Aims. Unicompartmental knee arthroplasty (UKA) is the preferred treatment for anterior medial knee osteoarthritis (OA) owing to the rapid postoperative recovery. However, the risk factors for UKA failure remain controversial. Methods. The clinical data of Oxford mobile-bearing UKAs performed between 2011 and 2017 with a minimum follow-up of five years were retrospectively analyzed. Demographic, surgical, and follow-up data were collected. The Cox proportional hazards model was used to identify the risk factors that contribute to UKA failure. Kaplan-Meier survival was used to compare the effect of the prosthesis position on UKA survival. Results. A total of 407 patients who underwent UKA were included in the study. The mean age of patients was 61.8 years, and the mean follow-up period of the patients was 91.7 months. The mean Knee Society Score (KSS) preoperatively and at the last follow-up were 64.2 and 89.7, respectively (p = 0.001). Overall, 28 patients (6.9%) with UKA underwent revision due to prosthesis loosening (16 patients), dislocation (eight patients), and persistent pain (four patients). Cox proportional hazards model analysis identified
Aims. This study aimed to determine outcomes of isolated tibial insert exchange (ITIE) during revision total knee arthroplasty (TKA). Methods. From 1985 to 2016, 270 ITIEs were performed at one institution for instability (55%, n = 148), polyethylene wear (39%, n = 105), insert fracture/dissociation (5%, n = 14), or stiffness (1%, n = 3). Patients with component loosening, implant
Approximately 10% to 20% of knee arthroplasty patients are not satisfied with the result, while a clear indication for revision surgery might not be present. Therapeutic options for these patients, who often lack adequate quadriceps strength, are limited. Therefore, the primary aim of this study was to evaluate the clinical effect of a novel rehabilitation protocol that combines low-load resistance training (LL-RT) with blood flow restriction (BFR). Between May 2022 and March 2024, we enrolled 45 dissatisfied knee arthroplasty patients who lacked any clear indication for revision to this prospective cohort study. All patients were at least six months post-surgery and had undergone conventional physiotherapy previously. The patients participated in a supervised LL-RT combined with BFR in 18 sessions. Primary assessments included the following patient-reported outcome measures (PROMs): Knee injury and Osteoarthritis Outcome Score (KOOS); Knee Society Score: satisfaction (KSSs); the EuroQol five-dimension five-level questionnaire (EQ-5D-5L); and the pain catastrophizing scale (PCS). Functionality was assessed using the six-minute walk Test (6MWT) and the 30-second chair stand test (30CST). Follow-up timepoints were at baseline, six weeks, three months, and six months after the start.Aims
Methods
Tibial tubercle osteotomy (TTO) facilitates surgical exposure and protects the extensor mechanism during revision total knee arthroplasty (rTKA). The purpose of this study was to determine the rates of bony union, complications, and reoperations following TTO during rTKA, to assess the functional outcomes of rTKA with TTO at two years’ minimum follow-up, and to identify the risk factors of failure. Between January 2010 and September 2020, 695 rTKAs were performed and data were entered into a prospective database. Inclusion criteria were rTKAs with concomitant TTO, without extensor mechanism allograft, and a minimum of two years’ follow-up. A total of 135 rTKAs were included, with a mean age of 65 years (SD 9.0) and a mean BMI of 29.8 kg/m2 (SD 5.7). The most frequent indications for revision were infection (50%; 68/135), aseptic loosening (25%; 34/135), and stiffness (13%; 18/135). Patients had standardized follow-up at six weeks, three months, six months, and annually thereafter. Complications and revisions were evaluated at the last follow-up. Functional outcomes were assessed using the Knee Society Score (KSS) and range of motion.Aims
Methods
Tibial fixation in revision total knee arthroplasty (rTKA) can present surgical challenges. It has been suggested that appropriate fixation in at least two of the three anatomical zones (epiphysis, metaphysis, and diaphysis) is essential for implant survival. However, supporting clinical data are lacking. In this retrospective case-control study, we investigated the relationship between zonal fixation of hybrid rTKA tibial components and re-revision total knee arthroplasty for aseptic loosening (rrTKA-AL). All consecutive rTKAs with hybrid tibial components (May 2006 to December 2020) were screened for subsequent rrTKA-AL. A control group was randomly selected from the remaining cohort. Postoperative radiographs of rTKAs were scored in random order by three blinded observers for zonal fixation in the epiphysis (bone resection level below, at, or above fibular head; 0 to 2), metaphysis (number of sufficiently cemented zones; 0 to 4), and diaphysis (canal filling ratio (CFR); %). The intraclass correlation coefficient (ICC) was calculated to quantify the agreement between observers. Multivariate logistic regression analysis was performed to assess the relationship between zonal fixation and rrTKA-AL.Aims
