Aims. Implant failure has become more common as the number of primary total ankle arthroplasties (TAAs) performed has increased. Although revision arthroplasty has gained attention for functional preservation, the long-term results remain unclear. This study aimed to assess the long-term outcomes of revision TAA using a mobile-bearing prosthesis in a considerably large cohort; the risk factors for failure were also determined. Methods. This single-centre retrospective cohort study included 116 patients (117 ankles) who underwent revision TAA for failed primary TAA between July 2000 and March 2010. Survival analysis and risk factor assessment were performed, and clinical performance and patient satisfaction were evaluated preoperatively and at last follow-up. Results. The mean duration from initial revision TAA to last follow-up was 15.0 years (SD 3.0; 11.2 to 20.5). The cumulative
Aims. Despite the increasing numbers of ankle arthroplasties, there are limited studies on their survival and comparisons between different implants. The primary aim of this study was to determine the failure rates of primary ankle arthroplasties commonly used in the UK. Methods. A data linkage study combined National Joint Registry (NJR) data and NHS Digital data. The primary outcome of failure was defined as the removal or exchange of any components of the implanted device. Life tables and Kaplan-Meier survival charts were used to illustrate survivorship. Cox proportional hazards regression models were fitted to compare failure rates between 1 April 2010 and 31 December 2018. Results. Overall, 5,562 primary ankle arthroplasties were recorded in the NJR. Linked data show a one-year survivorship of 98.8% (95% confidence interval (CI) 98.4% to 99.0%), five-year survival in 2,725 patients of 90.2% (95% CI 89.2% to 91.1%), and ten-year survival in 199 patients of 86.2% (95% CI 84.6% to 87.6%). The five-year survival for fixed-bearing implants was 94.3% (95% CI 91.3% to 96.3%) compared to 89.4% (95% CI 88.3% to 90.4%) for mobile-bearing implants. A Cox regression model for all implants with over 100 implantations using the implant with the best survivorship (Infinity) as the reference, only the STAR (hazard ratio (HR) 1.60 (95% CI 0.87 to 2.96)) and INBONE (HR 0.38 (95% CI 0.05 to 2.84)) did not demonstrate worse survival at three and five years. Conclusion. Ankle arthroplasties in the UK have a five-year
Aims. This study reports updates the previously published two-year clinical, functional, and radiological results of a group of patients who underwent transfibular total ankle arthroplasty (TAA), with follow-up extended to a minimum of five years. Methods. We prospectively evaluated 89 patients who underwent transfibular TAA for end-stage osteoarthritis. Patients’ clinical and radiological examinations were collected pre- and postoperatively at six months and then annually for up to five years of follow-up. Three patients were lost at the final follow-up with a total of 86 patients at the final follow-up. Results. A total of 86 patients were evaluated at a mean follow-up of 65.4 months (60 to 90). At five-year follow-up, statistically significant improvements (p < 0.001) were found in the mean American Orthopaedic Foot & Ankle Society Ankle Hindfoot Score (from 33.8 (SD 14.3) to 86.1 (SD 8.8)), visual analogue scale for pain (from 8.5 (SD 1.7) to 1.5 (SD 1.2)), Short Form-12 Physical and Mental Component Scores (from 29.9 (SD 6.7) and 43.3 (SD 8.6) to 47.3 (SD 7.5) and 52.2 (SD 8.0), respectively), and mean ankle dorsiflexion and plantarflexion (from 6.2° (SD 5.5°) and 9.6° (SD 5.8°) to 23.9° (SD 7.7°) and 16.9° (SD 7.2°), respectively). Radiologically, the implants maintained neutral alignment without subsidence. Tibial or talar radiolucency was found in eight patients, but none of these patients was symptomatic. At five-year follow up, 97.7% of implants (95% confidence interval 91.2 to 99.4) were free from revision or removal with 84 implants at risk. We recorded two cases (2.3%) of failure for septic loosening. Conclusion. Transfibular TAA is safe and effective with a high
