Advertisement for orthosearch.org.uk
Results 1 - 42 of 42
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 7 | Pages 889 - 894
1 Jul 2012
Burston BJ Barnett AJ Amirfeyz R Yates PJ Bannister GC

We prospectively followed 191 consecutive collarless polished tapered (CPT) femoral stems, implanted in 175 patients who had a mean age at operation of 64.5 years (21 to 85). At a mean follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were still alive. The fate of all original stems is known. The 16-year survivorship with re-operation for any reason was 80.7% (95% confidence interval 72 to 89.4). There was no loss to follow-up, with clinical data available on all 95 hips and radiological assessment performed on 90 hips (95%). At latest follow-up, the mean Harris hip score was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48). Stems subsided within the cement mantle, with a mean subsidence of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem (0.5%) has been revised due to aseptic loosening. In total seven stems were revised for any cause, of which four revisions were required for infection following revision of the acetabular component. A total of 21 patients (11%) required some sort of revision procedure; all except three of these resulted from failure of the acetabular component. Cemented acetabular components had a significantly lower revision burden (three hips, 2.7%) than Harris Galante uncemented components (17 hips, 21.8%) (p < 0.001). The CPT stem continues to provide excellent radiological and clinical outcomes at 15 years following implantation. Its results are consistent with other polished tapered stem designs


Bone & Joint Research
Vol. 11, Issue 5 | Pages 270 - 277
6 May 2022
Takegami Y Seki T Osawa Y Imagama S

Aims. Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems. Methods. We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless polished tapered (CPT) stem, Versys Advocate (Versys) stem, and Charnley-Marcel-Kerboull (CMK) stem. Experienced surgeons implanted each of these types of stems into six bone analogues, and the analogues were compressed and internally rotated until failure. Torque to fracture and fracture type were recorded. We also measured surface strain distribution using triaxial rosettes. Results. There was a significant difference in fracture torque between the three stem types (p = 0.036). Particularly, the median fracture torque for the CPT stem was significantly lower than that for the CMK stem (CPT vs CMK: 164.5 Nm vs 200.5 Nm; p = 0.046). The strain values for the CPT stem were higher than those for the other two stems at the most proximal site. The fracture pattern of the CPT and Versys stems was Vancouver type B, whereas that of the CMK stem was type C. Conclusion. Our study suggested that the cobalt-chromium alloy material, polished surface finish, acute-square proximal form, and the absence of a collar may be associated with lower fracture torque, which may be related to PPF. Cite this article: Bone Joint Res 2022;11(5):270–277


The Bone & Joint Journal
Vol. 97-B, Issue 6 | Pages 762 - 770
1 Jun 2015
Pennington MW Grieve R van der Meulen JH

There is little evidence on the cost effectiveness of different brands of hip prostheses. We compared lifetime cost effectiveness of frequently used brands within types of prosthesis including cemented (Exeter V40 Contemporary, Exeter V40 Duration and Exeter V40 Elite Plus Ogee), cementless (Corail Pinnacle, Accolade Trident, and Taperloc Exceed) and hybrid (Exeter V40 Trilogy, Exeter V40 Trident, and CPT Trilogy). We used data from three linked English national databases to estimate the lifetime risk of revision, quality-adjusted life years (QALYs) and cost. . For women with osteoarthritis aged 70 years, the Exeter V40 Elite Plus Ogee had the lowest risk of revision (5.9% revision risk, 9.0 QALYs) and the CPT Trilogy had the highest QALYs (10.9% revision risk, 9.3 QALYs). Compared with the Corail Pinnacle (9.3% revision risk, 9.22 QALYs), the most commonly used brand, and assuming a willingness-to-pay of £20 000 per QALY gain, the CPT Trilogy is most cost effective, with an incremental net monetary benefit of £876. Differences in cost effectiveness between the hybrid CPT Trilogy and Exeter V40 Trident and the cementless Corail Pinnacle and Taperloc Exceed were small, and a cautious interpretation is required, given the limitations of the available information. However, it is unlikely that cemented brands are among the most cost effective. Similar patterns of results were observed for men and other ages. The gain in quality of life after total hip arthroplasty, rather than the risk of revision, was the main driver of cost effectiveness. Cite this article: Bone Joint J 2015;97-B:762–70


The Bone & Joint Journal
Vol. 98-B, Issue 10 | Pages 1347 - 1354
1 Oct 2016
Palan J Smith MC Gregg P Mellon S Kulkarni A Tucker K Blom AW Murray DW Pandit H

Aims. Periprosthetic fracture (PF) after primary total hip arthroplasty (THA) is an uncommon but potentially devastating complication. This study aims to investigate the influence of cemented stem designs on the risk of needing a revision for a PF. Patients and Methods. We analysed data on 257 202 primary THAs with cemented stems and 390 linked first revisions for PF recorded in the National Joint Registry (NJR) of England, Wales and Northern Ireland to determine if a cemented femoral stem brand was associated with the risk of having revision for a PF after primary THA. All cemented femoral stem brands with more than 10 000 primary operations recorded in the NJR were identified. The four most commonly used cemented femoral stems were the Exeter V40 (n = 146 409), CPT (n = 24 300), C-Stem (n = 15 113) and Charnley (n = 20 182). We compared the revision risk ratios due to PF amongst the stems using a Poisson regression model adjusting for patient factors. Compared with the Exeter V40, the age, gender and ASA grade adjusted revision rate ratio was 3.89 for the cemented CPT stem (95% confidence interval (CI) 3.07 to 4.93), 0.89 for the C-Stem (95% CI 0.57 to 1.41) and 0.41 for the Charnley stem (95% CI 0.24 to 0.70). Conclusions. The limitations of the study include incomplete data capture, analysis of only PF requiring revision and that observation does not imply causality. Nevertheless, this study demonstrates that the choice of a cemented stem may influence the risk of revision for PF. Cite this article: Bone Joint J 2016;98-B:1347–54