Methods
The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).Aims
Methods
Patient dissatisfaction following primary total knee arthroplasty (TKA) with manual jig-based instruments has been reported to be as high as 30%. Robotic-assisted total knee arthroplasty (RA-TKA) has been increasingly used in an effort to improve patient outcomes, however there is a paucity of literature examining patient satisfaction after RA-TKA. This study aims to identify the incidence of patients who were not satisfied following RA-TKA and to determine factors associated with higher levels of dissatisfaction. This was a retrospective review of 674 patients who underwent primary TKA between October 2016 and September 2020 with a minimum two-year follow-up. A five-point Likert satisfaction score was used to place patients into two groups: Group A were those who were very dissatisfied, dissatisfied, or neutral (Likert score 1 to 3) and Group B were those who were satisfied or very satisfied (Likert score 4 to 5). Patient demographic data, as well as preoperative and postoperative patient-reported outcome measures, were compared between groups.Aims
Methods
In-hospital length of stay (LOS) and discharge dispositions following arthroplasty could act as surrogate measures for improvement in patient pathways, and have major cost saving implications for healthcare providers. With the ever-growing adoption of robotic technology in arthroplasty, it is imperative to evaluate its impact on LOS. The objectives of this study were to compare LOS and discharge dispositions following robotic arm-assisted total knee arthroplasty (RO TKA) and unicompartmental arthroplasty (RO UKA) versus conventional technique (CO TKA and UKA). This large-scale, single-institution study included patients of any age undergoing primary TKA (n = 1,375) or UKA (n = 337) for any cause between May 2019 and January 2023. Data extracted included patient demographics, LOS, need for post anaesthesia care unit (PACU) admission, anaesthesia type, readmission within 30 days, and discharge dispositions. Univariate and multivariate logistic regression models were also employed to identify factors and patient characteristics related to delayed discharge.Aims
Methods
To map literature on prognostic factors related to outcomes of revision total knee arthroplasty (rTKA), to identify extensively studied factors and to guide future research into what domains need further exploration. We performed a systematic literature search in MEDLINE, Embase, and Web of Science. The search string included multiple synonyms of the following keywords: "revision TKA", "outcome" and "prognostic factor". We searched for studies assessing the association between at least one prognostic factor and at least one outcome measure after rTKA surgery. Data on sample size, study design, prognostic factors, outcomes, and the direction of the association was extracted and included in an evidence map.Aims
Methods
Accurate identification of the ankle joint centre is critical for estimating tibial coronal alignment in total knee arthroplasty (TKA). The purpose of the current study was to leverage artificial intelligence (AI) to determine the accuracy and effect of using different radiological anatomical landmarks to quantify mechanical alignment in relation to a traditionally defined radiological ankle centre. Patients with full-limb radiographs from the Osteoarthritis Initiative were included. A sub-cohort of 250 radiographs were annotated for landmarks relevant to knee alignment and used to train a deep learning (U-Net) workflow for angle calculation on the entire database. The radiological ankle centre was defined as the midpoint of the superior talus edge/tibial plafond. Knee alignment (hip-knee-ankle angle) was compared against 1) midpoint of the most prominent malleoli points, 2) midpoint of the soft-tissue overlying malleoli, and 3) midpoint of the soft-tissue sulcus above the malleoli.Aims
Methods
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
Uncemented mobile bearing designs in medial unicompartmental knee arthroplasty (UKA) have seen an increase over the last decade. However, there are a lack of large-scale studies comparing survivorship of these specific designs to commonly used cemented mobile and fixed bearing designs. The aim of this study was to evaluate the survivorship of these designs. A total of 21,610 medial UKAs from 2007 to 2018 were selected from the Dutch Arthroplasty Register. Multivariate Cox regression analyses were used to compare uncemented mobile bearings with cemented mobile and fixed bearings. Adjustments were made for patient and surgical factors, with their interactions being considered. Reasons and type of revision in the first two years after surgery were assessed.Aims
