Advertisement for orthosearch.org.uk
Results 1 - 26 of 26
Results per page:
The Bone & Joint Journal
Vol. 96-B, Issue 2 | Pages 263 - 269
1 Feb 2014
Batta V Coathup MJ Parratt MT Pollock RC Aston WJ Cannon SR Skinner JA Briggs TW Blunn GW

We reviewed the outcome of 69 uncemented, custom-made, distal femoral endoprosthetic replacements performed in 69 patients between 1994 and 2006. There were 31 women and 38 men with a mean age at implantation of 16.5 years (5 to 37). All procedures were performed for primary malignant bone tumours of the distal femur. At a mean follow-up of 124.2 months (4 to 212), 53 patients were alive, with one patient lost to follow-up. All nine implants (13.0%) were revised due to aseptic loosening at a mean of 52 months (8 to 91); three implants (4.3%) were revised due to fracture of the shaft of the prosthesis and three patients (4.3%) had a peri-prosthetic fracture. Bone remodelling associated with periosteal cortical thinning adjacent to the uncemented intramedullary stem was seen in 24 patients but this did not predispose to failure. All aseptically loose implants in this series were diagnosed to be loose within the first five years. The results from this study suggest that custom-made uncemented distal femur replacements have a higher rate of aseptic loosening compared to published results for this design when used with cemented fixation. Loosening of uncemented replacements occurs early indicating that initial fixation of the implant is crucial. Cite this article: Bone Joint J 2014;96-B:263–9


The Bone & Joint Journal
Vol. 101-B, Issue 10 | Pages 1248 - 1255
1 Oct 2019
Pineda A Pabbruwe MB Kop AM Vlaskovsky P Hurworth M

Aims. The aim of this study was to conduct the largest low contact stress (LCS) retrieval study to elucidate the failure mechanisms of the Porocoat and Duofix femoral component. The latter design was voluntarily recalled by the manufacturer. Materials and Methods. Uncemented LCS explants were divided into three groups: Duofix, Porocoat, and mixed. Demographics, polyethylene wear, tissue ingrowth, and metallurgical analyses were performed. Results. In 104 implants, a decrease in the odds of loosening and an increase in metallosis and tissue staining in the Duofix group relative to Porocoat group was detected (p = 0.028). There was an increased presence of embedded metallic debris in the Duofix group (p < 0.001). Decreased tissue ingrowth was associated with the Duofix surface (p < 0.001). The attached beads had reduced microhardness, indicative of adverse thermal processing, which resulted in bead shedding, particulate debris, and metallosis. Conclusion. Hydroxyapatite coating of the LCS femoral component produced unexpected results and led to its recall. The root cause was likely a combination of retained alumina grit and a reduction in bead microhardness (mechanical strength) resulting in increased particle debris, metallosis, and early revision. The Duofix LCS femoral component was not equivalent to the Porocoat version despite its approval through the Food and Drug Administration (FDA) 510(k) equivalance approval process. Regulation of the introduction of modified existing devices needs to be improved and the Duofix LCS should have been considered to be a new device for which equivalence had not been demonstrated at the point of introduction. Cite this article: Bone Joint J 2019;101-B:1248–1255


Bone & Joint Research
Vol. 5, Issue 7 | Pages 307 - 313
1 Jul 2016
Sandgren B Skorpil M Nowik P Olivecrona H Crafoord J Weidenhielm L Persson A