Bone & Joint Research
Vol. 6, Issue 5 | Pages 351 - 357
1 May 2017
Takahashi E Kaneuji A Tsuda R Numata Y Ichiseki T Fukui K Kawahara N

Objectives. Favourable results for collarless polished tapered stems have been reported, and cement creep due to taper slip may be a contributing factor. However, the ideal cement thickness around polished stems remains unknown. We investigated the influence of cement thickness on stem subsidence and cement creep. Methods. We cemented six collarless polished tapered (CPT) stems (two stems each of small, medium and large sizes) into composite femurs that had been reamed with a large CPT rasp to achieve various thicknesses of the cement mantle. Two or three tantalum balls were implanted in the proximal cement in each femur. A cyclic loading test was then performed for each stem. The migration of the balls was measured three-dimensionally, using a micro-computed tomography (CT) scanner, before and after loading. A digital displacement gauge was positioned at the stem shoulder, and stem subsidence was measured continuously by the gauge. Final stem subsidence was measured at the balls at the end of each stem. Results. A strong positive correlation was observed between mean cement thickness and stem subsidence in the CT slices on the balls. In the small stems, the balls moved downward to almost the same extent as the stem. There was a significant negative correlation between cement thickness and the horizontal:downward ratio of ball movement. Conclusion. Collarless polished tapered stems with thicker cement mantles resulted in greater subsidence of both stem and cement. This suggests that excessive thickness of the cement mantle may interfere with effective radial cement creep. Cite this article: E. Takahashi, A. Kaneuji, R. Tsuda, Y. Numata, T. Ichiseki, K. Fukui, N. Kawahara. The influence of cement thickness on stem subsidence and cement creep in a collarless polished tapered stem: When are thick cement mantles detrimental? Bone Joint Res 2017;6:–357. DOI: 10.1302/2046-3758.65.BJR-2017-0028.R1


Aims

The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless polished taper slip stem (PTS) for management of an intracapsular hip fracture.

Methods

This retrospective study included hip fracture patients aged 50 years and older treated with Exeter (PTS) bipolar hemiarthroplasty between 2019 and 2022. Patient demographics, place of domicile, fracture type, delirium status, American Society of Anesthesiologists (ASA) grade, length of stay, and mortality were collected. Reoperation and mortality were recorded up to a median follow-up of 29.5 months (interquartile range 12 to 51.4). Cox regression was performed to evaluate independent risk factors associated with reoperation and mortality.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 466 - 475
8 Jul 2021
Jain S Lamb J Townsend O Scott CEH Kendrick B Middleton R Jones SA Board T West R Pandit H

Aims

This study evaluates risk factors influencing fracture characteristics for postoperative periprosthetic femoral fractures (PFFs) around cemented stems in total hip arthroplasty.

Methods

Data were collected for PFF patients admitted to eight UK centres between 25 May 2006 and 1 March 2020. Radiographs were assessed for Unified Classification System (UCS) grade and AO/OTA type. Statistical comparisons investigated relationships by age, gender, and stem fixation philosophy (polished taper-slip (PTS) vs composite beam (CB)). The effect of multiple variables was estimated using multinomial logistic regression to estimate odds ratios (ORs) with 95% confidence intervals (CIs). Surgical treatment (revision vs fixation) was compared by UCS grade and AO/OTA type.


Bone & Joint Open
Vol. 2, Issue 7 | Pages 476 - 485
8 Jul 2021
Scheerlinck T De Winter E Sas A Kolk S Van Gompel G Vandemeulebroucke J

Aims

Hip arthroplasty does not always restore normal anatomy. This is due to inaccurate surgery or lack of stem sizes. We evaluated the aptitude of four total hip arthroplasty systems to restore an anatomical and medialized hip rotation centre.

Methods

Using 3D templating software in 49 CT scans of non-deformed femora, we virtually implanted: 1) small uncemented calcar-guided stems with two offset options (Optimys, Mathys), 2) uncemented straight stems with two offset options (Summit, DePuy Synthes), 3) cemented undersized stems (Exeter philosophy) with three offset options (CPT, ZimmerBiomet), and 4) cemented line-to-line stems (Kerboul philosophy) with proportional offsets (Centris, Mathys). We measured the distance between the templated and the anatomical and 5 mm medialized hip rotation centre.


The Bone & Joint Journal
Vol. 103-B, Issue 8 | Pages 1339 - 1344
1 Aug 2021
Jain S Mohrir G Townsend O Lamb JN Palan J Aderinto J Pandit H

Aims

This aim of this study was to assess the reliability and validity of the Unified Classification System (UCS) for postoperative periprosthetic femoral fractures (PFFs) around cemented polished taper-slip (PTS) stems.

