Aims. The
Aims. The aim of this independent multicentre study was to assess the
mid-term results of
Polyethylene particulate wear debris continues to be implicated in the aetiology of aseptic loosening following knee arthroplasty. The Oxford unicompartmental knee arthroplasty employs a spherical femoral component and a fully congruous meniscal bearing to increase contact area and theoretically reduce the potential for polyethylene wear. This study measures the In this The results from this
This single-blinded randomised controlled trial
investigated whether one design of mobile-bearing (MB) total knee replacement
(TKR) has any advantage over a fixed-bearing (FB) design on long-term
fixation as measured by radiostereometry. The amount of wear underneath
the
There is conflicting evidence about the merits
of mobile bearings in total knee replacement, partly because most randomised
controlled trials (RCTs) have not been adequately powered. We report
the results of a multicentre RCT of mobile There was no significant difference between the groups pre-operatively:
mean OKS was 17.18 ( In this appropriately powered RCT, over the first five years
after total knee replacement functional outcomes, re-operation rates
and healthcare costs appear to be the same irrespective of whether
a mobile or fixed bearing is used. Cite this article:
Aims. Uncemented
It has been suggested that mobile-bearing total knee arthroplasty
(TKA) might lead to better outcomes by accommodating some femorotibial
rotational mismatch, thereby reducing contact stresses and polyethylene
wear. The aim of this study was to determine whether there is a
difference between fixed- and mobile-bearing versions of a contemporary
TKA with respect to durability, range of movement (ROM) and function,
ten years postoperatively. A total of 240 patients who were enrolled in this randomized
controlled trial (RCT) underwent a primary cemented TKA with one
of three tibial components (all-polyethylene fixed-bearing, modular
metal-backed fixed-bearing and mobile-bearing). Patients were reviewed
at a median follow-up of ten years (IQR 9.2 to 10.4).Aims
Patients and Methods
Aims. To report mid- to long-term results of Oxford
Aims. For many designs of total knee arthroplasty (TKA) it remains unclear whether cemented or uncemented fixation provides optimal long-term survival. The main limitation in most studies is a retrospective or non-comparative study design. The same is true for comparative trials looking only at the survival rate as extensive sample sizes are needed to detect true differences in fixation and durability. Studies using radiostereometric analysis (RSA) techniques have shown to be highly predictive in detecting late occurring aseptic loosening at an early stage. To investigate the difference in predicted long-term survival between cemented, uncemented, and hybrid fixation of TKA, we performed a randomized controlled trial using RSA. Methods. A total of 105 patients were randomized into three groups (cemented, uncemented, and hybrid fixation of the ACS
Aims. This study compared the cobalt and chromium serum ion concentration of patients treated with two different metal-on-metal (MoM) hinged total knee arthroplasty (TKA) systems, as well as a titanium nitride (TiN)-coated variant. Methods. A total of 63 patients (65 implants) were treated using either a MoM-coated (n = 29) or TiN-coated (n = 7) hinged TKA (GenuX
The incidence of anterior knee pain following
total knee replacement (TKR) is reported to be as high as 49%. The source
of the pain is poorly understood but the soft tissues around the
patella have been implicated. In theory circumferential electrocautery denervates the patella
thereby reducing efferent pain signals. However, there is mixed
evidence that this practice translates into improved outcomes. We aimed to investigate the clinical effect of intra-operative
circumpatellar electrocautery in patients undergoing TKR using the
LCS
Aims. The objectives of this study were to compare postoperative pain, analgesia requirements, inpatient functional rehabilitation, time to hospital discharge, and complications in patients undergoing conventional jig-based unicompartmental knee arthroplasty (UKA) versus robotic-arm assisted UKA. Patients and Methods. This prospective cohort study included 146 patients with symptomatic medial compartment knee osteoarthritis undergoing primary UKA performed by a single surgeon. This included 73 consecutive patients undergoing conventional jig-based
We conducted a randomised prospective study to evaluate the clinical and radiological results of a mobile- and fixed-bearing total knee replacement of similar design in 174 patients who had bilateral simultaneous knee replacement. The mean follow-up was for 5.6 years (5.2 to 6.1). The total knee score, pain score, functional score and range of movement were not statistically different (p >
0.05) between the two groups. Osteolysis was not seen in any knee in either group. Two knees (1%) in the mobile-bearing group required revision because of infection; none in the fixed-bearing group needed revision. Excellent results can be achieved with both mobile- and fixed-bearing prostheses of similar design at mid-term follow-up. We could demonstrate no significant clinical advantage for a