Objectives. Computed tomography (CT) plays an important role in evaluating wear and periacetabular osteolysis (PAO) in total hip replacements. One concern with CT is the high radiation exposure since standard pelvic CT provides approximately 3.5 millisieverts (mSv) of radiation exposure, whereas a planar radiographic examination with three projections totals approximately 0.5 mSv. The objective of this study was to evaluate the lowest acceptable radiation dose for dual-energy CT (DECT) images when measuring wear and periacetabular osteolysis in uncemented metal components. Materials and Methods. A porcine pelvis with bilateral uncemented hip prostheses and with known linear wear and acetabular bone defects was examined in a third-generation multidetector DECT scanner. The examinations were performed with four different radiation levels both with and without iterative reconstruction techniques. From the high and low peak kilo voltage acquisitions, polychrmoatic images were created together with virtual monochromatic images of energies 100 kiloelectron volts (keV) and 150 keV. Results. We could assess wear and PAO while substantially lowering the effective radiation dose to 0.7 mSv for a total pelvic view with an accuracy of around 0.5 mm for linear wear and 2 mm to 3 mm for PAO. Conclusion. CT for detection of prosthetic wear and PAO could be used with clinically acceptable accuracy at a radiation exposure level equal to plain radiographic exposures. Cite this article: B. Sandgren, M. Skorpil, P. Nowik, H. Olivecrona, J. Crafoord, L. Weidenhielm, A. Persson. Assessment of wear and periacetabular osteolysis using dual energy computed tomography on a pig cadaver to identify the lowest acceptable radiation dose. Bone Joint Res 2016;5:307–313. DOI: 10.1302/2046-3758.57.2000566


Bone & Joint Open
Vol. 3, Issue 12 | Pages 977 - 990
23 Dec 2022
Latijnhouwers D Pedersen A Kristiansen E Cannegieter S Schreurs BW van den Hout W Nelissen R Gademan M

Aims

This study aimed to investigate the estimated change in primary and revision arthroplasty rate in the Netherlands and Denmark for hips, knees, and shoulders during the COVID-19 pandemic in 2020 (COVID-period). Additional points of focus included the comparison of patient characteristics and hospital type (2019 vs COVID-period), and the estimated loss of quality-adjusted life years (QALYs) and impact on waiting lists.

Methods

All hip, knee, and shoulder arthroplasties (2014 to 2020) from the Dutch Arthroplasty Register, and hip and knee arthroplasties from the Danish Hip and Knee Arthroplasty Registries, were included. The expected number of arthroplasties per month in 2020 was estimated using Poisson regression, taking into account changes in age and sex distribution of the general Dutch/Danish population over time, calculating observed/expected (O/E) ratios. Country-specific proportions of patient characteristics and hospital type were calculated per indication category (osteoarthritis/other elective/acute). Waiting list outcomes including QALYs were estimated by modelling virtual waiting lists including 0%, 5% and 10% extra capacity.


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 888 - 894
1 Aug 2023
Murray J Jeyapalan R Davies M Sheehan C Petrie M Harrison T

Aims

Total femoral arthroplasty (TFA) is a rare procedure used in cases of significant femoral bone loss, commonly from cancer, infection, and trauma. Low patient numbers have resulted in limited published work on long-term outcomes, and even less regarding TFA undertaken for non-oncological indications. The aim of this study was to evaluate the long-term clinical outcomes of all TFAs in our unit.

Methods

Data were collected retrospectively from a large tertiary referral revision arthroplasty unit’s database. Inclusion criteria included all patients who underwent TFA in our unit. Preoperative demographics, operative factors, and short- and long-term outcomes were collected for analysis. Outcome was defined using the Musculoskeletal Infection Society (MSIS) outcome reporting tool.


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 649 - 656
1 Jun 2023
Dagneaux L Amundson AW Larson DR Pagnano MW Berry DJ Abdel MP

Aims

Nonagenarians (aged 90 to 99 years) have experienced the fastest percent decile population growth in the USA recently, with a consequent increase in the prevalence of nonagenarians living with joint arthroplasties. As such, the number of revision total hip arthroplasties (THAs) and total knee arthroplasties (TKAs) in nonagenarians is expected to increase. We aimed to determine the mortality rate, implant survivorship, and complications of nonagenarians undergoing aseptic revision THAs and revision TKAs.

Methods

Our institutional total joint registry was used to identify 96 nonagenarians who underwent 97 aseptic revisions (78 hips and 19 knees) between 1997 and 2018. The most common indications were aseptic loosening and periprosthetic fracture for both revision THAs and revision TKAs. Mean age at revision was 92 years (90 to 98), mean BMI was 27 kg/m2 (16 to 47), and 67% (n = 65) were female. Mean time between primary and revision was 18 years (SD 9). Kaplan-Meier survival was used for patient mortality, and compared to age- and sex-matched control populations. Reoperation risk was assessed using cumulative incidence with death as a competing risk. Mean follow-up was five years.