Methods

Radiographs of 71 patients with a PFF admitted consecutively at two centres between 25 February 2012 and 19 May 2020 were collated by an independent investigator. Six observers (three hip consultants and three trainees) were familiarized with the UCS. Each PFF was classified on two separate occasions, with a mean time between assessments of 22.7 days (16 to 29). Interobserver reliability for more than two observers was assessed using percentage agreement and Fleiss’ kappa statistic. Intraobserver reliability between two observers was calculated with Cohen kappa statistic. Validity was tested on surgically managed UCS type B PFFs where stem stability was documented in operation notes (n = 50). Validity was assessed using percentage agreement and Cohen kappa statistic between radiological assessment and intraoperative findings. Kappa statistics were interpreted using Landis and Koch criteria. All six observers were blinded to operation notes and postoperative radiographs.


Bone & Joint Open
Vol. 2, Issue 1 | Pages 16 - 21
1 Jan 2021
Kerzner B Kunze KN O’Sullivan MB Pandher K Levine BR

Aims

Advances in surgical technique and implant design may influence the incidence and mechanism of failure resulting in revision total hip arthroplasty (rTHA). The purpose of the current study was to characterize aetiologies requiring rTHA, and to determine whether temporal changes existed in these aetiologies over a ten-year period.

Methods

All rTHAs performed at a single institution from 2009 to 2019 were identified. Demographic information and mode of implant failure was obtained for all patients. Data for rTHA were stratified into two time periods to assess for temporal changes: 2009 to 2013, and 2014 to 2019. Operative reports, radiological imaging, and current procedural terminology (CPT) codes were cross-checked to ensure the accurate classification of revision aetiology for each patient.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1222 - 1230
1 Jul 2021
Slullitel PA Garcia-Barreiro GG Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims

We aimed to compare the implant survival, complications, readmissions, and mortality of Vancouver B2 periprosthetic femoral fractures (PFFs) treated with internal fixation with that of B1 PFFs treated with internal fixation and B2 fractures treated with revision arthroplasty.

Methods

We retrospectively reviewed the data of 112 PFFs, of which 47 (42%) B1 and 27 (24%) B2 PFFs were treated with internal fixation, whereas 38 (34%) B2 fractures underwent revision arthroplasty. Decision to perform internal fixation for B2 PFFs was based on specific radiological (polished femoral components, intact bone-cement interface) and clinical criteria (low-demand patient). Median follow-up was 36.4 months (24 to 60). Implant survival and mortality over time were estimated with the Kaplan-Meier method. Adverse events (measured with a modified Dindo-Clavien classification) and 90-day readmissions were additionally compared between groups.


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1215 - 1221
1 Jul 2021
Kennedy JW Ng NYB Young D Kane N Marsh AG Meek RMD

Aims

Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components.

Methods

From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 38 - 45
1 Jul 2021
Horberg JV Coobs BR Jiwanlal AK Betzle CJ Capps SG Moskal JT

Aims

Use of the direct anterior approach (DAA) for total hip arthroplasty (THA) has increased in recent years due to proposed benefits, including a lower risk of dislocation and improved early functional recovery. This study investigates the dislocation rate in a non-selective, consecutive cohort undergoing THA via the DAA without any exclusion or bias in patient selection based on habitus, deformity, age, sex, or fixation method.

Methods

We retrospectively reviewed all patients undergoing THA via the DAA between 2011 and 2017 at our institution. Primary outcome was dislocation at minimum two-year follow-up. Patients were stratified by demographic details and risk factors for dislocation, and an in-depth analysis of dislocations was performed.


The Bone & Joint Journal
Vol. 103-B, Issue 6 | Pages 1078 - 1087
1 Jun 2021
Awad ME Farley BJ Mostafa G Saleh KJ

Aims

It has been suggested that the direct anterior approach (DAA) should be used for total hip arthroplasty (THA) instead of the posterior approach (PA) for better early functional outcomes. We conducted a value-based analysis of the functional outcome and associated perioperative costs, to determine which surgical approach gives the better short-term outcomes and lower costs.

Methods

This meta-analysis was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) protocol and the Cochrane Handbook. Several online databases were searched. Non-stratified and stratified meta-analyses were conducted to test the confounding biases in the studies which were included. The mean cost and probability were used to determine the added costs of perioperative services.


Bone & Joint Open
Vol. 2, Issue 6 | Pages 443 - 456
28 Jun 2021
Thompson JW Corbett J Bye D Jones A Tissingh EK Nolan J

Aims

The Exeter V40 cemented polished tapered stem system has demonstrated excellent long-term outcomes. This paper presents a systematic review of the existing literature and reports on a large case series comparing implant fractures between the Exeter V40 series; 125 mm and conventional length stem systems.

Methods

A systematic literature search was performed adhering to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) criteria. In parallel, we performed a retrospective single centre study of Exeter V40 femoral stem prosthetic fractures between April 2003 and June 2020.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1565 - 1571
1 Dec 2018
Kristensen TB Dybvik E Furnes O Engesæter LB Gjertsen J

Aims

The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck.

Patients and Methods

From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis.


The Bone & Joint Journal
Vol. 102-B, Issue 7 Supple B | Pages 99 - 104
1 Jul 2020
Shah RF Bini S Vail T

Aims

Natural Language Processing (NLP) offers an automated method to extract data from unstructured free text fields for arthroplasty registry participation. Our objective was to investigate how accurately NLP can be used to extract structured clinical data from unstructured clinical notes when compared with manual data extraction.