Medial unicompartmental knee arthroplasty (mUKA) is an advised treatment for anteromedial knee osteoarthritis. While long-term survival after mUKA is well described, reported incidences of short-term surgical complications vary and the effect of surgical usage on complications is less established. We aimed to describe the overall occurrence and treatment of surgical complications within 90 days of mUKA, as well as occurrence in high-usage centres compared to low-usage centres. mUKAs performed in eight fast-track centres from February 2010 to June 2018 were included from the Lundbeck Foundation Centre for Fast-track Hip and Knee Replacement Database. All readmissions within 90 days of surgery underwent chart review and readmissions related to the surgical wound or the prosthesis were recorded. Centres were categorized as high-usage centres when using mUKA in ≥ 20% of annual knee arthroplasties. The occurrence of complications between high- and low-usage centres were compared using Fisher’s exact test.Aims
Methods
The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution. Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).Aims
Methods
The aim of this study was to determine whether obesity had a detrimental effect on the long-term performance and survival of medial unicompartmental knee arthroplasties (UKAs). This study reviewed prospectively collected functional outcome scores and revision rates of all medial UKA patients with recorded BMI performed in Christchurch, New Zealand, from January 2011 to September 2021. Patient-reported outcome measures (PROMs) were the primary outcome of this study, with all-cause revision rate analyzed as a secondary outcome. PROMs were taken preoperatively, at six months, one year, five years, and ten years postoperatively. There were 873 patients who had functional scores recorded at five years and 164 patients had scores recorded at ten years. Further sub-group analysis was performed based on the patient’s BMI. Revision data were available through the New Zealand Joint Registry for 2,323 UKAs performed during this time period.Aims
Methods
The low contact stress patellofemoral replacement consists of a trochlear component and a modular patellar component which has a metal-backed
The Oxford unicompartmental knee replacement (UKR) was designed to minimise wear utilising a fully-congruent,
We have studied the concept of posterior condylar offset and the importance of its restoration on the maximum range of knee flexion after posterior-cruciate-ligament-retaining total knee replacement (TKR). We measured the difference in the posterior condylar offset before and one year after operation in 69 patients who had undergone a primary cruciate-sacrificing
Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA. Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation.Aims
Methods
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
Higher osteoblastic bone activity is expected in aseptic loosening and painful unicompartmental knee arthroplasty (UKA). However, insights into normal bone activity patterns after medial UKAs are lacking. The aim of this study was to identify the evolution in bone activity pattern in well-functioning medial mobile-bearing UKAs. In total, 34 patients (13 female, 21 male; mean age 62 years (41 to 79); BMI 29.7 kg/m2 (23.6 to 42.1)) with 38 medial Oxford partial UKAs (20 left, 18 right; 19 cementless, 14 cemented, and five hybrid) were prospectively followed with sequential 99mTc-hydroxymethane diphosphonate single photon emission CT (SPECT)/CT preoperatively, and at one and two years postoperatively. Changes in mean osteoblastic activity were investigated using a tracer localization scheme with volumes of interest (VOIs), reported by normalized mean tracer values. A SPECT/CT registration platform additionally explored cortical tracer evolution in zones of interest identified by previous experimental research.Aims
Methods
Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.Aims
Methods
Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated.Aims
Methods
Debate continues regarding the optimum management of periprosthetic distal femoral fractures (PDFFs). This study aims to determine which operative treatment is associated with the lowest perioperative morbidity and mortality when treating low (Su type II and III) PDFFs comparing lateral locking plate fixation (LLP-ORIF) or distal femoral arthroplasty (DFA). This was a retrospective cohort study of 60 consecutive unilateral (PDFFs) of Su types II (40/60) and III (20/60) in patients aged ≥ 60 years: 33 underwent LLP-ORIF (mean age 81.3 years (SD 10.5), BMI 26.7 (SD 5.5); 29/33 female); and 27 underwent DFA (mean age 78.8 years (SD 8.3); BMI 26.7 (SD 6.6); 19/27 female). The primary outcome measure was reoperation. Secondary outcomes included perioperative complications, calculated blood loss, transfusion requirements, functional mobility status, length of acute hospital stay, discharge destination and mortality. Kaplan-Meier survival analysis was performed. Cox multivariate regression analysis was performed to identify risk factors for reoperation after LLP-ORIF.Aims