The Bone & Joint Journal
Vol. 106-B, Issue 4 | Pages 365 - 371
1 Apr 2024
Ledford CK Shirley MB Spangehl MJ Berry DJ Abdel MP

Aims

Breast cancer survivors have known risk factors that might influence the results of total hip arthroplasty (THA) or total knee arthroplasty (TKA). This study evaluated clinical outcomes of patients with breast cancer history after primary THA and TKA.

Methods

Our total joint registry identified patients with breast cancer history undergoing primary THA (n = 423) and TKA (n = 540). Patients were matched 1:1 based upon age, sex, BMI, procedure (hip or knee), and surgical year to non-breast cancer controls. Mortality, implant survival, and complications were assessed via Kaplan-Meier methods. Clinical outcomes were evaluated via Harris Hip Scores (HHSs) or Knee Society Scores (KSSs). Mean follow-up was six years (2 to 15).


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 284 - 293
1 Mar 2023
Li Y Zhang X Ji B Wulamu W Yushan N Guo X Cao L

Aims

Gram-negative periprosthetic joint infection (PJI) has been poorly studied despite its rapidly increasing incidence. Treatment with one-stage revision using intra-articular (IA) infusion of antibiotics may offer a reasonable alternative with a distinct advantage of providing a means of delivering the drug in high concentrations. Carbapenems are regarded as the last line of defense against severe Gram-negative or polymicrobial infection. This study presents the results of one-stage revision using intra-articular carbapenem infusion for treating Gram-negative PJI, and analyzes the characteristics of bacteria distribution and drug sensitivity.

Methods

We retrospectively reviewed 32 patients (22 hips and 11 knees) who underwent single-stage revision combined with IA carbapenem infusion between November 2013 and March 2020. The IA and intravenous (IV) carbapenem infusions were administered for a single Gram-negative infection, and IV vancomycin combined with IA carbapenems and vancomycin was applied for polymicrobial infection including Gram-negative bacteria. The bacterial community distribution, drug sensitivity, infection control rate, functional recovery, and complications were evaluated. Reinfection or death caused by PJI was regarded as a treatment failure.


The Bone & Joint Journal
Vol. 103-B, Issue 3 | Pages 522 - 529
1 Mar 2021
Nichol T Callaghan J Townsend R Stockley I Hatton PV Le Maitre C Smith TJ Akid R

Aims

The aim of this study was to develop a single-layer hybrid organic-inorganic sol-gel coating that is capable of a controlled antibiotic release for cementless hydroxyapatite (HA)-coated titanium orthopaedic prostheses.

Methods

Coatings containing gentamicin at a concentration of 1.25% weight/volume (wt/vol), similar to that found in commercially available antibiotic-loaded bone cement, were prepared and tested in the laboratory for: kinetics of antibiotic release; activity against planktonic and biofilm bacterial cultures; biocompatibility with cultured mammalian cells; and physical bonding to the material (n = 3 in all tests). The sol-gel coatings and controls were then tested in vivo in a small animal healing model (four materials tested; n = 6 per material), and applied to the surface of commercially pure HA-coated titanium rods.


Bone & Joint Open
Vol. 1, Issue 12 | Pages 743 - 748
1 Dec 2020
Mahon J McCarthy CJ Sheridan GA Cashman JP O'Byrne JM Kenny P

Aims

The Exeter V40 cemented femoral stem was first introduced in 2000. The largest single-centre analysis of this implant to date was published in 2018 by Westerman et al. Excellent results were reported at a minimum of ten years for the first 540 cases performed at the designer centre in the Exeter NHS Trust, with stem survivorship of 96.8%. The aim of this current study is to report long-term outcomes and survivorship for the Exeter V40 stem in a non-designer centre.

Methods

All patients undergoing primary total hip arthroplasty using the Exeter V40 femoral stem between 1 January 2005 and 31 January 2010 were eligible for inclusion. Data were collected prospectively, with routine follow-up at six to 12 months, two years, five years, and ten years. Functional outcomes were assessed using Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores. Outcome measures included data on all components in situ beyond ten years, death occurring within ten years with components in situ, and all-cause revision surgery.