Methods

A group of 1,000 randomly selected clinical and hospital notes from eight different surgeons were collected for patients undergoing primary arthroplasty between 2012 and 2018. In all, 19 preoperative, 17 operative, and two postoperative variables of interest were manually extracted from these notes. A NLP algorithm was created to automatically extract these variables from a training sample of these notes, and the algorithm was tested on a random test sample of notes. Performance of the NLP algorithm was measured in Statistical Analysis System (SAS) by calculating the accuracy of the variables collected, the ability of the algorithm to collect the correct information when it was indeed in the note (sensitivity), and the ability of the algorithm to not collect a certain data element when it was not in the note (specificity).


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1447 - 1458
1 Nov 2019
Chatziagorou G Lindahl H Kärrholm J

Aims

We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis.

Patients and Methods

The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.


The Bone & Joint Journal
Vol. 99-B, Issue 5 | Pages 585 - 591
1 May 2017
Buckland AJ Puvanesarajah V Vigdorchik J Schwarzkopf R Jain A Klineberg EO Hart RA Callaghan JJ Hassanzadeh H

Aims

Lumbar fusion is known to reduce the variation in pelvic tilt between standing and sitting. A flexible lumbo-pelvic unit increases the stability of total hip arthroplasty (THA) when seated by increasing anterior clearance and acetabular anteversion, thereby preventing impingement of the prosthesis. Lumbar fusion may eliminate this protective pelvic movement. The effect of lumbar fusion on the stability of total hip arthroplasty has not previously been investigated.

Patients and Methods

The Medicare database was searched for patients who had undergone THA and spinal fusion between 2005 and 2012. PearlDiver software was used to query the database by the International Classification of Diseases, 9th Revision, Clinical Modification (ICD-9-CM) procedural code for primary THA and lumbar spinal fusion. Patients who had undergone both lumbar fusion and THA were then divided into three groups: 1 to 2 levels, 3 to 7 levels and 8+ levels of fusion. The rate of dislocation in each group was established using ICD-9-CM codes. Patients who underwent THA without spinal fusion were used as a control group. Statistical significant difference between groups was tested using the chi-squared test, and significance set at p < 0.05.


The Bone & Joint Journal
Vol. 102-B, Issue 3 | Pages 293 - 300
1 Mar 2020
Zheng H Gu H Shao H Huang Y Yang D Tang H Zhou Y

Aims

Vancouver type B periprosthetic femoral fractures (PFF) are challenging complications after total hip arthroplasty (THA), and some treatment controversies remain. The objectives of this study were: to evaluate the short-to-mid-term clinical outcomes after treatment of Vancouver type B PFF and to compare postoperative outcome in subgroups according to classifications and treatments; to report the clinical outcomes after conservative treatment; and to identify risk factors for postoperative complications in Vancouver type B PFF.

Methods

A total of 97 consecutive PPFs (49 males and 48 females) were included with a mean age of 66 years (standard deviation (SD) 14.9). Of these, 86 patients were treated with surgery and 11 were treated conservatively. All living patients had a minimum two-year follow-up. Patient demographics details, fracture healing, functional scores, and complications were assessed. Clinical outcomes between internal fixation and revisions in patients with or without a stable femoral component were compared. Conservatively treated PPFs were evaluated in terms of mortality and healing status. A logistic regression analysis was performed to identify risk factors for complications.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 91 - 96
1 Jun 2019
Smith A Denehy K Ong KL Lau E Hagan D Malkani A

Aims

Cephalomedullary nails (CMNs) are commonly used for the treatment of intertrochanteric hip fractures. Total hip arthroplasty (THA) may be used as a salvage procedure when fixation fails in these patients. The aim of this study was to analyze the complications of THA following failed intertrochanteric hip fracture fixation using a CMN.

Patients and Methods

Patients who underwent THA were identified from the 5% subset of Medicare Parts A/B between 2002 and 2015. A subgroup involving those with an intertrochanteric fracture that was treated using a CMN during the previous five years was identified and compared with the remaining patients who underwent THA. The length of stay (LOS) was compared using both univariate and multivariate analysis. The incidence of infection, dislocation, revision, and re-admission was compared between the two groups, using multivariate analysis adjusted for demographic, hospital, and clinical factors.


The Bone & Joint Journal
Vol. 98-B, Issue 7 | Pages 917 - 924
1 Jul 2016
Whittaker RK Hothi HS Meswania JM Berber R Blunn GW Skinner JA Hart AJ

Aims

Surgeons have commonly used modular femoral heads and stems from different manufacturers, although this is not recommended by orthopaedic companies due to the different manufacturing processes.

We compared the rate of corrosion and rate of wear at the trunnion/head taper junction in two groups of retrieved hips; those with mixed manufacturers (MM) and those from the same manufacturer (SM).

Materials and Methods

We identified 151 retrieved hips with large-diameter cobalt-chromium heads; 51 of two designs that had been paired with stems from different manufacturers (MM) and 100 of seven designs paired with stems from the same manufacturer (SM). We determined the severity of corrosion with the Goldberg corrosion score and the volume of material loss at the head/stem junction. We used multivariable statistical analysis to determine if there was a significant difference between the two groups.


The Bone & Joint Journal
Vol. 101-B, Issue 4 | Pages 415 - 425
1 Apr 2019
Thewlis D Bahl JS Fraysse F Curness K Arnold JB Taylor M Callary S Solomon LB

Aims

The purpose of this exploratory study was to investigate if the 24-hour activity profile (i.e. waking activities and sleep) objectively measured using wrist-worn accelerometry of patients scheduled for total hip arthroplasty (THA) improves postoperatively.

