective randomized study involving one design of modular posterior-stabilized TKA. An analysis of 122 TKAs with an SP compression moulded liner and 123 with an XLP liner was performed, with a mean follow-up of six years (2 to 11). Patients were evaluated clinically using the Knee Society score, Lower Extremity Activity Score (LEAS), and the presence of an effusion, and standard radiographs were assessed for radiolucent lines and osteolytic lesions. Results. Osteolysis was present in four TKAs (3.3%) in the SP group, and no knees in the XLP group (p = 0.06). There were no significant differences between the Knee Society total score, change in total score, knee function score, change in function score, LEAS, and change in LEAS in the two groups. There was a significant difference in the presence of an effusion (10/122 with SP liners, 1/123 with XLP liners; p = 0.02). There was no significant difference in the rate of re-operation between the two groups (p = 0.36). There were no complications related to the XLP liner. Conclusion. At this length of follow-up, there were no advantages and no complications related to the use of this XLP tibial liner. The presence of effusion and small osteolytic lesions was more frequent with SP than XLP liners, but of unknown clinical significance. Cite this article: Bone Joint J 2019;101-B(7 Supple C):33–39


Bone & Joint Research
Vol. 13, Issue 5 | Pages 226 - 236
9 May 2024
Jürgens-Lahnstein JH Petersen ET Rytter S Madsen F Søballe K Stilling M

Aims

Micromotion of the polyethylene (PE) inlay may contribute to backside PE wear in addition to articulate wear of total knee arthroplasty (TKA). Using radiostereometric analysis (RSA) with tantalum beads in the PE inlay, we evaluated PE micromotion and its relationship to PE wear.

Methods

A total of 23 patients with a mean age of 83 years (77 to 91), were available from a RSA study on cemented TKA with Maxim tibial components (Zimmer Biomet). PE inlay migration, PE wear, tibial component migration, and the anatomical knee axis were evaluated on weightbearing stereoradiographs. PE inlay wear was measured as the deepest penetration of the femoral component into the PE inlay.


Bone & Joint Research
Vol. 11, Issue 8 | Pages 575 - 584
17 Aug 2022
Stoddart JC Garner A Tuncer M Cobb JP van Arkel RJ

Aims

The aim of this study was to determine the risk of tibial eminence avulsion intraoperatively for bi-unicondylar knee arthroplasty (Bi-UKA), with consideration of the effect of implant positioning, overstuffing, and sex, compared to the risk for isolated medial unicondylar knee arthroplasty (UKA-M) and bicruciate-retaining total knee arthroplasty (BCR-TKA).

Methods

Two experimentally validated finite element models of tibia were implanted with UKA-M, Bi-UKA, and BCR-TKA. Intraoperative loads were applied through the condyles, anterior cruciate ligament (ACL), medial collateral ligament (MCL), and lateral collateral ligament (LCL), and the risk of fracture (ROF) was evaluated in the spine as the ratio of the 95th percentile maximum principal elastic strains over the tensile yield strain of proximal tibial bone.


Bone & Joint Open
Vol. 4, Issue 10 | Pages 776 - 781
16 Oct 2023
Matar HE Bloch BV James PJ

Aims

The aim of this study was to evaluate medium- to long-term outcomes and complications of the Stanmore Modular Individualised Lower Extremity System (SMILES) rotating hinge implant in revision total knee arthroplasty (rTKA) at a tertiary unit. It is hypothesized that this fully cemented construct leads to satisfactory clinical outcomes.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using the fully cemented SMILES rotating hinge prosthesis between 2005 to 2018. Outcome measures included aseptic loosening, reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both prospectively collected local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


Bone & Joint Open
Vol. 5, Issue 10 | Pages 937 - 943
22 Oct 2024
Gregor RH Hooper GJ Frampton C

Aims

The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs).

Methods

This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 916 - 923
1 Sep 2024
Fricka KB Wilson EJ Strait AV Ho H Hopper, Jr RH Hamilton WG Sershon RA

Aims

The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution.

Methods

Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 610 - 621
1 Jun 2023
Prodromidis AD Chloros GD Thivaios GC Sutton PM Pandit H Giannoudis PV Charalambous CP

Aims

Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods

A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 10 | Pages 1366 - 1371
1 Oct 2012
Pijls BG Valstar ER Kaptein BL Nelissen RGHH

This single-blinded randomised controlled trial investigated whether one design of mobile-bearing (MB) total knee replacement (TKR) has any advantage over a fixed-bearing (FB) design on long-term fixation as measured by radiostereometry. The amount of wear underneath the mobile bearing was also evaluated. A series of 42 knees was randomised to MB or FB tibial components with appropriate polyethylene inserts and followed for between ten and 12 years, or until the death of the patient. The polyethylene in the MB group was superior in that it was gamma-irradiated in inert gas and was calcium-stearate free; the polyethylene in the FB group was gamma-irradiated in air and contained calcium stearate. In theory this should be advantageous to the wear rate of the MB group. At final follow-up the overall mean migration was 0.75 mm (. sd. 0.76) in the MB group and 0.66 mm (. sd. 0.4) in the FB group, with the FB group demonstrating more posterior tilt and the MB group more internal rotation. In the FB group there was one revision for aseptic loosening, but none in the MB group. There were no significant differences in clinical or radiological scores. . For the MB group, the mean linear wear rate on the under-surface was 0.026 mm/year (. sd.  0.014). This was significantly smaller than the wear rate of 0.11 mm/year (. sd. 0.06) in the MB between femur and polyethylene (p < 0.001). Nevertheless, even in a best-case setting the mobile bearings of this TKR design had no apparent advantage in terms of fixation over the FB knee prosthesis at ten to 12 years. The wear underneath the mobile bearing was small and is unlikely to be clinically relevant


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 367 - 373
1 Mar 2010
Kendrick BJL Longino D Pandit H Svard U Gill HS Dodd CAF Murray DW Price AJ

The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (. sd. 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/− signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly


The Bone & Joint Journal
Vol. 95-B, Issue 9 | Pages 1209 - 1216
1 Sep 2013
van der Voort P Pijls BG Nouta KA Valstar ER Jacobs WCH Nelissen RGHH

Mobile-bearing (MB) total knee replacement (TKR) was introduced to reduce the risk of aseptic loosening and wear of polyethylene inserts. However, no consistent clinical advantages of mobile- over fixed-bearing (FB) TKR have been found. In this study we evaluated whether mobile bearings have an advantage over fixed bearings with regard to revision rates and clinical outcome scores. Furthermore, we determined which modifying variables affected the outcome. A systematic search of the literature was conducted to collect clinical trials comparing MB and FB in primary TKR. The primary outcomes were revision rates for any reason, aseptic loosening and wear. Secondary outcomes included range of movement, Knee Society score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score and radiological parameters. Meta-regression techniques were used to explore factors modifying the observed effect. . Our search yielded 1827 publications, of which 41 studies met our inclusion criteria, comprising over 6000 TKRs. Meta-analyses showed no clinically relevant differences in terms of revision rates, clinical outcome scores or patient-reported outcome measures between MB and FB TKRs. It appears that theoretical assumptions of superiority of MB over FB TKR are not borne out in clinical practice. Cite this article: Bone Joint J 2013;95-B:1209–16


Bone & Joint Open
Vol. 3, Issue 5 | Pages 390 - 397
1 May 2022
Hiranaka T Suda Y Saitoh A Tanaka A Arimoto A Koide M Fujishiro T Okamoto K

The kinematic alignment (KA) approach to total knee arthroplasty (TKA) has recently increased in popularity. Accordingly, a number of derivatives have arisen and have caused confusion. Clarification is therefore needed for a better understanding of KA-TKA. Calipered (or true, pure) KA is performed by cutting the bone parallel to the articular surface, compensating for cartilage wear. In soft-tissue respecting KA, the tibial cutting surface is decided parallel to the femoral cutting surface (or trial component) with in-line traction. These approaches are categorized as unrestricted KA because there is no consideration of leg alignment or component orientation. Restricted KA is an approach where the periarthritic joint surface is replicated within a safe range, due to concerns about extreme alignments that have been considered ‘alignment outliers’ in the neutral mechanical alignment approach. More recently, functional alignment and inverse kinematic alignment have been advocated, where bone cuts are made following intraoperative planning, using intraoperative measurements acquired with computer assistance to fulfill good coordination of soft-tissue balance and alignment. The KA-TKA approach aims to restore the patients’ own harmony of three knee elements (morphology, soft-tissue balance, and alignment) and eventually the patients’ own kinematics. The respective approaches start from different points corresponding to one of the elements, yet each aim for the same goal, although the existing implants and techniques have not yet perfectly fulfilled that goal.


Aims

The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit.

Methods

A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 145 - 149
1 Jun 2021
Crawford DA Passias BJ Adams JB Berend KR Lombardi AV

Aims

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


The Journal of Bone & Joint Surgery British Volume
Vol. 84-B, Issue 5 | Pages 653 - 657
1 Jul 2002
Robinson BJ Rees JL Price AJ Beard DJ Murray DW Smith PM Dodd CAF

When the Oxford unicompartmental meniscal bearing arthroplasty is used in the lateral compartment of the knee, 10% of the bearings dislocate. A radiological review was carried out to establish if dislocation was related to surgical technique. The postoperative radiographs of 46 lateral unicompartmental arthroplasties were analysed. Five variables which related to the position and alignment of the components were measured. Dislocations occurred in six knees. Only one of the five variables, the proximal tibial varus angle, had a statistically significant relationship to dislocation. This variable quantifies the height of the lateral joint line. The mean proximal tibial varus angle for knees the bearings of which had dislocated was 9° and for those which had not it was 5°. In both groups it was greater than would be expected in the normal knee (3°). Our study suggests that a high proximal tibial varus angle is associated with dislocation. The surgical technique should be modified to account for this, with care being taken to avoid damage to or over-distraction of the lateral soft tissues


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 6 | Pages 983 - 989
1 Nov 1998
Murray DW Goodfellow JW O’Connor JJ

Retrieval studies have shown that the use of fully congruent meniscal bearings reduces wear in knee replacements. We report the outcome of 143 knees with anteromedial osteoarthritis and normal anterior cruciate ligaments treated by unicompartmental arthroplasty using fully congruous mobile polyethylene bearings. At review, 34 knees were in patients who had died and 109 were in those who were still living. The mean elapsed time since operation was 7.6 years (maximum 13.8). We established the status of all but one knee. There had been five revision operations giving a cumulative prosthetic survival rate at ten years (33 knees at risk) of 98% (95% CI 93% to 100%). Considering the knee lost to follow-up as a failure, the ‘worst-case’ survival rate was 97%. No failures were due to polyethylene wear or aseptic loosening of the tibial component. One bearing which dislocated at four years was reduced by closed manipulation. The ten-year survival rate is the best of those reported for unicompartmental arthroplasty and not significantly different from the best rates for total knee replacement


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims

The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs).

Methods

At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays.


The Bone & Joint Journal
Vol. 103-B, Issue 1 | Pages 105 - 112
1 Jan 2021
Lynch JT Perriman DM Scarvell JM Pickering MR Galvin CR Neeman T Smith PN

Aims

Modern total knee arthroplasty (TKA) prostheses are designed to restore near normal kinematics including high flexion. Kneeling is a high flexion, kinematically demanding activity after TKA. The debate about design choice has not yet been informed by six-degrees-of-freedom in vivo kinematics. This prospective randomized clinical trial compared kneeling kinematics in three TKA designs.

Methods

In total, 68 patients were randomized to either a posterior stabilized (PS-FB), cruciate-retaining (CR-FB), or rotating platform (CR-RP) design. Of these patients, 64 completed a minimum one year follow-up. Patients completed full-flexion kneeling while being imaged using single-plane fluoroscopy. Kinematics were calculated by registering the 3D implant models onto 2D-dynamic fluoroscopic images and exported for analysis.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims

A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA).