Methods
Tibial cones are often utilized in revision total knee arthroplasty (TKA) with metaphyseal defects. Because there are few studies evaluating mid-term outcomes with a sufficient cohort, the purpose of this study was to evaluate tibial cone survival and complications in revision TKAs with tibial cones at minimum follow-up of five years. A retrospective review was completed from September 2006 to March 2015, evaluating 67 revision TKAs (64 patients) that received one specific porous tibial cone during revision TKA. The final cohort was composed of 62 knees (59 patients) with five years of clinical follow-up or reoperation. The mean clinical follow-up of the TKAs with minimum five-year clinical follow-up was 7.6 years (5.0 to 13.3). Survivorship analysis was performed with the endpoints of tibial cone revision for aseptic loosening, tibial cone revision for any reason, and reoperation. We also evaluated periprosthetic joint infection (PJI), risk factors for failure, and performed a radiological review.Aims
Methods
Malalignment of the tibial component could influence the long-term survival of a total knee arthroplasty (TKA). The object of this study was to investigate the biomechanical effect of varus and valgus malalignment on the tibial component under stance-phase gait cycle loading conditions. Validated finite element models for varus and valgus malalignment by 3° and 5° were developed to evaluate the effect of malalignment on the tibial component in TKA. Maximum contact stress and contact area on a polyethylene insert, maximum contact stress on patellar button and the collateral ligament force were investigated.Objectives
Methods
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments. Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.Objectives
Materials and Methods
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
Although it has been suggested that the outcome
after revision of a unicondylar knee replacement (UKR) to total knee
replacement (TKR) is better when the mechanism of failure is understood,
a comparative study on this subject has not been undertaken. A total of 30 patients (30 knees) who underwent revision of their
unsatisfactory UKR to TKR were included in the study: 15 patients
with unexplained pain comprised group A and 15 patients with a defined
cause for pain formed group B. The Oxford knee score (OKS), visual
analogue scale for pain (VAS) and patient satisfaction were assessed before
revision and at one year after revision, and compared between the
groups. The mean OKS improved from 19 (10 to 30) to 25 (11 to 41) in
group A and from 23 (11 to 45) to 38 (20 to 48) in group B. The
mean VAS improved from 7.7 (5 to 10) to 5.4 (1 to 8) in group A
and from 7.4 (2 to 9) to 1.7 (0 to 8) in group B. There was a statistically
significant difference between the mean improvements in each group
for both OKS (p = 0.022) and VAS (p = 0.002). Subgroup analysis
in group A, performed in order to define a patient factor that predicts
outcome of revision surgery in patients with unexplained pain, showed
no pre-operative differences between both subgroups. These results may be used to inform patients about what to expect
from revision surgery, highlighting that revision of UKR to TKR
for unexplained pain generally results in a less favourable outcome
than revision for a known cause of pain. Cite this article:
We conducted a randomised controlled trial to assess the accuracy
of positioning and alignment of the components in total knee arthroplasty
(TKA), comparing those undertaken using standard intramedullary
cutting jigs and those with patient-specific instruments (PSI). There were 64 TKAs in the standard group and 69 in the PSI group. The post-operative hip-knee-ankle (HKA) angle and positioning
was investigated using CT scans. Deviation of >
3° from the planned
position was regarded as an outlier. The operating time, Oxford
Knee Scores (OKS) and Short Form-12 (SF-12) scores were recorded.Aims
Patients and Methods
We explored the literature surrounding whether
allergy and hypersensitivity has a clinical basis for implant selection
in total knee arthroplasty (TKA). In error, the terms hypersensitivity
and allergy are often used synonymously. Although a relationship
is present, we could not find any evidence of implant failure due
to allergy. There is however increasing basic science that suggests
a link between loosening and metal ion production. This is not an
allergic response but is a potential problem. With a lack of evidence
logically there can be no justification to use ‘hypoallergenic’
implants in patients who have pre-existing skin sensitivity to the
metals used in TKA. Cite this article:
The aim of this study was to examine the results of revision
total knee arthroplasty (TKA) undertaken for stiffness in the absence