Patients and Methods

A total of 51 THA patients with a mean age of 64 years (24 to 87) were recruited from a single public hospital. All patients underwent THA using the same surgical approach with the same prosthesis type. The 24-hour activity profiles were captured using wrist-worn accelerometers preoperatively and at 2, 6, 12, and 26 weeks postoperatively. Patient-reported outcomes (Hip Disability and Osteoarthritis Outcome Score (HOOS)) were collected at all timepoints except two weeks postoperatively. Accelerometry data were used to quantify the intensity (sedentary, light, moderate, and vigorous activities) and frequency (bouts) of activity during the day and sleep efficiency. The analysis investigated changes with time and differences between Charnley class.


The Bone & Joint Journal
Vol. 100-B, Issue 7 | Pages 867 - 874
1 Jul 2018
Makarewich CA Anderson MB Gililland JM Pelt CE Peters CL

Aims

For this retrospective cohort study, patients aged ≤ 30 years (very young) who underwent total hip arthroplasty (THA) were compared with patients aged ≥ 60 years (elderly) to evaluate the rate of revision arthroplasty, implant survival, the indications for revision, the complications, and the patient-reported outcomes.

Patients and Methods

We retrospectively reviewed all patients who underwent primary THA between January 2000 and May 2015 from our institutional database. A total of 145 very young and 1359 elderly patients were reviewed. The mean follow-up was 5.3 years (1 to 18). Logistic generalized estimating equations were used to compare characteristics and the revision rate. Survival was evaluated using Kaplan–Meier curves and hazard rates were created using Cox regression.


The Bone & Joint Journal
Vol. 97-B, Issue 8 | Pages 1038 - 1045
1 Aug 2015
Solomon LB Costi K Kosuge D Cordier T McGee MA Howie DW

The outcome of 219 revision total hip arthroplasties (THAs) in 98 male and 121 female patients, using 137 long length and 82 standard length cemented collarless double-taper femoral stems in 211 patients, with a mean age of 72 years (30 to 90) and mean follow-up of six years (two to 18) have been described previously. We have extended the follow-up to a mean of 13 years (8 to 20) in this cohort of patients in which the pre-operative bone deficiency Paprosky grading was IIIA or worse in 79% and 73% of femurs with long and standard stems, respectively.

For the long stem revision group, survival to re-revision for aseptic loosening at 14 years was 97% (95% confidence interval (CI) 91 to 100) and in patients aged > 70 years, survival was 100%. Two patients (two revisions) were lost to follow-up and 86 patients with 88 revisions had died. Worst-case analysis for survival to re-revision for aseptic loosening at 14 years was 95% (95% CI 89 to 100) and 99% (95% CI 96 to 100) for patients aged > 70 years. One additional long stem was classified as loose radiographically but not revised.

For the standard stem revision group, survival to re-revision for aseptic loosening at 14 years was 91% (95% CI 83 to 99). No patients were lost to follow-up and 49 patients with 51 hips had died. No additional stems were classified as loose radiographically.

Femoral revision using a cemented collarless double-taper stem, particularly with a long length stem, and in patients aged > 70 years, continues to yield excellent results up to 20 years post-operatively, including in hips with considerable femoral metaphyseal bone loss.

Cite this article: Bone Joint J 2015;97-B:1038–45.


The Bone & Joint Journal
Vol. 98-B, Issue 6 | Pages 754 - 760
1 Jun 2016
Malek IA Royce G Bhatti SU Whittaker JP Phillips SP Wilson IRB Wootton JR Starks I

Aims

We assessed the difference in hospital based and early clinical outcomes between the direct anterior approach and the posterior approach in patients who undergo total hip arthroplasty (THA).

Patients and Methods

The outcome was assessed in 448 (203 males, 245 females) consecutive patients undergoing unilateral primary THA after the implementation of an ‘Enhanced Recovery’ pathway. In all, 265 patients (mean age: 71 years (49 to 89); 117 males and 148 females) had surgery using the direct anterior approach (DAA) and 183 patients (mean age: 70 years (26 to 100); 86 males and 97 females) using a posterior approach. The groups were compared for age, gender, American Society of Anesthesiologists grade, body mass index, the side of the operation, pre-operative Oxford Hip Score (OHS) and attendance at ‘Joint school’. Mean follow-up was 18.1 months (one to 50).


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1304 - 1312
1 Oct 2017
Langton DJ Sidaginamale RP Joyce TJ Meek RD Bowsher JG Deehan D Nargol AVF Holland JP

Aims

We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system.

Patients and Methods

We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy.


The Bone & Joint Journal
Vol. 98-B, Issue 3 | Pages 301 - 306
1 Mar 2016
Reidy MJ Faulkner A Shitole B Clift B

Aims

The long-term functional outcome of total hip arthroplasty (THA) performed by trainees is not known. A multicentre retrospective study of 879 THAs was undertaken to investigate any differences in outcome between those performed by trainee surgeons and consultants.

Patients and Methods

A total of 879 patients with a mean age of 69.5 years (37 to 94) were included in the study; 584 THAs (66.4%) were undertaken by consultants, 138 (15.7%) by junior trainees and 148 (16.8%) by senior trainees. Patients were scored using the Harris Hip Score (HHS) pre-operatively and at one, three, five, seven and ten years post-operatively. Surgical outcome, complications and survival were compared between groups. The effect of supervision was determined by comparing supervised and unsupervised trainees. A primary univariate analysis was used to select variables for inclusion in multivariate analysis.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 2 | Pages 158 - 163
1 Feb 2011
Baker RP Pollard TCB Eastaugh-Waring SJ Bannister GC

We compared the medium-term clinical and radiological results of hybrid total hip replacement (THR) with metal-on-metal Birmingham hip resurfacing (BHR) in two groups of 54 young patients matched for age, gender, body mass index and pre-operative levels of activity.