Patients and Methods

Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1331 - 1347
1 Nov 2019
Jameson SS Asaad A Diament M Kasim A Bigirumurame T Baker P Mason J Partington P Reed M

Aims

Antibiotic-loaded bone cements (ALBCs) may offer early protection against the formation of bacterial biofilm after joint arthroplasty. Use in hip arthroplasty is widely accepted, but there is a lack of evidence in total knee arthroplasty (TKA). The objective of this study was to evaluate the use of ALBC in a large population of TKA patients.

Materials and Methods

Data from the National Joint Registry (NJR) of England and Wales were obtained for all primary cemented TKAs between March 2003 and July 2016. Patient, implant, and surgical variables were analyzed. Cox proportional hazards models were used to assess the influence of ALBC on risk of revision. Body mass index (BMI) data were available in a subset of patients.


Objectives

Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty for patients who require treatment of single-compartment osteoarthritis, especially for young patients. To satisfy this requirement, new patient-specific prosthetic designs have been introduced. The patient-specific UKA is designed on the basis of data from preoperative medical images. In general, knee implant design with increased conformity has been developed to provide lower contact stress and reduced wear on the tibial insert compared with flat knee designs. The different tibiofemoral conformity may provide designers the opportunity to address both wear and kinematic design goals simultaneously. The aim of this study was to evaluate wear prediction with respect to tibiofemoral conformity design in patient-specific UKA under gait loading conditions by using a previously validated computational wear method.

Methods

Three designs with different conformities were developed with the same femoral component: a flat design normally used in fixed-bearing UKA, a tibia plateau anatomy mimetic (AM) design, and an increased conforming design. We investigated the kinematics, contact stress, contact area, wear rate, and volumetric wear of the three different tibial insert designs.


The Bone & Joint Journal
Vol. 99-B, Issue 6 | Pages 793 - 798
1 Jun 2017
Anderson FL Koch CN Elpers ME Wright TM Haas SB Heyse TJ

Aims

We sought to establish whether an oxidised zirconium (OxZr) femoral component causes less loss of polyethylene volume than a cobalt alloy (CoCr) femoral component in total knee arthroplasty.

Materials and Methods

A total of 20 retrieved tibial inserts that had articulated with OxZr components were matched with 20 inserts from CoCr articulations for patient age, body mass index, length of implantation, and revision diagnosis. Changes in dimensions of the articular surfaces were compared with those of pristine inserts using laser scanning. The differences in volume between the retrieved and pristine surfaces of the two groups were calculated and compared.


Objectives

Posterior condylar offset (PCO) and posterior tibial slope (PTS) are critical factors in total knee arthroplasty (TKA). A computational simulation was performed to evaluate the biomechanical effect of PCO and PTS on cruciate retaining TKA.

Methods

We generated a subject-specific computational model followed by the development of ± 1 mm, ± 2 mm and ± 3 mm PCO models in the posterior direction, and -3°, 0°, 3° and 6° PTS models with each of the PCO models. Using a validated finite element (FE) model, we investigated the influence of the changes in PCO and PTS on the contact stress in the patellar button and the forces on the posterior cruciate ligament (PCL), patellar tendon and quadriceps muscles under the deep knee-bend loading conditions.


The Bone & Joint Journal
Vol. 98-B, Issue 10_Supple_B | Pages 34 - 40
1 Oct 2016
Emerson RH Alnachoukati O Barrington J Ennin K

Aims

Approved by the Food and Drug Administration in 2004, the Phase III Oxford Medial Partial Knee is used to treat anteromedial osteoarthritis (AMOA) in patients with an intact anterior cruciate ligament. This unicompartmental knee arthroplasty (UKA) is relatively new in the United States, and therefore long-term American results are lacking.

Patients and Methods

This is a single surgeon, retrospective study based on prospectively collected data, analysing a consecutive series of primary UKAs using the Phase III mobile-bearing Oxford Knee and Phase III instrumentation.

Between July 2004 and December 2006, the senior author (RHE) carried out a medial UKA in 173 patients (213 knees) for anteromedial osteoarthritis or avascular necrosis (AVN).

A total of 95 patients were men and 78 were women. Their mean age at surgery was 67 years (38 to 89) and mean body mass index 29.87 kg/m2 (17 to 62).

The mean follow-up was ten years (4 to 11).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1329 - 1334
1 Oct 2017
Lim JBT Chong HC Pang HN Tay KJD Chia SL Lo NN Yeo SJ

Aims

Little is known about the relative outcomes of revision of unicompartmental knee arthroplasty (UKA) and high tibial osteotomy (HTO) to total knee arthroplasty (TKA). The aim of this study is to compare the outcomes of revision surgery for the two procedures in terms of complications, re-revision and patient-reported outcome measures (PROMs) at a minimum of two years follow-up.

Patients and Methods

This study was a retrospective review of data from an institutional arthroplasty registry for cases performed between 2001 and 2014. A total of 292 patients were identified, of which 217 had a revision of HTO to TKA, and 75 had revision of UKA to TKA. While mean follow-up was longer for the HTO group compared with the UKA group, patient demographics (age, body mass index and Charlson co-morbidity index) and PROMs (Short Form-36, Oxford Knee Score, Knee Society Score, both objective and functional) were similar in the two groups prior to revision surgery. Outcomes included the rate of complications and re-operation, PROMS and patient-reported satisfaction at six months and two years post-operatively. We also compared the duration of surgery and the need for revision implants in the two groups.


The Bone & Joint Journal
Vol. 99-B, Issue 8 | Pages 996 - 1002
1 Aug 2017
Brown TS Van Citters DW Berry DJ Abdel MP

Advances in polyethylene (PE) in total hip arthroplasty have led to interest and increased use of highly crosslinked PE (HXLPE) in total knee arthroplasty (TKA). Biomechanical data suggest improved wear characteristics for HXLPE inserts over conventional PE in TKA. Short-term results from registry data and few clinical trials are promising. Our aim is to present a review of the history of HXLPEs, the use of HXLPE inserts in TKA, concerns regarding potential mechanical complications, and a thorough review of the available biomechanical and clinical data.