of sepsis or loosening. We present the results of revision surgery for stiff TKA in 48
cases (35 (72.9%) women and 13 (27.1%) men). The mean age at revision
surgery was 65.5 years (42 to 83). All surgeries were performed
by a single surgeon. Stiffness was defined as an arc of flexion
of <
70° or a flexion contracture of >
15°. The changes in the
range of movement (ROM) and the Western Ontario and McMasters Osteoarthritis
index scores (WOMAC) were recorded.Aims
Patients and Methods
Patellofemoral complications are common after
total knee replacement (TKR). Leaving the patellar unsurfaced after
TKR may lead to complications such as anterior knee pain, and re-operation
to surface it. Complications after patellar resurfacing include
patellar fracture, aseptic loosening, patellar instability, polyethylene
wear, patellar clunk and osteonecrosis. Historically, patellar complications
account for one of the larger proportions of causes of failure in
TKR, however, with contemporary implant designs, complication rates
have decreased. Most remaining failures relate to patellofemoral
tracking. Understanding the causes of patellofemoral maltracking
is essential to prevent these complications as well as manage them
when they occur. Cite this article:
Previous studies of failure mechanisms leading
to revision total knee replacement (TKR) performed between 1986 and
2000 determined that many failed early, with a disproportionate
amount accounted for by infection and implant-associated factors
including wear, loosening and instability. Since then, efforts have
been made to improve implant performance and instruct surgeons in
best practice. Recently our centre participated in a multi-centre evaluation
of 844 revision TKRs from 2010 to 2011. The purpose was to report
a detailed analysis of failure mechanisms over time and to see if
failure modes have changed over the past 10 to 15 years. Aseptic
loosening was the predominant mechanism of failure (31.2%), followed
by instability (18.7%), infection (16.2%), polyethylene wear (10.0%),
arthrofibrosis (6.9%) and malalignment (6.6%). The mean time to
failure was 5.9 years (ten days to 31 years), 35.3% of all revisions
occurred at less than two years, and 60.2% in the first five years.
With improvements in implant and polyethylene manufacture, polyethylene
wear is no longer a leading cause of failure. Early mechanisms of
failure are primarily technical errors. In addition to improving
implant longevity, industry and surgeons must work together to decrease
these technical errors. All reports on failure of TKR contain patients
with unexplained pain who not infrequently have unmet expectations.
Surgeons must work to achieve realistic patient expectations pre-operatively,
and therefore, improve patient satisfaction post-operatively. Cite this article:
The optimal timing of total knee replacement
(TKR) in patients with osteoarthritis, in relation to the severity
of disease, remains controversial. This prospective study was performed
to investigate the effect of the severity of osteoarthritis and
other commonly available pre- and post-operative clinical parameters
on the clinical outcome in a consecutive series of cemented TKRs.
A total of 176 patients who underwent unilateral TKR were included
in the study. Their mean age was 68 years (39 to 91), 63 (36%) were
male and 131 knees (74%) were classified as grade 4 on the Kellgren–Lawrence
osteoarthritis scale. A total of 154 patients (87.5%) returned for
clinical review 12 months post-operatively, at which time the outcome
was assessed using the Knee Society score. A low radiological severity of osteoarthritis was not associated
with pain 12 months post-operatively. However, it was significantly
associated with an inferior level of function (p = 0.007), implying
the need for increased focus on all possible reasons for pain in
the knee and the forms of conservative treatment which are available
for patients with lower radiological severity of osteoarthritis. Cite this article:
We prospectively assessed the benefits of using either a range-of-movement technique or an anatomical landmark method to determine the rotational alignment of the tibial component during total knee replacement. We analysed the cut proximal tibia intraoperatively, determining anteroposterior axes by the range-of-movement technique and comparing them with the anatomical anteroposterior axis. We found that the range-of-movement technique tended to leave the tibial component more internally rotated than when anatomical landmarks were used. In addition, it gave widely variable results (mean 7.5°; 2° to 17°), determined to some extent by which posterior reference point was used. Because of the wide variability and the possibilities for error, we consider that it is inappropriate to use the range-of-movement technique as the sole method of determining alignment of the tibial component during total knee replacement.
