We analysed 54 alumina ceramic-on-ceramic bearings
from total hip replacements retrieved at one centre after a mean
duration of 3.5 years (0.2 to 10.6) in situ. These
implants were obtained from 54 patients (16 men and 38 women) with
a mean age of 67 years (33 to 88) who underwent revision for a variety
of reasons. Posterior edge loading was found in the majority of
these retrievals (32 out of 54). Anterosuperior edge loading occurred
less often but produced a higher rate of wear. Stripe wear on the
femoral heads had a median volumetric wear rate of 0.2 mm3/year
(0 to 7.2). The wear volume on the femoral heads corresponded to
the width of edge wear on the matching liner. Anteversion of the
acetabular component was found to be a more important determinant
than inclination for wear in ceramic bearings. Posterior edge loading
may be considered to be a normal occurrence in ceramic-on-ceramic
bearings, with minimal clinical consequences. Edge loading should
be defined as either anterosuperior or posterior, as each edge loading
mechanism may result in different clinical implications.