The aim of this prospective multicentre study
was to report the patient satisfaction after total knee replacement (TKR),
undertaken with the aid of intra-operative sensors, and to compare
these results with previous studies. A total of 135 patients undergoing
TKR were included in the study. The soft-tissue balance of each
TKR was quantified intra-operatively by the sensor, and 18 (13%)
were found to be unbalanced. A total of 113 patients (96.7%) in
the balanced group and 15 (82.1%) in the unbalanced group were satisfied
or very satisfied one year post-operatively (p = 0.043). A review of the literature identified no previous study with
a mean level of satisfaction that was greater than the reported
level of satisfaction of the balanced TKR group in this study. Ensuring
soft-tissue balance by using intra-operative sensors during TKR
may improve satisfaction. Cite this article:
This retrospective study evaluated the midterm clinical and radiographic outcomes of a second-generation total knee replacement system. In a multicentre consecutive series of 1512 patients, 1970 knees were treated with the PFC Sigma knee system (Depuy, Warsaw, Indiana). The patients were reviewed for functional outcome, and underwent independent radiographic evaluation at a mean follow-up of 7.3 years (5 to 10). A total of 40 knees (2%) required revision, 17 (0.9%) for infection. The incidence of osteolysis was 2.2%. The ten-year survival with revision for any cause other than infection as the endpoint was 97.2% (95% CI 95.4 to 99.1). The PFC Sigma knee system appears to provide excellent results in the medium term.