The Bone & Joint Journal
Vol. 102-B, Issue 10 | Pages 1341 - 1348
3 Oct 2020
Scholten R Hannink G Willemsen K Mascini EM Somford MP Schreurs BW van Susante JLC

Aims

Preoperative nasal Staphylococcus aureus screening and eradication reduces surgical site infections (SSIs) but its impact on reducing early prosthetic joint infection (PJI) remains controversial. This study aims to assess the effect of preoperative nasal S. aureus screening and eradication on the incidence of early PJI in general and S. aureus-induced early PJI.

Methods

All primary total hip arthroplasties (THA) and total knee arthroplasties (TKA) performed from January 2006 to April 2018 were retrospectively reviewed for the incidence of early PJI. Demographic parameters, risk factors for PJI (American Society of Anaesthesiologists classification, body mass index, smoking status, and diabetes mellitus) and implant types were collected. A preoperative screening and eradication protocol for nasal colonization of S. aureus was introduced in October 2010. The incidence of early PJI was compared before and after the implementation of the protocol. Missing data were imputed via multiple imputation by chained equations. Inverse probability weighting was used to account for differences between patients in both groups. Weighted univariate logistic regression was used to evaluate the incidence of early PJI for both groups.


The Bone & Joint Journal
Vol. 101-B, Issue 8 | Pages 951 - 959
1 Aug 2019
Preston N McHugh GA Hensor EMA Grainger AJ O’Connor PJ Conaghan PG Stone MH Kingsbury SR

Aims

This study aimed to develop a virtual clinic for the purpose of reducing face-to-face orthopaedic consultations.

Patients and Methods

Anonymized experts (hip and knee arthroplasty patients, surgeons, physiotherapists, radiologists, and arthroplasty practitioners) gave feedback via a Delphi Consensus Technique. This consisted of an iterative sequence of online surveys, during which virtual documents, made up of a patient-reported questionnaire, standardized radiology report, and decision-guiding algorithm, were modified until consensus was achieved. We tested the patient-reported questionnaire on seven patients in orthopaedic clinics using a ‘think-aloud’ process to capture difficulties with its completion.


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 3 | Pages 431 - 433
1 May 1996
Niskanen RO Korkala O Pammo H

Any operation induces an elevation in the level of serum C-reactive protein (CRP). After hip and knee arthroplasty the maximal values are seen on the second and third postoperative days, after which the CRP decreases rapidly. There is no difference between patients with cemented or uncemented prostheses. Major postoperative complications may cause a further increase in CRP levels at one and two weeks


The Journal of Bone & Joint Surgery British Volume
Vol. 82-B, Issue 1 | Pages 103 - 107
1 Jan 2000
Pekkarinen J Alho A Lepistö J Ylikoski M Ylinen P Paavilainen T

We have reviewed retrospectively 68 revisions of the femoral component in arthroplasties of the hip in 65 patients, using impaction bone grafting, at a median of three years (1 month to 6 years). We employed the cemented Exeter X-Change technique in 36 patients and the uncemented Bi-Metric allografting method in 32. The 37 bone defects were grade 3 or grade 4 on the Endo-Klinik classification. The Mayo hip score improved from a mean of 32 (. sd. ± 18) to 62 (. sd. ± 15). Most (25) of the 34 complications occurred in grade-3 and grade-4 defects; nine were intraoperative diaphyseal fractures and eight fractures of the greater trochanter. All the fractures united. The risk of intraoperative fracture was prevented by supporting the bone with wires in 16 hips, with reinforcement mesh in 18 and by a plate in six. Early migration of the stem of more than 10 mm during the first year indicated rotational instability; it occurred in three cases. In difficult revision cases with large defects of the femoral bone, bone-impaction techniques carry a high risk of complications


The Journal of Bone & Joint Surgery British Volume
Vol. 79-B, Issue 3 | Pages 361 - 365
1 May 1997
Shih C Lee P Chen J Tai C Chen L Wu JS Chang WH

