The aim of this study was to report the long-term follow-up of cemented short Exeter femoral components when used in primary total hip arthroplasty (THA). We included all primary 394 THAs with a cemented short Exeter femoral component (≤ 125 mm) used in our tertiary referral centre between October 1993 and December 2021. A total of 83 patients (21%) were male. The median age of the patients at the time of surgery was 42 years (interquartile range (IQR) 30 to 55). The main indication for THA was a childhood hip disease (202; 51%). The median follow-up was 6.7 years (IQR 3.1 to 11.0). Kaplan-Meier survival analyses were performed to determine the rates of survival with femoral revision for any indication, for septic loosening, for fracture of the femoral component and for aseptic loosening as endpoints. The indications for revision were evaluated. Fractures of the femoral component were described in detail.Aims
Methods
When fracture of an extensively porous-coated
femoral component occurs, its removal at revision total hip arthroplasty
(THA) may require a femoral osteotomy and the use of a trephine.
The remaining cortical bone after using the trephine may develop
thermally induced necrosis. A retrospective review identified 11
fractured, well-fixed, uncemented, extensively porous-coated femoral
components requiring removal using a trephine with a minimum of
two years of follow-up. The mean time to failure was 4.6 years (1.7 to 9.1, standard
deviation ( A total of four patients (36.4%) required further revision: three
for instability and one for fracture of the revision component.
There was no statistically significant difference in the mean Harris
hip score before implant fracture (82.4; These findings suggest that removal of a fractured, well-fixed,
uncemented, extensively porous-coated femoral component using a
trephine does not compromise subsequent fixation at revision THA
and the patient’s pre-operative level of function can be restored.
However, the loss of proximal bone stock before revision may be associated
with a high rate of dislocation post-operatively. Cite this article:
Aims. Cement-in-cement revision of the femoral component represents a widely practised technique for a variety of indications in revision total hip arthroplasty. In this study, we compare the clinical and radiological outcomes of two polished tapered femoral components. Methods. From our prospectively collated database, we identified all patients undergoing cement-in-cement revision from January 2005 to January 2013 who had a minimum of two years' follow-up. All cases were performed by the senior author using either an Exeter short revision stem or the C-Stem AMT high offset No. 1 prosthesis. Patients were followed-up annually with clinical and radiological assessment. Results. A total of 97 patients matched the inclusion criteria (50 Exeter and 47 C-Stem AMT components). There were no significant differences between the patient demographic data in either group. Mean follow-up was 9.7 years. A significant improvement in Oxford Hip Score (OHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), and 12-item Short-Form Survey (SF-12) scores was observed in both cohorts. Leg lengths were significantly shorter in the Exeter group, with a mean of -4 mm in this cohort compared with 0 mm in the C-Stem AMT group. One patient in the Exeter group had early evidence of radiological loosening. In total, 16 patients (15%) underwent further revision of the femoral component (seven in the C-Stem AMT group and nine in the Exeter group). No femoral components were revised for aseptic loosening. There were two cases of
The aim of this study was to determine both the incidence of, and the reoperation rate for, postoperative periprosthetic femoral fracture (POPFF) after total hip arthroplasty (THA) with either a collared cementless (CC) femoral component or a cemented polished taper-slip (PTS) femoral component. We performed a retrospective review of a consecutive series of 11,018 THAs over a ten-year period. All POPFFs were identified using regional radiograph archiving and electronic care systems.Aims
Methods
One-stage revision hip arthroplasty for periprosthetic joint infection (PJI) has several advantages; however, resection of the proximal femur might be necessary to achieve higher success rates. We investigated the risk factors for resection and re-revisions, and assessed complications and subsequent re-revisions. In this single-centre, case-control study, 57 patients who underwent one-stage revision arthroplasty for PJI of the hip and required resection of the proximal femur between 2009 and 2018 were identified. The control group consisted of 57 patients undergoing one-stage revision without bony resection. Logistic regression analysis was performed to identify any correlation with resection and the risk factors for re-revisions. Rates of all-causes re-revision, reinfection, and instability were compared between groups.Aims