Methods
Cementless unicompartmental knee arthroplasty (UKA) has advantages over cemented UKA, including improved fixation, but has a higher risk of tibial plateau fracture, particularly in Japanese patients. The aim of this multicentre study was to determine when cementless tibial components could safely be used in Japanese patients based on the size and shape of the tibia. The study involved 212 cementless Oxford UKAs which were undertaken in 174 patients in six hospitals. The medial eminence line (MEL), which is a line parallel to the tibial axis passing through the tip of medial intercondylar eminence, was drawn on preoperative radiographs. Knees were classified as having a very overhanging medial tibial condyle if this line passed medial to the medial tibial cortex. They were also classified as very small if a size A/AA tibial component was used.Aims
Methods
To compare patients undergoing total knee arthroplasty (TKA) with ≤ 80° range of movement (ROM) operated with a 2 mm increase in the flexion gap with matched non-stiff patients with at least 100° of preoperative ROM and balanced flexion and extension gaps. In a retrospective cohort study, 98 TKAs (91 patients) with a preoperative ROM of ≤ 80° were examined. Mean follow-up time was 53 months (24 to 112). All TKAs in stiff knees were performed with a 2 mm increased flexion gap. Data were compared to a matched control group of 98 TKAs (86 patients) with a mean follow-up of 43 months (24 to 89). Knees in the control group had a preoperative ROM of at least 100° and balanced flexion and extension gaps. In all stiff and non-stiff knees posterior stabilized (PS) TKAs with patellar resurfacing in combination with adequate soft tissue balancing were used.Aims
Methods
The primary aim of the study was to compare the knee-specific functional outcome of robotic unicompartmental knee arthroplasty (rUKA) with manual total knee arthroplasty (mTKA) for the management of isolated medial compartment osteoarthritis. Secondary aims were to compare length of hospital stay, general health improvement, and satisfaction between rUKA and mTKA. A powered (1:3 ratio) cohort study was performed. A total of 30 patients undergoing rUKA were propensity score matched to 90 patients undergoing mTKA for isolated medial compartment arthritis. Patients were matched for age, sex, body mass index (BMI), and preoperative function. The Oxford Knee Score (OKS) and EuroQol five-dimension questionnaire (EQ-5D) were collected preoperatively and six months postoperatively. The Forgotten Joint Score (FJS) and patient satisfaction were collected six months postoperatively. Length of hospital stay was also recorded.Aims
Methods
The treatment of patients with allergies to metal in total joint arthroplasty is an ongoing debate. Possibilities include the use of hypoallergenic prostheses, as well as the use of standard cobalt-chromium (CoCr) alloy. This non-designer study was performed to evaluate the clinical outcome and survival rates of unicondylar knee arthroplasty (UKA) using a standard CoCr alloy in patients reporting signs of a hypersensitivity to metal. A consecutive series of patients suitable for UKA were screened for symptoms of metal hypersensitivity by use of a questionnaire. A total of 82 patients out of 1737 patients suitable for medial UKA reporting cutaneous metal hypersensitivity to cobalt, chromium, or nickel were included into this study and prospectively evaluated to determine the functional outcome, possible signs of hypersensitivity, and short-term survivorship at a minimum follow-up of 1.5 years.Aims
Patients and Methods
The aim of this study was to determine the polyethylene wear rate of Phase 3 Oxford Unicompartmental Knee Replacement bearings and to investigate the effects of resin type and manufacturing process. A total of 63 patients with at least ten years’ follow-up with three bearing types (1900 resin machined, 1050 resin machined, and 1050 resin moulded) were recruited. Patients underwent full weight-bearing model-based radiostereometric analysis to determine the bearing thickness. The linear wear rate was estimated from the change in thickness divided by the duration of implantation.Objectives
Methods
This prospective study reports the 15-year survival and ten-year
functional outcome of a consecutive series of 1000 minimally invasive
Phase 3 Oxford medial UKAs (818 patients, 393 men, 48%, 425 women,
52%, mean age 66 years; 32 to 88). These were implanted by two surgeons
involved with the design of the prosthesis to treat anteromedial
osteoarthritis and spontaneous osteonecrosis of the knee, which
are recommended indications. Patients were prospectively identified
and followed up independently for a mean of 10.3 years (5.3 to 16.6). At ten years, the mean Oxford Knee Score was 40 (standard deviation
( This is the only large series of minimally invasive UKAs with
15-year survival data. The results support the continued use of
minimally invasive UKA for the recommended indications. Cite this article:
The interest in unicompartmental knee arthroplasty (UKA) for
medial osteoarthritis has increased rapidly but the long-term follow-up
of the Oxford UKAs has yet to be analysed in non-designer centres.