Patients with skeletal dysplasia are prone to
developing advanced osteoarthritis of the knee requiring total knee replacement
(TKR) at a younger age than the general population. TKR in this
unique group of patients is a technically demanding procedure owing
to the deformity, flexion contracture, generalised hypotonia and ligamentous
laxity. We retrospectively reviewed the outcome of 11 TKRs performed
in eight patients with skeletal dysplasia at our institution using
the Stanmore Modular Individualised Lower Extremity System (SMILES)
custom-made rotating-hinge TKR. There were three men and five women
with mean age of 57 years (41 to 79). Patients were followed clinically
and radiologically for a mean of seven years (3 to 11.5). The mean
Knee Society clinical and function scores improved from 24 (14 to
36) and 20 (5 to 40) pre-operatively, respectively, to 68 (28 to
80) and 50 (22 to 74), respectively, at final follow-up. Four complications
were recorded, including a patellar fracture following a fall, a
tibial peri-prosthetic fracture, persistent anterior knee pain,
and aseptic loosening of a femoral component requiring revision.
Our results demonstrate that custom primary rotating-hinge TKR in
patients with skeletal dysplasia is effective at relieving pain,
with a satisfactory range of movement and improved function. It compensates
for bony deformity and ligament deficiency and reduces the likelihood
of corrective osteotomy. Patellofemoral joint complications are
frequent and functional outcome is worse than with primary TKR in
the general population.
The purpose of this study was to report the outcome
of ‘isolated’ anterior cruciate ligament (ACL) ruptures treated with
anatomical endoscopic reconstruction using hamstring tendon autograft
at a mean of 15 years (14.25 to 16.9). A total of 100 consecutive
men and 100 consecutive women with ‘isolated’ ACL rupture underwent
four-strand hamstring tendon reconstruction with anteromedial portal
femoral tunnel drilling and interference screw fixation by a single
surgeon. Details were recorded pre-operatively and at one, two,
seven and 15 years post-operatively. Outcomes included clinical
examination, subjective and objective scoring systems, and radiological
assessment. At 15 years only eight of 118 patients (7%) had moderate
or severe osteo-arthritic changes (International Knee Documentation
Committee Grades C and D), and 79 of 152 patients (52%) still performed
very strenuous activities. Overall graft survival at 15 years was
83% (1.1% failure per year). Patients aged <
18 years at the
time of surgery and patients with >
2 mm of laxity at one year had
a threefold increase in the risk of suffering a rupture of the graft
(p = 0.002 and p = 0.001, respectively). There was no increase in
laxity of the graft over time. ACL reconstructive surgery in patients with an ‘isolated’ rupture
using this technique shows good results 15 years post-operatively
with respect to ligamentous stability, objective and subjective
outcomes, and does not appear to cause osteoarthritis.
We studied the intra- and interobserver reliability of measurements of the position of the components after total knee replacement (TKR) using a combination of radiographs and axial two-dimensional (2D) and three-dimensional (3D) reconstructed CT images to identify which method is best for this purpose. A total of 30 knees after primary TKR were assessed by two independent observers (an orthopaedic surgeon and a radiologist) using radiographs and CT scans. Plain radiographs were highly reliable at measuring the tibial slope, but showed wide variability for all other measurements; 2D-CT also showed wide variability. 3D-CT was highly reliable, even when measuring rotation of the femoral components, and significantly better than 2D-CT. Interobserver variability in the measurements on radiographs were good (intraclass correlation coefficient (ICC) 0.65 to 0.82), but rotational measurements on 2D-CT were poor (ICC 0.29). On 3D-CT they were near perfect (ICC 0.89 to 0.99), and significantly more reliable than 2D-CT (p <
0.001). 3D-reconstructed images are sufficiently reliable to enable reporting of the position and orientation of the components. Rotational measurements in particular should be performed on 3D-reconstructed CT images. When faced with a poorly functioning TKR with concerns over component positioning, we recommend 3D-CT as the investigation of choice.
This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1). The PFC Sigma knee system appears to provide excellent results in the medium term.
Mechanical failure because of wear or fracture of the polyethylene tibial post in posteriorly-stabilised total knee replacements has been extensively described. In this study of 12 patients with a clinically and radiologically successful NexGen LPS posteriorly-stabilised prosthesis impingement of the anterior tibial post was evaluated in vivo in three dimensions during gait using radiologically-based image-matching techniques. Impingement was observed in all images of the patients during the stance phase, although the NexGen LPS was designed to accommodate 14° of hyperextension of the component before impingement occurred. Impingement arises as a result of posterior translation of the femur during the stance phase. Further attention must therefore be given to the configuration of the anterior portion of the femoral component and the polyethylene post when designing posteriorly-stabilised total knee replacements.