Aims. We report the medium-term outcomes of a consecutive series of 118 Zenith total ankle arthroplasties (TAAs) from a single, non-designer centre. Methods. Between December 2010 and May 2016, 118 consecutive Zenith prostheses were implanted in 114 patients. Demographic, clinical, and patient-reported outcome measures (PROMs) data were collected. The endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan-Meier survival curves were generated with 95% confidence intervals (CIs) and the rate of failure calculated for each year. Results. Eight patients (ten ankles) died during follow-up, but none required revision. Of the surviving 106 patients (108 ankles: rheumatoid arthritis (RA), n = 15; osteoarthritis (OA), n = 93), 38 were women and 68 were men, with a mean age of 68.2 years (48 to 86) at the time of surgery. Mean follow-up was 5.1 years (2.1 to 9.0). A total of ten implants failed (8.5%), thus requiring revision. The implant survival at seven years, using revision as an endpoint, was 88.2% (95% CI 100% to 72.9%). There was a mean improvement in Manchester-Oxford Foot and Ankle Questionnaire (MOXFQ) from 85.0 to 32.8 and visual analogue scale (VAS) scores from 7.0 to 3.2, and overall satisfaction was 89%. The three commonest complications were malleolar fracture (14.4%, n = 17), wound healing (13.6%, n = 16), and superficial infection (12.7%, n = 15). The commonest reason for revision was aseptic loosening. No patients in our study were revised for deep infection. Conclusion. Our results show that Zenith
When a total ankle arthroplasty (TAA) fails, it can be converted to a fusion or a revision arthroplasty. Despite the increasing numbers of TAAs being undertaken, there is little information in the literature about the management of patients undergoing fusion following a failed TAA. The primary aim of this study was to analyze the survival of fusions following a failed TAA using a large dataset from the National Joint Registry (NJR). A data linkage study combined NJR and NHS Digital data. Failure of a TAA was defined as a fusion, revision to a further TAA, or amputation. Life tables and Kaplan-Meier graphs were used to record survival. Cox proportional hazards regression models were fitted to compare the rates of failure.Aims
Methods
Aims. The number of revision total ankle arthroplasties (TAAs) which are undertaken is increasing. Few studies have reported the survival after this procedure. The primary aim of this study was to analyze the survival of revision ankle arthroplasties using large datasets. Secondary aims were to summarize the demographics of the patients, the indications for revision TAA, further operations, and predictors of survival. Methods. The study combined data from the National Joint Registry and NHS Digital to report the survival of revision TAA. We have previously reported the failure rates and risk factors for failure after TAA, and the outcome of fusion after a failed TAA, using the same methodology. Survival was assessed using life tables and Kaplan Meier graphs. Cox proportional hazards regression models were fitted to compare failure rates. Results. A total of 228 patients underwent revision TAA. The mean follow-up was 2.6 years (SD 2.0). The mean time between the initial procedure and revision was 2.3 years (SD 1.8). The most commonly used implant was the Inbone which was used in 81 patients. A total of 29 (12.7%) failed; nine (3.9%) patients underwent a further revision, 19 (8.3%) underwent a fusion, and one (0.4%) had an amputation. The
The Vantage Total Ankle System is a fourth-generation low-profile fixed-bearing implant that has been available since 2016. We aimed to describe our early experience with this implant. This is a single-centre retrospective review of patients who underwent primary total ankle arthroplasty (TAA) with a Vantage implant between November 2017 and February 2020, with a minimum of two years’ follow-up. Four surgeons contributed patients. The primary outcome was reoperation and revision rate of the Vantage implant at two years. Secondary outcomes included radiological alignment, peri-implant complications, and pre- and postoperative patient-reported outcomes.Aims
Methods
Calcaneal osteomyelitis remains a difficult condition to treat with high rates of recurrence and below-knee amputation, particularly in the presence of severe soft-tissue destruction. This study assesses the outcomes of single-stage orthoplastic surgical treatment of calcaneal osteomyelitis with large soft-tissue defects. A retrospective review was performed of all patients who underwent combined single-stage orthoplastic treatment of calcaneal osteomyelitis (01/2008 to 12/2022). Primary outcome measures were osteomyelitis recurrence and below-knee amputation (BKA). Secondary outcome measures included flap failure, operating time, complications, and length of stay.Aims
Methods