We have examined our ten- to 15-year clinical and radiological follow-up
data for the Oxford Phase III UKAs. Between January 1999 and January 2005 a total of 138 consecutive
Oxford Phase III arthroplasties were performed by a single surgeon
in 129 patients for medial compartment osteoarthritis (71 right
and 67 left knees, mean age 72.0 years (47 to 91), mean body mass
index 28.2 (20.7 to 52.2)). Both clinical data and radiographs were
prospectively recorded and obtained at intervals. Of the 129 patients,
32 patients (32 knees) died, ten patients (12 knees) were not able
to take part in the final clinical and radiological assessment due
to physical and mental conditions, but via telephone interview it
was confirmed that none of these ten patients (12 knees) had a revision
of the knee arthroplasty. One patient (two knees) was lost to follow-up.Aims
Patients and Methods
This systematic review and meta-analysis was conducted to determine
the mid- to long-term clinical outcomes for a medial-pivot total
knee replacement (TKR) system. The objectives were to synthesise
available survivorship, Knee Society Scores (KSS), and reasons for
revision for this system. A systematic search was conducted of two online databases to
identify sources of survivorship, KSS, and reasons for revision.
Survivorship results were compared with values in the National Joint
Registry of England, Wales, and Northern Ireland (NJR).Objectives
Methods
This study compared the outcome of total knee
replacement (TKR) in adult patients with fixed- and mobile-bearing prostheses
during the first post-operative year and at five years’ follow-up,
using gait parameters as a new objective measure. This double-blind
randomised controlled clinical trial included 55 patients with mobile-bearing (n
= 26) and fixed-bearing (n = 29) prostheses of the same design,
evaluated pre-operatively and post-operatively at six weeks, three
months, six months, one year and five years. Each participant undertook
two walking trials of 30 m and completed the EuroQol questionnaire,
Western Ontario and McMaster Universities osteoarthritis index,
Knee Society score, and visual analogue scales for pain and stiffness.
Gait analysis was performed using five miniature angular rate sensors
mounted on the trunk (sacrum), each thigh and calf. The study population
was divided into two groups according to age (≤ 70 years Improvements in most gait parameters at five years’ follow-up
were greater for fixed-bearing TKRs in older patients (>
70 years),
and greater for mobile-bearing TKRs in younger patients (≤ 70 years).
These findings should be confirmed by an extended age controlled
study, as the ideal choice of prosthesis might depend on the age
of the patient at the time of surgery.
The lateral compartment is predominantly affected
in approximately 10% of patients with osteoarthritis of the knee. The
anatomy, kinematics and loading during movement differ considerably
between medial and lateral compartments of the knee. This in the
main explains the relative protection of the lateral compartment
compared with the medial compartment in the development of osteoarthritis.