Methods
We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation.Aims
Methods
Cementless knee arthroplasty has seen a recent resurgence in popularity due to conceptual advantages, including improved osseointegration providing biological fixation, increased surgical efficiency, and reduced systemic complications associated with cement impaction and wear from cement debris. Increasingly younger and higher demand patients are requiring knee arthroplasty, and as such, there is optimism cementless fixation may improve implant survivorship and functional outcomes. Compared to cemented implants, the National Joint Registry (NJR) currently reports higher revision rates in cementless total knee arthroplasty (TKA), but lower in unicompartmental knee arthroplasty (UKA). However, recent studies are beginning to show excellent outcomes with cementless implants, particularly with UKA which has shown superior performance to cemented varieties. Cementless TKA has yet to show long-term benefit, and currently performs equivalently to cemented in short- to medium-term cohort studies. However, with novel concepts including 3D-printed coatings, robotic-assisted surgery, radiostereometric analysis, and kinematic or functional knee alignment principles, it is hoped they may help improve the outcomes of cementless TKA in the long-term. In addition, though cementless implant costs remain higher due to novel implant coatings, it is speculated cost-effectiveness can be achieved through greater surgical efficiency and potential reduction in revision costs. There is paucity of level one data on long-term outcomes between fixation methods and the cost-effectiveness of modern cementless knee arthroplasty. This review explores recent literature on cementless knee arthroplasty, with regards to clinical outcomes, implant survivorship, complications, and cost-effectiveness; providing a concise update to assist clinicians on implant choice. Cite this article:
Total hip replacement in patients with advanced osteonecrosis of the femoral head is often complicated by early loosening of the femoral component. Recent evidence has suggested that abnormal bone extending into the proximal femur may be responsible for the early failure of the femoral component. We aimed to identify which patients were at high risk of early failure by evaluating gadolinium-enhanced MR images of histologically-confirmed osteonecrotic lesions beyond the femoral head. Although the MR signal intensity has been shown to correlate well with osteonecrosis in the femoral head, it was found to be relatively insensitive at identifying lesions below the head, with a sensitivity of only 51% and a predictive value of a negative result of only 48%. However, the specificity was 90%, with the predictive value of a positive MRI finding being 86%. Only those patients with osteonecrosis of the femoral head secondary to sickle-cell disease, who are known to be at high risk of early loosening, had changes in the MR signal in the greater trochanter and the femoral shaft. This observation suggests that changes in the MR signal beyond the femoral head may represent osteonecrotic lesions in areas essential for the fixation of the femoral component. Pre-operative identification of such lesions in the neck of the femur may be important when considering hip resurfacing for osteonecrosis of the femoral head, following which early loosening of the
Vancouver type B periprosthetic femoral fractures (PFF) are challenging complications after total hip arthroplasty (THA), and some treatment controversies remain. The objectives of this study were: to evaluate the short-to-mid-term clinical outcomes after treatment of Vancouver type B PFF and to compare postoperative outcome in subgroups according to classifications and treatments; to report the clinical outcomes after conservative treatment; and to identify risk factors for postoperative complications in Vancouver type B PFF. A total of 97 consecutive PPFs (49 males and 48 females) were included with a mean age of 66 years (standard deviation (SD) 14.9). Of these, 86 patients were treated with surgery and 11 were treated conservatively. All living patients had a minimum two-year follow-up. Patient demographics details, fracture healing, functional scores, and complications were assessed. Clinical outcomes between internal fixation and revisions in patients with or without a stable femoral component were compared. Conservatively treated PPFs were evaluated in terms of mortality and healing status. A logistic regression analysis was performed to identify risk factors for complications.Aims