The clinical outcome was assessed by the University of California, Los Angeles (UCLA) activity score, the Oxford Hip Score (OHS) and the EuroQol scores. Radiologically, all hips were assessed for migration and osteolysis, the hybrid THRs for polyethylene wear and the BHRs for a pedestal sign. The mean follow-up of the patients with a hybrid THR was ten years and for those with a BHR, nine years. Four patients with a hybrid THR and one with a BHR had died. In each group five were lost to follow-up. The revision rate of the hybrid THRs was 16.7% (9 of 54) and of the BHRs 9.3% (5 of 54) (p = 0.195). Radiographs of a further eight hybrid THRs demonstrated wear and osteolysis, and they await revision (p = 0.008). Of the unrevised BHRs 90% had radiological changes, of which approximately 50% had progressed over the previous four years. All hybrid THRs demonstrated linear polyethylene wear with a mean of 1.24 mm (0.06 to 3.03). The BHRs recorded superior OHS (p = 0.013), UCLA (p = 0.008), and EuroQol visual analogue scores (p = 0.009).

After nine years, patients with BHRs remained more active and had a lower rate of revision than those with hybrid THRs. Both groups demonstrated progressive radiological changes at medium-term follow-up.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 10 | Pages 1299 - 1302
1 Oct 2007
Ramiah RD Ashmore AM Whitley E Bannister GC

We determined the ten-year life expectancy of 5831 patients who had undergone 6653 elective primary total hip replacements at a regional orthopaedic centre between April 1993 and October 2004. Using hospital, general practitioner and the local health authority records, we recorded the dates of death for those who died following surgery.

The mean age at operation was 67 years (13 to 96) with a male:female ratio of 2:3. Of 1154 patients with a ten-year follow-up 340 (29.5%) had died a mean of 5.6 years (0 to 10) after surgery. Using Kaplan-Meier curves, the ten-year survival was 89% in patients under 65 years at surgery, 75% in patients aged between 65 and 74 years, and 51% in patients over 75.

The standardised mortality rates were considerably higher for patients under 45 years, 20% higher for those between 45 and 64 years, and steadily reduced in patients aged 65 and over.

The survival of cemented hip replacement derived from the Swedish Hip Arthroplasty Register Annual Report 2004 exceeds the life expectancy of patients over the age of 60 in our area, suggesting that cemented hip replacement is the procedure of choice in this population.


The Bone & Joint Journal
Vol. 97-B, Issue 4 | Pages 463 - 472
1 Apr 2015
Panagiotidou A Meswania J Osman K Bolland B Latham J Skinner J Haddad FS Hart A Blunn G

The aim of this study was to assess the effect of frictional torque and bending moment on fretting corrosion at the taper interface of a modular femoral component and to investigate whether different combinations of material also had an effect. The combinations we examined were 1) cobalt–chromium (CoCr) heads on CoCr stems 2) CoCr heads on titanium alloy (Ti) stems and 3) ceramic heads on CoCr stems.

In test 1 increasing torque was imposed by offsetting the stem in the anteroposterior plane in increments of 0 mm, 4 mm, 6 mm and 8 mm when the torque generated was equivalent to 0 Nm, 9 Nm, 14 Nm and 18 Nm.

In test 2 we investigated the effect of increasing the bending moment by offsetting the application of axial load from the midline in the mediolateral plane. Increments of offset equivalent to head + 0 mm, head + 7 mm and head + 14 mm were used.

Significantly higher currents and amplitudes were seen with increasing torque for all combinations of material. However, Ti stems showed the highest corrosion currents. Increased bending moments associated with using larger offset heads produced more corrosion: Ti stems generally performed worse than CoCr stems. Using ceramic heads did not prevent corrosion, but reduced it significantly in all loading configurations.

Cite this article: Bone Joint J 2015;97-B:463–72.


The Bone & Joint Journal
Vol. 96-B, Issue 5 | Pages 604 - 608
1 May 2014
von Schewelov T Carlsson Å Sanzén L Besjakov J

In 2005, we demonstrated that the polished triple-tapered C-stem at two years had migrated distally and rotated internally. From that series, 33 patients have now been followed radiologically, clinically and by radiostereometric analysis (RSA) for up to ten years. The distal migration within the cement mantle had continued and reached a mean of 2 mm (0.5 to 4.0) at ten years. Internal rotation, also within the cement mantle, was a mean 3.8° (external 1.6° to internal 6.6°) The cement mantle did not show any sign of migration or loosening in relation to the femoral bone. There were no clinical or radiological signs indicating that the migration or rotation within the cement mantle had had any adverse effects for the patients.

Cite this article: Bone Joint J 2014;96-B:604–8.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 2 | Pages 155 - 157
1 Feb 2005
Konyves A Bannister GC

We assessed leg length discrepancy and hip function in 90 patients undergoing primary total hip arthroplasty before surgery and at three and 12 months after. Function was measured using the Oxford hip score (OHS). After surgery the mean OHS improved by 26 points after three months and by 30 points after 12.