The aetiology of lateral compartment osteoarthritis can be idiopathic,
usually affecting the femur, or secondary to trauma commonly affecting
the tibia. Surgical management of lateral compartment osteoarthritis
can include osteotomy, unicompartmental knee replacement and total
knee replacement. This review discusses the biomechanics, pathogenesis
and development of lateral compartment osteoarthritis and its management. Cite this article:
There are two techniques widely used to determine the rotational
alignment of the components in total knee arthroplasty (TKA); gap
balancing (GB) and measured resection (MR). Which technique is the
best remains controversial. We aimed to investigate this in a systematic
review and meta-analysis. In accordance with the methods of Cochrane, databases were searched
for all randomised controlled trials in the literature between January
1986 and June 2015 comparing radiographic and clinical outcomes
between the use of these two tecniques. Meta-analysis involved the
use of the Revman5.3 software provided by Cochrane collaboration.Aims
Materials and Methods
Since redesign of the Oxford phase III mobile-bearing unicompartmental
knee arthroplasty (UKA) femoral component to a twin-peg design,
there has not been a direct comparison to total knee arthroplasty
(TKA). Thus, we explored differences between the two cohorts. A total of 168 patients (201 knees) underwent medial UKA with
the Oxford Partial Knee Twin-Peg. These patients were compared with
a randomly selected group of 177 patients (189 knees) with primary
Vanguard TKA. Patient demographics, Knee Society (KS) scores and
range of movement (ROM) were compared between the two cohorts. Additionally,
revision, re-operation and manipulation under anaesthesia rates
were analysed.Aims
Patients and Methods
The purpose of the present study was to examine the long-term
fixation of a cemented fixed-bearing polished titanium tibial baseplate
(Genesis ll). Patients enrolled in a previous two-year prospective trial (n
= 35) were recalled at ten years. Available patients (n = 15) underwent
radiostereometric analysis (RSA) imaging in a supine position using
a conventional RSA protocol. Migration of the tibial component in
all planes was compared between initial and ten-year follow-up.
Outcome scores including the Knee Society Score, Western Ontario
and McMaster Universities Arthritis Index, 12-item Short Form Health
Survey, Forgotten Joint Score, and University of California, Los
Angeles Activity Score were recorded.Aims
Patients and Methods
This prospective study describes the complications and survival of the first 688 Phase 3 Oxford medial unicompartmental knee replacements implanted using a minimally-invasive technique by two surgeons and followed up independently. None was lost to follow-up. We had carried out 132 of the procedures more than five years ago. The clinical assessment of 101 of these which were available for review at five years is also presented. Nine of the 688 knees were revised: four for infection, three for dislocation of the bearing and two for unexplained pain. A further seven knees (1%) required other procedures: four had a manipulation under anaesthesia, two an arthroscopy and one a debridement for superficial infection. The survival rate at seven years was 97.3% (95% confidence interval 5.3). At five years, 96% of the patients had a good or excellent American Knee Society score, the mean Oxford knee score was 39 and the mean flexion was 133°. This study demonstrates that the minimally-invasive Oxford unicompartmental knee replacement is a reliable and effective procedure.
This prospective randomised controlled double-blind
trial compared two types of PFC Sigma total knee replacement (TKR),
differing in three design features aimed at improving flexion. The
outcome of a standard fixed-bearing posterior cruciate ligament-preserving
design (FB-S) was compared with that of a high-flexion rotating-platform
posterior-stabilised design (RP-F) at one year after TKR. The study group of 77 patients with osteoarthritis of the knee
comprised 37 men and 40 women, with a mean age of 69 years (44.9
to 84.9). The patients were randomly allocated either to the FB-S
or the RP-F group and assessed pre-operatively and at one year post-operatively.
The mean post-operative non-weight-bearing flexion was 107° (95%
confidence interval (CI) 104° to 110°)) for the FB-S group and 113°
(95% CI 109° to 117°) for the RP-F group, and this difference was
statistically significant (p = 0.032). However, weight-bearing range
of movement during both level walking and ascending a slope as measured
during flexible electrogoniometry was a mean of 4° lower in the RP-F
group than in the FB-S group, with 58° (95% CI 56° to 60°) Although the RP-F group achieved higher non-weight-bearing knee
flexion, patients in this group did not use this during activities
of daily living and reported more pain one year after surgery
There is a large amount of evidence available
about the relative merits of unicompartmental and total knee arthroplasty
(UKA and TKA). Based on the same evidence, different people draw
different conclusions and as a result, there is great variability
in the usage of UKA. The revision rate of UKA is much higher than TKA and so some
surgeons conclude that UKA should not be performed. Other surgeons
believe that the main reason for the high revision rate is that
UKA is easy to revise and, therefore, the threshold for revision
is low. They also believe that UKA has many advantages over TKA