Cite this article: Bone Joint J 2017;99-B:996–1002.


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 786 - 792
1 Jun 2015
Hutt JRB Farhadnia P Massé V Lavigne M Vendittoli P

This randomised trial evaluated the outcome of a single design of unicompartmental arthroplasty of the knee (UKA) with either a cemented all-polyethylene or a metal-backed modular tibial component. A total of 63 knees in 45 patients (17 male, 28 female) were included, 27 in the all-polyethylene group and 36 in the metal-backed group. The mean age was 57.9 years (39.6 to 76.9). At a mean follow-up of 6.4 years (5 to 9.9), 11 all-polyethylene components (41%) were revised (at a mean of 5.8 years; 1.4 to 8.0) post-operatively and two metal-backed components were revised (at one and five years). One revision in both groups was for unexplained pain, one in the metal-backed group was for progression of osteoarthritis. The others in the all-polyethylene group were for aseptic loosening. The survivorship at seven years calculated by the Kaplan–Meier method for the all-polyethylene group was 56.5% (95% CI 31.9 to 75.2, number at risk 7) and for the metal-backed group was 93.8% (95% CI 77.3 to 98.4, number at risk 16) This difference was statistically significant (p <  0.001). At the most recent follow-up, significantly better mean Western Ontario and McMaster Universities Arthritis Index Scores were found in the all-polyethylene group (13.4 vs 23.0, p = 0.03) but there was no difference in the mean Knee injury and Osteoarthritis Outcome scores (68.8; 41.4 to 99.0 vs 62.6; 24.0 to 100.0), p = 0.36). There were no significant differences for range of movement (p = 0.36) or satisfaction (p = 0.23).

This randomised study demonstrates that all-polyethylene components in this design of fixed bearing UKA had unsatisfactory results with significantly higher rates of failure before ten years compared with the metal-back components.

Cite this article: Bone Joint J 2015;97-B:786–92.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1493 - 1499
1 Nov 2015
Pandit H Hamilton TW Jenkins C Mellon SJ Dodd CAF Murray DW

This prospective study reports the 15-year survival and ten-year functional outcome of a consecutive series of 1000 minimally invasive Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women, 52%, mean age 66 years; 32 to 88). These were implanted by two surgeons involved with the design of the prosthesis to treat anteromedial osteoarthritis and spontaneous osteonecrosis of the knee, which are recommended indications. Patients were prospectively identified and followed up independently for a mean of 10.3 years (5.3 to 16.6).

At ten years, the mean Oxford Knee Score was 40 (standard deviation (sd) 9; 2 to 48): 79% of knees (349) had an excellent or good outcome. There were 52 implant-related re-operations at a mean of 5.5 years (0.2 to 14.7). The most common reasons for re-operation were arthritis in the lateral compartment (2.5%, 25 knees), bearing dislocation (0.7%, seven knees) and unexplained pain (0.7%, seven knees). When all implant-related re-operations were considered as failures, the ten-year rate of survival was 94% (95% confidence interval (CI) 92 to 96) and the 15-year survival rate 91% (CI 83 to 98). When failure of the implant was the endpoint the 15-year survival was 99% (CI 96 to 100).

This is the only large series of minimally invasive UKAs with 15-year survival data. The results support the continued use of minimally invasive UKA for the recommended indications.

Cite this article: Bone Joint J 2015;97-B:1493–99.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 129 - 132
1 Nov 2013
Berend KR Lombardi Jr AV Adams JB

Debate has raged over whether a cruciate retaining (CR) or a posterior stabilised (PS) total knee replacement (TKR) provides a better range of movement (ROM) for patients. Various sub-sets of CR design are frequently lumped together when comparing outcomes. Additionally, multiple factors have been proven to influence the rate of manipulation under anaesthetic (MUA) following TKR. The purpose of this study was to determine whether different CR bearing insert designs provide better ROM or different MUA rates. All primary TKRs performed by two surgeons between March 2006 and March 2009 were reviewed and 2449 CR-TKRs were identified. The same CR femoral component, instrumentation, and tibial base plate were consistently used. In 1334 TKRs a CR tibial insert with 3° posterior slope and no posterior lip was used (CR-S). In 803 there was an insert with no slope and a small posterior lip (CR-L) and in 312 knees the posterior cruciate ligament (PCL) was either resected or lax and a deep-dish, anterior stabilised insert was used (CR-AS). More CR-AS inserts were used in patients with less pre-operative ROM and greater pre-operative tibiofemoral deformity and flexion contracture (p < 0.05). The mean improvement in ROM was highest for the CR-AS inserts (5.9° (-40° to 55°) vs CR-S 3.1° (-45° to 70°) vs CR-L 3.0° (-45° to 65°); p = 0.004). There was a significantly higher MUA rate with the CR-S and CR-L inserts than CR-AS (Pearson rank 6.51; p = 0.04). Despite sacrificing or not substituting for the PCL, ROM improvement was highest, and the MUA rate was lowest in TKRs with a deep-dish, anterior-stabilised insert. Substitution for the posterior cruciate ligament (PCL) in the form of a PS design may not be necessary even when the PCL is deficient.

Cite this article: Bone Joint J 2013;95-B, Supple A:129–32.


The Bone & Joint Journal
Vol. 98-B, Issue 1_Supple_A | Pages 105 - 112
1 Jan 2016
Khan M Osman K Green G Haddad FS

Total knee arthroplasty (TKA) is a cost effective and extremely successful operation. As longevity increases, the demand for primary TKA will continue to rise. The success and survivorship of TKAs are dependent on the demographics of the patient, surgical technique and implant-related factors.

Currently the risk of failure of a TKA requiring revision surgery ten years post-operatively is 5%.

The most common indications for revision include aseptic loosening (29.8%), infection (14.8%), and pain (9.5%). Revision surgery poses considerable clinical burdens on patients and financial burdens on healthcare systems.