This scoping review aims to identify patient-related factors associated with a poorer outcome following total ankle arthroplasty (TAA). A scoping review was performed using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines. A computer-based literature search was performed in PubMed, Embase, Cochrane trials, and Web of Science. Two reviewers independently performed title/abstract and full-text screening according to predetermined selection criteria. English-language original research studies reporting patient-related factors associated with a poorer outcome following TAA were included. Outcomes were defined as patient-reported outcome measures (PROMs), perioperative complications, and failure.Aims
Methods
We present the outcomes in 38 consecutive patients who had total ankle replacement using the Ankle Evolution System with a minimum follow-up of four years. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) score and regular standardised anteroposterior and lateral weight-bearing radiographs were obtained. Patient satisfaction and complications were recorded and the survival of the implants was demonstrated by the Kaplan-Meier method. The mean follow-up was for 57.8 months (48 to 80). The cumulative
In our unit, we adopt a two-stage surgical reconstruction approach using internal fixation for the management of infected Charcot foot deformity. We evaluate our experience with this functional limb salvage method. We conducted a retrospective analysis of prospectively collected data of all patients with infected Charcot foot deformity who underwent two-stage reconstruction with internal fixation between July 2011 and November 2019, with a minimum of 12 months’ follow-up.Aims
Methods
This is a multicentre, non-inventor, prospective observational study of 503 INFINITY fixed bearing total ankle arthroplasties (TAAs). We report our early experience, complications, and radiological and functional outcomes. Patients were recruited from 11 specialist centres between June 2016 and November 2019. Demographic, radiological, and functional outcome data (Ankle Osteoarthritis Scale, Manchester Oxford Questionnaire, and EuroQol five-dimension five-level score) were collected preoperatively, at six months, one year, and two years. The Canadian Orthopaedic Foot and Ankle Society (COFAS) grading system was used to stratify deformity. Early and late complications and reoperations were recorded as adverse events. Radiographs were assessed for lucencies, cysts, and/or subsidence.Aims
Methods
To assess the effect of age on clinical outcome and revision rates in patients who underwent total ankle arthroplasty (TAA) for end-stage ankle osteoarthritis (OA). A consecutive series of 811 ankles (789 patients) that underwent TAA between May 2003 and December 2013 were enrolled. The influence of age on clinical outcome, including the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot score, and pain according to the visual analogue scale (VAS) was assessed. In addition, the risk for revision surgery that includes soft tissue procedures, periarticular arthrodeses/osteotomies, ankle joint debridement, and/or inlay exchange (defined as minor revision), as well as the risk for revision surgery necessitating the exchange of any of the metallic components or removal of implant followed by ankle/hindfoot fusion (defined as major revision) was calculated.Aims
Methods
Preoperative talar valgus deformity ≥ 15° is considered a contraindication for total ankle arthroplasty (TAA). We compared operative procedures and clinical outcomes of TAA in patients with talar valgus deformity ≥ 15° and < 15°. A matched cohort of patients similar for demographics and components used but differing in preoperative coronal-plane tibiotalar valgus deformity ≥ 15° (valgus, n = 50; 52% male, mean age 65.8 years (SD 10.3), mean body mass index (BMI) 29.4 (SD 5.2)) or < 15° (control, n = 50; 58% male, mean age 65.6 years (SD 9.8), mean BMI 28.7 (SD 4.2)), underwent TAA by one surgeon. Preoperative and postoperative radiographs, Ankle Osteoarthritis Scale (AOS) pain and disability and 36-item Short Form Health Survey (SF-36) version 2 scores were collected prospectively. Ancillary procedures, secondary procedures, and complications were recorded.Aims
Methods
Arthroplasty for end-stage hallux rigidus (HR) is controversial. Arthrodesis remains the gold standard for surgical treatment, although is not without its complications, with rates of up to 10% for nonunion, 14% for reoperation and 10% for metatarsalgia. The aim of this study was to analyze the outcome of a double-stemmed silastic implant (Wright-Medical, Memphis, Tennessee, USA) for patients with end-stage HR. We conducted a retrospective review of 108 consecutive implants in 76 patients, between January 2005 and December 2016, with a minimum follow-up of two years. The mean age of the patients at the time of surgery was 61.6 years (42 to 84). There were 104 females and four males. Clinical, radiological, patient reported outcome measures (PROMS) data, a visual analogue score (VAS) for pain, and satisfaction scores were collected.Aims