such as a faster recovery, lower morbidity and mortality and better
function. They therefore conclude that UKA should be undertaken
whenever appropriate. The solution to this argument is to minimise the revision rate
of UKA, thereby addressing the main disadvantage of UKA. The evidence
suggests that this will be achieved if surgeons use UKA for at least
20% of their knee arthroplasties and use implants that are appropriate
for these broad indications. Cite this article:
Mobile-bearing (MB) total knee replacement (TKR)
was introduced to reduce the risk of aseptic loosening and wear of
polyethylene inserts. However, no consistent clinical advantages
of mobile- over fixed-bearing (FB) TKR have been found. In this
study we evaluated whether mobile bearings have an advantage over
fixed bearings with regard to revision rates and clinical outcome
scores. Furthermore, we determined which modifying variables affected
the outcome. A systematic search of the literature was conducted to collect
clinical trials comparing MB and FB in primary TKR. The primary
outcomes were revision rates for any reason, aseptic loosening and
wear. Secondary outcomes included range of movement, Knee Society
score (KSS), Oxford knee score (OKS), Short-Form 12 (SF-12) score
and radiological parameters. Meta-regression techniques were used
to explore factors modifying the observed effect. Our search yielded 1827 publications, of which 41 studies met
our inclusion criteria, comprising over 6000 TKRs. Meta-analyses
showed no clinically relevant differences in terms of revision rates,
clinical outcome scores or patient-reported outcome measures between
MB and FB TKRs. It appears that theoretical assumptions of superiority
of MB over FB TKR are not borne out in clinical practice. Cite this article:
Unicompartmental knee arthroplasty (UKA) has
advantages over total knee arthroplasty but national joint registries report
a significantly higher revision rate for UKA. As a result, most
surgeons are highly selective, offering UKA only to a small proportion
(up to 5%) of patients requiring arthroplasty of the knee, and consequently
performing few each year. However, surgeons with large UKA practices
have the lowest rates of revision. The overall size of the practice
is often beyond the surgeon’s control, therefore case volume may
only be increased by broadening the indications for surgery, and
offering UKA to a greater proportion of patients requiring arthroplasty
of the knee. The aim of this study was to determine the optimal UKA usage
(defined as the percentage of knee arthroplasty practice comprised
by UKA) to minimise the rate of revision in a sample of 41 986 records
from the for National Joint Registry for England and Wales (NJR). UKA usage has a complex, non-linear relationship with the rate
of revision. Acceptable results are achieved with the use of 20%
or more. Optimal results are achieved with usage between 40% and
60%. Surgeons with the lowest usage (up to 5%) have the highest
rates of revision. With optimal usage, using the most commonly used
implant, five-year survival is 96% (95% confidence interval (CI)
94.9 to 96.0), compared with 90% (95% CI 88.4 to 91.6) with low
usage (5%) previously considered ideal. The rate of revision of UKA is highest with low usage, implying
the use of narrow, and perhaps inappropriate, indications. The widespread
use of broad indications, using appropriate implants, would give
patients the advantages of UKA, without the high rate of revision. Cite this article:
The contraindications for unicompartmental knee replacement (UKR) remain controversial. The views of many surgeons are based on Kozinn and Scott’s 1989 publication which stated that patients who weighed more than 82 kg, were younger than 60 years, undertook heavy labour, had exposed bone in the patellofemoral joint or chondrocalcinosis, were not ideal candidates for UKR. Our aim was to determine whether these potential contraindications should apply to patients with a mobile-bearing UKR. In order to do this the outcome of patients with these potential contraindications was compared with that of patients without the contraindications in a prospective series of 1000 UKRs. The outcome was assessed using the Oxford knee score, the American Knee Society score, the Tegner activity score, revision rate and survival. The clinical outcome of patients with each of the potential contraindications was similar to or better than those without each contraindication. Overall, 678 UKRs (68%) were performed in patients who had at least one potential contraindication and only 322 (32%) in patients deemed to be ideal. The survival at ten years was 97.0% (95% confidence interval 93.4 to 100.0) for those with potential contraindications and 93.6% (95% confidence interval 87.2 to 100.0) in the ideal patients. We conclude that the thresholds proposed by Kozinn and Scott using weight, age, activity, the state of the patellofemoral joint and chondrocalcinosis should not be considered to be contraindications for the use of the Oxford UKR.
The purpose of this study was to undertake a
meta-analysis to determine whether there is lower polyethylene wear and
longer survival when using mobile-bearing implants in total knee
replacement when compared with fixed-bearing implants. Of 975 papers
identified, 34 trials were eligible for data extraction and meta-analysis
comprising 4754 patients (6861 knees). We found no statistically
significant differences between the two designs in terms of the incidence
of radiolucent lines, osteolysis, aseptic loosening or survival.