We present a current concepts review on the epidemiology of failed TKAs using data from worldwide National Joint Registries.

Cite this article: Bone Joint J 2016;98-B(1 Suppl A):105–12.


Bone & Joint Research
Vol. 5, Issue 11 | Pages 552 - 559
1 Nov 2016
Kang K Koh Y Son J Kwon O Baek C Jung SH Park KK

Objectives

Malrotation of the femoral component can result in post-operative complications in total knee arthroplasty (TKA), including patellar maltracking. Therefore, we used computational simulation to investigate the influence of femoral malrotation on contact stresses on the polyethylene (PE) insert and on the patellar button as well as on the forces on the collateral ligaments.

Materials and Methods

Validated finite element (FE) models, for internal and external malrotations from 0° to 10° with regard to the neutral position, were developed to evaluate the effect of malrotation on the femoral component in TKA. Femoral malrotation in TKA on the knee joint was simulated in walking stance-phase gait and squat loading conditions.


The Bone & Joint Journal
Vol. 96-B, Issue 1 | Pages 59 - 64
1 Jan 2014
Weston-Simons JS Pandit H Kendrick BJL Jenkins C Barker K Dodd CAF Murray DW

Mobile-bearing unicompartmental knee replacements (UKRs) with a flat tibial plateau have not performed well in the lateral compartment, owing to a high dislocation rate. This led to the development of the Domed Lateral Oxford UKR (Domed OUKR) with a biconcave bearing. The aim of this study was to assess the survival and clinical outcomes of the Domed OUKR in a large patient cohort in the medium term.

We prospectively evaluated 265 consecutive knees with isolated disease of the lateral compartment and a mean age at surgery of 64 years (32 to 90). At a mean follow-up of four years (sd 2.2, (0.5 to 8.3)) the mean Oxford knee score was 40 out of 48 (sd 7.4). A total of 12 knees (4.5%) had re-operations, of which four (1.5%) were for dislocation. All dislocations occurred in the first two years. Two (0.8%) were secondary to significant trauma that resulted in ruptured ligaments, and two (0.8%) were spontaneous. In four patients (1.5%) the UKR was converted to a primary TKR. Survival at eight years, with failure defined as any revision, was 92.1% (95% confidence interval 81.3 to 100).

The Domed Lateral OUKR gives good clinical outcomes, low re-operation and revision rates and a low dislocation rate in patients with isolated lateral compartmental disease, in the hands of the designer surgeons.

Cite this article: Bone Joint J 2014;96-B:59–64.


The Bone & Joint Journal
Vol. 97-B, Issue 10_Supple_A | Pages 9 - 15
1 Oct 2015
Parratte S Ollivier M Lunebourg A Abdel MP Argenson J

Partial knee arthroplasty (PKA), either medial or lateral unicompartmental knee artroplasty (UKA) or patellofemoral arthroplasty (PFA) are a good option in suitable patients and have the advantages of reduced operative trauma, preservation of both cruciate ligaments and bone stock, and restoration of normal kinematics within the knee joint. However, questions remain concerning long-term survival. The goal of this review article was to present the long-term results of medial and lateral UKA, PFA and combined compartmental arthroplasty for multicompartmental disease. Medium- and long-term studies suggest reasonable outcomes at ten years with survival greater than 95% in UKA performed for medial osteoarthritis or osteonecrosis, and similarly for lateral UKA, particularly when fixed-bearing implants are used. Disappointing long-term outcomes have been observed with the first generation of patellofemoral implants, as well as early Bi-Uni (ie, combined medial and lateral UKA) or Bicompartmental (combined UKA and PFA) implants due to design and fixation issues. Promising short- and med-term results with the newer generations of PFAs and bicompartmental arthroplasties will require long-term confirmation.

Cite this article: Bone Joint J 2015;97-B(10 Suppl A):9–15.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 58 - 63
1 Jan 2009
Hooper G Rothwell A Frampton C

We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time.

At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p < 0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees.

There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p < 0.001).

The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 12 | Pages 1641 - 1648
1 Dec 2012
Baker PN Jameson SS Deehan DJ Gregg PJ Porter M Tucker K

Current analysis of unicondylar knee replacements (UKRs) by national registries is based on the pooled results of medial and lateral implants. Consequently, little is known about the differential performance of medial and lateral replacements and the influence of each implant type within these pooled analyses. Using data from the National Joint Registry for England and Wales (NJR) we aimed to determine the proportion of UKRs implanted on the lateral side of the knee, and their survival and reason for failure compared with medial UKRs. By combining information on the side of operation with component details held on the NJR, we were able to determine implant laterality (medial versus lateral) for 32 847 of the 35 624 unicondylar replacements (92%) registered before December 2010. Of these, 2052 (6%) were inserted on the lateral side of the knee. The rates of survival at five years were 93.1% (95% confidence interval (CI) 92.7 to 93.5) for medial and 93.0% (95% CI 91.1 to 94.9) for lateral UKRs (p = 0.49). The rates of failure remained equivalent after adjusting for patient age, gender, American Society of Anesthesiologists (ASA) grade, indication for surgery and implant design using Cox’s proportional hazards method (hazard ratio for lateral relative to medial replacement = 0.88 (95% CI 0.69 to 1.13); p = 0.32). Aseptic loosening/lysis and unexplained pain were the main reasons for revision in both groups, although the reasons did vary depending on whether a mobile- or a fixed-bearing design was used. At a maximum of eight years the mid-term survival rates of medial and lateral UKRs are similar.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1493 - 1497
1 Nov 2005
Price AJ Short A Kellett C Beard D Gill H Pandit H Dodd CAF Murray DW

Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the in vivo ten-year linear wear of the device, using a roentgenstereophotogrammetric technique.

In this in vivo study, seven medial Oxford unicompartmental prostheses, which had been implanted ten years previously were studied. Stereo pairs of radiographs were acquired for each patient and the films were analysed using a roentgen stereophotogrammetric analysis calibration and a computer-aided design model silhouette-fitting technique. Penetration of the femoral component into the original volume of the bearing was our estimate of linear wear. In addition, eight control patients were examined less than three weeks post-insertion of an Oxford prosthesis, where no wear would be expected. The control group showed no measured wear and suggested a system accuracy of 0.1 mm. At ten years, the mean linear wear rate was 0.02 mm/year.