We made a clinical study of polyethylene wear in 240 hips of 187 patients having primary total hip arthroplasties from 1989 to 1990, using uncemented Osteonics components, with a head size of 26 mm. We excluded cups with anteversion of over 20° and measured linear wear by a new method using a digitiser and special software of our design. Follow-up was from two to five years (mean 4.3). The mean age at operation was 50.3 years, with more men than women (1.4:1). The mean linear wear per year was 0.15 mm; this did not increase with the longevity of the prosthesis (p = 0.54). In 59 hips showing evidence of osteolysis, the mean linear wear rate was significantly higher at 0.23 mm/year (p < 0.001). The mean linear wear rate also correlated significantly with age at the time of operation (p = 0.008), but we found no significant correlations with body-weight, gender, aetiology of the disease, thickness of polyethylene, or cup position. Our new method of measurement is time-saving and reproducible. The results confirm the greater rate of linear wear of polyethylene in patients showing osteolysis and in those who are younger


The Bone & Joint Journal
Vol. 99-B, Issue 12 | Pages 1618 - 1628
1 Dec 2017
Hunt LP Blom A Wilkinson JM

Aims

To investigate whether elective joint arthroplasty performed at the weekend is associated with a different 30-day mortality versus that performed between Monday and Friday.

Patients and Methods

We examined the 30-day cumulative mortality rate (Kaplan-Meier) for all elective hip and knee arthroplasties performed in England and Wales between 1st April 2003 and 31st December 2014, comprising 118 096 episodes undertaken at the weekend and 1 233 882 episodes performed on a weekday. We used Cox proportional-hazards regression models to assess for time-dependent variation and adjusted for identified risk factors for mortality.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1592 - 1599
1 Dec 2018
Galea VP Connelly JW Shareghi B Kärrholm J Sköldenberg O Salemyr M Laursen MB Muratoglu O Bragdon C Malchau H

Aims

The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure.

Patients and Methods

A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery.


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 921 - 926
1 Jul 2017
Märdian S Perka C Schaser K Gruner J Scheel F Schwabe P

Aims

Periprosthetic fracture is a significant complication of total hip and knee arthroplasty. This study aimed to describe the survival of patients sustaining periprosthetic femoral fractures and compare this with that of the general population, as well as to identify the factors that influence survival.

Patients and Methods

A total of 151 patients (women: men 116:35, mean age 74.6 years, standard deviation 11.5) that sustained a periprosthetic fracture between January 2005 and October 2012 were retrospectively analysed. Epidemiological data, comorbidities, type of surgical management, type of implant, and mortality data were studied.


Bone & Joint Research
Vol. 5, Issue 4 | Pages 122 - 129
1 Apr 2016
Small SR Rogge RD Malinzak RA Reyes EM Cook PL Farley KA Ritter MA

Objectives

Initial stability of tibial trays is crucial for long-term success of total knee arthroplasty (TKA) in both primary and revision settings. Rotating platform (RP) designs reduce torque transfer at the tibiofemoral interface. We asked if this reduced torque transfer in RP designs resulted in subsequently reduced micromotion at the cemented fixation interface between the prosthesis component and the adjacent bone.

Methods

Composite tibias were implanted with fixed and RP primary and revision tibial trays and biomechanically tested under up to 2.5 kN of axial compression and 10° of external femoral component rotation. Relative micromotion between the implanted tibial tray and the neighbouring bone was quantified using high-precision digital image correlation techniques.


The Bone & Joint Journal
Vol. 96-B, Issue 12 | Pages 1663 - 1668
1 Dec 2014
Bottle A Aylin P Loeffler M

The aim of this study was to define return to theatre (RTT) rates for elective hip and knee replacement (HR and KR), to describe the predictors and to show the variations in risk-adjusted rates by surgical team and hospital using national English hospital administrative data.

We examined information on 260 206 HRs and 315 249 KRs undertaken between April 2007 and March 2012. The 90-day RTT rates were 2.1% for HR and 1.8% for KR. Male gender, obesity, diabetes and several other comorbidities were associated with higher odds for both index procedures. For HR, hip resurfacing had half the odds of cement fixation (OR = 0.58, 95% confidence intervals (CI) 0.47 to 0.71). For KR, unicondylar KR had half the odds of total replacement (OR = 0.49, 95% CI 0.42 to 0.56), and younger ages had higher odds (OR = 2.23, 95% CI 1.65 to 3.01) for ages < 40 years compared with ages 60 to 69 years). There were more funnel plot outliers at three standard deviations than would be expected if variation occurred on a random basis.

Hierarchical modelling showed that three-quarters of the variation between surgeons for HR and over half the variation between surgeons for KR are not explained by the hospital they operated at or by available patient factors. We conclude that 90-day RTT rate may be a useful quality indicator for orthopaedics.

Cite this article: Bone Joint J 2014; 96-B:1663–8.


The Bone & Joint Journal
Vol. 96-B, Issue 4 | Pages 479 - 485
1 Apr 2014
Pedersen AB Mehnert F Sorensen HT Emmeluth C Overgaard S Johnsen SP

We examined the risk of thrombotic and major bleeding events in patients undergoing total hip and knee replacement (THR and TKR) treated with thromboprophylaxis, using nationwide population-based databases. We identified 83 756 primary procedures performed between 1997 and 2011. The outcomes were symptomatic venous thromboembolism (VTE), myocardial infarction (MI), stroke, death and major bleeding requiring hospitalisation within 90 days of surgery.

A total of 1114 (1.3%) and 483 (0.6%) patients experienced VTE and bleeding, respectively. The annual risk of VTE varied between 0.9% and 1.6%, and of bleeding between 0.4% and 0.8%. The risk of VTE and bleeding was unchanged over a 15-year period. A total of 0.7% of patients died within 90 days, with a decrease from 1% in 1997 to 0.6% in 2011 (p < 0.001). A high level of comorbidity and general anaesthesia were strong risk factors for both VTE and bleeding, with no difference between THR and TKR patients. The risk of both MI and stroke was 0.5%, which remained unchanged during the study period.

In this cohort study of patients undergoing THR and TKR patients in routine clinical practice, approximately 3% experienced VTE, MI, stroke or bleeding. These risks did not decline during the 15-year study period, but the risk of dying fell substantially.

Cite this article: Bone Joint J 2014;96-B:479–85.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 9 | Pages 1217 - 1222
1 Sep 2009
King RJ Makrides P Gill JA Karthikeyan S Krikler SJ Griffin DR

We have developed a novel method of calculating the radiological magnification of the hip using two separate radio-opaque markers. We recruited 74 patients undergoing radiological assessment following total hip replacement. Both the new double marker and a conventional single marker were used by the radiographer at the time of x-ray. The predicted magnification according to each marker was calculated, as was the true radiological magnification of the components. The correlation between true and predicted magnification was good using the double marker (r = 0.90, n = 74, p < 0.001), but only moderate for the single marker (r = 0.50, n = 63, p < 0.001). The median error was significantly less for the double marker than for the single (1.1% vs 4.8%, p < 0.001). The double marker method demonstrated excellent validity (intraclass correlation coefficient = 0.89), in contrast to the single marker (0.32).

The double marker method appears to be superior to the single marker method when used in the clinical environment.


The Bone & Joint Journal
Vol. 96-B, Issue 6 | Pages 765 - 771
1 Jun 2014
Thomassen BJW den Hollander PHC Kaptijn HH Nelissen RGHH Pilot P

We hypothesised there was no clinical value in using an autologous blood transfusion (ABT) drain in either primary total hip (THR) or total knee replacement (TKR) in terms of limiting allogeneic blood transfusions when a modern restrictive blood management regime was followed. A total of 575 patients (65.2% men), with a mean age of 68.9 years (36 to 94) were randomised in this three-arm study to no drainage (group A), or to wound drainage with an ABT drain for either six hours (group B) or 24 hours (group C). The primary outcome was the number of patients receiving allogeneic blood transfusion. Secondary outcomes were post-operative haemoglobin (Hb) levels, length of hospital stay and adverse events.

This study identified only 41 transfused patients, with no significant difference in distribution between the three groups (p = 0.857). The mean pre-operative haemoglobin (Hb) value in the transfused group was 12.8 g/dL (9.8 to 15.5) versus 14.3 g/dL (10.6 to 18.0) in the non-transfused group (p < 0.001, 95% confidence interval: 1.08 to 1.86). Post-operatively, the median of re-transfused shed blood in patients with a THR was 280 mL (Interquartile range (IQR) 150 to 400) and in TKR patients 500 mL (IQR 350 to 650) (p <  0.001). ABT drains had no effect on the proportion of transfused patients in primary THR and TKR. The secondary outcomes were also comparable between groups.