There is thus currently no evidence to suggest that the use of mobile-bearing
designs reduce polyethylene wear and prolong survival after total
knee replacement. Cite this article:
We measured the contact areas and contact stresses at the post-cam mechanism of a posterior-stabilised total knee arthroplasty when a posterior force of 500 N was applied to the Kirschner Performance, Scorpio Superflex, NexGen LPS Flex Fixed, and NexGen LPS Flex Mobile knee systems. Measurements were made at 90°, 120°, and 150° of flexion both in neutral rotation and 10° of internal rotation of the tibial component. Peak contact stresses at 90°, 120°, and 150° were 24.0, 33.9, and 28.8 MPa, respectively, for the Kirschner; 26.0, 32.4, and 22.1 MPa, respectively, for the Scorpio; and 34.1, 31.5, and 32.5 MPa, respectively, for the NexGen LPS Flex Fixed. With an internally rotated tibia, the contact stress increased significantly with all the fixed-bearing arthroplasties but not with the NexGen LPS Flex Mobile arthroplasty. The post-cam design should be modified in order to provide a larger contact area whilst avoiding any impingement and edge loading.
The most common reasons for revision of unicompartmental
knee arthroplasty (UKA) are loosening and pain. Cementless components
may reduce the revision rate. The aim of this study was to compare
the fixation and clinical outcome of cementless and cemented Oxford
UKAs. A total of 43 patients were randomised to receive either a cemented
or a cementless Oxford UKA and were followed for two years with
radiostereometric analysis (RSA), radiographs aligned with the bone–implant
interfaces and clinical scores. The femoral components migrated significantly during the first
year (mean 0.2 mm) but not during the second. There was no significant
difference in the extent of migration between cemented and cementless
femoral components in either the first or the second year. In the
first year the cementless tibial components subsided significantly
more than the cemented components (mean 0.28 mm ( As second-year migration is predictive of subsequent loosening,
and as radiolucency is suggestive of reduced implant–bone contact,
these data suggest that fixation of the cementless components is
at least as good as, if not better than, that of cemented devices. Cite this article:
The cementless Oxford unicompartmental knee replacement
has been demonstrated to have superior fixation on radiographs and
a similar early complication rate compared with the cemented version.
However, a small number of cases have come to our attention where,
after an apparently successful procedure, the tibial component subsides into
a valgus position with an increased posterior slope, before becoming
well-fixed. We present the clinical and radiological findings of
these six patients and describe their natural history and the likely
causes. Two underwent revision in the early post-operative period,
and in four the implant stabilised and became well-fixed radiologically with
a good functional outcome. This situation appears to be avoidable by minor modifications
to the operative technique, and it appears that it can be treated
conservatively in most patients. Cite this article:
Mobile-bearing unicompartmental knee replacements
(UKRs) with a flat tibial plateau have not performed well in the
lateral compartment, owing to a high dislocation rate. This led
to the development of the Domed Lateral Oxford UKR (Domed OUKR)
with a biconcave bearing. The aim of this study was to assess the
survival and clinical outcomes of the Domed OUKR in a large patient
cohort in the medium term. We prospectively evaluated 265 consecutive knees with isolated
disease of the lateral compartment and a mean age at surgery of
64 years (32 to 90). At a mean follow-up of four years ( The Domed Lateral OUKR gives good clinical outcomes, low re-operation
and revision rates and a low dislocation rate in patients with isolated
lateral compartmental disease, in the hands of the designer surgeons. Cite this article:
Bone loss in the proximal tibia and distal femur
is frequently encountered in revision knee replacement surgery.
The various options for dealing with this depend on the extent of
any bone loss. We present our results with the use of cementless
metaphyseal metal sleeves in 103 patients (104 knees) with a mean
follow-up of 43 months (30 to 65). At final follow-up, sleeves in
102 knees had good osseointegration. Two tibial sleeves were revised
for loosening, possibly due to infection. The average pre-operative Oxford Knee Score was 23 (11 to 36)
and this improved to 32 (15 to 46) post-operatively. These early
results encourage us to continue with the technique and monitor
the outcomes in the long term. Cite this article: