Aims. Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes. Methods. Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries. Results. In total, 1,312 Thompson’s hemiarthroplasties were analyzed (mean age at surgery 82.8 years); 125 complications were recorded, necessitating 82 returns to theatre. These included 14 patients undergoing aspiration or manipulation under anaesthesia, 68 reoperations (5.2%) for debridement and implant retention (n = 12), haematoma evacuation (n = 2), open reduction for dislocation (n = 1), fixation of periprosthetic fracture (n = 5), and 48 revised stems (3.7%), for infection (n = 13), dislocation (n = 12),
Aims. For displaced femoral neck fractures (FNFs) in geriatric patients, there remains uncertainty regarding the effect of total hip arthroplasty (THA) compared with hemiarthroplasty (HA) in the guidelines. We aimed to compare 90-day surgical readmission, in-hospital complications, and charges between THA and HA in these patients. Methods. The Hospital Quality Monitoring System was queried from 1 January 2013 to 31 December 2019 for displaced FNFs in geriatric patients treated with THA or HA. After propensity score matching, which identified 33,849 paired patients, outcomes were compared between THA and HA using logistic and linear regression models. Results. The HA group had a lower incidence of 90-day surgical readmission than the THA group (odds ratio (OR) 0.75 (95% CI 0.68 to 0.83)). Meanwhile, the HA group had lower incidence of dislocation (OR 0.42 (95% CI 0.33 to 0.52)),
Femoral periprosthetic fractures are rising in incidence. Their management is complex and carries a high associated mortality. Unlike native hip fractures, there are no guidelines advising on time to theatre in this group. We aim to determine whether delaying surgical intervention influences morbidity or mortality in femoral periprosthetic fractures. We identified all periprosthetic fractures around a hip or knee arthroplasty from our prospectively collated database between 2012 and 2021. Patients were categorized into early or delayed intervention based on time from admission to surgery (early = ≤ 36 hours, delayed > 36 hours). Patient demographics, existing implants, Unified Classification System fracture subtype, acute medical issues on admission, preoperative haemoglobin, blood transfusion requirement, and length of hospital stay were identified for all patients. Complication and mortality rates were compared between groups.Aims
Methods
The aim of this study was to investigate the rate of revision for distal femoral arthroplasty (DFA) performed as a primary procedure for native knee fractures using data from the Australian Orthopaedic Association National Joint Arthroplasty Registry (AOANJRR). Data from the AOANJRR were obtained for DFA performed as primary procedures for native knee fractures from 1 September 1999 to 31 December 2020. Pathological fractures and revision for failed internal fixation were excluded. The five prostheses identified were the Global Modular Arthroplasty System, the Modular Arthroplasty System, the Modular Universal Tumour And Revision System, the Orthopaedic Salvage System, and the Segmental System. Patient demographic data (age, sex, and American Society of Anesthesiologists grade) were obtained, where available. Kaplan-Meier estimates of survival were used to determine the rate of revision, and the reasons for revision and mortality data were examined.Aims
Methods
Hip fractures in patients < 60 years old currently account for only 3% to 4% of all hip fractures in England, but this proportion is increasing. Little is known about the longer-term patient-reported outcomes in this potentially more active population. The primary aim is to examine patient-reported outcomes following isolated hip fracture in patients aged < 60 years. The secondary aim is to determine an association between outcomes and different types of fracture pattern and/or treatment implants. All hip fracture patients aged 18 to 60 years admitted to a single centre over a 15-year period were used to identify the study group. Fracture pattern (undisplaced intracapsular, displaced intracapsular, and extracapsular) and type of operation (multiple cannulated hip screws, angular stable fixation, hemiarthroplasty, and total hip replacement) were recorded. The primary outcome measures were the Oxford Hip Score (OHS), the EuroQol five-dimension questionnaire (EQ-5D-3L), and EQ-visual analogue scale (VAS) scores. Preinjury scores were recorded by patient recall and postinjury scores were collected at a mean of 57 months (9 to 118) postinjury. Ethics approval was obtained prior to study commencement.Aims
Methods
Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses. A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented polished tapered stem hemiarthroplasty or an uncemented Furlong hydroxyapatite-coated hemiarthroplasty. Follow-up was conducted by a nurse blinded to the implant at set intervals for up to one year from surgery.Aims
Methods
The aim of this study was to describe implant and patient-reported outcome in patients with a unilateral transfemoral amputation (TFA) treated with a bone-anchored, transcutaneous prosthesis. In this cohort study, all patients with a unilateral TFA treated with the Osseointegrated Prostheses for the Rehabilitation of Amputees (OPRA) implant system in Sahlgrenska University Hospital, Gothenburg, Sweden, between January 1999 and December 2017 were included. The cohort comprised 111 patients (78 male (70%)), with a mean age 45 years (17 to 70). The main reason for amputation was trauma in 75 (68%) and tumours in 23 (21%). Patients answered the Questionnaire for Persons with Transfemoral Amputation (Q-TFA) before treatment and at two, five, seven, ten, and 15 years’ follow-up. A prosthetic activity grade was assigned to each patient at each timepoint. All mechanical complications, defined as fracture, bending, or wear to any part of the implant system resulting in removal or change, were recorded.Aims
Methods
To evaluate the outcomes of cemented total hip arthroplasty (THA)
following a fracture of the acetabulum, with evaluation of risk
factors and comparison with a patient group with no history of fracture. Between 1992 and 2016, 49 patients (33 male) with mean age of
57 years (25 to 87) underwent cemented THA at a mean of 6.5 years
(0.1 to 25) following acetabular fracture. A total of 38 had undergone
surgical fixation and 11 had been treated non-operatively; 13 patients
died at a mean of 10.2 years after THA (0.6 to 19). Patients were
assessed pre-operatively, at one year and at final follow-up (mean
9.1 years, 0.5 to 23) using the Oxford Hip Score (OHS). Implant
survivorship was assessed. An age and gender-matched cohort of THAs
performed for non-traumatic osteoarthritis (OA) or avascular necrosis
(AVN) (n = 98) were used to compare complications and patient-reported outcome
measures (PROMs).Aims
Patients and Methods
The aim of this study was to compare the cost-effectiveness of
treatment with an osseointegrated percutaneous (OI-) prosthesis
and a socket-suspended (S-) prosthesis for patients with a transfemoral
amputation. A Markov model was developed to estimate the medical costs and
changes in quality-adjusted life-years (QALYs) attributable to treatment
of unilateral transfemoral amputation over a projected period of
20 years from a healthcare perspective. Data were collected alongside
a prospective clinical study of 51 patients followed for two years.Aims
Patients and Methods
The anterior pelvic internal fixator is increasingly used for
the treatment of unstable, or displaced, injuries of the anterior
pelvic ring. The evidence for its use, however, is limited. The
aim of this paper is to describe the indications for its use, how
it is applied and its complications. We reviewed the case notes and radiographs of 50 patients treated
with an anterior pelvic internal fixator between April 2010 and
December 2015 at a major trauma centre in the United Kingdom. The
median follow-up time was 38 months (interquartile range 24 to 51).Aims
Patients and Methods
This is a prospective randomised study which compares the radiological
and functional outcomes of ring and rail fixators in patients with
an infected gap (>
3 cm) nonunion of the tibia. Between May 2008 and February 2013, 70 patients were treated
at our Institute for a posttraumatic osseocutaneous defect of the
tibia measuring at least 3 cm. These were randomised into two groups
of 35 patients using the lottery method. Group I patients were treated
with a ring fixator and group II patients with a rail fixator. The
mean age was 33.2 years (18 to 64) in group I and 29.3 years (18
to 65) in group II. The mean bone gap was 5.84 cm in group I and 5.78
cm in group II. The mean followup was 33.8 months in group I and 32.6 months
in group II. Bone and functional results were assessed using the
classification of the Association for the Study and Application
of the Method of Ilizarov (ASAMI). Functional results were also
assessed at six months using the short musculoskeletal functional
assessment (SMFA) score.Aims
Patients and Methods
In a series of 450 patients over 70 years of age with displaced fractures of the femoral neck sustained between 1995 and 1997 treatment was randomised either to internal fixation or replacement. Depending on age and level of activity the latter was either a total hip replacement or a hemiarthroplasty. Patients who were confused or bed-ridden were excluded, as were those with rheumatoid arthritis. At ten years there were 99 failures (45.6%) after internal fixation compared with 17 (8.8%) after replacement. The rate of mortality was high at 75% at ten years, and was the same in both groups at all times. Patient-reported pain and function were similar in both groups at five and ten years. Those with successfully healed fractures had more hip pain and reduction of mobility at four months compared with patients with an uncomplicated replacement, and they never attained a better outcome than the latter patients regarding pain or function. Primary replacement gave reliable long-term results in patients with a displaced fracture of the femoral neck.