After operation 56 (62%) limbs were long by a mean of 9 mm and this was perceived by 24 (43%) patients after three months and by 18 (33%) after 12. The mean OHS in patients who perceived true lengthening was 27% worse than the rest of the population after three months and 18% worse after 12. In 55 (98%) patients, lengthening occurred in the femoral component. Appropriate placement of the femoral component could significantly reduce a patient’s perception of discrepancy of length.


The Bone & Joint Journal
Vol. 95-B, Issue 12 | Pages 1610 - 1616
1 Dec 2013
Epinette J Asencio G Essig J Llagonne B Nourissat C

We report a multicentre prospective consecutive study assessing the long-term outcome of the proximally hydroxyapatite (HA)-coated ABG II monobloc femoral component in a series of 1148 hips in 1053 patients with a mean age at surgery of 64.77 years (22 to 80) at a mean follow-up of 10.84 years (10 to 15.25). At latest follow-up, the mean total Harris hip score was 94.7 points (sd; 6.87; 49 to 100), and the mean Merle d’Aubigné–Postel score was 17.6 points (sd 1.12; 7 to 18). The mean total Engh radiological score score was 21.54 (sd 5.77; 3.5 to 27), with 95.81% of ‘confirmed ingrowth’, according to Engh’s classification. With aseptic loosening or pain as endpoints, three AGB II stems (0.26%) failed, giving a 99.7% survival rate (se 0.002; 95% confidence interval (CI) 0.994 to 1) at 14 years’ follow-up. The survival of patients ≤ 50 years of age (99.0% (se 0.011; 95% CI 0.969 to 1)) did not differ significantly from those of patients aged > 50 years (99.8% (se 0.002; 95% CI 0.994 to 1)). This study confirmed the excellent long-term results currently achieved with the ABG II proximally HA-coated monobloc stem.

Cite this article: Bone Joint J 2013;95-B:1610–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 3 | Pages 322 - 327
1 Mar 2012
Morley JR Blake SM Hubble MJW Timperley AJ Gie GA Howell JR

The removal of all prosthetic material and a two-stage revision procedure is the established standard management of an infected total hip replacement (THR). However, the removal of well-fixed femoral cement is time-consuming and can result in significant loss of bone stock and femoral shaft perforation or fracture. We report our results of two-stage revision THR for treating infection, with retention of the original well-fixed femoral cement mantle in 15 patients, who were treated between 1989 and 2002. Following partial excision arthroplasty, patients received local and systemic antibiotics and underwent reconstruction and re-implantation at a second-stage procedure, when the infection had resolved.

The mean follow-up of these 15 patients was 82 months (60 to 192). Two patients had positive microbiology at the second stage and were treated with six weeks of appropriate antibiotics; one of these developed recurrent infection requiring further revision. Successful eradication of infection was achieved in the remaining 14 patients.

We conclude that when two-stage revision is used for the treatment of peri-prosthetic infection involving a THR, a well-fixed femoral cement mantle can be safely left in situ, without compromising the treatment of infection. Advantages of this technique include a shorter operating time, reduced loss of bone stock and a technically more straightforward second-stage procedure.


The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 608 - 615
1 May 2011
Bolland BJRF Culliford DJ Langton DJ Millington JPS Arden NK Latham JM

This study reports the mid-term results of a large-bearing hybrid metal-on-metal total hip replacement in 199 hips (185 patients) with a mean follow-up of 62 months (32 to 83).

Two patients died of unrelated causes and 13 were lost to follow-up. In all, 17 hips (8.5%) have undergone revision, and a further 14 are awaiting surgery. All revisions were symptomatic. Of the revision cases, 14 hips showed evidence of adverse reactions to metal debris. The patients revised or awaiting revision had significantly higher whole blood cobalt ion levels (p = 0.001), but no significant difference in acetabular component size or position compared with the unrevised patients. Wear analysis (n = 5) showed increased wear at the trunnion-head interface, normal levels of wear at the articulating surfaces and evidence of corrosion on the surface of the stem.

The cumulative survival rate, with revision for any reason, was 92.4% (95% confidence interval 87.4 to 95.4) at five years. Including those awaiting surgery, the revision rate would be 15.1% with a cumulative survival at five years of 89.6% (95% confidence interval 83.9 to 93.4).

This hybrid metal-on-metal total hip replacement series has shown an unacceptably high rate of failure, with evidence of high wear at the trunnion-head interface and passive corrosion of the stem surface. This raises concerns about the use of large heads on conventional 12/14 tapers.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 595 - 600
1 May 2009
Leonardsson O Rogmark C Kärrholm J åkesson K Garellick G

Between 1999 and 2005, 10 264 patients who had undergone total hip replacement (THR) for subcapital fracture of the hip were compared with 76 520 in whom THR had been performed for other reasons. All the cases were identified through the Swedish Hip Arthroplasty Register. The THRs performed as primary treatment for fracture were also compared with those done after failure of internal fixation.

After seven years the rate of revision was higher in THR after fracture (4.4% vs 2.9%). Dislocation and periprosthetic fracture were the most common causes of revision. The risk was higher in men than in women. The type of femoral component and the surgical approach influenced the risk. After correction for gender, type of component and the surgical approach the revision rates were similar in the primary and secondary fracture THR groups.

Total hip replacement is therefore a safe method for both the primary and secondary management of fracture of the hip.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 5 | Pages 583 - 588
1 May 2009
Rasch A Byström AH Dalén N Martinez-Carranza N Berg HE

Muscle atrophy has been demonstrated in patients suffering from osteoarthritis of the hip, but little is known about muscular recovery after total hip replacement (THR). A total of 20 patients with unilateral osteoarthritis of the hip were assessed before, six months and two years after THR. The cross-sectional area and radiological density of the muscles of the hip, thigh, calf and back were measured using CT. We hypothesised that the muscles would not recover fully after operation.