The results from this in vivo study confirm that the device has low ten-year linear wear in clinical practice. This may offer the device a survival advantage in the long term.


The Bone & Joint Journal
Vol. 95-B, Issue 5 | Pages 636 - 642
1 May 2013
Gøthesen Ø Espehaug B Havelin L Petursson G Lygre S Ellison P Hallan G Furnes O

We evaluated the rates of survival and cause of revision of seven different brands of cemented primary total knee replacement (TKR) in the Norwegian Arthroplasty Register during the years 1994 to 2009. Revision for any cause, including resurfacing of the patella, was the primary endpoint. Specific causes of revision were secondary outcomes.

Three posterior cruciate-retaining (PCR) fixed modular-bearing TKRs, two fixed non-modular bearing PCR TKRs and two mobile-bearing posterior cruciate-sacrificing TKRs were investigated in a total of 17 782 primary TKRs. The median follow-up for the implants ranged from 1.8 to 6.9 years. Kaplan-Meier 10-year survival ranged from 89.5% to 95.3%. Cox’s relative risk (RR) was calculated relative to the fixed modular-bearing Profix knee (the most frequently used TKR in Norway), and ranged from 1.1 to 2.6. The risk of revision for aseptic tibial loosening was higher in the mobile-bearing LCS Classic (RR 6.8 (95% confidence interval (CI) 3.8 to 12.1)), the LCS Complete (RR 7.7 (95% CI 4.1 to 14.4)), the fixed modular-bearing Duracon (RR 4.5 (95% CI 1.8 to 11.1)) and the fixed non-modular bearing AGC Universal TKR (RR 2.5 (95% CI 1.3 to 5.1)), compared with the Profix. These implants (except AGC Universal) also had an increased risk of revision for femoral loosening (RR 2.3 (95% CI 1.1 to 4.8), RR 3.7 (95% CI 1.6 to 8.9), and RR 3.4 (95% CI 1.1 to 11.0), respectively). These results suggest that aseptic loosening is related to design in TKR.

Cite this article: Bone Joint J 2013;95-B:636–42.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 904 - 910
1 Jul 2007
Kim Y Kim D Kim J

We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1).

The total knee score, pain score, functional score and range of movement were not statistically different (p > 0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a mobile bearing.


Bone & Joint Research
Vol. 3, Issue 10 | Pages 289 - 296
1 Oct 2014
van IJsseldijk EA Harman MK Luetzner J Valstar ER Stoel BC Nelissen RGHH Kaptein BL

Introduction

Wear of polyethylene inserts plays an important role in failure of total knee replacement and can be monitored in vivo by measuring the minimum joint space width in anteroposterior radiographs. The objective of this retrospective cross-sectional study was to compare the accuracy and precision of a new model-based method with the conventional method by analysing the difference between the minimum joint space width measurements and the actual thickness of retrieved polyethylene tibial inserts.

Method

Before revision, the minimum joint space width values and their locations on the insert were measured in 15 fully weight-bearing radiographs. These measurements were compared with the actual minimum thickness values and locations of the retrieved tibial inserts after revision.


Bone & Joint Research
Vol. 3, Issue 10 | Pages 297 - 304
1 Oct 2014
Fitch DA Sedacki K Yang Y

Objectives

This systematic review and meta-analysis was conducted to determine the mid- to long-term clinical outcomes for a medial-pivot total knee replacement (TKR) system. The objectives were to synthesise available survivorship, Knee Society Scores (KSS), and reasons for revision for this system.

Methods

A systematic search was conducted of two online databases to identify sources of survivorship, KSS, and reasons for revision. Survivorship results were compared with values in the National Joint Registry of England, Wales, and Northern Ireland (NJR).


The Bone & Joint Journal
Vol. 95-B, Issue 10 | Pages 1339 - 1347
1 Oct 2013
Scott CEH Eaton MJ Nutton RW Wade FA Pankaj P Evans SL

As many as 25% to 40% of unicompartmental knee replacement (UKR) revisions are performed for pain, a possible cause of which is proximal tibial strain. The aim of this study was to examine the effect of UKR implant design and material on cortical and cancellous proximal tibial strain in a synthetic bone model. Composite Sawbone tibiae were implanted with cemented UKR components of different designs, either all-polyethylene or metal-backed. The tibiae were subsequently loaded in 500 N increments to 2500 N, unloading between increments. Cortical surface strain was measured using a digital image correlation technique. Cancellous damage was measured using acoustic emission, an engineering technique that detects sonic waves (‘hits’) produced when damage occurs in material.

Anteromedial cortical surface strain showed significant differences between implants at 1500 N and 2500 N in the proximal 10 mm only (p < 0.001), with relative strain shielding in metal-backed implants. Acoustic emission showed significant differences in cancellous bone damage between implants at all loads (p = 0.001). All-polyethylene implants displayed 16.6 times the total number of cumulative acoustic emission hits as controls. All-polyethylene implants also displayed more hits than controls at all loads (p < 0.001), more than metal-backed implants at loads ≥ 1500 N (p < 0.001), and greater acoustic emission activity on unloading than controls (p = 0.01), reflecting a lack of implant stiffness. All-polyethylene implants were associated with a significant increase in damage at the microscopic level compared with metal-backed implants, even at low loads. All-polyethylene implants should be used with caution in patients who are likely to impose large loads across their knee joint.