Methods
We investigated patient characteristics and outcomes of Vancouver type B periprosthetic fractures treated with femoral component revision and/or osteosynthesis. The study utilized data from the Swedish Hip Arthroplasty Register (SHAR) and information from patient records. We included all primary total hip arthroplasties (THAs) performed in Sweden since 1979, and undergoing further surgery due to Vancouver type B periprosthetic femoral fracture between 2001 and 2011. The primary outcome measure was any further reoperation between 2001 and 2013. Cross-referencing with the National Patient Register was performed in two stages, in order to identify all surgical procedures not recorded on the SHAR.Aims
Patients and Methods
We describe six patients with aggressive granulomatous lesions around cementless total hip prostheses. Two patients previously had a cemented prosthesis in the same hip. The Lord prosthesis was used in five patients, the PCA in one. Both prostheses were made of chrome-cobalt alloy. Pain on weight-bearing occurred on average 3.2 years after the cementless arthroplasty, and at that time radiography revealed aggressive granulomatosis around the proximal femoral stem and the acetabular component in five of the patients; one had a large solitary granuloma in the proximal femur. Revision was performed on average 4.8 years after the cementless arthroplasty. At that time all granulomas had grown large in size; while waiting for revision operation, two
This study describes and compares the operative management and outcomes in a consecutive case series of patients with dislocated hemiarthroplasties of the hip, and compares outcomes with those of patients not sustaining a dislocation. Of 3326 consecutive patients treated with hemiarthroplasty for fractured neck of femur, 46 (1.4%) sustained dislocations. Of the 46 dislocations, there were 37 female patients (80.4%) and nine male patients (19.6%) with a mean age of 83.8 years (66 to 100). Operative intervention for each, and subsequent dislocations, were recorded. The following outcome measures were recorded: dislocation; mortality up to one-year post-injury; additional surgery; residential status; mobility; and pain score at one year.Aims
Patients and Methods
The goals of this study were to define the risk factors, characteristics,
and chronology of fractures in 5417 revision total hip arthroplasties
(THAs). From our hospital’s prospectively collected database we identified
all patients who had undergone a revision THA between 1969 and 2011
which involved the femoral stem. The patients’ medical records and
radiographs were examined and the relevant data extracted. Post-operative
periprosthetic fractures were classified using the Vancouver system.
A total of 5417 revision THAs were identified.Aims
Patients and Methods
We report on the outcome of the Exeter Contemporary flanged cemented
all-polyethylene acetabular component with a mean follow-up of 12
years (10 to 13.9). This study reviewed 203 hips in 194 patients.
129 hips in 122 patients are still A retrospective review was undertaken of a consecutive series
of 203 routine primary cemented total hip arthroplasties (THA) in
194 patients.Aims
Patients and Methods
We assessed the difference in hospital based and early clinical
outcomes between the direct anterior approach and the posterior
approach in patients who undergo total hip arthroplasty (THA). The outcome was assessed in 448 (203 males, 245 females) consecutive
patients undergoing unilateral primary THA after the implementation
of an ‘Enhanced Recovery’ pathway. In all, 265 patients (mean age:
71 years (49 to 89); 117 males and 148 females) had surgery using
the direct anterior approach (DAA) and 183 patients (mean age: 70
years (26 to 100); 86 males and 97 females) using a posterior approach.
The groups were compared for age, gender, American Society of Anesthesiologists
grade, body mass index, the side of the operation, pre-operative
Oxford Hip Score (OHS) and attendance at ‘Joint school’. Mean follow-up
was 18.1 months (one to 50).Aims
Patients and Methods
The goals of this study were to define the risk factors, nature,
chronology, and treatment strategies adopted for periprosthetic
femoral fractures in 32 644 primary total hip arthroplasties (THAs). There were 564 intra-operative fractures (1.7%); 529 during uncemented
stem placement (3.0%) and 35 during cemented stem placement (0.23%).