Methods
The purpose of this study was to determine the functional outcome and implant survivorship of mobile-bearing total ankle arthroplasty (TAA) performed by a single surgeon. We reviewed 205 consecutive patients (210 ankles) who had undergone mobile-bearing TAA (205 patients) for osteoarthritis of the ankle between January 2005 and December 2015. Their mean follow-up was 6.4 years (2.0 to 13.4). Functional outcome was assessed using the Ankle Osteoarthritis Scale, American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot score, 36-Item Short-Form Health Survey (SF-36) score, visual analogue scale, and range of movement. Implant survivorship and complications were also evaluated.Aims
Patients and Methods
We report the long-term clinical and radiological outcomes of a consecutive series of 200 total ankle arthroplasties (TAAs, 184 patients) at a single centre using the Scandinavian Total Ankle Replacement (STAR) implants. Between November 1993 and February 2000, 200 consecutive STAR prostheses were implanted in 184 patients by a single surgeon. Demographic and clinical data were collected prospectively and the last available status was recorded for further survival analysis. All surviving patients underwent regular clinical and radiological review. Pain and function were assessed using the American Orthopaedic Foot and Ankle Society (AOFAS) hindfoot scoring system. The principal endpoint of the study was failure of the implant requiring revision of one or all of the components. Kaplan–Meier survival curves were generated with 95% confidence intervals and the rate of failure calculated for each year.Aims
Patients and Methods
Total ankle arthroplasty (TAA) surgery is complex and attracts a wide variety of complications. The literature lacks consistency in reporting adverse events and complications. The aim of this article is to provide a comprehensive analysis of each of these complications from a literature review, and to compare them with rates from our Unit, to aid clinicians with the process of informed consent. A total of 278 consecutive total ankle arthroplasties (251 patients), performed by four surgeons over a six-year period in Wrightington Hospital (Wigan, United Kingdom) were prospectively reviewed. There were 143 men and 108 women with a mean age of 64 years (41 to 86). The data were recorded on each follow-up visit. Any complications either during initial hospital stay or subsequently reported on follow-ups were recorded, investigated, monitored, and treated as warranted. Literature search included the studies reporting the outcomes and complications of TAA implants.Aims
Patients and Methods
Total ankle arthroplasty (TAA) has become the most reliable surgical solution for patients with end-stage arthritis of the ankle. Aseptic loosening of the talar component is the most common complication. A custom-made artificial talus can be used as the talar component in a combined TAA for patients with poor bone stock of the talus. The purpose of this study was to investigate the functional and clinical outcomes of combined TAA. Ten patients (two men, eight women; ten ankles) treated using a combined TAA between 2009 and 2013 were matched for age, gender, and length of follow-up with 12 patients (one man, 11 women; 12 ankles) who underwent a standard TAA. All had end-stage arthritis of the ankle. The combined TAA features a tibial component of the TNK ankle (Kyocera, Kyoto, Japan) and an alumina ceramic artificial talus (Kyocera), designed using individualized CT data. The mean age at the time of surgery in the combined TAA and standard TAA groups was 71 years (Aims
Patients and Methods
The purpose of this study was to compare the clinical and radiographic
outcomes of total ankle arthroplasty (TAA) in patients with pre-operatively
moderate and severe arthritic varus ankles to those achieved for
patients with neutral ankles. A total of 105 patients (105 ankles), matched for age, gender,
body mass index, and follow-up duration, were divided into three
groups by pre-operative coronal plane tibiotalar angle; neutral
(<
5°), moderate (5° to 15°) and severe (>
15°) varus deformity.
American Orthopaedic Foot and Ankle Society (AOFAS) ankle-hindfoot
score, a visual analogue scale (VAS), and Short Form (SF)-36 score
were used to compare the clinical outcomes after a mean follow-up period
of 51 months (24 to 147).Aims
Patients and Methods