Cite this article: Bone Joint J 2013;95-B:1339–47.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 123 - 125
1 Nov 2012
Munro JT Garbuz DS Masri BA Duncan CP

Antibiotic impregnated articulating spacers are used in two-stage revision total knee arthroplasty to deliver local antibiotic therapy while preserving function. We have observed infection control in greater than 95% of cases with functional outcomes approaching those seen in revision for aseptic loosening. Higher failure has been observed with methicillin resistant organisms.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 354 - 359
1 Mar 2013
Chareancholvanich K Narkbunnam R Pornrattanamaneewong C

Patient-specific cutting guides (PSCGs) are designed to improve the accuracy of alignment of total knee replacement (TKR). We compared the accuracy of limb alignment and component positioning after TKR performed using PSCGs or conventional instrumentation. A total of 80 patients were randomised to undergo TKR with either of the different forms of instrumentation, and radiological outcomes and peri-operative factors such as operating time were assessed. No significant difference was observed between the groups in terms of tibiofemoral angle or femoral component alignment. Although the tibial component in the PSCGs group was measurably closer to neutral alignment than in the conventional group, the size of the difference was very small (89.8° (sd 1.2) vs 90.5° (sd 1.6); p = 0.030). This new technology slightly shortened the bone-cutting time by a mean of 3.6 minutes (p < 0.001) and the operating time by a mean 5.1 minutes (p = 0.019), without tangible differences in post-operative blood loss (p = 0.528) or need for blood transfusion (p = 0.789). This study demonstrated that both PSCGs and conventional instrumentation restore limb alignment and place the components with the similar accuracy. The minimal advantages of PSCGs in terms of consistency of alignment or operative time are unlikely to be clinically relevant.

Cite this article: Bone Joint J 2013;95-B:354–9.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 295 - 300
1 Mar 2013
Cawley DT Kelly N McGarry JP Shannon FJ

The optimum cementing technique for the tibial component in cemented primary total knee replacement (TKR) remains controversial. The technique of cementing, the volume of cement and the penetration are largely dependent on the operator, and hence large variations can occur. Clinical, experimental and computational studies have been performed, with conflicting results. Early implant migration is an indication of loosening. Aseptic loosening is the most common cause of failure in primary TKR and is the product of several factors. Sufficient penetration of cement has been shown to increase implant stability.

This review discusses the relevant literature regarding all aspects of the cementing of the tibial component at primary TKR.

Cite this article: Bone Joint J 2013;95-B:295–300.


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 436 - 444
1 Apr 2013
Scott CEH Nutton RW Biant LC

The lateral compartment is predominantly affected in approximately 10% of patients with osteoarthritis of the knee. The anatomy, kinematics and loading during movement differ considerably between medial and lateral compartments of the knee. This in the main explains the relative protection of the lateral compartment compared with the medial compartment in the development of osteoarthritis. The aetiology of lateral compartment osteoarthritis can be idiopathic, usually affecting the femur, or secondary to trauma commonly affecting the tibia. Surgical management of lateral compartment osteoarthritis can include osteotomy, unicompartmental knee replacement and total knee replacement. This review discusses the biomechanics, pathogenesis and development of lateral compartment osteoarthritis and its management.

Cite this article: Bone Joint J 2013;95-B:436–44.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1317 - 1323
1 Oct 2007
Kim Y Yoon S Kim J

We compared the results of 146 patients who received an anatomic modular knee fixed-bearing total knee replacement (TKR) in one knee and a low contact stress rotating platform mobile-bearing TKR in the other. There were 138 women and eight men with a mean age of 69.8 years (42 to 80). The mean follow-up was 13.2 years (11.0 to 14.5). The patients were assessed clinically and radiologically using the rating systems of the Hospital for Special Surgery and the Knee Society at three months, six months, one year, and annually thereafter.

The assessment scores of both rating systems pre-operatively and at the final review did not show any statistically significant differences between the two designs of implant. In the anatomic modular knee group, one knee was revised because of aseptic loosening of the tibial component and one because of infection. In addition, three knees were revised because of wear of the polyethylene tibial bearing. In the low contact stress group, two knees were revised because of instability requiring exchange of the polyethylene insert and one because of infection.

The radiological analysis found no statistical difference in the incidence of radiolucent lines at the final review (Student’s t-test, p = 0.08), most of which occurred at tibial zone 1. The Kaplan-Meier survivorship for aseptic loosening of the anatomic modular knee and the low contact stress implants at 14.5 years was 99% and 100%, respectively, with a 95% confidence interval of 94% to 100% for both designs.

We found no evidence of the superiority of one design over the other at long-term follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 5 | Pages 648 - 655
1 May 2012
Jolles BM Grzesiak A Eudier A Dejnabadi H Voracek C Pichonnaz C Aminian K Martin E

This study compared the outcome of total knee replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses during the first post-operative year and at five years’ follow-up, using gait parameters as a new objective measure. This double-blind randomised controlled clinical trial included 55 patients with mobile-bearing (n = 26) and fixed-bearing (n = 29) prostheses of the same design, evaluated pre-operatively and post-operatively at six weeks, three months, six months, one year and five years. Each participant undertook two walking trials of 30 m and completed the EuroQol questionnaire, Western Ontario and McMaster Universities osteoarthritis index, Knee Society score, and visual analogue scales for pain and stiffness. Gait analysis was performed using five miniature angular rate sensors mounted on the trunk (sacrum), each thigh and calf. The study population was divided into two groups according to age (≤ 70 years versus > 70 years).

Improvements in most gait parameters at five years’ follow-up were greater for fixed-bearing TKRs in older patients (> 70 years), and greater for mobile-bearing TKRs in younger patients (≤ 70 years). These findings should be confirmed by an extended age controlled study, as the ideal choice of prosthesis might depend on the age of the patient at the time of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 4 | Pages 484 - 489
1 Apr 2011
Charalambous CP Abiddin Z Mills SP Rogers S Sutton P Parkinson R

The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed mobile polyethylene bearing. We present the early results of the use of this prosthesis for established isolated patellofemoral arthritis in 51 consecutive patellofemoral replacements in 35 patients. The mean follow-up was 25 months (5 to 60). The estimated survival rate at three years was 63% (95% confidence interval 47 to 80) with revision as the endpoint and 46% (95% confidence interval 30 to 63) with revision and ongoing moderate or severe pain as the endpoint.