Radiological evidence of post-traumatic osteoarthritis
(PTOA) after fracture of the tibial plateau is common but end-stage arthritis
which requires total knee arthroplasty is much rarer. The aim of this study was to examine the indications for, and
outcomes of, total knee arthroplasty after fracture of the tibial
plateau and to compare this with an age and gender-matched cohort
of TKAs carried out for primary osteoarthritis. Between 1997 and 2011, 31 consecutive patients (23 women, eight
men) with a mean age of 65 years (40 to 89) underwent TKA at a mean
of 24 months (2 to 124) after a fracture of the tibial plateau.
Of these, 24 had undergone ORIF and seven had been treated non-operatively.
Patients were assessed pre-operatively and at 6, 12 and >
60 months
using the Short Form-12, Oxford Knee Score and a patient satisfaction
score. Patients with instability or nonunion needed total knee arthroplasty
earlier (14 and 13.3 months post-injury) than those with intra-articular
malunion (50 months, p <
0.001). Primary cruciate-retaining implants
were used in 27 (87%) patients. Complication rates were higher in
the PTOA cohort and included wound complications (13% Total knee arthroplasty undertaken after fracture of the tibial
plateau has a higher rate of complications than that undertaken
for primary osteoarthritis, but patient-reported outcomes and satisfaction
are comparable. Cite this article:
Endoprosthetic replacement of the proximal femur may be required to treat primary bone tumours or destructive metastases either with impending or established pathological fracture. Modular prostheses are available off the shelf and can be adapted to most reconstructive situations for this purpose. We have assessed the clinical and functional outcome of using the METS (Stanmore Implants Worldwide) modular tumour prosthesis to reconstruct the proximal femur in 100 consecutive patients between 2001 and 2006. We compared the results with the published series for patients managed with modular and custom-made endoprosthetic replacements for the same conditions. There were 52 males and 48 females with a mean age of 56.3 years (16 to 84) and a mean follow-up of 24.6 months (0 to 60). In 65 patients the procedure was undertaken for metastases, in 25 for a primary bone tumour, and in ten for other malignant conditions. A total of 46 patients presented with a pathological fracture, and 19 presented with failed fixation of a previous pathological fracture. The overall patient survival was 63.6% at one year and 23.1% at five years, and was significantly better for patients with a primary bone tumour than for those with metastatic tumour (82.3% vs 53.3%, respectively at one year (p = 0.003)). There were six early dislocations of which five could be treated by closed reduction. No patient needed revision surgery for dislocation. Revision surgery was required by six (6%) patients, five for pain caused by acetabular wear and one for tumour progression. Amputation was needed in four patients for local recurrence or infection. The estimated five-year implant survival with revision as the endpoint was 90.7%. The mean Toronto Extremity Salvage score was 61% (51% to 95%). The implant survival and complications resulting from the use of the modular system were comparable to the published series of both custom-made and other modular proximal femoral implants. We conclude that at intermediate follow-up the modular tumour prosthesis for proximal femur replacement provides versatility, a low incidence of implant-related complications and acceptable function for patients with metastatic tumours, pathological fractures and failed fixation of the proximal femur. It also functions as well as a custom-made endoprosthetic replacement.
Gunshot injuries to the shoulder are rare and
difficult to manage. We present a case series of seven patients
who sustained a severe shoulder injury to the non-dominant side
as a result of a self-inflicted gunshot wound. We describe the injury
as ‘suicide shoulder’ caused by upward and outward movement of the
gun barrel as the trigger is pulled. All patients were male, with
a mean age of 32 years (21 to 48). All were treated at the time
of injury with initial repeated debridement, and within four weeks
either by hemiarthroplasty (four patients) or arthrodesis (three patients).
The hemiarthroplasty failed in one patient after 20 years due to
infection and an arthrodesis was attempted, which also failed due
to infection. Overall follow-up was for a mean of 26 months (12
to 44). All four hemiarthroplasty implants were removed with no
feasible reconstruction ultimately possible, resulting in a poor functional
outcome and no return to work. In contrast, all three primary arthrodeses
eventually united, with two patients requiring revision plating
and grafting. These patients returned to work with a good functional
outcome. We recommend arthrodesis rather than replacement as the
treatment of choice for this challenging injury. Cite this article:
Using data from the Norwegian Hip Fracture Register,
8639 cemented and 2477 uncemented primary hemiarthroplasties for
displaced fractures of the femoral neck in patients aged > 70 years
were included in a prospective observational study. A total of 218
re-operations were performed after cemented and 128 after uncemented
procedures. Survival of the hemiarthroplasties was calculated using
the Kaplan-Meier method and hazard rate ratios (HRR) for revision
were calculated using Cox regression analyses. At five years the
implant survival was 97% (95% confidence interval (CI) 97 to 97)
for cemented and 91% (95% CI 87 to 94) for uncemented hemiarthroplasties.
Uncemented hemiarthroplasties had a 2.1 times increased risk of
revision compared with cemented prostheses (95% confidence interval
1.7 to 2.6, p < 0.001). The increased risk was mainly caused
by revisions for peri-prosthetic fracture (HRR = 17), aseptic loosening
(HRR = 17), haematoma formation (HRR = 5.3), superficial infection
(HRR = 4.6) and dislocation (HRR = 1.8). More intra-operative complications,
including intra-operative death, were reported for the cemented
hemiarthroplasties. However, in a time-dependent analysis, the HRR
for re-operation in both groups increased as follow-up increased. This study showed that the risk for revision was higher for uncemented
than for cemented hemiarthroplasties.
United Kingdom National Institute for Health
and Clinical Excellence guidelines recommend the use of total hip replacement
(THR) for displaced intracapsular fractures of the femoral neck
in cognitively intact patients, who were independently mobile prior
to the injury. This study aimed to analyse the risk factors associated
with revision of the implant and mortality following THR, and to
quantify risk. National Joint Registry data recording a THR performed
for acute fracture of the femoral neck between 2003 and 2010 were
analysed. Cox proportional hazards models were used to investigate
the extent to which risk of revision was related to specific covariates.
Multivariable logistic regression was used to analyse factors affecting
peri-operative mortality (<
90 days). A total of 4323 procedures
were studied. There were 80 patients who had undergone revision
surgery at the time of censoring (five-year revision rate 3.25%, 95%
confidence interval 2.44 to 4.07) and 137 patients (3.2%) patients
died within 90 days. After adjusting for patient and surgeon characteristics,
an increased risk of revision was associated with the use of cementless
prostheses compared with cemented (hazard ratio (HR) 1.33, p = 0.021).
Revision was independent of bearing surface and head size. The risk
of mortality within 90 days was significantly increased with higher
American Society of Anesthesiologists (ASA) grade (grade 3: odds
ratio (OR) 4.04, p <
0.001; grade 4/5: OR 20.26, p <
0.001;
both compared with grades 1/2) and older age (≥ 75 years: OR 1.65,
p = 0.025), but reduced over the study period (9% relative risk reduction
per year). THR is a good option in patients aged <
75 years and with
ASA 1/2. Cementation of the femoral component does not adversely
affect peri-operative mortality but improves survival of the implant
in the mid-term when compared with cementless femoral components.
There are no benefits of using head sizes >
28 mm or bearings other
than metal-on-polyethylene. More research is required to determine
the benefits of THR over hemiarthroplasty in older patients and
those with ASA grades >
2.