Cite this article: Bone Joint J 2014;96-B:765–71.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 1 | Pages 123 - 129
1 Jan 2010
Jameson SS Bottle A Malviya A Muller SD Reed MR

The National Institute for Clinical Excellence (NICE) produces recommendations on appropriate treatment within the National Health Service (NHS) in England and Wales. The NICE guidelines on prophylaxis for venous thromboembolism in orthopaedic surgery recommend that all patients be offered a low molecular weight heparin (LMWH). The linked hospital episode statistics of 219 602 patients were examined to determine the rates of complications following lower limb arthroplasty for the 12-month periods prior to and following the publication of these guidelines. These were compared with data from the National Joint Registry (England and Wales) regarding the use of LMWH during the same periods. There was a significant increase in the reported use of LMWH (59.5% to 67.6%, p < 0.001) following the publication of the guidelines. However, the 90-day venous thromboembolism events actually increased slightly following total hip replacement (THR, 1.69% to 1.84%, p = 0.06) and remained unchanged following total knee replacement (TKR, 1.99% to 2.04%). Return to theatre in the first 30 days for infection did not show significant changes. There was an increase in the number of patients diagnosed with thrombocytopenia, which was significant following THR (0.11% to 0.16%, p = 0.04). The recommendations from NICE are based on predicted reductions in venous thromboembolism events, reducing morbidity, mortality and costs to the NHS.

The early results in orthopaedic patients do not support these predictions, but do show an increase in complications.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 6 | Pages 776 - 783
1 Jun 2009
Rajesparan K Biant LC Ahmad M Field RE

Tranexamic acid is a fibrinolytic inhibitor which reduces blood loss in total knee replacement. We examined the effect on blood loss of a standardised intravenous bolus dose of 1 g of tranexamic acid, given at the induction of anaesthesia in patients undergoing total hip replacement and tested the potential prothrombotic effect by undertaking routine venography. In all, 36 patients received 1 g of tranexamic acid, and 37 no tranexamic acid. Blood loss was measured directly per-operatively and indirectly post-operatively.

Tranexamic acid reduced the early post-operative blood loss and total blood loss (p = 0.03 and p = 0.008, respectively) but not the intraoperative blood loss. The tranexamic acid group required fewer transfusions (p = 0.03) and had no increased incidence of deep-vein thrombosis. The reduction in early post-operative blood loss was more marked in women (p = 0.05), in whom this effect was dose-related (r = −0.793).

Our study showed that the administration of a standardised pre-operative bolus of 1 g of tranexamic acid was cost-effective in reducing the blood loss and transfusion requirements after total hip replacement, especially in women.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 8 | Pages 1079 - 1083
1 Aug 2008
Moonen AFCM Thomassen BJW Knoors NT van Os JJ Verburg AD Pilot P

This prospective randomised clinical trial evaluated the effect of alternatives for allogeneic blood transfusions after total hip replacement and total knee replacement in patients with pre-operative haemoglobin levels between 10.0 g/dl and 13.0 g/dl. A total of 100 patients were randomly allocated to the Eprex (pre-operative injections of epoetin) or Bellovac groups (post-operative retransfusion of shed blood). Allogeneic blood transfusions were administered according to hospital policy.

In the Eprex group, 4% of the patients (two patients) received at least one allogeneic blood transfusion. In the Bellovac group, where a mean 216 ml (0 to 700) shed blood was retransfused, 28% (14 patients) required the allogeneic transfusion (p = 0.002). When comparing Eprex with Bellovac in total hip replacement, the percentages were 7% (two of 30 patients) and 30% (nine of 30 patients) (p = 0.047) respectively, whereas in total knee replacement, the percentages were 0% (0 of 20 patients) and 25% (five of 20 patients) respectively (p = 0.042).

Pre-operative epoetin injections are more effective but more costly in reducing the need for allogeneic blood transfusions in mildly anaemic patients than post-operative retransfusion of autologous blood.