The early results of the use of the low contact stress patellofemoral replacement are disappointing with a high rate of revision. We cannot therefore recommend its use.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 198 - 204
1 Feb 2011
Pandit H Jenkins C Gill HS Barker K Dodd CAF Murray DW

This prospective study describes the outcome of the first 1000 phase 3 Oxford medial unicompartmental knee replacements (UKRs) implanted using a minimally invasive surgical approach for the recommended indications by two surgeons and followed up independently. The mean follow-up was 5.6 years (1 to 11) with 547 knees having a minimum follow-up of five years. At five years their mean Oxford knee score was 41.3 (sd 7.2), the mean American Knee Society Objective Score 86.4 (sd 13.4), mean American Knee Society Functional Score 86.1 (sd 16.6), mean Tegner activity score 2.8 (sd 1.1). For the entire cohort, the mean maximum flexion was 130° at the time of final review.

The incidence of implant-related re-operations was 2.9%; of these 29 re-operations two were revisions requiring revision knee replacement components with stems and wedges, 17 were conversions to a primary total knee replacement, six were open reductions for dislocation of the bearing, three were secondary lateral UKRs and one was revision of a tibial component. The most common reason for further surgical intervention was progression of arthritis in the lateral compartment (0.9%), followed by dislocation of the bearing (0.6%) and revision for unexplained pain (0.6%). If all implant-related re-operations are considered failures, the ten-year survival rate was 96% (95% confidence interval, 92.5 to 99.5). If only revisions requiring revision components are considered failures the ten-year survival rate is 99.8% (confidence interval 99 to 100).

This is the largest published series of UKRs implanted through a minimally invasive surgical approach and with ten-year survival data. The survival rates are similar to those obtained with a standard open approach whereas the function is better. This demonstrates the effectiveness and safety of a minimally invasive surgical approach for implanting the Oxford UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 1 | Pages 52 - 57
1 Jan 2009
Newman J Pydisetty RV Ackroyd C

Between 1989 and 1992 we had 102 knees suitable for unicompartmental knee replacement (UKR). They were randomised to receive either a St Georg Sled UKR or a Kinematic modular total knee replacement (TKR). The early results demonstrated that the UKR group had less complications and more rapid rehabilitation than the TKR group. At five years there were an equal number of failures in the two groups but the UKR group had more excellent results and a greater range of movement.

The cases were reviewed by a research nurse at 8, 10 and 12 years after operation. We report the outcome at 15 years follow-up. A total of 43 patients (45 knees) died with their prosthetic knees intact. Throughout the review period the Bristol knee scores of the UKR group have been better and at 15 years 15 (71.4%) of the surviving UKRs and 10 (52.6%) of the surviving TKRs had achieved an excellent score. The 15 years survivorship rate based on revision or failure for any reason was 24 (89.8%) for UKR and 19 (78.7%) for TKR. During the 15 years of the review four UKRs and six TKRs failed.

The better early results with UKR are maintained at 15 years with no greater failure rate. The median Bristol knee score of the UKR group was 91.1 at five years and 92 at 15 years, suggesting little functional deterioration in either the prosthesis or the remainder of the joint. These results justify the increased use of UKR.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1016 - 1021
1 Aug 2006
Delport HP Banks SA De Schepper J Bellemans J

Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the in vivo kinematics of this new group of implants. We investigated 31 patients who had undergone a total knee replacement with a similar prosthetic design but with three different options: fixed-bearing posterior cruciate ligament-retaining, fixed-bearing posterior-stabilised and mobile-bearing posterior-stabilised. To do this we used a three-dimensional to two-dimensional model registration technique. Both the fixed- and mobile-bearing posterior-stabilised configurations used the same femoral component. We found that fixed-bearing posterior stabilised and mobile-bearing posterior-stabilised knee replacements demonstrated similar kinematic patterns, with consistent femoral roll-back during flexion. Mobile-bearing posterior-stabilised knee replacements demonstrated greater and more natural internal rotation of the tibia during flexion than fixed-bearing posterior-stabilised designs. Such rotation occurred at the interface between the insert and tibial tray for mobile-bearing posterior-stabilised designs. However, for fixed-bearing posterior-stabilised designs, rotation occurred at the proximal surface of the bearing. Posterior cruciate ligament-retaining knee replacements demonstrated paradoxical sliding forward of the femur.

We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 11 | Pages 1488 - 1492
1 Nov 2005
Price AJ Dodd CAF Svard UGC Murray DW

We present a comparison of the results of the Oxford unicompartmental knee arthroplasty in patients younger and older than 60 years of age. The ten-year all-cause survival of the < 60 years of age group (52) was 91% (95% confidence interval (CI) 12), while in the ≥ 60 years of age group (512), the figure was 96% (95% CI 3). For the younger group, the mean Hospital for Special Surgery score at ten-year follow-up (n = 21) was 94 of 100, compared with a mean of 86 of 100 for the older group (n = 135). The results show that the Oxford unicompartmental arthroplasty can achieve ten-year results that are comparable to total knee arthroplasty in patients < 60 years of age. We conclude that for patients aged over 50, age should not be considered a contraindication for this procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 1 | Pages 54 - 60
1 Jan 2006
Pandit H Jenkins C Barker K Dodd CAF Murray DW

This prospective study describes the complications and survival of the first 688 Phase 3 Oxford medial unicompartmental knee replacements implanted using a minimally-invasive technique by two surgeons and followed up independently. None was lost to follow-up. We had carried out 132 of the procedures more than five years ago. The clinical assessment of 101 of these which were available for review at five years is also presented.

Nine of the 688 knees were revised: four for infection, three for dislocation of the bearing and two for unexplained pain. A further seven knees (1%) required other procedures: four had a manipulation under anaesthesia, two an arthroscopy and one a debridement for superficial infection. The survival rate at seven years was 97.3% (95% confidence interval 5.3). At five years, 96% of the patients had a good or excellent American Knee Society score, the mean Oxford knee score was 39 and the mean flexion was 133°. This study demonstrates that the minimally-invasive Oxford unicompartmental knee replacement is a reliable and effective procedure.