Intra-operative fractures were more common in females and patients
over 65 years (p <
0.001). The majority occurred during placement
of the femoral component (60%), and involved the calcar (69%). There
were 557 post-operative fractures (20-year probability: 3.5%; 95%
confidence interval (CI) 3.2 to 3.9); 335 fractures after placement
of an uncemented stem (20-year probability: 7.7%; 95% CI 6.2 to
9.1) and 222 after placement of a cemented stem (20-year probability:
2.1%; 95% CI 1.8 to 2.5). The probability of a post-operative fracture
within 30 days after an uncemented stem was ten times higher than
a cemented stem. The most common post-operative fracture type was
a Vancouver AG (32%; n = 135), with 67% occurring after
a fall. In all, 36% (n = 152) were treated with revision arthroplasty. Aim and Methods
Results
The primary purpose of this study of metal-on-metal
(MoM) hip resurfacing was to compare the effect of using a cementless
or cemented femoral component on the subsequent bone mineral density
(BMD) of the femoral neck. This was a single-centre, prospective, double-blinded control
trial which randomised 120 patients (105 men and 15 women) with
a mean age of 49.4 years (21 to 68) to receive either a cemented
or cementless femoral component. Follow-up was to two years. Outcome
measures included total and six-point region-of-interest BMD of
the femoral neck, radiological measurements of acetabular inclination,
neck-shaft and stem-shaft angles, and functional outcome scores
including the Harris hip score, the Western Ontario and McMaster
Universities Osteoarthritis Index and the University of California
at Los Angeles activity scale. In total, 17 patients were lost to follow-up leaving 103 patients
at two years. There were no revisions in the cementless group and
three revisions (5%) in the cemented group (two because of hip pain
and one for pseudotumour). The total BMD was significantly higher in the cementless group
at six months (p <
0.001) and one year (p = 0.01) than in the
cemented group, although there was a loss of statistical significance
in the difference at two years (p = 0.155). All patient outcomes improved significantly: there were no significant
differences between the two groups. The results show better preservation of femoral neck BMD with
a cementless femoral component after two years of follow-up. Further
investigation is needed to establish whether this translates into
improved survivorship. Cite this article:
In a prospective study, 93 unselected consecutive uncemented hip arthroplasties were performed in 80 patients using the titanium-coated RM acetabular component and the CLS femoral component. The mean age of the patients at operation was 52 years (28 to 81). None were lost to follow-up. In the 23 patients who had died (26 hips) only one acetabular component had been revised. In the 57 living patients (67 hips), 13 such revisions had been performed. Of the 14 revisions, seven were for osteolysis, five for loosening and two for infection. Survival analysis of this implant showed a total probability of survival of 83% (95% confidence interval 73 to 90), with all revisions as the endpoint, and a probability of 94% (95% confidence interval 87 to 98) with revision for aseptic loosening as the endpoint, indicating reliable long-term fixation of the titanium-coated RM acetabular component.
We present a retrospective series of 170 cemented titanium straight-stem femoral components combined with two types of femoral head: cobalt-chromium (CoCr) alloy (114 heads) and alumina ceramic (50 heads). Of the study group, 55 patients (55 stems) had died and six (six stems) were lost to follow-up. At a mean of 13.1 years (3 to 15.3) 26 stems had been revised for aseptic loosening. The mean follow-up time for stable stems was 15.1 years (12.1 to 16.6). Survival of the stem at 15 years was 75.4% (95% confidence interval (CI) 67.3 to 83.5) with aseptic failure (including radiological failure) as the end-point, irrespective of the nature of the head and the quality of the cement mantle. Survival of the stem at 15 years was 79.1% (95% CI 69.8 to 88.4) and 67.1% (95% CI 51.3 to 82.9) with the CoCr alloy and ceramic heads, respectively. The quality of the cement mantle was graded as a function of stem coverage: stems with complete tip coverage (type 1) had an 84.9% (95% CI 77.6 to 92.2) survival at 15 years, compared with those with a poor tip coverage (type 2) which had a survival of only 22.4% (95% CI 2.4 to 42.4). The poor quality of the cement mantle and the implantation of an alumina head substantially lowered the survival of the stem. In our opinion, further use of the cemented titanium alloy straight-stem femoral components used in our series is undesirable.