We have examined the outcome of 400 consecutive patients who underwent total knee replacement with the Low Contact Stress mobile-bearing system between 1993 and 1994 and were followed up for a minimum of ten years. All operations were performed by surgeons in Christchurch, New Zealand, who used no other knee prosthesis during the study time. At ten years after operation 238 patients (244 knees) remained for independent clinical and radiological assessment. There was a significant improvement (p <
0.001) in the postoperative knee scores at one, three, seven and ten years, although a slight deterioration in the score occurred between seven and ten years which did not reach statistical significance. The survival for polyethylene wear or loosening was 97% (95% CI 96 to 98) and survival using reoperation for any cause was 92% (95% CI 90 to 94) at 12 years. Polyethylene wear was more common in the meniscal-bearing component, with five knees requiring revision and a further eight demonstrating early wear. Osteolysis was not seen in the rotating platform component, but was present in three of the meniscal-bearing knees. There was no association between the radiological alignment at one year and those knees that subsequently required revision. However, there was an association between the overall limb alignment and the Western Ontario McMasters University score (p <
0.001). The Low Contact Stress mobile-bearing total knee replacement has proved to be a reliable implant at ten years when used in primary knee replacement irrespective of the deformity and diagnosis.
We reviewed 87 patients who underwent revision reconstruction of the anterior cruciate ligament. The incidence of meniscal tears and degenerative change was assessed and related to the interval between failure of the primary graft and revision reconstruction. Patients were divided into two groups: early revision surgery within six months of graft failure, and delayed revision. Degenerative change was scored using the French Society of Arthroscopy system. There was a significantly higher incidence of articular cartilage degeneration in the delayed group (Mann-Whitney U-test, 53.2% We conclude that revision reconstruction should be carried out within six months of primary graft failure, in order to minimise the risk of degenerative change.
Stiffness is an uncommon but potentially debilitating complication following total knee replacement (TKR). The treatment of this condition remains difficult and controversial. We present the results of 13 patients who underwent open arthrolysis for stiffness. The mean time between TKR and arthrolysis was 14 months. The mean follow-up was 7.2 years (2 to 10). The mean range of movement prior to arthrolysis was 55°. This increased to 91°, six months after arthrolysis (p <
0.005). The improved range of movement was maintained during the follow-up period. No patient has required revision of their components. We have found arthrolysis to be a useful and successful approach to post-TKR stiffness.
We report the clinical and radiological results of a two- to three-year prospective randomised study which was designed to compare a minimally-invasive technique with a standard technique in total knee replacement and was undertaken between January 2004 and May 2007. The mini-midvastus approach was used on 50 patients (group A) and a standard approach on 50 patients (group B). The mean follow-up in both groups was 23 months (24 to 35). The functional outcome was better in group A up to nine months after operation, as shown by statistically significant differences in the mean function score, mean total score and the mean Oxford knee score (all, p = 0.05). Patients in group A had statistically significant greater early flexion (p = 0.04) and reached their greatest mean knee flexion of 126.5° (95° to 135°) 21 days after operation. However, at final follow-up there was no significant difference in the mean maximum flexion between the groups (p = 0.08). Technical errors were identified in six patients from group A (12%) on radiological evaluation. Based on these results, the authors currently use minimally-invasive techniques in total knee replacement in selected cases only.
We have examined the differences in clinical outcome of total knee replacement (TKR) with and without patellar resurfacing in a prospective, randomised study of 181 osteoarthritic knees in 142 patients using the Profix total knee system which has a femoral component with features considered to be anatomical and a domed patellar implant. The procedures were carried out between February 1998 and November 2002. A total of 159 TKRs in 142 patients were available for review at a mean of four years (3 to 7). The patients and the clinical evaluator were blinded in this prospective study. Evaluation was undertaken annually by an independent observer using the knee pain scale and the Knee Society clinical rating system. Specific evaluation of anterior knee pain, stair-climbing and rising from a seated to a standing position was also undertaken. No benefit was shown of TKR with patellar resurfacing over that without resurfacing with respect to any of the measured outcomes. In 22 of 73 knees (30.1%) with and 18 of 86 knees (20.9%) without patellar resurfacing there was some degree of anterior knee pain (p = 0.183). No revisions related to the patellofemoral joint were performed in either group. Only one TKR in each group underwent a re-operation related to the patellofemoral joint. A significant association between knee flexion contracture and anterior knee pain was observed in those knees with patellar resurfacing (p = 0.006).