After two years comparison of the limb with the THR with the healthy limb showed that there was such a reduction in the cross-sectional area in iliopsoas (7.0%; p = 0.006) and the hip adductors (8.4%, p = 0.003) and in the radiological density in gluteus maximus (10.1 Hounsfield units; p < 0.001), gluteus medius/minimus (5.6 Hounsfield units; p = 0.011), iliopsoas (3.9 Hounsfield units; p < 0.001) and the adductors (2.4 Hounsfield units; p = 0.022).

Thus, there was persistent muscle atrophy in muscles acting about the hip two years after THR. We suggest that an earlier operation or a more intensive rehabilitation may reverse these changes.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 5 | Pages 556 - 561
1 May 2008
Glyn-Jones S McLardy-Smith P Gill HS Murray DW

The creep and wear behaviour of highly cross-linked polyethylene and standard polyethylene liners were examined in a prospective, double-blind randomised, controlled trial using radiostereometric analysis.

We randomised 54 patients to receive hip replacements with either highly cross-linked polyethylene or standard liners and determined the three-dimensional penetration of the liners over three years.

After three years the mean total penetration was 0.35 mm (SD 0.14) for the highly cross-linked polyethylene group and 0.45 mm (SD 0.19) for the standard group. The difference was statistically significant (p = 0.0184). From the pattern of penetration it was possible to discriminate creep from wear. Most (95%) of the creep occurred within six months of implantation and nearly all within the first year. There was no difference in the mean degree of creep between the two types of polyethylene (highly cross-linked polyethylene 0.26 mm, SD 0.17; standard 0.27 mm, SD 0.2; p = 0.83). There was, however, a significant difference (p = 0.012) in the mean wear rate (highly cross-linked polyethylene 0.03 mm/yr, SD 0.06; standard 0.07 mm/yr, SD 0.05). Creep and wear occurred in significantly different directions (p = 0.01); creep was predominantly proximal whereas wear was anterior, proximal and medial.

We conclude that penetration in the first six months is creep-dominated, but after one year virtually all penetration is due to wear. Highly cross-linked polyethylene has a 60% lower rate of wear than standard polyethylene and therefore will probably perform better in the long term.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 7 | Pages 879 - 886
1 Jul 2007
Howie DW Wimhurst JA McGee MA Carbone TA Badaruddin BS

We investigated 219 revisions of total hip replacement (THR) in 211 patients using a collarless double-taper cemented femoral component. The mean age of the patients was 72 years (30 to 90). The 137 long and 82 standard length stems were analysed separately. The mean follow-up was six years (2 to 18), and no patient was lost to follow-up.

Survival of the long stems to re-revision for aseptic loosening at nine years was 98% (95% confidence interval (CI) 94 to 100), and for the standard stems was 93% (95% CI 85 to 100). At five years, one long stem was definitely loose radiologically and one standard stem was probably loose. Pre-operative femoral bone deficiency did not influence the results for the long stems, and corrective femoral osteotomy was avoided, as were significant subsidence, major stress shielding and persistent thigh pain.

Because of these reliable results, cemented long collarless double-taper femoral components are recommended for routine revision THR in older patients.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 3 | Pages 298 - 303
1 Mar 2006
Bhan S Pankaj A Malhotra R

We compared the safety and outcome of one-stage bilateral total hip arthroplasty with those of a two-stage procedure during different admissions in a prospective, randomised controlled trial in an Asian population. Of 168 patients included in the study, 83 had a single- and 85 a two-stage procedure. Most of the patients (59.9%) suffered from inflammatory arthritis.

The intra-operative complications, early systemic complications, the operating time, positioning of the components, the functional score, restoration of limb length and survival rates at 96 months were similar in the two groups. The total estimated blood loss was significantly lower in patients undergoing a one-stage procedure than in patients who had a two-stage procedure, but the transfusion requirements were significantly higher in the former group (p = 0.001). The hospital stay was significantly shorter in the one-stage group, 7.25 days (sd 1.30; 5 to 20) compared with 10 days (sd 1.65; 8 to 24) in the two-stage group (p = 0.023). We believe that a one-stage procedure is safe and appropriate in our population.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 5 | Pages 592 - 600
1 May 2006
Pollard TCB Baker RP Eastaugh-Waring SJ Bannister GC

We compared the five- to seven-year clinical and radiological results of the metal-on-metal Birmingham hip resurfacing with a hybrid total hip arthroplasty in two groups of 54 hips, matched for gender, age, body mass index and activity level.

Function was excellent in both groups, as measured by the Oxford hip score, but the Birmingham hip resurfacings had higher University of California at Los Angeles activity scores and better EuroQol quality of life scores. The total hip arthroplasties had a revision or intention-to-revise rate of 8%, and the Birmingham hip resurfacings of 6%. Both groups demonstrated impending failure on surrogate end-points. Of the total hip arthroplasties, 12% had polyethylene wear and osteolysis under observation, and 8% of Birmingham hip resurfacings showed migration of the femoral component. Polyethylene wear was present in 48% of the hybrid hips without osteolysis. Of the femoral components in the Birmingham hip resurfacing group which had not migrated, 66% had radiological changes of unknown significance.